Download - KR2000 SERIES GRAPHIC RECORDER - CHINO
1 PSE-353I
GRAPHIC RECORDERKR2000 SERIES
KR2000 series are network-compatible paperless recorders with high performance and high operating function employed high visibility 5.7" TFT color LCD display. High speed of sampling rate 100ms and high accuracy of ±0.1% were realized, and measured data is stored into internal memory and maximum 8GB compact flash card (CF card). As it can be monitored by a web browser display on several computers on intranet or internet, FTP transfer of data file and E-mail notification are also available.
●Employing clear 5.7"TFT color LCD display. Large-sized high visibility display with various display functions.
Real time/historical trend screen, bar-graph screen, data screenare selectable for various applications.●Large capacity of data memory and various recording method. Compact flash card (CF card) slot is equipped as standard exter-
nal memory.Large capacity storage of maximum 8GB is available.
. Various data storing methods are selectable such as schedule pro-gramming by time of day and time of date, recording start-up by ex-ternal signal, and event and data logging of before and after trig-ger points for alarm.●Multi points recording with high speed/accuracy. High-speed recording of approximately 100ms and high accuracy
of ±0.1% were realized. Stable measuring and recording are possi-ble with high speed.
. High withstand voltage of 1000V AC between input channels. (Ex-cept resistance thermometer input)●Easy operating and programming without manual. Easy operating by dedicated keys for each function●USB port prepared in front compartment. USB port is prepared for connecting maximum 8GB USB memory
and PC. . Readout of data and files are possible by connecting the panel
mounted recorder.●LAN network capability. Various networked environment such as remote monitoring by
browser, FTP server and E-mail notification are applied as Ether-net is equipped as standard.●Safety system and reliability. No battery backup needed for external memory for recorded data
storage.●Analyzing/data acquisition application software (option). It is easy to replay and edit the recorded data file. Replay display
has various mode of vertical/horizontal trend, circular trend, and also has wave-analyzing and marking by using the cursor.●Custom graphic screen for per each applications (NEW). By using optional custom graphic screen function, it can display
the graphic screen which the user created by PC software KR Screen Designer (optional). Create letters, rectangle, oval, line, etc by drawing tool and allocate KR measuring data while making the background by JPEG or other images. By lower communication, controller SV, MV, PID can also be changed. Register up to 5 screens and its screens are switchable.
■ FEATURES
■ MODELS KR21□□□□□□-□□□
Measuring points/sampling rate*1
60 : 6 points/100ms 20 : 12 points/100ms 61 : 6 points/1s 21 : 12 points/1sUSB port M : USB memory stick (type A) - : PC (type B)Communications interface (option) N : None R : High-order (RS232C/RS485) Q : High-order (RS232C/RS485) + Low-order (RS485)Digital input/ alarm output (option) 0 : None 1 : Mechanical relay output -12 points
(a contact) 2 : Mechanical relay output - 6 points
(c contact) 7 : Digital input - 8 points + MOS relay output 8 points
Others (option)(Blank) : None-1NN : Custom graphic screen
*1 1 to 4 channels input (4 points) when setting faster than 500ms sampling rate with model of 1sec sampling rate.
Carrying handle and feet (option) A : None T : With carrying handle and feet *2
*2 Non-conformance to CE
2
KR2000 SERIES
■ SCREENS●Real-time trend screen
Displays data (measured and virtual)of selected group.Vertical trend and horizontal trendselectable.
●Bar-graph screenDisplays data (measured and virtual)of selected group.Combination display with real-timetrend is available.
●Data screenDisplays data (measured and virtual)of selected group.Simultaneous display of alarm status.
●Dual trend screen2 split display for real-time trend andhistorical trend.Scroll available for historical trend.
●Information screen
●HOME setting screen ●Channel setting screen ●Schedule setting screen
●Graphic screenEnable to create custom display for each user*.
*Graphic screen feature is provided optionally. BMP image has to be prepared by customer.
3
■ INPUT SPECIFICATIONSMeasuring points: 6 points, 12 pointsInput types: Universal
DC voltage --- ±13.8mV, ±27.6mV, ±69.0mV±200mV, ±500mV, ±2V±5V*, ±10V*, ±20V*, ±50V* (*with built-in voltage divider)
DC current --- With external shunt resistor (sold separately)Thermocouple --- B, R, S, K, E, J, T, N, PtRh40-PtRh20,
W-WRe26, WRe5-WRe26, PlatinelⅡ, NiMo-Ni,CR-AuFe, U, L
Resistance thermometer --- Pt100, JPt100, Pt-Co, Pt50Accuracy ratings: Refer to the table of measuring range and accuracy
ratingsReference junction compensation accuracy:
K, E, J, T, N, PlatinelⅡ--- ±0.5°C or lessR, S, W-WRe26, WRe5-WRe26, NiMo-Ni, CR-AuFe,U, L --- ±1.0°C or less
Sampling rate: 100ms --- Approximately 100ms for all points1s --- Approximately 300ms for all points**100ms/4 points when setting faster than 0.5s (0.1s to 0.5s) sampling rate with model of KR2161/KR2121
Burnout: Disconnection of input signal is detected on thermocouple and resistance thermometer input.UP/DOWN/DISABLE is selectable.
Scaling: Range/scale is selectable.Digital filter: Programming FIR filter for each point (common to
all points)Allowable signal source resistance:
Thermocouple input (burnout disable)/DC voltage input (±2V or less) ---1kΩ or lessDC voltage input (±5V or more) ---100Ω or lessResistance thermometer --- Per wire 10Ω or less (same resistance for 3 wires)
Input resistance: DC voltage, thermocouple input --- Approximately 1 MΩ Maximum input voltage:
DC voltage input (±2V or less)/thermocouple input (burnout disable) --- ±10VDCDC voltage input (±5V to ±50V) --- ±60VDC
Dielectric strength between channels:1000V AC or more between each channel(High strength semiconductor relay used)(B terminal of resistance thermometer is shorted inside between channels)
Common mode rejection ratio:120dB or more (50 or 60Hz)
Series mode rejection ratio:50dB or more (50 or 60Hz)
■ DISPLAY SPECIFICATIONSDisplay: 5.7"TFT color LCDDisplay types: Measured data display (Trend screen, Data screen, Bar-graph
screen)Historical trend display (simultaneous display with Real-time trend is available) Information display (alarm display, marker list, file list)Setting screen (alarm, computation, memory, system, maintenance, communication, etc.)
Trend screen: 12 colors selectableDisplay screen--- 5 screens (5 groups)Display points --- Maximum 44 points/screenTime axis direction --- Vertical or horizontalLine width --- 1/3/5 dot selectableScale display --- 4 scalesTag/data display --- Show/hide selectableMarker display
Data screen: Display screen --- 5 screens (5 groups)Display points --- Maximum 44 points/screenDisplay contents --- Measured value, channel/tag, unit, alarm status
Bargraph screen: 12 colors selectableDisplay screen --- 5 screens (5 groups)Display points --- Maximum 44 points/screenDisplay direction --- Vertical or horizontalScale display --- 1 scale
Information display: Alarm display (alarm activation/released history display)Marker listFile list (group data file list display)
LCD back light: Auto/manual OFF functionUnit information (Model, Serial no., option, etc.)Brightness --- 4 levels adjustment
*The LCD display may contain some pixels that always or never illuminate, and the brightness of some areas of the display may appear uneven. There are typical LCD performance characteristics and do not constitute malfunctions.
■ COMMUNICATION FUNCTIONS
■ CONNECTIVITY
●NetworkCommunication type:
Ethernet (10BASE-T/100BASE-TX)FTP server: Data file can be read from the network computerFTP client: Transfer a data file to a network serverSNTP client: The time can be synchronized to the time of SNTP serverWeb server: Conformed to HTTP1.0 --- Displays the alarm, information of
maintenance by browser software (InternetExplorer5.0 or later, NetScape6.0 or later, Opera7 or later)* User's ID and password registration available
E-Mail: E-Mail notification at specified time for alarm activationReport data at specified time is selectable from all registered data.Corresponds to SSL and TLS.Notification address --- Maximum 8 contacts
MODBUS TCP: Read and write the data of compatibles units.●USB CommunicationsUSB: Communication type --- USB1.1
Transfer systems --- Bulk transfer, control transferFile transfer by connecting as removable disk drive
■ RECORDING SPECIFICATIONSMemory for history: 132MBAdditional memory: CF card (Up to 8GB)
128MB standard attached, Apacer Technology made recommendedUSB memory stick (Up to 8GB)Not all USB memory stick allowable
Recording cycle: 100, 200, 500ms1, 2, 3, 5, 10, 15, 20, 30s1, 2, 3, 5, 10, 15, 20, 30, 60min
Logging data: Measured data --- File name (group name), time of day,month and year of recording start, tag, measured data,alarm status/types and marker textSetting parameterOperation result data
Storing types: Binary/CSVStoring methods: Manual start/stop (dedicated key operation)
Schedule (designation for time of day and date)Trigger signal (alarm event, digital input)*Pre-trigger is selectableMeasuring numbers of pre-trigger --- Maximum 950 data
Recording group: Recording cycle 500ms or faster --- up to 3 groups of 12 points/group can be programmedRecording cycle 1s or slower --- up to 5 groups of 44 points/group can be programmed(Up to total of 100 points)
■ COMPUTATION SPECIFICATIONSComputation points: Maximum 44 pointsComputation types: Arithmetic operations --- Addition, subtraction,
multiplication, division, remainder, exponential
Comparison operations --- Equality, inequality, great, less, equality /great, equality / less
Logical operations --- AND, OR, XOR, NOTGeneral functions --- Round-up, round-down, absolute
value, square root, exponent of e, natural logarithm, common logarithm
Integration operations --- Analog integration, digital integration
Channel data operations --- Measured data computation, calculated data computation
Others --- Dew point, relative humidity, F-valueRemaining amount of CF card,linearization table, data communications input
■ ALARM SPECIFICATIONSSetups: Up to 4 alarms can be programmed per channelAlarm types: Upper limit, lower limit, differential upper limit, differential lower
limit (deadband is selectable), abnormal dataDelay function: Setup range of alarm delay --- 1 to 3600 seconds Alarm settings: AND/OR selectableAlarm outputs: Refer to option specification
Application
Ethernet
File readout
Data storing in USB memory stickUSB port for PC
USB port for USB memory stick
RS485(Low-order communication) *Option
When 6 channels recorded in sampling mode (real data)Recording cycle
0.1 sec1sec
60 sec
128MB6.32 days63.2 days
10 yrs
256MB12.6 days126 days
21 yrs
512MB25.3 days253 days
42 yrs
1GB50.6 days
1.4 yrs83 yrs
2GB101 days2.8 yrs166 yrs
When 12 channels recorded in sampling mode (real data).Recording cycle
0.1 sec1sec
60 sec
128MB3.16 days31.6 days
5.2 yrs
256MB6.32 days63.2 days
10 yrs
512MB12.6 days126 days
21 yrs
1GB25.3 days253 days
42 yrs
2GB50.6 days1.4 yrs83 yrs
TYPE A TYPE B
KEPLC DB
KR3000
PLC PLC
Web browser FTP server
Ethernet *RS422A/RS485 cannot be parallel used.
MODBUS TCP compatible unit
4
■ PROGRAMMING/OPERATION SPECIFICATIONSOperation key: HOME, MENU, DISP, MARKER, SCROLL, CURSOR,
START, STOP, DIRECTION keys, ENTER, ESCHOME settings: Simple recording settings --- Common setting to all channels
Parameter programming for all channels together, recording cycle, selection settings
MENU settings: Input/computation programming --- Input parameter, computation parameterDISP settings --- Data channel parameter, group parameter, common parameter (combination display, trend vertical/horizontal)Alarm settings File settings (5 individual files) --- Storing method settingsMarker text settingsSystem settings --- Communication, clock, maintenance,key lock, password, screen, etc.
DISP operations: Operating screen selection --- Trend, data, bar-graph,historical trend, alarm display, maker listDisplay selection on each screen --- Group 1 to 5 selectable
■ STANDARDSCE : EMC directive --- EN61326-1: 2006 Class A
EN61000-3-2EN61000-3-3
Low voltage directive --- EN61010-1(2001)Protection: Conformed to IEC60529 IP65 (recorder front bezel)
■ OPTION SPECIFICATIONS
■ GENERAL SPECIFICATIONSRated power voltage: 100 to 240V AC (universal power supply) 50/60HzMaximum power consumption:
50VAReference operating condition:
Ambient temperature --- 21 to 25℃, Ambient humidity --- 45 to 65%RHPower voltage --- 100V AC±1.0%Power frequency --- 50/60Hz±0.5%Attitude --- Left/right 0°, forward/backward 0°Warm-up time --- Longer than 30 minutes
Normal operating condition:Ambient temperature --- 0 to 50℃Ambient humidity --- 20 to 80%RHPower voltage --- 90 to 264V ACPower frequency --- 50/60Hz±2%Attitude --- left/right 0°, forward tilting 0°,Backward tilting 0° to 20°
Transport condition (at the packed condition on shipment from our factory):Ambient temperature --- -20 to 60℃Ambient humidity --- 5 to 90%RH (No dew condensation)Vibration --- 10 to 60Hz 0.5G (4.9m/S2) or lessImpact --- 40G (392m/ S2) or less
Storage condition: Ambient temperature --- -20 to 60℃Ambient humidity --- 5 to 90%RH (No dew condensation)
Power failure protection:Setups and data are backed up by flash memoryClock --- Lithium battery backs up RAM(Minimum 5 years)
Insulation resistance: Secondary terminals and protective conductor terminals --- 20MΩ or more at 500V DCPrimary terminals and protective conductor terminals --- 20MΩ or more at 500V DCPrimary and secondary terminals --- 20MΩ or more at 500V DCPrimary terminals: power terminals (L,N), alarm output terminalsSecondary terminals: measuring input terminals, digital input terminals, communications terminals
Dielectric strength: Secondary terminals and protective conductor terminals --- 1 minute at 500V ACPrimary terminals and protective conductor terminals ---1 minute at 1500V ACPrimary and secondary terminals --- 1 minute at 2300V ACPrimary terminals: power terminals (L,N), alarm output terminalsSecondary terminals: measuring input terminals, digital input terminals, communications terminals
Case assembly material:Front bezel --- ABS resin Case --- Steel
Color: Front bezel --- Black (equivalent to Munsell N3.0)Case --- Painting color, gray (equivalent to Munsell N7.0)
Weight: 2.2kgMounting: Panel mountingTerminal screws: Power terminals/protective conductor
terminals/communicationsterminals --- M4.0Measuring input terminals/alarm output terminals/digital input terminals --- M3.5
■ OPERATION KEYS
KR2000 SERIES
START SCROLL MARKER HOME ESC
STOP CURSOR DISP MENU ENTER
DIRECTION keys
Options Specifications
Mechanical relayalarm output
Communicationsinterface
Mechanical relay contact output for abnormal input and alarm activationOutput: 12 points (a contact), 6 points (c contact)Contact ratings: Mechanical relay
--- 100V AC 0.5A, 240V AC 0.2A, 30V DC 0.3A
Communications interface for high-order unitsRS232C/RS485 (MODBUS) switchableEthernet is standard equipped
Digital inputs
ON/OFF signal
MOS relay contact output for abnormal input and alarm activationOutput: 8 pointsContact ratings: MOS relay --- 240V (DC, AC) 50mA
MOS relay alarmoutput
Handle and feet, white front bezel, point indication cardOthers
Pulse input
Remote contact
High-ordercommunications
Low-ordercommunications
ON/OFF input recording
Maximum 10Hz pulse inputUsed for flow, operating time and frequencyInput system:Photocoupler isolation (Common
use for contact and pulse input)Built-in isolated power supply (approx. 5V)Input type: Non-power contact, open collector
(TTL or transistor)
The following operations are available by contactinput 8 points and common signal 4 points(Selectable by parameter).・Data memory triggering
Start data recording by conductive signal fromOFF to ONData recording while conductive signal is ON・Marker display
Registered makers display by conductive signalfrom OFF to ON・Integration operations
Reset data for integration operations(all channels simultaneously)
Communications interface for low-order unitsInput data storing of units connected to low-orderRS485 (MODBUS)Recording points: 6 points recorder --- 30 points
12 points recorder --- 24 pointsSampling rate: 1s per connected unitModels: KE, SE3000, KR2000, KR3000, LE5000,AL3000, AH3000, AL4000, AH4000LT230, 830, 350, 450, 470,DB1000, 2000, KP1000/KP2000/DP-G (dataacquisition only),JU, JW. Transfer input data of KR2000 to PLC. The input data can be written on PLC only. Data writing points: 44 pointsConnectable PLC: Mitsubishi Electric
CorporationMELSEC AnA, QnA,QnAS, FX seriesOMRON CorporationSYSMAC series
Note) Separate purchase of protocol converter SC8-10 (optional) is required for connection to OMRON PLC.
■ KR SCREEN DESIGNER (sold separately) (NEW)
Model:KS3200-000OS:Windows Vista/7/8Others:Your OS recommended requirements or better
Custom GraphicScreen
By KR Screen Designer (optional), create graphic screen byPC and display to KR screen via CF card. KR measuring valuecan be located to the screen.
5
■ MEASURING RANGES ■ APPLICATION SOFTWARE ZAILA(sold separately)
Input type Accuracy ratingsMeasuring range
±0.1%±1digit
±0.1%±1digit*-200 to 0℃:±0.2%±1digit
±0.15%±1digit*-200 to 0℃:±0.3%±1digit
±0.1%±1digit*0 to 400℃:±0.2%±1digit
±0.1%±1digit*0 to 400℃:Out ofaccuracy ratings*400 to 800℃:0.15%±1digit
±0.15%±1digit*0 to 100℃:±4%±1digit*100 to 400℃:±0.5%±1digit
-13.80-27.60-69.00
totototototo
totototo
tototo
tototo
tototo
tototo
toto
toto
to
200.0mV500.0mV
2.000V
13.80mV27.60mV69.00mV
-5.00010.00V20.00V50.00V
5.000V
-200.0-200.0-200
300.0℃600.0℃
1370℃
-200.0-200.0-200
-2.000
-200.0-500.0
-10.00-20.00-50.00
200.0℃350.0℃900℃
-200.0-200.0-200
250.0℃500.0℃
1200℃
-200.0-200.0-200
400.0℃750.0℃
1300℃
to
to
to
2315℃
-200.0-200.0
250.0℃400.0℃
00
toto
00
1300℃1760℃
0 1820℃
1200℃1760℃
DC voltage
(with built-involtage divider)
K
E
J
T
R
S
B
N
W-WRe26
±0.2%±1digit2315℃WRe5-WRe26
0
0
0
±0.2%±1digit*0 to 300℃:±1.5%±1digit*300 to 800℃:±0.8%±1digit
±0.1%±1digit*-140.0 to 150.0℃700 to 850℃:±0.15%±1digit
1888℃
tototo
tototo
tototo
tototo
tototo
tototo
to
-50.0-50.0-50
±0.2%±1digit
±0.2%±1digit*0 to 20K:±0.5%±1digit*20 to 50K:±0.3%±1digit
290.0℃600.0℃
1310℃
PtRh40-PtRh20
NiMo-Ni
CR-AuFe
0.00.00
±0.15%±1digit350.0℃650.0℃
1395℃
0.0 280.0K
PlatinelⅡ
-200.0-200.0-200.0
±0.15%±1digit*-200 to 0℃:±0.3%±1digit
250.0℃500.0℃600.0℃
U
-200.0-200.0-200
±0.1%±1digit*-200 to 0℃:±0.2%±1digit
250.0℃500.0℃900℃
-140.0-200.0-200.0
±0.1%±1digit*-140.0 to 150.0℃:±0.15%±1digit
150.0℃300.0℃649.0℃
to-200.0 649.0℃
-140.0-200.0-200.0
150.0℃300.0℃850.0℃
L
Pt100
JPt100
±0.1%±1digitPt50
to4.0±0.15%±1digit*4 to 50K:±0.3%±1digit
374.0KPt-Co
RTD
T/C
The software is applied for replay display/wave editing operation of recorded data in KR2000 series. It has replay display of vertical/horizontal trend and circular trend function, and also analyzing function such as magnify/reduce/partially magnify of graphs and message insert.
■Trend displaySelectable from trend display window (vertical flow, horizontal flow) andcircular trend display window.■Continuous replay display window
Trend is scrolled continuously (automatically).Scroll changes by speed and renewal data no.■Data list display window
Displays registered data as list display.■Bar-graph
Displays by bar. Message can be inserted into bar-graph.■Data between markers
Displays date/time, time difference between 2 data, data difference,maximum, minimum, average, standard deviation and median amongall data.■Alarm display
Points for alarm activation at each level are displayed on a trend graph.■Settings
Cursor, trend line, scale axis, time axis, title input on the graph, graphassistant and magnify/reduce/rotation of graphs■Data conversion
Exporting to Excel, and converting to CSV file or TEXT file areavailable.
●Display examples
●Main functions
Trend display window (vertical flow) Trend display window (horizontal flow)
Trend display window (circular trend) Bar-graph
Note: The accuracy ratings are converted into the measuring range under reference operating condition. Thermocouple input does not contain reference junction compensation accuracy.K,E,J,T,R,S,B,N:IEC584,JIS C1602-1995W-WRe26,WRe5-WRe26,PtRh40-PtRh20,PlatinelⅡ,NiMo-Ni,Cr-AuFe:ASTM Vol14.03U(Cu-CuNi),L(Fe-CuNi):DIN43710Pt100:IEC751(1995),JIS C1604-1997JPt100:JIS C1606-1989
■ ENVIRONMENT
CPU
OS
Others
Disk drive
Language
Your OS recommended CPU and/or upper grade
Windows VISTA / 7 / 8*Internet Explorer 6.0 or later
Your OS recommended memory or larger
CD-ROM drive: 1 drive or moreHard disk drive: Disk space of 1 drive or more for 100MB or more
Japanese, English, Chinese (simplified and traditional characters), Korean
■ TERMINAL ARRANGEMENT
■ DIMENSIONS ● PANEL CUTOUT
Communications terminals (option)
Ethernet connectorHigh-order: RS232C/RS485 (switchable)Low-order: RS485
●Alarm mechanical relay alarm output 12 points (option)
[Contact types]
a contact: Normally open
c contact: Switch contact
N.O
N.O
N.C
COM
COM
N.O.terminals
COM terminals(Mechanical relay a contact output)
Alarm output terminals (option)
Measuring input terminals
Thermocouple DC voltage (+)/Resistance thermometer (A) terminalsThermocouple DC voltage (-)/Resistance thermometer (B) terminals
Resistance thermometer (B) terminals
Protective conductor terminalsPower terminals
Communications terminals (option)
Ethernet connectorHigh-order: RS232C/RS485 (switchable)Low-order: RS485
●Digital input 8 points + MOS relay alarm output 8points (option)
N.O.terminals
Contact input terminals
COM terminals(MOS relay a contact output)
Alarm output terminals (option)
Protective conductor terminalsPower terminalsContact input COM
Channels 1 to 6 Channels 7 to 12
Communications terminals (option)
Ethernet connectorHigh-order: RS232C/RS485 (switchable)Low-order: RS485
●Mechanical relay alarm output 6 points (option)
N.C(M3.5) terminals
COM(M3.5) terminalsAlarm output terminals (option)
Measuring input terminals
Thermocouple DC voltage (+)/Resistance thermometer (A) terminalsThermocouple DC voltage (-)/Resistance thermometer (B) terminals
Resistance thermometer (B) terminals
Measuring input terminals
Thermocouple DC voltage (+)/Resistance thermometer (A) terminalsThermocouple DC voltage (-)/Resistance thermometer (B) terminals
Protective conductor terminalsPower terminals
Channels 1 to 6 Channels 7 to 12
Channels 1 to 6 Channels 7 to 12
N.O.(M3.5) terminals(Mechanical relay a contact output)
Resistance thermometer (B) terminals
24.5 203.6
144 29.8 200.5
63 6318
144
137
161
200
200
138+1 0
138+
1
0
Unit: mm
Specifications subject to change without notice. Printed in Japan (I) 2020. 3
32-8 KUMANO-CHO,ITABASHI-KU,TOKYO 173-8632Telephone : +81-3-3956-2171Facsimile : +81-3-3956-0915E-mail : [email protected] : www.chino.co.jp/