device presentation kris dan · 3/2/18 2 obtaining a strong pulse oximetry trace •most clinical...

7
3/2/18 1 Guidelines – monitoring, evacuation and equipment Daniel Roberts (senior radiographer) Kristopher Knott (CMR fellow) Effective Monitoring Why do we need monitoring? CMR is reliant on ECG gating Direct monitoring of patient ECG is an additional necessity when scanning device patients Of primary concern is any deviation from expected heart rhythm/rate Dependent on chosen pre-scan device settings To be discussed with all relevant team members pre-scan (cardiologist/s, cardiac physiologist/s and radiographer/s) Further team discussion: Type of device (PPM/ICD, conditional/non- conditional), leads, underlying rhythm, expected response to stress…etc BHC SOP: Scanning of Patients with PPM Two forms of monitoring via ECG and pulse oximetry Typical 4 lead ECG used for clinical CMR is sufficient for purpose To be set up as soon after re-programming of device as possible Can be done outside scan room itself where wireless ECG probes are utilised Pulse oximetry used as a surrogate for rhythm analysis in situations when electrocardiography exhibits artefact Realtime monitoring of O2 saturation levels Factors effecting ECG trace Vectorcardiographic (VCG) techniques are the most common clinical norm for ECG monitoring/gating: Overcomes the magnetohydrodynamic effect Higher field strength = greater effect RF and Gradients Low SAR mode or manual reduction of gradients if absolutely necessary Movement of patient within static field: Closely monitor ECG when moving the patient into the scanner Watch for rhythm pausing and adjust device settings if necessary Obtaining a strong ECG trace Adequate skin preparation: Removal of naturally occurring oils and body hair Mildly abrasive preparation gel ECG electrode positioning: Creative electrode placement may be necessary for a strong ECG trace depending on body habitus and device position Trial and error approach may need to be adopted regarding electrode positioning and channel selection Excessive patient movement may also illicit artefact

Upload: others

Post on 20-Oct-2019

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Device presentation kris dan · 3/2/18 2 Obtaining a strong Pulse Oximetry trace •Most clinical CMR scanners will allow for pulse oximetry, however: •For ECG gating only •Dedicated

3/2/18

1

Guidelines– monitoring,evacuationandequipment

DanielRoberts(seniorradiographer)KristopherKnott(CMRfellow) EffectiveMonitoring

Whydoweneedmonitoring?

• CMRisreliantonECGgating

• DirectmonitoringofpatientECGisanadditionalnecessitywhenscanningdevicepatients

• Ofprimaryconcernisanydeviationfromexpectedheartrhythm/rate• Dependentonchosenpre-scandevicesettings• Tobediscussedwithallrelevantteammemberspre-scan(cardiologist/s,cardiacphysiologist/sandradiographer/s)

• Furtherteamdiscussion:Typeofdevice(PPM/ICD,conditional/non-conditional),leads,underlyingrhythm,expectedresponsetostress…etc

BHCSOP:ScanningofPatientswithPPM• TwoformsofmonitoringviaECGandpulseoximetry

• Typical4leadECGusedforclinicalCMRissufficientforpurpose• Tobesetupassoonafterre-programmingofdeviceaspossible• CanbedoneoutsidescanroomitselfwherewirelessECGprobesareutilised

• Pulseoximetryusedasasurrogateforrhythmanalysisinsituationswhenelectrocardiographyexhibitsartefact

• RealtimemonitoringofO2saturationlevels

FactorseffectingECGtrace

• Vectorcardiographic (VCG)techniquesarethemostcommonclinicalnormforECGmonitoring/gating:• Overcomesthemagnetohydrodynamiceffect• Higherfieldstrength=greatereffect

• RFandGradients• LowSARmodeormanualreductionofgradientsifabsolutelynecessary

• Movementofpatientwithinstaticfield:• CloselymonitorECGwhenmovingthepatientintothescanner• Watchforrhythmpausingandadjustdevicesettingsifnecessary

ObtainingastrongECGtrace

• Adequateskinpreparation:• Removalofnaturallyoccurringoilsandbodyhair• Mildlyabrasivepreparationgel

• ECGelectrodepositioning:• CreativeelectrodeplacementmaybenecessaryforastrongECGtracedependingonbodyhabitusanddeviceposition

• Trialanderrorapproachmayneedtobeadoptedregardingelectrodepositioningandchannelselection

• Excessivepatientmovementmayalsoillicitartefact

Page 2: Device presentation kris dan · 3/2/18 2 Obtaining a strong Pulse Oximetry trace •Most clinical CMR scanners will allow for pulse oximetry, however: •For ECG gating only •Dedicated

3/2/18

2

ObtainingastrongPulseOximetrytrace• MostclinicalCMRscannerswillallowforpulseoximetry,however:

• ForECGgatingonly• Dedicatedpulseoximetryrequiredtomeasureoxygenation• Audiblealarmwhenoxygensaturationdropsbelowgiventhreshold

• Signaldropoutmayrequirestrialanderrorwithalternativefingersorprobetype:• Peripheralvasoconstriction(coldextremities)• Poorcardiacoutput• Excessivemovement;tremors• Nailpolish(darklypigmented)

EvacuationProcedures

BHCEvacuationProcedure

1. OnlyauthorisedCMRstaffaretoenterthescanroomintheeventofanemergency

2. CMRstaffmustremovepatientfromwithintheboreandtransferpatientontoaCMRcompatibletrolley(tobekeptinthescanroom)

3. 2222iscalledbyamemberofstaff4. Patientisremovedfromscanroomandtakentorecoverybay

wherecrashteamwilladministeremergencymeasures

1.CMRauthorisedstaffonlyinthescanroom• Authorisedmeans:

• CertifiedtoenteranMRcontrolledarea• AmemberofMRcentrepersonnel• MusthavereadandunderstoodlocalrulesregardingMRsafety• Completedasafetyscreeningquestionnaireannually

• Keepalltraininguptodateandwelldocumented

• Asalways,itistheresponsibilityofauthorised/responsiblestaffmemberstoensurenounauthorisedpersonsenterthescanroom

2.PatientremovedfromscannertoCMRcompatibletrolley• MRcompatibletrolleytobepresentatalltimesandwithoutfail• Keeptrolleyclearofclutter• Trolleybreakstobekeptunlockedinordertofacilitateswiftevacuation• Evacuationsheetisofparticularimportance

• For‘highrisk’patients– earlyECGmonitoringandexternalpacingequipmentinpreparationroom

• Allinvolvedstafftobefullytrainedinsafemanualhandlingtechniques• Evacuationroutesshouldbeclearofobstructions/triphazards

3.Callingforappropriatesupport

• StandardBLSprocedureincaseofcardiacarrest• Crashcall:2222orequivalent• Informsupervisingconsultant/s

• Staffwhoshouldalreadybepresent:• Radiographer(ALSqualified)• Cardiologist(ALSqualified)ifnon-conditionaldevice• Cardiacphysiologist• +/- MRPhysicistifgreatdealofartefactencountered

Page 3: Device presentation kris dan · 3/2/18 2 Obtaining a strong Pulse Oximetry trace •Most clinical CMR scanners will allow for pulse oximetry, however: •For ECG gating only •Dedicated

3/2/18

3

Evacuationsupportinpractice

• Ensurethatrequiredlevelsoflifesupporttrainingaremaintained:• Mandatorytraininguptodate• Supplementarytrainingifrequired• Documentation

• Thescanningofdevicesrequirestheattentionofallinvolvedmembersofstaff• Minimumstaffinglevelsmaintainedatalltimesincaseofemergency• Allinvolvedstaffmemberstobeinimmediatevicinityoreasilyandreliablycontactable

• 1deviceCMRperteam

4.Evacuationfromscanroom

• Incaseofemergency,patienttobemovedtoZone3asquicklyaspossible• Zone4– Synonymouswithscanroomitself• Zone3– Outsidescanroomdoor.Accessforsafetyscreenedpatientsandstaffonly.Oftencontrolroom/anteroom.

• Zone2– Nodetectiblemagneticfieldbutrestrictedtostaff,patientandfamilies

• Zone1– Norestriction

• Therefore,ensureallthoseresponsibleforfirstresponseinanemergencyhavebeensafetyscreened

Recoveryareaindetail

• Zone3• Dedicated/aptforpurpose:

• Appropriatesize• Unoccupied• Clearofobstruction• Allowsforprivacy

• Withineasyreachofcrashtrolley:• AEDwithexternalpacingcapacity• Atropineandadrenaline

• Deviceprogrammerforrelevantmanufacturers• Devicemagnet

Pre-scan– knowyourpatient

• Whatdevicedoesyourpatienthave?

• Terminology:• PPM,CRT-P,CRT-D,ICD,S-ICD

• Arealldevicesthesame?

• Whywasthedeviceputin?

Pre-scan– knowyourpatient

• Whatisthedevicedoinginthispatient?

• Whatdomagnetsdotothedevice?

• Whatcangowronginthescanner?

• IsthedeviceMRconditional?

• Howhasthedevicebeenprogrammedpre-scan?

Page 4: Device presentation kris dan · 3/2/18 2 Obtaining a strong Pulse Oximetry trace •Most clinical CMR scanners will allow for pulse oximetry, however: •For ECG gating only •Dedicated

3/2/18

4

Terminology

• PPM– permanentpacemaker• ICD– implantablecardioverterdefibrillator• CRT-P- Cardiacresynchronisationtherapy– pacemaker• CRT-D- Cardiacresynchronisationtherapy– defibrillator• S-ICD– subcutaneousICD

Pre-scan– settinguptheequipment

• MRIsafetrolley• Patslide/sheet• Resuscitationtrolley• Defibrillator

Pre-scan– whoishighrisk?

• Pacingdependent

• Highthresholdsinleadspre-scan

• Oldorbrokenleads

• Historyofrecentventriculararrhythmia

Page 5: Device presentation kris dan · 3/2/18 2 Obtaining a strong Pulse Oximetry trace •Most clinical CMR scanners will allow for pulse oximetry, however: •For ECG gating only •Dedicated

3/2/18

5

Pre-scan– whoneedstobethere?

• Radiographer

• Physiologist– deviceprogramming,monitoringinhighrisk

• Cardiologist– non-conditional,highriskpatients

Pacemakers

• Isthepatient“pacingdependent”?• Whathappensifyouturnoffthepacemaker?!

• Thepacemakeriseithera“back-up”,incasetheheart’sconductionsystemfails….

…..Orthethepacemakerispacingmost/allofthetime.

Scenario1

Keypoint:knowwhatratethepacemakerwasprogrammedtopre-scan

Scenario1– bradycardiaduringthescan

• Ifpacinginhibitionisobservedduringthescan1. Abortthescan2. Movethepatienttozone33. Checkifpacinghasresumed4. Applyamagnettothepacemaker&/orprogramtoasynchronous

pacing5. Applyexternaldefibrillationpadsfortranscutaneouspacing6. ALS

Whatdoesamagnetdotoapacemaker?

• Typicallyasynchronouspacing• Paceatmagnetrate(forexample100bpm,90bpmor85bpm– dependentonmanufacturer,model,batterylifeetc)

• DOO,VOO,AOO

Page 6: Device presentation kris dan · 3/2/18 2 Obtaining a strong Pulse Oximetry trace •Most clinical CMR scanners will allow for pulse oximetry, however: •For ECG gating only •Dedicated

3/2/18

6

Scenario2

Scenario2– ventriculararrhythmias

• Ifsustainedventriculararrhythmiaorhaemodynamiccompromise1. Abortthescan2. Movethepatienttozone33. Re-activatetheICD4. Applyexternaldefibrillationpads5. ALS

Whatdoesamagnetdotoadefibrillator?

• Typicallyinhibitsarrhythmiadetection/therapydelivery• Noasynchronouspacing• Willnotre-activatethedefibrillatorinsomeonehavingVT• Willnotresumepacinginsomeoneasystolic

Resuscitation

• ABC• (CPR)• Applytheexternaldefibrillatorpads• Physiologistinterrogatesthedeviceconcurrently

Page 7: Device presentation kris dan · 3/2/18 2 Obtaining a strong Pulse Oximetry trace •Most clinical CMR scanners will allow for pulse oximetry, however: •For ECG gating only •Dedicated

3/2/18

7

Lessobviouscases Pragmaticapproach– whentocutshortthescan?• Infrequent,fewbeatsofVT

• Thisisahighriskpatient– havethingssetupaccordingly(physiologist/cardiologistpresent)

• Scanquicklyanddothepoint• Abortifmoresustainedrunsoccur

• Patientsmaydevelopsymptomsduetoincreasedoutput/diaphragmaticpacing.• Thisisoftenwelltoleratedbutmaynotbe.

Summary

1. Knowyourpatient2. Knowyourpatient’sdevice3. Ensuretheresuscitationequipmentissetupcorrectlyandyou

knowwhereitis4. Haveanevacuationprocedureinplace5. Knowhowtotreatbrady andtachyarrhythmias inpatientswith

devices

Thanksforyourattention

• Anyquestions?