Development of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures by Mohammad Hassan Marvasti A thesis submitted in conformity with the requirements for the degree of Doctor of Philosophy (PhD) Department of Mechanical and Industrial Engineering University of Toronto © Copyright by Mohammad Hassan Marvasti 2014

Upload: phungkhanh

Post on 06-Feb-2018

236 views

Category:

Documents


4 download

TRANSCRIPT

Page 1: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

Development of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures

by

Mohammad Hassan Marvasti

A thesis submitted in conformity with the requirements for the degree of Doctor of Philosophy (PhD)

Department of Mechanical and Industrial Engineering University of Toronto

© Copyright by Mohammad Hassan Marvasti 2014

Page 2: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

ii

Development of Ultrasound Phased Array System for Weld

Inspections at Elevated Temperatures

Mohammad Hassan Marvasti

Doctor of Philosophy (PhD)

Mechanical and Industrial Engineering

University of Toronto

2014

Abstract

Interruption of plant operation can be avoided if non destructive testing inspections are

performed on-line at operating temperatures, which may be up to several hundred degrees

Celsius in a petrochemical or electric power generating plant. However, there are operational

temperature limits of the phased array transducers and associated plastic wedges used for

ultrasonic inspections. In addition, there is a major gap in terms of professionally-approved high-

temperature inspection techniques. In this project, the design and operation of an ultrasonic

phased array system are described for inspections of engineering components such as pipe welds

at elevated temperatures of up to 350oC. Wedges are built from plastics resistant to high

temperature degradation, and equipped with a cooling jacket around the array. A model of the

ultrasonic beam skew pattern due to thermal gradients inside a wedge is developed. The model is

used in a separate algorithm to calculate transmission and reception time delays on individual

array elements for generation of plane waves or focused beams in a hot test piece, while

compensating for thermal gradient effects inside the wedge.

The algorithm is also used to investigate the magnitude of thermal gradient effect on the

calculated time delays of the phased array elements. The algorithm results for inspections of test

Page 3: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

iii

pieces at 150oC demonstrate that application of conventional element time delays can lead to

serious phase errors. This results in major distortion of the desired beam profile, and very poor

imaging resolution. However, experimental trials indicate that plane waves and focused beams

can be generated in a hot test piece using the new focal law algorithm with appropriate timing

delays applied to all active array elements.

Page 4: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

iv

Acknowledgments

I would like to express my sincere gratitude to my supervisor, Professor Anthony Sinclair, for his

guidance and support without which this research work thesis would not have been possible to be

completed.

I am grateful to National Science and Engineering Research Council of Canada (NSERC), and

Eclipse Scientific for sponsoring my research and more importantly, for giving me a

distinguished opportunity to work on a rewarding academic/industrial collaborative research

project.

I wish to thank my colleagues Jonathan Lesage, Hossein Amini, Babak Shakibi and Jill Bond at

Ultrasonic Nondestructive Evaluation Laboratory at the University of Toronto for their help and

assistance in this project.

I would also like to express my gratitude to Robert Ginzel, Edward Ginzel, Jeff van Heumen and

the team at Eclipse Scientific for their kind helps in this project. Working with them was an

invaluable experience for me.

Finally, I am particularly thankful to my wonderful wife, my parents and my family for their

support and encouragement throughout the course of my thesis.

Page 5: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

v

Table of Contents

Abstract……………………………………………………………………………………………ii

Acknowledgments .......................................................................................................................... iv

Table of Contents ............................................................................................................................ v

List of Tables ................................................................................................................................ vii

List of Figures .............................................................................................................................. viii

1 Introduction .............................................................................................................................. 1

2 Background Theory and Literature Review.......................................................................... 4

2.1 Ultrasonic Non-Destructive Testing ................................................................................... 4

2.2 Fundamentals of Ultrasonic Wave Propagation in Solid Media ......................................... 4

2.2.1 Wave Propagation Modes ....................................................................................... 4

2.2.2 Speed of Propagation .............................................................................................. 5

2.2.3 Reflection, Refraction and Mode Conversion ........................................................ 6

2.2.4 Ultrasonic NDE System .......................................................................................... 6

2.2.5 Display Modes ........................................................................................................ 7

2.3 Principles of Phased Array Ultrasound ............................................................................... 9

2.3.1 Beam Steering and Focusing Using Phased Arrays .............................................. 10

2.3.2 Phased Array ultrasonic Inspection System .......................................................... 11

2.3.3 Phased Array Scanning Configuration .................................................................. 13

2.3.4 Phased Array Display Modes ................................................................................ 15

2.4 Weld Inspections with Phased Arrays .............................................................................. 16

2.5 Phased Array Inspection at Elevated Temperatures ......................................................... 17

3 High Temperature Wedges ................................................................................................... 20

4 Temperature Distribution Model ......................................................................................... 23

5 Velocity Measurements at Elevated Temperatures ............................................................ 26

Page 6: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

vi

6 Focal Law Algorithm – Planar Waves ................................................................................. 31

6.1 Room Temperature Inspection .......................................................................................... 31

6.2 Elevated Temperature Inspection ..................................................................................... 35

7 Experimental Validation – Planar Waves............................................................................ 44

7.1 Concept for Experimental Validation ............................................................................... 44

7.2 Experimental Details ......................................................................................................... 46

7.3 Experimental Results and Analysis - Room Temperature ................................................ 47

7.3.1 Systematic Errors .................................................................................................. 49

7.3.2 Random Errors ...................................................................................................... 51

7.4 Experimental Results and Analysis - Elevated Temperature ............................................ 53

8 Focal Law Algorithm – Focused Beam ................................................................................ 64

8.1 Algorithm Details .............................................................................................................. 65

8.2 Magnitude of Thermal Gradient Effect ............................................................................. 68

8.3 Experimental Evaluation ................................................................................................... 73

9 Summary, Conclusions and Future Work ........................................................................... 82

References ..................................................................................................................................... 84

Page 7: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

vii

List of Tables

Table 7-1 Inputs to the focal law calculating algorithm at room temperature: the levels of

uncertainty in the measured values of inputs are listed and the magnitude of their resulting effect

on the bias error on the first element of the aperture is presented. ............................................... 50

Table 7-2 Magnitude of the experimental results associated with generation of a 45o shear plane

wave at 150oC as depicted in Figure 7.5. The results of 15 repeated time delay measurements of

echo signals of the first and central elements of the active aperture are presented for two cases:

active aperture = elements 1-16; and active aperture = elements 49-64. ..................................... 59

Table 7-3 Experimental results associated with generation of a 60o shear plane wave at 150

oC as

depicted in Figure 6. The results of 15 repeated time delay measurements of echo signals of

selected elements of the active aperture are listed in separate columns. ...................................... 62

Page 8: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

viii

List of Figures

Figure 2.1 Ultrasonic wave propagation modes: (a) Longitudinal (Compression) mode and (b)

Shear (Transverse) mode [5]. .......................................................................................................... 5

Figure 2.2 Refraction and mode conversion of a longitudinal wave at the boundary of two media

[6]. ................................................................................................................................................... 6

Figure 2.3 Major components of a typical ultrasonic inspection set up in pulse/echo configuration

mode where one transducer is used for both transmission and reception of the ultrasonic waves

[8]. ................................................................................................................................................... 7

Figure 2.4 Representation of A-scan display: (a) transducer position, (b) signal display [9]. ....... 8

Figure 2.5 Representation of B-scan: (a) test configuration, (b) scan display [9]. ......................... 8

Figure 2.6 Representation of C-scan display: (a) transducer movement pattern (b) C-scan display

[9]. ................................................................................................................................................... 9

Figure 2.7 Phased array probe cross-sectional view [13]. ............................................................ 10

Figure 2.8 Beam steering with phased arrays, (a) unfocused beam (plane wave) and (b) focused

beam [14]. ..................................................................................................................................... 11

Figure 2.9 Typical components of a phased array inspection system. .......................................... 12

Figure 2.10 Electronic (linear) scan performed by focused beam normal to array elements [17].14

Figure 2.11 Sectorial scanning with phased arrays: the sound beam sweeps through a series of

angles [17]. .................................................................................................................................... 14

Figure 2.12 S-scan of 3 side drilled holes in a steel block through phased array sectorial scanning

[18]. ............................................................................................................................................... 15

Page 9: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

ix

Figure 2.13 Scan plan of a phased array ultrasonic inspection of a weld in a steel block, as

generated by a ray-tracing algorithm. The scan plan covers a range of inspection angles (46o-72

o)

inside the steel piece. The blue lines represent the direction normal to the plane wave fronts

generated by the active phased array elements. ............................................................................ 17

Figure 3.1 PEI high temperature wedge schematic with sloped front and dampening material at

the top of the wedge: ray 1 (solid line) and ray 2 (dashed line) illustrate reflection pattern of steep

and shallow beam incidence at the wedge bottom. As a result of wedge elongation and angled

front, the internally reflected beams reflect to the top of the wedge where they are absorbed by

the dampener. ................................................................................................................................ 22

Figure 3.2 High temperature phased array inspection system including: Eclipse WA10-HT55S-

IH-B PBI wedge prototype based on the geometry of the Olympus linear phased array model

5L16-A10 with 16 elements and center frequency of 5MHz, water cooling jacket, coolant tubing

system and mounting arms. ........................................................................................................... 22

Figure 4.1 PEI wedge temperature distribution modeling result, the color palate represents the

temperature distribution inside the wedge between 25oC and 150

oC. Surface temperature

measurements located on the dashed lines were used to validate the model results. ................... 25

Figure 4.2 Comparison of COMSOL and experimental results for the temperature distribution on

the surface of the PEI wedge mounted on a 150oC steel pipe. Solid lines on the graph and the

error bars represent average temperature values of 5 experiments and standard deviations of the

measurements: the red circles represent COMSOL model results at the specified locations on the

wedge surface. ............................................................................................................................... 25

Figure 5.1 Phase velocity measurement from two successive backwall echoes. The phase of

each pulse is determined relative to left side of the pulse acquisition window. ........................... 27

Figure 5.2 Experimental set up for measuring phase velocity at high temperatures. The sample

and the delay line were wrapped in high temperature insulation to lessen the temperature gradient

inside the plastic block. The mean block temperature was estimated from thermocouples placed

on top and bottom of the plastic block. ......................................................................................... 28

Page 10: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

x

Figure 5.3 Phase velocity data of PEI (a) and PBI (b) plastic blocks at selected elevated

temperatures. Solid lines and the error bars represent the mean and standard deviation of 5

repeated experimental measurements. .......................................................................................... 29

Figure 5.4 Phase velocity of PEI (a) and PBI (b) as a function of temperature at 5 MHz

frequency. The empirical relations shown on the graphs indicate the functional change of phase

velocity results with temperature T expressed in degrees Centigrade. ......................................... 30

Figure 6.1 Wave propagation pattern for a shear plane wave at angle Фs in a steel block at a

uniform temperature of 25oC. The ray traces are applicable for a wave transmitted from the array

into the test piece, or equivalently for a wave travelling from the test piece back to the array. In

the latter case, the waves associated with sample points along an arbitrary plane wave front

propagate in a direction perpendicular to the wave front and reach the piece-wedge interface at

locations labeled as interface points. Then they refract into the wedge based on Snell’s law and

propagate through the wedge to the array-wedge interface at points labeled as array-line points.

....................................................................................................................................................... 33

Figure 6.2 Wave propagation pattern for a shear plane wave at angle Фs in a steel block at a

uniform temperature of 150oC. The ray traces are applicable for a wave transmitted from the

array into the test piece, or equivalently for a wave traveling from the test piece back to the array.

In the latter case, the waves associated with sample points along an arbitrary plane wave front

propagate in a direction perpendicular to the wave front and reach the piece-wedge interface at

locations labeled as interface points. Then they refract into the wedge based on Snell’s law and

propagate through the wedge to the array-wedge interface at points labeled as array-line points.

The waves propagate along arced paths (red lines) in the wedge due to thermal gradient induced

velocity variation. ......................................................................................................................... 37

Figure 6.3 Calculated relative element time delays for the generation of a 60o shear plane wave

for Eclipse WA12-HT55S-IH-G PEI wedge and the Olympus linear phased array model 5L64-

A12 one a) the first 16 elements of the array (elements 1-16) were used and b) the last 16

elements of the array (elements 49-64) were used. The blue and red data points represent the

results for room (25oC) and elevated temperature (150

oC) inspection condition respectively and

the delays are expressed relative to the first element of the used active aperture......................... 39

Page 11: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

xi

Figure 6.4 Temperature profile inside an Eclipse WA12-HT55S-IH-G PEI wedge for inspection

of test piece at 150oC, the color palate represents the temperature distribution inside the wedge

between 25oC and 150

oC as obtained by our COMSOL finite element model. Waves emitted

from elements 1-16 propagate along the shortest path length in the wedge with a relatively steep

temperature gradient, compared to the waves emitted from elements 49-64. .............................. 40

Figure 6.5 Calculated element time delays for the generation of a 45o shear plane wave for an

Eclipse WA12-HT55S-IH-G PEI wedge and the Olympus linear phased array model 5L64-A12

one a) the first 16 elements of the array (elements 1-16) were used and b) the last 16 elements of

the array (elements 49-64) were used. The blue and red data points represent the results for room

(25oC) and elevated temperature (150

oC) inspection condition respectively and the delays are

expressed relative to the last element of the used active aperture. ............................................... 43

Figure 7.1 propagation of a wavefornt corresponding to a shear plane wave at an angle Фs in a

steel test block at elevated temperature: the wavefront is not straight inside the wedge due to

non-uniformity of the medium. However, it turns into a planar surface when the wave refracts

into the steel test piece at the wedge-piece interface. ................................................................... 45

Figure 7.2 Experimental results for the case of generating a 45o shear plane wave in the 45

o-

angled steel test block at a uniform temperature of 25oC using the first 16 elements of the array

(elements 1-16): the blue data points represent transmitting time delays applied to the elements in

transmission mode as calculated by our model. The solid red line represents the mean of 10

measurements of receiving delays at the same individual array elements. The error bars indicate

the standard deviation of the measurements at each individual element. ..................................... 48

Figure 7.3 Deviation times between the measured received echo time delays on each individual

elements and their associated theoretical value as calculated by our model for the case of

generating a 45o shear plane wave in the 45

o-angled steel test block at a uniform temperature of

25oC using the first 16 elements (data illustrated in Figure 7.2). The solid line and error bars are

associated with the mean and standard deviation of the measurements on each individual array

element. Note the very fine time scale on the vertical axis. ......................................................... 49

Figure 7.4 Experimental set up for evaluation of model results for generation of shear plane

waves in an angled steel test block at 150oC. An Eclipse WA12-HT55S-IH-G PEI wedge was

Page 12: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

xii

used along with an Olympus 5L64-A12 array for plane wave generation with a cooling jacket

around the array. The steel test block was located on a temperature-controlled hot plate and

wrapped in high temperature insulation which led to temperature variation over the entire volume

of the test piece of less than 10oC. ................................................................................................ 54

Figure 7.5 Experimental results for generation of a 45o shear plane wave in a 45

o-angled steel test

block at 150oC using an Olympus 5L64-A12 array and Eclipse WA12-HT55S-IH-G PEI wedge

when the first 16 elements (a) and last 16 elements (b) of the array were used. The red and green

data points represent the relative time delays calculated by our algorithm, and the mean of the

measured echo time delays, respectively. The green points are each the average of 15 repeated

experiments and error bars indicate the uncertainty level in the value of the green points. Blue

data points represent the calculated time delays for room temperature inspection. ..................... 56

Figure 7.6 Experimental results for generation of a 60o shear plane wave at 150

oC when the first

16 elements (a) and last 16 elements (b) of the array were used. The red and green data points

represent the relative time delays calculated by our algorithm, and mean of the experimentally

measured received echo time delays, respectively. The green points are each the average of 15

repeated experiments; the associated error bars indicate the uncertainty level in the value of the

green points. Blue data points represent the calculated time delays for room temperature

inspection, presented for sake of comparison. .............................................................................. 61

Figure 8.1 Wave propagation pattern for focusing an ultrasonic beam at an arbitrary focal point

in a steel block at a uniform temperature of 25oC. Wave propagation paths are shown from

various array-line points along the phased array to the selected focal point. The propagation paths

are straight lines in both the wedge and the steel block, each with a single wave propagation

velocity at 25oC. The refraction angle at each interface point is determined by Snell’s Law. ..... 66

Figure 8.2 Wave propagation pattern for focusing an ultrasonic beam at an arbitrary focal point

in a steel block at a uniform temperature of 150oC. Wave propagation paths are shown from

various array-line points along the phased array to the selected focal point. The propagation paths

are straight lines in the steel block based on the assumption of uniform temperature inside the

block (blue lines). After the waves refract into the wedge, they propagate along arced paths (red

lines) due to velocity variations across the heated wedge that are induced by thermal gradients. 68

Page 13: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

xiii

Figure 8.3 Location of the selected focal points and calculated near field points in a steel block

on the sound paths representing the waves with refraction angles of 40o, 55

o and 70

o using an

Eclipse WA12-HT55S-IH-G PEI wedge-5L64-A12 Olympus array system. The first 16 elements

of the aperture are used for beam formation and the each sound path is represented by a straight

line emitted from the point located at the center of the active aperture. ....................................... 69

Figure 8.4 The calculated time delays of individual elements for focusing an ultrasonic beam at

the selected focal points located on the wave sound paths representing (a) 45o, (b) 55

o and (c) 70

o

refraction angles inside a steel test block using an Eclipse WA12-HT55S-IH-G PEI wedge-5L64-

A12 Olympus array system (as depicted in Figure 8.3). The first 16 elements of the aperture are

used for beam formation. The blue and red data points represent the results for room (25oC) and

elevated temperature (150oC) inspection condition respectively. In each case, delays are

expressed relative to the element associated with the longest travel time from array to the focal

point. This element is fired first with delay time defined as time zero; all other elements are then

fired at their relative delay times. ................................................................................................. 71

Figure 8.5 Calibration block for steering assessment of phased array probe based on the

amplitudes of the echo signals received from a series of 2-mm diameter side drilled holes. The

holes are located at 5o interval on an arc 50 mm from the surface location where the midpoint of

the array’s active aperture is located. A focused beam is electronically swept along the arc with

1o angular steps. Amplitudes of the received echo signals are analyzed. High amplitudes are

expected to correspond to the locations of the holes. The clarity of the amplitude peaks as a

function of steering angles indicates the angular range of the array. ............................................ 74

Figure 8.6 Design specifications of a steel block for performance assessment of a wedge-array

system. The block was specifically designed for use with Eclipse WA12-HT55S-IH-G PEI

wedge and first 16 elements of the 5L16-A12 linear phased array probe. The blue straight lines

represent the wave paths associated with refraction angles ranging from 40o up to 70

o in 1

o

increments. The red arc represents the boundary of the near field on the specified sound paths.

The black arc links the locations of the focal points on the presented sound paths. These are

located at the same travel time distance from the center of the active aperture, and are all within

the near field. ................................................................................................................................ 76

Page 14: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

xiv

Figure 8.7 Configuration of the designed steel block for angular resolution and steering

assessment of an Eclipse WA12-HT55S-IH-G PEI Wedge-5L64-A12 Olympus array system: a

set of four side drilled holes are selected as reflectors. The holes are centered on the focal points

which lie along the sound paths representing the refraction angles s of 44o, 49

o, 56

o and 66

o.

The straight lines represent the wave paths associated with refraction angles ranging from 40o up

to 70o in 1

o increments. The black and red points represent the location of focal points (steering

samples) and the near field points on the specified sound paths. ................................................. 77

Figure 8.8 Experimental set up for evaluation of model results for focusing the ultrasonic beam

in a steel test block at 25oC. An Eclipse WA12-HT55S-IH-G PEI wedge was used with a 5L64-

A12 Olympus array for generation of focused beams. The steel test block contains four side

drilled holes at locations specified in Figure 8.7. ......................................................................... 77

Figure 8.9 Assessment results of an Eclipse WA12-HT55S-IH-G PEI wedge-5L64-A12 Olympus

array system at ambient temperature of 25oC. The focused beam was swept electronically along

the focal points as depicted in Figure 8.7 and Figure 8.8. The maximum amplitude of the

analytical form of the echo signals was plotted versus the refraction angles (blue points). The

local amplitude peaks correspond to refraction angles of 44o, 49

o, 56

o and 66

o inside the steel test

block (dashed lines). These represent the location of the side drilled holes. ................................ 78

Figure 8.10 Assessment results of an Eclipse WA12-HT55S-IH-G PEI wedge-5L64-A12

Olympus array system at 150oC. The focused beam was swept electronically along the focal

points as depicted in Figure 8.7 and Figure 8.8 and the maximum amplitude of the analytical

form of the echo signals was plotted versus the refraction angles. Two sets of focal laws were

used for beam formation: First, the focal laws calculated for ambient temperature condition

(25oC) and then the focal laws calculated by our new model for the actual experimental condition

(150oC test block). The results corresponding to these cases are presented by the blue and red

points, respectively. The dashed lines represent refraction angles which correspond to location of

the side drilled holes in the steel test block. .................................................................................. 80

Page 15: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

1

1 Introduction

Many industrial sectors such as power generation, petrochemical production, and metal

processing have systems that operate at elevated temperatures for extended time periods, during

which some failure mechanisms can be significantly accelerated. Regular Non-Destructive

Testing (NDT) of high-temperature systems is therefore required. In particular, ultrasonic NDT

with phased array transducers has been found to be an accurate and convenient method for

inspection of many industrial piping systems and welds.

Phased array ultrasonic transducers create a user-specified beam profile by a controlled pulse

sequencing of the array elements determined by software known as a focal law calculator. A

phased array’s beam forming capability introduces unique benefits for weld inspection:

electronic beam steering provides rapid and accurate scanning of the weld sections, and

inspection angles can be tailored based on weld geometry to increase scanning efficiency

specifically for complex-shaped welds.

For a weld inspection, the beam travels from the array through a plastic wedge, which has the

primary function of coupling sound energy from the array transducer to the test piece at a

prescribed angle. An ultrasonic beam refracts from the wedge into the test piece where it

interacts with flaws; echoes are then returned to the same wedge and transducer, or a separate

receiver for analysis to determine flaw locations and sizes.

Interruption of plant operation can be avoided if NDT inspections can be performed on-line at

operating temperatures; this may be up to several hundred degrees Celsius. This capability

avoids costly downtime and reduces the risk of damage from thermal cycling associated with

periodic shutdowns. However, high temperature phased array inspections cannot be routinely

performed: there are operational temperature limits of the phased array transducers and plastic

wedges; in addition, there is a gap in terms of approved and appropriate high temperature

inspection techniques.

Wedges for high temperature inspections can be built from materials that have low ultrasonic

attenuation, appropriate values of acoustic impedance, and resistance to high temperature

degradation. A cooling jacket can be placed on the wedge to protect the array. However, this

Page 16: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

2

introduces new problems: inside the wedge, thermal gradients lead to variations in temperature-

dependent wave velocity and skewing of the waves; the focal law calculator, which is based on

the assumption of wave propagation in homogeneous propagation media, would become

inaccurate. Refraction angles of ultrasound into the test piece become altered by elevated wedge

temperature and skewed ultrasound propagation paths. To date, the significance of these factors

has not been thoroughly evaluated.

The objectives of this research are to:

1) Model the beam skew induced by thermal gradients inside the heated wedge,

2) Develop an algorithm to calculate appropriate time delays for individual array elements for

generation of a shear plane wave or a focused beam in a hot test piece,

3) Investigate the amount of distortion to the focal law calculator at elevated temperature, as a

function of key scanning parameters; these parameters include the selection of the active array

aperture, plus the magnitude and orientation of the thermal gradient inside the high temperature

wedges, and

4) Validate the new focal law calculator experimentally.

In Chapter 2, we present a brief overview of the fundamentals of ultrasonic wave propagation in

solid media, basics of non-destructive testing, a description of a typical phased array weld

inspection system and its limitations at elevated temperature.

In Chapters 3 to 8 of this thesis, we progress through 6 major steps of this project. First, the

design specifications of high-temperature wedges with a cooling jacket around the array are

described in Chapter 3. The temperature distribution inside the wedges is modeled using

COMSOL Multiphysics1 finite element software package and validated experimentally (Chapter

4). Next, the dependence of compression wave velocity on temperature in wedge materials is

measured experimentally, and the results are presented in Chapter 5.

1 1 New England Executive Park, Burlington, USA.

Page 17: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

3

The material from Chapters 3-5 is then combined with a numerical ray-tracing technique to

model the skewed path of waves propagating across thermal gradients in the heated wedge. This

is followed by development of an original algorithm to calculate appropriate time delays for

individual array elements for generation (or reception) of a shear plane wave in a hot test piece

based on the beam skew model results. The newly developed algorithm is described in Chapter 6

and its results for generation of planar shear waves at refraction angles of 45o and 60

o inside a

steel test piece at elevated temperature are presented.

In chapter 7, a set of experiments are conducted to evaluate the quality of the shear plane wave

generated in a hot steel test piece using the new focal law calculator described in Chapter 6.

Finally, in Chapter 8, a second set of experiments is conducted to evaluate the performance of

the new focal law calculator. This time, the goal is to focus the array at a single user-specified

point inside the hot test piece. The experimental results are presented and analyzed.

Conclusions are presented in Chapter 9. This includes a summary of results and suggestions for

future work.

Page 18: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

4

2 Background Theory and Literature Review

2.1 Ultrasonic Non-Destructive Testing

Many industrial sectors have systems that operate under severe conditions such as elevated

temperatures and corrosive environments. Failure mechanisms can be significantly accelerated

under these conditions [1]. Regular non-destructive testing (NDT) is therefore required to avoid

costly failures [2]. In particular, ultrasonic NDT has been found to be an accurate and convenient

method for inspection of many industrial piping systems and welds [3] [4].

In ultrasonic inspections, high-frequency mechanical waves travel through materials and

partially reflect from discontinuities. The reflected or transmitted signal is detected and displayed

as an electric signal. Analysis of the obtained signals provides information on the location and

size of the discontinuities [5].

A brief review of the fundamentals of ultrasonic wave propagation in solid media and the basics

of ultrasonic non-destructive testing will be presented in the following sections.

2.2 Fundamentals of Ultrasonic Wave Propagation in Solid Media

2.2.1 Wave Propagation Modes

An ultrasonic wave can be generated by applying a time-varying mechanical stress to a medium.

This leads to vibrational motion of particles (atoms) in the medium about their equilibrium

positions. The sound waves propagate through the medium in different modes based on the way

the particles oscillate. Longitudinal and shear modes are the two modes of propagation in bulk

media that are widely used in ultrasonic testing [5].

A longitudinal (compression) wave or L-wave is generated by applying a normal stress to the

material surface. In this mode, particle vibrations occur in the direction of wave propagation and

cause compression and refraction regions along the propagation path (Figure 2.1.a).

Shear (transverse) wave or S-wave is excited by applying a shear force at the surface of a

medium; this causes wave propagation through the medium by shear stresses. This mode results

in atomic oscillations perpendicular to the direction of wave propagation (Figure 2.1.b).

Page 19: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

5

Figure 2.1 Ultrasonic wave propagation modes: (a) Longitudinal (Compression) mode and (b) Shear (Transverse)

mode [5].

2.2.2 Speed of Propagation

The speed at which the wave propagates depends on the properties of the medium. In an elastic,

isotropic, solid medium the speed of ultrasound is determined by the elastic constants (Young’s

and Shear modulus) and density of the medium. Variations of these parameters result in changes

to wave speed as it propagates through the medium [5] [6].

Temperature variation is frequently a cause of non-homogeneity in a wave propagation medium.

A material’s elastic moduli and its density are both temperature dependent properties. Therefore,

the presence of a thermal gradient in a medium results in regions with different local mechanical

properties. As a wave propagates in such media, its velocity varies due to a dependency on local

temperature-dependent mechanical properties [7].

Page 20: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

6

2.2.3 Reflection, Refraction and Mode Conversion

When an ultrasonic wave reaches an interface between two media with different mechanical

properties (Figure 2.2), part of the sound is reflected back into the first medium and the rest is

refracted and transmitted into the second medium according to Snell’s Law. Ultrasonic waves

can also undergo mode conversion between shear and longitudinal modes upon oblique

incidence at an interface. In the case of spatially continuous variation of material properties (e.g.,

due to temperature gradients), there can be a continuous change in the propagation amplitude and

direction (skewing) of the wave [6].

Figure 2.2 Refraction and mode conversion of a longitudinal wave at the boundary of two media [6].

2.2.4 Ultrasonic NDE System

Major components of a typical ultrasonic NDE system are shown in Figure 2.3. An ultrasonic

pulser generates short high-voltage pulses at regular intervals. The pulses excite an ultrasonic

transducer which contains a piezoelectric element that converts electrical pulses into mechanical

pulses and vice versa. Mechanical pulses propagate through the engineering component as

ultrasound waves and reflect back to the transducer after interaction with flaws. The transducer

now acts as a receiver and the piezoelectric element converts the received ultrasonic waves to

electric signals. The signals are then amplified, digitized and displayed on an oscilloscope or a

computer screen for further analysis [8].

Page 21: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

7

Figure 2.3 Major components of a typical ultrasonic inspection set up in pulse/echo configuration mode where one

transducer is used for both transmission and reception of the ultrasonic waves [8].

2.2.5 Display Modes

Ultrasonic signals generated by single-element transducers are commonly collected and

displayed in one of the three following formats [6] [7]:

A-Scan: in this type of display, the amplitude of the received signal is plotted versus time as

shown in Figure 2.4. Information on the location of the flaws can be provided by analysis of

time-of-arrival of ultrasonic reflector echoes.

B-Scan: to form a B-scan display, a sequence of A-scans is generated by moving the transducer

along the top surface of the specimen (x-axis). The A-scans are each converted to an intensity

graph (gray-scale pattern) and placed side-by-side. A cross sectional view of the specimen can be

obtained using a B-scan (Figure 2.5).

Page 22: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

8

Figure 2.4 Representation of A-scan display: (a) transducer position, (b) signal display [9].

Figure 2.5 Representation of B-scan: (a) test configuration, (b) scan display [9].

Page 23: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

9

C-Scan: The C-scan provides a two dimensional view of specimen flaw echoes within a specified

depth range z. The transducer is moved in a raster-like manner on the top x-y surface plane of

the specimen and captures A-scans at each step. The magnitude of the largest echo within the

specified z depth range for each A-scan is then converted to a gray-scale pattern for display. An

example of the transducer movement pattern and a C-scan display are shown in Figure 2.6.

Figure 2.6 Representation of C-scan display: (a) transducer movement pattern (b) C-scan display [9].

2.3 Principles of Phased Array Ultrasound

Ultrasonic phased arrays were initially introduced in the field of medical imaging. More recently,

numerous advantages of utilizing phased array ultrasound technology over conventional

ultrasonic transducers have led to their increasing use for many industrial settings such as weld

inspections [10] [11] [12].

Conventional ultrasonic transducers commonly consist of a single piezoelectric element which

can function as transmitter, receiver or both. A phased array transducer, however, contains

Page 24: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

10

multiple individual piezoelectric elements. The elements are connected to separate electric

channels and can be pulsed individually. Figure 2.7 illustrates a linear array with rectangular

footprint which is a common configuration of a phased array probe.

Figure 2.7 Phased array probe cross-sectional view [13].

The piezoelectric elements in a phased array transducer can be fired at different times. After

being pulsed, the waves generated by different elements undergo constructive and destructive

interference with each other to create a new composite wave. The element firing times can be

adjusted to steer the composite beam at different angles, or to create a focused beam at a point

located at a depth specified by the user.

2.3.1 Beam Steering and Focusing Using Phased Arrays

To generate a planar wave which propagates at a certain angle in a homogenous and isotropic

medium, the piezoelectric elements should be fired with specific relative time delays. In this case

the acoustic fields generated by all the individual array elements should interfere constructively

to produce uniform acoustic pressure along each plane that is normal to the desired direction of

propagation. The appropriate delay pattern and its magnitude depend on factors such as speed of

ultrasonic waves in the medium, spacing between the array elements and desired propagation

angle of the beam with respect to the array line [7] [10] [14].

Page 25: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

11

A set of delay times to be applied to the elements for composition of a desired beam is known as

the focal law. By changing the focal law applied to array elements, the angle of the generated

beam can be controlled electronically. A sequence of incremental changes in the focal law results

in steering the composite beam generated by the array, i.e., sweeping out a fan-shaped sector.

The array can also be used to generate a composite beam focused at a specified point in the

medium. To achieve this, an appropriate focal law must be applied to the elements such that

waves generated by individual elements all interfere constructively at the desired location. Travel

times of the waves from each element to the focal point are evaluated to obtain the appropriate

focal law. Figure 2.8 illustrates beam steering and focusing with phased arrays.

Figure 2.8 Beam steering with phased arrays, (a) unfocused beam (plane wave) and (b) focused beam [14].

2.3.2 Phased Array Ultrasonic Inspection System

The major components of a typical industrial phased ultrasonic inspection system can be seen in

Figure 2.9. The phased array probe is connected to a pulser/receiver instrument, and mounted on

a plastic wedge in direct contact with the test piece. The ultrasonic beam propagates through the

wedge and refracts into the test piece where it interacts with flaws; echoes are then returned to

the same wedge and transducer (or a second wedge and transducer) and received by the phased

array instrument for further analysis to determine flaw locations and sizes.

Page 26: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

12

Figure 2.9 Typical components of a phased array inspection system.

The phased array instrument contains software known as a focal law calculator, which

determines the relative pulse delay time for each array element based on the desired beam angle,

focal distance, probe and wedge characteristics as well as acoustical properties of the test piece

material. The focal law calculator software is connected to electrical channels that apply the

appropriate excitation sequence to the individual array elements [10] [14].

Phased array probes can be found in a wide range of sizes, shapes, frequencies and number of

elements. However, the linear array with a rectangular footprint is the most common

configuration for industrial use. The resonant frequency of the probe, number of elements,

element spacing and element dimensions are the main factors to be considered in probe selection.

These parameters influence a probe’s capability in focusing and steering and it’s functionality in

generating desirable beam profiles. Therefore, probe selection should be driven by the

application [15].

Page 27: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

13

A phased array probe assembly usually includes a plastic wedge. The ultrasonic wedge is a key

component of the inspection system and has the primary function of coupling sound energy from

the phased array transducer to the test piece at a specified contact angle. At the wedge-test piece

interface, the wave refracts according to Snell’s law, and may also convert between shear and

compression modes according to the wedge design [16]. The wedge typically has an added

component called a dampener (Figure 2.9), which absorbs any incident ultrasonic pulses so that

they do not bounce back into the wedge and generate spurious “ghost echo” signals. The wedges

come in many different sizes and styles to be compatible with different probe arrays.

2.3.3 Phased Array Scanning Configurations

Focal laws applied to individual elements in a multi-element phased array probe can be

controlled by computer. The ability of applying fast-changing automated computer–controlled

excitation to the array elements and the capability to generate beams with defined parameters

such as angle and focal distance, enable unique scanning configurations such as electronic

(linear) scanning and sectorial scanning.

In electronic (linear) scanning configuration, the same focal law is multiplexed across a group of

array elements and fixed beam angles are produced and scanned electronically along the array.

This is equivalent to mechanically scanning a conventional ultrasonic transducer along a distance

equal to the length of the phased-array probe. Using phased arrays, both focused and unfocused

beams can be generated for scanning electronically along a desired length in the medium. Figure

2.10 shows a linear scan performed by a focused beam with constant angle along the probe

length [10] [14].

Page 28: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

14

Figure 2.10 Electronic (linear) scan performed by focused beam normal to array elements [17].

In the sectorial scanning configuration however, ultrasonic pulses are launched sequentially at a

sequence of angles into the test specimen by automated application of multiple focal laws. Focal

laws are produced to generate beams which sweep through an angular range with a specific

increment. Figure 2.11 illustrates the principle of sectorial scan.

Figure 2.11 Sectorial scanning with phased arrays: the sound beam sweeps through a series of angles [17].

Page 29: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

15

With these scanning configurations, a large area can be covered for inspection in a very short

time.

2.3.4 Phased Array Display Modes

A-scan, B-scan and C-scan modes obtained with conventional single-element transducers can

also be generated by phased array equipment. In addition to these conventional display modes,

another mode known as a sectorial scan (S-scan) can be generated by the application of phased

arrays. The A-scan echo signal corresponding to each angle (focal law) is recorded and

transformed to a colour intensity graph of signal intensity vs. travel distance. The A-scans are

then assembled together sequentially to form a cross sectional view of a pre-shaped sector of the

specimen [10] [14]. Figure 2.12 illustrates an S-scan representation of three side-drilled holes in

a steel block.

Figure 2.12 S-scan of 3 side drilled holes in a steel block through phased array sectorial scanning [18].

Page 30: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

16

2.4 Weld Inspections with Phased Arrays

Defects are often generated through the welding process. These defects may propagate under

operational service conditions and cause mechanical failures leading to severe consequences.

Regular non-destructive testing is therefore required.

Ultrasonic NDT with phased arrays has been found to be an accurate and convenient method for

inspection of many industrial weld systems. A phased array’s beam forming capability

introduces unique benefits for weld inspection: 1) weld sections are scanned rapidly and

accurately by electronic beam steering, 2) sectorial scanning provides a large field of coverage

from a single probe location, using beams with a sequence of refraction angles, 3) inspection

angles can be tailored based on weld geometry to increase scanning efficiency specifically for

complex-shaped welds, and 4) defect detection probability can be maximized by optimizing

beam shape and size through electronic focusing [19] [20].

Application of phased array ultrasound for weld inspection has increased in recent years due to

the above mentioned advantages. Therefore, many codes and standards have been developed

describing requirements and criteria for weld inspections using phased arrays. A key point

emphasized in these codes is that appropriate settings and scanning parameters must be selected

that ensure full inspection coverage of the weld area [21].

A document known as a “scan plan” is normally prepared before an inspection to lay out the

inspection methodology. Potential scanning configurations and beam angles are developed

through the scan plan and beam directions are identified to check that full coverage of the weld

section is achieved.

Ray tracing software is commonly used for scan plan generation. These computer programs

provide a simplified pictorial view of the projected path of the selected ultrasonic beams inside

the wedge and the test piece, along with a graphical representation of the weld overlay. An

example of use of such a ray tracing program is illustrated in Figure 2.13.

In this Figure, each beam is approximated by a straight line emitted from a point located at the

center of the active array elements to show its propagation direction. For the example of Figure

Page 31: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

17

2.13, it is demonstrated that full volume coverage of the weld section is obtained using a set of

beams with refraction angles ranging from 46o to 72

o in a sectorial scan configuration. It is seen

that visualization of beam path helps the inspector to optimize the scanning parameters for

efficient inspection of a specific weld geometry.

Once the scanning parameters have been finalized through the scan plan, an appropriate time

delay for excitation of each active array element is obtained by the focal law calculator. Plane

waves with desired propagation directions are then generated. Ultrasonic waves refract from the

wedge into the test piece where then interact with flaws; echoes may then be returned to the

same wedge and transducer for analysis to determine flaw locations. When the weld

discontinuity has been detected and located, electronic focusing can be used to optimize beam

shape and size at the expected defect location for accurate sizing and flaw characterization [14].

Figure 2.13 Scan plan of a phased array ultrasonic inspection of a weld in a steel block, as generated by a ray-tracing

algorithm. The scan plan covers a range of inspection angles (46o-72

o) inside the steel piece. The blue lines represent

the direction normal to the plane wave fronts generated by the active phased array elements.

2.5 Phased Array Inspection at Elevated Temperatures

Interruption of plant operation can be avoided if NDT inspections can be performed on-line at

operating temperatures; this may be up to several hundred degrees Celsius. This capability

Page 32: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

18

avoids costly downtime and reduces the risk of damage from thermal cycling associated with

periodic shutdowns [22]. However, high temperature phased array inspections cannot be

routinely performed. There are operational temperature limits of the phased array transducers

and plastic wedges; in addition, there is a gap in terms of approved, appropriate high temperature

inspection techniques. These limitations are explained briefly in the following paragraphs.

High temperature application of phased array transducers is limited due to two important factors:

the piezoelectric element and internal bonding of the multi-layered structure. PZT is the most

common piezoelectric material used in phased array transducers and its maximum operational

temperature is approximately 270oC [22]. However, the upper operational temperature limit for

phased array transducers is about 50oC [15]. At higher temperatures, the transducer’s internal

bonds and plastic-based materials can undergo permanent damages due to thermal expansion and

distortion.

Efforts have been made to develop single-element ultrasonic transducers to operate at elevated

temperatures, up to several hundred degrees Celsius. Many different candidate piezoelectric

materials have been tested to check their functionality at these temperatures. Also, efforts have

been made to optimize techniques used for bonding transducer’s internal layers in a manner such

that they can withstand high temperatures. These efforts have led to custom manufacturing of

high temperature conventional ultrasound probes. However, manufacturing processes have

generally been reported to be challenging and costly [22].

Only a few attempts were reported on manufacturing high temperature ultrasonic arrays. Kirk et

al. designed a piezoelectric linear array structure to operate at up to 400oC. They replaced

conventional piezoelectric materials with lithium niobate and used a single plate of piezoelectric

element to avoid the costs and challenges of manufacturing multiple elements. A series of

discrete narrow electrodes were defined and precisely clamped on the lithium niobate plate by a

patterning technique. Array operation was then achieved by activating the electrodes in

sequence. Once an electrode was activated, an electric field was applied to the small part of the

piezoelectric plate in contact with the electrode leading to generation of an ultrasound pulse. The

array was attached to a steel block with simulated defects and its capability was demonstrated by

scanning for defects at elevated temperatures [23].

Page 33: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

19

A similar approach was used by Shih et al. They reported the development of an ultrasonic array

transducer for non-destructive testing at 150oC. The array was made by spraying a newly

developed piezoelectric composite film on a titanium foil by the sol-gel method followed by

colloidal silver spraying to deposit electrodes. The manufactured array was attached to a steel

pipe and used for thickness measurements through post-processing of the signals stored in pulse-

echo mode using each single element as a transmitter and multi elements as receivers [24].

The arrays described above are intended to be fixed permanently to the test subjects at a single

location. Their use is generally confined to sites where defects have already been found, to

monitor their growth behavior over a period of time. This application is further limited by piping

and pressure vessel codes which prohibit attachment of inspection instruments to some

engineering components to avoid the possible promotion of flaw generation.

The limited potential application of high temperature arrays precludes their high development

and manufacturing costs. Therefore there is a great interest in developing systems which use

conventional phased arrays probes and wedges for inspection at elevated temperatures.

This project describes the design of a phased array system for inspection of test pieces at

elevated temperatures of up to 350oC using conventional phased array probes and specially

designed wedges. Wedges are built from plastics resistant to high temperature attack and

equipped with a cooling jacket around the array. However, inside the wedge, thermal gradients

lead to variations in temperature-dependent wave velocity and skewing of the ultrasonic waves.

Therefore, a conventional focal law calculator, which is based on the assumption of wave

propagation in straight lines at a single velocity, would yield inaccurate results.

A model of the beam skew pattern in the presence of thermal gradients inside the wedge is

developed to predict the magnitude of this effect. The model is then used to redesign the focal

law calculator to accommodate the non-homogeneity of wave speeds in the heated wedge. Beam

forming accuracy of the redesigned focal law calculator is evaluated experimentally to

demonstrate the applicability of the system for performing phased array inspections at elevated

temperatures. Details are provided in Chapters 3 to 8 of this thesis.

Page 34: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

20

3 High Temperature Wedges

The ultrasonic wedge is a key component of the inspection system and has the primary function

of coupling sound energy from the phased array to the test piece. At the wedge-test piece

interface, the wave refracts according to Snell’s law, and may also convert between shear and

compression modes according to the wedge design.

Rexolite2 is a commonly used wedge material for phased array inspection of pipelines and

pressure vessels. It is an isotropic, inexpensive, polystyrene plastic with a moderately low level

of attenuation. However, use of Rexolite is limited to temperatures well below 100oC, which is

close to its glass transition temperature [25].

For elevated temperature inspection purposes, ultrasonic wedges must be manufactured using

plastics with a high glass temperature and suitable acoustic properties. Polyetherimide (PEI) and

polybenzimidazole (PBI) plastics are two such candidate materials. PEI is a polyetherimide

thermoplastic high heat resistant polymer, and PBI is a polybenzimidazole engineering plastic,

with glass temperatures of 210oC and 413

oC, respectively [26] [27].

A prototype wedge with water cooling jacket was designed out of PEI based on the geometry of

the Olympus3 linear phased array model 5L64-A12 with 64 elements and center frequency of

5MHz. The wedge was designed to generate shear waves with a nominal 55o angle inside the test

piece (steel) using Snell’s law and an initial assumption of a uniform temperature of 20oC inside

the wedge.

The path of propagating waves can be monitored using ray-tracing software such as the

BeamTool4 package that is based on the assumption of linear wave propagation in each

homogeneous, isotropic material, with Snell’s law to determine the refraction angles at each

material interface.

2 C-LEC Plastics Inc., 6800 State Rd, Philadelphia, PA 19135, United States.

3 Olympus NDT Canada, 505, boul du parc-Technologique Quebec City, QC, Canada.

4 Eclipse Scientific Inc., Waterloo, ON, Canada.

Page 35: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

21

Conventional wedges have a dampening material embedded at the front of the wedge to diffuse

any internally reflected signals. Because of the low heat tolerance of the dampening material, this

traditional approach is not practical in a high temperature environment as the dampening

material could melt off the wedge or, even combust during inspection.

In order to be able to provide adequate damping within high temperature wedges three major

design elements were implemented by engineers at Eclipse Scientific Inc. First, the dampening

material was installed at top of the wedge rather than on the front of the wedge. This provides

adequate isolation from the high-temperature test piece to protect the dampening material.

Second, the wedge itself was elongated. Elongation of the wedge provides a beam path for

reflected signals generated from steep incident angles to be reflected to the top of the wedge and

diffused by the dampener and third, the front of the wedge was angled to cause signals generated

at a shallow incident angle to be reflected up to the top of the wedge and diffused by the

dampener. These three innovations allow the high temperature wedge to eliminate internally

reflected signals as effectively as a traditional wedge. These design concepts are illustrated in a

PEI wedge schematic picture shown in Figure 3.1.

A similar procedure was followed for designing PBI high temperature wedges with a water

cooling jacket to protect the arrays. Room temperature water circulates through the jacket to

maintain the probe-wedge interface well below the critical temperature (<40oC) [15].

The final assembled system for high temperature inspection with a PBI wedge is seen in Figure

3.2.

Page 36: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

22

Figure 3.1 PEI high temperature wedge schematic with sloped front and dampening material at the top of the wedge:

ray 1 (solid line) and ray 2 (dashed line) illustrate reflection pattern of steep and shallow beam incidence at the

wedge bottom. As a result of wedge elongation and angled front, the internally reflected beams reflect to the top of

the wedge where they are absorbed by the dampener.

Figure 3.2 High temperature phased array inspection system including: Eclipse WA10-HT55S-IH-B PBI wedge

prototype based on the geometry of the Olympus linear phased array model 5L16-A10 with 16 elements and center

frequency of 5MHz, water cooling jacket, coolant tubing system and mounting arms.

Page 37: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

23

4 Temperature Distribution Model

The compression wave speed in an isotropic material is determined by its elastic modulus and its

density, all of which are temperature dependent [6]. Therefore, the temperature distribution must

be determined in order to characterize ultrasonic wave propagation in a wedge. The procedure

for obtaining this temperature distribution is described in this Chapter, followed by temperature-

dependent sound velocity measurements in the next Chapter.

The temperature distribution inside the wedge was modeled using the COMSOL finite element

package. A critical part of the model was to specify appropriate boundary conditions to represent

the various surfaces of the wedge.

The bottom of the wedge is heated when placed on the hot surface of a test piece, such as a

heated pipe. Therefore, a temperature-based boundary condition can be used to describe the

bottom of the wedge.

The array-wedge contact boundary is cooled by a circulating water jacket. Room temperature

water circulates through the channel with a minimum flow rate of 100 mL/min absorbing heat

from the boundary area. Laboratory measurements showed that the water maintained a

temperature of 25oC +/- 0.5

oC. Therefore, a temperature-based boundary condition was also

specified on this boundary of the wedge.

The remaining wedge surfaces are cooled by the surrounding air through natural convection.

Natural convection boundary conditions can be very challenging to model accurately due to the

complex nature of the cooling process and sensitivity to minor changes in the environmental

conditions. This involves the combination of heat transfer and laminar flow in a fluid (air) with

coupled temperature-velocity fields, accompanied by heat transfer within the solid. This leads to

density gradients in the air that cause buoyancy forces, which then lead to air flow [28].

To avoid the challenges described above, an alternative procedure is commonly used for

modeling natural convection in which an “average” heat transfer coefficient is specified on the

boundaries that interface with the surrounding fluid [29]. This heat transfer coefficient can be

difficult to specify; its value depends on several factors such as geometry and orientation of a

surface. In addition, the temperature of the surrounding fluid may not drop off uniformly to an

Page 38: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

24

ambient level far from the wedge. For example, a large heated pipe under inspection can greatly

distort the temperature distribution in the surrounding air and hence its flow pattern and cooling

capacity.

Empirical and theoretical correlations have been provided in the literature to estimate the heat

transfer coefficient for common geometries such as a vertical wall, inclined wall and horizontal

plate. Using such correlations, COMSOL provides built-in functions for estimating the average

convective heat transfer coefficient of a surface based on its geometry and the ambient

temperature [29].

These correlations were used to solve for the temperature distribution inside a wedge made of

PEI on a hot steel pipe at a steady state condition, where the steel pipe surface temperature, the

surrounding air temperature and cooling water temperature were 150oC (PEI’s maximum

operating temperature), 23oC and 25

oC respectively.

Model results can be seen in Figure 4.1. Only half the wedge was modeled due to symmetry.

Temperature values calculated on the wedge surface lines specified in Figure 4 were compared to

experimental data obtained with thermocouple measurements averaged over five trials. Results

of the comparison are shown in Figure 4.2. According to this figure, the maximum difference

between the model and experiment results is not larger than 5oC. This discrepancy will have a

negligible effect on estimates of compression wave velocity (see Chapter 5).

Page 39: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

25

Figure 4.1 PEI wedge temperature distribution modeling result, the color palate represents the temperature

distribution inside the wedge between 25oC and 150

oC. Surface temperature measurements located on the dashed

lines were used to validate the model results.

Figure 4.2 Comparison of COMSOL and experimental results for the temperature distribution on the surface of the

PEI wedge mounted on a 150oC steel pipe. Solid lines on the graph and the error bars represent average temperature

values of 5 experiments and standard deviations of the measurements: the red circles represent COMSOL model

results at the specified locations on the wedge surface.

Page 40: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

26

5 Velocity Measurements at Elevated Temperatures

The compression wave velocities of both PEI and PBI were measured experimentally from room

temperature up to their maximum operating temperature. The experimental procedure and results

are described in this Chapter. These data will then be combined with the calculated temperature

distribution inside a wedge to allow determination of beam paths and refraction angles.

In attenuative materials, ultrasonic waves are dispersive in nature, meaning that the phase

velocity and group velocity are frequency dependent [6] [30]. This has the effect of distorting the

wavefront, such that highly localized pulses become increasingly spread out as the pulse

propagates. Plastics are well-known to be dispersive media due to their viscoelastic nature [31]

[32].

Phase velocities can be obtained from two successive backwall echoes in a sample with two

parallel faces. The roundtrip travel time as a function of frequency is obtained from a comparison

of the phase spectra (f) of the two successive backwall echoes [33] [34] [35]. An example of

these calculations is illustrated in Figure 5.1 for two successive backwall signals (labeled first

backwall and second backwall) of a sample of thickness The governing equations are shown

below where and represent the time difference of the two backwall echoes and

the associated phase velocity respectively.

represents the time difference between two signal acquisition windows, as illustrated in Figure

5.1.

Page 41: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

27

Figure 5.1 Phase velocity measurement from two successive backwall echoes. The phase of each pulse is

determined relative to left side of the pulse acquisition window.

The phase velocities of PEI and PBI plastics were measured at room temperature using 10 mm

thick sample blocks with smooth parallel surfaces. A 10 MHz Panametrics5 highly-damped

ultrasonic contact probe was used with a useful frequency band of 3-11 MHz. Two successive

backwall echoes of the plastic blocks were obtained in the contact pulse-echo mode.

After the room temperature measurements, the procedure was modified to accommodate

measurements at elevated temperatures. The plastic test blocks were located on a temperature-

controlled hot plate. A glass delay line was placed between the ultrasound probe and the block

surface to avoid direct exposure of the piezoelectric elements to temperatures higher than 50oC,

the recommended operational temperature limit of the transducer [36]. The sample and the delay

line were then wrapped in high temperature insulation to minimize the temperature gradient

inside the plastic block. Two 0.001’’ diameter thermocouples were placed on either side of the

plastic block to measure the surface temperatures. The mean of these two temperature values was

used for the calculations, assuming an approximately linear dependence of phase velocity on

temperature within a narrow temperature range. The experimental set up can be seen in Figure

5.2.

5 Olympus, Waltham, Massachusetts, USA.

Page 42: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

28

Measurements were performed at 10oC increments at mean temperatures ranging from 30

oC up

to 120oC for the block made of PEI, and up to 300

oC for the PBI block. Each test was repeated 5

times; results are shown in Figure 5.3 for selected temperatures.

Figure 5.4 shows the temperature dependence of the phase velocity at 5 MHz which is the center

frequency of the phased array probe used with the high-temperature wedge prototypes. It can be

seen that the phase velocity decreases in both high-temperature wedge materials as temperature

increases. The data were fit to a linear function shown in Figure 5.4-a and Figure 5.4-b for PEI

and PBI, respectively. According to these Figures, the maximum difference of 5oC between the

COMSOL temperature distribution model and experiment results presented in the previous

Chapter can lead to less than 0.4% error in estimation of compression wave velocity.

Figure 5.2 Experimental set up for measuring phase velocity at high temperatures. The sample and the delay line

were wrapped in high temperature insulation to lessen the temperature gradient inside the plastic block. The mean

block temperature was estimated from thermocouples placed on top and bottom of the plastic block.

Page 43: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

29

Figure 5.3 Phase velocity data of PEI (a) and PBI (b) plastic blocks at selected elevated temperatures. Solid lines and

the error bars represent the mean and standard deviation of 5 repeated experimental measurements.

Page 44: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

30

Figure 5.4 Phase velocity of PEI (a) and PBI (b) as a function of temperature at 5 MHz frequency. The empirical

relations shown on the graphs indicate the functional change of phase velocity results with temperature T expressed

in degrees Centigrade.

Page 45: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

31

6 Focal Law Algorithm – Planar Waves

Thermal gradients inside a high temperature wedge lead to variations in the temperature-

dependent wave velocity and skewing of the direction of ultrasonic wave propagation; such

conditions invalidate conventional calculation of relative delay times on individual elements of a

phased array that are based on a homogenous propagation medium.

In this project we use a numerical ray-tracing technique to approximate the arced path of waves

propagating across thermal gradients in an isotropic wedge. The results are then used in a

separate algorithm to modify the phased array focal law; this will yield the required delays for

individual array elements to generate a plane wave or a focused beam, while compensating for

thermal gradient effects inside the wedge.

The numerical ray-tracing technique and modified focal law algorithm are described in this

chapter.

6.1 Room Temperature Inspection

Ray tracing is an established technique to estimate the wave path as it propagates in a medium

based on tracking a point on the wavefront rather than the complete wave field. This is achieved

based on the mechanical properties of each region and interface through which the wave passes.

Application of ray tracing techniques for calculations of wave paths through a medium with non-

uniform propagation velocity has been reported by several researchers. In seismology, for

example, a ray tracing technique is used to record ground motion caused by the passage of

seismic waves [37] [38].

Application of ray-tracing techniques to the field of NDT has also been reported by several

researchers [39] [40], e.g., Nowers et al who introduced ray-tracing algorithms for the inspection

of anisotropic weld sections using ultrasonic arrays [41].

As described in Chapter 2, a scan plan is normally prepared before any industrial phased array

inspection is initiated. The scan plan ensures total inspection coverage of the region of interest in

a test piece, such as a weld section. This is achieved by an optimized selection of scanning

parameters such as wedge geometry and range of scanning angles. Planar waves are typically

used to locate any flaws and determine their approximate size quickly; plane wave generation is

Page 46: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

32

therefore a very common use of phased arrays. Once discontinuities have been located using the

plane waves, a focused beam may be used at the defect locations for more accurate sizing and

flaw characterization.

For a planar wave, the particles lying on any plane that is perpendicular to the wave vector are all

vibrating in phase with each other and along parallel paths. For a plane wave to be generated by

a phased array, the acoustic fields generated by all the individual array elements should interfere

constructively to produce uniform acoustic pressure along each plane that is normal to the

direction of propagation. (This is mathematically possible only for a single two-dimensional

transducer of infinite area. However, the effect can be approximated over a small wavefront

using a linear array with a finite number of elements). To generate this pseudo plane wave with a

specified direction of propagation, the array elements must be excited with appropriate relative

time delays.

Consider first the case of generating a shear plane wave in a steel test block in contact with an

array-wedge system at a uniform room temperature of 25oC (Figure 6.1). This Figure illustrates

an Eclipse WA12-HT55S-IH-G PEI wedge designed to mate with a linear phased array with 64

elements and center frequency of 5 MHz (Olympus model 5L64-A12); this system is configured

to generate planar shear waves with nominal angle Фs=55o inside a steel test piece at 20

oC if all

phased array elements are fired simultaneously. Also shown in Figure 6.1 is an arbitrary

wavefront in the steel, on which all sample points are vibrating in phase with each other.

Page 47: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

33

Figure 6.1 Wave propagation pattern for a shear plane wave at angle Фs in a steel block at a uniform temperature of

25oC. The ray traces are applicable for a wave transmitted from the array into the test piece, or equivalently for a

wave travelling from the test piece back to the array. In the latter case, the waves associated with sample points

along an arbitrary plane wave front propagate in a direction perpendicular to the wave front and reach the piece-

wedge interface at locations labeled as interface points. Then they refract into the wedge based on Snell’s law and

propagate through the wedge to the array-wedge interface at points labeled as array-line points.

To calculate the appropriate element time delays for generation of such plane waves, the travel

time of the wave emitted from each element to the specified wave front must be obtained. To

achieve this, wave propagation paths from various “array-line points” spaced at intervals along

the phased array to the plane wave front are first selected, as shown schematically by ray traces

in Figure 6.1.

It will be convenient to conduct the calculation of the focal law in the reverse direction, i.e.,

model the waves originating from a few “sample points” on the selected plane wave front in the

test piece, and traveling back along the ray traces to their respective “interface points”, and then

to the “array line points”. This is shown schematically by the arrow directions on the ray traces

of Figure 6.1.

Page 48: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

34

The propagation paths are straight lines inside each homogenous isotropic medium, i.e., the

wedge and the test piece, each with a single wave propagation velocity corresponding to 25oC.

The waves propagate in a direction perpendicular to the wave front and reach the piece-wedge

interface (at locations labeled interface points in Figure 6.1). At the interface they refract into the

wedge according to Snell’s law and then propagate through the wedge to the array-wedge

interface (array-line points). Note that the same propagation paths are applicable in both

directions, i.e., from array to plane wave front, and from the plane wave front back to the array.

The lengths of each shown wave propagation path can be calculated geometrically in both the

test piece and the wedge sections. Given the ultrasonic shear wave velocity in the steel and

compression wave speed of the wedge, the travel times along each propagation path in the two

materials can be calculated. Finally, the travel time inside the piece and the wedge are added

together to find the total travel time along each propagation path. Note that these travel times and

propagation paths also apply in the opposite direction as well, for waves originating from the

array and traveling into the steel test piece.

The total travel time will vary linearly along the array-line points and can be interpolated to yield

travel times as a continuous function of position along the array-wedge boundary. The function

can be used to find the travel time associated with each element on the array. These yield the

required relative time delays for excitation of each element to ensure that all points on the

wavefront in the steel are in phase with each other. The same time delays are also applicable to

the summing of received signals at the array elements, originating from a planar wave traveling

through the test piece and impinging on the transducer assembly.

This simple procedure can be quickly applied to construct a conventional focal law calculator for

room temperature inspections. However, for inspections of test pieces at elevated temperatures

with high-temperature wedges and cooling jacket around the array, this simple focal law

calculator will no longer work. The presence of thermal gradients inside the wedge leads to

variations in temperature-dependent wave velocity and skewing of the wave directions inside the

wedge; this invalidates the conventional calculation procedure. Account must also be taken of

the change in wave velocity associated with elevated temperature in the test piece. These issues

are explored in the next Section.

Page 49: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

35

6.2 Elevated Temperature Inspection

The beam skew pattern caused by thermal gradients inside wedges mounted on test pieces at

elevated temperature can be approximated utilizing ray-tracing techniques. However, the

problem is now more complex. We use a discretized analytical ray-tracing technique utilizing

temperature distribution model data and phase velocity measurements of the wedge materials at

elevated temperatures.

Ultrasonic waves traveling in a two-dimensional isotropic medium with smoothly varying

mechanical properties are described by a system of first-order ordinary differential equations. To

simplify the mathematics, we track the progression of a single point on the wave front, with

propagation vector in the x-y plane. The wave front moves at the local compression wave speed

V(x,y) at an angle with respect to the x-axis [42] [43].

Combining equations (6.1) and (6.2) yields:

The above equations can be integrated numerically over fine time steps using Matlab [44], from

any starting point {xo,yo} and initial wave trajectory inside an ultrasonic wedge. This then yields

the propagation path and travel time. The velocity profile V(x,y) is based on the temperature

profile as calculated by the finite element program COMSOL. Note that the parameter

represents the amount of skew in the beam per unit time.

The results of these wave propagation simulations, performed for a specific series of wave

starting points and trajectories, are then fed into a separate algorithm to develop a new phased

array focal law – analogous to the simulation described in Section 6.1 for an inspection

performed at room temperature. This in turn will yield the required time delays for individual

Page 50: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

36

array elements to generate any desired beam profile inside a high temperature wedge and test

piece, even in the presence of strong thermal gradients in the wedge.

Figure 6.2 illustrates the wave paths for generation of the same shear plane wave depicted in

Figure 6.1 when the steel block is at a uniform temperature of 150oC and the wedge is cooled by

ambient air at 25oC plus the array’s water jacket. This arrangement will cause the wedge

temperature to range from close to 150oC where it contacts the test piece, down to approximately

30oC adjacent to the cooling jacket. The same sample points on the plane wave front in the test

piece shown in Figure 6.1 are again selected here to demonstrate the focal law calculation, using

the same ray tracing procedure used in Section 6.1. The following differences between Figure

6.1 and Figure 6.2 should be noted:

1) Wave path lengths in the steel test piece are the same in both Figure 6.1 and Figure 6.2,

given that the test piece is at a uniform temperature in both Figures; however, the

associated wave velocities and travel times in the test piece are now changed due to its

elevated temperature of 150oC.

2) The angle Фs of the waves at the wedge-piece interface is the same in both Figures.

However, the associated angles of the waves inside the wedge will be different; this is a

result of Snell’s Law and variations in the temperature-dependent sound velocities in both

the test piece and bottom of the wedge.

3) Once the waves refract into the wedge, their paths will no longer be straight lines, as the

thermal gradients inside the wedge lead to skewing of the waves. The wave paths and

their travel times inside the wedge are simulated with the ray-tracing algorithm of

Equations 6.1 to 6.3, utilizing temperature distribution data from our COMSOL finite

element model, and phase velocity data of the wedge material at elevated temperatures.

The solid red lines in Figure 6.2 represent the skewed paths followed by rays originating

from the sample points.

It is noted that the amount of skew is not equal for all the waves. The reason is that the

amount of skew depends on the amount of misalignment between the propagation vector

(at angle in equation 6.3) and the gradient of the temperature or velocity field (gradient

=

.

Page 51: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

37

4) Once the new travel times of the waves have been obtained in both the wedge and test

piece, the total travel time and the appropriate delay times are calculated as described

before in Section 6.1.

Figure 6.2 Wave propagation pattern for a shear plane wave at angle Фs in a steel block at a uniform temperature of

150oC. The ray traces are applicable for a wave transmitted from the array into the test piece, or equivalently for a

wave traveling from the test piece back to the array. In the latter case, the waves associated with sample points along

an arbitrary plane wave front propagate in a direction perpendicular to the wave front and reach the piece-wedge

interface at locations labeled as interface points. Then they refract into the wedge based on Snell’s law and

propagate through the wedge to the array-wedge interface at points labeled as array-line points. The waves

propagate along arced paths (red lines) in the wedge due to thermal gradient induced velocity variation.

The model results for the generation of the 60o shear plane wave are illustrated in Figure 6.3 for

the Olympus linear phased array model 5L64-A12 with 64 elements and center frequency of 5

MHz. Typically, only a subset of 16 elements on the array are selected to be used at one time.

Figure 6.3-a and Figure 6.3-b illustrate the calculated delays vs. their corresponding element

numbers when the first 16 elements (elements 1 to 16) and last 16 elements (elements 49 to 64)

of the array are used. The blue and red data points represent the results for room (25oC) and

elevated temperature (150oC) inspection conditions respectively. Delay times are expressed

relative to the first element of the 16 elements used to form the plane wave. The reason is that

delay times are usually expressed relative to the element associated with the longest travel time

Page 52: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

38

from array to the plane wave front inside the test piece. This element is the first element of the

aperture for the case of generation of a 60o shear plane wave in a test piece. This element should

be fired first at a time defined as t=0; all other elements are then fired at their respective relative

delay times.

For the simulation results shown in Figure 6.3-a, application of conventional room-temperature

focal laws for inspections of steel test pieces at 150oC lead to incorrect time delays on array

elements of up to 100 ns; this is approximately one half a period of the pulse central frequency of

5 MHz (period of 200 ns) in the test piece. The associated phase delay difference of up to 50% of

the wave period would lead to destructive interference, major distortion of the desired beam

profile, and consequently poor imaging resolution. The corresponding timing errors are much

lower for the pulse generated by firing array elements 49-64; Figure 6.3-b model results show a

minor difference of up to 10 ns (5% of the wave period) between the delay times generated by

the conventional focal law (designed for room temperature) and our new algorithm for elevated

temperature.

Figure 6.3 results illustrate how the effects of thermal gradients on the optimal element delay

times have a complicated dependence on several test parameters, such as test temperature, wedge

geometry & thermal gradients, locations of array elements, etc. This dependence is explored in

the following paragraphs.

For the example results depicted in Figure 6.3 (inspection of the test piece at 150oC), the wedge

temperature varies from close to 150oC where it contacts the test piece, down to approximately

30oC on the wedge-array interface adjacent to the cooling jacket. Therefore, the emitted waves

from all elements propagate through a similar temperature drop of about 120oC when

propagating through the wedge to the test piece. However, the path length and steepness of the

gradient varies based on the location of each transmitting array element. This variation is

considerably larger if the first 16 elements of the array are used, as opposed to the last 16

(elements 49-64). That variation over the 16 elements contributes to a non-linear delay pattern in

the timing of array element excitation, compared to the linear pattern obtained by room

temperature delays (Figure 6.3-a). The temperature profile inside the wedge associated with the

described inspection condition and the locations of elements 1-16 and 49-64 are depicted in

Page 53: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

39

Figure 6.4. The color palate shows the temperature distribution inside the wedge according to the

COMSOL finite element results.

Figure 6.3 Calculated relative element time delays for the generation of a 60o shear plane wave for Eclipse WA12-

HT55S-IH-G PEI wedge and the Olympus linear phased array model 5L64-A12 one a) the first 16 elements of the

array (elements 1-16) were used and b) the last 16 elements of the array (elements 49-64) were used. The blue and

red data points represent the results for room (25oC) and elevated temperature (150

oC) inspection condition

respectively and the delays are expressed relative to the first element of the used active aperture.

Page 54: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

40

Figure 6.4 Temperature profile inside an Eclipse WA12-HT55S-IH-G PEI wedge for inspection of test piece at

150oC, the color palate represents the temperature distribution inside the wedge between 25

oC and 150

oC as

obtained by our COMSOL finite element model. Waves emitted from elements 1-16 propagate along the shortest

path length in the wedge with a relatively steep temperature gradient, compared to the waves emitted from elements

49-64.

Beam skewing is caused by ultrasonic wave propagation across a thermal gradient. The angle

between the thermal gradient and the ultrasound propagation vector varies strongly with the

choice of array elements used (Figure 6.4). This angle (and resultant beam skew) is generally

much larger if the upper array elements are used (e.g., 49-64) as opposed to the lower array

elements. This angle also tends to be larger if the system is set to generate a plane wave in the

test piece with a large refracted angle s (e.g., 65 degrees) as opposed to a small refracted angle.

These various effects combine in a non-linear manner to determine the total correction to the

optimal array element delay sequence for a high temperature inspection. In some cases, as shown

in Figure 6.3-b, the various effects may largely cancel out such that the optimal delay pattern is

similar to that for a room temperature inspection. However, it would be expected that

Page 55: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

41

modifications to the focal law would become increasingly important if inspections were carried

out at even higher temperatures than those shown in Figure 6.3, e.g., 350oC.

To confirm these observations, the model results for the generation of a 45o shear plane wave are

illustrated in Figure 6.5 for the same array-wedge configuration used in Figure 6.2, and using the

same interface points on the piece-wedge interface as shown in Figure 6.2. Sample points are

selected on an arbitrary 45o plane wavefront in the test piece based on straight line propagation

paths inside the piece at a uniform temperature.

Figure 6.5-a and Figure 6.5-b illustrate the relative element delay times calculated by our new

algorithm when the first 16 elements (elements 1 to 16) and last 16 elements (elements 49 to 64)

of the array are used. The blue and red data points represent the results for room (25oC) and

elevated temperature (150oC) of the test piece respectively. Note that in the case of generation of

a 45o shear plane wave, delay times are expressed relative to the last element of the 16 elements

used to form the plane wave. In this case, it takes the longest time for the wave emitted from the

last element of the aperture to travel from the array to the plane wave front selected for

calculation. Therefore, this element should be fired first at a time defined as t=0; all other

elements are then fired at their respective relative delay times with respect to this reference

element.

Model results depicted in Figure 6.5 demonstrate similar trends to those seen in Figure 6.3, and

show the necessity of our new high-temperature array focal law. For the case of using elements

1-16 of the array (Figure 6.5-a) the required correction is up to 120 ns – over half a cycle period

in the 5 MHz burst. However, once again the magnitude of the correction is far smaller if the last

16 elements of the array are used (Figure 6.5-b) – a maximum of 25 ns.

Comparison of the results illustrated in Figure 6.3-a and Figure 6.5-a demonstrates how the

distortional effect of the temperature gradient depends on the selected shear plane wave angle Фs

in the test piece. This dependency on Фs is complicated as the following effects must be

considered:

1) Path lengths vary linearly across the width of the plane wave front in the steel test piece;

this variation, and the associated travel times along each ray depend on Фs.

2) A change to Фs leads to an associated change in the trajectories of the waves inside the

wedge through Snell’s Law. This leads to a complete change of the path of each ray

Page 56: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

42

through the wedge, as determined by our ray tracing algorithm of Equations 6.1 to 6.3.

This leads to a change in the misalignment between each propagation vector and the

gradient of the temperature field (and associated velocity field). This in turn causes a

change in the amount of beam skew.

All of these changes, precipitated by a decision to perform an inspection at a different beam

angle Фs, lead to changes in beam travel times in both the wedge and test piece. Combining all

these effects, our wave propagation algorithm determines the appropriate delay time for each

array element to yield a pseudo-plane wave in the test piece at the desired Фs.

An experiment has been designed to validate our algorithm for determining appropriate element

time delays on a heated wedge. The results obtained from our algorithm for generation of shear

plane waves with 45o and 60

o refraction angles inside a steel test piece at 150

oC were used to

experimentally evaluate the generation of a plane wave in the test piece. The details of the

experiments and results are described in the following chapter.

Page 57: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

43

Figure 6.5 Calculated element time delays for the generation of a 45o shear plane wave for an Eclipse WA12-

HT55S-IH-G PEI wedge and the Olympus linear phased array model 5L64-A12 one a) the first 16 elements of the

array (elements 1-16) were used and b) the last 16 elements of the array (elements 49-64) were used. The blue and

red data points represent the results for room (25oC) and elevated temperature (150

oC) inspection condition

respectively and the delays are expressed relative to the last element of the used active aperture.

Page 58: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

44

7 Experimental Validation – Planar Waves

In the previous chapter, we described an algorithm to calculate a new array focal law for plane

wave inspections of test pieces at elevated temperatures using high temperature wedges. The

algorithm was based on a numerical technique to determine the arced path of ultrasonic waves

propagating across thermal gradients in an isotropic wedge; this yielded the required delays for

individual array elements to generate a plane wave in a hot test piece.

In order to validate the algorithm, we will now use the results for element time delays in the

examples described in chapter 6 to experimentally evaluate the quality of shear plane wave

generation in a test piece at 150oC. First, we describe the overall concept of how experimental

validation will be accomplished. This is then followed by a detailed description of the

experiment itself. These are followed by results and analysis.

7.1 Concept for Experimental Validation

Consider the generation of a shear plane wave at an angle Фs in a steel test block, as shown in

Figure 7.1. Note that the right side of the block is also cut at the same angle Фs as the desired

angle of wave propagation. The array element delay times calculated in Chapter 6 are now

applied to the individual array elements, using a phased array instrument to provide the

excitation pulses.

Once the array elements are fired with these calculated relative time delays, their individual

acoustic fields interfere with each other to create an acoustic field which propagates through the

wedge. The exact shape of the field depends on both the individual element delay times and the

thermal gradients. However, at each instant of time, a surface can be found along which all the

particles are vibrating in-phase with each other; such a surface is called a wave front (Figure

7.1).

It should be noted that this wavefront is not straight inside the wedge, due to the non-uniformity

of the medium; the wavefront shape undergoes continuous change as it propagates through the

wedge due to the temperature dependence of the wave speed. However, if the array element

times have been correctly implemented, the wave front should turn into a planar surface when

the wave refracts into the steel test piece at the wedge-piece interface. The particles which are

Page 59: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

45

vibrating in phase with each other along any selected wavefront in the steel will then lie on

planes perpendicular to the propagation vector (arrows shown in Figure 7.1).

Figure 7.1 propagation of a wavefornt corresponding to a shear plane wave at an angle Фs in a steel test block at

elevated temperature: the wavefront is not straight inside the wedge due to non-uniformity of the medium. However,

it turns into a planar surface when the wave refracts into the steel test piece at the wedge-piece interface.

From the geometry shown in Figure 7.1, the shear plane wave will strike the slanted right hand

surface of the test block perpendicularly (wave vector perpendicular to this block boundary).

Therefore, the plane wave reflects straight back towards the wedge-piece interface on the same

propagation path, with an 180o change in phase caused by reflection at a free surface. The

reflected wave should then refract into the wedge with the same incident-refraction angle system

as the original transmitted wave. i.e., the reflected wave should retrace its exact path back into

the wedge and back to the individual elements of the array. This implies that the echo signals

should be received back at the individual elements with the same relative time delays as those

applied in transmission mode, provided that an appropriate focal law calculator was used to

generate the original wave. Measurement of the relative echo delays at the array elements can

therefore be used as one criterion to evaluate our algorithm as a focal law calculator. Details are

given in the following Section.

Page 60: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

46

7.2 Experimental Details

Two angled steel test blocks of the general shape shown in Figure 7.1 were prepared with slanted

face angles Фs of 45o and 60

o, to be used in two separate algorithm validation experiments. An

Eclipse WA12-HT55S-IH-G PEI wedge was used along with a 5L64-A12 Olympus linear array

for wave generation, matching the simulation parameters described in the two cases presented in

Chapter 6. The array was connected to a Focus LT6 phased array instrument which can apply

user-specified relative element time delays to the selected active elements of the array using

Tomoview6 software. The wedge-array system was placed sequentially on the two test blocks for

generation of the desired plane waves. The following experiment was performed individually for

each block.

The relative element time delays determined in Chapter 6 were applied to the selected active

array elements (transmitting delays). Tests were performed using the first 16 elements (element

numbers 1 to 16), and then the last 16 elements (element numbers 49 to 64) of the array; the

orientation of the array element numbering system is shown in Figure 7.1. After transmitting

pulses from the selected 16 elements, the system was then switched to receiving mode, and the

echo signals were captured at the same 16 elements as used in transmission. The relative time

delays among the received echo signals at the 16 array elements are measured and compared to

the time delays applied to their associated elements in the transmitting mode.

Theoretically, if the plane wave has been generated properly in the test piece using the element

time delays calculated by our new focal law calculator, then the measured receiving delays

should be equal to the applied transmitting delays. For example, the array element that was fired

first should be the last one to receive an echo pulse. However, this experimental result may not

exactly match the model predictions due to possible inaccuracies in the new theoretical model

and/or the experiments. In order to investigate the validity of our model, it is important to

identify the sources of experimental and modeling errors, and quantify their effect on this series

of tests.

6 Olympus, Quebec City, Canada

Page 61: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

47

To this end, we first perform all of the experiments with the test piece and wedge at room

temperature, using the conventional focal laws described in Section 6.1. The associated sources

of errors for this relatively simple case are identified, and their magnitudes estimated as

accurately as possible in Section 7.3. The experiments are then repeated with the wedge on a hot

steel test piece at 150oC; the wedge is allowed to reach an equilibrium temperature distribution

(as estimated by the COMSOL finite element model) before any measurements are taken.

Estimates are made of the additional sources of error introduced to the experiments by the

elevated temperature of the test piece. The experimental results are then used to evaluate the

validity of our algorithm as a focal law calculator; deviations in our experimental results from

those predicted by our new algorithm are compared to the various sources of error. This analysis

provides one measure of the validity of our new model, and the possible consequences of

ignoring the effects of elevated temperatures when using phased arrays for inspections at

elevated temperatures. Details are given in the following sections.

7.3 Experimental Results and Analysis - Room Temperature

Figure 7.2 shows the results of a set of experiments performed for the case of generating a 45o

shear plane wave in the 45o-angled steel test block at a uniform temperature of 25

oC, using the

first 16 elements of the linear array. The blue data points represent the relative time delays for

excitation of each element in transmission mode, as determined in Chapter 6 using our

conventional focal law calculating algorithm (Section 6.1). The delays are specified relative to

the excitation time of element #16. By convention, a positive delay time for an individual

element indicates that it is pulsed after the excitation of reference element #16. Elements #1-15

all have positive delay times in Figure 7.2; this implies that element #16 was pulsed first.

The solid red line in Figure 7.2 represents the relative time delays of the echo pulses received at

the same 16 individual array elements (receiving delays). Each point on this line is the average of

10 repeated experiments. The error bars indicate the random fluctuations associated with

measured data of each individual element as described by the standard deviation of the

measurements.

Page 62: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

48

If the desired plane wave has been properly generated in the angled block, the measured echo

pulse delays corresponding to each array element should be equal to the applied transmitting

delays as calculated in Chapter 6. However, slight deviations are seen between the experimental

results and their associated theoretical expected values in Figure 7.2. These deviation times are

depicted on a more detailed scale in Figure 7.3. Note that because element #16 is used as the

reference for timing purposes, its delay times in both transmission of the outgoing pulse, and

reception of the echo pulse, are zero by definition. Therefore its deviation between theory and

experiment is always zero.

It is important to understand the sources of the deviations shown in Figure 7.3, and quantify the

overall impact of these sources of error on the objective of generating a planar shear wave in a

hot test specimen. We find it convenient to categorize the deviations illustrated in Figure 7.3 into

two independent groups: systematic errors and random errors. Such a categorization represents a

slight simplification of the true situation, but makes the analysis more tractable.

Figure 7.2 Experimental results for the case of generating a 45o shear plane wave in the 45

o-angled steel test block at

a uniform temperature of 25oC using the first 16 elements of the array (elements 1-16): the blue data points represent

transmitting time delays applied to the elements in transmission mode as calculated by our model. The solid red line

represents the mean of 10 measurements of receiving delays at the same individual array elements. The error bars

indicate the standard deviation of the measurements at each individual element.

Page 63: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

49

Figure 7.3 Deviation times between the measured received echo time delays on each individual element and their

associated theoretical value as calculated by our model for the case of generating a 45o shear plane wave in the 45

o-

angled steel test block at a uniform temperature of 25oC using the first 16 elements (data illustrated in Figure 7.2).

The solid line and error bars are associated with the mean and standard deviation of the measurements on each

individual array element. Note the very fine time scale on the vertical axis.

7.3.1 Systematic Errors

Systematic errors are biases in measurements, which are seen each time the same experimental

measurement is repeated. Therefore, their magnitude cannot be reduced by repeating a

measurement or the entire experiment and using the average result. Such errors result in a

deviation between the mean of very many separate repeated measurements of a parameter and its

theoretically expected value. Therefore, the solid line in Figure 7.3 can be used to provide a

rough estimate for the magnitude of the systematic errors in the above described experiment. The

magnitude of the systematic errors varies for results associated with different elements, and

reaches a maximum of approximately 18 ns at element #1 (Figure 7.3). These systematic errors

could arise from repeatable imperfections in the performance of the equipment (e.g., calibration

issues), geometrical errors in the shape of the block, and uncertainties in the value of the inputs

used for focal law model calculations, etc. As expected, there is a general trend for the

magnitude of the bias error to increase as one moves away from the reference element #16,

where the bias error is zero by definition.

Page 64: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

50

Inputs to the algorithm for the focal law calculation at room temperature include sound velocity

in both the wedge and the steel test piece, plus the wedge and block angles. The wedge and steel

blocks angles were measured by a digital angle gauge and the wave velocities were measured by

the procedure described in Chapter 4. The levels of uncertainty in these measured values are

listed in Table 7-1. In order to verify whether the uncertainties in these data values are consistent

with the magnitudes of bias errors seen in Figure 7.3, these data input values were individually

varied in our model by an amount equal to their estimated level of uncertainty. The resulting

effects on array element delay times are shown in Table 7-1.

Table 7-1 Inputs to the new focal law algorithm at room temperature: the levels of uncertainty in the measured

values of inputs are listed and the magnitude of their resulting effect on the bias error on the first element of the

aperture is presented.

Model Input Measured Value Measurement

Uncertainty

Associated Bias Error

on Element #1

Steel Block Angle 45o ± 0.1

o 8 ns

Wedge Angle 38o ± 0.1

o 12 ns

Sound Velocity in the Wedge 2460 m/s ± 10 m/s 15 ns

Sound Velocity in the Steel Block 3230 m/s ± 10 m/s 14 ns

The total interactive effect of these systematic errors on the bias errors seen in Figure 7.3 will

depend on each experimental set-up, such that it can’t be uniquely quantified. However,

experiments were conducted on several systems such as those shown in Figure 7.2 using a 64-

element 5-MHz linear array at room temperature, using various selections of 16 active elements

and two frequently-employed refraction angles Фs: 45o and 60

o. The results all had bias

deviations in the element time delays that were of the same order as those shown in Figure 7.3,

i.e., bias error ranging from zero for the reference element, up to a maximum value of the order

Page 65: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

51

of 20 ns for elements most distant from the reference element. These bias deviations are of the

same order of magnitude as the combined measurement uncertainties shown in Table 7-1.

Although these experiments were conducted at room temperature, the results can be used to help

estimate the magnitudes of bias error sources in experiments conducted at elevated temperature.

It is noted that the observed bias deviation of up to 20 ns represents only 10% of the wave period

for a 5 MHz transducer. That is, the phase difference variation among the wave vectors

associated with the 16 active array elements is of the order of +/- 18 degrees. This is consistent

with the requirement of a phased array that echoes received at the active elements should be

approximately in phase with each other, once the appropriate time delay is applied to each

element echo pulse. This then allows the 16 echoes to be added together constructively, to

achieve an enhanced signal-to-noise ratio compared to that seen individually at each element.

Last, we note the jagged nature of the solid line in Figure 7.3. This would not be due to any of

the bias error sources noted in Table 7-1, which would affect the deviations for all elements in a

smoothly varying manner. The jagged nature might be reduced somewhat by increasing the

number of measurements – this would further reduce any contribution of random errors, such as

noise that is affecting our estimate of the mean deviation at each element, as represented by the

solid line in Figure 7.3. However, the jagged nature of the line may also indicate that there are

additional systematic errors (beyond those listed in Table 7-1) that are individually dependent on

each element, e.g., imperfections in the individual array element alignments. In particular, it is

noted that the limited timing resolution of 10 ns of the Focus LT system for applying individual

delay times to each element, is of the same order of magnitude as the jagged oscillations in the

solid line of Figure 7.3.

7.3.2 Random Errors

Random errors are caused by unpredictable variations in the experiment, measuring system or

surrounding environment. The statistical distribution of random errors may be known

theoretically, or approximated through repeated measurement. In our case, the effects of these

random errors on the measured arrival times of echo pulses at each array element are indicated

by the error bars in Figure 7.2 and Figure 7.3. It is noted that in many cases, these “random”

Page 66: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

52

errors may not be completely random – there may actually be some pattern, bias, or deterministic

component to the fluctuations in experimental results that we label as “random”. In the case of

our experiments, three factors can be identified as the main sources of the random errors in the

results:

1) Instrument resolution; all instruments have finite precision that limits their measurement

resolution. For example, the Focus LT instrument applies a maximum sampling frequency of 100

MHz for digitizing the ultrasonic echo signals; this leads to a timing resolution of 10 ns in the

digitized echo signal of each array element, yielding a random temporal rounding error of the

order of ± 5 ns.

2) Instrument noise errors; noise is an unwanted disturbance caused by inherently unpredictable

electric fluctuations in the circuit of an electrical instrument, e.g., the effects of stray electrical

fields in the air or from the electrical power source.

3) Couplant effect; a couplant layer is required between the wedge and the test block to transmit

the ultrasonic energy efficiently through the interface. Ultrasonic waves generated by the

elements travel through the thickness of such a layer before entering the test piece. However, the

couplant thickness distribution between the wedge and the test piece varies from one experiment

to the next, and among various locations on the wedge/piece interface. These seemingly random

fluctuations are due to pressure-induced flow. These transients in couplant thickness lead to

changes in wave travel times through the interfacial layer, and random variation in the delay time

of echo arrival at each array element.

The magnitude of the combined effects of all these random errors can be estimated by repeating

the experiments and calculating the standard deviation of the obtained results. Note that the

magnitude of random error is zero on element #16 by definition as it is used as the reference for

calculations.

The random component in the timing deviations in the results of Figure 7.3 was also assessed for

all of the room temperature experiments, with different combinations of block angles and active

array elements. The standard deviation of the random fluctuations in our echo arrival times was

of the order of 15 ns (or less) for all test cases. This can be used as a benchmark for evaluating

the magnitude of random errors in the experiments performed at elevated temperatures.

Page 67: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

53

However, it is noted that some sources of random fluctuations, particularly those related to

couplant flow may be larger at elevated temperature. For laboratory experiments, the influence

of random errors can be minimized by performing each measurement many times and calculating

the mean result of many measurements, as done in Figure 7.3. However, this may not be

practical for field NDT inspections where time is precious, such that both random errors and

systematic errors will play a role.

7.4 Experimental Results and Analysis - Elevated Temperature

In the previous section, we evaluated our model results experimentally for the case of generating

shear plane waves in a steel test piece at room temperature (25oC). The possible sources of error

on delay times for individual array elements were described briefly and their maximum

magnitudes were estimated to be of the order of 20 ns (systematic errors) and 15 ns (random

errors).

These same experiments are now repeated at elevated temperatures for the much more

challenging case of the steel blocks at 150oC, using the new focal law algorithm to calculate the

optimal time delay for each array element. Experimental details are described in the following

paragraphs, followed by presentation and analysis of the results. It is noted that limitations in

experimental equipment prevented us from conducting these experiments at higher temperatures.

The same wedge-array system was used for the elevated temperature experiments as used for the

room temperature experiments. The two angled steel blocks were placed on a temperature-

controlled hot plate, and the block-wedge surface was kept at 150oC. The steel test pieces were

wrapped in high temperature insulation to minimize heat loss (Figure 7.4). This led to

temperature variations over the entire volume of each test piece of less than 10oC, as indicated by

experimental measurement at selected locations.

Page 68: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

54

Figure 7.4 Experimental set up for evaluation of model results for generation of shear plane waves in an angled steel

test block at 150oC. An Eclipse WA12-HT55S-IH-G PEI wedge was used along with an Olympus 5L64-A12 array

for plane wave generation with a cooling jacket around the array. The steel test block was located on a temperature-

controlled hot plate and wrapped in high temperature insulation which led to temperature variation over the entire

volume of the test piece of less than 10oC.

Once the temperature was stabilized in the test piece and wedge system, the relative element time

delays determined in Chapter 6 were applied to the selected active array elements (transmitting

delays). Tests were again performed using the first 16 element (element numbers 1 to 16) and

last 16 elements (element numbers 49 to 64) of the 5 MHz array. After transmitting pulses from

the selected 16 elements, the system immediately switched to receiving mode.

The received echo signals were then analyzed. Only a limited number of detailed analyses of

echo signals are presented here; equipment limitations forced an extensive amount of time and

data processing to be undertaken for echo analysis at each element. A particular emphasis was

therefore placed on only three elements of the active aperture: elements 1, 8 and 16 when the

first 16 elements of the array were used for beam generation, and elements 49, 54 and 64 when

Page 69: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

55

the last 16 elements of the array were active. However, the presented results and analysis are

sufficient to assess the overall performance of the system for all array elements.

The relative time delays of the received three echo signals were measured and compared to the

time delays applied to their associated elements in the transmitting mode. The experimental

results for the generation of the 45o shear plane wave at 150

oC are illustrated in Figure 7.5-a and

Figure 7.5-b when the first 16 elements and last 16 elements of the array were used, respectively.

The red data points represent the relative time delays of the elements calculated by our algorithm

for this elevated temperature inspection condition (Chapter 6). As for the case of the room

temperature experiments of Section 7.3, these delay times are expressed relative to the last

element of the selected aperture and were applied to the aperture elements in the transmitting

mode to form a plane wave with propagation angle Фs in the steel test piece.

The green data points on the graphs represent the experimentally measured time delays of the

received echo signals at the array elements that are being analyzed in detail. These points are

each the average of 15 repeated experiments and the error bars indicate the level of uncertainty in

calculated averages (green data points) due to the presence of random errors.

The dark green data point represents the average of the measured data corresponding to the first

element of the aperture and the light green data point represents the same for the middle element

of the aperture. These time delays were calculated relative to the last element of the active

aperture, which served as a reference as done for the room temperature measurements.

To illustrate the effects of temperature changes on phased array timing considerations, the

calculated time delays for room temperature (25oC) inspection used in Section 7.3 are also

shown in blue for comparison.

Page 70: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

56

Figure 7.5 Experimental results for generation of a 45o shear plane wave in a 45

o-angled steel test block at 150

oC

using an Olympus 5L64-A12 array and Eclipse WA12-HT55S-IH-G PEI wedge when the first 16 elements (a) and

last 16 elements (b) of the array were used. The red and green data points represent the relative time delays

calculated by our algorithm, and the mean of the measured echo time delays, respectively. The green points are each

the average of 15 repeated experiments and error bars indicate the uncertainty level in the value of the green points.

Blue data points represent the calculated time delays for room temperature inspection.

If our new algorithm for calculating time delays of individual array elements worked perfectly,

the two green dots in each of Figure 7.5-a and Figure 7.5-b would fall directly on top of the red

line. That is, the experimentally measured time delays of the echo signals would exactly match

the calculated time delays. Instead, we see some deviation, which can be analyzed in terms of

Page 71: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

57

systematic and random errors. Several possible sources of systematic and random sources were

described in Section 7.3 for the case of room temperature experiments. These, plus additional

error sources are present for the experiments conducted at elevated temperatures. These

additional error sources include:

Uncertainties in the temperature distribution in the wedges as estimated by the finite

element model (systematic error)

Uncertainties in the temperature dependence of compression wave velocity in the wedge

material, and the temperature dependence of shear waves in the steel test specimen

(systematic error)

Time-dependent fluctuations in the temperature distribution in the wedge, test piece, and

couplant (random error)

Our strategy is first to quantify the magnitude of the systematic and random errors. Then we can

analyze whether there is any significant deviation between our experimental results (e.g., green

data points shown in Figure 7.5) and the theoretical delay times predicted by our new algorithm

(red line).

Raw data values corresponding to results presented in Figure 7.5 are presented in Table 7-2, in

four separate columns. Each column represents the experimentally measured time delays of the

received echo signals at an individual element. The first two columns represent data

corresponding to the first and middle elements of the aperture, when the first 16 elements of the

array (elements 1-16) were used; these data are depicted by the dark and light green points,

respectively, in Figure 7.5-a. The third and fourth column of Table 7-2 represent the data

corresponding to the first and middle elements of the active aperture when the last 16 elements of

the array (elements 49-64) were used, as depicted by the dark and light green points,

respectively, in Figure 7.5-b. It is noted again that these delay times are all expressed relative to

the last element of the selected aperture, which is the first element fired in transmission mode

(Figure 7.5).

Each measurement was performed 15 times. The mean values of the 15 repeated experiments are

presented in the first row of Table 7-2. The associated standard deviations are shown in the

second row, and represent the magnitude of random fluctuation associated with each individual

test. From these data, the maximum magnitude of random errors in high temperature experiments

Page 72: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

58

is of the order of 24 ns. Note that this value is slightly larger than the 15 ns reported for room

temperature experiments. This increase is believed to be due to the influence of temperature

fluctuations in the wedge, couplant, or test piece, and thickness variations in the couplant.

The magnitudes of the uncertainty in the measured mean values are presented in the third row of

Table 7-2, as illustrated by the error bars in Figure 7.5. These are estimated from the standard

deviations (second row) and the number of measurements (N=15), based on statistical

distribution associated with the T-test with a 95% confidence interval [45].

The fourth row of the Table shows the time delays calculated by our model (red dots in Figure

7.5) for specific array elements. Finally, the deviations between the mean of the 15 repeated

measurements (row 1) and their associated theoretical values (row 4) are presented in the last

(fifth) row of the Table.

It is seen that in three of the presented cases, the deviation between the mean of the

measurements and its associated theoretical value is within the range of the uncertainty in the

experimental measurement’s mean value (comparison of data in the third and fifth row of the

Table). This means that in these cases there is no significant deviation between our experimental

results (e.g., green data points shown in Figure 7.5) and the theoretical delay times predicted by

our new algorithm (red data points in Figure 7.5).

For the case presented in the second column of Table 7-2, the magnitude of the deviation

between the mean and its associated theoretical value exceeds the range of the uncertainty in the

measurement’s mean (12 ns > ± 7 ns). This is the case associated with echo time delays of the

middle element of the aperture when the first 16 elements of the array were used. However, the

magnitude of deviation is within the systematic error range of up to +/-20 ns in the focal law

calculator, noted to be present in the room temperature tests due to uncertainties in input data.

Page 73: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

59

Table 7-2 Magnitude of the experimental results associated with generation of a 45o shear plane wave at 150

oC as

depicted in Figure 7.5. The results of 15 repeated time delay measurements of echo signals of the first and central

elements of the active aperture are presented for two cases: active aperture = elements 1-16; and active aperture =

elements 49-64.

Фs=45o

N=15

Tests Using Elements 1-16

(Figure 7.5-a)

Tests Using Elements 49-64

(Figure 7.5-b)

Element No. 1

(dark-green)

Element No. 8

(light-green)

Element No. 49

(dark-green)

Element No. 56

(light-green)

1 Experimental Mean

(green dots in Fig. 7.5)

475 ns 258 ns 395 ns 203 ns

2 Standard Deviation of

Individual

Measurements

20 ns 13 ns 24 ns 20 ns

3 Uncertainty in the

Mean (error bars)

± 11 ns ± 7 ns ± 13 ns ± 11 ns

4 Model Prediction

(Transmitting Delay)

479 ns 246 ns 383 ns 205 ns

5 Deviation of the Mean

from the Model

prediction

-4 ns +12 ns +12 ns -2 ns

The above results indicate that the experimentally measured echo delay times agree with our

calculated element time delays, within the range of experimental systematic and random errors

for generation of a 45o shear plane wave in a steel test piece at a uniform temperature of 150

oC.

Figure 7.6 is similar to Figure 7.5, but for the case of generation of a 60o shear plane wave in a

steel test piece at 150oC. Figure 7.6-a and Figure 7.6-b illustrate the results when the first 16 and

last 16 elements of the array were used, respectively. The red data points represent the relative

time delays of the elements calculated by our algorithm for this elevated temperature (150oC)

inspection condition, as described in Chapter 6. These delay times are expressed relative to the

first element of the selected aperture and were applied to the aperture elements in the

transmitting mode to form a plane wave with propagation angle Фs in the steel test piece.

Page 74: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

60

Continuing to follow the format of Figure 7.5, the green data points in Figure 7.6 represent the

experimentally measured time delays of the received echo signals at the array elements that are

being analyzed in detail. These points are each the average of 15 repeated experiments and the

error bars indicate the level of uncertainty in calculated averages (green data points) due to the

presence of random errors.

The dark green data point represents the average of the measured data corresponding to the first

element of the aperture and the light green data point represents the same for the middle element

of the aperture. These time delays were calculated relative to the first element of the aperture,

which served as a reference as done for the room temperature measurements. For comparison

purposes, the calculated time delays for room temperature (25oC) inspection used in Section 7.3

are also shown in blue.

Raw data values corresponding to results presented in Figure 7.6 are presented in Table 7-3 with

the same format as used for Table 7-2. The first two columns represent data corresponding to the

middle and last elements of the aperture, when the first 16 elements of the array (elements 1-16)

were used; these data are depicted by the dark and light green points, respectively, in Figure 7.6-

a. The third and fourth column of Table 7-3 represent the data corresponding to the middle and

last elements of the active aperture when the last 16 elements of the array (elements 49-64) were

used, as depicted by the dark and light green points, respectively, in Figure 7.6-b. It is noted that

these delay times are expressed relative to the first element of the selected aperture.

Page 75: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

61

Figure 7.6 Experimental results for generation of a 60o shear plane wave at 150

oC when the first 16 elements (a) and

last 16 elements (b) of the array were used. The red and green data points represent the relative time delays

calculated by our algorithm, and mean of the experimentally measured received echo time delays, respectively. The

green points are each the average of 15 repeated experiments; the associated error bars indicate the uncertainty level

in the value of the green points. Blue data points represent the calculated time delays for room temperature

inspection, presented for sake of comparison.

Page 76: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

62

Table 7-3 Experimental results associated with generation of a 60o shear plane wave at 150

oC as depicted in Figure

7.6. The results of 15 repeated time delay measurements of echo signals of selected elements of the active aperture

are listed in separate columns.

Фs=60o

N=15

Tests Using Elements 1-16

(Figure 6-a)

Tests Using Elements 49-64

(Figure 6-b)

Element No. 8

(dark-green)

Element No. 16

(light-green)

Element No. 56

(dark-green)

Element No. 64

(light-green)

1

Experimental Mean

(green dots in Fig. 7.6)

55 ns

109 ns

83 ns

175 ns

2

Standard

Deviation

20 ns

23 ns

13 ns

15 ns

3

Uncertainty in the

Mean (error bars)

± 11 ns

± 13 ns

± 7 ns

± 8 ns

4

Model Prediction

(Transmitting Delay)

49 ns

119 ns

99 ns

215 ns

5

Deviation of the mean

from the Model

prediction

+6 ns

-10 ns

-16 ns

-40 ns

The results presented in the first two columns of Table 7-3 indicate that there is no significant

statistical deviation between the experimental echo time delay measurements and the theoretical

delay times calculated by our algorithm when the first 16 elements of the array were used for

plane wave generation in a piece at 150oC. The deviation between the mean of the measured time

delays and their associated theoretical values are within the range of the estimated uncertainty.

When the last 16 elements of the array are used for plane wave generation, the deviation between

the mean of measured time delays and their associated theoretical values exceed the estimated

uncertainty. These cases are presented in the third and fourth column of Table 7-3. The deviation

is within the estimated systematic error of up to 20 ns for the results associated with the middle

element of the aperture (element number 56, as presented in the third column of Table 7-3).

However the deviation magnitude is about 40 ns for the measurements associated with the last

element of the aperture (element number 64, as presented in the fourth column). This is

Page 77: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

63

significantly larger than the maximum estimated systematic error magnitude of 20 ns observed

for room temperature experiments.

Larger systematic errors are expected for the case of high temperature experiments due to the

presence of additional bias error sources, i.e., errors in the convection cooling estimates for the

heated wedge; inaccuracies in the calculation of temperature distribution in the heated wedge;

errors in the temperature dependence of wave velocities. These additional error sources would

combine with the systematic errors of +/- 20 ns estimated for room temperature experiments, to

yield larger deviations between experimental and theoretical determinations of element timing

delays.

The experimental results indicate that the overall magnitude of the bias errors is dependent on

the shear plane wave angle and the selection of active elements. The largest deviation between

experimental and theoretical results occurs for the most challenging case of generating a shear

plane wave with a large refraction angle in a test piece. The deviation is also larger when the last

elements of the array are used for plane wave generation as compared to elements 1-16. Element

numbers 49-64 are the most distant from the wedge-piece interface, such that waves associated

with those elements undergo the longest path through the heated wedge. In addition, waves

associated with this last 16 elements traverse the thermal gradient at the widest angle, and

therefore experience far more beam skew, as described in Section 6.2.

In summary, the experimental results indicate that the new focal law calculator determines

element time delays for shear planar wave inspection of a heated test piece, that are not

inconsistent with experimental results. In the next Chapter we modify the focal law calculator to

calculate element time delays for generation of a beam focused at a single point in a hot piece.

Page 78: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

64

8 Focal Law Algorithm – Focused Beam

As described in Chapter 2, a scan plan is prepared before any industrial phased array inspection

is initiated. This is achieved by an optimized selection of scanning parameters such as wedge

geometry and range of scanning angles. Typically, planar waves are first used for the inspection

to locate any flaws and determine their approximate size. Once discontinuities have been located

using plane waves, a focused beam may be used at the defect locations for more accurate sizing

and flaw characterization. Focusing of the ultrasound beam at a particular location has the effect

of narrowing the beam diameter and increasing the local ultrasound energy per unit area. This

results in higher sensitivity to small reflectors, improved signal-to-noise ratio, and better image

resolution.

In Chapter 6 we introduced a numerical ray-tracing technique to model the arced path of waves

propagating across thermal gradients in an isotropic wedge. The results were then used in a

separate algorithm to modify the phased array focal law for generation of desired shear plane

waves in a hot piece, while compensating for thermal gradient effects inside the wedge. The

calculated relative time delays for individual array elements were presented for generation of

shear plane waves with two different angles in a steel test piece at 150oC. These were then used

to experimentally evaluate the quality of shear plane wave generation in a hot steel test piece.

The experimental results were presented and analyzed in Chapter 7.

In this chapter we introduce a second set of validation trials for our new focal law calculator, but

this time for the generation of a beam focused at a single point in the hot test piece. To this end,

we follow parallel steps to those described in Chapters 6 and 7 for the generation of a plane wave

in a hot test piece.

In Section 8.1, we develop the overall algorithm to calculate a focal law for generation of a

focused beam in a hot test piece, while compensating for thermal gradient effects inside the

wedge. In Section 8.2, the algorithm will be applied to a number of trial focal points in a

hypothetical hot test piece; the magnitude of the thermal gradient effects on the focal law

calculation will be quantified for those focal points and analyzed. In Section 8.3, experiments

will be carried out to validate the numerical model results of Section 8.2 for generating a focused

beam at any user-specified point in a hot test piece. Chapter 8 will then conclude with an analysis

of the results.

Page 79: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

65

8.1 Algorithm Details

Figure 8.1 illustrates the same inspection system as described in Chapter 6 (Figures 6-1 and 6-2)

with an Eclipse WA12-HT55S-IH-G PEI wedge and an Olympus linear phased array model

5L64-A12 with 64 elements and center frequency of 5 MHz. Also shown in Figure 8.1 is an

arbitrary focal point in a test piece at 25oC.

In order to calculate the appropriate element time delays for focusing the ultrasonic beam at the

specified focal point, the travel time of the wave emitted from each element to the focal point

must be obtained. To achieve this, the same steps and principles are followed as described in

Chapter 6. First, wave propagation paths from various “array-line points” along the phased array

to the focal point are determined, as shown schematically by ray traces in Figure 8.1. Also

labeled in Figure 8.1 are “interface points” on the wedge-piece interface; note that a single

interface point is applicable for each wave path between the focal point and an array-line point.

The propagation paths are straight lines inside each homogenous isotropic medium, i.e., the

wedge and the test piece, each with a single wave propagation velocity corresponding to 25oC.

The refraction angle at each Interface Point is determined by Snell’s Law. Note that the same

propagation paths are applicable in both directions, i.e., from array to focal point, and from focal

point back to the array.

Page 80: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

66

Figure 8.1 Wave propagation pattern for focusing an ultrasonic beam at an arbitrary focal point in a steel block at a

uniform temperature of 25oC. Wave propagation paths are shown from various array-line points along the phased

array to the selected focal point. The propagation paths are straight lines in both the wedge and the steel block, each

with a single wave propagation velocity at 25oC. The refraction angle at each interface point is determined by

Snell’s Law.

It will be convenient to conduct the calculation of the focal law in the “reverse” direction, i.e.,

model a pulse originating at the selected focal point and travelling along the various propagation

paths to the Interface points, and then on the Array Line Points. This is shown schematically by

the arrow directions on the ray traces of Figure 8.1.

The travel distances in both the test piece and the wedge sections can be calculated geometrically

along each of the shown propagation paths. These are divided by the ultrasonic shear wave

velocity in the steel and compression wave speed in the wedge to calculate the pulse travel times

along each propagation path in each of the two materials. The two travel-time components are

then added together to find the total travel time along each propagation path; these travel times

are each applicable to pulses travelling in either direction along that propagation path.

Knowing the total travel time associated with each of the array line points, those data can be

interpolated to yield travel time as a continuous function of position along the array-wedge

boundary. Specifically, one can interpolate the total travel time from the focal point associated

with each element on the array. These travel times then represent the relative element time delays

Page 81: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

67

required for excitation of the array to yield constructive interference of all waves arriving at the

specified focal point in the test piece: the same relative time delays apply to both transmitted

waves from array to focal point, and echo signals travelling from focal point to the array

elements.

This procedure for calculation of an array focal law becomes a bit more complex for inspection

of test pieces at elevated temperatures using a high temperature wedge with a cooling jacket

around the array. The presence of thermal gradients inside the wedge leads to variations in

temperature-dependent wave velocity and skewing of the wave directions inside the wedge. Our

intent is to utilize the ray-tracing technique introduced in Section 6.2 to model the beam skewing

effect in the generation of a focused beam.

Figure 8.2 illustrates schematically the wave paths for generation of a focused beam at the same

focal point as depicted in Figure 8.1, but now with the steel test block at a uniform temperature

of 150oC and the wedge cooled by ambient air at 25

oC plus the array’s water jacket. This

arrangement causes the wedge temperature to range from close to 150oC where it contacts the

test piece, down to approximately 30oC adjacent to the cooling jacket. The same interface points

on the test piece-wedge interface shown in Figure 8.1 are again selected here to illustrate the

focal law calculation, using the same ray tracing procedure as described for the ambient

temperature case moving from the focal points towards the array. The following differences

between Figure 8.1 and Figure 8.2 are noted:

1) Wave path lengths in the steel test piece are the same in Figure 8.1 and Figure 8.2, given

that the test piece is at a uniform temperature in both Figures; however, the associated

wave velocities and travel times in the test piece are now changed due to its elevated

temperature of 150oC.

2) The incident angles of the waves in the steel specimen at the wedge-piece interface are

the same in both Figures. However, the associated angles of the waves inside the wedge

will be different; this is a result of Snell’s Law and the temperature-dependent sound

velocities in both the test piece and bottom surface of the wedge.

3) Once the waves refract into the wedge, their paths will no longer be straight lines, as the

thermal gradients inside the wedge lead to skewing of the waves. The wave paths and

their travel times inside the wedge are simulated with the ray-tracing algorithm

introduced in Section 6.2, utilizing temperature distribution data from our COMSOL

Page 82: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

68

finite element model, and phase velocity data of the wedge material at elevated

temperatures. The solid red lines in Figure 8.2 represent the skewed ray paths originating

from the interface points.

4) Once the new travel times associated with each ray path have been obtained in both the

wedge and test piece, the total travel time and the appropriate relative delay times are

calculated as described in Chapter 6.

Figure 8.2 Wave propagation pattern for focusing an ultrasonic beam at an arbitrary focal point in a steel block at a

uniform temperature of 150oC. Wave propagation paths are shown from various array-line points along the phased

array to the selected focal point. The propagation paths are straight lines in the steel block based on the assumption

of uniform temperature inside the block (blue lines). After the waves refract into the wedge, they propagate along

arced paths (red lines) due to velocity variations across the heated wedge that are induced by thermal gradients.

In the next Section, we provide examples to demonstrate the magnitude of the thermal gradient

effects on the elements delay times calculated by the above described algorithms.

8.2 Magnitude of Thermal Gradient Effect

In order to quantify the magnitude of the thermal gradient effect on the focal law calculations,

we consider a case of focusing the ultrasonic beam at a series of three specific focal points in a

steel test block using the wedge-array system described in Section 8.1. The focal points lie along

Page 83: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

69

the sound paths representing the waves with refraction angles of 40o, 55

o, and 70

o inside the test

block when using elements #1-16 of the probe array, as shown in Figure 8.3. As is standard

practice in simplified training manuals for industrial phased array systems, the three test cases

are each represented by a straight line emitted from a point located at the center of the active

array region.

Figure 8.3 Location of the selected focal points and calculated near field points in a steel block on the sound paths

representing the waves with refraction angles of 40o, 55

o and 70

o using an Eclipse WA12-HT55S-IH-G PEI wedge-

5L64-A12 Olympus array system. The first 16 elements of the aperture are used for beam formation and each sound

path is represented by a straight line emitted from the point located at the center of the active aperture.

Theoretically, the desired focal points can be selected at any distance along the specified sound

paths. This means that focal laws can be calculated and applied to the elements for generation of

a focused beam at any distance along the sound paths. However, the focusing efficiency

decreases beyond the near field distance along each sound path. The near field distance is the

distance from the face of the transducer to the location with the last local maximum in the sound

field pressure, which then decreases monotonically due to beam spreading beyond this point. For

a wedge-array system, the near field distance depends on the size of the active aperture, the

center frequency of the probe, refraction angle (40o – 70

o), plus the ultrasonic sound velocity

inside the wedge and steel test specimen [14] [46]. The red points on Figure 8.3 represent the

location of the near field distance on each sound path.

Page 84: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

70

The black points shown in Figure 8.3 represent the location of focal points selected along each

sound path for characterizing the new focal law calculator. They are all located at the same travel

time distance from the center of the active array aperture, and within the near field. For the case

of the ray representing a 55o refraction angle path in the test piece (the nominal refraction angle

of the wedge), the selected focal point lies at approximately 90% of the near field distance from

the center of the active array.

(a)

Page 85: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

71

Figure 8.4 The calculated time delays of individual elements for focusing an ultrasonic beam at the selected focal

points located on the wave sound paths representing (a) 45o, (b) 55

o and (c) 70

o refraction angles inside a steel test

block using an Eclipse WA12-HT55S-IH-G PEI wedge-5L64-A12 Olympus array system (as depicted in Figure

8.3). The first 16 elements of the aperture are used for beam formation. The blue and red data points represent the

results for room (25oC) and elevated temperature (150

oC) inspection condition, respectively. In each case, delays are

expressed relative to the element associated with the longest travel time from array to the focal point. This element

is fired first with delay time defined as time zero; all other elements are then fired at their relative delay times.

(b)

(c)

Page 86: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

72

Figure 8.4-a, Figure 8.4-b and Figure 8.4-c show the calculated relative delay times associated

with ultrasonic waves focused at the selected focal points on the 40o, 55

o, and 70

o refraction

sound paths respectively. The blue and red data points represent the results corresponding to

room (25oC) and elevated temperature (150

oC) of the steel test piece, respectively. Delay times

are expressed relative to the element associated with the longest travel time from array to focal

point. This element should be fired first at a time defined as t=0; all other elements should then

be fired at their respective relative delay times.

Application of conventional (i.e., room-temperature) array focal laws for generation of a focused

beam at 40o yielded time delay errors for array elements of up to 174 ns when applied to a test

block at 150oC, as seen in Figure 8.4-a; this timing error is approximately 87% of one period at

the pulse central frequency of 5 MHz (period of 200 ns) in the test piece. Such large percentage

errors imply that the ultrasound pulses arriving from all the array elements would not all

constructively add together at the focal point. Instead there would be appreciable destructive

interference of the beam, significant distortion of the desired profile and loss of ultrasound

amplitude at the focal point, and non-optimal imaging resolution.

The effects of thermal gradients on the optimal element time delays is even more severe for the

case of generating a focused beam at larger refraction angles in a test piece, such as the focal

point located on a 55o refraction sound path as depicted in Figure 8.4-b. In this case, the

calculated time delays for the case of generating a focused beam in a hot test piece show a

significantly different profile from the delays calculated for inspection at room temperature. It is

noted that the high temperature time delays are expressed relative to element number 16 and the

room temperature time delays are expressed relative to element number 1. This indicates that the

wave associated with element number 1 undergoes a longer travel time to the focal point

compared to the waves emitted from all the other elements at room temperature condition.

However, at an elevated temperature condition the wave emitted by this element reaches the

focal point more quickly compared to all other elements, due to wave skew and the temperature

profiles in the wedge and test piece.

For the case of generating a focused beam at a focal point located on a 70o refraction path line

(Figure 8.4-c), the maximum discrepancy between element delay times calculated by the

conventional room-temperature focal law calculator and the new algorithm is 147 ns; this is

Page 87: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

73

approximately 73% of a period of the pulse central frequency of 5 MHz. A lack of proper

focusing at the desired focal point would result.

The three graphs of Figure 8.4 illustrate how the effects of thermal gradients on the optimal

element delay times have a complicated dependence on the location of the focal point in the steel

test piece. The following points are noted:

1) The general profile of the optimal time delay pattern is dependent on the refraction angle

associated with the focal point in the test piece.

2) The magnitude of thermal gradient effect on the array focal law for generation of a

focused beam in a hot test piece depends on the refraction angle associated with the focal

point in the test piece. The effect is larger when there is a significant deviation between

the direction of the thermal gradient and the direction of wave propagation: larger

deviations between these two directions lead to greater beam skewing.

The next step is to design an experiment to evaluate the accuracy of the focal laws calculated by

our model for the generation of a focused beam in a hot test piece. This is described in Section

8.3.

8.3 Experimental Evaluation

In this section, we develop an experiment to evaluate the effectiveness of the new algorithm with

respect to the focusing performance of a phased array system at elevated temperatures. Sectorial

scans are commonly used for inspection of welds in pipes, plates, and pressure vessels.

Furthermore, a focused beam can be used in a sectorial configuration, i.e., focused beams sweep

along a range of desired angles and depths in the test piece. Use of focused beams increases the

resolution of the scan image by increasing the local sound pressure and decreasing the beam spot

size.

Several methods of phased array performance assessment are described in ultrasonic inspection

guides. In particular, standardized calibration blocks are often used for checking the capabilities

of a phased array ultrasonic system. These blocks contain a series of reflectors such as side

drilled holes which are distributed within the block in a specific pattern. The echo signals

Page 88: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

74

received from these reflectors are analyzed for calibration purposes based on established codes

[47] [48].

Figure 8.5 illustrates an example of such a block for assessment of phased array steering

capability [47]. This block is designed for phased array probe steering assessment when the

probe is used in direct contact with the block (no wedge). A series of 2 mm diameter side drilled

holes are located at 5o intervals on an arc, 50 mm from the surface location where the midpoint

of the active aperture is located. Once the probe is located at the proper position, a focused beam

is electronically swept along the 50 mm arc with 1o angular steps. Amplitudes of the received

echo signals are plotted versus their corresponding steering angle. High amplitudes would be

expected to correspond to the locations of the side drilled holes. Therefore, the clarity of these

amplitude peaks as a function of steering angle indicates the angular range of the array.

Figure 8.5 Calibration block for steering assessment of phased array probe based on the amplitudes of the echo

signals received from a series of 2-mm diameter side drilled holes. The holes are located at 5o interval on an arc 50

mm from the surface location where the midpoint of the array’s active aperture is located. A focused beam is

electronically swept along the arc with 1o angular steps. Amplitudes of the received echo signals are analyzed. High

amplitudes are expected to correspond to the locations of the holes. The clarity of the amplitude peaks as a function

of steering angles indicates the angular range of the array.

Page 89: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

75

The block depicted in Figure 8.5 can also be used for resolution assessment of the phased array

probe in a sectorial scanning configuration. The resolution capability in a sectorial scan is

identified as “angular resolution” [48]. The angular resolution is the minimum angular separation

at which two identical targets located at the same sound path distance (same wave travel time

between the source and the targets) can be separated. By one definition, two identical targets at

the same sound path distance are considered resolved in angle when the echo amplitude from a

location midway between the two targets is at least 6 dB less than the echo from either target

[48].

This type of calibration block cannot be used for performance assessment of a wedge-array

system, in which the apparent beam exit point from the wedge varies strongly with refraction

angle. In such a case, the length of the sound path would be different for each side-drilled hole

and distort the results. In this project, we have therefore designed a modified steel block for

angular resolution and steering assessment of our wedge-array system, using a set of only four

side drilled holes as reflectors. The locations of the reflectors were selected to compensate for the

dependence of path length in the wedge and test block on refraction angle.

The block was specifically designed for use with the Eclipse WA12-HT55S-IH-G PEI wedge

and first 16 elements of the model 5L64-A12 Olympus linear phased array probe. A series of

focal points was selected along an arc in the steel block; the focal points were associated with

refraction angles ranging from 40o up to 70

o in 1

o increments (31 focal points), as shown in

Figure 8.6. Each sound path is approximated by a single straight line originating at the center of

the active array aperture. The red arc in Figure 8.6 represents the boundary of the near field, and

the black arc links the locations of the 31 focal points. Locations of both arcs were determined

using the same procedure as described in Section 8.2.

Four side drilled holes were located on the black line. The diameter of the doles was selected to

be 1.5 mm which is higher than the wavelength of the wave propagating in the steel block as

specified by the phased array performance assessment code [47]. The side drilled holes were

centered on the focal points which lie along the sound paths representing the refraction angles of

44o, 49

o, 56

o and 66

o. The block configuration, focal points and side drilled holes are depicted in

Figure 8.7. Also illustrated in this Figure are the wave sound paths in the steel corresponding to

the various refraction angles inside the block as depicted in Figure 8.6.

Page 90: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

76

Figure 8.6 Design specifications of a steel block for performance assessment of a wedge-array system. The block

was specifically designed for use with Eclipse WA12-HT55S-IH-G PEI wedge and first 16 elements of the 5L16-

A12 linear phased array probe. The blue straight lines represent the wave paths associated with refraction angles

ranging from 40o up to 70

o in 1

o increments. The red arc represents the boundary of the near field on the specified

sound paths. The black arc links the locations of the focal points on the presented sound paths. These are located at

the same travel time distance from the center of the active aperture, and are all within the near field.

The designed block was first used for assessment of our wedge-array system at ambient

temperature of 25oC. The appropriate element time delays were calculated for focusing the

ultrasonic beam at the selected focal points based on the conventional room temperature focal

law calculation procedure as described in Section 8.1. The array was connected to a Focus LT

phased array instrument which can apply any user-specified relative element time delays to the

selected active elements of the array using TomoView software. The wedge-array system was

placed on the designed steel test block for generation of the focused beams, as shown in Figure

8.8.

Page 91: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

77

Figure 8.7 Configuration of the designed steel block for angular resolution and steering assessment of an Eclipse

WA12-HT55S-IH-G PEI Wedge-5L64-A12 Olympus array system: a set of four side drilled holes are selected as

reflectors. The holes are centered on the focal points which lie along the sound paths representing the refraction

angles s of 44o, 49

o, 56

o and 66

o. The straight lines represent the wave paths associated with refraction angles

ranging from 40o up to 70

o in 1

o increments. The black and red points represent the location of focal points (steering

samples) and the near field points on the specified sound paths respectively.

Figure 8.8 Experimental set up for evaluation of model results for focusing the ultrasonic beam in a steel test block

at 25oC. An Eclipse WA12-HT55S-IH-G PEI wedge was used with a 5L64-A12 Olympus array for generation of

focused beams. The steel test block contains four side drilled holes at locations specified in Figure 8.7.

Page 92: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

78

For each of the 31 focal points, the relative element time delays were calculated for the selected

active array elements (elements 1 to 16). The focused beam was then swept electronically along

the focal points, representing refraction angles of 40o to 70

o in 1

o increments. For each of the 31

focal points, the echo signals received at the 16 active array elements were summed

electronically, using the same relative time delays as applied to the transmitted signals. The

resultant total echo signals were then converted to their analytical form. The maximum

amplitudes of the obtained analytical signals were plotted versus the corresponding refraction

angles, Figure 8.9. The experiment repeated three times to ensure the results were reproducible

with negligible difference.

Figure 8.9 Assessment results of an Eclipse WA12-HT55S-IH-G PEI wedge-5L64-A12 Olympus array system at

ambient temperature of 25oC. The focused beam was swept electronically along the focal points as depicted in

Figure 8.7 and Figure 8.8. The maximum amplitude of the analytical form of the echo signals was plotted versus the

refraction angles (blue points). The local amplitude peaks correspond to refraction angles of 44o, 49

o, 56

o and 66

o

inside the steel test block (dashed lines). These represent the location of the side drilled holes.

The local amplitude peaks in Figure 8.9 correspond to refraction angles of approximately 44o,

49o, 56

o and 66

o inside the steel test block - the locations of the side drilled holes were

successfully identified through the sectorial scanning with a focused beam. The following points

are noted:

Page 93: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

79

1) The amplitude peaks corresponding to all the side drilled holes are not identical in

magnitude. This is partly because the sound path length from the array to the focal points

varies with refraction angle. Therefore, the echo signals corresponding to the holes

undergo different level of attenuation and beam spreading as they travel through the

wedge and steel test block. In addition the transmission coefficient from the wedge to the

test block varies with the refraction angle.

2) The echo profile in Figure 8.9 does not consistently meet the resolution criterion of a 6

dB drop in signal amplitude between reflectors located with 5o angular separation. This

point is particularly evident for the target holes on sound paths corresponding to 44o and

49o refraction angles (see Figure 8.7). However, a skilled technician would have little

difficulty in resolving the adjacent target reflectors. The results indicate that the angular

resolution limit of the inspection system is close to 5o for ambient temperature (25

oC)

inspection conditions.

Once the focusing effectiveness of our system was evaluated for the case of room temperature

inspection, the same experiment was repeated at elevated temperature for the more challenging

case of the steel block at 150oC. The same wedge-array system was used for the elevated

temperature experiment as used for the room temperature experiment. The designed block was

placed on a temperature-controlled hot plate, and the block-wedge surface was kept at 150oC.

The steel block was wrapped in high-temperature insulation to minimize heat loss and

temperature gradients in the block. This led to temperature variations over the entire volume of

the test piece of less than 5oC, as indicated by experimental measurement at selected locations.

Once the temperature stabilized in the test piece and wedge system, the calculated relative

element time delays were applied to the selected active array elements (elements 1 to 16) to

generate the focused beams. The experiment was performed twice: First, with time delays

calculated for ambient temperature (25oC) conditions with no compensation for the effect of

thermal gradients inside the wedge. The experiment was then repeated using the focal laws

calculated by our new model for the actual experimental condition of elevated temperature

(150oC test block) and thermal gradient in the wedge. The motivation for running the experiment

twice in this manner was to help assess whether use of the high-temperature focal law actually

yields a significant improvement in ultrasonic inspection results.

Page 94: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

80

For both cases the relative element time delays were applied to the selected active array elements

(elements 1 to 16) for signal transmission. The received elements echo signals were summed

using the same relative time delays. The amplitudes of the summed echo signals were plotted

versus refraction angle, in the same manner as was done for the ambient temperature results of

Figure 8.9.

The results of the elevated-temperature experiments are shown in Figure 8.10. The blue and red

data points represent the results corresponding to the application of the room temperature

(incorrect) and elevated temperature (correct) focal laws, respectively, for generation of focused

beams in the steel test piece.

Figure 8.10 Assessment results of an Eclipse WA12-HT55S-IH-G PEI wedge-5L64-A12 Olympus array system at

150oC. The focused beam was swept electronically along the focal points as depicted in Figure 8.7 and Figure 8.8

and the maximum amplitude of the analytical form of the echo signals was plotted versus the refraction angles. Two

sets of focal laws were used for beam formation: First, the focal laws calculated for ambient temperature condition

(25oC) and then the focal laws calculated by our new model for the actual experimental condition (150

oC test block).

The results corresponding to these cases are presented by the blue and red points, respectively. The dashed lines

represent refraction angles which correspond to location of the side drilled holes in the steel test block.

As seen in Figure 8.10, in both cases four peaks can be identified in the amplitude-angle curve

which represents the four side-drilled holes in the sample block. As expected, the peaks do not

Page 95: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

81

match exactly with the actual location of the holes when the incorrect focal laws were used (blue

points). This implies that application of incorrect relative element delay times led to non-optimal

focusing of the beam at the desired location inside the piece.

When the correct (high-temperature) focal laws were used with accommodation for beam skew,

there is a slight shift of the peak amplitudes toward the expected locations. This would help in

determining the location of flaws detected in an ultrasonic inspection. The change in signal

amplitude and resolvability is negligible; a major source of error is believed to be uncertainty in

the temperature profile in the wedge and the associated velocity profile. It is noted that the test

piece temperature considered in this experiment was only 150oC. Use of the new focal law is

expected to become increasingly important as temperatures are raised further. Although the new

wedge materials can tolerate temperatures up to 350oC, limitations in the laboratory experimental

system prevented conducting our tests at this more demanding temperature.

Page 96: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

82

9 Summary, Conclusions and Future Work

The design and operation of an ultrasonic phased array system were described for inspections of

engineering components such as pipe welds at elevated temperatures of up to 350oC. Wedges

were built from plastics resistant to high temperature degradation, and equipped with a cooling

jacket around the array. The temperature distribution inside the wedges was modeled with finite

element software and validated experimentally. Ultrasonic compression wave velocities of

wedge materials were measured as a function of temperature. These were used to model the

skewed path of waves propagating across thermal gradients in the heated wedge utilizing a

numerical ray-tracing technique.

A new algorithm was developed based on the beam skew model, to calculate transmission and

reception time delays on individual array elements for generation of plane waves or focused

beams in test pieces at elevated temperatures, while compensating for thermal gradient effects

inside the wedge. The following results and key points were noted:

1. The amount of beam skew depends on the amount of misalignment between the beam

propagation vector and the gradient of the temperature or velocity field inside the wedge.

2. Application of conventional array focal law for phased array inspections at elevated

temperatures leads to incorrect time delays on array elements, significant distortion of the

desired beam profile, and a reduction in imaging resolution.

3. The magnitude of the thermal gradient effect on the array focal law for generation of a

shear plane wave in a test piece strongly depends on the location of the active elements

on the array, the angle s of plane wave propagation in the test piece, and the magnitude

and orientation of the thermal gradient inside the high temperature wedge. These items

are all inter-related; for example, the angle between the thermal gradient and the

ultrasound propagation vector varies strongly with the choice of array elements used for

plane wave formation. That angle and resultant beam skew are generally larger if the

upper array elements are used as opposed to the lower array elements. That angle also

Page 97: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

83

tends to be larger if the system is set to generate a plane wave in the test piece with a

large refracted angle s as opposed to a small refracted angle.

4. The magnitude of the thermal gradient effect on the array focal law for generation of a

focused beam depends on the refraction angle associated with the focal point in the hot

test piece. The effect is larger when there is a significant deviation between the direction

of the thermal gradient and the direction of wave propagation: larger deviations between

these two directions lead to greater beam skewing.

Element time delays calculated by the new algorithm were validated experimentally for

generation of shear plane waves and focused beams in a steel block at 150oC. The experimental

results indicate that the new algorithm does determine appropriate element time delays for

inspection of a heated piece using both shear planar waves and focused beams.

Several further steps can be carried out to complement this thesis:

1. The new wedge materials can tolerate temperatures up to 350oC, and laboratory

verification tests should be carried out at such temperatures. Experimental validation of

the new focal law was not possible in our case for temperatures greater than 150oC, due

to limitations of laboratory equipment. The effects of the thermal gradients should be

much larger at 350oC, and therefore the improvements enabled by the new focal law

should be more evident.

2. The new algorithm for focal law generation should be implemented in an automated

scanning system to perform scans at elevated temperatures.

3. The high temperature scanning system can be used for inspection of blocks which contain

defects with known size and location at elevated temperatures. The scan result can be

used to further evaluate the effects of temperature on signal amplitude.

4. The results in step 3 can be used for developing standards for performance of high

temperature inspections with phased array systems.

Page 98: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

84

References

[1] R. Manzini, A. Regattieri, H. Pham and E. Ferrari, Maintenance for Industrial Systems,

Springer, 2010.

[2] M. Brendt, "Non-Destructive Testing Methods for Geothermal Piping," Brookhaven

National Lab, New York, 2001.

[3] J. Wessels, "Ultrasonic Testing of Thin-Walled Austenitic Welds," International Journal

of Pressure Vessles & Piping, vol. 39, pp. 295-312, 1989.

[4] T. Arakawa, S. Hirose and T. Senda, "The Detection of Weld Cracks Using Ultrasonic

Testing," NDT International, vol. 18, no. 1, 1985.

[5] J. Krautramer and H. Krautramer, Ultrasonic Testing of Materials, New York: Springer-

Verlag, 1983.

[6] J. L. Rose, Ultrasonic Waves in Solid Media, New York: Cambridge University Press,

1999.

[7] R. Cobald, Foundations of Biomedical Ultrasound, New York: Oxford University Press,

2007.

[8] J. Schmerr and W. Lester, Fundamentals of Nondestructive Evaluation, New York:

Plenum Press, 1998.

[9] K. Mirkhani, Characterization of Interfacial Degradation in Adhesive Joints Using

EMATs, Toronto: PhD Thesis, Department of Mechanical and Industrial Engineering,

University of Toronto, 2002.

[10] N. Dube, Introduction to Phased Array Ultrasonic Technology Applications, Olympus

NDT, 2007.

Page 99: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

85

[11] A. Erhad, G. Schenk, T. Hause and U. Volz, "New Applications Using Phased Array

Techniques," Nuclear Engineering and Design, pp. 325-336, 2001.

[12] M. Moles, N. Dube, S. Labbe and E. Ginzel, "Pipeline Girth Weld Inspections Using

Ultrasonic Phased Arrays," in International Pipeline Conference, Calgary, 2002.

[13] "Phased Array tutorial," Olympus NDT, [Online]. Available: http://www.olympus-

ims.com/en/ndt-tutorials/transducers/inside.

[14] E. Ginzel, Phased Array Ultrasonic Technology, Waterloo: Eclipse Scientific Inc., 2013.

[15] O. NDT, "Phased Array Probes and Wedges," [Online]. Available: http://static1.olympus-

ims.com/data/File/PA_Probes/PA_Probe_Catalog.en.pdf?rev=F9D0.

[16] M. Anderson, Phased Array Testing: Basic Theory for Industrial Applications, Olympus

NDT, 2010.

[17] Optel, "Phased Array Description," Optel, [Online]. Available:

http://www.optel.pl/manual/english/oppa.htm.

[18] O. NDT, "Phased Array Tutorial," Olympus NDT, [Online]. Available:

http://www.olympus-ims.com/en/ndt-tutorials/instrumententation/pa-scan/.

[19] M. Moles and G. Fortier, "Phased Arrays for Pipeline Girth Weld Inspections," in IV

Conferencia Panamericana de END, Buenos Aires, 2007.

[20] M. Moles, "Applications of Industrial Phased Arrays," in 2nd American Welding Society

Conference on NDT, Orlando, 2003.

[21] M. Moles, "Code Inspection of Wleds Using Phased Arrays," Energy Materials:

Scienceand Engineering for Energy systems, vol. 4, pp. 108-112, 2009.

[22] I. Atkinson, C. Gregory, S. Kelly and K. Kirk, "Ultrasmart: Developments in Ultrasonic

Flaw Detection and Monitoring for High Temperature Plant Applications," in ASME

Page 100: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

86

Pressure Vessels and Piping Conference, Texas, 2007.

[23] K. Kirk, A. McNab, A. Hall and G. Hayward, "Ultrasonic Arrays for monitoring Cracks

in an Industrial Plant at High Temperatures," IEEE Transactions on Ultrasonic,

Ferroelectrics, and Frequency Control, vol. 46, no. 2, 1999.

[24] J. Shih, K. Wu, C. Jen, C. Chiu, J. Tzeng and J. Liaw, "Application of Flexible Ultrasonic

Transducer Array for Defect Determination at 150oC," Sensors, vol. 13, pp. 975-983,

2013.

[25] L. Polymer Plastics Company, "Rexolite," Polymer Plastics Company, LC, [Online].

Available: http://www.polymerplastics.com/mechanical_rex.shtml.

[26] "MatWeb Material Property Data," [Online]. Available:

http://www.matweb.com/search/DataSheet.aspx?MatGUID=65baf7a4f90c4a54a6ace03e1

6b1125b.

[27] "MatWeb Material Property Data," Quadrant EPP, [Online]. Available:

http://www.matweb.com/search/DataSheet.aspx?MatGUID=0b5b8eb71f444db8a3d7b23d

f6db0782.

[28] F. Incropera, D. Dewitt, T. Bergman and A. Lavine, Fundamentals of Heat and Mass

Transfer, John wiley $ Sons Inc., 2007.

[29] Heat Transfer Module User's Guide, COMSOL Multiphysics, 2012.

[30] M. O' Donnell, E. Jaynes and J. Miller, "General Relationships Between Ultrasonic

Attenuation and Dispersion," Journal of Acoustical Society of America, vol. 63, 1978.

[31] Y. Chevalier and J. Tuong, Mechanics of Viscoelastic Materials and Wave Dispersion,

Hoboken, NJ: John Wiley & Sons, Inc., 2010.

[32] M. Felix, "Attenuation and Dispersion Characteristics of Various Plastics in the

Frequency range 1-10 MHz," Journal of Composite Materials, vol. 8, pp. 275-287, 1974.

Page 101: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

87

[33] W. Sachse and Y. Pao, "On the Determination of Phase and Group Velocities of

Dispersive Waves in Solids," Journal of Applied Physics, vol. 49, 1978.

[34] T. Cutard, D. Fargeot, C. Gault and M. Hunger, "Time Delay and Phase Shift

Measurements for ultrasonic Pulses Using Autocorrelation Methods," Journal of Applied

Physics, vol. 75, pp. 1909-1913, 1994.

[35] H. Hachiya and S. Ohtsuki, "Determination of Sound Speed in Biological Tissue Based on

Frequency Analysis of Pulse Response," Journal of Acoustical Society of America, vol.

92, 1992.

[36] O. NDT, "Panametrics Ultrasonic Transducers," [Online]. Available:

http://www.olympus-ims.com/data/File/panametrics/panametrics-UT.en.pdf.

[37] H. Gjoystdl, E. Iversen, R. Laurain, I. Lecomte, V. Vinje and K. Astebol, "Review of Ray

Theory Applications in Modeling and Imaging of Seismic Data," Studia Geophysica &

Geodaetica, vol. 46, no. 2, pp. 293-320, 2002.

[38] V. Cerveny, Seismic Ray Theory, UK: Cambridge University Press, 2001.

[39] J. Ogilvy, "Computerized Ultrasonic Ray Tracing in Austenitic Steel," NDT International,

vol. 18, no. 2, pp. 67-77, 1985.

[40] S. Kolkoori, M. Rahaman, P. Chinta and J. Prager, "Computationally Efficient Ray

Tracing Algorithm for Simulation of Transducer Fields in Anisotropic Materials," in

Proceeding for the National Seminar & Exhibition on NDE, 2011.

[41] O. Nowers, D. Duxbury, J. Zhang and B. Drinkwater, "Novel Ray-Tracing algorithms in

NDE: Application of Dijkstra and A* Algorithms to the Inspection of an Anisotropic

Weld," NDT&E International, vol. 61, pp. 58-66, 2014.

[42] C. Zelt and R. Ellis, "Practical and Efficient Ray Tracing in Two-Dimensional Media for

Rapid Travel Time and Amplitude Forward Modeling," Canadian Journal of Exploration

Page 102: Development of Ultrasound Phased Array System for · PDF fileDevelopment of Ultrasound Phased Array System for Weld Inspections at Elevated Temperatures ... Figure 2.13 Scan plan of

88

Geophysics, vol. 24, no. 1, pp. 16-31, 1988.

[43] V. Cerveny, "Seismic Rays and Ray Intensities in Inhomogeneous Anisotropic Media,"

Geophysical Journal International, vol. 29, pp. 1-13, 1972.

[44] MathWorks, "Documentation Center, ode45," MathWorks, [Online]. Available:

http://www.mathworks.com/help/matlab/ref/ode45.html.

[45] D. Wackerly, W. Mendenhall and R. Scheaffer, Mathematical Statistics with

Applications, Thomson Learning Inc., 2008.

[46] E. Ginzel, "Near Field and Focusing Consideration with Wedges," Materials Research

Institute, Waterloo, ON.

[47] "Standard Guide of Evaluation Performance Characteristics of Phased Array Ultrasonic

Testing Instruments and Sytems," ASTM International, Vols. E2491-08, 2008.

[48] E. Ginzel and D. Johnson, "Phased Array Resolution Assessment Techniques," The e-

Journal of Nondestructive Testing, 2008.

.