detectors and preamplifiers detector deliveries customer acceptance tests available detectors agata...

16
Detectors and Preamplifiers • Detector deliveries • Customer Acceptance Tests • Available detectors • AGATA triple detectors • Scanning program ter IKP University Cologne AGATA week 8-11 July 2008 U

Upload: alexandrina-butler

Post on 31-Dec-2015

248 views

Category:

Documents


3 download

TRANSCRIPT

Page 1: Detectors and Preamplifiers Detector deliveries Customer Acceptance Tests Available detectors AGATA triple detectors Scanning program P. Reiter IKP University

Detectors and Preamplifiers

• Detector deliveries

• Customer Acceptance Tests

• Available detectors

• AGATA triple detectors

• Scanning program

P. Reiter IKP University Cologne AGATA week 8-11 July 2008 Uppsala

Page 2: Detectors and Preamplifiers Detector deliveries Customer Acceptance Tests Available detectors AGATA triple detectors Scanning program P. Reiter IKP University

Detector Delivery Owner Status

A001* 11/2005 GANIL rejected – hole trapping

A001 06/2006 GANIL accepted

A002* 04/2006 TU Munich rejected – leakage currents

A002 06/2007 TU Munich accepted

A003* 11/2006 Liverpool rejected – leakage currents

A003 10/2007 Liverpool accepted

A004 09/2007 Ankara rejected – segments out of specification

A004* 07/2008 Ankara CAT at IKP, Cologne

A005 12/2007 Stockholm CAT at CEA, Saclay & IKP

B001* 11/2005 Padova rejected – segments out of specification

B001 06/2007 Padova rejected – leakage currents

B002 08/2006 GANIL accepted

B003 06/2007 Liverpool accepted

B004 12/2007 Ankara rejected – missing segment in capsule

B004* 06/2008 Ankara CAT at IKP, Cologne

B005 12/2007 Stockholm CAT at CEA, Saclay

C001 12/2005 Padova accepted

C002 07/2006 GANIL accepted

C003 03/2007 Liverpool accepted, broken, leakage currents

C004 09/2007 Ankara accepted

C005 12/2007 Stockholm CAT at CEA, Saclay

Page 3: Detectors and Preamplifiers Detector deliveries Customer Acceptance Tests Available detectors AGATA triple detectors Scanning program P. Reiter IKP University

Detector Delivery Owner Status

A001 06/2006 GANIL Accepted, available

A002 06/2007 TU Munich Accepted, available

A003 10/2007 Liverpool Accepted, available

B002 08/2006 GANIL Accepted, available

B003 06/2007 Liverpool Accepted, available

C001 12/2005 Padova Accepted, available

C002 07/2006 GANIL Accepted, available

C004 09/2007 Ankara accepted

Page 4: Detectors and Preamplifiers Detector deliveries Customer Acceptance Tests Available detectors AGATA triple detectors Scanning program P. Reiter IKP University

Recent detector tests and future deliveries

Turkish detectors

• B004 (repaired) delivered to IKP, Cologne, test started

• A004 (repaired) delivered July, 1st 2008

• shipment of A004, B004, C004 to Turkey autumn 2008

Swedish detectors

• A005, B005, C005 tests ongoing at CEA

- test results of A005 to be confirmed at IKP, Cologne

- test of B005 started

Liverpool detector C003: leakage current, out of warranty

Italian detectors

• A006 announced delivery July (week 29) 2008

• B006 announced delivery July (week 31) 2008

• B001 announced delivery July 2008

Page 5: Detectors and Preamplifiers Detector deliveries Customer Acceptance Tests Available detectors AGATA triple detectors Scanning program P. Reiter IKP University

Summary available detector

• 8 detectors accepted (A1,A2,A3 B2,B3 C1,C2,C4) • 1 detector to be repaired at Canberra (B1)• 5 detectors ongoing test (A4, A5, B4, B5, C5)• 1 detector broken (C3) ----------------------------------------------------------------------------------------• 3 (Sweden A005, B005, C005) ongoing CAT at Saclay• 2 (Turkey A004, B004) ongoing CAT at Cologne

• 1 (Italy C001) scanning at Liverpool until July

• 1 (Italy B001) not accepted after CAT, not repaired• 1 (UK C003) is broken after CAT, needs repair

• 3 (Turkey A004, B004, C004) are shipped to Turkey fall 2008-----------------------------------------------------------------------------------------• At the moment 6 detectors are available for AGATA triple clusters

A001, A002, A003 B002, B003 C002

Page 6: Detectors and Preamplifiers Detector deliveries Customer Acceptance Tests Available detectors AGATA triple detectors Scanning program P. Reiter IKP University

AGATA triple cluster detectors

First AGATA triple *A001 : B002 : C002 :

*Goes to LegnaroJuly 2008

First AGATA triple *A001 : B002 : C002 :

*Goes to LegnaroJuly 2008

Second AGATA tripleA003 : B003 : C004 : *

*Goes to Turkey in fall 2008,Replaced by C005

Second AGATA tripleA003 : B003 : C004 : *

*Goes to Turkey in fall 2008,Replaced by C005

Third AGATA tripleA002 : B005 or B001: * C001: **

*B005 after test or B001 after delivery & test** C001 after scanning

Third AGATA tripleA002 : B005 or B001: * C001: **

*B005 after test or B001 after delivery & test** C001 after scanning

Page 7: Detectors and Preamplifiers Detector deliveries Customer Acceptance Tests Available detectors AGATA triple detectors Scanning program P. Reiter IKP University

Scanning program

LiverpoolC001 * A005

* Goes to IKP,Assembly of secondTriple cluster by CTTJuly 2008

LiverpoolC001 * A005

* Goes to IKP,Assembly of secondTriple cluster by CTTJuly 2008

OrsayS002 *B005 :

* Commissioning, confirmLiverpool results

OrsayS002 *B005 :

* Commissioning, confirmLiverpool results

See talks by A. Boston C. UnsworthA. KorichiF. CrespiThursday, 14:00-15:40

Page 8: Detectors and Preamplifiers Detector deliveries Customer Acceptance Tests Available detectors AGATA triple detectors Scanning program P. Reiter IKP University

Future detector tests and triple assembly

Detectors for period fall and winter 2008/09• 13 detectors may be available A1, A2, A3, A5, A6 B1, B2, B3, B5, B6 C1, C2, C5 • A6, B1, B6 delivery July 2008

• redelivery of turkish detectors A4, B4, C4 ?

• January - June 2009 italian detectors C6, A7 ?

• detector repair of C3 ?

• summer 2008 order of german detectors A8, B7, C7 ?

Page 9: Detectors and Preamplifiers Detector deliveries Customer Acceptance Tests Available detectors AGATA triple detectors Scanning program P. Reiter IKP University

Talks by • A. Wiens, IKP Cologne

• A. Goergen, CEA Saclay

• D. Quirion, Canberra

• B. Bruyneel, IKP Cologne

• G. Pascovici, IKP Cologne

Page 10: Detectors and Preamplifiers Detector deliveries Customer Acceptance Tests Available detectors AGATA triple detectors Scanning program P. Reiter IKP University

1st Tripel Cluster Energies: Single vs Triple cryostat

0

0.2

0.4

0.6

0.8

1

1.2

1.4

1.6

Core A1A2A3A4A5A6 B1B2B3B4B5B6 C1C2C3C4C5C6 D1D2D3D4D5D6 E1E2E3E4E5E6 F1 F2 F3 F4 F5 F6

59,6 keV Single 59.6 keV Triple Specification low energy

0

0.5

1

1.5

2

2.5

3

Core A1A2A3A4A5A6 B1B2B3B4B5B6 C1C2C3C4C5C6 D1D2D3D4D5D6 E1E2E3E4E5E6 F1 F2 F3 F4 F5 F6

1332,5 keV Single 1332,5 keV Triple Specification high energy

A0

01

– F

WH

M [

keV

]B

00

2 –

FW

HM

[ke

V]

0

0.5

1

1.5

2

2.5

3

Core A1A2A3A4A5A6 B1B2B3B4B5B6 C1C2C3C4C5C6 D1D2D3D4D5D6 E1E2E3E4E5E6 F1 F2 F3 F4 F5 F6

1332,5 keV Single 1332,5 keV Triple Specification high energy

0

0.2

0.4

0.6

0.8

1

1.2

1.4

1.6

Core A1A2A3A4A5A6 B1B2B3B4B5B6 C1C2C3C4C5C6 D1D2D3D4D5D6 E1E2E3E4E5E6 F1 F2 F3 F4 F5 F6

59.6 keV Single 59.6 keV Triple Specification low energy

0

0.5

1

1.5

2

2.5

3

Core A1A2A3A4A5A6 B1B2B3B4B5B6 C1C2C3C4C5C6 D1D2D3D4D5D6 E1E2E3E4E5E6 F1 F2 F3 F4 F5 F6

1332,5 keV Single 1332,5 keV Triple Specification high energy

0

0.2

0.4

0.6

0.8

1

1.2

1.4

1.6

Core A1A2A3A4A5A6 B1B2B3B4B5B6 C1C2C3C4C5C6 D1D2D3D4D5D6 E1E2E3E4E5E6 F1 F2 F3 F4 F5 F6

59,6 keV Single 59,6 keV Triple Specification low energy

C0

02

– F

WH

M [

keV

]

Resolution 60keV line Resolution 1.33MeV line

Page 11: Detectors and Preamplifiers Detector deliveries Customer Acceptance Tests Available detectors AGATA triple detectors Scanning program P. Reiter IKP University

0

0,2

0,4

0,6

0,8

1

1,2

1,4

1,6

] F

WH

M[k

eV

59,6 keV Single 59,6 keV Triple Specification low energy

2nd Tripel Cluster Energies: Single vs Triple cryostatA

003

FW

HM

]ke

V[

0

0,2

0,4

0,6

0,8

1

1,2

1,4

1,6

] F

WH

M[k

eV

59,6 keV Single 59,6 keV Triple Specification low energy

0

0,2

0,4

0,6

0,8

1

1,2

1,4

1,6

C0

04

F

WH

M ]k

eV

[

59,6 keV Single 59,6 keV Triple Specification low energy

C00

4 F

WH

M [k

eV]

B

003

FW

HM

[keV

]

A00

3 F

WH

M [k

eV]

Resolution:

60keV line

Page 12: Detectors and Preamplifiers Detector deliveries Customer Acceptance Tests Available detectors AGATA triple detectors Scanning program P. Reiter IKP University

Tripel Cluster Energy resolution: Single vs Triple cryostat

Excellent energy resolution in two AGATA triple

clusters

Mean values segments @ 60 keV :

especially at low energies (electronic noise):

- glued detectors less microphonics

- HEKO AGATA preamp

- electronic assembly (grounding, cabling)

First AGATA tripleA001 : 1011 +/- 53 eVB002 : 1039 +/- 70 eVC002 : 965 +/- 63 eV

First AGATA tripleA001 : 1011 +/- 53 eVB002 : 1039 +/- 70 eVC002 : 965 +/- 63 eV

Second AGATA tripleA003 : 1052 +/- 65 eVB003 : 995 +/- 73 eV C004 : 1148 +/- 104 eV *

Second AGATA tripleA003 : 1052 +/- 65 eVB003 : 995 +/- 73 eV C004 : 1148 +/- 104 eV *

* high frequency oscillation to be removed

Page 13: Detectors and Preamplifiers Detector deliveries Customer Acceptance Tests Available detectors AGATA triple detectors Scanning program P. Reiter IKP University

For any 1406keV single event in the detector:

-3.0

-2.5

-2.0

-1.5

-1.0

-0.5

0.0

0 6 12 18 24 30 36

Hit segment number

Ba

selin

e S

hift

in A

1 (

keV

)Cross talk measurements

A2..A6

B1 F1

Co

re to

se

g X

talk

= u

nd

erst

ood

!

Segment labeling:

Sectors: A...F

Rings: 1...6

X

Bart Bruyneel, Andreas Wiens

Page 14: Detectors and Preamplifiers Detector deliveries Customer Acceptance Tests Available detectors AGATA triple detectors Scanning program P. Reiter IKP University

B002 in Triple Cryostat @ 5000 V

-0.25%

-0.20%

-0.15%

-0.10%

-0.05%

0.00%

0 200 400 600 800 1000 1200

segment "i" + 36x segment "j"

Xta

lk a

mpl

itude

s fro

m s

eg "

i" to

seg

"j"

[%]

Measured Cross talk

Core to Seg Xtalk (theory)

Smallest capacities in ring 2

A2

B2 C2 D2 E2 F2

Page 15: Detectors and Preamplifiers Detector deliveries Customer Acceptance Tests Available detectors AGATA triple detectors Scanning program P. Reiter IKP University

Cross talk in first AGATA Triple Cryostat

-0.25%

-0.20%

-0.15%

-0.10%

-0.05%

0.00%

0.05%

0 200 400 600 800 1000 1200 1400 1600

-0.25%

-0.20%

-0.15%

-0.10%

-0.05%

0.00%

0.05%

0 200 400 600 800 1000 1200 1400 1600

-0.25%

-0.20%

-0.15%

-0.10%

-0.05%

0.00%

0.05%

0 200 400 600 800 1000 1200 1400 1600

A001 B002 C002

Co

re B

Co

re C

Co

re A

Co

re C

Co

re A

Co

re B

Sector A Sector F

Sector A Sector F Sector A Sector F

A00

1B

002

C00

2

Page 16: Detectors and Preamplifiers Detector deliveries Customer Acceptance Tests Available detectors AGATA triple detectors Scanning program P. Reiter IKP University

AGATA Triple Clusters: Summary & outlookTwo AGATA Triple Cluster Detectors operational:- first & second triple: cryostat, mechanics, vacuum, thermal properties of preamps o.k.- seg-preamps, core preamps & pulser with high dynamic range operational- very good energy resolution at low energies - no cross talk between detectors observed in 1st ATC between detectors- commissioning of 2nd ATC ongoing

Next steps:- transport of 1st ATC to Legnaro, first tests with digital electronics at Legnaro July 2008-- finish commissioning of 2nd ATC, - cross talk tests-- assembly of 3rd ATC constrained by -available detectors

ATC1

ATC2

ATC3