developing real world tests and assays using gold particles arista biologicals inc
TRANSCRIPT
Developing Real World Tests and Assays Using Gold Particles
Arista Biologicals Inc.
Developing Real World Tests and Assays
1. Biological Reagent Concerns
2. Assay Characteristics
3. Conjugate Optimization
4. Membrane Optimization
5. Conjugate Pads
6. Sample Pads
7. Problem Solving
8. Clinical and Stability Testing
Arista Biologicals Inc.
Biological Reagent Concerns
Monoclonal antibodies Most reproducible - True since no change in affinity or specificity
theoretically Most sensitive and specific - Chosen for best optimal behavior Easiest to conjugate - Since you are dealing with a homogeneous
protein
Polyclonal antibodies Fastest to produce - Good response in 45-90 days Generally least expensive - Unless affinity purified Least reproducible - Variation from animal to animal and over time in
the same animal Most cross-reactivity - Usually not desirable except for looking for
variants etc.
Arista Biologicals Inc.
Biological Reagent Concerns Cont…
Antigens Hapten - Carrier conjugates, small molecule assays Recombinant antigens - Very reproducible but may not cover variants and do not
always fold as native protein Native antigens - Some variation from lot to lot and usually a mixture of proteins.
In many cases less sensitive assay is produced. Many times are denatured Synthetic peptides - Inexpensive but less like native forms. Usually need to be
conjugated to a carrier
Matching Compatibilities Cross-reactivity - Epitope overlapping, not enough specificity in the reagents,
too much specificity in the reagents
Interspecies interactions - Usually observed in assay systems containing
antibodies from different species
Prozoning - Usually observed with non affinity purified antibodies or assays
requiring extremely high levels of antigen detection
Arista Biologicals Inc.
Biological Reagent Concerns Cont…
Purity Methods - Protein A and G, ion exchange, affinity
purifications, and other chromatographic methods Effects on producing conjugate - Irreproducible
sensitivities, aggregation of some conjugates etc. Effects on specificity - Background reactions sometimes
immunological in nature and sometimes protein-protein interactions are observed.
Effects on sensitivity - Less sensitive due to less binding sites available
Testing - Generally electophoresis and chromatographic methods acceptable. HPLC especially useful looking for aggregation of the proteins
Arista Biologicals Inc.
Biological Reagent Concerns Cont…
Buffers Compatibility with conjugation method - Several
types of buffers promote aggregation of the
conjugate during formation. Many conjugate to the
probe directly.
Compatibility with membrane coating - Many
buffers contain salts or other materials which
compete for binding sites with the membrane
Arista Biologicals Inc.
Biological Reagent Concerns Cont…
Concentrations Solubility Issues - Typically problems with
recombinant, native, and synthetic peptides because
of hydrophobic characteristics
Typical Ranges 1-4 mg/ml for antibodies
1 mg/ml for protein antigens
4mg/ml do to ug/ml for Haptens
Arista Biologicals Inc.
Assay Characteristics
Format - Flow through or lateral flow (dipstick, device, or midstream). May require
specific components including pads, membranes, pretreatments
Serum, Urine, Whole Blood, Feces, Dirt, Air, etc. - Test may be
desired to work with several for the same assay.
Dilution of sample - Creates another step but allows standardization of the
sample
Sensitivity - May require large amounts of reagents or slower assay development
times
Time of assay - If your requirements are for a faster assay, larger
amounts of reagents may be required
Cost - Limitations on the amounts of reagents, type of components, etc.
Arista Biologicals Inc.
Conjugate Optimization
Type of Conjugate Latex or colloidal gold - Color desired, type of assay,
sensitivity, amounts of reagents required, etc. Covalent or passive absorption
Size of Particle Type of particle Sensitivity desired - Larger particles generally give stronger
signals but more nonspecific interaction problems. Conversely, smaller particles may be used to cure a non-specific binding problem
Cost of protein to be conjugated
Arista Biologicals Inc.
Conjugate Optimization Cont…
Sensitivity Desired Coating levels - Passive absorption vs. covalent
attachment issues
Method to Immobilize Conjugate Dispensing - Can use high concentration with low
volumes. Higher conjugation ratios may also be
required
Soaking - May require vacuum drying or lyophilization
depending on how critical conjugate quantity per test is
Arista Biologicals Inc
Membrane Optimization Quality and Cost
Quality and cost vary from manufacturer to manufacturer and from material to material.
Concentration Ranges Typically in sandwich assays 1-4 mg/ml, protein antigens
usually range from 0.1-1.0 mg/ml, and Hapten assays may require 1 microgram to 4 mg/ml.
Sensitivity Desired A higher flow rate membrane will generally develop faster
but require more reagents to obtain the same sensitivity. Smaller pore size membranes also bind more capture protein
.Arista Biologicals Inc.
Membrane Optimization Cont…
Flow Rate Generally the higher the flow rate required the larger the
pore size needed. Larger pore sizes generally have less chance of aggregation of the reagents causing flow problems
Blocking or Not Blocking Blocking takes care of any nonspecific binding and can be
used to change flow rates and other characteristics of the native membrane. It is also time consuming in many cases. Several blocking agents may be required to be tested including BSA, PEG, PVA, PVP, etc.
Arista Biologicals Inc.
Conjugate Pad Optimization
Considerations Polyester, rayon, glass fiber materials most widely used Most of these materials are hydrophobic naturally and need to
be treated during or prior to conjugate immobilization Flow rates may vary by the material or the treatment Treatment usually includes addition of polymer and surfactant Sugars may or may not be required in the conjugate upon
immobilization depending on the method Conjugate can be dispensed up to approximately 3 to 4
microliters per cm. Dispensing is much more accurate than soaking methods
Arista Biologicals Inc.
Conjugate Pad Optimization Cont…
Considerations Cont… Treatments can be varied to make the product more or
less sensitive to humidity and temperature conditions
Conjugate can be formulated to vary the rate of release
of the conjugate during actual assay. This is sometimes
key in determining sensitivity. It also can save on reagent
Blocking reagents may be included in this pad if
membrane is not blocked
Assay buffer may be included in the pad if compatible
with conjugate stability and resolubilization.
Arista Biologicals Inc.
Sample Pad Optimization
Is a Sample Pad Required? If the conjugate is stable and resoluble in
the assay buffer a one bottom pad system
can be utilized. The material must also have
acceptable flow and liquid capacity
characteristics
Arista Biologicals Inc.
Sample Pad Optimization Cont…
Material Requirements Flow characteristics - Does it flow at the proper rate in order not to
inhibit the rate of assay development
Volume capacity - Does it hold enough buffer for the components to
be in high enough amounts to have the proper effect for the entire
range of clinical samples
Nonspecific binding of antigen - Does the analyte interact with the
material nonspecifically
Ease of handling - Can the material be processed with out
damaging it during the normal manufacturing conditions
Arista Biologicals Inc.
Sample Pad Optimization Cont… Buffer Components
pH buffering capacity - the pad must control the pH of the assay during use.
Ionic strength control - the pad must contain proper amounts of salts to control the ionic strength of the assay
The pad material - must be blocked of any nonspecific binding of the analyte if the material does not already have that characteristic
The pad material- must contain any surfactants required for specificity requirements of the assay. It must reduce these type of nonspecific interactions
The pad material - must contain any polymers or protein needed to block either immunological reactions from occurring nonspecifically or protein-protein interactions not desired
Note: If membrane blocking reagent is required the sample pad normally will contain this reagent and/or the conjugate pad
Arista Biologicals Inc.
Typical Problem Observations and Solving
Nonspecific Binding Heterophile Reactions
absorption addition of IgG addition of surfactants antibody fragmentation pH changes addition of “hard” buffers
Arista Biologicals Inc.
Typical Problem Observations and Solving Cont…
Nonspecific Binding Cont… General Nonspecific Reactions
blocking changing blocking agent on membrane change blocking agent on conjugate surfactant addition smaller particle production pH changes buffer changes decreasing reagents or strength change in conjugation procedure (method)
Arista Biologicals Inc.
Typical Problem Observations and Solving Cont…
Lack of Sensitivity Increase capture material Increase conjugate Increase strength of conjugate Increase particle size Decrease or change blocking agent Decrease flow rate
decrease pore size increase blocking agent
Decrease rate of conjugate release Use higher affinity antibodies or stronger antigens Increase activity of present reagents through repurification Add activity enhancer - such as PEG Decrease or change surfactants Check for analyte absorption onto assay component
Arista Biologicals Inc.
Typical Problem Observations and Solving Cont…
Large Volumes of Reagent Required Increase particle size Decrease flow rate Increase activity or purity of reagent Decrease rate of conjugate release Dispense conjugate
Lack of Specificity (Immunological) Repurify (affinity chromatography, absorption) Scavenger antibodies and antigens Find new reagents
Arista Biologicals Inc.
Typical Problem Observations and Solving Cont…
Migration Problems Remake conjugate
smaller particle different particle change protocol for production (chemistry changes) change or increase blocking agent add surfactants block membrane directly look for leaching of reagents from one material to
another
Arista Biologicals Inc.
Typical Problem Observations and Solving Cont…
Stability Problems Change conjugate pad material Change preparation of conjugate pad
increase or add surfactants increase or add polymers change pH or buffer add or increase sugars to conjugate change sugars in conjugate
Look for moisture problems Change membrane material Add preservatives to membrane block Add preservatives to capture protein prior to dispensing Look for leaching of reagents from one material to another Look for degradation of materials on sample pad
Arista Biologicals Inc.
Clinical and Stability Testing Clinical Testing
Highly probable problem samples should be tested first. Suspected crossreactant containing samples, heterophile type samples, low sensitivity samples, variants, lipemic, hemolyzed, etc. samples should be tested early before large complete final clinical tests are performed. Reformulation should be performed as required
If new formulations or materials are used than before especially new blocking formulations, conjugate pad formulations, membranes, etc. then some accelerated stability studies should be performed. Both biological stability and changes in physical characteristics such as flow should be observed closely
During clinical testing several technicians should be used especially when performing in house testing looking for slight variations in users techniques
Enough clinical testing should be performed using challenging samples which would be expected to be encountered in the field
Arista Biologicals Inc.
Clinical Testing and Stability Cont…
Stability Proper stability testing should be carried out. If real time stability cannot be
used the product should be tested at several different temperatures. At least
three different temperatures is recommended. These assays are
multicomponent biological systems and cannot be treated a simple chemical
reactions. However, we have observed using identical formulations and
procedures for key components as a rule do follow a pattern
We have also found that fresh product sometimes does not perform the
same as slightly aged product. Several drying or curing steps seem to take a
few days to complete. In some cases performance increases. This is
especially important when producing assays which require high precision
optimization for every batch such as DOA tests
Arista Biologicals Inc.