Improved Analysis of Fatty Acids by Capillary GC columns based on Ionic Liquids
•F. Michel, L.M. Sidisky, M.D. Buchanan,
•G.A. Baney, Y. Ni,J.L. Desorcie, and K.K. Stenerson
2
Introduction
OH
O
CH3
OH
O
CH3
OH
O
CH3
☺☺☺☺
����
����
Estimation: replacing of trans-fats by plant oils may prevent up to 100.000 premature fatal incidents per year (http://www.hsph.harvard.edu/reviews/transfats.html)
3
Introduction
•Fatty acids are typically separated by GC as their methyl
esters (FAME)
•Capillary column choice for FAME analysis will depend
upon the information required from the analysis.
• Non-polar columns are routinely used to provide results on the amount of saturated versus unsaturated fatty acids.
• Polyethylene glycol based columns are used to provide carbon chain length & degree of saturation resolution.
• The analysis of cis and trans FAMEs, along with resolving their positional isomers, requires the use of highly polar biscyanopropyl polysiloxane columns [1].
[1] AOCS Method Ce 1h-05, “Determination of cis-, trans-, Saturated, Monounsaturated and Polyunsaturated Fatty Acids in Vegetable or Non-ruminant Animal Oils and Fats by Capillary GLC” AOCS Official Methods (2005) American Oil Chemists Society.
4
Fatty Acid / FAME ChemistryFAME Structures
Why use derivatization?
• The active carboxyl group has a much stronger interaction with the stationary phase than the remainder of the analyte
• Neutralizing the carboxyl group interactions allows slight differences in interaction strengths resulting from saturation level and double bond orientation to be observed
•Derivatization via an esterification reaction
sigma-aldrich.com/fame
Short chain fatty acids are often analyzed in their free form using GC.
Long chain fatty acids are often analyzed as FAMEs using GC.
5
Derivatization of Fatty Acids
www.sigmaaldrich.com/derivatization
www.sigmaaldrich.com/derivatization
6� Click on „Technical Literature“
Technical Literature on Derivatization
7
8
The limit of classic method for FAME analysisExtracted fat from margarine
20 30 40 50Time (min)
35.0 36.0 37.0 38.0
Time (min)
16:0 18:0
18:1
18:2
18:3
6t 9t10t
11t 12t
13t6c7c
9c
10c11c 12c13c
Peaks from 18:1 trans and cis isomers overlap
Overlap of trans/cis monounsaturated octadecenoic fatty acids on the GC chromatogram of a biscyanopropyl siloxane phase
SPE approach to overcome co-elution of cis- andtrans-FAMEs
•Use of silver-ion loaded SPE material for separation of cis-
/trans-FAMEs
• W. Christie: Ag-ion SPE for separation of FAMEs according to their degree of saturation in 1989
• Japanese research group (Food Analysis): Custom made Ag-ion SPE cartridge for separation of cis-/trans isomer up to trienes in 2005
How does it work?
•Silver ions immobilized onto
SCX SPE phase
•The FAMEs sample extract is
passed through the cartridge
•Ag ions interact with the
double-bonds of the FAMEs
•The more double bonds, the greater the retention
•Cis-isomers retain stronger than trans-isomers
How Is It Used?
•Fatty acids (FA) extracted from food sample
•FA converted to FAMEs using BF3
•FAMEs are extracted into hexane
•Hexane sample applied to Discovery Ag-ION SPE cartridge
•FAMEs separated using different mixtures of
hexane:acetone to extract from cartridge
•Fractions analyzed by GC-MS (SP-2560 capillary column).
Discovery Ag-ION Protocol
Fractionation of the standard FAME mixture
Standard sample, total FAMEs at 1 mg/ml
6 8 10 12 14 16 18 20
Time (min)
Standard Mixture
16:0 18:0
18:118:2 18:3
14:0
18:1 cSPE fraction 2
7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 15.0 16.0 17.0 18.0 19.0 20.0
Time (min)
Hexane: Acetone 90:10
18:1 t
6 8 10 12 14 16 18 20
Time (min)
SPE fraction1Hexane: Acetone 96:4
6 8 10 12 14 16 18 20
Time (min)
SPE fraction 3 Acetone
6 8 10 12 14 16 18 20
Time (min)
SPE fraction 4 Acetonitrile
GC Results of Cis/Trans Fractionation of Potato Chips
column: SP-2560, 75 m x 0.18 mm I.D., 0.14 µm (23348-U)
oven: 180 °°°°C, isothermalinj.: 220 °°°°C
det.: FID, 220 °°°°Ccarrier gas: hydrogen, 40 cm/sec. at 180 °°°°C
injection: 0.5 µL, 100:1 splitliner: 4 mm I.D., split, cup design
Discovery Ag-Ion SPE
•Technical Report T406062 (IRV)
• 54225-U
–750 mg/6 mL SPE Tube
• 54226-U
–750 mg/1 mL Rezorian™
Improvements in the GC Analysis of FAMEs
17
18
IntroductionStructure of Polysiloxane and PEG Polymer Phase
•Drawbacks
• Active hydroxyl (-OH) groups at the polymer termini allow a back-biting reaction
– Resulting in phase degradation
– Contributing to column bleed
• Chemistry modifications are limited to pendent group changes
• PEG/Wax: Limited to 280 °C maximum temperatureR = methyl, phenyl, fluoropropyl, and/or cyanopropyl (least polar to most polar).
x,y = percentage in the overall polymer composition.
HO Si O Si O H
R1 R3
R2 R4
x y
HO CH2 CH2 O CH2 CH2 O CH2 CH2 OH
n
19
Definition, Early Literature, and Use of Ionic Liquids
•Ionic liquids are a class of solvents with low melting points
that consist of organic cations associated with (inorganic or
organic) anions
•Ethyl ammonium nitrate (C2H5NH3+)(NO3
-), which has a
melting point of 12 oC, was described in 1914
• P. Walden, Bull. Acad. Imper. Sci. (St. Petersburg) 1800 (1914)
•Today, they are used as solvents, electrically conducting
fluids, and sealants
Ionic Liquids
20
Water
Ionic Liquid
CHCl3
Source: Prof. Jared Anderson,
University of Toledo, USA
21
Ionic Liquids in GC
•Several properties make ionic liquids desirable as GC
stationary phases
• Very low volatility…
– should result in columns with lower bleed and longer life
• Remain in the liquid state over a wide temperature range…
– should result in columns with extended temperature ranges
• No active hydroxyl (-OH) groups at their termini…
– should result in columns resistant to damage from moisture/oxygen
22
Ionic Liquids in GC
•Several properties make ionic liquids desirable as GC
stationary phases
• Are inherently highly polar…
– should result in columns with higher polarity that have lower elution temperatures plus an increased selectivity for polarizable analytes
• Have the broadest range of physical-chemical solvation interactions of any solvent…
– should result in columns with unique selectivity
• High viscosity…
– should be easy to coat columns
24
Example Structure of an Ionic Liquid Phase
• Numerous combinations of cations, anions and linkers are possible allowing for “tailored” selectivity or application
– Dicationic (shown) or polycationic
– Cations (imidazolium, phosphonium, pyrrolidinium, …), anions (NTf2-, triflat, tetrafluoroborate, hexafluorophosphate, …), and linkages (alkanes, polyethylenglycols, different lengths, …) can be changed
– Pendant groups can be added to cations and/or linkages
Phase used to make the SLB-IL100
1,9-di(3-vinylimidazolium) nonane bis(trifluoromethyl) sulfonyl imide
26
GC Column Polarity ScaleVisual Representation
Above the scale: positions/maximum temperatures of non-ionic liquid columns.
Below the scale: positions/maximum temperatures of Supelco ionic liquid columns.
Non-Polar
Intermediate Polar
Polar Highly Polar Extremely Polar
0 65 1003210
Ionic LiquidPhases
Non-IonicLiquid Phases
27
Selectivity: Test Mix Peak IDs and Conditions
•Peak IDs (listed in boiling point order)
• Toluene
• Ethylbenzene
• p-Xylene
• Isopropylbenzene (Cumene)
• Cyclohexanone
• 1,2,4-Trimethylbenzene
• 1,2,4,5-Tetramethylbenzene
• n-Tridecane (C13)
•Conditions
• columns: 30 m x 0.25 mm I.D., 0.20 µm
• oven: 110 °C• inj.: 250 °C• det.: FID, 250 °C• carrier gas: helium, 26 cm/sec
• injection: 1.0 µL, 100:1 split
• sample: each analyte at various concentrations in isooctane
28
Selectivity: Test Mix on Ionic Liquid Columns30 m Columns, 110 °°°°C Isothermal
2.0 3.0 4.0 5.0Time (min)
2.0 3.0 4.0 5.0Time (min)
2.0 3.0 4.0 5.0Time (min)
2.0 3.0 4.0 5.0Time (min)
2.0 3.0 4.0 5.0Time (min)
2.0 3.0 4.0 5.0Time (min)
8
8
8
8
8
8 5
5
5
5
5
5SLB-IL59
SLB-IL61
SLB-IL76
SLB-IL82
SLB-IL100
SLB-IL111
29
Rapeseed Oil FAME Peak IDs and Conditions
•Peak IDs and composition
• Myristic (C14:0) @ 1.0%
• Palmitic (C16:0) @ 4.0%
• Stearic (C18:0) @ 3.0%
• Oleic (C18:1n9c) @ 60.0%
• Linoleic (C18:2) @ 12.0%
• Linolenic (C18:3) @ 5.0%
• Arachidic (C20:0) @ 3.0%
• cis-11-Eicosenoic (C20:1) @ 1.0%
• Behenic (C22:0) @ 3.0%
• Erucic (C22:1) @ 5.0%
• Lignoceric (C24:0) @ 3.0%
•Conditions
• columns: 30 m x 0.25 mm x 0.20 µm
• oven: 180 °C isothermal
• inj.: 250 °C• det.: FID, 250 °C• carrier gas: helium, 25 cm/sec
• injection: 1.0 µL, split 100:1
• liner: 4 mm I.D., split type, cup design
• sample: Rapeseed oil FAME mix (O7756-1AMP) diluted to 10 mg/mL in methylene chloride
30
0 10 20
Time (min)
Rapeseed Oil FAMEs on Ionic Liquid Columns30 m Columns, 180 °°°°C Isothermal
0 10 20Time (min)
0 10 20Time (min)
0 10 20Time (min)
0 10 20
Time (min)
0 10 20Time (min)
6
6
6
6
6
7
7
7
7
7
8
8
8
8
8
7
8
6
SLB-IL59
SLB-IL61
SLB-IL76
SLB-IL82
SLB-IL100
SLB-IL111
6: C18:37: C20:08: C20:1
31
Partially Hydrogenated Vegetable Oil (PHVO) FAMEs
SP-2560 SLB-IL111
32
Partially Hydrogenated Vegetable Oil (PHVO) FAMEs
•When comparing FAME isomers with the same degree of
unsaturation, the SLB-IL111 provides increased retention of cis
isomers relative to trans isomers with the double bond at the
same location
•The SLB-IL111 was able to provide resolution of C18:1∆9c
from C18:1∆15t, a separation not possible with the SP-2560.
•The SLB-IL111 offered improved resolution of some isomers
that cannot be completely resolved with the SP-2560 either
• C18:1∆10t from C18:1∆11t
• The pair C18:1∆13t/C18:1∆14t from other isomers
33
CLA FAMEs in Ruminant Fats
•The SLB-IL111 is able to resolve C18:2∆9c,11t from
C18:2∆7t,9c – the two most abundant conjugated linoleic acid (CLA) isomers found in ruminant fats [5]
• No other column can resolve these
•SLB-IL111 as complementary column (to biscyanopropyl polysiloxane) for complete analysis of cis- and trans-fat
[5] P. Delmonte, A.-R.F. Kia, J.K.G. Kramer, M.M. Mossoba, L. Sidisky, J.I. Rader, J. Chrom. A, 1218 (2011) 545
34
To Learn More…
•Visit the Ionic Liquid GC column landing page (sigma-aldrich.com/il-gc) at the Sigma-Aldrich web site
•Download or request “Supelco Ionic Liquid GC Columns:
Applications”
•Download or request “Supelco Ionic Liquid GC Columns:
Bibliography”
sigma-aldrich.com/il-gc
35
Summary
• Ag-Ion SPE provides a separation of cis and trans FAMEs
– Allow for a complete separation of the critical mono-unsaturated C18 FAMEs
– Correct calculation of amount of cis/trans fatty acids
– Alternative approach to using SLB-IL111 and biscyanopropylpolysiloxane phases (e.g. SP-2560) as complementary column
• SLB-IL111 as complementary column to biscyanopropyl
polysiloxane phases allows the complete analysis of cis and
trans fatty acids
–Separation of several isomers that could not be resolved before
–Further publications on FAME analysis on Ionic Liquid GC [6]
[6] F. Destaillats, M. Guitard, C. Cruz-Hernandez , J. Chrom. A, 1218 (2011) 9384
Fatty Acid/FAME Application Guide (T408126, KUK)
•24-page brochure
•Includes columns, standards, and other
products for the:
• Analysis of free fatty acids
• Derivatization of fatty acids to FAMEs
• SPE fractionation of FAMEs
• Analysis of FAMEs by boiling point elution
• Analysis of FAMEs by degree of unsaturation
• Analysis of omega 3 and omega 6 FAMEs
• Analysis of cis/trans FAME isomers
•Also includes:
• A list of non-Supelco references
• A list of our product literature 36