ECG(Electrocardiogram)
Achmad RizalBioSPIN
ARL-EL4703-Instrumentasi Biomedis 2
Cardiac Conduction System (1)
ARL-EL4703-Instrumentasi Biomedis 3
Cardiac Conduction System (2)
ARL-EL4703-Instrumentasi Biomedis 4
The ECG (1)
ARL-EL4703-Instrumentasi Biomedis 5
The ECG (1)
ARL-EL4703-Instrumentasi Biomedis 6
The ECG (1)Electrocardiography1. Standard clinical ECG (12 leads)2. VCG (3 orthogonal leads)3. Monitoring ECG (1 or 2 leads)
2
ARL-EL4703-Instrumentasi Biomedis 7
Modern ECG
Figures above show how the earliest ECGs were recorded by Einthoven at 1900.Vats of salt water provided the electrical connection to the body. The string galvanometer served as the measurement instrument for recording the ECG. ARL-EL4703-Instrumentasi Biomedis 8
Electrocardiograph
ARL-EL4703-Instrumentasi Biomedis 9
Holter Monitor
ARL-EL4703-Instrumentasi Biomedis 10
Dipole Model
ARL-EL4703-Instrumentasi Biomedis 11
Standard Location ?
ARL-EL4703-Instrumentasi Biomedis 12
ECG Lead Configurations • A pair of electrodes, or a combination of several electrodes
through a resistive network that gives an equivalent pair is called a lead.
• A vector that connects a lead electrode pair is the lead vector.• If the cardiac vector M is known, the voltage generated at any
lead can be easily computed.
3
ARL-EL4703-Instrumentasi Biomedis 13
Frontal Plane Bipolar LeadsEinthovenTriangle
ARL-EL4703-Instrumentasi Biomedis 14
The Einthoven Triangle
ARL-EL4703-Instrumentasi Biomedis 15
Lead I
ARL-EL4703-Instrumentasi Biomedis 16
Lead II
ARL-EL4703-Instrumentasi Biomedis 17
Lead III
ARL-EL4703-Instrumentasi Biomedis 18
Cardiac Vector Through the Cardiac Cycle As Seen By Leads I, II and III
4
ARL-EL4703-Instrumentasi Biomedis 19
Cardiac Vector Through the Cardiac Cycle As Seen By Leads I, II and III
ARL-EL4703-Instrumentasi Biomedis 20
Cardiac Vector Through the Cardiac Cycle As Seen By Leads I, II and III
ARL-EL4703-Instrumentasi Biomedis 21
Cardiac Vector Through the Cardiac Cycle As Seen By Leads I, II and III
ARL-EL4703-Instrumentasi Biomedis 22
UnipolarMeasurements:Wilson Central Terminal
• Three additional leads are used for frontal plane measurements.
• These are the measurements at the specific electrode, with respect to a reference electrode
• One commonly used reference electrode is the Wilson Central Terminal, obtained through a resistive network, combining limb electrodes
• The new set of leads obtained by combining the standard limb electrodes to the Wilson terminal form the augmented leads
• These leads provide additional vector views of cardiac depolarization in the frontal plane. Unlike leads I, II, III, the augmented leads utilize WCT, a central negative terminal. This virtual "electrode" is calculated by the EKG computer to measure vectors originating roughly at the center of the heart.
Note that the voltage at Wilson’s terminal is zero (why?)
ARL-EL4703-Instrumentasi Biomedis 23
Wilson’s Central Terminal
ARL-EL4703-Instrumentasi Biomedis 24
Augmented Leads: aVR
21IIIVVaVR −−=
5
ARL-EL4703-Instrumentasi Biomedis 25
Augmented Leads: aVL
21IIVVaVL −=
ARL-EL4703-Instrumentasi Biomedis 26
Augmented Leads: aVF
2I
IIVVaVF −=
ARL-EL4703-Instrumentasi Biomedis 27
All Frontal Leads
ARL-EL4703-Instrumentasi Biomedis 28
Frontal Lead
ARL-EL4703-Instrumentasi Biomedis 29
Transversal Lead
ARL-EL4703-Instrumentasi Biomedis 30
12 Leads ECG Recording
6
ARL-EL4703-Instrumentasi Biomedis 31 ARL-EL4703-Instrumentasi Biomedis 32
ECG Intrumentation
• Single Channel ECG
ARL-EL4703-Instrumentasi Biomedis 33
ECG Intrumentation
Electrode TransducerCharacteristic :
acquiring signal with amplitude between 0.01mV – 3mVanticipate impedance till > 5 MΩsmall current leakages < 1 µ-Aresponse frequency 0.05 – 150 Hz
ARL-EL4703-Instrumentasi Biomedis 34
ECG Intrumentation
ECG Amplifier schematicsRight leg driven loop- noise reducing- patient protection- Reduce Common Mode
Voltage
ARL-EL4703-Instrumentasi Biomedis 35
ECG Intrumentation
• Instrumentation Amplifier
ARL-EL4703-Instrumentasi Biomedis 36
ECG Intrumentation
• AD620 single chip instrumentation amplifier
7
ARL-EL4703-Instrumentasi Biomedis 37
ECG IntrumentationHigh Pass Filter Second Amplifier
Low Pass Filter 20 Hz
Clamper
Low Pass Filter 40 Hz
Analog subsystemARL-EL4703-Instrumentasi Biomedis 38
ECG Intrumentation
Digital Subsystem
ARL-EL4703-Instrumentasi Biomedis 39
ECG Intrumentation
Display Unit• Galvanometer• Osciloskop• CRT• LCD• PC + Monitor• Paper chart• PDA/HP display
ARL-EL4703-Instrumentasi Biomedis 40
Some Problems in ECG
Noise & interferences
ARL-EL4703-Instrumentasi Biomedis 41
Some Problems in ECG
• Souces of interference in an ECG trace– base line wander : patient movement, electrode
reposition, dry electrode, static buildup around patient
– AC noise (50 Hz) : powerline/common mode interfrence
– intermittent signal : lead connection not tight and secured, dry electrode, damage cable or lead wire
– Other Electrical interference (EI)
Another ECG System
8
ARL-EL4703-Instrumentasi Biomedis 43
Glossary of Terms• AED : Automated External Defibrillator – Auto diagnosis with voice
directed treatment.• EKG/ECG: ElectroCardioGram – skin surface, heart monitor• Holter Monitor: Portable device for continuously monitoring the electrical
activity of the heart for at least 24 hours. Can be external or implantable.
• Defibrillation : Delivering a therapeutic dose of electrical energy to the heart.• T-wave alterans : A change in the shape of the heartbeat signal on an every
other beat basis that is not apparent to the naked eye.• Right Leg Drive : An operational amplifier deriving common mode voltage is
used to invert the common-mode signal and drive it back into thepatient through the right leg.
• RFI : Radio Frequency Interference• ESD : Electro Static Discharge
ARL-EL4703-Instrumentasi Biomedis 44
ECG Markets ECG
Professional Healthcare Consumer
Diagnostic
Above Diagnostic
AED
Defibrillators
Commercial Fitness Equipment
Home Products
line battery line battery
Holter Monitor
ARL-EL4703-Instrumentasi Biomedis 45
Diagnostic ECG (Clinical Use)
AD8220
InstrumentationAmplifier
+5V
-5V
A B
C
24.9kΩ24.9kΩ
AD7980
ADum240X
BlackFin
ADC
DSP
iCoupler
REF 22μF
ReferenceADR435
4.12kΩ
Additional Channels
Additional Channels
AD8618
+5V
4.7μF
14kΩ
+5V
+5V
-5V +5V
-5V
REF/2
REF/2
1MΩ
47nF
10kΩ
10kΩ
2.2pF
2.2pF
10pF
G=14
12.7kΩ 12.7kΩ
866kΩ
68pF
499kΩ
OP2177
+5V
OP2177+5V
op amps-5V
-5V
AD8618
+5V
14.5kΩ
14.5kΩ
68nF
22nF
REF/2
Low Pass 5th Order Filter at 157Hz
switchADG601
From Lead I
From Lead IIFrom Lead III
ADum240X
AD8618
+5V
19.3kΩ
19.3kΩ
33nF
33nF
220pF
REF/2Pacing Artifact
Detection Circuitry
15kΩ
Lead Detection
50kHz
866kΩ
ADG620
REF/2
AD8618
+5V
57.6kΩ1.18kΩ 14kΩ
High Pass Filter 0.033Hz G=50
1.15kΩ
4.99kΩ
33Ω
10nF
ARL-EL4703-Instrumentasi Biomedis 46
Diagnostic ECG In-amp + Right Leg Drive
AD8220
InstrumentationAmplifier
+5V
-5V
A B
C
24.9kΩ24.9kΩ
4.12kΩ
4.7μF
+5V
-5V +5V
-5V
REF/2
10kΩ
10kΩ
2.2pF
2.2pF
10pF
G=14
12.7kΩ 12.7kΩ
866kΩ
68pF
499kΩ
OP2177
+5V
OP2177+5V
op amps-5V
-5V
switchADG601
From Lead I
From Lead IIFrom Lead III
220pF
15kΩ
866kΩ
ADG620
ARL-EL4703-Instrumentasi Biomedis 47
Diagnostic ECG Filters + Pace Detection
AD8618
+5V
4.7μF
14kΩ
REF/2
REF/2
1MΩ
47nF
12.7kΩ
AD8618
+5V
14.5kΩ
14.5kΩ
68nF
22nF
REF/2
Low Pass 5th Order Filter at 157Hz
switchADG601
From Lead IIFrom Lead III
AD8618
+5V
19.3kΩ
19.3kΩ
33nF
33nF
220pF
REF/2Pacing Artifact
Detection Circuitry
Lead Detection
50kHz
866kΩ
ADG620
REF/2
AD8618
+5V
57.6kΩ1.18kΩ 14kΩ
High Pass Filter 0.033Hz G=50
1.15kΩ
4.99kΩ
33Ω
10nF
ARL-EL4703-Instrumentasi Biomedis 48
Diagnostic ECG ADC + Isolation + Processing
AD7980
ADum240X
BlackFin
ADC
DSP
iCoupler
REF 22μF
ReferenceADR435
Additional Channels
Additional Channels
+5V
ADum240X
33Ω
10nF
9
ARL-EL4703-Instrumentasi Biomedis 49
Low Power, Low Cost, Wireless ECG (Holter Monitor)
ESDDefibrillation
Protection
ADuC7022
ADF702x*
ESDDefib
Protect
RLElectrode
RAElectrode
RLElectrode
AD623G=10.6
High-Pass Pole0.5Hz
0.47μF
5pF
681kΩ
499kΩ
5.23kΩ 5.23kΩ
475Ω 10kΩ
108pF
Anti-alias Filter-3dB, 1592 Hz
AdditionalDigital Filter
Level ShiftAnti-Alias
Total Gain ~ 220
G = 22.05
AD8641
AD8603
Dependent on Frequency Band Used (WMTS or ISM)
ARL-EL4703-Instrumentasi Biomedis 50
Low Cost, Low Power AED
Filterand
ProtectionNetwork
ADR365
AD7685
HighVoltage
Discharge
PowerControl
ADP1712
AD8220/4AD8606
AD8605
PatientPads
Li IonBattery
ADuC70xxor
Blackfin®
LCD Display
Single poleanti-alias
filter
Digital Filtering
Blanking
ARL-EL4703-Instrumentasi Biomedis 51
Exercise Heart Rate Monitor(Example: treadmill with heart rate monitor)
AD8692AD8692
REF
REF REF
AD8553
AD8691
High Pass Filter
ADuC70xxREF
InputProtection
+5V
40 Hz Low Pass Filter