Page 1: Digital frequency meter

Digital Frequency Meter

By : Parcha Amit.K

Roll No: 2K13E21

Department of electronics

University of pune

Page 2: Digital frequency meter

Contents• Introduction to Digital Frequency

Meter

• Basic Building Blocks of Frequency Meter

• Typical specifications of Frequency meter

• Applications

Page 3: Digital frequency meter

Introduction To Frequency Meter

(UNIVERSAL COUNTER)

• Need Of Frequency Meter• A Digital Frequency Meter is a general-purpose, basic, digital

counter for measuring, setting, and monitoring frequencies, for counting random events, and for industrial counting applications.

• Among its many frequency measurement applications are the test, adjustment , and calibration of oscillators, telemetering equipment, i-f amplifiers and filters.

• With appropriate transducers, it can be used to measure such physical quantities as pressure, temperature, strain, and weight.

• In conjunction with an optical or magnetic pickup, it will measure rotational speed also.

Page 4: Digital frequency meter
Page 5: Digital frequency meter
Page 6: Digital frequency meter

Basic principal of operation • A simplified block diagram, indicates the

principal circuits of the instrument and the signal flow paths.

• The input amplifier, pulse-forming circuits, and the five ring counting units perform the counting operation.

• The input amplifier increases the instrument sensitivity and input impedance.

Page 7: Digital frequency meter

• The ring counting units are cascaded scale-of-ten circuits.

• Each unit is coupled to a Numeric indicator, which provides the digital display.

• The main gate acts as a switch either to pass pulses (open gate) or to stop the flow of pulses (closed gate).

• The gate is driven by the main-gate flip-flop.

Page 8: Digital frequency meter

• Time-base pulses from the time-base divider switch the flip -flop between its two stable states to open and close the main gate.

• The crystal-oscillator output is shaped into a pulse-type waveform to drive the time-base divider circuits.

• The divider produces a train of pulses derived from the oscillator signal with a frequency of 10, 1 or 0.1 pulses per second as selected by the COUNTING TIME control.

Page 9: Digital frequency meter

• The pulse train is fed through the time-base gate to the main gate flip flop.

• The time-base gate and time-base gate flip flop operate in a manner similar to the main -gate flip-flop and main gate.

• The time-base gate prevents a divider pulse from opening the main gate during the display-time interval.

Page 10: Digital frequency meter

• A counting interval starts when the ring counting units have been reset, the time-base gate has opened, and a divider pulse arrives at the main-gate flip-flop to open the main gate.

• The counting interval ends when the next divider pulse is received.

• This pulse closes the main gate, starts the display interval by switching the time-base gate flip-flop to close the time-base gate, and activates the display-time generator.

Page 11: Digital frequency meter

• In this condition the divider pulses are locked out and the indicators display the measurement.

• After a time interval determined by the DISPLAY TIME control, the reset-pulse generator is triggered, and the ring counters are reset.

Page 12: Digital frequency meter
Page 13: Digital frequency meter

Synchronies Gating For Counting

Page 14: Digital frequency meter

Main Gate Requirements:

• As with any physical gate, the main gate of the counter does exhibit propagation delays and takes some finite time to both switch ON and OFF.

• This finite amount of switching time is reflected in the total amount of time the gate is open for counting.

• If this switching time is significant compared to the period of the highest frequency counted, errors in the count will result.

• However, if this switching time is significantly less compared to the period of the highest frequency counted, the error is not appreciable.

Page 15: Digital frequency meter

Types of Measurement Error

• The ±1 Count Error(quantization error)• When an electronic counter makes a measurement, a ±1

count ambiguity can exist in the least significant digit. This is often referred to as quantization error. This ambiguity can occur because of the non-coherence between the internal clock frequency and the input signal.

Page 16: Digital frequency meter

Quantization error

Page 17: Digital frequency meter

• The Time Base Error

• Any error resulting from the difference between the actual time base oscillator frequency and its nominal frequency is directly translated into a measurement error.

• This difference is the cumulative effect of all the individual time base oscillator errors described previously and may be expressed as dimensionless factor such as so many parts per million.

Page 18: Digital frequency meter

Trigger Error

• Trigger error is a random error caused by noise on the input signal and noise from the input channels of the counter.

• In period and time interval measurements, the input signal(s) control the opening and closing of the counter’s gate.

• This causes the main gate to be open for an incorrect period of time.

• This results in a random timing error for period and time interval measurements.

Page 19: Digital frequency meter

Frequency Measurement Error

• The accuracy of an electronic counter is dependent on the mode of operation.

• The total frequency measurement error may be defined as the sum of its ±1 count error and its total time base error.

• The relative frequency measurement error due to ±1 count ambiguity is

• Df/f = 1/ fin

where fin is the input signal frequency.

Page 20: Digital frequency meter

SPECIFICATIONS

• Hp - Model 5240A

(12.4 GHz Digital Frequency Meter)• FREQUENCY MEASUREMENTS:

• RANGE: 10 Hz to 12.5 MHz

• GATE TIME: 0.1. 1.0 s: 10 s on special order.

• SELF CHECK: Counts 1 MHz for gate time chosen.

• SIGNALINPUT :

• SENSITIVITY: 100 mV rms.

• MAX. INPUT: 2 V rms.

• IMPEDANCE: 1 MΩ shunted by 10 pF.

Page 21: Digital frequency meter

• TIME BASE:FREQUENCY: 1 MHz

EXTERNAL INPUT: 1 V rms into 1kΩOUTPUT: 1 MHz , 2 V square wave into 6k

• REMOTE RESET: By grounding center of BNC back panel.

• ACCURACY : ± 1 count ± 1 timebaseaccuracy

• READSIN : MHz or GHz with positional decimal point .

• IMPEDANCE: 20kΩ each line.

Page 22: Digital frequency meter

• BCD REFERENCE LEVELS:

GROUND: 5 V , 1kΩ source.

PRINT COMMAND: 0 V to 10 V step, dc coupled.

HOLD-OFF REQUIREMENTS:

+15 V max+ 2.5 V min.

PRICE:

Model 5240A. $4750

Page 23: Digital frequency meter

Applications:

• Frequency Counting

• Precise frequency measurements in radar based application.

• Pressure, temperature, strain, and weight measurements with proper transducer.

• Motor speed measurements are also possible with proper arrangement.

• Microwave frequency measurements.

Page 24: Digital frequency meter

Reference:

• “Basic Electronic Instrument Handbook”, Clyde Coombs, Editor, McGraw-Hill, 1972.

• Hewlett-Packard Journal, June 1974.

• GENERAL RADIO COMPANY

model TYPE 1150·A Manual

Page 25: Digital frequency meter