determination of elemental impurities – challenges of a screening method
TRANSCRIPT
DETERMINATION OF ELEMENTAL IMPURITIES
– CHALLENGES OF A SCREENING METHOD
THOMAS DEVADDERTHOMAS DEVADDERSGS LIFE SCIENCE SERVICES TEAM LEADER QC AAS / ICP / Particles / TGALaboratory Services
� Challenges
� Strategy
� Potential interferences
� Sample preparation / determination by ICP-MS
� Validation of a screening method
AGENDA
2SGS LIFE SCIENCE SERVICES – CPHI/ICSE Madrid October 2015
� Validation of a screening method
� Screenings on excipients, APIs and drug products
� Where are the boundaries in the applicability?
CHALLENGES OF A SCREENING METHOD
� Method must provide
• Valid information about APIs, Excipients and FinishedProducts
• Applicability for a broad variability of sample materials
• Acceptance criteria from EP 5.20 / USP <233>
• A validated basis
3SGS LIFE SCIENCE SERVICES – CPHI/ICSE Madrid October 2015
• A validated basis
� Challenges
• Interferences from different sample materials
STRATEGY
• Worst case limits by EP/USP/ICH Oral/Parenteral/Inhalation (Late 2013)
• Daily dose of 10g/day
• Quantitative procedure
Basis
• Worst case matrix (Omega-3 fish oil)
• Closed vessle microwave digestion (nitric acid)Method
Development
4SGS LIFE SCIENCE SERVICES – CPHI/ICSE Madrid October 2015
• Closed vessle microwave digestion (nitric acid)
• Determination by ICP-MSDevelopment
• According to USP<233>/EP2.4.20 (Omega-3 fish oil)
• Simulation of potential Interferences from different sample materials (K, Na, Ca, Mg, Cl)
MethodValidation
• Fullfilled acceptance critera (USP<233>/EP2.4.20)
• Method verification on new samples by spikingexperiment
Routine Testing
INTERFERENCES
� Physical Interferences
• Viscosity
• Density
• Matrix
• Sediments on Cones
� Dilution, Internal Standard, Sample Preparation
5SGS LIFE SCIENCE SERVICES – CPHI/ICSE Madrid October 2015
� Dilution, Internal Standard, Sample Preparation
� Chemical Interferences
• Absorption effects during sample introduction
• Nebulization effects
• Stability of solution
• Contaminations
• Carry over
� Method-Optimization, Stabilization
� Spectral Interferences
• Oxid Formation
• Double charged Ions
• Polyatomic combinations
� System-Optimization
INTERFERENCES
6SGS LIFE SCIENCE SERVICES – CPHI/ICSE Madrid October 2015
� Isobare Interferences
• Elements share isotopic masses (Resolution 0.7 amu)
� Choice of an appropriate isotope, Corrective equation
� Polyatomic Interferences
• Combinations of Ar40, O16, Cl35, N14, H1
� Dynamic Reaction Cell (NH3, CH4, H2, O2, He)
It‘s good to know, that…
� Most interferences show up in a mass range below 80 amu (atomic mass units) because of polyatomic
INTERFERENCES SUMMARY
IT IS GOOD TO KNOW THAT…
7SGS LIFE SCIENCE SERVICES – CPHI/ICSE Madrid October 2015
80 amu (atomic mass units) because of polyatomic compounds from Ar40, O16, Cl35, N14, H1 and their combination.
� Many interferences end at a concentration of 20 µg/l
INTERFERENCES
� To avoid Physical Interferences
→ Add 20 µg/l Indium / Internal Standard in all solutions
� To avoid Chemical Interferences
� Signal Os Standard / Os Standard after CVMD ≈ 1:10
→ Adding of a complexing agent
� Hg Stabilization
8SGS LIFE SCIENCE SERVICES – CPHI/ICSE Madrid October 2015
� Hg Stabilization
→ Add 500 µg/l Gold for (preserves Hg <10µg/l)
� Carry over effect
→ 60 sec rinsing step between samples
� To avoid Polyatomic Interferences
� Ar40Cl35 interferes As75
� O16Cl35 interferes V51
→ DRC using 0.6 ml/min O2
→ Polyatomic Combination of As75 to AsO91 and V51 to VO67
600
700
800
900
1000
[cps] As75 in Water
50500
51000
51500
52000
[cps] As75 in HCl 1.5%
500
600
700
800
900
[cps] AsO91 in HCl 1.5%(DRC O2 0.6 ml/min)
Ar40Cl 35 / As75 INTERFERENCE
9SGS LIFE SCIENCE SERVICES – CPHI/ICSE Madrid October 2015
0
100
200
300
400
500
0 0.5 1
[µg/l]
48000
48500
49000
49500
50000
0 0.5 1
[µg/l]
Ar40Cl35 Background
0
100
200
300
400
500
0 0.5 1
[µg/l]
MICROWAVE DIGESTION
� 300 mg sample material digested (6 ml HNO3/50 ml)
� Max. 80 bar / max. 280°C
Sample material +6 ml HNO3 after digestion transferred filled up
10SGS LIFE SCIENCE SERVICES – CPHI/ICSE Madrid October 2015
Sample material +6 ml HNO3 after digestion transferred filled up
# ElementLimit
[µg/g]
Range
[µg/l]
Range
[µg/g]
1 As 0.15 2.25 0.375
2 Cd 0.15 2.25 0.375
3 Hg 0.12 1.8 0.3
4 Pb 0.5 7.5 1.25
5 V 0.12 1.8 0.3
6 Cr 0.29 4.35 0.725
7 Ni 0.15 2.25 0.375
8 Cu 1.3 19.5 3.25
9 Mo 0.76 11.4 1.9
10 Ru 0.14 2.1 0.35
CALIBRATION
11SGS LIFE SCIENCE SERVICES – CPHI/ICSE Madrid October 2015
11 Rh 0.14 2.1 0.35
12 Pd 0.1 1.5 0.25
13 Ir 0.14 2.1 0.35
14 Pt 0.14 2.1 0.35
15 Os 0.14 2.1 0.35
16 Fe 130 1950 325
17 Zn 130 1950 325
18 Mn 25 375 62.5
19 Co 0.29 4.35 0.725
20 Se 8.5 127.5 21.25
21 Ag 0.69 10.35 1.725
22 Sb 2.2 33 5.5
23 Tl 0.8 12 2
24 Ba 34 510 85
25 Li 2.5 37.5 6.25
26 Sn 6.4 96 16
� Calibration up to 250% of target limit
� Correlation coefficient r ≥ 0,998
� Recovery QC Standard 80-120%
� Sample: 300 mg in 50 ml final solution via microwave digestion
USP <233> EP 2.4.20Quantitative Procedure
ProcedureAcceptanceCriteria
ProcedureAcceptanceCriteria
Specificity Method must show reliable measurements for target elements in the matrix and components including other target elements
- Method must show reliable measurements for target elements in the matrix
Demonstrating compliance with Acceptance Criteria from Accuracy
Linearity, Range - Demonstrating by meeting the Accuracy requirement
- Demonstrating compliance with Acceptance Criteria from Recovery
Accuracy Standard solutions within a range of 50% – 150% of the specification limit in triplicate
Mean recovery of 3 individual replicates must be within
Spiking Experiment in 3 Levels within a range of 50% – 150%
Mean recovery of 3 individual replicates must be within 70% - 150% for each level
VALIDATION REQUIREMENTS
12SGS LIFE SCIENCE SERVICES – CPHI/ICSE Madrid October 2015
specification limit in triplicate and Spiking Experiment within a range of 50% – 150% of the specification limit in triplicate (+ e.g. spiking Experiment at 10% Specification limit for LOQ)
be within 70% - 150% for each level
– 150% of the specification limit in triplicate (+ e.g. spiking Experiment at
10% Specification limit for
LOQ)
70% - 150% for each level
Repeatability 6 spiking experiments at specification limit
RSD <= 20% 6 spiking experiments at specification limit or procedure of Accuracy
RSD <= 20%
Ruggedness Experiments of Repeatability on a different day, or with a different instrument or by different analyst. Minimum 1 of these 3 choices.
RSD <= 25% Experiments of Repeatability on a different day, or with a different instrument or by different analyst. Minimum 1 of these 3 choices.
RSD <= 25%
Quantification Limit (LOQ)
- Demonstrating by meeting the Accuracy requirement
Determine the lowest concentration meeting the Acceptance Criteria from Accuracy
LOQ < Specification limit
# Element Limit Selectivity Linearity Method precision Intermediate precision
[µg/g] Isotope Ratio: 0,8-1,2 Criteria: r≥0,998 RSD n=6 (100%): ≤20% RSD n=12 (100%): ≤25%
1 As 0.15 Reaction cell 0.99976 10.0% 7.1%
2 Cd 0.15 complies 0.99910 2.4% 2.2%
3 Hg 0.12 complies 0.99972 15.3% 16.6%
4 Pb 0.5 complies 0.99983 11.3% 12.6%
5 V 0.12 Reaction cell 0.99996 7.3% 5.5%
6 Cr 0.29 Reaction cell 0.99983 1.9% 2.6%
7 Ni 0.15 complies 0.99994 8.6% 6.5%
8 Cu 1.3 complies 0.99964 2.4% 1.7%
9 Mo 0.76 complies 0.99950 4.8% 3.4%
10 Ru 0.14 complies 0.99972 3.2% 3.5%
VALIDATION RESULTS 1/2
13SGS LIFE SCIENCE SERVICES – CPHI/ICSE Madrid October 2015
10 Ru 0.14 complies 0.99972 3.2% 3.5%
11 Rh 0.14 Mono Isotope 0.99993 2.1% 3.3%
12 Pd 0.1 complies 0.99976 2.3% 3.0%
13 Ir 0.14 complies 0.99994 9.8% 10.9%
14 Pt 0.14 complies 0.99807 12.3% 13.5%
15 Os 0.14 complies 0.99993 10.2% 9.5%
16 Fe 130 complies 1.00000 1.5% 1.3%
17 Zn 130 complies 0.99976 7.6% 5.7%
18 Mn 25 Mono Isotope 0.99986 1.0% 1.0%
19 Co 0.29 Mono Isotope 0.99991 1.3% 1.5%
20 Se 8.5 complies 0.99837 11.6% 9.4%
21 Ag 0.69 complies 0.99839 1.9% 12.5%
22 Sb 2.2 complies 0.99998 8.9% 8.6%
23 Tl 0.8 complies 0.99996 12.3% 11.1%
24 Ba 34 complies 0.99993 3.6% 2.5%
25 Li 2.5 complies 0.99996 2.9% 3.3%
26 Sn 6.4 complies 1.00000 3.1% 3.2%
# Element Limit Accuracy / Mean Recovery 70-150% for each spiking level LOQ [Level]
[µg/g] 10% Level 20% Level 50% Level 100% Level 200% Level Target: ≤50%
1 As 0.15 104.2% 98.8% 101.1% 98.5% 100.5% 10%
2 Cd 0.15 93.1% 97.1% 97.4% 99.4% 100.3% 10%
3 Hg 0.12 69.7% 78.5% 74.6% 84.3% 88.5% 10%
4 Pb 0.5 89.7% 95.7% 92.1% 76.8% 82.8% 10%
5 V 0.12 92.5% 93.3% 92.3% 93.7% 95.3% 10%
6 Cr 0.29 98.3% 92.7% 104.3% 102.6% 100.8% 10%
7 Ni 0.15 66.4% 110.3% 93.0% 95.0% 100.8% 20%
8 Cu 1.3 100.3% (RSD:32.7%) 92.5% 102.8% 102.4% 106.3% 20%
9 Mo 0.76 118.1% 113.5% 120.1% 111.2% 104.9% 10%
10 Ru 0.14 109.5% 110.7% 117.5% 110.0% 109.3% 10%
VALIDATION RESULTS 2/2
14SGS LIFE SCIENCE SERVICES – CPHI/ICSE Madrid October 2015
10 Ru 0.14 109.5% 110.7% 117.5% 110.0% 109.3% 10%
11 Rh 0.14 106.4% 106.8% 114.0% 107.5% 106.7% 10%
12 Pd 0.1 75.0% 88.0% 103.7% 102.6% 103.7% 10%
13 Ir 0.14 76.2% 78.5% 74.6% 84.3% 88.5% 10%
14 Pt 0.14 79.8% 72.3% 78.5% 79.4% 84.4% 10%
15 Os 0.14 84.5% 82.5% 74.9% 84.4% 80.0% 10%
16 Fe 130 97.2% 98.6% 102.0% 99.8% 100.4% 10%
17 Zn 130 91.4% 93.2% 94.6% 98.0% 103.0% 10%
18 Mn 25 107.6% 108.8% 113.8% 109.9% 110.8% 10%
19 Co 0.29 106.2% 107.6% 114.2% 110.4% 111.1% 10%
20 Se 8.5 92.5% 93.1% 98.8% 97.3% 104.5% 10%
21 Ag 0.69 72.7% 86.2% 101.4% 106.0% 106.7% 10%
22 Sb 2.2 92.4% 90.0% 90.5% 92.3% 79.6% 10%
23 Tl 0.8 74.8% 76.3% 73.8% 81.7% 86.2% 10%
24 Ba 34 93.7% 95.7% 93.6% 98.8% 101.6% 10%
25 Li 2.5 106.8% 108.8% 113.9% 115.6% 116.7% 10%
26 Sn 6.4 95.3% 92.9% 97.4% 95.0% 81.4% 10%
ROUTINE SCREENING
Organic
Samples
APIs
Solvents, Polyol
Artificial flavours
Cellulose
Fatty oils
Clear sample solution
Method verificationcomplies
Salts
Clear Sample solution containing
high saltconcentration
PhysicalInterferences /
False negative
15SGS LIFE SCIENCE SERVICES – CPHI/ICSE Madrid October 2015
Inorganic
Samples
high saltconcentration
False negative results
Pigments, InkSiO2 , Talc, TiO2
Glue Potential forundisolved
components
Filtration,
Potential forphysical
Interferences/
False negative results
Finished
Products
Capsules
Tabletes
Protein
Solutions
Clear sample solution
Method verificationcomplies
POTENTIAL IMPACT ON RESULTS
Salts
PotassiumMagnesium
Potential forPhysical Interferences
SodiumCu63 / Na23Ar40
False positive results
16SGS LIFE SCIENCE SERVICES – CPHI/ICSE Madrid October 2015
CalciumNi60 / Ca44O16
False positive results
Pigments, InkSilicon
componentsGlue
CapsulesTabletes
Ferric Oxide Major Physical Interferences
SiO2, Talc, TiO2 Analyte loss within filtration
USP EP ICP-MS
Excipient ProcedureTarget Elements
ProcedureTarget Elements
MethodImprovement
Ferric Oxide AAS
ColorimetricLimit Test
Hg, Pb
As
- - Reduced sample concentration,Improved Digestion
Talc Flame AAS Al, Ca, Fe, Pb(Impurity)
Flame AAS Al, Ca, Fe, Pb(Impurity)
Digestion with nitric acid + hydrofluoric acid
WAYS TO CONTROL CRITICAL EXCIPIENTS
17SGS LIFE SCIENCE SERVICES – CPHI/ICSE Madrid October 2015
Mg (Assay) Mg (Assay)
TitaniumDioxide
ColorimetricLimit Test
As <211> ColorimetricLimit Test
Sb, As, Ba, Fe
Heavy metals (2.4.8)
Digestion with nitric acid + hydrofluoric acid
Silicon Dioxide ColorimetricLimit Test
As <211>
Heavy metals<231>
- - Digestion with nitric acid + hydrofluoric acid
Salts Reduced sample concentration, Improved Interference control
SGS SOLUTIONS 1/2
� Latest equipment and techniques
� ICP-MS
� ICP-OES
� Flame / Graphite Furnace-AAS
� FIMS (Hydrid System) and combination to AAS
18SGS LIFE SCIENCE SERVICES – CPHI/ICSE Madrid October 2015
� Large geographic coverage and sites with ICP-MS
� Europe Berlin (Germany), Clichy (France)
� Asia Chennai (India), Taipei (China)
� North America Fairfield (USA), Lincolnshire (USA)
� Wide experiences in:
� Method Development and Validation
� Verification of pharmacopeia methods
� ICH Q3D, USP <232> / <233> and EP 5.20 / 2.4.20
SGS SOLUTIONS 2/2
19SGS LIFE SCIENCE SERVICES – CPHI/ICSE Madrid October 2015
� Extractable / Leachable Studies
� Determination of Silicon Oil traces
Life Science Services Thomas Devadder
Team Leader QC AAS/ICP/Particles/TGA
Laboratory Services
THANK YOU FOR YOUR ATTENTION
+ 41 22 739 9548
+ 1 866 SGS 5003
20SGS LIFE SCIENCE SERVICES – CPHI/ICSE Madrid October 2015
SGS Institut Fresenius GmbH t: + 49 30 34607 659Tegeler Weg 33, f: + 49 30 34607 600D-10589 Berlin Germany E-mail : [email protected]
Web : www.sgs.com/lifescience
+ 1 866 SGS 5003
+ 65 637 90 111
+ 33 1 41 24 87 87
+ 1 877 677 2667