Issues with low concentration gas detection in ambient air:
Sub-ppm and ppb detection are not just
business as usual.
![Page 2: Issues with low concentration gas detection in ambient air: Sub-ppm and ppb detection are not just business as usual](https://reader035.vdocument.in/reader035/viewer/2022081515/551a35c6550346cb358b51f3/html5/thumbnails/2.jpg)
Presented by:
ETA Associatesand
Interscan Corporation
![Page 3: Issues with low concentration gas detection in ambient air: Sub-ppm and ppb detection are not just business as usual](https://reader035.vdocument.in/reader035/viewer/2022081515/551a35c6550346cb358b51f3/html5/thumbnails/3.jpg)
KEY POINT
Once you enter the world of low concentration measurements, issues that previously were hidden are now magnified—and are thus all too apparent.
![Page 4: Issues with low concentration gas detection in ambient air: Sub-ppm and ppb detection are not just business as usual](https://reader035.vdocument.in/reader035/viewer/2022081515/551a35c6550346cb358b51f3/html5/thumbnails/4.jpg)
Compliance levels are always being lowered
So this is… Far from just an “academic” concern
![Page 5: Issues with low concentration gas detection in ambient air: Sub-ppm and ppb detection are not just business as usual](https://reader035.vdocument.in/reader035/viewer/2022081515/551a35c6550346cb358b51f3/html5/thumbnails/5.jpg)
Calibration issues
![Page 6: Issues with low concentration gas detection in ambient air: Sub-ppm and ppb detection are not just business as usual](https://reader035.vdocument.in/reader035/viewer/2022081515/551a35c6550346cb358b51f3/html5/thumbnails/6.jpg)
At best, it will be difficult!
Permeation tubes, usually not cylinder gas
![Page 7: Issues with low concentration gas detection in ambient air: Sub-ppm and ppb detection are not just business as usual](https://reader035.vdocument.in/reader035/viewer/2022081515/551a35c6550346cb358b51f3/html5/thumbnails/7.jpg)
A permeation tube is a sealed permeable membrane containing solid, liquid or liquefiable gases that permeate through the walls of the membrane at a constant rate.
![Page 8: Issues with low concentration gas detection in ambient air: Sub-ppm and ppb detection are not just business as usual](https://reader035.vdocument.in/reader035/viewer/2022081515/551a35c6550346cb358b51f3/html5/thumbnails/8.jpg)
The device is maintained at a constant temperature to establish constant vapor pressure inside the device.
The vapor escapes through the walls of the permeable membrane at a constant rate as long as the set point temperature is maintained.
![Page 9: Issues with low concentration gas detection in ambient air: Sub-ppm and ppb detection are not just business as usual](https://reader035.vdocument.in/reader035/viewer/2022081515/551a35c6550346cb358b51f3/html5/thumbnails/9.jpg)
A measured flow of an inert gas is passed through the permeation chamber resulting in known volumetric concentration in ppm/ppb.
By varying the dilution flow rate one is able to generate a wide range of concentrations using a single device.
![Page 10: Issues with low concentration gas detection in ambient air: Sub-ppm and ppb detection are not just business as usual](https://reader035.vdocument.in/reader035/viewer/2022081515/551a35c6550346cb358b51f3/html5/thumbnails/10.jpg)
![Page 11: Issues with low concentration gas detection in ambient air: Sub-ppm and ppb detection are not just business as usual](https://reader035.vdocument.in/reader035/viewer/2022081515/551a35c6550346cb358b51f3/html5/thumbnails/11.jpg)
There is an alternative
Electronic Calibration Service (ECS)
![Page 12: Issues with low concentration gas detection in ambient air: Sub-ppm and ppb detection are not just business as usual](https://reader035.vdocument.in/reader035/viewer/2022081515/551a35c6550346cb358b51f3/html5/thumbnails/12.jpg)
A certified spare sensor is kept on hand, to be put into the instrument, while the presently used sensor is sent back to the factory for re-certification.
The ECS certification details zero and span adjustments that are to be made on the instrument, to set it up with the specified newly certified sensor.
![Page 13: Issues with low concentration gas detection in ambient air: Sub-ppm and ppb detection are not just business as usual](https://reader035.vdocument.in/reader035/viewer/2022081515/551a35c6550346cb358b51f3/html5/thumbnails/13.jpg)
For some gases…
Even permeation tubes are not available
e.g. H2O2
![Page 14: Issues with low concentration gas detection in ambient air: Sub-ppm and ppb detection are not just business as usual](https://reader035.vdocument.in/reader035/viewer/2022081515/551a35c6550346cb358b51f3/html5/thumbnails/14.jpg)
In such cases….
Specialized lab techniques will be required
With ECS or a surrogate to the customer
![Page 15: Issues with low concentration gas detection in ambient air: Sub-ppm and ppb detection are not just business as usual](https://reader035.vdocument.in/reader035/viewer/2022081515/551a35c6550346cb358b51f3/html5/thumbnails/15.jpg)
Zero gas issues
![Page 16: Issues with low concentration gas detection in ambient air: Sub-ppm and ppb detection are not just business as usual](https://reader035.vdocument.in/reader035/viewer/2022081515/551a35c6550346cb358b51f3/html5/thumbnails/16.jpg)
Zero gas must
Contain ≤ 1% of expected value of target gas
Exclude potential interferents, as necessary
![Page 17: Issues with low concentration gas detection in ambient air: Sub-ppm and ppb detection are not just business as usual](https://reader035.vdocument.in/reader035/viewer/2022081515/551a35c6550346cb358b51f3/html5/thumbnails/17.jpg)
BUT
This may not always be possible
![Page 18: Issues with low concentration gas detection in ambient air: Sub-ppm and ppb detection are not just business as usual](https://reader035.vdocument.in/reader035/viewer/2022081515/551a35c6550346cb358b51f3/html5/thumbnails/18.jpg)
Consider best availablezero gas Total hydrocarbons < 0.01 ppmCO < 0.01 ppmNOx < 0.001 ppm
SO2 < 0.001 ppm
N2O < 0.001 ppm
![Page 19: Issues with low concentration gas detection in ambient air: Sub-ppm and ppb detection are not just business as usual](https://reader035.vdocument.in/reader035/viewer/2022081515/551a35c6550346cb358b51f3/html5/thumbnails/19.jpg)
Thermo Model 48i offers 0-1 ppm measuring range
for COBut, a reading of 0.2 ppm would be subject to a 5% error--just from the zero gas
And what about the calibration error?mn
![Page 20: Issues with low concentration gas detection in ambient air: Sub-ppm and ppb detection are not just business as usual](https://reader035.vdocument.in/reader035/viewer/2022081515/551a35c6550346cb358b51f3/html5/thumbnails/20.jpg)
With gases such as SO2
The situation is even worse, for very low range measurements
![Page 21: Issues with low concentration gas detection in ambient air: Sub-ppm and ppb detection are not just business as usual](https://reader035.vdocument.in/reader035/viewer/2022081515/551a35c6550346cb358b51f3/html5/thumbnails/21.jpg)
Check out our Knowledge Base
article
“In search of zero”
![Page 22: Issues with low concentration gas detection in ambient air: Sub-ppm and ppb detection are not just business as usual](https://reader035.vdocument.in/reader035/viewer/2022081515/551a35c6550346cb358b51f3/html5/thumbnails/22.jpg)
Interference issues
![Page 23: Issues with low concentration gas detection in ambient air: Sub-ppm and ppb detection are not just business as usual](https://reader035.vdocument.in/reader035/viewer/2022081515/551a35c6550346cb358b51f3/html5/thumbnails/23.jpg)
Interferences that pose negligible problems at higher
concentration measurements…
Can become significant at lower range measurements
![Page 24: Issues with low concentration gas detection in ambient air: Sub-ppm and ppb detection are not just business as usual](https://reader035.vdocument.in/reader035/viewer/2022081515/551a35c6550346cb358b51f3/html5/thumbnails/24.jpg)
Consider this example…
The effect of NO2 on SO2 measurements
OSHA PEL for SO2 is 5 ppm
![Page 25: Issues with low concentration gas detection in ambient air: Sub-ppm and ppb detection are not just business as usual](https://reader035.vdocument.in/reader035/viewer/2022081515/551a35c6550346cb358b51f3/html5/thumbnails/25.jpg)
In our sensor…
24 ppm NO2 will produce a 1 ppm negative effect on SO2
And, that’s not so bad
![Page 26: Issues with low concentration gas detection in ambient air: Sub-ppm and ppb detection are not just business as usual](https://reader035.vdocument.in/reader035/viewer/2022081515/551a35c6550346cb358b51f3/html5/thumbnails/26.jpg)
But, if you wanted to read 0.1 ppm
The effect has been magnified by a factor of ten
And, can now become a problem
![Page 27: Issues with low concentration gas detection in ambient air: Sub-ppm and ppb detection are not just business as usual](https://reader035.vdocument.in/reader035/viewer/2022081515/551a35c6550346cb358b51f3/html5/thumbnails/27.jpg)
•Calibration issues1•Zero gas issues2•Interference issues3
![Page 28: Issues with low concentration gas detection in ambient air: Sub-ppm and ppb detection are not just business as usual](https://reader035.vdocument.in/reader035/viewer/2022081515/551a35c6550346cb358b51f3/html5/thumbnails/28.jpg)
Joel Myerson
Special thanks to…
Caroline Ronten