Calibration and Specification Considerations
When Using Modular Instrumentation
Mike Dobbert
![Page 2: Calibration and Specification Considerations When Using Modular Instrumentation](https://reader038.vdocument.in/reader038/viewer/2022103103/5681638c550346895dd48015/html5/thumbnails/2.jpg)
Environmental ConditionsOperating SoftwareMulti-module Instruments
Modular Instruments
Configuration FlexibilityInterchangeabilityMeasurement SpeedSize
Traceable Measurements!(calibration required)
![Page 3: Calibration and Specification Considerations When Using Modular Instrumentation](https://reader038.vdocument.in/reader038/viewer/2022103103/5681638c550346895dd48015/html5/thumbnails/3.jpg)
Modules have neighbors.
Environmental Conditions
![Page 4: Calibration and Specification Considerations When Using Modular Instrumentation](https://reader038.vdocument.in/reader038/viewer/2022103103/5681638c550346895dd48015/html5/thumbnails/4.jpg)
Fixed Configuration.
Ambient environmental conditions are a good predictor of the internal operating conditions.
Environmental Conditions
![Page 5: Calibration and Specification Considerations When Using Modular Instrumentation](https://reader038.vdocument.in/reader038/viewer/2022103103/5681638c550346895dd48015/html5/thumbnails/5.jpg)
Environmental Conditions
![Page 6: Calibration and Specification Considerations When Using Modular Instrumentation](https://reader038.vdocument.in/reader038/viewer/2022103103/5681638c550346895dd48015/html5/thumbnails/6.jpg)
Calibration Laboratory
Ambient environmental conditions outside the chassis are not necessarily a good indication of the internal operating conditions!
Environmental Conditions
![Page 7: Calibration and Specification Considerations When Using Modular Instrumentation](https://reader038.vdocument.in/reader038/viewer/2022103103/5681638c550346895dd48015/html5/thumbnails/7.jpg)
Environmental Conditions
![Page 8: Calibration and Specification Considerations When Using Modular Instrumentation](https://reader038.vdocument.in/reader038/viewer/2022103103/5681638c550346895dd48015/html5/thumbnails/8.jpg)
Environmental Conditions
![Page 9: Calibration and Specification Considerations When Using Modular Instrumentation](https://reader038.vdocument.in/reader038/viewer/2022103103/5681638c550346895dd48015/html5/thumbnails/9.jpg)
Instrument Drivers (reside outside the module)Installed Drivers at the Calibration Laboratory
Ideally, the same software driver revision is used for both calibration and during intended operation.
Operating Software
![Page 10: Calibration and Specification Considerations When Using Modular Instrumentation](https://reader038.vdocument.in/reader038/viewer/2022103103/5681638c550346895dd48015/html5/thumbnails/10.jpg)
Operating Software
![Page 11: Calibration and Specification Considerations When Using Modular Instrumentation](https://reader038.vdocument.in/reader038/viewer/2022103103/5681638c550346895dd48015/html5/thumbnails/11.jpg)
Set of modules working together as a single instrument.Instrument accuracy may depend on more than one module.
Multi-Module Instruments
Calibrate the multi-module instrument as a single instrument.Multi-module instruments require a unique identification(when calibrated as a set).
![Page 12: Calibration and Specification Considerations When Using Modular Instrumentation](https://reader038.vdocument.in/reader038/viewer/2022103103/5681638c550346895dd48015/html5/thumbnails/12.jpg)
Calibration Manager
Multi-Module Instruments
![Page 13: Calibration and Specification Considerations When Using Modular Instrumentation](https://reader038.vdocument.in/reader038/viewer/2022103103/5681638c550346895dd48015/html5/thumbnails/13.jpg)
Multi-Module Instruments
![Page 14: Calibration and Specification Considerations When Using Modular Instrumentation](https://reader038.vdocument.in/reader038/viewer/2022103103/5681638c550346895dd48015/html5/thumbnails/14.jpg)
Model multi-module instruments as a function of individual modules.
(GUM method of uncertainty propagation.)
Multi-Module Instruments
![Page 15: Calibration and Specification Considerations When Using Modular Instrumentation](https://reader038.vdocument.in/reader038/viewer/2022103103/5681638c550346895dd48015/html5/thumbnails/15.jpg)
Multi-Module Instruments
饾挌=饾拠 ( 饾挋饾煆 , 饾挋饾煇 ,饾挋饾煈 )
Tradeoff: modular flexibility versus accuracy.
![Page 16: Calibration and Specification Considerations When Using Modular Instrumentation](https://reader038.vdocument.in/reader038/viewer/2022103103/5681638c550346895dd48015/html5/thumbnails/16.jpg)
In-Situ Calibration
Address environment and controlling software issues.
![Page 17: Calibration and Specification Considerations When Using Modular Instrumentation](https://reader038.vdocument.in/reader038/viewer/2022103103/5681638c550346895dd48015/html5/thumbnails/17.jpg)
Conclusion
Environmental Conditions
Module environment highly dependent on system configuration.
Operating Software
Instrument driver resides outside the module.
Multi-module Instruments
Calibration of a set of modules. Calibration of individual modules
utilizing the method of the GUM.