doppler ultrasonic technique

Upload: waseemjabarin

Post on 30-May-2018

215 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/14/2019 Doppler Ultrasonic Technique

    1/11

    MEDICAL ENGNEERING SYSTMS

    Doppler ultrasonic technique

    Doppler ultrasound technique, was originally applied in the medical fieldand dates back more then 30 years. The use of pulsed emissions hasextended this technique to other fields and has open the way to newmeasuring techniques in fluid dynamics. The term "Doppler ultrasoundvelocimetry" implies that the velocity is measured by finding theDoppler frequency in the received signal, as it is the case in LaserDoppler velocimetry. In fact, in ultrasonic pulsed Doppler velocimetry,this is never the case. Velocities are derived from shifts in positionsbetween pulses, and the Doppler effect plays a minor role.Unfortunately, many publications, even recent ones, fails to make thedistinction, resulting in erroneous system description and fallacious

    interpretation of the influence from various physical effects.

    Functioning principles of pulsed Doppler ultrasound

    In pulsed Doppler ultrasound, instead of emitting continuous ultrasonicwaves, an emitter sends periodically a short ultrasonic burst and areceiver collects continuously echoes issues from targets that may bepresent in the path of the ultrasonic beam. By sampling the incomingechoes at the same time relative to the emission of the bursts, the shiftof positions of scatterers are measured. Let assume a situation, asillustrated in the figure below, where only one particle is present along

    the ultrasonic beam.

    From the knowledge of the time delay Td between an emitted burst andthe echo issue from the particle, the depth p of this particle couldcomputed by:

  • 8/14/2019 Doppler Ultrasonic Technique

    2/11

    where c is the sound velocity of the ultrasonic wave in the liquid.If theparticle is moving at an angle q regarding the axis of the ultrasonicbeam, its velocity could be measured by computing the variation of itsdepth between two emissions separated in time by Tprf:

    The time difference (T2-T1) is always very short, most of the time lowerthan a microsecond. It is advantageous to replace this timemeasurement by a measurement of the phase shift of the received echo.

    where fe is the emitting frequency. With this information the velocity ofthe target is expressed by:

    This last equation gives the same result as the Doppler equation. Butone should always be aware that the phenomena involved are not thesame. assume that the particles are randomly distributed inside theultrasonic beam. The echoes issue from each particle are thencombined together in a random fashion, giving a random echo signal.Hopefully, a high degree of correlation exists between differentemissions. This high correlation degree is put in advance in all digitalprocessing techniques used in Signal Processing's Ultrasonic Dopplervelocimeter to extract information, such as the velocity.

    Doppler EffectINTODUCTION:

    B-mode imaging is based on the reflection and scatter of

    ultrasound at interfaces and intrinsic organic structures. Theinformation carriers that make Doppler sonography possible are the

    red blood cells (erythrocytes) inside the vessels, which move towards

    the transducer or away from it at various speed. The Doppler effect

    is named after the physicist Christian Johann Doppler (1803 1853).

    Doppler physically interpreted and mathematically expressed the

    fact that the light of stars moving towards the earth undergoes a shift

    towards blue (shorter wavelength), and that the light of stars moving

    away from the earth undergoes a shift towards red (longer

    wavelength).

  • 8/14/2019 Doppler Ultrasonic Technique

    3/11

    In a similar way, the sound waves generated from the erythrocytesand the echo signals returning to the transducer experience a slight

    frequency shift f with regard to the transmit frequency f. The

    frequency shift f depends on the magnitude and direction of the

    blood flow velocity v. The scatter echoes from within the vessels are

    weaker than the signals obtained from the interfaces of vessels,

    organs, and tissues, by a factor of 100 to 1000. For their detection

    and processing a very sensitive system technique is needed.

    The Doppler equation (forSpectral Dopplerand Color Doppler)describes the relation between f and v:

    f = 2 f/c v cos

    where c is the velocity of sound (at an average of 1540 m/s in tissue), v

    the blood flow velocity to be analyzed and the angle of incidence in

    relation to the axis of the vessel. The factor 2 takes into account that

    the Doppler effect is observed twice: once when the moving blood

  • 8/14/2019 Doppler Ultrasonic Technique

    4/11

    cells receive the transmit signal from the probe, and the second time,

    when the probe receives the echo from the blood cells.

    The frequency shift f measured by Doppler techniques in the

    following also referred to asDoppler frequency is a direct measurefor v. For a given magnitude of v, the frequency shift f will be

    higher with the selection of a higher transmit frequency f.

    f also depends on the angle of incidence: f is at its highest when the

    incident sound beam is directed parallel to the vessel axis. At

    perpendicular incidence, cos = 0, i.e., no Doppler signal is obtained.

    To calculate the velocity v from the Doppler frequency f, the angle

    must be measured in the B-mode or Color Doppler image and an

    angle correction must be made.

    The Doppler effect increases or reduces the echo signal frequency f

    coming from the vessel by f, depending on the direction of blood

    flow. That means that the echo signal frequency also contains the

    information on the direction of flow (bi-directional Doppler). For the

    evaluation of echo signals, the flow velocities towards the transducer

    are usually displayed on the positive axis in the Spectral Doppler

    curve (encoded red when Color Doppler is used). They are displayed

    on the negative axis and encoded blue in Color Doppler, when theflow is directed away from the probe.

    The Doppler frequencies f measured within the frequency range of

    2 10 MHz and at physiological flow velocities of several mm/s up to

    1 2 m/s, are in the audible range between 50 Hz and 16 kHz and are

    thus directly accessible to the ear as audio signals.

    Advantages and limitations

    The main advantage of pulsed Doppler ultrasound is its capability tooffer spatial information associated to velocity values. Unfortunately, asthe information is available only periodically, this technique suffers fromthe Nyquist theorem. This means that a maximum velocity exists foreach pulse repetition frequency (Prf):

  • 8/14/2019 Doppler Ultrasonic Technique

    5/11

    In addition to the velocity limitation, there is a limitation in depth. Theultrasonic burst travels in the liquid at a velocity which depends on thephysical properties of the liquid. The pulse repetition frequency givesthe maximum time allowed to the burst to travel to the particle and backto the transducer. This gives a maximum depth of:

    From the above two equations, we could see that increasing the timebetween pulses (TPRF) will increase the maximum measurable depth,but will also reduce the maximum velocity which can be measured. Themaximum velocity and maximum depth are thus related according to thefollowing equation:

  • 8/14/2019 Doppler Ultrasonic Technique

    6/11

    Ultrasound scattering

    The ultrasonic waves generated by the transducer are more or lessconfined in a narrow cone. As they travel in this cone they may bereflected or scattered when they touch a particle having a different

    acoustic impedance. The acoustic impedance is defined by:

    where is the density and c the sound velocity.

    If the size of the particle is bigger than the wave length, the ultrasonicwaves are reflected and refracted by the particle. In such a case thedirection of propagation and the intensity of the ultrasonic waves areaffected. But if the size of the particle is much smaller than the wavelength an other phenomena appears, which is named scattering. In sucha case, a very small amount of the ultrasonic energy is reflected in alldirection. The intensity and the direction of propagation of the incomingwaves are practically not affected by the scattering phenomena.Ultrasonic Doppler velocimetry needs therefore particles smaller thanthe wave length.

  • 8/14/2019 Doppler Ultrasonic Technique

    7/11

    Doppler Waveforms and Vascular Resistance

    The velocity vs. time display of the Doppler spectrum or the velocity

    curves deduced from it yield information on the differences and

    changes of peripheral vascular resistance. At a constant blood

    pressure P, conclusions may be drawn on the perfusion (indirect

    method) from the relation between volume flow Q and vascular

    resistance R (see figure 2006).

    The waveform of a low resistance vesselis marked by a higher end-

    diastolic flow velocity and low pulsatility. Pulsatility is defined as therelation between systolic and end-diastolic velocity. The lower the

    end-diastolic velocity, the higher is the vascular resistance.

    The waveform of a high resistance vesselshows a low to disappearing

    end-diastolic velocity and a high pulsatility. The triphasic flow profile

    is typical of peripheral vessels in the extremities.

    From the envelope curves of the Spectral Doppler, vascular resistance

    andpulsatility indices can be deduced to obtain the actual vascularresistance .

  • 8/14/2019 Doppler Ultrasonic Technique

    8/11

    Flow Velocity and Doppler Effect

    Ultrasound imaging techniques for the detection of blood flow are

    based mainly on the fact that the pulses generated by the transducerare reflected not only by motionless organ interfaces and structures

    but also by the formations of red blood cells flowing in the vessels. If

    it were possible to measure the spatial displacement s of this

    formation within a time interval, e.g., the time 1 / PRF between two

    pulses, the flow velocity could be calculated directly, but this direct

    approach is not yet possible.

    Therefore, ultrasound imaging makes use of theDoppler effect

    (Christian Johann Doppler, 1803 1853): The frequency perceivedby an observer of light or sound traveling towards him is higher then

    the perceived frequency from that same light or sound at rest. The

    frequency from a source of light or sound moving away from the

    observer is perceived as lower. The higher the relative velocity the

    greater the difference in perceived frequencies. This frequency

    difference is calledDoppler frequency shiftorDoppler frequency.

    In ultrasound imaging, the transducer is both a source of sound

    and an observer. If the flow is directed towards the transducer, i.e.,

    the incident sound direction is opposite to the direction of flow (from

    upper right to lower left on the graphic), the pressure wave generated

    by the moving blood cells is compressed. Its wavelength is shorter

    and consequently, its frequency is higher. If the flow is directed away

    from the transducer (incedent sound direction from upper left to

    lower right on the graphic), the echo sound wave is extended, i.e., its

    frequency is lower.

    The Doppler techniques in ultrasound imaging are basedupon the measurement and evaluation of the Doppler

    frequency shift which, subsequently, makes it possible tocalculate the flow velocities that go with it

    CW- and PW-Doppler

    CW-Doppler (continuous wave) and PW-Doppler (pulsed wave) differ

    in the way of signal acquisition, but have some similarities in signal

  • 8/14/2019 Doppler Ultrasonic Technique

    9/11

    processing and presentation of results. With CW-Doppler, the

    piezoceramic elements of the (phased array) transducer are divided

    into two groups, one of them continuously transmitting while the

    other one simultaneously receives the incoming signals. With pencil

    probes, the transducer element is divided into two. Due to thecontinuous operation, it is not possible to assign an echos point of

    origin. On the other hand, very high flow velocities (e.g. a high-grade

    valvular stenosis) can be analyzed unambiguously with CW-Doppler.

    In simple Doppler units without duplex operation (i.e. without

    support by B-mode imaging), CW-Doppler further simplifies the

    detection of vessels supplied with blood.

    PW-Doppler meets the requirement for measuring flow in user

    selected areas of interest. One single group of array elements is usedboth for receiving and transmitting. They transmit sequences of

    short pulses into the body, just as in B-mode imaging. After the pulse

    has traveled to the selected sample site and back (travel time T), the

    sample gate is opened for a short period of time, TR, to receive the

    echoes. The size and depth of the sample volume in the B-mode or the

    Color Doppler image can be controlled on the monitor and are

    adjustable by the sonographer.

    The pulse travel time T determines the shortest possible time interval

    between two successive transmit pulses. Therefore thepulse repetition

    frequencyPRFfor the transmit pulse cannot be set higher than 1/T

    without jeopardizing the unambiguous depth assignment. Since the

    applicable PRF values are also in the range of the Doppler

    frequencies f, depth discrimination with the PW-Doppler is

    achieved at the price of an ambiguous evaluation of higher flow

    velocities, resulting in aliasingeffects. Also Color Doppler is a PW-

    technique to which both the Doppler equation and the restrictions

    concerning PRF are applicable.

  • 8/14/2019 Doppler Ultrasonic Technique

    10/11

    Direct Measurement of Volume Flow

    TAV mean is equivalent to the average constant flow velocity and is the

    basis for the direct, quantitative calculation of the volume flow Q:

    Q = A TAVmean

    where A = r2, the cross sectional area of the vessel at the samplingsite which is obtained from the vessel diameter 2 r measured in the B-

    mode image. When interpreting these values with regard to the

    volume flow Q, two critical sources of error must be considered:

    First, the limited measuring accuracy of the vessel diameter,

    especially in the case of very small vessels and second, the angle

    correction error which depends on the angle of incidence.

    Nota bene: TAVmax , i.e. the time averagedmaximum velocity, cannot

    be utilized for the determination of the volume flow Q, since it would

  • 8/14/2019 Doppler Ultrasonic Technique

    11/11

    overestimate it. TAVmax is frequentlyused for the indirectevaluation

    of organ perfusion.