    User’s Guide

    Shop online at

    DDPP6633000000AA-IIDDPP6633000000BB-II

    Digital PPanel MMeter

  • OMEGAnet® Online Serviceomega.com

    Internet [email protected]

    Servicing North America:

    U.S.A.: One Omega Drive, P.O. Box 4047ISO 9001 Certified Stamford, CT 06907-0047

    TEL: (203) 359-1660 FAX: (203) 359-7700e-mail: [email protected]

    Canada: 976 BergarLaval (Quebec) H7L 5A1, CanadaTEL: (514) 856-6928 FAX: (514) 856-6886e-mail: [email protected]

    For immediate technical or application assistance:

    U.S.A. and Canada: Sales Service: 1-800-826-6342/1-800-TC-OMEGA®Customer Service: 1-800-622-2378/1-800-622-BEST®Engineering Service: 1-800-872-9436/1-800-USA-WHEN®

    Mexico: En Español: (001) 203-359-7803 e-mail: [email protected]: (001) 203-359-7807 [email protected]

    Servicing Europe:

    Czech Republic: Frystatska 184, 733 01 Karviná, Czech RepublicTEL: +420 (0)59 6311899 FAX: +420 (0)59 6311114Toll Free: 0800-1-66342e-mail: [email protected]

    Germany/Austria: Daimlerstrasse 26, D-75392 Deckenpfronn, GermanyTEL: +49 (0)7056 9398-0 FAX: +49 (0)7056 9398-29Toll Free in Germany: 0800 639 7678e-mail: [email protected]

    United Kingdom: One Omega Drive, River Bend Technology CentreISO 9002 Certified Northbank, Irlam, Manchester

    M44 5BD United KingdomTEL: +44 (0)161 777 6611 FAX: +44 (0)161 777 6622Toll Free in United Kingdom: 0800-488-488e-mail: [email protected]

    It is the policy of OMEGA Engineering, Inc. to comply with all worldwide safety and EMC/EMIregulations that apply. OMEGA is constantly pursuing certification of its products to the European NewApproach Directives. OMEGA will add the CE mark to every appropriate device upon certification.

    The information contained in this document is believed to be correct, but OMEGA accepts no liability for any errors it contains, and reserves the right to alter specifications without notice.

    WARNING : These products are not designed for use in, and should not be used for, human applications.

  • 3

    FOUR SELECTABLE D.C. RANGES200 µA, 2 mA, 20 mA, 200 mA

    MINIMUM AND MAXIMUM DISPLAY CAPTURE

    LCD, REFLECTIVE OR RED/GREEN LED BACKLIGHTING

    0.48" (12.2 mm) HIGH DIGITS

    OPTIONAL SETPOINT OUTPUT MODULES

    OPTIONAL SERIAL COMMUNICATIONS MODULES (RS232 or RS485)

    OPERATES FROM 9 TO 28 VDC POWER SOURCE

    FRONT PANEL OR SOFTWARE PROGRAMMABLE

    DISPLAY COLOR CHANGE CAPABILITY AT SETPOINT OUTPUT

    NEMA 4X/IP65 SEALED FRONT BEZEL

    GENERAL DESCRIPTIONThe DP63000 provides the user the ultimate in flexibility, from its complete

    user programming to the optional setpoint control and communicationcapability. This unit accepts a DC Current input signal and provides a display inthe desired unit of measure. The meter also features minimum and maximumdisplay capture, display offset, units indicator, and programmable user input.The display can be toggled either manually or automatically between theselected displays.

    The DP63000 display has 0.48" (12.2 mm) high digits. The LCD is availablein two versions, reflective and red/green backlight. The backlight version is userselectable for the desired color and also has variable display intensity.

    The capability of the DP63000 can be easily expanded with the addition ofoption modules. Setpoint capability is field installable with the addition of thesetpoint output modules. Serial communications capability for RS232 or RS485is added with a serial option module.

    The DP63000 can be powered from an optional Power Supply (ModelNumber DP6-MLPS1), that attaches directly to the back of a DP63000. TheDP6-MLPS1 is powered from 85 to 250 VAC and provides up to 400 mA todrive the unit and sensors.

    CURRENTThis unit is the DC Current meter. It features 4 current input ranges, that are

    selected by the user via a programming jumper and software input rangeselection. The ranges consist of following: 200 µA, 2 mA, 20 mA, or 200 mA.Users should select the appropriate current range that covers their maximumsignal input.

    SAFETY SUMMARYAll safety related regulations, local codes and instructions that appear in this

    literature or on equipment must be observed to ensure personal safety and toprevent damage to either the instrument or equipment connected to it. Ifequipment is used in a manner not specified by the manufacturer, the protectionprovided by the equipment may be impaired.

    Do not use this meter to directly command motors, valves, or other actuatorsnot equipped with safeguards. To do so can be potentially harmful to persons orequipment in the event of a fault to the meter.

    CAUTION: Risk of Danger.Read complete instructions prior to

    installationand operation of the unit.

    CAUTION: Risk of electric shock.

    DIMENSIONS In inches (mm)Note: Recommended minimum clearance (behind the panel) for mounting clip installation is 2.15" (54.6) H x 3.00" (76.2) W.

    C US LISTEDUS LISTEDULRIND. CONT. EQ.

    51EB

  • 1. DISPLAY: 5 digit LCD 0.48" (12.2 mm) high digitsDP63000A-I: Reflective LCD with full viewing angleDP63000B-I: Transmissive LCD with selectable red or green LED backlight,viewing angle optimized. Display color change capability with output statewhen using an output module.

    2. POWER: Input voltage range is +9 to +28 VDC with short circuit and inputpolarity protection. Must use a DP6-MLPS1 or a Class 2 or SELV rated powersupply.

    3. INPUT RANGES: Jumper SelectableD.C. Currents: 200 µA, 2 mA, 20 mA, or 200 mA

    4. SIGNAL INPUTS:

    5. OVERRANGE RATINGS, PROTECTION & INDICATION:9 to 28 VDC power circuit is not isolated from the signal circuit.Input Overrange Indication: “OLOL”.Input Underrange Indication: “ULUL”.Display Overrange/Underrange Indication: “.....”/“-.....”

    6. DISPLAY RESPONSE TIME: 500 msec min.7. NORMAL MODE REJECTION: 60 dB 50/60 Hz8. USER INPUT (USR): Programmable input. Connect terminal to common

    (USR COMM) to activate function. Internal 10KΩ pull-up resistor to +9 to28 VDC.Threshold Levels: VIL = 1.0 V max; VIH = 2.4 V min; VMAX = 28 VDC Response Time: 5 msec typ.; 50 msec debounce (activation and release)

    9. MEMORY: Nonvolatile E2PROM memory retains all programmingparameters and max/min values when power is removed.

    10. ENVIRONMENTAL CONDITIONS:Operating Temperature Range for DP63000A-I: -35 to 75°COperating Temperature Range for DP63000B-I depends on display color

    and intensity level as per below:

    Storage Temperature: -35 to 85°COperating and Storage Humidity: 0 to 85% max. relative humidity (non-

    condensing)

    Vibration According to IEC 68-2-6: 5 to 500 Hz, in X, Y, Z direction for 1.5hours, 5 g’s.

    Shock According to IEC 68-2-27: Operational 30 g, 11 msec in 3 directionsAltitude: Up to 2000 meters

    11. CONNECTIONS: Wire clamping screw terminalsWire Strip Length: 0.3" (7.5 mm)Wire Gage: 30-14 AWG copper wireTorque: 5 inch-lbs (0.565 N-m) max.

    12. CONSTRUCTION: This unit is rated for NEMA 4X/IP65 requirements forindoor use. Installation Category I, Pollution Degree 2. High impact plasticcase with clear viewing window. Panel gasket and mounting clip included.

    13. CERTIFICATIONS AND COMPLIANCES:SAFETYUL Recognized Component, File #E313607, UL61010A-1, CSA22.2 No. 61010-1

    Recognized to U.S. and Canadian requirements under the ComponentRecognition Program of Underwriters Laboratories, Inc.

    UL Listed, File # E313547, UL508, CSA C22.2 No. 14-M95LISTED by Und. Lab. Inc. to U.S. and Canadian safety standardsIEC 61010-1, EN 61010-1: Safety requirements for electrical equipment

    for measurement, control, and laboratory use, Part 1.IP65 Enclosure rating (Face only), IEC 529Type 4X Indoor Enclosure rating (Face only), UL50

    ELECTROMAGNETIC COMPATIBILITYEmissions and Immunity to EN 61326: Electrical Equipment for Measurement,

    Control and Laboratory use.

    Notes:1. Criterion A: Normal operation within specified limits.

    14. WEIGHT: 3.2 oz (100 g)

    4

    GENERAL METER SPECIFICATIONS

    MODELNO. DISPLAY COLOR

    INPUT CURRENT@ 9 VDC WITH

    DP6-RLY0

    DP63000A-I --- 10 mA 40 mADP63000B-I Red (max intensity) 85 mA 115 mADP63000B-I Green (max intensity) 95 mA 125 mA

    TEMPERATURE1 & 2 -35 to 75°C

    3 -35 to 70°C4 -35 to 60°C5 -35 to 50°C

    1 & 2 -35 to 75°C3 -35 to 65°C

    54

    -35 to 35°C-35 to 50°C

    INTENSITY LEVEL

    Class AEN 55011EmissionsEmissions:

    30 A/m

    3 V/rms Criterion A

    Criterion A

    EN 61000-4-8

    EN 61000-4-6RF conducted interference

    1 kV L-L, Criterion AEN 61000-4-5Surge 1 kV signal2 kV powerCriterion AEN 61000-4-4Fast transients (burst)

    2 kV L&N-E power

    10 V/mCriterion AEN 61000-4-3Electromagnetic RF fields8 kV air discharge4 kV contact dischargeCriterion AEN 61000-4-2Electrostatic discharge

    Immunity to Industrial Locations:

    Power frequency magnetic fields

    Green Display

    Red Display

    INPUT CURRENT@ 9 VDC WITHOUT

    DP6-RLY0

    INPUT RANGE

    ACCURACY@23 °C, less than 85% RH

    INPUT IMPEDANCE

    MAX INPUT

    SIGNALRESOLUTION TEMP.COEFFICIENT

    200 µA 0.1% of span 1.111 KΩ 15 mA 10 nA 70 ppm / °C2 mA 0.1% of span 50 mA .1 µA 70 ppm / °C

    20 mA 0.1% of span 11 Ω 150 mA 1 µA 70 ppm / °C200 mA 0.1% of span 1 Ω 500 mA 10 µA 70 ppm / °C

    111 Ω

    PART NUMBER INFORMATION

    DP6-SNK0Dual Sinking Open Collector Output card

    DP6-SOFTDP6-485-CABLERS485 Programming Cable (DB9-RJ11)

    DP63000B-I

    PC Configuration Software for Windows 98, ME, 2000, XP *

    DC Current Meter with backlight display

    DP6-232-CABLE

    DP6-RLY0

    DP6-MLPS1Micro-Line Power Supply, 85 to 250 VAC

    Single Relay Output Card

    RS232 Programming Cable (DB9-RJ11)

    DP6-COM2RS232 Serial Communications CardDP6-COM1RS485 Serial Communications Card

    DP63000A-IDC Current Meter with reflective display

    PART NUMBERDESCRIPTION

  • 5

    ADDING OPTION CARDSThe DP63000 meters can be fitted with optional output cards and/or serial

    communications cards. The details for the plug-in cards can be reviewed in thespecification section below. The plug-in cards, that are sold separately, can beinstalled initially or at a later date.

    WARNING: Disconnect all power to the unit beforeinstalling Plug-in card.

    Note: Measurement errors may occur if signal input common is shared withanother circuit common (ie, serial common, Dual Sinking Output option card,or Power Supply common) on multiple units.

    SINGLE RELAY CARDType: Single FORM-C relayIsolation To Sensor & User Input Commons: 1400 Vrms for 1 min.

    Working Voltage: 150 VrmsContact Rating: 1 amp @ 30 VDC resistive; 0.3 amp @ 125 VAC resistiveLife Expectancy: 100,000 minimum operationsResponse Time:

    Turn On Time: 4 msec max.Turn Off Time: 4 msec max.

    DUAL SINKING OUTPUT CARDType: Non-isolated switched DC, N Channel open drain MOSFETCurrent Rating: 100 mA max. VDS ON: 0.7 V @ 100 mAVDS MAX: 30 VDCOffstate Leakage Current: 0.5 mA max.

    RS485 SERIAL COMMUNICATIONS CARDType: RS485 multi-point balanced interface (non-isolated)Baud Rate: 300 to 38.4kData Format: 7/8 bits; odd, even, or no parityBus Address: 0 to 99; max 32 meters per lineTransmit Delay: Selectable (refer to DP6-COM bulletin)

    RS232 SERIAL COMMUNICATIONS CARDType: RS232 half duplex (non-isolated)Baud Rate: 300 to 38.4kData Format: 7/8 bits; odd, even, or no parity

    OPTIONAL PLUG-IN CARDS

    1.0 INSTALLING THE METERINSTALLATION

    The meter meets NEMA 4X/IP65 requirements when properly installed. Theunit is intended to be mounted into an enclosed panel. Prepare the panel cutoutto the dimensions shown. Remove the panel latch from the unit. Slide the panelgasket over the rear of the unit to the back of the bezel. The unit should beinstalled fully assembled. Insert the unit into the panel cutout.

    While holding the unit in place, push the panel latch over the rear of the unitso that the tabs of the panel latch engage in the slots on the case. The panel latchshould be engaged in the farthest forward slot possible. To achieve a proper seal,tighten the latch screws evenly until the unit is snug in the panel (Torque toapprox. 28 to 36 in-oz [0.202 to 0.26 N-m]). Do not over-tighten the screws.

    INSTALLATION ENVIRONMENTThe unit should be installed in a location that does not exceed the operating

    temperature and provides good air circulation. Placing the unit near devices thatgenerate excessive heat should be avoided.

    The bezel should only be cleaned with a soft cloth and neutral soap product.Do NOT use solvents. Continuous exposure to direct sunlight may accelerate theaging process of the bezel.

    Do not use tools of any kind (screwdrivers, pens, pencils, etc.) to operate thekeypad of the unit.

    2.0 SETTING THE JUMPERSINPUT RANGE JUMPER

    This jumper is used to select the proper input range. The input range selectedin programming must match the jumper setting. Select a range that is highenough to accommodate the maximum signal input to avoid overloads. Toaccess the jumper, remove the rear cover of the meter.

    Warning: Exposed line voltage exists on the circuit boards. Removeall power to the meter and load circuits before accessing inside ofthe meter.

    REMOVING THE REAR COVERTo remove the rear cover, locate the cover locking tab below the 2nd and 3rd

    input terminals. To release the tab, insert a small, flat blade screwdriver betweenthe tab and the plastic wall below the terminals. Inserting the screwdriver willprovide enough pressure to release the tab locks. To replace the cover, align thecover with the input terminals and press down until the cover snaps into place.

  • WIRING OVERVIEWElectrical connections are made via screw-clamp terminals located on the

    back of the meter. All conductors should conform to the meter’s voltage andcurrent ratings. All cabling should conform to appropriate standards of goodinstallation, local codes and regulations. It is recommended that the powersupplied to the meter (DC or AC) be protected by a fuse or circuit breaker.

    Strip the wire, leaving approximately 0.3" (7.5 mm) bare lead exposed(stranded wires should be tinned with solder.) Insert the lead under the correctscrew-clamp terminal and tighten until the wire is secure. (Pull wire to verifytightness.) Each terminal can accept up to one #14 AWG (2.55 mm) wire, two#18 AWG (1.02 mm), or four #20 AWG (0.61 mm).

    EMC INSTALLATION GUIDELINESAlthough this meter is designed with a high degree of immunity to Electro-

    Magnetic Interference (EMI), proper installation and wiring methods must befollowed to ensure compatibility in each application. The type of the electricalnoise, source or coupling method into the meter may be different for variousinstallations. The meter becomes more immune to EMI with fewer I/Oconnections. Cable length, routing, and shield termination are very importantand can mean the difference between a successful or troublesome installation.Listed below are some EMC guidelines for successful installation in anindustrial environment.1. The meter should be mounted in a metal enclosure, which is properly

    connected to protective earth.2. Use shielded (screened) cables for all Signal and Control inputs. The shield

    (screen) pigtail connection should be made as short as possible. Theconnection point for the shield depends somewhat upon the application.Listed below are the recommended methods of connecting the shield, in orderof their effectiveness.a. Connect the shield only at the panel where the unit is mounted to earth

    ground (protective earth).b. Connect the shield to earth ground at both ends of the cable, usually when

    the noise source frequency is above 1 MHz.

    c. Connect the shield to common of the meter and leave the other end of theshield unconnected and insulated from earth ground.

    3. Never run Signal or Control cables in the same conduit or raceway with ACpower lines, conductors feeding motors, solenoids, SCR controls, andheaters, etc. The cables should be ran in metal conduit that is properlygrounded. This is especially useful in applications where cable runs are longand portable two-way radios are used in close proximity or if the installationis near a commercial radio transmitter.

    4. Signal or Control cables within an enclosure should be routed as far as possiblefrom contactors, control relays, transformers, and other noisy components.

    5. In extremely high EMI environments, the use of external EMI suppressiondevices, such as ferrite suppression cores, is effective. Install them on Signaland Control cables as close to the unit as possible. Loop the cable through thecore several times or use multiple cores on each cable for additional protection.Install line filters on the power input cable to the unit to suppress power lineinterference. Install them near the power entry point of the enclosure. Thefollowing EMI suppression devices (or equivalent) are recommended:

    Ferrite Suppression Cores for signal and control cables:Fair-Rite # 0443167251TDK # ZCAT3035-1330ASteward # 28B2029-0A0

    Line Filters for input power cables:Schaffner # FN610-1/07Schaffner # FN670-1.8/07Corcom # 1 VR3

    Note: Reference manufacturer’s instructions when installing a line filter.6. Long cable runs are more susceptible to EMI pickup than short cable runs.

    Therefore, keep cable runs as short as possible.7. Switching of inductive loads produces high EMI. Use of snubbers across

    inductive loads suppresses EMI.

    4.0 WIRING THE METER

    4.1 POWER WIRINGDC Power+9 to +28 VDC: +VDCPower Common: -VDC

    CAUTION: 9 to 28 VDCpower circuit is not isolatedfrom the signal circuit.

    4.2 USER INPUT WIRINGSinking Logic USR COMMUSR

    The user input of the meter isinternally pulled up to +9 to +28 Vwith 10 K resistance. The input isactive when it is pulled low (

  • 7

    4.3 INPUT WIRINGCAUTION: Power input common is NOT isolated from user input

    common. In order to preserve the safety of the meter application, thepower input common must be suitably isolated from hazardous live earthreferenced voltage; or input common must be at protective earth ground

    potential. If not, hazardous voltage may be present at the User Inputs and User InputCommon terminals. Appropriate considerations must then be given to the potential ofthe user input common with respect to earth ground; and the common of the plug-incards with respect to input common.

    Before connecting signal wires, the Input Range Jumper should be verified for properposition.

    4.4 SETPOINT (OUTPUT) WIRING

    4.5 SERIAL COMMUNICATION WIRING

    SINGLE SETPOINT RELAY PLUG-IN CARD

    SERIAL COMMUNICATIONS PLUG-IN CARD RJ11 CONNECTOR PIN OUTS

    Current Signal(self powered)

    ELECTRICAL CONNECTIONS

    DUAL SETPOINT N-FET OPEN DRAIN PLUG-IN CARD ELECTRICAL CONNECTIONS

    JUMPERPOSITION

    MAX INPUTCURRENT

    200 µA 15 mA

    200 mA

    20 mA

    2 mA

    500 mA

    150 mA

    50 mA

  • 8

    5.0 REVIEWING THE FRONT BUTTONS AND DISPLAY

    PROGRAMMING MODE ENTRY (SEL BUTTON)It is recommended all programming changes be made off line, or before

    installation. The meter normally operates in the Display Mode. No parameterscan be programmed in this mode. The Programming Mode is entered bypressing and holding the SEL button. If it is not accessible then it is locked byeither a security code, or a hardware lock.

    MODULE ENTRY (SEL & RST BUTTONS)The Programming Menu is organized into separate modules. These modules

    group together parameters that are related in function. The display will alternatebetween Pro and the present module. The RST button is used to select the desiredmodule. The displayed module is entered by pressing the SEL button.

    MODULE MENU (SEL BUTTON)Each module has a separate module menu (which is shown at the start of each

    module discussion). The SEL button is pressed to advance to a particularparameter to be changed, without changing the programming of precedingparameters. After completing a module, the display will return to Pro NO.Programming may continue by accessing additional modules.

    SELECTION / VALUE ENTRYFor each parameter, the display alternates between the present parameter and

    the selections/value for that parameter. The RST button is used to move throughthe selections/values for that parameter. Pressing the SEL button, stores andactivates the displayed selection/value. This also advances the meter to the nextparameter.

    For numeric values, press the RST button to access the value. The right handmost digit will begin to flash. Pressing the RST button again increments thedigit by one or the user can hold the RST button and the digit will automaticallyscroll. The SEL button will advance to the next digit. Pressing and holding theSEL button will enter the value and move to the next parameter.

    PROGRAMMING MODE EXIT (SEL BUTTON)The Programming Mode is exited by pressing the SEL button with Pro NO

    displayed. This will commit any stored parameter changes to memory andreturn the meter to the Display Mode. (If power loss occurs before returning tothe Display Mode, verify recent parameter changes.)

    PROGRAMMING TIPSIt is recommended to start with Module 1 and proceed through each module in

    sequence. When programming is complete, it is recommended to record theparameter programming and lock out parameter programming with the user inputor programming security code.

    FACTORY SETTINGSFactory Settings may be completely restored in Module 2. This is useful

    when encountering programming problems.

    ALTERNATING SELECTION DISPLAYIn the explanation of the modules, the following dual display with arrows will

    appear. This is used to illustrate the display alternating between the parameteron top and the parameter’s Factory Setting on the bottom. In most cases,selections and values for the parameter will be listed on the right.

    6.0 PROGRAMMING THE METEROVERVIEW

    PROGRAMMING MENU

    Indicates Program Mode Alternating Display

    Factory Settings are shown.

    Parameter

    Selection/Value

    BUTTON DISPLAY MODE OPERATION ENTERING PROGRAM MODE PROGRAMMING MODE OPERATIONSEL Index display through enabled values Press and hold for 2 seconds to activate Store selected parameter and index to next parameter

    RST Resets values (MIN/MAX) or outputs Advances through the program menuIncrements selected parameter value or selection

    OPERATING MODE DISPLAY DESIGNATORSMAX - Maximum display capture valueMIN - Minimum display capture value

    “1” - To the right of the display indicates setpoint 1 output activated.“2” - To the right of the display indicates setpoint 2 output activated.

    Pressing the SEL button toggles the meter through the selected displays. If display scroll is enabled, the display will toggle automatically every four seconds betweenthe enabled display values.

    N0

    USrIN

  • 9

    6.1 MODULE 1 - SIGNAL INPUT PARAMETERS (1-INP)PARAMETER MENU

    0.02A

    rAN6E

    INPUT RANGE

    Select the input range that corresponds to the external signal. This selectionshould be high enough to avoid input signal overload but low enough for thedesired input resolution. This selection and the position of the Input RangeJumper must match.

    0.000

    dECPt 0.000 0.00000.000.00

    DISPLAY DECIMAL POINT

    Select the decimal point location for the Input, MIN and MAX displays. Thisselection also affects the dSP1 and dSP2 parameters and setpoint values.

    1

    FILtr

    FILTER SETTING

    If the displayed value is difficult to read due to small process variations ornoise, increased levels of filtering will help to stabilize the display. Softwarefiltering effectively combines a fraction of the current input reading with afraction of the previous displayed reading to generate the new display.

    Filter values represent no filtering (0), up to heavy filtering (3). A value of 1for the filter uses 1/4 of the new input and 3/4 of the previous display to generatethe new display. A filter value of 2 uses 1/8 new and 7/8 previous. A filter valueof 3 uses 1/16 new and 15/16 previous.

    0,1 2 3

    10

    bANd

    FILTER BAND

    The filter will adapt to variations in the input signal. When the variationexceeds the input filter band value, the filter disengages. When the variationbecomes less than the band value, the filter engages again. This allows for astable readout, but permits the display to settle rapidly after a large processchange. The value of the band is in display units, independent of the DisplayDecimal Point position. A band setting of ‘0’ keeps the filter permanentlyengaged at the filter level selected above.

    0 to 199 display units

    KEy

    StYLE

    SCALING STYLE

    If Input Values and corresponding Display Values are known, the Key-in(KEY) scaling style can be used. This allows scaling without the presence orchanging of the input signal. If Input Values have to be derived from the actualinput signal source or simulator, the Apply (APLY) scaling style must be used.

    KEy APLY

    0.000

    INP 1

    INPUT VALUE FOR SCALING POINT 1

    For Key-in (KEY) style, enter the first Input Value using the front panel buttons.(The Input Range selection sets the decimal location for the Input Value).

    For Apply (APLY) style, the meter shows the previously stored Input Value. Toretain this value, press the SEL button to advance to the next parameter. Tochange the Input Value, press the RST button and apply the input signal to themeter. Adjust the signal source externally until the desired Input Value appears.Press the SEL button to enter the value being displayed.

    0 to 29999

    0.000

    dSP 1

    DISPLAY VALUE FOR SCALING POINT 1

    -19999 to 99999

    Enter the first Display Value by using the front panel buttons. This is the samefor KEY and APLY scaling styles. The decimal point follows the dECPt selection.

    10.000

    INP 2

    INPUT VALUE FOR SCALING POINT 2

    0 to 29999

    For Key-in (KEY) style, enter the known second Input Value using the frontpanel buttons.

    For Apply (APLY) style, the meter shows the previously stored Input Value forScaling Point 2. To retain this value, press the SEL button to advance to the nextparameter. To change the Input Value, press the RST button and apply the inputsignal to the meter. Adjust the signal source externally until the desired InputValue appears. Press the SEL button to enter the value being displayed.

    10.000

    dSP 2

    DISPLAY VALUE FOR SCALING POINT 2

    -19999 to 99999

    Enter the second Display Value by using the front panel buttons. This is thesame for KEY and APLY scaling styles.

    200uA 200.00 µA0.002A 2.0000 mA

    20.000 mA200.00 mA

    0.02A

    0.2A

    SELECTION RANGERESOLUTIONRANGE

    RESOLUTIONSELECTION

    0.000

    OFSEt

    DISPLAY OFFSET VALUE

    The display can be corrected with an offset value. This can be used tocompensate for signal variations or sensor errors. This value is automaticallyupdated after a Zero Display to show how far the display is offset. A value ofzero will remove the effects of offset.

    -19999 to 19999

    General Notes on Scaling1. When using the Apply (APLY) scaling style, input values for scaling points

    must be confined to the range limits shown.2. The same Input Value should not correspond to more than one Display Value.

    (Example: 20 mA can not equal 0 and 20.)3. For input levels beyond the programmed Input Values, the meter extends the

    Display Value by calculating the slope from the two coordinate pairs (INP1 /dSP1 & INP2 / dSP2).

  • Entering Code 66 will overwrite all user settings withthe factory settings. The meter will display rESEt and thenreturn to CodE 00. Press the SEL button to exit the module.

    10

    6.2 MODULE 2 - SECONDARY FUNCTION PARAMETERS (2-SEC)PARAMETER MENU

    MIN DISPLAY ENABLE

    USER INPUT ASSIGNMENT

    Select the value(s) to which the User Input Function is assigned. The UserInput Assignment only applies if a selection of reset, display hold, or print andreset is selected in the User Input Function menu.

    MODEDISPLAY

    No FunctionNO

    DESCRIPTION

    User Input disabled.

    Program Mode Lock-outP-Loc

    Zero Input (Edge triggered)ZErO

    Zero the Input Display value causingDisplay Reading to be Offset.

    COLOr

    d-SEL

    d-LEV

    d-HLd

    rESEtResets the assigned value(s) to thecurrent input value.

    HI-LO HI

    dSPLO

    See Programming Mode Access chart(Module 3).

    YESNO

    MODE DESCRIPTIONDISPLAY

    Setpoint 1 and 2 Reset

    Setpoint 1 Reset

    Setpoint 2 Reset

    Print and Reset

    Reset both setpoint 1 and 2 outputs.

    Resets setpoint 1 output.

    Resets setpoint 2 output.

    Same as Print Request followed by amomentary reset of the assigned value(s).

    rSt12

    rSt-1

    rSt-2

    P-r5t

    Print RequestSerial transmit of the active parametersselected in the Print Options menu(Module 5).

    Print

    2.0

    HI-t

    NO

    LO-En

    NO

    FCS

    MAX CAPTURE DELAY TIME

    When the Input Display is above the present MAX value for the entereddelay time, the meter will capture that display value as the new MAX reading.A delay time helps to avoid false captures of sudden short spikes.

    2.0

    LO-tMIN CAPTURE DELAY TIME

    When the Input Display is below the present MIN value for the entered delaytime, the meter will capture that display value as the new MIN reading. A delaytime helps to avoid false captures of sudden short spikes.

    0.0 to 999.9 sec.

    USER INPUT FUNCTION

    NO

    USrIN

    dSP

    U-ASN

    NO

    HI-EnMAX DISPLAY ENABLE

    0.0 to 999.9 sec.

    Select yES to perform either of the Factory Service Operations shown below.

    FACTORY SERVICE OPERATIONS

    yESNO

    YESNO

    Enables the Maximum Display Capture capability.

    Enables the Minimum Display Capture capability.

    The unit uses stored current calibration values toprovide accurate current measurements. Over time, theelectrical characteristics of the components inside the unitwill slowly change with the result that the stored

    calibration values no longer accurately define the input circuit. For mostapplications, recalibration every 1 to 2 years should be sufficient.

    Calibration of the unit involves a current calibration which should only beperformed by individuals experienced in calibrating electronic equipment.Allow 30 minute warm up before performing any calibration related procedure.The following procedures should be performed at an ambient temperature of 15to 35 °C (59 to 95 °F).

    CAUTION: The accuracy of the calibration equipment will directly affect theaccuracy of the unit.

    Current Calibration1. Connect the negative lead of a precision DC current source with an accuracy

    of 0.01% or better to the COMM terminal. Leave the positive lead of the DCcurrent source unconnected.

    2. With the display at CodE 48, press and hold the SEL button for 2 seconds. Unitwill display CAL NO

    3. Press the RST button to select the range to be calibrated.4. Press the SEL button. Display reads 0.0A5. With the positive lead of the DC current source unconnected, press SEL.

    Display reads CALC for about 8 seconds.6. When the display reads the selected range, connect the positive lead of the DC

    current source to INP+ and apply full-scale input signal for the range. (Note:For 200 mA range, apply 100 mA as indicated on the display.)

    7. Repeat steps 3 through 6 for each input range to be calibrated. When displayreads CAL NO, press the SEL button to exit calibration.

    CALIBRATION

    48

    CodE

    66

    CodE

    RESTORE FACTORY DEFAULT SETTINGS

    Backlight Color (Edge Triggered)

    Display Select (Edge Triggered)Display Intensity Level (Edge Triggered)

    Display Hold

    Reset (Edge triggered)

    Change backlight color with eachactivation (backlight version only).

    Advance once for each activation.

    Increase intensity one level for eachactivation (backlight version only).

    Holds the assigned display, but all othermeter functions continue as long asactivated (maintained action).

    Entering Code 50 will display the version (x.x) of themeter. The display the returns to CodE 00. Press the SELbutton to exit the module. 50

    CodE

    VIEW VERSION DISPLAY

  • 11

    The Security Code determines the programming mode and the accessibilityof programming parameters. This code can be used along with the ProgramMode Lock-out (P-Loc) in the User Input Function parameter (Module 1).

    Two programming modes are available. Full Programming mode allows allparameters to be viewed and modified. Quick Programming mode permits onlythe Setpoint values to be modified, but allows direct access to these valueswithout having to enter Full Programming mode.

    Programming a Security Code other than 0, requires this code to be enteredat the CodE prompt in order to access Full Programming mode. Depending on thecode value, Quick Programming may be accessible before the CodE promptappears (see chart).

    PROGRAMMING SECURITY CODE

    000 to 999

    USER INPUTFUNCTION

    USER INPUTSTATE

    SECURITYCODE

    MODE WHEN “SEL”BUTTON IS PRESSED

    FULL PROGRAMMINGMODE ACCESS

    0 Full Programming Immediate Access

    not P-Loc ______ 1-99 Quick ProgrammingAfter Quick Programming

    with correct code entryat CodE prompt *

    100-999 CodE promptWith correct code entry

    at CodE prompt *

    0 Programming Lock No Access

    Active 1-99 Quick Programming No AccessP-Loc

    100-999 CodE promptWith correct code entry

    at CodE prompt *

    Not Active 0-999 Full Programming Immediate Access

    * Entering Code 222 allows access regardless of security code.

    6.3 MODULE 3 - DISPLAY AND FRONT PANEL BUTTONPARAMETERS (3-dSP)

    PARAMETER MENU

    The yES selection allows the SEL button to toggle through the enableddisplays.

    FRONT PANEL DISPLAY SELECT ENABLE (SEL)

    This selection allows the RST button to reset the selected value(s).

    FRONT PANEL RESET ENABLE (RST)

    The yES selection allows the display to automatically scroll through theenabled displays. The scroll rate is every 4 seconds. This parameter only appearswhen the MAX or MIN displays are enabled.

    DISPLAY SCROLL ENABLE

    NOyES

    dSPNO

    HI

    LO

    HI-LO

    NOyES

    This parameter activates the Units Indicator on the display. There are twomethods of selecting the Indicator. List will present a group of Unitspreprogrammed into the meter. Segments allows the user to choose which of thesegments should light.

    UNITS INDICATOR SELECTION

    LISt SEGSOFF

    Enter the desired display color, red or green. This parameter is active forbacklight units only.

    DISPLAY COLOR (BACKLIGHT UNIT ONLY)

    6rnrEd

    Enter the desired Display Intensity Level (1-5). The display will actively dimor brighten as levels are changed. This parameter is active for backlight units only.

    DISPLAY INTENSITY LEVEL (BACKLIGHT UNIT ONLY)

    1 to 5

    1

    dSP-t

    yES

    SEL

    dSP

    rSt

    NO

    ScroL

    This parameter enables the RST button or user input to zero the input displayvalue, causing the display reading to be offset.

    Note: For this parameter to operate, the RST button or User Input being usedmust be set to dSP and the Input value must be displayed. If these conditions arenot met, the display will not zero.

    ZERO DISPLAY WITH DISPLAY RESET

    NOyESNO

    ZErO

    OFF

    UNItS

    rEd

    COLOr

    5

    d-LEV

    000

    CodE

    DISPLAY UPDATE TIME

    This parameter sets the display update time in seconds.

    10.5 2 seconds

  • Enter the reset action of the output. See figure for details.

    Auto = Automatic action; This action allows the output to automatically reset offat the trigger points per the Setpoint Action shown in Setpoint OutputFigures. The “on” output may be manually reset (off) immediately by thefront panel RST button or user input.The output remains off until the triggerpoint is crossed again.

    LAtCH = Latch with immediate reset action; This action latches the output on atthe trigger point per the Setpoint Action shown in Setpoint Output Figures.Latch means that the output can only be turned off by the front panel RST

    12

    6.4 MODULE 4 - SETPOINT OUTPUT PARAMETERS (4-SPt)

    PARAMETER MENU

    The Setpoint Output Parameters are only active when an optional outputmodule is installed in the meter.

    Enter the setpoint (output) to be programmed. The n in the followingparameters will reflect the chosen setpoint number. After the chosen setpointis completely programmed, the display will return to SPSEL. Repeat steps foreach setpoint to be programmed. Select NO to exit the module. The number ofsetpoints available is setpoint output card dependent.

    SETPOINT SELECT

    SP-2SP-1NO

    NO

    SPSEL

    2

    HYS-n

    HYSTERESIS VALUE

    0.0

    tOF-n

    0.0

    tON-n

    OFF TIME DELAY

    ON TIME DELAY

    OUTPUT RESET ACTION

    LAtCHAuto L-dLY

    0.0 to 599.9 Sec

    0.0 to 599.9 Sec

    Auto

    rSt-n

    1 to 59999

    Enter desired hysteresis value. See Setpoint Output Figures for visualexplanation of how setpoint output actions (balanced and unbalanced) areaffected by the hysteresis. When the setpoint is a control output, usuallybalanced hysteresis is used. For alarm applications, usually unbalancedhysteresis is used. For unbalanced hysteresis modes, the hysteresis functions onthe low side for high acting setpoints and functions on the high side for lowacting setpoints.Note: Hysteresis eliminates output chatter at the switch point, while time delay

    can be used to prevent false triggering during process transient events.

    Enter the time value in seconds that the output is delayed from turning onafter the trigger point is reached. A value of 0.0 allows the meter to update theoutput status per the response time listed in the Specifications.

    Enter the time value in seconds that the output is delayed from turning offafter the trigger point is reached. A value of 0.0 allows the meter to update theoutput status per the response time listed in the Specifications.

    Select YES to enable Setpoint 2 and access the setup parameters. If NO isselected, the unit returns to SPSEL and setpoint 2 is disabled.

    SETPOINT 2 ENABLE

    NOYES

    NO

    Enb-2

    Enter the action for the selected setpoint (output). See Setpoint OutputFigures for a visual detail of each action.

    SETPOINT ACTION

    HI-Ub

    Act-n LO-UbHI-UbLO-bLHI-bL

    LO-Ub =HI-Ub =LO-bL =HI-bL =

    Low Acting, with unbalanced hysteresisHigh Acting, with unbalanced hysteresisLow Acting, with balanced hysteresisHigh Acting, with balanced hysteresis

    High Acting (Balanced Hys) = HI-bL

    Low Acting (Unbalanced Hys) = LO-Ub

    Low Acting (Balanced Hys) = LO-bL

    High Acting (Unbalanced Hys) = HI-Ub

    SETPOINT VALUE

    Enter the desired setpoint value. The decimal point position for the setpointand hysteresis values follow the selection set in Module 1.

    -19999 to 9999910000

    SPt-n

  • 13

    6.5 MODULE 5 - SERIAL SETUP PARAMETERS (5-SSEr)

    The Serial Setup Parameters are only active when the optional RS232 or RS485 serial communications module is installed in the meter. Refer to the DP6-COMbulletin for complete details on DP63000 serial communications.

    PARAMETER MENU

    button or user input manual reset, serial reset command or meter powercycle. When the user input or RST button is activated (momentary action),the corresponding “on” output is reset immediately and remains off until thetrigger point is crossed again. (Previously latched alarms will be off if powerup Display Value is lower than setpoint value.)

    L-dLY = Latch with delay reset action; This action latches the output on at thetrigger point per the Setpoint Action shown in Setpoint Output Figures. Latchmeans that the output can only be turned off by the front panel RST buttonor user input manual reset, serial reset command or meter power cycle. Whenthe user input or RST button is activated (momentary action), the meterdelays the event until the corresponding “on” output crosses the trigger offpoint. (Previously latched outputs are off if power up Display Value is lowerthan setpoint value. During a power cycle, the meter erases a previous L-dLYreset if it is not activated at power up.)

    Setpoint Output Reset Actions

    This parameter enables the RST button or user input to reset the output whenthe display is reset.

    Note: For this parameter to operate, the RST button or User Input being usedmust be set to dSP and the Input value must be displayed. If these conditions arenot met, the output will not reset.

    YES

    rEn-n

    OUTPUT RESET WITH DISPLAY RESET

    YESNO

    When YES, the output is disabled (after a power up) until the trigger point iscrossed. Once the output is on, the output operates normally per the SetpointAction and Output Reset Action.

    NO

    Stb-n

    STANDBY OPERATION

    YESNO

    This parameter enables the backlight DP63000B-I to switch the backlight colorwhen the output state changes. This parameter is only active for the backlightversion.

    CHANGE DISPLAY COLOR w/OUTPUT STATE

    NO

    ChC-nYESNO

  • PROGRAMMING QUICK OVERVIEW

    Pre

    ss a

    nd h

    old

    SEL

    butto

    nto

    ent

    er P

    rogr

    amm

    ing

    Mod

    e.

  • WARRANTY/DISCLAIMEROMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for aperiod of 13 months from date of purchase. OMEGA’s WARRANTY adds an additional one (1) month graceperiod to the normal one (1) year product warranty to cover handling and shipping time. This ensuresthat OMEGA’s customers receive maximum coverage on each product. If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer ServiceDepartment will issue an Authorized Return (AR) number immediately upon phone or written request. Uponexamination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no charge.OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including butnot limited to mishandling, improper interfacing, operation outside of design limits, improper repair, orunauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tamperedwith or shows evidence of having been damaged as a result of excessive corrosion; or current, heat,moisture or vibration; improper specification; misapplication; misuse or other operating conditions outsideof OMEGA’s control. Components in which wear is not warranted, include but are not limited to contactpoints, fuses, and triacs.OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGAneither assumes responsibility for any omissions or errors nor assumes liability for anydamages that result from the use of its products in accordance with information provided byOMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by thecompany will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES ORREPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OFTITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITYAND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OFLIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability ofOMEGA with respect to this order, whether based on contract, warranty, negligence,indemnification, strict liability or otherwise, shall not exceed the purchase price of thecomponent upon which liability is based. In no event shall OMEGA be liable for consequential,incidental or special damages.CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “BasicComponent” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medicalapplications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity,medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forthin our basic WARRANTY/DISCLAIMER language, and, additionally, purchaser will indemnify OMEGA andhold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) insuch a manner.

    RETURN REQUESTS/INQUIRIESDirect all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORERETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR)NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSINGDELAYS). The assigned AR number should then be marked on the outside of the return package and on anycorrespondence.The purchaser is responsible for shipping charges, freight, insurance and proper packaging to preventbreakage in transit.

    OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords ourcustomers the latest in technology and engineering.OMEGA is a registered trademark of OMEGA ENGINEERING, INC.© Copyright 2006 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied,reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without theprior written consent of OMEGA ENGINEERING, INC.

    FOR WARRANTY RETURNS, please have thefollowing information available BEFORE contacting OMEGA:1. Purchase Order number under which the product

    was PURCHASED,2. Model and serial number of the product under

    warranty, and3. Repair instructions and/or specific problems

    relative to the product.

    FOR NON-WARRANTY REPAIRS, consult OMEGAfor current repair charges. Have the followinginformation available BEFORE contacting OMEGA:1. Purchase Order number to cover the COST of the

    repair,2. Model and serial number of the product, and3. Repair instructions and/or specific problems

    relative to the product.

  • Where Do I Find Everything I Need forProcess Measurement and Control?

    OMEGA…Of Course!Shop online at omega.com

    TEMPERATURE] Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies] Wire: Thermocouple, RTD & Thermistor] Calibrators & Ice Point References] Recorders, Controllers & Process Monitors] Infrared Pyrometers

    PRESSURE, STRAIN AND FORCE] Transducers & Strain Gages] Load Cells & Pressure Gages] Displacement Transducers] Instrumentation & Accessories

    FLOW/LEVEL] Rotameters, Gas Mass Flowmeters & Flow Computers] Air Velocity Indicators] Turbine/Paddlewheel Systems] Totalizers & Batch Controllers

    pH/CONDUCTIVITY] pH Electrodes, Testers & Accessories] Benchtop/Laboratory Meters] Controllers, Calibrators, Simulators & Pumps] Industrial pH & Conductivity Equipment

    DATA ACQUISITION] Data Acquisition & Engineering Software] Communications-Based Acquisition Systems] Plug-in Cards for Apple, IBM & Compatibles] Datalogging Systems] Recorders, Printers & Plotters

    HEATERS] Heating Cable] Cartridge & Strip Heaters] Immersion & Band Heaters] Flexible Heaters] Laboratory Heaters

    ENVIRONMENTALMONITORING AND CONTROL] Metering & Control Instrumentation] Refractometers] Pumps & Tubing] Air, Soil & Water Monitors] Industrial Water & Wastewater Treatment] pH, Conductivity & Dissolved Oxygen Instruments M4487/0108