END TIDAL CO2
(ETCO2) MONITORING
![Page 2: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/2.jpg)
EtCO2 Monitoring
“Snapshot in time” Assists with patient assessment BUT:
–Do NOT replace eyes-on/hands-on care–Are just one piece of clinical judgment–ALL have pitfalls/malfunctions/limitations–Is more complex than ever
![Page 3: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/3.jpg)
EtCO2 Monitoring
Non-invasive method of determining Carbon Dioxide levels in
intubated and non-intubated patients
Uses infra-red technology, to monitor exhaled breath to determine CO2 levels numerically and by
waveform (capnogram).
![Page 4: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/4.jpg)
EtCO2 Monitoring
EtCO2 is directly related to the ventilation status of the patient (as opposed to SAo2, which relates oxygenation of the patient)
Capnography can be used to verifyendotracheal tube/Combi-Tube & King Airway placement and monitor its position, assess ventilation and treatments, and to evaluate resuscitative efforts during CPR
![Page 5: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/5.jpg)
EtCO2 Monitoring Review of Pulmonary
Anatomy & Physiology
The primary function of therespiratory system is toexchange carbon dioxide foroxygen.
During inspiration, air enters theupper airway via the
nose where it is warmed, filtered, and humidified
The inspired air flows throughthe trachea and bronchial
tree to enter the pulmonary alveoli where the oxygen diffuses across the alveolar capillary membrane into the blood.
Nasal Passages
Roof of the Mouth
Epiglottis
Trachea (windpipe)Esophagus (food tube)
Alveoli
BronchiPulmonary VeinBronchiole
![Page 6: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/6.jpg)
Cellular Ventilation
EtCO2 Monitoring
![Page 7: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/7.jpg)
EtCO2 Monitoring
Alveolar Ventilation
![Page 8: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/8.jpg)
EtCO2 Monitoring
Measurement methods Single, one-point-in-time (Easy-Cap).
Electronic devices Continuous information Utilize infrared (IR) spectroscopy to measure the
CO2 molecules’ absorption of IR light as the light passes through a gas sample.
![Page 9: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/9.jpg)
EtCO2 Monitoring Electronic Devices:
Mainstream Located directly on the patient’s endotracheal tube
Sidestream Remote from the patient.
Mainstream sampling Occurs at the airway of an intubated patient Was not originally intended for use on non-intubated
patients. Heavy and bulky adapter and sensor assemblies may
make this method uncomfortable for non-intubated patients.
![Page 10: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/10.jpg)
EtCO2 Monitoring Sidestream sampling
Exhaled CO2 is aspirated (at 50ml/min) via ETT, cannula, or mask through a 5–10 foot long
sampling tube connected to the instrument for analysis
Both mainstream and sidestream technologies calculate the CO2 value and waveform.
![Page 11: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/11.jpg)
EtCO2 Monitoring A new technology, Microstream, utilizes a
modified sidestream sampling method, and employs a microbeam IR sensor that
specifically isolates the CO2 waveform.
Microstream can be used on both intubatedand non-intubated patients.
![Page 12: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/12.jpg)
EtCO2 Monitoring• Continuous EtCO2 monitoring = changes are immediately seen
(CO2 diffuses across the capillary-alveolar membrane <½ second)• Sa02 monitoring is also continuous, but relies on trending.
- and -• The oxygen content in blood can maintain for
several minutes after apnea (especially w/ pre-oxygenation)
![Page 13: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/13.jpg)
EtCO2 Monitoring Definitions
Tachypnea Abnormally rapid respiration
Hyperventilation Increased minute volume that results in lowered CO2
levels (hypocapnia)
Hypoventilation Reduced rate & depth of breathing that causes an
increase in carbon dioxide (hypercapnia)
![Page 14: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/14.jpg)
EtCO2 Monitoring EtCO2 Numerical Values
(Ventilatory Assessment)
Normal = 35-45mmHg
< 35mmHg = Hyperventilation Respiratory alkalosis
> 45mmHg = Hypoventilation Respiratory acidosis
![Page 15: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/15.jpg)
EtCO2 Monitoring EtCO2 Numerical Values
(Metabolic Assessment)
Normal = 35-45mmHg < 35mmHg = Metabolic Acidosis > 45mmHg = Metabolic Alkalosis
Dependant on 3 variables CO2 production Delivery of blood to lungs Alveolar ventilation
![Page 16: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/16.jpg)
EtCO2 Monitoring Increased EtCO2
Decreased CO2 clearance Decreased central drive Muscle weakness Diffusion problems
Increased CO2 Production Fever Burns Hyperthyroidism Seizure Bicarbonate Rx ROSC Release of tourniquet/Reperfusion
![Page 17: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/17.jpg)
EtCO2 Monitoring Decreased EtCO2
Increased CO2 Clearance Hyperventilation Acidosis ( ↓ HCO3 levels 2° to ↑ Hydrogen) Decreased CO2 production
Hypothermia Sedation Paralysis
Decreased Delivery to Lungs Decreased cardiac output
V/Q Mismatch Ventilating non-perfused lungs (pulmonary
edema)
![Page 18: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/18.jpg)
EtCO2 Monitoring
Ventilation/Perfusion Ratio (V/Q)• Effective pulmonary gas exchange depends on
balanced V/Q ratio• Alveolar Dead Space (atelectasis/pneumonia)
(V > Q = CO2 content)• Shunting (blood bypasses alveoli w/o picking up o2)
(V < Q = CO2 content)• 2 types of shunting:
• Anatomical – blood moves from right to left heart w/o passing through lungs (congenital)
• Physiological – blood shunts past alveoli w/o picking up o2
![Page 19: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/19.jpg)
EtCO2 Monitoring
![Page 20: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/20.jpg)
EtCO2 Monitoring Ventilation/Perfusion Ratio (V/Q)
V/Q Mismatch Inadequate ventilation, perfusion or both 3 types
Physiological Shunt (V<Q) Blood passes alveoli Severe hypoxia w/ > 20% bypassed blood Pneumonia, atalectasis, tumor, mucous plug
Alveolar Dead Space (V>Q) Inadequate perfusion exists Pulmonary Embolus, Cardiogenic shock, mechanical ventilation w/
tidal volumes Silent Unit ( V & Q)
Both ventilation & perfusion are decreased Pneumothorax & ARDS
![Page 21: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/21.jpg)
EtCO2 Monitoring
![Page 22: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/22.jpg)
More Air Less BloodV > Q
Equal Air and BloodV = Q
More Blood Less Air
V < Q
EtCO2 Monitoring
![Page 23: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/23.jpg)
EtCO2 Monitoring Components of the normal
capnogram
![Page 24: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/24.jpg)
EtCO2 Monitoring
A - B =respiratory baseline CO2-free gas in the deadspace of the
airways
![Page 25: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/25.jpg)
EtCO2 Monitoring
B-C (expiratory upstroke) Alveolar air mixes with dead space air
![Page 26: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/26.jpg)
EtCO2 Monitoring
C-D (expiratory plateau) Exhalation of mostly alveolar gas
(should be straight) Point D = measurement point
(35-45mmHg)
![Page 27: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/27.jpg)
EtCO2 Monitoring
D-E =inspiration Inhalation of CO2-free gas
![Page 28: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/28.jpg)
EtCO2 Monitoring
![Page 29: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/29.jpg)
EtCO2 Monitoring
Changes in the capnogram or EtCO2 levels: Changes in ventilation Changes in metabolism Changes in circulation Equipment failure
![Page 30: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/30.jpg)
EtCO2 Monitoring
EtCO2 in specific settings
Non-Intubated patients Asthma & COPD CHF/Pulmonary Edema Pulmonary Embolus Head Injury Metabolic Illnesses
![Page 31: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/31.jpg)
EtCO2 Monitoring
Asthma and COPD Provides information on the ventilatory status of the
patient Combined with other assessments, can guide treatment
![Page 32: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/32.jpg)
EtCO2 Monitoring
Asthma and COPD (Cont’d) Shark fin waveform
![Page 33: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/33.jpg)
EtCO2 Monitoring
Asthma and COPD (Cont’d) Ventilatory assistance and/or intubation
may be considered with severe dyspneaand respiratory acidosis (EtCO2 >50mmHg)
18% of ventilated asthma patients suffer atension pneumothorax
New ACLS standards recommend ETI forasthma patients who deteriorate despiteaggressive treatment.
![Page 34: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/34.jpg)
EtCO2 Monitoring
Emphysema
![Page 35: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/35.jpg)
EtCO2 Monitoring
EtCO2 & CHF/Pulmonary Edema Wave forms will be normal (there is no
bronchospasm) Values may be increased (hypoventilation) or
decreased (hyperventilation)
![Page 36: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/36.jpg)
EtCO2 Monitoring
Pulmonary Embolus
“Normal” waveform but low numerical value(why?)
Look for other signs and symptoms
![Page 37: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/37.jpg)
EtCO2 Monitoring Pulmonary Embolus Note near “normal” waveform, but angled C-
D section (indicates alveolar dead space)
![Page 38: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/38.jpg)
EtCO2 Monitoring
EtC02 is very useful in monitoring intubated head- injured patients.
Hyperventilation = Hypocapnia = Cerebral Ischemia
Target EtC02 value of 35-38 mmHg
Head Injury
![Page 39: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/39.jpg)
EtCO2 Monitoring
Hypothermia
![Page 40: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/40.jpg)
EtCO2 Monitoring
Hyperventilation
![Page 41: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/41.jpg)
EtCO2 Monitoring
Hypoventilation
![Page 42: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/42.jpg)
![Page 43: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/43.jpg)
EtCO2 Monitoring
EtCO2 in the Intubated Patient
Identifies esophageal intubations & accidental extubations (head/neck motion can cause ETT movement of 5 cm)
Waveforms/numerical values are absentor greatly diminished
Do not rely on capnography alone to assureintubation!
![Page 44: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/44.jpg)
EtCO2 Monitoring
Tracheal –vs- Esophageal Intubation
![Page 45: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/45.jpg)
EtCO2 Monitoring
Esophageal Intubation
![Page 46: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/46.jpg)
EtCO2 Monitoring Esophageal Intubation w/carbonated
beverages
![Page 47: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/47.jpg)
EtCO2 Monitoring EtCO2 and cardiac output
Values <20mmHg = unsuccessful resuscitation Low (20-30mmHg) = good CPR or recovering heart
![Page 48: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/48.jpg)
EtCO2 Monitoring• EtCO2 and cardiac output• Sudden increase in value = ROSC
Cardiac arrest survivors had an average ETCO2 of 18mmHg, 20 minutes into an arrest while non survivors averaged 6. In another study, survivors averaged 19, and non-survivors 5.
![Page 49: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/49.jpg)
EtCO2 MonitoringEtCO2 and cardiac output
Successful defibrillation = pulses & EtcO2
![Page 50: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/50.jpg)
EtCO2 MonitoringEtCO2 and cardiac outputBecause ETCO2 measures cardiac output, rescuer fatigue during CPR will show up as decreasing ETCO2.
Change in rescuers – Note values w/ non-fatigued compressor
![Page 51: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/51.jpg)
EtCO2 Monitoring
Right Mainstem Bronchus Intubation
Numerical Values and Waveforms may/may notchange, but SAo2 will drop
![Page 52: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/52.jpg)
EtCO2 Monitoring
Kinked ET Tube
No alveolar plateau – very limited gas exchange
![Page 53: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/53.jpg)
EtCO2 Monitoring Spontaneous Respirations in the
paralyzed patient (Curare Cleft)
![Page 54: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/54.jpg)
Metabolic States Diabetes/Dehydration
EtCO2 tracks serum HCO3 & degree ofacidosis ( EtcO2 = metabolic
acidosis) Helps to distinguish DKA from NKHHC and
dehydration
EtCO2 Monitoring
![Page 55: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/55.jpg)
Metabolic StatesEtCO2 Monitoring
![Page 56: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/56.jpg)
EtCO2 monitoring will show you thecorrect respiratory rate, too.
Define “Synypnea”
EtCO2 Monitoring
![Page 57: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/57.jpg)
Synypnea is seen across the country and is defined as when emergency
department waiting room patients have the same respiratory rate.
EtCO2 Monitoring
![Page 58: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/58.jpg)
EtCO2 MonitoringTroubleshooting
Sudden increase in EtCO2
Malignant HyperthermiaVentilation of previously unventilated lung
Increase of blood pressureRelease of tourniquet
Bicarb causes a temporary <2 minute rise in ETCO2
![Page 59: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/59.jpg)
EtCO2 Monitoring
EtCO2 values 0
Extubation/Movement into hypopharynxVentilator disconnection or failure
EtCO2 defectETT kink
Troubleshooting
![Page 60: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/60.jpg)
EtCO2 Monitoring
Sudden decrease EtCO2 (not to 0)
Leak or obstruction in systemPartial disconnect
Partial airway obstruction (secretions)High-dose epi can cause a decrease (unk why)
Troubleshooting
![Page 61: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/61.jpg)
EtCO2 Monitoring
Change in Baseline
Calibration errorMechanical failureWater in system
Troubleshooting
![Page 62: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/62.jpg)
EtCO2 Monitoring
Continual, exponential decrease in EtCO2
Pulmonary EmbolismCardiac Arrest
Sudden hypotension/hypovolemiaSevere hyperventilation
Troubleshooting
![Page 63: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/63.jpg)
EtCO2 Monitoring
Gradual increase in EtCO2
Rising body temperatureHypoventilation
Partial airway obstruction (foreign body)Reactive airway disease
Troubleshooting
![Page 64: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/64.jpg)
![Page 65: End Tidal CO 2 (EtCO 2 ) Monitoring](https://reader035.vdocument.in/reader035/viewer/2022081507/568165c1550346895dd8ca33/html5/thumbnails/65.jpg)
Many special thanks to: JEMS Magazine (http://www.jems.com/) Peter Canning, EMT-P (http://emscapnography.blogspot.com/) Dr. Baruch Krauss ([email protected]) Bhavani-Shankar Kodali MD (http://www.capnography.com/) Bob Page, AAS, NREMT-P, CCEMT-P Steve Berry (https://www.iamnotanambulancedriver.com/mm5/merchant.mvc?) Dr. Reuben Strayer ([email protected]) UTSW/BIOTEL EMS SYSTEM (http://www.utsouthwestern.edu/) Oridion Medical Systems (http://www.oridion.com/global/english/home.html) Blogborgymi (http://blogborygmi.blogspot.com/) University of Adelaide, South Australia
(http://www.health.adelaide.edu.au/paed-anaes/talks/CO2/capnography.html)