dose reduction technique in ct scan
TRANSCRIPT
DOSE REDUCTION TECHNIQUE IN
CT SCANMohd Aiman bin Azmardi
Pegawai Pengimejan,Jabatan Pengimejan Diagnostik
Hospital Kulim11th August 2015
PRESENTATION OUTLINE Introduction
Dose in CT Scan Dose Reduction Technique
Equipment Design What is dose in CT Scan..? Acquisition Parameters
mAs ATM kVp Image Reconstruction
FBP Iterative Reconstruction
Scan Technique Recommendation Conclusion References
INTRODUCTION CT scan advancement in technology
Near to isotropic pixel – high resolutionFast scan – cover ROI in short timeAdvance post processing (e.g.: MIP, 3D
reconstruction, VRT and many more) Weakness?
DOSE IN CT SCAN CT scan lead high dose compared to
other imaging modalities 50 times riskier from plain abdomen
radiograph MSCT compared to single slice
Increase 10% and 34% of effective dose per patient
CT scan induce cancer arise rapidly ALARA
The most powerful dose reduction tool in radiation protection -justification of a study,
However, where an examination is undertaken, the emphasis must be on dose optimization, achieving the required image quality at the
lowest possible dose level. This can be approached in two ways
design of dose efficient equipment, through the optimization of scan protocols.
DOSE REDUCTION TECHNIQUE
EQUIPMENT DESIGN (TUBE DESIGN) Siemens intoduced in the SOMATOM
Sensation 64 CT-system is equipped with the 0 MHU STRATON x-ray tube.
This newly developed X-ray tube offerssignificantly reduced cooling times for
shorter interscan delays increased power reserves.
CONVENTIONAL X-RAY TUBE
SIEMENS STRATON X-RAY TUBE
DOUBLE Z-SAMPLING Double Z sampling is two overlapping
projections result doubling the scan (more data) without an increase in dose.
AdvantagesSlice widht can be reduce e.g: it is possible to
reconstruct 0.6mm slices at any pitch less than 1.5) with best image quality
Improve image qualitydecreasing slice widthand removing windmill artifacts, without
increasing the radiation dose
Z axis direction
Windmili artifact can be reduced which commonly found in MSCT images in the vicinity of sharp contrast in axial direction.
DUAL ENERGY CT Dual energy CT scans are a relatively
new form of CT scanning that use separate X-ray energies to make images. Images can be generated:by the simultaneous use of two X-ray tubes
(‘dual source’);by using an X-ray detector with separate
layers to detect two different energy ranges (‘dual layer’); or
by using a single scanner to scan twice using two different energy levels (electronic kVp switching).
Advantages CT angiography. Dual energy scans can amplify
the iodine signal of contrast agents, improving the delineation of arteries. They can also better distinguish iodine from calcium, therefore allowing better bone subtraction around vessels; for example, at the skull base.
CT of the kidney, ureter and bladder (CT KUB). Dual energy CT KUB scans can reliably distinguish urate from non-urate calculi.
CT imaging around metal implants. Dual energy CT can significantly reduce the streak artefact normally associated with metal implants and allow better visualisation; for example, around spinal rods or hip replacements.
TYPE OF DUAL ENERGY With fast kV-switching, Dual
Energy data can be acquired by rapidly switching the tube voltage between CT projections.
Disadvantages lower number of projections are
available to create each image; reduced image quality
In addition, only the kVp can be modulated between individual projections. Resulting over-exposure in the highkV projections or under-exposure in the low-kV projections
Idealized dual layer detector technology: In reality a certain amount of high and low-energy photons are registered in both layers which significantly reduces spectral separation
Disadvantages Detector not able to distinguish
between high and low energy photons. Both high and low energy photons are absorbed in both layers
The construction of this detector requires two photodiodes, which significantly increases electronic noise. Leads to inferior image quality for dual and single energy images.
Slow kV-switching: Both kV and mA are switched between half rotations of the gantry, either in sequence or in spiral modes
Disadvantages The time needed to switch from
80 kV to 140 kV and adjust the mA is typically in the order of 100 ms.
During this time, the patient is exposed to radiation that does not provide useful information.
Thus, this method does not follow the ALARA (“as low as reasonably achievable”) principle.
DUAL ENERGY SOURCE
Dual Energy imaging means that the system uses two X-ray sources simultaneously at different energy levels.
This makes it possible to differentiatebetween fat, soft tissue, and bone, and also between calcifications and contrast
material (iodine) on the basis of their unique energy-dependent attenuation profiles.
WHAT IS DOSE IN CT SCAN..?
WHAT IS DOSE? Volume Computed Tomography Dose Index
(CTDIvol) is a standardized parameter to measure Scanner Radiation Output CTDIvol is NOT patient dose CTDIvol is reported in units of mGy for either a 16-cm
(for head exams) or 32-cm (for body exams) diameter acrylic phantom
For the same technique settings, the CTDIvol reported for the 16-cm phantom is about twice that of the 32-cm phantom
The reported CTDIvol is based on measurements made by the manufacturer in a factory setting
In these slides, the term "patient dose" is used to describe the absorbed dose to a patient, while the generic term "dose" refers to CTDIvol
HOW IS CTDIVOL RELATED TO PATIENT DOSE? CTDIvol is not patient dose The relationship between the two
depends on many factors, including patient size and composition
AAPM Report 204 introduces a parameter known as the Size Specific Dose Estimate (SSDE) to allow estimation of patient dose based on CTDIvol and patient size
For the same CTDIvol, a smaller patient will tend to have a higher patient dose than a larger patient
HOW IS CTDIVOL RELATED TO PATIENT DOSE?
Both patients scanned with the same CTDIvol Patient dose will be
higher for the smaller patient
CTDIvol = 20 mGy CTDIvol = 20 mGy
120 kVp at 200 mAs
120 kVp at 200 mAs
32 cm Phantom
32 cm Phanto
m
HOW IS CTDIVOL RELATED TO PATIENT DOSE?
Smaller patient scanned with a lower CTDIvol Patient doses will be
approximately equal
CTDIvol = 10 mGy CTDIvol = 20 mGy
120 kVp at 100 mAs
120 kVp at 200 mAs
32 cm Phantom
32 cm Phanto
m
WHY USE CTDIVOL? CTDIvol provides information about the amount
of radiation used to perform the study CTDIvol is a useful index to track across patients
and protocols for quality assurance purposes CTDIvol can be used as a metric to compare
protocols across different practices and scanners when related variables, such as resultant image quality, are also taken in account
The ACR Dose Index Registry (DIR) allows comparison across institutions of CTDIvol for similar exam types (e.g., routine head exam)
DOSE LENGTH PRODUCT The Dose Length Product (DLP) is also
calculated by the scanner DLP is the product of the length of the
irradiated scan volume and the average CTDIvol over that distance
DLP has units of mGy*cm
USEFUL CONCEPTS/TERMS The relationships between acquisition parameters and CTDIvol
described in the following slides assume all other parameters are held constant
The relationship between a parameter and CTDIvol is often described as proportional in some way The symbol µ is used to indicate “proportional to”
Directly proportional means that a change in the parameter results in the same change in CTDIvol Example: Doubling the rotation time from 0.5 to 1.0 seconds will
double the CTDIvol
Inversely proportional means that a change in a parameter has the opposite effect on CTDIvol
Example: Doubling the pitch from 1 to 2 will reduce the CTDIvol by half
ACQUISITION PARAMETER
ACQUISITION PARAMETER SETTINGS Acquisition Parameters define the
technique that will be used and how the scan will proceed
Acquisition Parameters are set in the user interface where scans are prescribed
Changing a single Acquisition Parameter while holding everything else constant will typically affect the CTDIvol for that scan
The following slides describe what that affect is for each parameter
SCAN MODE CT Scanners offer a variety of Scan Modes
which describe how the table moves during an exam
Scan Modes includeAxialHelical or Spiral Dynamic
The Acquisition Parameters that affect CTDIvol may change amongst
different Scan Modes
DYNAMIC SCAN MODE NOTES In the Dynamic Scan Mode multiple
acquisitions covering the same body region are acquired. Examples of these study types include:Perfusion StudiesBolus Tracking StudiesTest Bolus Studies
Dynamic Scans often have large CTDIvol values because the scanner reports the sum of the CTDIvol values from each rotation
The reported CTDIvol is NOT skin dose or organ dose
TABLE FEED/INCREMENT Is the movement of the table through the bore of
the scanner over a full 360 degree rotation Units: millimeters/rotation or millimeters/second The parameter is known both as Table Feed
(helical/spiral acquisition) & Table Increment (axial acquisition)
Table Feed affects CTDIvol through its inclusion in Pitch (discussed later)
DETECTOR CONFIGURATION Is the combination of the number of data channels and the
width of the detector associated with each data channel The Detector Configuration determines the Beam Width or
Beam Collimation (nT), which is the number of channels (n) times the detector width associated with each data channel (T)
For a selected detector width per data channel, a smaller total Beam Collimation usually has a higher CTDIvol than a larger Beam Collimation Example: On a 16 slice scanner with a detector width per channel
of 1.25 mm, a collimation of 4x1.25mm is generally less dose efficient than a collimation of 16x1.25mm
Users should monitor CTDIvol values when changing detector
configuration
Acquisition Parameter Settings
DETECTOR CONFIGURATION
PITCH Is the Table Feed per gantry rotation divided by
the beam width/collimation Pitch is the ratio of two distances and therefore
has no units Users should monitor other parameters when
changing Pitch. The scanner may or may not automatically compensate for changes in Pitch (for example, by changing the tube current) to maintain the planned CTDIvol.
CTDIvol µ 1/Pitch: Hitachi, Toshiba (no AEC)
CTDIvol independent of Pitch: GE, Siemens, Philips, Neusoft, Toshiba (AEC)
PITCH CTDIvol may not change in the expected
manner if the scanner automatically adjust other parameters when the pitch is changed
The relationships between CTDIvol and pitch for the different vendors are described below CTDIvol inversely proportional to change in pitch:
Hitachi, NeuroLogica CTDIvol constant when pitch is changed due to
changes to other parameters: GE, Neusoft, Philips and Siemens
The relationship between CTDIvol and pitch depends on scan mode or Software version: Toshiba
Pitch < 1Beam Width
has some overlap at each view angle from rotation to
rotation
Pitch = 1No overlap of
Beam Width at each view angle
and no view angles not covered at certain table
positions
Pitch > 1Some view angles are not covered
by the beam width at certain table positions
Acquisition Parameter Settings
PITCH
TUBE CURRENT Determines the number of electrons
accelerated across the x-ray tube per unit time
Units: milliAmperes (mA) CTDIvol is directly proportional to
Tube Current
CTDIvol µ Tube Current
TUBE POTENTIAL Is the electrical potential applied across
the x-ray tube to accelerate electrons toward the target material
Units: kiloVolts (kV or kVp) CTDIvol is approximately proportional to
the square of the percentage change in Tube Potential
n
old
new
kVkV
µvolCTDI
TUBE POTENTIAL (KVP) Research done by reducing kVp improved
enhancement of iodinated contrast media as a dye in the vascular and no significance increase in image noise.
Contrast to noise ratio(CNR) will increase at lower tube potentials, radiation dose may be reduced to achieve similar or improved iodine CNR compared to high kV. This will lead to reducing the dose to the patient.
With phantom 10cm the require dose at 80kV is only 35% of the amount required at 120kV and at 100kV the required dose is 62% of the amount required at 120kV.
With the 25cm phantom, the doses are 46% of the 120kV dose at 80kV and 63% of the 120-kV dose at 100 kV
Graph shows the relative radiation dose required at each tube potential to obtain the same iodine CNR For all three phantoms. For the 10-cm phantom, 35% of the 120-kV dose is required at 80 kV to achieve the same iodine CNR as at 120 kV, and 62% of the 120kV Dose is required At 100 kV. For the 25-cm Phantom, 46% of the 120kV dose Is requiredAt 80 kV, and 63% of the 120-kV dose is required at 100 kV
Axial contrast material–enhanced multidetector CT images obtained in a 58-year-old man during late hepatic arterial phase with (left) protocol A (140 kVp), (right) protocol B (80 kVp)((Marin et al., 2010)
TUBE POTENTIAL (KVP) However selecting the appropriate kVp
is not straightforward task Scanning speed, motion artifacts,
patient size and diagnostic task must be considered and carefully evaluated before the patients undergo CT scan examination
TUBE CURRENT TIME PRODUCT Is the product of Tube Current and
the Exposure Time per Rotation Units: milliAmpere-seconds (mAs) CTDIvol is directly proportional to
Tube Current Time Product
CTDIvol µ Tube Current Time Product
X-RAY TUBE CURRENT (MAS) dose ~ tube current(constant tube
potential, scanning time & slice thickness)
Weight or size based – 50% dose reduction
One research done reduction from 100mAs to 40mAs in sinus examination (findings – chronic and acute sinusitis)
Low dose (40mAs) can be clearly visualize; no significant difference dspite increase in noise (graininess)
Reduction should be made without degrade the image quality
Figure 1.Axial and Sagittal images of MDCT scans obtained at 100 mAs (A, B) and 40 mAs (C, D) at the level of the maxillary sinuses show complete opacification of the right maxillary sinus (star) and air-fluid levelin the left maxillary sinus (arrow). No significant difference in the diagnostic image quality of these tworadiological findings of these two scans.
Figure 2.Coronal reformatted images of MDCT scans obtained at 100 mAs (A) and 40 mAs (B) at the level of osteomeatal complex showing a normal right osteomeatal complex and a blocked left osteomeatal complex (arrow). This structure is clearly identified and correctly assessed in these two scans of this patient.
DOSE MODULATION AND REDUCTION Many CT scanners automatically adjust
the technique parameters (and as a result the CTDIvol) to achieve a desired level of image quality and/or to reduce dose
Dose Modulation and Reduction techniques vary by scanner manufacturer, model and software version
AUTOMATIC EXPOSURE CONTROL (AEC) Automatically adapts the Tube Current or Tube Potential
according to patient attenuation to achieve a specified image quality Automatic adjustment of Tube Current may not occur when Tube
Potential is changed Centering the patient in the gantry is VITAL for most AEC
systems AEC aims to deliver a specified image quality across a
range of patient sizes. It tends to increase CTDIvol for large patients and decrease it for small patients relative to a reference patient sizeThe use of Automatic Exposure Control
may decrease or increase CTDIvol depending on the patient size and body
area imaged and image quality requested
IMAGE QUALITY REFERENCE PARAMETER Is the AEC parameter that is set by
the user to define the desired level of image quality
Changing the Image Quality Reference Parameter will affect the CTDIvol
The effect on CTDIvol when changing the Image Quality Reference
Parameter is vendor dependentDose Modulation and
Reduction
IMAGE QUALITY REFERENCE PARAMETER A change in the Image Quality
Reference Parameter will affect the CTDIvol
Setting the parameter for “increased” image quality (e.g., lower noise) will result in more dose
Setting the parameter for “decreased” image quality (e.g., more noise) will result in less dose
Dose Modulation and Reduction
ANGULAR TUBE CURRENT MODULATION Is an AEC feature that adjusts the Tube
Current as the x-ray tube rotates around the patient to compensate for attenuation changes with view angle
Angular Tube Current Modulation is used to adjust the Tube Current to attempt to deliver similar dose to the detector at all view angles
The use of Angular Tube Current Modulation may decrease or increase CTDIvol depending on the patient size
and body area imaged and image quality requested
Dose Modulation and Reduction
ANGULAR TUBE CURRENT MODULATION Angular Tube Current Modulation uses
information from one or two view localizers
LONGITUDINAL TUBE CURRENT MODULATION Is an AEC feature that adjusts the
Tube Current as patient attenuation changes in the longitudinal direction
The CT Localizer Radiograph is used to estimate patient attenuation
The use of Longitudinal Tube Current Modulation may decrease or increase
CTDIvol depending on the patient size and body area imaged and image quality
requestedDose Modulation and
Reduction
LONGITUDINAL TUBE CURRENT MODULATION Longitudinal Tube Current Modulation
uses information from one or two view localizers
Dose Modulation and Reduction
ANGULAR AND LONGITUDINAL TUBE CURRENT MODULATION
Is an AEC feature that incorporates the properties of both Angular and Longitudinal Tube Current Modulation to Adjust the Tube Current based on the patient’s
overall attenuation Modulate the Tube Current in the angular (X-Y) and
longitudinal (Z) dimensions to adapt to the patient’s shape
The use of Angular and Longitudinal Tube Current Modulation may decrease or
increase CTDIvol depending on the patient size and body area imaged and image
quality requestedDose Modulation and
Reduction
ANGULAR AND LONGITUDINAL TUBE CURRENT MODULATION
Dose Modulation and Reduction
AUTOMATED TUBE MODULATION (ATM) - AEC AEC is available in modern CT scan
machine Reduction dose 40 – 50% dose The penetration adjust according to
patient specific attenuation an all three planes
Technique; depend on patient size, attenuation profile and scanning parameter
Figure. Graph of tube current (mA) superimposed on a CT projection radiograph shows the variation in tube current as a function of time (and, hence, table position along the z-axis) at spiral CT in a 6-year-old child. An adult scanning protocol and an AEC system.
ORGAN-BASED TUBE CURRENT MODULATION Is an AEC feature that allows for the tube
current to be decreased or turned off over radiosensitive organs on the patient periphery, such as the breasts or eye lenses
To maintain image quality, tube current may need to be increased at other view anglesThe use of Organ-Based Tube Current Modulation may reduce the absorbed dose to organs at the surface of the body but may increase the absorbed
dose to other organsDose Modulation and
Reduction
Gantry Gantry
Conventional Organ-Based Modulation
Dose Modulation and Reduction
ORGAN-BASED TUBE CURRENT MODULATION
AUTOMATIC TUBE POTENTIAL SELECTION Is an AEC feature that selects the
tube potential according to the diagnostic task and patient size in order to achieve the desired image quality at a lower CTDIvol
The use of Automatic Tube Potential Selection is intended to decrease CTDIvol while achieving the image
quality required for a specific diagnostic task and patient attenuation
Dose Modulation and Reduction
AUTOMATIC TUBE POTENTIAL SELECTION Tube Potential is not modulated in the
same fashion as Tube Current It does not change with different tube
positions (view angles) around the patient
The Tube Potential for a specific patient, anatomic region and diagnostic tasks is selected and held constant for that acquisition, though it may be changed to a different tube potential for a different diagnostic task
Dose Modulation and Reduction
62
INTRODUCTION TO FBPReconstruct an image from tomographic scan data
Schematic of Projection and Reconstruction
63
METHODOLOGY How are we going to solve this problem ? Filtered Back Projection
64
THEORY BEHIND BACK PROJECTION How does BP work ?
65
WHY DO WE APPLY A FILTER ?
Another diagram that illustrates this concept
66
WHY DO WE APPLY A FILTER ?
Without filter:• Point object becomes widely spread out
67
WHY DO WE APPLY A FILTER ?
With Filter: • Cancellations in the vicinity of the object
• The result is a sharper reconstructed image
68
WHY DO WE APPLY A FILTER ?
Process is linear - if it works for a point object (at any location),
then it will work for any image
ITERATIVE RECONSTRUCTION Is a feature that uses the information
acquired during the scan and repeated reconstruction steps to produce an image with less “noise” or better image quality (e.g., higher spatial resolution or decreased artifacts) than is achievable using standard reconstruction techniques
The use of Iterative Reconstruction by itself may not decrease CTDIvol; with use of Iterative
Reconstruction, image quality will change and this may allow a reduction in the CTDIvol by
adjusting the acquisition parameters used for the exam
Dose Modulation and Reduction
ITERATIVE RECONSTRUCTION Iterative Reconstruction may be completed using
data in Image Space, Sinogram Space or a Model Based Approach
Changing/Turning On the %/Level of the iterative reconstruction used may or may not affect the CTDIvol of the scan and will affect the image quality of the final set of images
In consultation, the Radiologists and Medical Physicists at an institution may adjust the acquisition parameters for studies reconstructed using iterative reconstruction based on the imaging task, the patient population, the desired image quality, dose concerns and the needs of the interpreting Radiologist
Dose Modulation and Reduction
ITERATIVE RECONSTRUCTION According to the Dictionary.com the iteration is define as
problem solving or computational method in which a the most close to the approximation
First an initial estimate of the x-ray photon from the tube is made. Second an estimate is made of the x-ray detector counts that would be acquired in each projection with the use of forward projection (Nelson et al., 2011). Then the estimated forward projection before will be compared with the actual measured projection acquired by the CT system detector array (Nelson et al., 2011). The comparison between the estimate and the actual measured will be use to update the original estimate (Nelson et al., 2011). Then the whole process will be continuously repeated that would result from the revised x-ray photon distribution. The process done will results in the estimation of x-ray photon closer to the actual photon distribution
ITERATIVE RECONSTRUCTION Siemens
IRIS: Iterative Reconstruction in Image SpaceSAFIRE: Sinogram-Affirmed Iterative
Reconstruction Thosiba
AIDR 3D: Adaptive Iterative Dose Reduction 3D
GEASIR: Adaptive Statistical Iterative
ReconstructionMBIR: Model-Based Iterative Reconstruction
Philips iDose4
(A) FBP image of the right shoulder in the coronal plane reconstructed with a soft tissue algorithm. Note the horizontal streak artifacts because of aliasing from low x-ray photon counts through the bones. (B) FBP image of the right shoulder in the coronal plane reconstructed with a bone algorithm. Note the horizontal streak artifacts because of aliasing are even more noticeable with this reconstruction. (C) MBIR image of the right shoulder in the coronal plane depicted with a soft tissue window and level. Note excellent depiction of the soft tissue about the shoulder and significantly fewer streak artifacts. (D) MBIR image of the right shoulder in the coronal plane depicted with a bone window and level. Note excellent depiction of the comminuted fracture of the coronoid process (arrow)
Filtered back projection (FBP) (a) and SAFIRE (b) reconstructions at the same level of the ascending aorta. Image noise expressed as the standard deviation of the attenuation (HU) in the region of interest was significantly lower in images reconstructed using iterative reconstruction (circle in B) compared with those reconstructed using FBP (circle in A).
DOSE DISPLAY Information about the CTDIvol planned for
each scan is typically displayed before the exam on the user console
Information about the CTDIvol delivered by each scan is typically reported in a data page or DICOM structured dose report
Dose information provided after the exam typically also includes the DLP and the CTDI phantom size. These may also be included in information displayed before the scan.
DISPLAY OF PLANNED CTDIVOL CTDIvol is displayed before a study is
performed based on the selected technique parameters
It is important to check CTDIvol before a study is performed to ensure that the output of the scanner is appropriate for the specific patient and diagnostic task
CTDIvol is displayed for each planned acquisition
Dose Display
RECOMMENDATION Development of technology in CT scan
make essential for radiographer cope with new skills and techniques.
Training should be provided Increase knowledge Know your THINGS. Avoid becoming
PUSH BUTTON RADIOGRAPHER..!!
CONCLUSION Radiographers are integral part of a
multidisciplinary team whose responsibility is to treat and manage patient came to the imaging department.
The role of the radiographer is to take care of the patients’ imaging before proceed to other treatment or investigation.
Current advances in technology have made it essential for nowadays radiographers to constantly learn and cope with new skills.
The increasing usage of CT scan facilities increases radiation doses to the staff and the population and this create a full awareness for continuous efforts in reducing the dose level.
REFERENCES FLEISCHMANN, D. & BOAS, F. E. 2011. Computed tomography—old ideas
and new technology. Eur Radiol, 21, 510-517. GERVAISE, A., OSEMONT, B., LECOCQ, S., NOEL, A., MICARD, E.,
FELBLINGER, J. & BLUM, A. 2012. CT image quality improvement using adaptive iterative dose reduction with wide-volume acquisition on 320-detector CT. Eur Radiol, 22, 295-301.
GRANT, K. & RAUPACH, R. 2012. SAFIRE: Sinogram Affirmed Iterative Reconstruction. Siemens Medical Solutions Whitepaper. Available via http://www. medical. siemens. com/siemens/en_US/gg_ct_FBAs/files/Definition_AS/Safire. pdf. Accessed Nov.
IRWAN, B., NAKANISHI, S. & BLUM, A. 2012. AIDR 3D—Reduces Dose and Simultaneously Improves Image Quality. Toshiba Medical Systems Whitepaper. Available via http://www. toshiba-medical. eu/upload/TMSE_CT/White% 20Papers/White% 20Papers/Toshiba_White% 20paper% 20CT_nov11. pdf.
KATSURA, M., MATSUDA, I., AKAHANE, M., SATO, J., AKAI, H., YASAKA, K., KUNIMATSU, A. & OHTOMO, K. 2012. Model-based iterative reconstruction technique for radiation dose reduction in chest CT: comparison with the adaptive statistical iterative reconstruction technique. Eur Radiol, 22, 1613-1623.
MARIN, D., NELSON, R. C., SCHINDERA, S. T., RICHARD, S., YOUNGBLOOD, R. S., YOSHIZUMI, T. T. & SAMEI, E. 2010. Low-tube-voltage, high-tube-current multidetector abdominal CT: improved image quality and decreased radiation dose with adaptive statistical iterative reconstruction algorithm--initial clinical experience. Radiology, 254, 145-53.
MOSCARIELLO, A., TAKX, R. A., SCHOEPF, U. J., RENKER, M., ZWERNER, P. L., O’BRIEN, T. X., ALLMENDINGER, T., VOGT, S., SCHMIDT, B. & SAVINO, G. 2011. Coronary CT angiography: image quality, diagnostic accuracy, and potential for radiation dose reduction using a novel iterative image reconstruction technique—comparison with traditional filtered back projection. Eur Radiol, 21, 2130-2138.
NELSON, R. C., FEUERLEIN, S. & BOLL, D. T. 2011. New iterative reconstruction techniques for cardiovascular computed tomography: how do they work, and what are the advantages and disadvantages? Journal of Cardiovascular Computed Tomography, 5, 286-292.
NOËL, P. B., FINGERLE, A. A., RENGER, B., MÜNZEL, D., RUMMENY, E. J. & DOBRITZ, M. 2011. Initial Performance Characterization of a Clinical Noise–Suppressing Reconstruction Algorithm for MDCT. American Journal of Roentgenology, 197, 1404-1409.
PARK, E.-A., LEE, W., KIM, K. W., KIM, K. G., THOMAS, A., CHUNG, J. W. & PARK, J. H. 2012. Iterative reconstruction of dual-source coronary CT angiography: assessment of image quality and radiation dose. The International Journal of Cardiovascular Imaging, 28, 1775-1786.
SAGARA, Y., HARA, A. K., PAVLICEK, W., SILVA, A. C., PADEN, R. G. & WU, Q. 2010. Abdominal CT: comparison of low-dose CT with adaptive statistical iterative reconstruction and routine-dose CT with filtered back projection in 53 patients. AJR Am J Roentgenol, 195, 713-9.
SCHEFFEL, H., STOLZMANN, P., SCHLETT, C. L., ENGEL, L.-C., MAJOR, G. P., KÁROLYI, M., DO, S., MAUROVICH-HORVAT, P. & HOFFMANN, U. 2012. Coronary artery plaques: cardiac CT with model-based and adaptive-statistical iterative reconstruction technique. European Journal of Radiology, 81, e363-e369.
SILVA, A. C., LAWDER, H. J., HARA, A., KUJAK, J. & PAVLICEK, W. 2010. Innovations in CT dose reduction strategy: application of the adaptive statistical iterative reconstruction algorithm. American Journal of Roentgenology, 194, 191-199.
THIBAULT, J.-B., SAUER, K. D., BOUMAN, C. A. & HSIEH, J. 2007. A three-dimensional statistical approach to improved image quality for multislice helical CT. Med Phys, 34, 4526.
URIOL, M. V. Reconstruction from Projections. WILLEMINK, M. J., DE JONG, P. A., LEINER, T., DE HEER, L. M., NIEVELSTEIN, R.
A., BUDDE, R. P. & SCHILHAM, A. M. 2013. Iterative reconstruction techniques for computed tomography part 1: technical principles. Eur Radiol, 1-9.
WINKLEHNER, A., KARLO, C., PUIPPE, G., SCHMIDT, B., FLOHR, T., GOETTI, R., PFAMMATTER, T., FRAUENFELDER, T. & ALKADHI, H. 2011. Raw data-based iterative reconstruction in body CTA: evaluation of radiation dose saving potential. Eur Radiol, 21, 2521-2526.
XU, J., MAHESH, M. & TSUI, B. M. 2009. Is iterative reconstruction ready for MDCT? Journal of the American College of Radiology: JACR, 6, 274.
THANK YOU
Mohd Aiman bin AzmardiPeg. Pengimejan
Hospital Kulim