43.35” (LENGTH OF INNER LIQUID HE RESERVIOR)
CONNECTION CHAMBER: FF 4.75” FF: FLANGE-TO-FLANGE DISTANCE
10.45” TO 12.45”
6.35” TO 8.35” (JACK -2” TO 0) (BOTTOM OF HE COLD FINGER TO A-A FOCAL POINT)
COARSE-Z JACK + RNN FF 5.7” TO 7.7”
NIPPLE 3.56”
2.47”
50.4” (BETWEEN BOTTOM OF CAGE AND MIDDLE OF CONNECTION CHAMBER) = 42.55” + (7.85” BASED ON SCANNING POSITION AT JACK @ -0.5”)
PROBE/SCANNER HEIGHT 1.42”
(36MM SPECIFIED BY
INTEMATIX) BOTTOM PLATE THICKNESS
0.25”
~ 3.33” TO 5.33”
SCANNING POSITION: Z-STAGE (ATTOCUBE – FULL RANGE
15MM) EXTENSION ~ 0.43” (11MM) JACK AT -0.5” (THIS DIMENSION
DETERMINES HEIGHT OF SPM CAGE – POSSIBLE RANGE -0.2” TO -0.8”)
TRANSFER POSITION: Z-STAGE ~ 0 => JACK NEEDS TO GO DOWN ~1” (@ -1.5”)
10-PIN FEEDTHROUGH FOR SIGNALS (2 FOR TUNNELING CURRENT)HV CABLES (PZT) GO UP THROUGH THE CONNECTION CHAMBER
3.86”
LIQUID N2 CRYO-SHIELD/ANCHOR
COARSE-Z JACK: FF 4.27” TO 6.27” (-2” TO 0)
RNN: DIFFERENTIALLY PUMPED ROTARY PLATFORM – POLAR ROTATION
~ 7.33 TO 8.19”
0.84”
FEMALE SMA CONNECTOR
LENGTH OF RF CABLE:53.9” (FROM BOTTOM OF PROBE CAGE TO FACE OF MALE SMA CONNECTOR) 50.4” + ~0.5” (BENT SECTION IN MIDDLE OF CONNECTION CHAMBER) + 3” (STRAIGHT SECTION BEYOND BENT TO FACE OF MALE CONNECTOR)
A U-SHAPED SECTION NEEDS TO BE ADDED TO EACH OF THE RF CABLES BELOW THE CAGE TO CONNECT TO THE PROBETHE “U’ SECTIONS NEED TO FIT WITHIN A CYLINDER OF HEIGHT 2” BY ID 2.37”
LENGTH OF RF CABLE:53.9” (FROM BOTTOM OF PROBE CAGE TO FACE OF MALE SMA CONNECTOR) 50.4” + ~0.5” (BENT SECTION IN MIDDLE OF CONNECTION CHAMBER) + 3” (STRAIGHT SECTION BEYOND BENT TO FACE OF MALE CONNECTOR)
A U-SHAPED SECTION NEEDS TO BE ADDED TO EACH OF THE RF CABLES BELOW THE CAGE TO CONNECT TO THE PROBETHE “U’ SECTIONS NEED TO FIT WITHIN A CYLINDER OF HEIGHT 2” BY ID 2.37”
FEMALE SMA CONNECTOR
RF CABLE
2.47”
43.35” (LENGTH OF INNER LIQUID HE RESERVIOR)
CONNECTION CHAMBER: FF 4.75” FF: FLANGE-TO-FLANGE DISTANCE
10.45” TO 12.45”
6.35” TO 8.35” (JACK -2” TO 0) (BOTTOM OF HE COLD FINGER TO A-A FOCAL POINT)
COARSE-Z JACK + RNN FF 5.7” TO 7.7”
NIPPLE 3.56”
2.47”
PROBE/SCANNER HEIGHT 1.42”
(36MM SPECIFIED BY
INTEMATIX) BOTTOM PLATE THICKNESS
0.4”
~ 3.33” TO 5.33”
SCANNING POSITION: Z-STAGE (ATTOCUBE – FULL RANGE
15MM) EXTENSION ~ 0.39” (10MM) JACK AT -0.5” (THIS DIMENSION
DETERMINES HEIGHT OF SPM CAGE – POSSIBLE RANGE -0.2” TO -0.8”)
TRANSFER POSITION: Z-STAGE ~ 0 => JACK NEEDS TO GO DOWN ~1” (@ -1.5”)
10-PIN FEEDTHROUGH FOR SIGNALS (2 FOR TUNNELING CURRENT)HV CABLES (PZT) GO UP THROUGH THE CONNECTION CHAMBER
3.86”
LIQUID N2 CRYO-SHIELD/ANCHOR
COARSE-Z JACK: FF 4.27” TO 6.27” (-2” TO 0)
RNN: DIFFERENTIALLY PUMPED ROTARY PLATFORM – POLAR ROTATION
~ 7.33 TO 8.19”
0.84”
FEMALE SMA CONNECTOR
LENGTH OF RF CABLE:53.9” (FROM BOTTOM OF PROBE CAGE TO FACE OF MALE SMA CONNECTOR) 50.4” + ~0.5” (BENT SECTION IN MIDDLE OF CONNECTION CHAMBER) + 3” (STRAIGHT SECTION BEYOND BENT TO FACE OF MALE CONNECTOR)
A U-SHAPED SECTION NEEDS TO BE ADDED TO EACH OF THE RF CABLES BELOW THE CAGE TO CONNECT TO THE PROBETHE “U’ SECTIONS NEED TO FIT WITHIN A CYLINDER OF HEIGHT 2” BY ID 2.37”
LENGTH OF RF CABLE:53.9” (FROM BOTTOM OF PROBE CAGE TO FACE OF MALE SMA CONNECTOR) 50.4” + ~0.5” (BENT SECTION IN MIDDLE OF CONNECTION CHAMBER) + 3” (STRAIGHT SECTION BEYOND BENT TO FACE OF MALE CONNECTOR)
A U-SHAPED SECTION NEEDS TO BE ADDED TO EACH OF THE RF CABLES BELOW THE CAGE TO CONNECT TO THE PROBETHE “U’ SECTIONS NEED TO FIT WITHIN A CYLINDER OF HEIGHT 2” BY ID 2.37”
FEMALE SMA CONNECTOR
RF CABLE
Updated version 8/23/07
9.67” =(50.55”+2.47”-43.35”)
50.55” (BASED ON SCANNING POSITION AT JACK @ -0.5”)
5.6
5”
39.25”
~ 17” TO THE TABLE