BAS100248SQ.05 09/13
Beckman Coulter System Reagent, AU400/ AU400e
DRI® Oxycodone Assay, Semi-Quantitative
Cat. No.: Microgenics 100248,100249 Reagent ID:
Specific Test Parameters
General LIH ISE Range
Test Name: OXY Type: URINE Operation: Yes
Sample: Volume 10.5 L Dilution 0 L Pre-Dilution Rate: 1
Reagents: R1 Volume 87 L Dilution 0 L Min OD Max OD
R2 Volume 87 L Dilution 0 L L 0.00 H 2.50
Reagent OD limit:
Wavelength: Pri. 340 Sec. 410 First L 0.00 First H 2.50
Method: RATE1* Last L 0.00 Last H 2.50
Reaction slope: + Dynamic Range:
Measuring Point 1: First 17 Last 21 L -99999 H 1000
Measuring Point 2: First Last Correlation Factor:
Linearity: 100 % A 1 B 0
No Lag Time: NO On-board stability period: 30
Specific Test Parameters
General LIH ISE Range
Test Name: OXY Type: URINE
Value/Flag: VALUE Level L: -99999 Level H: #
Normal Ranges: Age L Age H Sex Year Month Year Month L H
1. # # # # # # #
2. # # # # # # #
3. # # # # # # #
4. # # # # # # #
5. # # # # # # #
6. # # # # # # #
7. None Selected # #
8. Out of Range L H # #
Panic Value: # # Unit: ng/mL Decimal places: #
Calibration Specific
General ISE
Test Name: OXY Type URINE
Calibration Type: 5AB Formula: POLYGONAL Counts: 2 Process: CONC
Cal. No. OD CONC† Factor/OD-L Factor/OD-H
Point 1: # 0 -2.00 2.500
Point 2: # 100 -2.00 2.500
Point 3: # 300 -2.00 2.500
Point 4: # 500 -2.00 2.500
Point 5: # 1000 -2.00 2.500
Point 6:
Point 7:
1-Point Cal. Point: With CONC-0 Slope Check None Advanced Calibration: ##
MB Type Factor: Calibration Stability Period: 14
# User Defined ## Lot or Lot + Bottle † DRI
® Calibrators (Cat. No.: 1644 – NegCal; Cat. No.: 100250-100Cal; Cat. No.: 100251-300Cal;
Cat. No.: 100252-500Cal; and Cat. No.: 100253-1000Cal.) * Can also be run as FIXED1
BAS100248SQ.05 09/13
Beckman Coulter System Reagent, AU640/AU640e
DRI® Oxycodone Assay, Semi-Quantitative
Cat. No.: Microgenics 100248,100249
Reagent ID:
Specific Test Parameters
General LIH ISE Range
Test Name: OXY Type: URINE Operation: Yes
Sample: Volume 10.5 L Dilution 0 L Pre-Dilution Rate: 1
Reagents: R1 Volume 87 L Dilution 0 L Min OD Max OD
R2 Volume 87 L Dilution 0 L L 0.00 H 2.50
Reagent OD limit:
Wavelength: Pri. 340 Sec. 410 First L 0.00 First H 2.50
Method: RATE1* Last L 0.00 Last H 2.50
Reaction slope: + Dynamic Range:
Measuring Point 1: First 17 Last 21 L -99999 H 1000
Measuring Point 2: First Last Correlation Factor:
Linearity: 100 % A 1 B 0
No Lag Time: NO On-board stability period: 30
Specific Test Parameters
General LIH ISE Range
Test Name: OXY Type: URINE
Value/Flag: VALUE Level L: -99999 Level H: #
Normal Ranges: Age L Age H Sex Year Month Year Month L H
1. # # # # # # #
2. # # # # # # #
3. # # # # # # #
4. # # # # # # #
5. # # # # # # #
6. # # # # # # #
7. None Selected # #
8. Out of Range L H # #
Panic Value: # # Unit: ng/mL Decimal places: #
Calibration Specific
General ISE
Test Name: OXY Type URINE
Calibration Type: 5AB Formula: POLYGONAL Counts: 2 Process: CONC
Cal. No. OD CONC† Factor/OD-L Factor/OD-H
Point 1: # 0 -2.00 2.500
Point 2: # 100 -2.00 2.500
Point 3: # 300 -2.00 2.500
Point 4: # 500 -2.00 2.500
Point 5: # 1000 -2.00 2.500
Point 6:
Point 7:
1-Point Cal. Point: With CONC-0 Slope Check None Advanced Calibration: ##
MB Type Factor: Calibration Stability Period: 14
# User Defined ## Lot or Lot + Bottle † DRI
® Calibrators (Cat. No.: 1644 – NegCal; Cat. No.: 100250-100Cal; Cat. No.: 100251-300Cal;
Cat. No.: 100252-500Cal; and Cat. No.: 100253-1000Cal.) * Can also be run as FIXED1
BAS100248SQ.05 09/13
DRI® OXYCODONE ASSAY, Semi-Quantitative, AU480
Beckman Coulter System Reagent: Microgenics 100248, 100249 Reagent ID: Specific Test Parameters General LIH ISE Range Test Name: OXY Type: URINE Operation: Yes Sample Volume 10.5 L Dilution 0 L OD Limit Pre-Dilution Rate 1 Min. OD 0.00 Max. OD 3.00 Reagents Volume: R1(R1-1) 87 L Dilution 0 L Reagent OD limit: First Low 0.00 High 3.00 Last Low 0.00 High 3.00 R2 (R2-1) 87 L Dilution 0 L Dynamic Range Low -99999 High 99999 Correlation Factor A 1 B 0 Wavelength: Pri. 340 nm Sec. 410 nm Factor for Maker A 1 B 0 Method: RATE* Reaction slope: + Onboard Stability 30 Days 0 Hour Measuring Point 1: First 17 Last 21 Measuring Point 2: First Last Linearity: 100 % No Lag Time: NO
Specific Test Parameters
General ISE Range
Test Name: OXY Type: URINE Value/Flag: VALUE Level L: -99999 Level H: # Specific Ranges: Panic Value From To
Sex Year Month Year Month Low High Low High 1. # # # # # # # # # 2. # # # # # # # 3. # # # # # # # 4. # # # # # # # 5. # # # # # # # 6. # # # # # # # 7. No demographics # # 8. Not within expected values # # Unit ng/mL Decimal Places #
Calibration Specific General ISE Test Name: OXY Type URINE Use Serum Cal. Calibration Type: 5AB Formula: POLYGONAL Counts: 2 <Calibrator Parameters> Slope Check None Calibrator † OD Conc † Factor Range Low High Allowable Range Check Point 1: # 0 -2.00 2.50 Reagent Blank
Point 2: # 100 -2.00 2.50 Calibration
Point 3: # 300 -2.00 2.50 Advanced Calibration
Point 4: # 500 -2.00 2.50 Operation NO
Point 5: # 1000 -2.00 2.50 Interval (RB/ACAL)
Point 6:
Point 7:
Point 8:
Point 9:
Point10:
<Point Cal. For Master Curve> No. of Correction Points Use Master Curve Lot Calibration Calibrator OD Conc OD Range Stability Low High Reagent Blanks 14 Day 0 Hour
Point 1: Calibration 14 Day 0 Hour Point 2: MB Type Factor: 1-Point Calibration Point With CONC-0
# User Defined † DRI
® Calibrators (Cat. No. 1644 – NegCal; Cat. No.: 100250-100Cal; Cat. No.: 100251-300Cal;
Cat. No.: 100252-500Cal; and Cat No.: 100253-1000Cal.)) * Can also be run as FIXED
BAS100248SQ.05 09/13
DRI® OXYCODONE ASSAY, Semi-Quantitative, AU680
Beckman Coulter System Reagent: Microgenics 100248, 100249 Reagent ID: Specific Test Parameters General LIH ISE Range Test Name: OXY Type: URINE Operation: Yes Sample Volume 10.5 L Dilution 0 L OD Limit Pre-Dilution Rate 1 Min. OD 0.00 Max. OD 3.00 Reagents Volume: R1(R1-1) 87 L Dilution 0 L Reagent OD limit: First Low 0.00 High 3.00 Last Low 0.00 High 3.00 R2 (R2-1) 87 L Dilution 0 L Dynamic Range Low -99999 High 99999 Common Reagent Type None Name Correlation Factor A 1 B 0 Wavelength: Pri. 340 nm Sec. 410 nm Factor for Maker A 1 B 0 Method: RATE* Reaction slope: + Onboard Stability 30 Days 0 Hour Measuring Point 1: First 17 Last 21 Measuring Point 2: First Last Linearity: 100 % No Lag Time: NO
Specific Test Parameters
General ISE Range
Test Name: OXY Type: URINE Value/Flag: VALUE Level L: -99999 Level H: # Specific Ranges: Panic Value From To
Sex Year Month Year Month Low High Low High 1. # # # # # # # # # 2. # # # # # # # 3. # # # # # # # 4. # # # # # # # 5. # # # # # # # 6. # # # # # # # 7. No demographics # # 8. Not within expected values # # Unit ng/mL Decimal Places #
Calibration Specific General ISE Test Name: OXY Type URINE Use Serum Cal. Calibration Type: 5AB Formula: POLYGONAL Counts: 2 <Calibrator Parameters> Slope Check None Calibrator † OD Conc † Factor Range Low High Allowable Range Check Point 1: # 0 -2.00 2.50 Reagent Blank
Point 2: # 100 -2.00 2.50 Calibration
Point 3: # 300 -2.00 2.50 Advanced Calibration
Point 4: # 500 -2.00 2.50 Operation NO
Point 5: # 1000 -2.00 2.50 Interval (RB/ACAL)
Point 6:
Point 7:
Point 8:
Point 9:
Point10:
<Point Cal. For Master Curve> No. of Correction Points Use Master Curve Lot Calibration Calibrator OD Conc OD Range Stability Low High Reagent Blanks 14 Day 0 Hour
Point 1: Calibration 14 Day 0 Hour Point 2: MB Type Factor: 1-Point Calibration Point With CONC-0
# User Defined † DRI
® Calibrators (Cat. No. 1644 – NegCal; Cat. No.: 100250-100Cal; Cat. No.: 100251-300Cal;
Cat. No.: 100252-500Cal; and Cat No.: 100253-1000Cal.) * Can also be run as FIXED
BAS100248SQ.05 09/13
Beckman Coulter System Reagent, AU600
DRI® Oxycodone Assay, Semi-Quantitative
Cat. No.: Microgenics 100248, 100249
Reagent ID:
Specific test parameters
Test No # Test name OXY Sample type URINE Page 1/2
Sample vol. 10.5 Dil. vol 0 l Min. OD Max. OD
Reagent 1 vol 87 Dil. vol 0 l L 0.00 H 2.500
Reagent 2 vol 87 Dil. vol 0 l Reagent OD limit
Fst. L 0.00 Fst. H 2.500
Wave Main 340 Sub 410 Lst. L 0.00 Lst. H 2.500
Method RATE1* Dynamic range
Reaction + L -99999 H 1000
Point 1 Fst 17 Lst 21 Correlation factor A 1.00
Point 2 Fst Lst B 0.00
Linearity Fst 100 % Sec %
No lag time NO On-board stability period 30
Select using Space key, or select from list displayed by Guide key
Test No # Test name OXY Sample type URINE Page 2/2
Level L Level H
Value/flag VALUE # #
Normal range Sex Age L Age H L H
1 # # Y # M # Y # M # #
2 # # Y # M # Y # M # #
3 # # Y # M # Y # M # #
4 # # Y # M # Y # M # #
5 # # Y # M # Y # M # #
6 # # Y # M # Y # M # #
7 Non select # #
8 Out of range # #
L H
Panic value # #
Select the function using the Function key or the Mouse
Calibration specific
Test No # Name OXY
Cal type 13 5AB Count 2
Formula 3 POLYGONAL Process Conc
Selection calibrator Cal No OD Conc † Factor/OD-L Factor/OD-H
Point 1 # 0 -2.00 2.500
Point 2 # 100 -2.00 2.500
Point 3 # 300 -2.00 2.500
Point 4 # 500 -2.00 2.500
Point 5 # 1000 -2.00 2.500
Point 6
Point 7
1-point cal. point
MB type factor
Calibrator stability period 14
Select the function using the Function key or the Mouse
# User Defined † DRI
® Calibrators (Cat. No.: 1644 – NegCal; Cat. No.: 100250-100Cal; Cat. No.: 100251-300Cal;
Cat. No.: 100252-500Cal; and Cat. No.: 100253-1000Cal.) * Can also be run as FIXED1
BAS100248SQ.05 09/13
Beckman Coulter System Reagent, AU2700/AU5400
DRI® Oxycodone Assay, Semi-Quantitative
Cat. No.: Microgenics 100248,100249
Reagent ID:
Specific Test Parameters
General LIH ISE Range
Test Name: OXY Type: URINE Operation: Yes
Sample: Volume 10.5 L Dilution 0 L Pre-Dilution Rate: 1
Reagents: R1 Volume 87 L Dilution 0 L Min OD Max OD
R2 Volume 87 L Dilution 0 L L 0.00 H 2.50
Reagent OD limit:
Wavelength: Pri. 340 Sec. 410 First L 0.00 First H 2.50
Method: RATE1* Last L 0.00 Last H 2.50
Reaction slope: + Dynamic Range:
Measuring Point 1: First 17 Last 21 L -99999 H 1000
Measuring Point 2: First Last Correlation Factor:
Linearity: 100 % A 1 B 0
No Lag Time: NO On-board stability period: 30
Specific Test Parameters
General LIH ISE Range
Test Name: OXY Type: URINE
Value/Flag: VALUE Level L: -99999 Level H: #
Normal Ranges: Age L Age H
Sex Year Month Year Month L H
1. # # # # # # #
2. # # # # # # #
3. # # # # # # #
4. # # # # # # #
5. # # # # # # #
6. # # # # # # #
7. None Selected # #
8. Out of Range L H # #
Panic Value: # # Unit: ng/mL Decimal places: #
Calibration Specific
General ISE
Test Name: OXY Type URINE
Calibration Type: 5AB Formula: POLYGONAL Counts: 2 Process: CONC
Cal. No. OD CONC† Factor/OD-L Factor/OD-H
Point 1: # 0 -2.00 2.500
Point 2: # 100 -2.00 2.500
Point 3: # 300 -2.00 2.500
Point 4: # 500 -2.00 2.500
Point 5: # 1000 -2.00 2.500
Point 6:
Point 7:
1-Point Cal. Point: With CONC-0 Slope Check None Advanced Calibration: ##
MB Type Factor: Calibration Stability Period: 14
# User Defined ## Lot or Lot + Bottle † DRI
® Calibrators (Cat. No.: 1644 – NegCal; Cat. No.: 100250-100Cal; Cat. No.: 100251-300Cal;
Cat. No.: 100252-500Cal; and Cat. No.: 100253-1000Cal.) * Can also be run as FIXED1
BAS100248SQ.05 09/13
DRI® OXYCODONE ASSAY, Semi-Quantitative, AU5800
Beckman Coulter System Reagent: Microgenics 100248, 100249 Reagent ID: Parameters Specific Test Parameters
General LIH ISE Calculated Test Range Test Name: OXY Type: Urine Operation: Yes Sample Volume 10.5 L Dilution 0 L OD Limit Pre-Dilution Rate 1 Diluent Bottle # Min. OD 0.00 Max. OD 3.00 Reagents Volume: R1(R1-1) 87 L Dilution 0 L Reagent OD limit: R1-2 L Dilution L First Low 0.00 High 3.00 Last Low 0.00 High 3.00 R2 (R2-1) 87 L Dilution 0 L Dynamic Range Low -99999 High 99999 Common Rgt. Type None Name None Correlation Factor A 1 B 0 Wavelength: Pri. 340 nm Sec. 410 nm Factor for Maker A 1 B 0 Method: RATE* Reaction slope: + Onboard Stability 30 Day 0 Hour Measuring Point 1: First 17 Last 21 Measuring Point 2: First Last Linearity Limit: 100 % Lag Time Check: NO
Parameters Specific Test Parameters
General LIH ISE Calculated Tests Range
Test Name: OXY Type: Urine Value/Flag: VALUE Level Low -99999 High # Specific Ranges: From To
Sex Year Month Year Month Low High 1. # # # # # # # 2. # # # # # # # 3. # # # # # # # 4. # # # # # # # 5. # # # # # # # 6. # # # # # # # 7. Standard demographics # # 8. Not within expected values # # Panic Value Low # High # Unit ng/mL
Decimal Places #
Parameters Calibration Parameters
Calibrators Calibration Specific General ISE
Test Name: OXY Type Urine Cuvette Use Serum Cal. Calibration Type: 5AB Formula: POLYGONAL Counts: 2 <Calibrator Parameters> Slope Check None Calibrator † OD Conc † Factor Range Low High Point 1: # 0.0 -2.000 2.500
Point 2: # 100.0 -2.000 2.500 Allowable Range Check
Point 3: # 300.0 -2.000 2.500 Reagent Blank
Point 4: # 500.0 -2.000 2.500 Calibration
Point 5: # 1000.0 -2.000 2.500
Point 6: Advanced Calibration
Point 7: Operation No
Point 8: Interval (RB/ACAL)
Point 9:
Point10: Lot Calibration
<Point Cal. For Master Curve> No. of Correction Points Use Master Curve
Calibrator OD Conc OD Range Stability
Low High Reagent Blank 14 Day 0 Hour
Point 1: Calibration 14 Day 0 Hour Point 2: MB Type Factor: 1-Point Calibration Point With Conc-0
# User Defined † DRI
® Calibrators (Cat. No. 1644 – NegCal; Cat. No.: 100250-100Cal; Cat. No.: 100251-300Cal;
Cat. No.: 100252-500Cal; and Cat No.: 100253-1000Cal.) * Can also be run as FIXED