Course Objectives Understanding the importance of your role in the care
and handling of microsurgical instruments
Understand the importance of the CSPD in patient
safety, surgeon satisfaction and cost controls
Review of instrument categories
Discuss the what and why of the inspection process
Identifying instrument issues…
![Page 3: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/3.jpg)
History
• Ophthalmology was first mentioned in 2250 B.C.
• Hippocrates was the father of Ophthalmology
![Page 4: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/4.jpg)
History Cataract set circa Early 18th Century
2. Exhibition ophthalmologic surgical set by Aubry, (ca. 1880)
Images displayed courtesy of www.phisick.com
![Page 5: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/5.jpg)
Today
•All Microsurgical Instruments to greater or
lesser extent are Handmade.
•Forgings
•Stampings
•Machining
![Page 6: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/6.jpg)
Why is this important? • PATIENT CARE
• Cost of general/specialty instruments – General surgery set
• $4,000 average
– Handheld Ortho, CV, Neuro instruments • $300 average each – including rongeurs and bone biters
• Cost of microsurgical instruments – Middle ear set
• $6,000 average
– Cataract set • $6,000 average
– Single microsurgical instrument • $500 average – delicate needle holders, forceps
![Page 7: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/7.jpg)
Inspection of Micro Instruments
Inspection Basics
High quality lighted magnifying lamp
Lighted magnifying loupes
Hand held lighted magnifying glass
Testing media for scissors
![Page 8: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/8.jpg)
Micro Instruments
• Must use appropriate
magnification (after all, this is
how the surgeon sees them!)
• Two types of grip = flat handle
and knurled round
handle…why?
![Page 9: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/9.jpg)
Inspection of Ophthalmic Instruments
Inspection Basics
High quality lighted magnifying lamp
Lighted magnifying loupes
Hand held lighted magnifying glass
Testing media for scissors It is CRITICAL that the CSPD professionals
inspecting and assembling micro instrument sets are
thoroughly trained and regularly assessed for competency
Ongoing mandatory training should be scheduled to address
new technology and current events in the industry
Communications between the OR and SPD should be
established sharing information…
We will review inspection criteria on the following slides.
![Page 10: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/10.jpg)
Inspection
Always use Magnifier
– Burrs on tips?
– Nicks on cutting edges?
– Improper alignment of jaws, shafts, blades?
– Condition of tungsten carbide inserts?
Remember, what you are not seeing with your naked
eyes, the surgeon is seeing under a microscope or a
magnifier
![Page 11: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/11.jpg)
Inspection Guidelines
New instruments should always be checked before use • Damage in shipment
• Damage in packaging
• MUST BE PROCESSED AS ‘DIRTY’ BEFORE USE
• Ring-handled forceps • Minute cracks in the box locks
• Misaligned jaws
• If serrated, assure proper mesh on closure
• If toothed, ensure teeth are not broken, mesh properly
• Assure ratchets/shanks are set properly. • Tips should meet on first ratchet with a definite click
![Page 12: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/12.jpg)
Guidelines continued
• Scissors
– Roughness or burs at the scissor tips
– Broken or bent tips
– Nicks in the cutting edges
– Worn or loose fulcrum screws
• Can prevent smooth cutting action
![Page 13: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/13.jpg)
Inspection
Micro Scissors:
• Theraband (yellow) to test scissors
• Use a cotton ball to test for burrs in
cutting edge
NOTE: If using a cotton ball to test for
burrs, ensure you remove any cotton
fibers left on instrument from the
inspection!
![Page 14: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/14.jpg)
Guidelines continued
• Needle Holders
– Signs of wear on the serrations • Clamp a needle in the jaws by locking
the second ratchet
• If needle can be turned, then repair of holder needed
– Other inspections same as for ring handled instruments
NOTE: Check to ensure locking
mechanism in handle engages
properly
![Page 15: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/15.jpg)
Guidelines continued
• Picks, knives
– Inspect for burrs and nicks
– Check for bends or cracks
• Hooks
– Misalignment
– Burrs or breaks
![Page 16: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/16.jpg)
Guidelines continued
• Rongeurs
– Proper alignment
– Fulcrum screws, as applicable
– Smooth cutting action
– Sharpness of blades, cups
![Page 17: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/17.jpg)
Guidelines continued
• Suctions/Irrigators – Smooth tips
• Even microscopic sharp/nicked tips can damage delicate patient tissue (check with cotton ball)
– Stylet available
• Retractors – Wire uncompromised
![Page 18: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/18.jpg)
Inspection of Micro Instruments Due to the size of microsurgical
instruments, special equipment
must be used by repair specialists to
examine and determine necessary
repairs.
The following slides will help us better
understand the need for close
inspection of all instruments and
periodic tray maintenance by a qualified
repair specialist
![Page 19: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/19.jpg)
Inspection of Micro Instruments
![Page 20: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/20.jpg)
Inspection of Micro Instruments
![Page 21: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/21.jpg)
Inspection of Micro Instruments
![Page 22: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/22.jpg)
Inspection of Micro Instruments
![Page 23: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/23.jpg)
Inspection of Micro Instruments
![Page 24: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/24.jpg)
Inspection of Micro Instruments
![Page 25: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/25.jpg)
Inspection of Micro Instruments
![Page 26: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/26.jpg)
Inspection of Ophthalmic Instruments
Scissor points are extremely delicate; the tips
should not be touched.
![Page 27: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/27.jpg)
Inspection of Micro Instruments
Note the difference in size of ophthalmic scissors.
![Page 28: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/28.jpg)
Inspection of Ophthalmic Instruments
Note the bi-polar cautery forceps.
![Page 29: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/29.jpg)
Inspection of Ophthalmic Instruments
Ophthalmic retractors and fixation instruments
![Page 30: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/30.jpg)
Inspection of Ophthalmic Instruments
Retractors, knife handles and various nerve hooks
must be inspected for wear and damage that may cause
unintended trauma during use.
![Page 31: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/31.jpg)
Golden Rule!!!
Always follow the device manufacturers written
instructions!
![Page 32: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/32.jpg)
Cleaning
• Must begin in the O.R. – Wipe debris as used
– Immediately after surgery • Keep moist
• Utilize enzymatic or pre-cleaning solution
– Assure proper transportation • Lighter instruments on top
• Segregate sharps
![Page 33: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/33.jpg)
Cleaning
• Decontamination – Enzymatic
– Sonic: dB level is different than with general instruments
– Washer Decontaminator • Always follow manufacturers instructions specific
to micro-instruments
– Standard instrument practices • Take apart whenever possible
– Special care for knurled of milled grooved handles
• Lubrication is critical!
![Page 34: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/34.jpg)
The ASORN Cleaning Process: An Overview1
30,31
22,23
30,31
22,23
The instrument cleaning process needs to begin in
the operating room
Instruments must be wiped off after use to keep
ophthalmic viscosurgical solutions from drying
Hand pieces should be flushed with, then soaked
in, sterile water off the sterile field in the OR until
they can be properly cleaned and decontaminated
3,4. Schuyer, D. The risk of TASS makes proper instrument handling procedures more important than
ever. ESCRS EuroTimes, July-September 2007.
![Page 35: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/35.jpg)
The ASORN Cleaning Process: An Overview1
30,31
22,23
30,31
22,23
The instrument cleaning process needs to begin in the operating
room
Instruments must be wiped off after use to keep ophthalmic
viscosurgical solutions from drying
Hand pieces should be flushed with, then soaked in, sterile water off
the sterile field in the OR until they can be properly cleaned and
decontaminated
5. Schuyer, D. The risk of TASS makes proper instrument handling procedures more important than ever. ESCRS
EuroTimes, July-September 2007.
Ophthalmic viscosurgical solution, which can
dry and harden within minutes, should not be
allowed to dry on the instruments
Special
Consideration!
![Page 36: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/36.jpg)
The ASORN Cleaning Process: An Overview1
30,31
22,23
30,31
22,23
• The next step is decontamination, which means
bringing the instruments to a soiled utility room for
mechanical cleaning with a soft bristled brush.
• Hand pieces should be flushed with 120 to150cc of
water according to the manufacturer's instructions
and dried with forced air or medical-grade nitrogen.
• There should also be a visual inspection of each
instrument.
7 Schuyer, D. The risk of TASS makes proper instrument handling procedures more important than ever.
ESCRS EuroTimes, July-September 2007.
![Page 37: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/37.jpg)
The ASORN Cleaning Process: An Overview1
30,31
22,23
30,31
22,23
• After decontamination the instruments should be
cleaned according to the manufacturer's
instructions.
• For most phacoemulsification and I/A hand pieces,
this means avoiding the use of ultrasound machines,
enzymatic cleaners and detergents and simply flushing
with sterile, distilled or deionized water.
8. Schuyer, D. The risk of TASS makes proper instrument handling procedures more important than ever.
ESCRS EuroTimes, July-September 2007.
![Page 38: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/38.jpg)
The ASORN Cleaning Process: An Overview
30,31
22,23
30,31
22,23
1. Adequate time for thorough cleaning and sterilization
of instrumentation should be established.
2. For each piece of equipment, the manufacturer’s
directions for use pertaining to cleaning and
sterilization should be followed.
3. Whether they are used or not, instruments opened for
a procedure should be transported from the OR in a
closed container to the decontamination area and
cleaned immediately
![Page 39: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/39.jpg)
The ASORN Cleaning Process: An Overview
30,31
22,23
30,31
22,23
5. Disposable cannula and tubing should be used
whenever possible, and they should be discarded
after each use.
6. Devices labeled for “single use only” should not be
reused. (Single-use devices do not include
instructions for reuse or reprocessing. Third-party
and hospital re-processors of single-use devices are
subject to additional regulatory requirements as
specified by the FDA.)
![Page 40: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/40.jpg)
The ASORN Cleaning Process: An Overview
30,31
22,23
30,31
22,23
7. To avoid contamination with bio-burden and cleaning
chemicals, intraocular instruments should be cleaned
separately from non-ophthalmologic surgical
instruments.
8. The importance of enzymatic detergents for the
cleaning of soiled intraocular instruments has not been
established. Inappropriate use and incomplete rinsing
of enzymatic detergents have been associated with
outbreaks of TASS (Toxic Anterior Segment Syndrome).
![Page 41: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/41.jpg)
The ASORN Cleaning Process: An Overview
30,31
22,23
30,31
22,23
7. To avoid contamination with bio-burden and cleaning
chemicals, intraocular instruments should be cleaned
separately from non-ophthalmologic surgical
instruments.
8. The importance of enzymatic detergents for the
cleaning of soiled intraocular instruments has not been
established. Inappropriate use and incomplete rinsing
of enzymatic detergents have been associated with
outbreaks of TASS.
When flushing is used as part of a cleaning
technique,
the effluent should be discharged into
a sink or separate basin so the fluid is not reused
Don’t Double-Dip!!!
Special
Consideration!
![Page 42: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/42.jpg)
Packaging
• Assure instruments are dry
• Utilize tip guards/protective pouches wherever
possible
• Utilize silicone mats in bottom of trays
• Racks specific for holding microsurgical
instruments
• Trays designed for holding microsurgical
instruments
![Page 43: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/43.jpg)
The ASORN Cleaning Process: An Overview1
30,31
22,23
30,31
22,23
The final step is sterilization •Sterilization of intraocular instruments
must be done according to manufacturers
DFU’s
•Verification of sterilizer function should be
completed at least weekly, preferably daily
•Preventive maintenance, cleaning, and
inspection performed on a scheduled basis,
according to the manufacturer’s written
instructions
6 Schuyer, D. The risk of TASS makes proper instrument handling procedures more important than ever.
ESCRS EuroTimes, July-September 2007.
![Page 44: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/44.jpg)
The ASORN Cleaning Process: An Overview1
30,31
22,23
30,31
22,23
The final step is sterilization
•Sterilization of intraocular instruments
should be done according to manufacturers
DFU’s
•Verification of sterilizer function should be
completed at least weekly, preferably daily
•Preventive maintenance, cleaning, and
inspection performed on a scheduled basis,
according to the manufacturer’s written
instructions
6 Schuyer, D. The risk of TASS makes proper instrument handling procedures more important than ever.
ESCRS EuroTimes, July-September 2007.
Due to the delicate nature of these
instruments, it is extremely important to have
them maintained by a qualified and highly-
experienced repair specialist with the
capability to bring the instruments
back to like-new condition quickly
and cost effectively.
Special
Consideration!
![Page 45: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/45.jpg)
47
Sterilization; The Spaulding System
Adapted from Recommended Practices for Selection and Use of Packaging Systems, Association of periOperative
Registered Nurses, 2007 Standards, Recommended Practices, and Guidelines.
Classification Definition Processing Type
Critical
* Surgical instruments
* Implants
* Needles
* Cardiac and urinary catheters
Contacts sterile tissue or is
introduced into the bloodstream
Sterilization
Semi critical
* Bronchoscopes
* Gastrointestinal endoscopes
* Anesthesia equipment
* Cystoscopes
Contacts
intact mucous
membranes
High-level disinfection
Non critical
* Blood pressure cuffs
* Tourniquets
* Utensils such as basins
* Equipment and furnishings
Contacts intact
skin
Low-level disinfection
![Page 46: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/46.jpg)
Sterilization
• Low heat is best
– High temps shorten the life of delicate instruments
• IUSS very harmful
– Dulls sharp edges
– Causes rapid degeneration of micro-parts
• Mixing metals
– Microsurgical instruments are more susceptible to
electrolytic chemical reactions
– Instruments will appear to have rusting
– Segment any chrome plated or titanium
instruments from stainless steel
![Page 47: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/47.jpg)
![Page 48: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/48.jpg)
Review (Class Participation)
![Page 49: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/49.jpg)
CSPD professionals inspecting and assembling
microsurgical instrument sets must be:
Thoroughly trained at hire
Have their training, initial
and ongoing, documented
Regularly assessed for
competency
All the above
![Page 50: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/50.jpg)
CSPD professionals inspecting and assembling
microsurgical instrument sets must be:
Thoroughly trained at hire
Have their training, initial
and ongoing, documented
Regularly assessed for
competency
All the above
![Page 51: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/51.jpg)
Due to technology, virtually all surgical instruments today
are:
Stain Free
Replaced cheaper than
repair
Easily repaired by SPD
None of the above
![Page 52: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/52.jpg)
Due to technology, virtually all surgical instruments today
are:
Stain Free
Replaced cheaper than
repair
Easily repaired by SPD
None of the above
![Page 53: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/53.jpg)
Why should ophthalmic scissors not be touched with bare
hands during the inspection process?
Cut your self
Leave behind microscopic
biological debris
Bend the tips
You can touch the tips
![Page 54: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/54.jpg)
Why should ophthalmic scissors not be touched with bare
hands during the inspection process?
Cut your self
Leave behind microscopic
biological debris
Bend the tips
You can touch the tips
![Page 55: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/55.jpg)
When cleaning microsurgical instruments, what
documentation processes should be followed?
Instructions from detergent
Established facility training
program
Instrument manufacturer
DFU’s
SDS and fact sheet
![Page 56: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/56.jpg)
When cleaning microsurgical instruments, what
documentation processes should be followed?
Instructions from detergent
Established facility training
program
Instrument manufacturer
DFU’s
SDS and fact sheet
![Page 57: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/57.jpg)
When flushing is used as part of a cleaning technique the
effluent…
Should be re-used until
visibly dirty
Can be recycled
Should be discharged into
a sink or separate basin
None of the above
![Page 58: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/58.jpg)
When flushing is used as part of a cleaning technique the
effluent…
Should be re-used until
visibly dirty
Can be recycled
Should be discharged into
a sink or separate basin
None of the above
![Page 59: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/59.jpg)
Decontamination of micro instruments begins…
In the decontamination
area
Only by CSPD staff
when the instruments can
be mechanically cleaned
At the point of use
![Page 60: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/60.jpg)
Decontamination of micro instruments begins…
In the decontamination
area
Only by CSPD staff
when the instruments can
be mechanically cleaned
At the point of use
![Page 61: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/61.jpg)
What media would be best suited to test for ‘burrs’ in the
cutting edges of micro-instruments?
Cellophane
Yellow Theraband
Cotton ball
Tissue paper
![Page 62: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/62.jpg)
What media would be best suited to test for ‘burrs’ in the
cutting edges of micro-instruments?
Cellophane
Yellow Theraband
Cotton ball
Tissue paper
![Page 63: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/63.jpg)
Ophthalmic visco-surgical solution is harmless to the
instrument and easily removed during the cleaning
process…
True False
![Page 64: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/64.jpg)
Ophthalmic visco-surgical solution is harmless to the
instrument and easily removed during the cleaning
process…
True False
![Page 65: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/65.jpg)
To best serve our patients, it is strongly suggested that a
coordinated training effort be established between the OR
and CSPD professional staff.
True False
![Page 66: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/66.jpg)
To best serve our patients, it is strongly suggested that a
coordinated training effort be established between the OR
and CSPD professional staff.
True False
![Page 67: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/67.jpg)
Surgical Instrument, Processing and Inspection
Summary
Microsurgical instruments are a large investment and with
proper care and handling, we can support better patient
procedure outcomes, extend the life of the instruments and
support the cost concerns of our facility
SPD professionals with a keen understanding and continued
education can have a major impact on microsurgical
instrument processing and the safety of the patient.
![Page 68: Are Your Microsurgical Instruments Patient Ready?...after each use. 6. Devices labeled for “single use only” should not be reused. (Single-use devices do not include instructions](https://reader035.vdocument.in/reader035/viewer/2022072902/5f2ec52d4996ec5fb33863fc/html5/thumbnails/68.jpg)
Safety in CSPD
References 1. Aubry(ca. 1880)Exhibition ophthalmologic surgical set, Images displayed
courtesy of www.phisick.com
3-9. Schuyer(D. ca. July-September 2007, ESCRS EuroTimes)The risk of TASS
makes proper instrument handling procedures more important than ever.