detection of muons at 150 gev/c and absolute energy calibration of a cms preshower prototype

23
Detection of muons at 150 GeV/c Detection of muons at 150 GeV/c and absolute energy calibration and absolute energy calibration of a CMS Preshower Prototype of a CMS Preshower Prototype Panagiotis Kokkas Univ. of Ioannina On behalf of the CMS ECAL Collaboration IPRD06 10 th Topical Seminar on Innovative Particle and Radiation Detectors 1-5 October 2006 Siena, Italy

Upload: nay

Post on 24-Jan-2016

28 views

Category:

Documents


0 download

DESCRIPTION

Detection of muons at 150 GeV/c and absolute energy calibration of a CMS Preshower Prototype. Panagiotis Kokkas Univ. of Ioannina On behalf of the CMS ECAL Collaboration IPRD06 10 th Topical Seminar on Innovative Particle and Radiation Detectors 1-5 October 2006 Siena, Italy. - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Detection of muons at 150 GeV/c and absolute energy calibration of a CMS Preshower Prototype

Detection of muons at 150 GeV/c and Detection of muons at 150 GeV/c and absolute energy calibration of a CMS absolute energy calibration of a CMS

Preshower PrototypePreshower Prototype

Panagiotis KokkasUniv. of Ioannina

On behalf of the CMS ECAL Collaboration

IPRD06 10th Topical Seminar on Innovative Particle and Radiation

Detectors1-5 October 2006 Siena, Italy

Page 2: Detection of muons at 150 GeV/c and absolute energy calibration of a CMS Preshower Prototype

22Siena IPRD06 Panagiotis Kokkas, Univ. of Ioannina

Talk OverviewTalk Overview

• The Preshower detector in CMS.• Preshower calibration (why?)• Test Beam Setup• Muon Data Analysis

• Pedestal Subtraction• Common Mode correction• Single strip signal extraction• Total signal extraction• GEANT 4 simulation• Energy calibration • Data over MC

• Electron Analysis• Summary

Page 3: Detection of muons at 150 GeV/c and absolute energy calibration of a CMS Preshower Prototype

33Siena IPRD06 Panagiotis Kokkas, Univ. of Ioannina

The Preshower Detector in CMSThe Preshower Detector in CMS

Page 4: Detection of muons at 150 GeV/c and absolute energy calibration of a CMS Preshower Prototype

44Siena IPRD06 Panagiotis Kokkas, Univ. of Ioannina

The Preshower Detector The Preshower Detector

Preshower fiducial coverage:1.653<η<2.6 or 10˚<θ<20˚

Barrel ECAL

End

cap E

CA

L

Pre

show

er

• The Preshower is a sandwich of two orthogonal layers of silicon strip sensors positioned behind two planes of lead absorbers.

• Provides the position of the foot of the shower in the endcap ECAL.• Main task is to identify the photons from π0→γ γ and distinguish

them from single photons (π0/γ separation).

Page 5: Detection of muons at 150 GeV/c and absolute energy calibration of a CMS Preshower Prototype

55Siena IPRD06 Panagiotis Kokkas, Univ. of Ioannina

The Preshower DetectorThe Preshower Detector

Single γ detection Double γ detection from π0 decay

Signal : Higgs decay to two gammas Background : one gamma faked by π0 decay

Page 6: Detection of muons at 150 GeV/c and absolute energy calibration of a CMS Preshower Prototype

66Siena IPRD06 Panagiotis Kokkas, Univ. of Ioannina

The Preshower DetectorThe Preshower Detector

32 strip

s

6.3 cm

6.3

cm

μmodule Ladder

layout of the ladders layout of the ladders

for one half of an ES planefor one half of an ES plane

Page 7: Detection of muons at 150 GeV/c and absolute energy calibration of a CMS Preshower Prototype

77Siena IPRD06 Panagiotis Kokkas, Univ. of Ioannina

Preshower calibration (Why?)Preshower calibration (Why?)

• Total Energy Deposition in ECAL: ETotal = ECrystals + EPreshower

• Preshower should provide EPreshower with an accuracy of ~5%.

• Signal produced by the sensors depends on:• Incident particle type and energy.• Angle of incidence (varies with η).• Sensor capacitance (vary by ±5% due to variation of

thickness 320±15 μm).• Charge collection efficiency of the sensors, which decreases

with neutron - proton fluence (10 y for High Luminosity ~1.6x1014cm-2 neutrons and ~0.4x1014cm-2 protons for large η).

• Intrinsic gain of pre-amplifier (PACE3).

Page 8: Detection of muons at 150 GeV/c and absolute energy calibration of a CMS Preshower Prototype

88Siena IPRD06 Panagiotis Kokkas, Univ. of Ioannina

Preshower calibrationPreshower calibration

• PACE 3: • 32 channel

Preamplifier shaper• 32x192 analogue

memory for event pipelining

PACE 3 modes of operation: • “Calibration mode”: average gain ~ 20 mV /mip• “Normal mode”: average gain ~ 3.2 mV /mip

(1 mip = 83.7 KeV = 3.7 fC for a 320 μm thick sensor)

PACE3 hybrid

Silicon Sensor

Page 9: Detection of muons at 150 GeV/c and absolute energy calibration of a CMS Preshower Prototype

99Siena IPRD06 Panagiotis Kokkas, Univ. of Ioannina

Test Beam SetupTest Beam Setup

CERN H4 experimental hall.• Muon data at 150 GeV/c• Electron data at 20, 35, 50, 80 and 120 GeV/c• Pion data at 20, 30, 50, 80 and 120 GeV/c

Page 10: Detection of muons at 150 GeV/c and absolute energy calibration of a CMS Preshower Prototype

1010Siena IPRD06 Panagiotis Kokkas, Univ. of Ioannina

One Plane of the Preshower PrototypeOne Plane of the Preshower Prototype

Page 11: Detection of muons at 150 GeV/c and absolute energy calibration of a CMS Preshower Prototype

1111Siena IPRD06 Panagiotis Kokkas, Univ. of Ioannina

Pedestal SubtractionPedestal Subtraction

• Use of dedicated pedestal runs.

• Pedestal evaluation per strip.

• For every channel raw values fitted to a Gaussian distribution.

• Extraction of central values and sigmas.

)9.2(6.6 mVADCcountsX =σ )7.3(5.8 mVADCcountsY =σ

Page 12: Detection of muons at 150 GeV/c and absolute energy calibration of a CMS Preshower Prototype

1212Siena IPRD06 Panagiotis Kokkas, Univ. of Ioannina

Common Mode correctionCommon Mode correction

• Small displacement of base line, differing on an event by event basis, due to external sources of noise.

• Correction on an event by event basis.• Method: Projection of the event on the y axis. Fitting of the

peak to a Gaussian distribution. Evaluation of common mode.

Page 13: Detection of muons at 150 GeV/c and absolute energy calibration of a CMS Preshower Prototype

1313Siena IPRD06 Panagiotis Kokkas, Univ. of Ioannina

Single strip signal extractionSingle strip signal extraction

• Pedestal region fitted to a Gaussian Distribution (sigma = noise N).

• Signal region fitted to a Landau distribution convoluted with a Gaussian one (MPE = signal S).

X sensor : S/N≈9 Y sensor : S/N≈7

Page 14: Detection of muons at 150 GeV/c and absolute energy calibration of a CMS Preshower Prototype

1414Siena IPRD06 Panagiotis Kokkas, Univ. of Ioannina

Total signal extractionTotal signal extraction

• Total signal (ADC counts) = The sum of all strips after cutting 5σ (pedestal σ) on every strip.

• Fitting data to a Landau distribution convoluted by a Gaussian:

• σL : Width parameter of Landau density• MPE : Most Probable Energy of Landau density• Integral : Normalization constant• σG : Width of convoluted Gaussian distribution

Page 15: Detection of muons at 150 GeV/c and absolute energy calibration of a CMS Preshower Prototype

1515Siena IPRD06 Panagiotis Kokkas, Univ. of Ioannina

GEANT 4 simulationGEANT 4 simulation

• For the analysis of the Test Beam Data a Monte Carlo was developed based on Geant 4.

• Fit MC data on a Landau distribution.

Simulation DATA (keV) (ADC counts)

Sensor X (MPE) 89.0±0.1 49.0±0.1Sensor Y (MPE) 89.3±0.1 43.7±0.1

Page 16: Detection of muons at 150 GeV/c and absolute energy calibration of a CMS Preshower Prototype

1616Siena IPRD06 Panagiotis Kokkas, Univ. of Ioannina

Data over MCData over MC

• Electronic noise added in MC in the form of a Gaussian distribution with widths:• X sensor: 5.4 ADC counts correspond to 9.8 keV• Y sensor: 5.9 ADC counts correspond to 12.1 keV

DataMC without noiseMC with noise

DataMC without noiseMC with noise

Page 17: Detection of muons at 150 GeV/c and absolute energy calibration of a CMS Preshower Prototype

1717Siena IPRD06 Panagiotis Kokkas, Univ. of Ioannina

Electron DetectionElectron Detection

• Analysis of electron Data.

• Pedestal Subtraction.• Common Mode correction.• Total signal (ADC counts) =

The sum of all strips after cutting 5σ (pedestal σ) on every strip.

• Use Absolute Energy Calibration (muons).

• Going from “calibration” to “normal” mode.

• Plot electron Energy Deposition in X and Y sensor.

Preliminary

Preliminary

Page 18: Detection of muons at 150 GeV/c and absolute energy calibration of a CMS Preshower Prototype

1818Siena IPRD06 Panagiotis Kokkas, Univ. of Ioannina

Electron DetectionElectron Detection

• Electrons Mean Energy Deposition per sensor and for the various momentum.

• Red points Monte Carlo.

Prelim

inar

y

Page 19: Detection of muons at 150 GeV/c and absolute energy calibration of a CMS Preshower Prototype

1919Siena IPRD06 Panagiotis Kokkas, Univ. of Ioannina

SummarySummary

• The Muon signal per silicon sensor is extracted, after pedestal subtraction and common mode correction.

• The absolute calibration of the Preshower sensors is demonstrated.

• MC in excellent agreement with data.• Electron – pion analysis in progress.

• Calibration before detector construction.• Calibration during construction of the ladders using cosmic

muons.

• Calibration during data taking.• An absolute calibration will be performed at regular intervals

in-situ in CMS using muons or even pions.

Page 20: Detection of muons at 150 GeV/c and absolute energy calibration of a CMS Preshower Prototype

2020Siena IPRD06 Panagiotis Kokkas, Univ. of Ioannina

Spare TransparenciesSpare Transparencies

Page 21: Detection of muons at 150 GeV/c and absolute energy calibration of a CMS Preshower Prototype

2121Siena IPRD06 Panagiotis Kokkas, Univ. of Ioannina

Preshower calibrationPreshower calibration

Most probable energy loss in silicon, scaled to the mean loss of a minimumionising particle, 388 eV/μm (1.66 MeV g-1 cm2) (PDG 2004)

Muons with momentum ~150 GeV/c are on the most probable energy loss plateau. Energy deposition ~91 KeV for a 320μm thick sensor.

Page 22: Detection of muons at 150 GeV/c and absolute energy calibration of a CMS Preshower Prototype

2222Siena IPRD06 Panagiotis Kokkas, Univ. of Ioannina

Calibration using cosmic muonsCalibration using cosmic muons

• Stack of 6 ladders.• Two scintillators for

triggering.• Trigger rate ~1Hz.• ~1 week data taking.

• One year operation for 536 ladders (4300 μmodules).

μ

Page 23: Detection of muons at 150 GeV/c and absolute energy calibration of a CMS Preshower Prototype

2323Siena IPRD06 Panagiotis Kokkas, Univ. of Ioannina

In-situ Absolute CalibrationIn-situ Absolute Calibration

• Using muon or jet events.• Use tracker/muon system to predict ES strips that have been

traversed.

ES plane #2, (Y-plane)

XX

YY

strip orientation

strip orientation

ZZ

ΔΧ

ΔΥ

vertexES plane #1, (X-plane)

Single track

<2 weeks needed for calibration at low luminosity