development and validation of the 3500 series genetic...

29
© 2010 Life Technologies Corporation 11-Apr-11 Title of the presentation, editable in the slide master 1 Development and Validation of the 3500 Series Genetic Analyzer for Human Identification Jeff Sailus Forensic Science Applications Group

Upload: others

Post on 18-Jul-2020

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Development and Validation of the 3500 Series Genetic ...tools.thermofisher.com/content/sfs/manuals/cms_093153.pdf · 3500 Series Genetic Analyzer for Human Identification Jeff Sailus

© 2010 Life Technologies Corporation11-Apr-11

Title of the presentation, editable in the slide master

1

Development and Validation of the 3500 Series Genetic Analyzer for Human Identification

Jeff SailusForensic Science Applications Group

Page 2: Development and Validation of the 3500 Series Genetic ...tools.thermofisher.com/content/sfs/manuals/cms_093153.pdf · 3500 Series Genetic Analyzer for Human Identification Jeff Sailus

© 2010 Life Technologies Corporation11-Apr-11

Title of the presentation, editable in the slide master

2

Development of a Next Generation Genetic AnalyzerA Whole System Approach

New Laser and Polymer Pump

Block

Redesigned Consumables with Tracking

Real-time System and Consumable

Information

Predefined Protocols and HID

Templates

Improved Temperature

Control Shorter Run Times

Optional Signal Normalization

Data Analysis QC Tools and

Flags

Simplified Reinjection

Scheme

Instrument Hardware

Consumables

Data Collection

QC Analysis

GMID-X v1.2

ID-X 1.1.1 Upgrade Supports Windows XP

and Vista

Analyzes .fsa & New .hid File Format

Consumable Data Tracking

Page 3: Development and Validation of the 3500 Series Genetic ...tools.thermofisher.com/content/sfs/manuals/cms_093153.pdf · 3500 Series Genetic Analyzer for Human Identification Jeff Sailus

© 2010 Life Technologies Corporation11-Apr-11

Title of the presentation, editable in the slide master

3

Improved Data Quality and Sample Throughput

Smaller Oven

More Consistent Migration for Better

Sizing Precision

Faster Run Time Higher Throughput

Flat Oven Door Seal and New Locking

Mechanism

Detection cell heater

Page 4: Development and Validation of the 3500 Series Genetic ...tools.thermofisher.com/content/sfs/manuals/cms_093153.pdf · 3500 Series Genetic Analyzer for Human Identification Jeff Sailus

© 2010 Life Technologies Corporation11-Apr-11

Title of the presentation, editable in the slide master

4

Instrument Setup and Performance

Elimination of Lower Polymer Block and Polymer Delivery

Tubing

Reduced Polymer WasteDirect Channels

Promote Efficient Polymer Flow

Redesigned Array Port

Minimized Potential for Bubbles and Leaks

Page 5: Development and Validation of the 3500 Series Genetic ...tools.thermofisher.com/content/sfs/manuals/cms_093153.pdf · 3500 Series Genetic Analyzer for Human Identification Jeff Sailus

© 2010 Life Technologies Corporation11-Apr-11

Title of the presentation, editable in the slide master

5

Instrument Setup and Performance

• Pre-Filled, Quality-Controlled Reagents

• Information Recorded via RFID• Lot numbers• Part Numbers• Serial numbers• Dates (expiration and installation)• Capacity/Usage

• Per Sample Running Cost Comparable to 31xx

Page 6: Development and Validation of the 3500 Series Genetic ...tools.thermofisher.com/content/sfs/manuals/cms_093153.pdf · 3500 Series Genetic Analyzer for Human Identification Jeff Sailus

© 2010 Life Technologies Corporation11-Apr-11

Title of the presentation, editable in the slide master

6

Data Collection: QC Analysis Tools for Preliminary Data Analysis

Review data quality flags, electropherograms and sizing tables - Select samples for reinjection

Page 7: Development and Validation of the 3500 Series Genetic ...tools.thermofisher.com/content/sfs/manuals/cms_093153.pdf · 3500 Series Genetic Analyzer for Human Identification Jeff Sailus

© 2010 Life Technologies Corporation11-Apr-11

Title of the presentation, editable in the slide master

7

3500 Developmental Validation Studies

Page 8: Development and Validation of the 3500 Series Genetic ...tools.thermofisher.com/content/sfs/manuals/cms_093153.pdf · 3500 Series Genetic Analyzer for Human Identification Jeff Sailus

© 2010 Life Technologies Corporation11-Apr-11

Title of the presentation, editable in the slide master

8

Developmental Validation ScopeHardware and Reagents

• Instruments• (3) 8 Capillary Instruments (3500)• (3) 24 Capillary Instruments (3500xL)

• Size Standard• GeneScan™ 600 LIZ® Size Standard v2.0• GeneScan™ 500 ROX™ Size Standard

• 4 Dye Amplification Kits do not use the normalization functionality

• Polymer and Array• POP-4™ and 36 cm Capillary Array

Page 9: Development and Validation of the 3500 Series Genetic ...tools.thermofisher.com/content/sfs/manuals/cms_093153.pdf · 3500 Series Genetic Analyzer for Human Identification Jeff Sailus

© 2010 Life Technologies Corporation11-Apr-11

Title of the presentation, editable in the slide master

9

Developmental Validation ScopeSample Summary

Test Name Samples Input Replicates

Genotype Concordance and Reproducibility

• 40 male and 42 female gDNA samples• 4 racial groups• Controls

1ng total DNA • 1 replicate per sample• 2 injections

Sizing Precision and Accuracy

Allelic ladder 1 µL per well • See below

Sensitivity and Normalization

• 007 control DNA• 3 gDNA samples

Kit dependent 0.125ng, 0.25ng, 0.5ng, 0.75, 1ng, 6ng

• 5 replicates for DNA samples• 3 replicates for NTC• 4 injections

Mixture Analysis • 4 pairs of male and female gDNA• Mixture ratios• 1:0, 1:1, 1:3, 1:5, 1:9, and 0:1

• Recommended concentrations per kit

1 replicate per sample

Resolution • Allelic Ladder• Research material

• 1 µL per well 24 replicates

Page 10: Development and Validation of the 3500 Series Genetic ...tools.thermofisher.com/content/sfs/manuals/cms_093153.pdf · 3500 Series Genetic Analyzer for Human Identification Jeff Sailus

© 2010 Life Technologies Corporation11-Apr-11

Title of the presentation, editable in the slide master

10

Validation ScopeAmpFℓSTR® Kits

• Identifiler®

• Minifiler™

• SGM Plus®

• Yfiler®

• Sinofiler™

• SEfiler Plus™

• Profiler Plus®

• Cofiler®

• Identifiler® Direct• Identifiler® Plus• NGM™

Page 11: Development and Validation of the 3500 Series Genetic ...tools.thermofisher.com/content/sfs/manuals/cms_093153.pdf · 3500 Series Genetic Analyzer for Human Identification Jeff Sailus

© 2010 Life Technologies Corporation11-Apr-11

Title of the presentation, editable in the slide master

11

Precision and Accuracy Studies

Page 12: Development and Validation of the 3500 Series Genetic ...tools.thermofisher.com/content/sfs/manuals/cms_093153.pdf · 3500 Series Genetic Analyzer for Human Identification Jeff Sailus

© 2010 Life Technologies Corporation11-Apr-11

Title of the presentation, editable in the slide master

12

3500 Sizing Precision Study:Within an Injection of Identifiler Allelic Ladder

Page 13: Development and Validation of the 3500 Series Genetic ...tools.thermofisher.com/content/sfs/manuals/cms_093153.pdf · 3500 Series Genetic Analyzer for Human Identification Jeff Sailus

© 2010 Life Technologies Corporation11-Apr-11

Title of the presentation, editable in the slide master

13

3500 Sizing Precision Study:Across Multiple Injections of Identifiler Allelic Ladder Samples

Page 14: Development and Validation of the 3500 Series Genetic ...tools.thermofisher.com/content/sfs/manuals/cms_093153.pdf · 3500 Series Genetic Analyzer for Human Identification Jeff Sailus

© 2010 Life Technologies Corporation11-Apr-11

Title of the presentation, editable in the slide master

14

Resolution and Baseline Evaluation

Page 15: Development and Validation of the 3500 Series Genetic ...tools.thermofisher.com/content/sfs/manuals/cms_093153.pdf · 3500 Series Genetic Analyzer for Human Identification Jeff Sailus

© 2010 Life Technologies Corporation11-Apr-11

Title of the presentation, editable in the slide master

15

Resolution

• SEfiler Plus™ (SE33 triplet) and Identifiler® (TH01 9.3/10) Allelic Ladders

• 160 injections per instrument on 6 instruments

• Research STR markers greater than 300 base pairs were also developed with single base pair differences (data not shown)

Page 16: Development and Validation of the 3500 Series Genetic ...tools.thermofisher.com/content/sfs/manuals/cms_093153.pdf · 3500 Series Genetic Analyzer for Human Identification Jeff Sailus

© 2010 Life Technologies Corporation11-Apr-11

Title of the presentation, editable in the slide master

16

Baseline Noise - Developmental Evaluation G5 Average Noise (PAT=Avg Noise+10x SD)

0

20

40

60

80

100

120

140

160

180

200

FAM VIC NED PET LIZ TOTAL

10x

Ave

rage

Std

. Dev

.

Data generated during the Sensitivity Studies were then used to confirm the level of pull up and an optimal peak amplitude threshold setting.

175 RFU

Page 17: Development and Validation of the 3500 Series Genetic ...tools.thermofisher.com/content/sfs/manuals/cms_093153.pdf · 3500 Series Genetic Analyzer for Human Identification Jeff Sailus

© 2010 Life Technologies Corporation11-Apr-11

Title of the presentation, editable in the slide master

17

Introduction to Normalization

Page 18: Development and Validation of the 3500 Series Genetic ...tools.thermofisher.com/content/sfs/manuals/cms_093153.pdf · 3500 Series Genetic Analyzer for Human Identification Jeff Sailus

© 2010 Life Technologies Corporation11-Apr-11

Title of the presentation, editable in the slide master

18

Normalization: What is it?

A method to attenuate signal variations associated with instrument, capillary array, sample salt load, and injection variability

• Factory Standardization: Hardware-based calibration to help enable more consistent instrument to instrument performance

• (Optional) Internal Standard Normalization:Chemistry and software based method to help enable more consistent performance across injections and instruments

• GeneScan™ 600 LIZ® Size Standard v2.0• 3500 Data Collection and GMID-X v1.2

Before normalization

After normalization

Page 19: Development and Validation of the 3500 Series Genetic ...tools.thermofisher.com/content/sfs/manuals/cms_093153.pdf · 3500 Series Genetic Analyzer for Human Identification Jeff Sailus

© 2010 Life Technologies Corporation11-Apr-11

Title of the presentation, editable in the slide master

19

Sources of Variation by Normalization Method

No NormFactory only

Int Std onlyFactor+IntStd

Injection to Injection

Cap to Cap

Within

Instrument to Instrument

Total

39%

33%

25%

14%

31%

24% 24%

13%

20%19%

5%5%

12%11%

4%4%3%

3%

0%0%

0%

5%

10%

15%

20%

25%

30%

35%

40%

%CV

Levels of Normalization

Sources of Variation

Instrument to Instrument largest source of variation

Factory + Internal Standard

Normalization greatest reduction

in variation

Page 20: Development and Validation of the 3500 Series Genetic ...tools.thermofisher.com/content/sfs/manuals/cms_093153.pdf · 3500 Series Genetic Analyzer for Human Identification Jeff Sailus

© 2010 Life Technologies Corporation11-Apr-11

Title of the presentation, editable in the slide master

20

Sensitivity and Internal Standard Normalization Studies

Page 21: Development and Validation of the 3500 Series Genetic ...tools.thermofisher.com/content/sfs/manuals/cms_093153.pdf · 3500 Series Genetic Analyzer for Human Identification Jeff Sailus

© 2010 Life Technologies Corporation11-Apr-11

Title of the presentation, editable in the slide master

21

Sensitivity Study:Average Peak Height across a Range of DNA Input Amounts AmpFℓSTR® MiniFilerTM PCR Amplification Kit

Normalization MethodFactory OnlyFactory + Internal Std

0 750 50 250 125

H1126

H1152

H1161 OO7

10000

5000

15000

00.125ng 0.25ng 0.50ng 0.75ng

10000

5000

15000

0

0.125ng 0.25ng 0.50ng 0.75ngGenomic DNA #2

Control DNA Standard 007

Genomic DNA #1

Genomic DNA #3

Peak

Hei

ght

Page 22: Development and Validation of the 3500 Series Genetic ...tools.thermofisher.com/content/sfs/manuals/cms_093153.pdf · 3500 Series Genetic Analyzer for Human Identification Jeff Sailus

© 2010 Life Technologies Corporation11-Apr-11

Title of the presentation, editable in the slide master

22

Size Standard Peak Height Consistency Study: Five AmpFℓSTR® PCR Amplification Kits GeneScan™ 600 LIZ® Size Standard v2.0

6000

5000

4000

3000

2000

1000

0

Lot 1 Lot 2 Lot 3

GS600 Peak Height

Identifiler®Identifiler® DirectIdentifiler® PlusMinifilerTM

NGMTM

7000

GS600 Lot Number

Peak

Hei

ght

Page 23: Development and Validation of the 3500 Series Genetic ...tools.thermofisher.com/content/sfs/manuals/cms_093153.pdf · 3500 Series Genetic Analyzer for Human Identification Jeff Sailus

© 2010 Life Technologies Corporation11-Apr-11

Title of the presentation, editable in the slide master

23

Size Standard Peak Height Consistency Study:Injection to Injection Peak Height Consistency

% CV

Normalization MethodFactory OnlyFactory + Internal Std

40

30

20

10

0Lot 1 Lot 2 Lot 3

Size Standard Peak Height Consistency · 6 instruments · 3 lots of size standard

Page 24: Development and Validation of the 3500 Series Genetic ...tools.thermofisher.com/content/sfs/manuals/cms_093153.pdf · 3500 Series Genetic Analyzer for Human Identification Jeff Sailus

© 2010 Life Technologies Corporation11-Apr-11

Title of the presentation, editable in the slide master

24

Size Standard Peak Height Consistency Study:%CV for 5 Different AmpFℓSTR® KitsGeneScan™ 600 LIZ® Size Standard v2.0

% CV

Factory + Internal Std Normalization · 6 instruments · 3 lots of size standard

Page 25: Development and Validation of the 3500 Series Genetic ...tools.thermofisher.com/content/sfs/manuals/cms_093153.pdf · 3500 Series Genetic Analyzer for Human Identification Jeff Sailus

© 2010 Life Technologies Corporation11-Apr-11

Title of the presentation, editable in the slide master

25

Additional Developmental Validation Studies

Page 26: Development and Validation of the 3500 Series Genetic ...tools.thermofisher.com/content/sfs/manuals/cms_093153.pdf · 3500 Series Genetic Analyzer for Human Identification Jeff Sailus

© 2010 Life Technologies Corporation11-Apr-11

Title of the presentation, editable in the slide master

26

Additional Developmental Studies

• Genotype Concordance• 100% genotype concordance with samples run on the 3130xL

• GeneMapper® ID-X v1.2 Verification• 100% Concordant with historical data set collected from legacy CE

instruments using a variety of AmpFℓSTR® kits analyzed and GeneMapper® ID-X v1.1 software versions (including mixtures)

• Normalization and RFID documentation functionality performed as expected

Page 27: Development and Validation of the 3500 Series Genetic ...tools.thermofisher.com/content/sfs/manuals/cms_093153.pdf · 3500 Series Genetic Analyzer for Human Identification Jeff Sailus

© 2010 Life Technologies Corporation11-Apr-11

Title of the presentation, editable in the slide master

27

In Summary• 100% genotype concordance between the 3500(xl) and 3130xl instrument and

GMID-X software

• Average Sizing precision data with GeneScan™ 600 LIZ® Size Standard v2.0 was 0.05 or less within an injections and across injections

• Linear relationship between input amounts and peak heights, with average peak heights ~6,000 rfu for 1ng input reactions Identifiler® Plus and NGMTM

• Internal Standard Normalization significantly reduced injection-to-injection and instrument-to-instrument variation, but not the absolute values of the allele peak heights

• No reproducible PCR artifacts were observed within the read regions - spectral pull up of less than 5% was observed

Page 28: Development and Validation of the 3500 Series Genetic ...tools.thermofisher.com/content/sfs/manuals/cms_093153.pdf · 3500 Series Genetic Analyzer for Human Identification Jeff Sailus

© 2010 Life Technologies Corporation11-Apr-11

Title of the presentation, editable in the slide master

28

FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.

©2009 Life Technologies Corporation. All rights reserved. The trademarks mentioned herein are the property of Life Technologies Corporation or their respective owners.

For those who require IVD-marked devices, the 3500 Dx and 3500xL Dx Genetic Analyzers and system accessories meet the requirements of the In Vitro Diagnostic Medical Devices Directive (98/79/EC). The 3500 Dx and 3500xL Dx systems are for distribution and use in specific European countries only. For more information about the 3500 Dx Series Systems, contact your Applied Biosystems representative.

Legal

Page 29: Development and Validation of the 3500 Series Genetic ...tools.thermofisher.com/content/sfs/manuals/cms_093153.pdf · 3500 Series Genetic Analyzer for Human Identification Jeff Sailus

© 2010 Life Technologies Corporation11-Apr-11

Title of the presentation, editable in the slide master

29

Thank you!

www.appliedbiosystems.com/3500HID