Introduction Fluorescence Oil in Water Calibration
BlueTrace Product Solutions
Functions & Features
Application Areas
BlueTraceFluorescence Sensor for BTEX, PAH & Refined Oils
Solutions by GO Systemelektronik
GO Systemelektronik GmbH 1 www.go-sys.de
Fluorescence spectral data
There is no universal fluorescence spectrum for all oils. Rather, the spectrum depends on the composition of the oil. Refined oils consist mainly of aromatic hydrocarbons, which in solution in water are often indicated in the BTEX collection parameter. The graph shows examples for typical fluorescence spectra of some refined oils. The BlueTrace oil in water sensor is suitable for the measurement of refined oils, BTEX and aromatic hydrocarbons.
Excitation
Hydraulic oil
Rela
tive
light
inte
nsity
Wavelength (nm)250 275 300 325 350 375 400
0
1
S1
Ener
gy
Ligh
t sou
rce
Abso
rptio
n
Fluorescence
Relaxation
Relaxation
Phosphorescence
IntercombinationT1
S0
Light source Detector
Oil particles
When light of a certain wavelength hits an oil particle, the oil emits light of a different wavelength shortly after excitation. This effect is called fluorescence. Fluo-rescence occurs not only in oils, but also in other substances. The BlueTrace oil in water sensor uses this effect to determine the concentration of refined oils in water.A transmitter installed in the sensor emits light at around 280 nm. The oil particles in the water absorb this energy and then emit light in a range from 300 to 400 nm. This light is measured by a detector.
The Jablonski diagram shows the fluorescence effect in detail. The oil particle absorbs the energy of the light, changes to a higher, unstable energy level and then falls back to the lower energy level. Part of the energy is released by the fluorescence effect. The intensity of the fluorescence is directly dependent on the concentration. By measuring the intensity at the detector, the concentration of the oil in the water can be determined.
Introduction Fluorescence - Oil in Water
Rela
tive
light
inte
nsity
Wavelength (nm)250 275 300 325 350 375 400
0
1
Excitation
DieselCooling lubricant
Rela
tive
light
inte
nsity
Wavelength (nm)250 275 300 325 350 375 400
0
1
Excitation
GO Systemelektronik GmbH 1 www.go-sys.de
Fluorescence spectral data
There is no universal fluorescence spectrum for all oils. Rather, the spectrum depends on the composition of the oil. Refined oils consist mainly of aromatic hydrocarbons, which in solution in water are often indicated in the BTEX collection parameter. The graph shows examples for typical fluorescence spectra of some refined oils. The BlueTrace oil in water sensor is suitable for the measurement of refined oils, BTEX and aromatic hydrocarbons.
Excitation
Hydraulic oil
Rela
tive
light
inte
nsity
Wavelength (nm)250 275 300 325 350 375 400
0
1
S1
Ener
gy
Ligh
t sou
rce
Abso
rptio
n
Fluorescence
Relaxation
Relaxation
Phosphorescence
IntercombinationT1
S0
Light source Detector
Oil particles
When light of a certain wavelength hits an oil particle, the oil emits light of a different wavelength shortly after excitation. This effect is called fluorescence. Fluoafter excitation. This effect is called fluorescence. Fluo-rescence occurs not only in oils, but also in other substances. The BlueTrace oil in water sensor uses this effect to determine the concentration of refined oils in water.A transmitter installed in the sensor emits light at around 280 nm. The oil particles in the water absorb this energy and then emit light in a range from 300 to 400 nm. This light is measured by a detector.400 nm. This light is measured by a detector.
The Jablonski diagram shows the fluorescence effect The Jablonski diagram shows the fluorescence effect in detail. The oil particle absorbs the energy of the in detail. The oil particle absorbs the energy of the light, changes to a higher, unstable energy level and light, changes to a higher, unstable energy level and light, changes to a higher, unstable energy level and then falls back to the lower energy level. Part of the energy is released by the fluorescence effect. The intensity of the fluorescence is directly dependent on the concentration. By measuring the intensity at the detector, the concentration of the oil in the water can be determined.
Introduction Fluorescence - Oil in Water
Rela
tive
light
inte
nsity
Wavelength (nm)250 275 300 325 350 375 400
0
1
Excitation
DieselCooling lubricant
Rela
tive
light
inte
nsity
Wavelength (nm)250 275 300 325 350 375 400
0
1
Excitation
GO Systemelektronik GmbH 2 www.go-sys.de
Specific calibration to specific oil typesSince the fluorescence spectrum is extremely depend-ent on the refined oil, the sensor functions optimally with a specific calibration to this oil. Also, the absorp-tion of the water has an influence on the measure-ment, so that for the calibration the oil should be mixed in the water, which is to be expected locally. In low concentrations a linear behaviour can be expect-ed, therefore, a two-point calibration can be sufficient. However, at least a three-point calibration is recom-mended.
Easy calibrationThe BlueTrace can easily be optimally calibrated to the specific application. All you have to do is hold the sensor in the prepared samples and then perform a multi-point calibration. For this purpose, only the reference values have to be entered into the BlueBox controller or the corresponding software.
Calibration
Dark non-reflective surface
Prepared sample
Sensor
Magnetic stirrer
Oil in water concentration (ppm)
Out
put s
igna
l of d
etec
tor (
mV)
0
0 50 100 150 200 250 300
450
400
350
300
250
200
150
100
50
Functions & Features
Selectable Measuring Range
The sensitivity of the receiver can easily be changed either directly on the controllers of GO Systemelek-tronik, or with the help of the freely available PC program. Thus, the sensor can optimally be adjusted to the required measuring range.
Modbus Interface
The BlueTrace features a Modbus RTU interface. This means that the sensor can not only be connected to a GO Systemelektronik controller, but can also be inte-grated into third-party controllers or directly to a PLC. The necessary Modbus protocol is freely available. Settings or calibrations are stored directly on the sensor and can be adapted with the freely available PC program. The sensor only needs to be connected to a PC, for example with a Modbus USB converter.
Robust Design
The BlueTrace is available in stainless-steel and titani-um. For most applications, stainless steel is sufficient. However, if the sensor is used in corrosive media or harsh environmental conditions, it is recommended to use the BlueTrace in a titanium housing.
Hydraulic oil
Diesel
Cooling lubricant
www.go-sys.de
The BlueTrace is available in stainless-steel and titani-um. For most applications, stainless steel is sufficient. However, if the sensor is used in corrosive media or harsh environmental conditions, it is recommended to
GO Systemelektronik GmbH
Specific calibration to specific oil typesSince the fluorescence spectrum is extremely dependent on the refined oil, the sensor functions optimally with a specific calibration to this oil. Also, the absorption of the water has an influence on the measurement, so that for the calibration the oil should be mixed in the water, which is to be expected locally. In low concentrations a linear behaviour can be expected, therefore, a two-point calibration can be sufficient. However, at least a three-point calibration is recommended.
Easy calibrationThe BlueTrace can easily be optimally calibrated to the specific application. All you have to do is hold the sensor in the prepared samples and then perform a multi-point calibration. For this purpose, only the reference values have to be entered into the BlueBox controller or the corresponding software.
CalibrationCalibration
Prepared sample
Out
put s
igna
l of d
etec
tor (
mV)
Out
put s
igna
l of d
etec
tor (
mV)
Out
put s
igna
l of d
etec
tor (
mV)
0
0 50
450
400
350
300
250
200
150
100
50
GO Systemelektronik GmbH 3 www.go-sys.deGO Systemelektronik GmbH
Application Areas - BlueTrace
■ Cooling water (e.g. in power plants)■ Leakage detection
Process Monitoring
■ Detection of contamination in rivers and lakes■ Maritime applications
Environmental Monitoring
■ Influent of dirnking water plants■ Influent to desalination plants
Drinking Water
■ Influent of WWTP■ Monitoring of wastewater of industrial plants (e.g. refineries, tire manufacturers, petrochemical industry)
Wastewater
Application Areas - BlueTraceApplication Areas - BlueTrace
www.go-sys.de3
Application Areas - BlueTrace
■ Cooling water Cooling water (e.g. in power plants)■ Leakage detection
Process Monitoring
■ Detection of contamination in rivers and lakes■ Maritime applications
Environmental Monitoring
■ Influent of dirnking water plants■ Influent to desalination plants
Drinking Water
■ Influent of WWTP■ Monitoring of wastewater of industrial plants(e.g. refineries, tire manufacturers, petrochemical industry)
Wastewater
Application Areas - BlueTrace
GO Systemelektronik GmbH 4 www.go-sys.de
Power supply 10 - 32 V DC
Power consumption (typical) 0.5 W
Measuring principle Fluorescence measurementEvaluation (300 - 400 nm)
Measuring range (typical) 0 - 30 / 100 / 300 ppm
Measuring accuracy 3 % FS
Detection limit (typical) 0.1 ppm
Measuring interval ≥ 1 s
Light source < 300 nm
Material Stainless steel 1.4404 / Titanium [optional]
Operation temperature range 0 °C to +55 °C
Weight 0.8 kg
Dimensions Length 150 mm; Ø 38 mm
Interface Modbus [RTU]
Maximum pressure 6 bar
Art. no. 461 6200
Technical dataParameters■ BTEX■ PAH■ Refined oil in water
Application Areas
Environmental Monitoring■ River water■ Surface water
Wastewater■ Effluent monitoring■ Trend analysis■ Early detection of discharge
Process Measurement & Control Technology■ Process monitoring in industrial facilities■ Control of process water treatment■ Process optimization
Drinking Water■ Quality control■ Alarm systems
The BlueTrace Oil in Water Sensor is a compact fluorescence probe for the measurement of BTEX, PAH and refined oil in water. The robust design of the Blue-Trace Sensor allows for applications in harsh condi-tions, such as corrosive media or at high pressure.
BlueTrace Oil in Water Sensor
Main Functions & Features
Selectable Measuring Range
Robust & Non-corrosive Modbus InterfaceCompact Design
Easy Calibration
Low Power Demand
Insensitive to Ambiant Light High Pressure Resistance
GO Systemelektronik GmbH 4 www.go-sys.de
Power supply 10 - 32 V DC
Power consumption (typical) 0.5 W
Measuring principle Fluorescence measurementEvaluation (300 - 400 nm)
Measuring range (typical) 0 - 30 / 100 / 300 ppm
Measuring accuracy 3 % FS
Detection limit (typical) 0.1 ppm
Measuring interval ≥ 1 s
Light source < 300 nm
Material Stainless steel 1.4404 / Titanium [optional]
Operation temperature range 0 °C to +55 °C
Weight 0.8 kg
Dimensions Length 150 mm; Ø 38 mm
Interface Modbus [RTU]
Maximum pressure 6 bar
Art. no. 461 6200
Technical dataParameters■ BTEX■ PAH■ Refined oil in water
Application Areas
Environmental Monitoring■ River water■ Surface water
Wastewater■ Effluent monitoring■ Trend analysis■ Early detection of discharge
Process Measurement & Control Technology■ Process monitoring in industrial facilities■ Control of process water treatment■ Process optimization
Drinking Water■ Quality control■ Alarm systems
The The BlueTrace Oil in Water Sensor is a BlueTrace Oil in Water Sensor is a BlueTrace Oil in Water Sensor compact fluorescencefluorescence probe for the measurement of BTEX, PAH and refined oil in water. The robust design of the Blueand refined oil in water. The robust design of the Blueand refined oil in water. The robust design of the Blue-Trace Sensor allows for applications in harsh condi-tions, such as corrosive media or at high pressure.
BlueTrace Oil in Water SensorBlueTrace Oil in Water Sensor
Main Functions & Features
Selectable Measuring Range
Robust & Non-corrosive Modbus InterfaceCompact Design
Easy Calibration
Low Power Demand
Insensitive to Ambiant Light High Pressure Resistance
GO Systemelektronik GmbH [email protected]+49 431 5 80 80-0 www.go-sys.de+49 431 5 80 80-11
Main Functions & Features
Cloud Data Service
SD Card Data Logger
Modular & Expandable
PID-Controller FunctionMonitoring Function
CAN bus, Modbus & ProfibusIntelligent Event Handling
1
2 Sensor Inputs
Parameters / Sensors■ Blue-green algae■ Chlorine■ Chlorine dioxide■ Chlorophyll a■ Conductivity■ Dissolved organic substances■ Dissolved oxygen■ ISE■ Level■ Oil in water
■ Ozone■ pH■ Redox (ORP)■ Salinity■ Temperature■ Turbidity■ Water pressure
Application Areas
Environmental Monitoring■ River water■ Surface water
Wastewater■ Effluent monitoring■ Trend analysis■ Early detection of discharge (fingerprint)
Process Measurement & Control Technology■ Process monitoring in industrial facilities■ Control of process water treatment■ Process optimization
Drinking Water■ Quality control■ Alarm systems
The BlueSense transducer facilitates the processing and calculation of measurement values of physical and chemical sensors. In addition, the transducer features a full controller function that can cover all process-oriented tasks.The BlueSense transducer is compatible with analo-gue and digital sensors and actuators of all manufac-turers.
BlueSense Transducer
GO Systemelektronik GmbH [email protected]+49 431 5 80 80-0 www.go-sys.de+49 431 5 80 80-11
Technical data
Power consumption (typical) 8 W
Dimensions (wxhxd) 23.5 x 18.5 x 11.9 cm
IP protection class IP 65
Graphic touch display 240 x 128 pixel
12 V DC (9 - 18 V)Power supply24 V DC (18 - 36 V)
230 V AC (90 - 260 V)
Ambient temperature -10 to +45 °C
Optional interfaces1x RS-232, RS-485, var. protocols e.g. Modbus
1x Ethernet [TCP/IP], Modbus [TCP/IP]
Profibus
GPRS / UMTS / LTE modem
1x CAN bus for connection to ISA, BlueBox, BlueMon & MPS
Inputs2x Sensor input for selectable parameters
2x Pulse-In (PNP/NPN selectable)
2x Digital-In (static) potential-free contacts
Outputs2x Current output 4-20 mA
2x Relay with a switching capacity of 24 V AC /DC; 0,5 A
2x Relay with a switching capacity of 230 V AC/DC; 2 A
Features & functionsPID controller
Signal compensation
Data storage on SD card
Multipoint calibration
The BlueSense transducer can optimally be adapted to the customer specific application through the modular config-uration of the sensor inputs.
Art. no. 485 0001BlueSense
Product configuration