7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
1/51
LOCKING COMPRESSION PLATE IN
FRACTURES OF OSTEOPOROTIC BONES
BY
DR.YOGESH SHIVMURTI KHANDALKAR
M.B.B.S.,D.ORTHO,M.Ch(ORTHO)(USAIM)
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
2/51
INTRODUCTION
LIFE IS MOVEMENT, MOVEMENT IS LIFE
Life is defined here is said to be present only if active movement is there. To
put this principle into practice is what every orthopedic surgeon concerned
about.
To achieve full movement at a particular joint after a fracture is most
important is early mobilization. This has to be pain free mobilization to have
full support from patients side. Mobilization restores rapid return of blood
flow to bone and soft tissues and also prevents fracture disease.
Plate is an implant which is fastened to the bone for the purpose of fixation.
It can be protection or neutralization or tension band plate. The shape of
plate is an adaptation of plate to the local anatomy and does not denote
any function.
THE AIM OF ANY SURGICAL FRACTURE TREATMENT IS TO RECONSTRUCT
THE ANATOMY AND RESTORE THE FUNCTION.
1
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
3/51
OSTEOPOROSIS AND FRACTURES
Osteoporotic fracture treatment remains a challenge for the surgeon, often
with unpredictable outcomes. In this series of cases highlights current
aspects of these fractures and focuses on advances in implant design and
surgical techniques. Osteoporosis is a skeletal disorder characterized by
compromised bone strength, predisposing to an increased risk of fracture in
the clinical setting, it can be defined as a reduction in bone mass of > 2.5 SD
below the mean for the young adult.1-4
This is confirmed in this study by
DEXA SCAN.
More than 40% of women and 14% of men over the age of 50 years will
experience fracture due to osteoporosis.4
Globally approximately 200
million people are at risk of sustaining an osteoporotic fracture each year.5
It is expected that osteoporosis will become epidemic to come as a result of
the increasing number of elderly people.
6
By 2012, 25% of the Europeanpopulation will be over the age of 65 years and by 2020, 52 million people
will be over this in United States.
The most frequent osteoporotic fractures seen in men arise in the spine,
and wrist.7
In women the common fracture sites include the wrist, spine,
humors femur and ribs. In the rising incidence of proximal fracture of the
femur represents the most important socio-economic impact of
osteoporosis.8
In 2000, there were approximately 424000 hip fractures
worldwide and 1098000 in women. Based upon altering demographics and
the increase in life expectancy is estimated that by 2025 there will be an
increase of 89% in men, resulting in 800000 hip fractures per year, and in
women the numbers will rise by 69% to 1.8 million.8
2
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
4/51
Osteoporosis is characterized by thinner cross- linking connections within
trabecular bone.9
Endosteal diaphyseal restoration and medullary
expansion are common in both men and women changes in diameter of the
inner and outer cortices affect the bending and torsional characteristics of
the entire bone and predispose to low-energy fractures, which often have a
complex pattern at tissue level, there is a decrease in the cancellous bone
mineral density.10,19
There is also a decrease in the density of cortical bone,
because of an increase in porosity, which can affect the holding of screws.11
The healing of a fracture in osteoporotic bone passes through the normal
stages and concerned with union of the fracture although the healing
process is prolonged.12
There is evidence of animal models. Namkung-
Matthail et al.13
, showed a 40% reduction of callus in the cross-sectional
area and a 23% reduction in bone mineral density in the healing femur of an
osteoporotic bone (postoophorectomy and low calcium diet). Similar results
demonstrating that healing took longer in rats, and both stiffness and
strength remain below the values of controls, where found Meyer et al.14
The major technical problem facing the surgeon is difficult secure fixation,
less corticocancelleous screw purchase leading to decrease pull-out
strength of implants. BMD correlates linearly with holding power of the
screws
15-16
. In osteoporotic bone microfracture and loosening of implant,resorption of the bone is due to reduced strength tolerance.
17-18The
operative treatment of metaphyseal fractures in the elderly is associated
with an increased rate of complications; and implant failure occur in 2% to
10% of fractures, malunion in 4% to 40% and re-operation to 23%.19-24
3
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
5/51
The high rate of complications has encouraged extensive research into the
development of which can improve the bone-implant interface by
preventing high stress and distributing the load transmitted to bone in a
load-sharing, rather than load-bearing way.
Techniques of the internal fixation which aim to provide absolute stability
with lag screws are usually inappropriate in Buttress-plate fixation of
metaphyseal fractures is effective in osteoporotic bone as it avoids strains
at a single screw while the implant provides a large contact area at the
bone-implant surface again reducing strain.22
Fixed-angle devices, such as the angled blade plate, are very useful as they
resist angular deformation and torsion, and the strain is reduced because
the blade has a large surface and initial success with fixed- angle implants
such as the blade plates23-24
has led t the development of screws which are
rigidly fixed to the plate. This was first achieved by adding Schuhli nuts (S
Paoli, Pennsylvania) to standard plates and, more recently, with threaded
holes incorporated into the plate, the so-called LOCKING COMPRESSION
PLATED (LCP)25-26
. Plated with locking- head screws also produced a fixed-
angle device and have similar mechanical properties. The holding power of
implant can be increased further by having locked screws at multiple fixed
angles.27
Thus LCP has achieved the purpose of angle stability and bone to
plate interface and secured fixation in fractures of osteoporotic bones. Thus
LCP is a plate of new era. An early example of this tried was the pint contact
fixator.28 4
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
6/51
The main advantage of the locking-plate device is the mechanical couple
between the screw and the plate (fixed-angle device) so that even if the
screw-bone interface fails, the screw plate interface remains intact.
Complete failure of fixation is still possible, and is seen in very severe
osteoporosis, but all screws on one side of the fracture fixation must fail
simultaneously. The implants such as the locking compression plate and the
less invasive stabilization system significant advantages in osteoporotic
bone.22
Comparison between conventional and locking plated have been
conducted in distal29
and diaphyseal30
fractures of the humans and were to
be better suited to provide stable and reliable fixation. A review of the
available literature in the field of plate osteosynthesis31
came to the same
conclusion. Similar developments to include principle of angular stability in
intramedullary nails are now underway.32-33
The treatment of unstable
fractures of the proximal humerus using the se nails has demonstrated thata stable osteosyntheis is achievable in very old patients.
34
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
7/51
LOCKING COMPRESSION PLATE:35-36
It is equivalent to external fixator, which is shifted close to the bone. The
Combination hole present in the LCP was designed by Mr. Michael Wagner
& Mr. Robert Frigg in order to incorporate a threaded hole alongside a
dynamic compression unit. A LCP-combination hole has two parts:
1. DCP unit- for cortical and cancellous screws2. Conical thread unit- for L.H.S. (Locking Head Screw)5
THUS LCP IS A NEW ERA IN THE TREATMENT OF FRACTURES OF
OSTEOPOROTIC BONES.
It is an advantageous in patients with severe osteoporosis.
For treating fractures with open reduction and internal fixation. ASIF
(Association for Study of Problems in Internal Fixation) have propounded a
1. Accurate and anatomical reduction.2. Rigid internal fixation3. Atraumatic technique on bone soft tissues. Working hypothesiscomprising of four principles.
4. Early pain free active mobilization especially during first tenpostoperative days.
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
8/51
AIMS AND OBJESCTIVES:-
1)To study efficacy of L.C.P. as a treatment modality in case of fractures ofosteoporotic bones.
2)To evaluate the progression of fracture healing in osteoporotic bones.
3)To study effects of fixation fractures by locking compression plate .
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
9/51
MATERIALS AND METHODS
During the period from Aug 2007 to Dec 2010, 35 cases of osteoporoticfractures confirmed by DEXA scan were selected and prospective study has
done.
Type of study Prospective
Sample Size 35 Patients included in the study
Study Duration 40 Months
Selection of Patients:
Any patient with # of osteoporotic bones was included in the study.
Compound Grade 1; 1/ Gustillo Andersons # type and closed # were also
treated. According to the requirement of the patient the surgical procedure
was done as:
- Open reduction & Int. Fix.OR
- Minimally invasive plate osteosynthesis (MIPO)
Individual case study
X-rays classified all #s according to type & BMD (Bone Mineral Density)
Confirmed by DEXA machine.
7
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
10/51
DEXA MACHINE
Each patient underwent a detailed history taking, clinical examination
and assessment of # site. (As per attached proforma)
Fractures were classified whether intraarticular or extraarticular.
8
MANAGEMENT:-
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
11/51
a)Management in causality:-
On admission, the detailed history was asked regarding mode of injury, time
since injury, any primary treatment taken, and any significant past history.
Patient was thoroughly assessed clinically. Patients vital parameters were
checked, associated-head, chest, abdominal injuries were looked for.
On local examination skin condition, presence of Haemarthrosis, instability
at joint and any distal neurovascular compromise was looked. Any other
associated limb or bone fracture was checked.
Depending on the vital parameters and general condition, an intravenous
access was sought for and intravenous fluids were given whenever
necessary. All other injuries and associated fractures were immobilized and
taken appropriate care.
If associated contused, lacerated wounds were present, they were given a
thorough wash and cleaned with savlon, hydrogen peroxide, normal saline,
betadine and dressing was applied.Bolus dose of intravenous broad-spectrum antibiotic was given when
associated wounds. Intravenous/ Intramuscular NSAIDs were given after
systemic injuries were ruled out.
Anteroposterior, lateral, and oblique, views radiographs of affected part
were taken.
9
b) Management thereafter:-
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
12/51
As required, the patients with Haemarthrosis, associated dislocation,
compartment syndrome, were subsequently taken to operation theatre.
In patients with associated unclean wounds, at other sites, surgical
debridement under appropriate anesthesia was done. The wound was
cleaned with savlon, hydrogen peroxide, normal saline, betadine and
dressed. Broad-spectrum antibiotic covering gram positive, gram negative
and anaerobes were started. Tetanus toxoid and antitetanus serum were
given.
Investigation:
The patients who were admitted and were subsequently operated had
following investigation done
a) Routine Investigations:Hemoglobin %, Bleeding time, Clotting time,Peripheral blood smear, Urine- routine and microscopy.
b) For operative point of view:Blood sugar level, blood urea level, serum electrolytes, electrocardiogramand chest skiagram were done.
Radiographs of affected part in anteroposterior and lateral view formed a
major part in deciding the modality of treatment for a particular fracture.
The following questions arose:
Q. What was the type and degree of comminution of fracture?
Q. Was the depression anterior of posterior/
Q. Will bone grafting be required?
Q. Were the collateral and cruciate ligaments avulsed in case of knee injury
and ankle injury?
Answer the above questions and you will be on the right path
10
INSTRUMENTS AND IMPLANT:-
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
13/51
- Full Swiss AO- synthes instrumentation and implant set- Screwdrivers having a torque limiting devices that avoids stripping of thescrew heads after they are locked in the plate.
- Plates also permit standard self tapping titanium screws which can beused in normal compression mode with all the advantages of compression
technique.
- Minimal invasive technique required image intesifier and radiolucentoperation table. All the operations were performed in operating room of
high standards.
INSTRUMENTS:-
A.O.Standard instruments were used as follows:
1. Guide sleeve
2. A.O. drill bit
3. A.O. depth gauge
4. A.O. drill guid
5. A.O. torque limiting screw driver
6. A.O. quick coupling system
7. A.O. plate bender
8. A.O. standard screw driver
11
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
14/51
STANDARD A.O. (LARGE FRAGMENT) SET
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
15/51
STANDARD A.O. (SMALL FRAGMENT) INSTRUMENT
Diagnosis of osteoporosis:
Diagnosis of osterporosis was confirmed by DEXA machine scan. Patient
selected for study was screened for osteoporosis.
12
OPERATIVE PROCEDURE:-
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
16/51
Preoperative planning:
It was done for every patient managed by this technique. Preoperative
roentgenogram of affected part was taken in anteroposterior, lateral and if
required oblique views.
These roentgenograms were assessed to determine type of fracture
amount of displacement or depression present, probable length of screws
required for particular case; particular direction of the screw insertion
required to achieve compression at fracture site and to avoid neurovascular
damage.
Preoperative planning in each case was helpful in minimizing intra operative
decision-making, shortens the operative time and hence improves the
results. Preoperative PHYSICIAN check up was done in every case.
Anesthesia:-
The operative procedure was performed under spinal anesthesia, I.V.regional anesthesia or general anesthesia depending up on patients
physical fitness and requirement under pneumatic tourniquet control.
Positioning the patient:
Patient was in supine position. The image intensifier was placed on the
opposite side of the table and positioned perpendicular to the table.
Patients Preparation:-
Preoperative scrubbing with savlon and 7.5% povidine iodine scrub for at
least 10 min was done and mopped off with spirit and painting was done
with 10% povidines iodine and spirit.
13
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
17/51
Appropriate surgical approach for affected fracture part was made
depending upon upper or lower extremity bone under all aseptic
precautions. Whenever necessary MIPPO technique was used, otherwise
fracture site was exposed.
As tissue respect is important in every surgery tissue handling was kept to
minimum. periosteum was stripped only from the site where the LCP was
to be applied. Open reduction was achieved using bone holding forceps and
LCP was fixed on the reduced fragment with plate holding device.
Whenever necessary few k wires were used to maintain the reduction
Anatomical reduction again confirmed under Image-Intensifier- A.O.
Principle. Drilling done using standard A.O. drill for LCP. Now 3.5 or 4.5 mm
depending upon type of the plate screws are inserted in combihole of LCP.
There was no need of tapping A.O. titanium screws as they are self tapping
screws. Closure was done using vicryl 1-0 and Ethilon 2-0. Negative suction
drain was kept depending upon need and site of fracture.Dressing done on 5
th, 9
th& 11
thpost operative day.
Alternate sutures were removed on 9th
and all sutures removed on 11th
post
operative day. Every patient was taught post operative exercises.
For lower limb walking without weight bearing with the aid of crutches was
started a few days afterwards. Exact timing of beginning of partial and full
weight bearing was judged individually but generally this was not before 6-8
weeks. Follow up with detailed objective assessment was carried out.
Radiographs were taken at 6, 12, 14, 16, 24, 48, 72 weeks. Early and late
complications were recorded and adequately managed.
Pain assessment of each patient done with VISUAL ANAGLOUGUE SCALE &
grading done according as excellent, good, fair, poor.
14
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
18/51
Range of movement compared with previous normal movement with each
patient and graded as excellent, good, fair, and poor.
VISUAL ANALOUGE SCALE
15
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
19/51
OBSERVATIONS AND RESULTS
Total number of cases were - 35
Table No.1
Age in years No. of cases Percentage
60-70 25 71.42
70-80 08 22.86
80-90 02 5.72
Most of cases were old age group of 60 to 70 years.
Table No. 2
Sex- Male 09 25.72
Sex- Female 26 74.28
Female predominated as commonly involved in home activity and had fall.
16
Table No. 3
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
20/51
Bone involved No. of cases Percentage
Femur 06 17.14
Humerus 12 34.28
Radius-Ulna 14 40.00
Tibia- Fibula 03 08.58
Total 35 100
Table No. 4
Type of fracture No. of cases Percentage
Transverse 20 57.15
Oblique 11 31..42
Comminuted 04 11.43
Total 35 100
Transverse fractures were most suitable for compression. For comminuted
fractures with large fragments fixation done with interfragmentary screws.
17
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
21/51
Table no.5
Type of fractures No. of cases Percentage
Simple fracture 31 88.58
Compound fractures 04 11.42
Gustilo- Anderson
type I
03
Gustilo- Anderson
type II
01
Type I & Type II compound fractures were primarily treated with
debridement under anesthesia. These patients were temporarily
immobilized with slab support or external fixator and antibiotics were
administered. These were taken for definitive surgery after wound healing.
Table No. 6
Mechanism of injury
Type of injury No. of cases Percentage
Vehicular accidents 04 11.43
Fall 31 88.57
From stair-case 14
Fall in bathroom 13
Fall on road 04
Patients predominantly had fall.
18
Table No. 7
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
22/51
Fractures in upper
limb
23 65.72%
Fractures in lower
limb
12 34.28%
Upper limb fractures predominated compared to lower limb.
Table no. 8
Radiological features considered
1. Anatomical reduction2. Compression3. Position of plate4. Placement of screws5. Interfragmentary compression6.
Primary union
7. Presence/ Absence of infective changes8. Disuse osteoporosis9. Implant complications
Table No. 9
No. of cases Percentage
Screw loosening 00 00
No screw loosening 35 100
19
Table no. 10
Minimum cortices secured in plating different bones
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
23/51
Humerus 6
Radius 5
Ulna 5
Femur 7
Tibia 6
Fibula 5
Table No. 11
No. of cases Percentage
Postoperative
immobilization
07 20
No immobilization 28 80
Patients having comminution and compound injuries were immobilized
temporarily.
20
Table No. 12
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
24/51
Weight bearing in lower limb fracture
Early - 10th
post operative day 6 50
Delayed - 10th
day of 5 weeks 3 25
Late - More than 6 weeks 3 25
Table No. 13
Post operative stiffness in adjacent joints
Restriction upto 10 deg. Mild 04
Restriction upto 30 deg. Moderate 02
Restriction more than 30
deg.
Severe 02
Table No. 14
Stiffness in UL 05
Stiffness in LL 03
Postoperative stiffness: It was observed in (10%) cases. Two cases had
ipsilateral fractures of femur, tibia and patella.
21
Table No. 15
UNION
Metaphysis
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
25/51
No. of weeks 0-2 2-6 6-
12
12-14 14-16 16-18
No. of cases 00 00 00 19 00 00
Percentage 54.28
Diaphysis
No. of weeks 0-2 2-6 6-
12
12-14 14-16 16-18
No. of cases 00 00 00 00 16 00
Percentage 45.72
Healing in 12-14 weeks for metaphyseal and 14-16 for diaphyseal fractures
is considered as excellent.
22
Table no. 16
Infection No. of cases Percentage
Superficial 03 8.58
Deep 01 2.86
Table No. 17
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
26/51
Complications
1) Postoperative stiffness 08
2) Infection Superficial 03
Deep 01
3) Delayed union
4) Malunion
5) Nonunion
6) Implant failure
7) DNVC
8) Re-fracture
23
Table no. 18
BMD Scores
T-SCORE No. of patients
-2.5 to -3.5 21
-3.5 to -4.0 14
Table No. 19
Range of movement compared to previous level
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
27/51
Range of movement Excellent
(75%)
Good
(50-
75%)
Fair
(25-
50%)
Poor
(
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
28/51
25
DISCUSSION
The objective of fracture management is the restoration of optimal pre-
injury status by safes and most reliable method. Early mobilization and
return to pre-injury environment also provides a psychological stimulus to
healing.
Every fracture leads to a complex tissue injury involving the bone and
surrounding soft tissue envelope. Immediate after fracture and during the
repair phase, the following are evident:
1. Local circulatory disturbances,2. Local inflammation,3. Pain,4. Reflex immobilization/ adjustment of joints and surrounding muscles.
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
29/51
These factors lead to development of fracture disease which is described
by Lucas-Championniere in 1907.
Fracture disease is evident clinically as chronic edema, soft tissue injury and
patchy osteoporosis. Edema, per se, induces intermuscular fibrosis and
atrophy. These fibrotic process cause muscle to develop unphysiological
adhesions to bone and fascia and therefore joint stiffness and contractures.
In case of osteoporotic bones internal fixation of fractured bone is
challenge with regular implant due to weal holding capacity. There is
loosening of screws of back firing of screws. There may be loosening of
plate also. But LCP has revolutionized the treatment by giving angular
stability in internal fixation of fractures and locking head screws.
26
LCP while fixed to bone acts as monoblock unit so chances of implant
loosening are very less as with previous plates.Blood supply of bone beneath the LCP is well maintained as it acts like
internal ex-fixator and plate is well away from bone surface which is very
important in case of osteoporotic bones. So it does not hamper local blood
healing of fixed fractures of osteoporotic bones which is confirmed in our
series of 30 cases so also helps in early pain free mobilization of operated
fractures of osteoporotic bones.
LCP is made up of titanium which is a tuff material and resistant to
corrosive effect. This reduces chances of implant failure and infection due
to stresses across the plate and even after fall after fixation in case old
people. So cost of resurgery and stress of resurgery is avoided in case old
osteoporotic people. Titanium is also resistant to infection and this
important fact proved in our series with cases of two superficial and one
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
30/51
deep infection. Thus LCP is ideal implant in compound fractures of
osteoporotic bones. So also added advantage of this material is it does not
interfere with investigation like M.R.I. study in case of elderly patients with
fractures of osteoporotic bones.
LCP is bio-friendly and body-friendly material is it can be kept in body
lifelong. So cost of resurgery, stress of resurgery and stress of implant
removal in osteoporotic bones is avoided. Thus it has helped to avoid
socioeconomic burden in our community.
Significant factors which affect union are initial displacement,
communication, associated soft tissue wounds, infection and distraction.
Lambotte (1907), Dannis (1949), Kuntscher (1935), Charnley (1948) and
others have already demonstrated importance of compression and its
tolerance by bone. 27
Muller (1963), Willenger, Bagby and others have advocated use of rigid
internal fixation of diaphyseal fracture.
The significant factors which affect union of fracture are displacement,
comminution, associated soft tissue injuries, infection, and distraction of
fracture fragments and adequacy of blood supply.
By treating these patients with LCP it was possible to mobilize then early
and reduce the changes of fracture diseases. By early rehabilitation, good
physiotherapy these patients could return to their work early and could
ones again start their earning. This not only helps patients and their family
financially but psychologically also. The patient is much happier and
satisfied. In the present series treatment of fractures in these patients with
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
31/51
LCP was of immense benefit as it facilitated early pain free post operative
mobilization of the patients, without a cumbersome plaster. This prevented
joint stiffness and muscle wasting to a large extent and helped nursing care
and prevention of bedsores and plaster sores.
At operation, when dealing with the fracture, the most important aspects
taken into consideration were perfect anatomical reduction.
This was relatively easy in short oblique and transverse fractures. For
comminuted fractures, perfect anatomical reduction was difficult, but could
be achieved with the help of interfragmentary screws in cases with large
butterfly fragments.
28
The technique of applying a LCP is quite an exacting technique, Requiring
considerable technical skill and a well equipped instrument trolley.
Operating time required was shortened as more experience was gained.
The results of our study were quite comparable with the standard series,
where the average period required for union of the fracture was 12 weeks.
In case of primary union, where no callus is seen it is difficult to assess
union. Minimal callus was seen in few cases mainly due to periosteal
strapping, and did not take part in consolidation of the fracture.
Follow up X-rays were also looked up carefully for presence of infective
changes, osteoporosis and implant complications.
The signs of nonunion, such as increasing fracture gap on roentgenograms
and pain on walking were not encountered in the present series. Also no
case of stress fracture of the plate was seen.
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
32/51
In present series 35 cases were included.
Most of patients in our series were females from 60-70 years age group;
from low socio economic and middle class family; some of them were dailywedge earners. Period of immobilization and hospitalization mattered a lot
for them from economic point of view as well.
Most commonly involved bones were humerus and radius-ulna.
Mostly transverse and oblique fractures predominated in our series.
Compression can be easily applied to these fractures after achieving good
anatomical reduction. 29
As against that is comminuted fractures compression is difficult to achieve
unless comminution is restricted to only one or two butterfly fragments. In
case of bad comminution we used MIPO technique which shows good
results.
88.58% were simple fractures and 11.42% were compound fractures. These
were Gustilo-Anderson type I and II compound fractures which were
initially treated with debridement, suturing, temporary immobilization and
antibiotics. After complete wound healing, on an average after 7-10 days
these were taken for LCP fixation. Most of cases were fall i.e. 31 of which 14
had fall in bathroom, 13 over stair-case and 04 on road. There were 04
cases of RTA who had fracture due to minor trauma because of
osteoporosis as later on confirmed by DEXA scan.
Because of post operative pain and inflammation response to
physiotherapy was poor in the first 3 days. But later on relative pain free
mobilization could be started in these patients. 07 patients were required
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
33/51
immobilization for 4-6 weeks due to comminution of fracture and
associated ipsilateral injuries.
65.72% patients had fractures in upper limbs and 34.28% in lower limbs.After rigid fixation of fractures with LCP active mobilization is very easily
isolated and single limb fractures.
30
Important fractures considered:
Plate was positioned on tension side of fractures so as to convert tension
forces into compression forces whenever possible.
Interfragmentary compression could be achieved with separate screws as
well as passing through the plate with over drilling of proximal cortex.
Radiological union is said to exist when the fracture line was totally
obliterated. In case of primary union where no callus was seen it is difficult
to access the union. Minimal callus was present in many cases which was
mainly due to periosteal stripping and did not take part in consolidation of
the fracture.
Follow up X-rays were looked at carefully for presence of infective changes,
osteoporosis and implant complications.
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
34/51
Radiological union was seen after a period of average 12 weeks for
metaphyseal fractures and 14 weeks for diaphyseal fractures. Total
obliteration of fracture line was taken as radiological union.
Superficial infection was observed in 03 patients. This was restricted to
suture tracts only and could be taken care of easily with proper wound care
and use of antibiotics. 01 case of deep infection was seen in patients who
had compound injury graded as Gustilo-Anderson type II. They were initially
treated by through debridement followed by antibiotics and internal
fixation after wound healing.
31
Post operative stiffness was observed in 08 cases which were put on
physiotherapy and subsequent gain of good range of movement achieved.
The complications can be avoided in case of fractures of osteoporotic bones
by observing following points learned over period of two years-
1) Proper selection of patients2) Thorough post operative preparation3) Aseptic and atraumatic operative technique4) Selecting proper size screws and plates
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
35/51
32
SUMMARY & CONCLUSIONS
Our series included 35 cases of fractures of osteoporotic bones.
Osteoporosis was confirmed by DEXA machine.
Most of the patients were female.
Common age group of patient was 60-80 years.
All patients were without plaster after surgery which was a problem with
previous implants but not in case with LCP.
All patients had pain free, early and effective mobilization after surgery.
LCP provides optimum hold and stability in case of osteoporotic fractures as
LCP gives angular stability with locking head screws. It acts as monoblock
construct. Locking minimizes the compressive forces exerted by the plate
on the bone.
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
36/51
As LCP has combination hole, depending upon fracture situation, it can be
used in either a conventional technique (compression principle), bridging
technique (internal fixator principle), or a combination technique
(compression and bridging principles).
Thus LCP has improved patient compliance in case of fractures of
osteoporotic . bones.
33
In the modern day Orthopedic practice with widened indication of internal
fixation LCP is an excellent device in osteoporotic bone. Moreover it
decreases morbidity in old osteoporotic patients.
As function of affected limb is very much important in old people, LCP
offers early mobility and union by fracture site compression compared to
other devices. It does not require post operative external support hence
increases patients compliance.
It has following advantages:
1) Less surgical exposure2) Rigid internal fixation3) Interfragmentary compression4) Early post operative mobilization5) Less chances of infection6) Less chances of stiffness7) Early and good post operative pain relief8) Less chances of delayed union, nonunion, and malunion9) Decrease hospital stay
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
37/51
10) Early occupational rehabilitation
11) Implant can be kept for life long unless complication
12) Body friendly material
THUS LOCKING COMPRESSION PLATE IN FRACTURES OF OSTEOPOROTIC
BONES IS THE BEST MODALITY OF TREATMENT AT PRESENT TIME. 34
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
38/51
Recent Advances
1)Failure of proximal Femoral LCP with open reduction and internal
fixation,
2)Good use in periprosthetic fracture Vancouver B1 femoral
fracture- JBJS SEPT.2007 89(9) 1964-9
3)Elbow arthrodesis using two LCP Sept.2007 vol.8 issue 3 page 141
to 146-Techniques in elbow & shoulder surgery
4)Free vascularised fibular bone grafting combined with LCP for
massive bone defects in lower limbJournal of international
orthopaedics 26 jan.2012
5)FUTURE DEVELOPEMENT FEB.2012= POLYAXIAL LOCKING
COMPRESSION PLATES TO FIX SCREWS IN VARIOUS DIRECTIONS.
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
39/51
BIBLIOGRAPHY
1) Kanis JA, Oden A, Johnell O, et at. The burden of osteoporotic fractures:a method for setting intervention thresholds. Osteoporos Int 2001; 12:417 -
2 LC.
2) Mori S, Harruf R, Ambrosius W, Burr DB. Trabecular bone volume andmicrodam-a~ accumulation in the femoral heads of women with and
without femoral neck fractures. 1997; 21: 521-6.3) Bassey EJ, Rothwell MC, Littlewood JJ, Pye OW. Pre- andpostmenopausal women different bone mineral density responses to the
same high- impact exercise. J Bone Mi 1998; 13: 1805-
4) Eastell R, Lambert H. Strategies for skeletal health in the elderly. ProcMutr Sac 200L 80.
5) Cummings SR, Kelsey JL, Nevitt MC, ODowd KJ. Epidemiology ofosteoporosis an osteoporotic fractures. Epidemiol Rev 1985;7-178.
6) Smith DP, Enderson BL, Maull KI. Trauma in the elderly: determinanatsof outcome. ; Med J 1990; 83:171-7.
7) Seeman E, Bianchi G, Adami S, et at. Osteoporosis in men: consensus inpremature tissue Int 2004; 75:120-2.
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
40/51
8) Gullberg B, Johnell O, Kanis JA. World-wide projections for hip fracture.Osteoporos 1997; 7:407-13.
9) Goldstein SA, Goulet Z, McCubbrey D. measurement and significance ofthree-dimensional architecture to the mechanical intergrity of trabecular
bone. Calcify Tissue Int 1995; 53 (Suppl1): 127-33.
35
10) Prevrhal S, Njeh CF, Jergas M, Genant HK. Imaging and bone
densitometry of steoporotic In: An YH, ed. Internal fixation in osteoporotic
bone. New York: Thieme Inc., 2002:51.
11) seebeck J, Goldhahn J, Morlock MM, Schneider E. Mechanical behavior
of screws normal and osteoporotic bone. Osteoporos Int 2005; 16
(Suppl2):101-1
12) Lopez MJ, Edwards III RB, Markel MD. Healing of normal and
osteoporotic bone. If ed. Orthopaedic issues in osteoporosis. Boca Raton:
CRC Press, 2003:55-70.
13) Namkung-Matthail H, Appleyard R, Jansen J, et at. Osteoporosis
influences the ease of fracture healing in a rat osteoporotic model Bone
2001; 28:80-6.
14) Meyer RA Jr, Tsahakis PJ, Martin OF, et al. Age and ovarietomy impair
both the normalization of mechanical properties and the accretin of mineral
by the fracture cal rats. J Orthop Res 2001; 19:428-35.
15) Perren SM. Backgrounds of the technology of internal fixators. Injury
200;34 (Suppl 2).
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
41/51
16) Chapman JR, Harrington RM, Lee JM, et at. Factors affecting the pullout
strength of cancellous bone screws. J Biomech Eng 1996; 118:391-8.
17) Koval KJ, Meek R, Schemitsch E, et al. An AOA critical issue: geriatrictrauma: youn J Bone Joint Surgery (Am) 2003; 85-A: 1380-8.
18) Liu JG, Zu XX. Stress shielding and fracture healing. Zhonghua Yi Xue Za
Zhi 1994; 5, 51 9.
36
19) Stover M. Distal femoral fracture: current treatment, results and
problems. Injury 2001 (Suppl 2): 3-13.
20) Syed AA, Agarwal M, Giannoudis PV, Matthews SJ, Smith RM. Distal
femoral fracture long-term outcome following stabilization with L1SS. Injury
2004; 35:599-607.
21) Bergman RJ, Gazit O, Kahn AJ, et at. Age-related changes in osteogenic
stem cells J Bone Miner Res 1996; 11:568-77.
22) DIppolito G, Schiller PC, Ricordi C, Roos BA, Howard GA. Age-related
osteogenic of mescenchymal stromal stem cells from human vertebral bone
marrow. J Bone Miner 1999; 14: 1115-22.
23) Instrum K, Fennell C, Shrive N, et al. Semitubular blade plate fixation in
proximal humerus fractures: a biomechanical study in a cadaveric model. J
Shoulder Elbow Surg 1998; 7:1.
24) Jupiter JB, Mullaji AB. Blade plate fixation of proximal humeral non-
unions. Injury 1994; 25:30 1-3.
25) Kolodziej P, Lee FS, Patel A, et al. Biomechanical evaluation of the
schuhli nut. Clin 1998; 347:79-85.
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
42/51
26) Simon JA, Dennis MG, Kummer FJ, Koval KJ. Schuhli augmentation of
plate and screw fixation for humeral shaft fractures: a laboratory study. J
Orlhop Trauma 1999; 13:1961.
27) Schutz M, Sudkamp NP. Revolution inplate osteosynthesis: new fixator
systems. J (Sci 2003; 8:252-8. 37
28) Haas N, Hauke C, Schutz M, Kaab M, Perren SM. Treatment of
diaphyseal fractures forearm using the point Contact Fixator (PC-Fix):
results of 387 fractures of a prospec multcentric study (PC-Fix II). Injury
2001; 32 (Suppl 2): 51-62.
29) Korner J, Diederichs G, Arzforf M, et al. A biomechanical evaluation of
methods of c humerus fracture fixation using locking compression plates
versus conventional reconstruction plates. J Orlhop Trauma 2004; 18:286-
93
30) Wenzl ME, Porte T, Fuchs S, Faschingbauer M, Jurgens C. Delayed and
non-union humeral diaphysis: compression plate or internal plate fixator?
Injury 2004; 35:55-60
31) Egol KA, Kubiack EN, Fulkerson E, Kummer FJ, Koval KJ. Biomechanics of
locked plates and screws. J Orlhop Trauma 2004; 18:488-93.
32) Plecko M, Kraus A. Internal fixation of proximal humerus fractures using
tile locking humerus plate. Oper Orlhop Traumato 2005; 17:25-50.
33) Stannard JP, Harris HW, McGuin G Jr, Volgas DA, Alonso JE.
Intramedullary nailing humeral shaft fractures with a locking flexible nail. J
Bone Joint Surg (Am) 2003; 85-A: 10.
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
43/51
34) Mathews J, Lobenhoffer P. Results of the provision of unstable proximal
humeral fractures geriatric patients with a new angle stabilizing antegrade
nail system. Unfallchirung 2004: 1 07:372-80.
35) AO/ASIF Principles of fracture management, Ruedi T. et. Al, Thieme
Verlag, 2000.
38
36) Schatzker j, changes in AO/ASIF principles and methods
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
44/51
39
Proforma
Name:
Age: Sex:
Case No.:
Date of admission:
Date of discharge:
Chief complaints:
History:
Mode of injury
Whether case of polytrauma
Hypovolemia
History of head injury
On exam:
Inspection Compound/Closed
Deformity/Scar/Sinuses/Swelling
Palpation
Abnormal bony mobility
DNVC
Rom at adjacent joint
Pain at # Site
Osteoporosis + / -
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
45/51
Post operative management:
Period for which pt- immobility 40
Period for which pt is on antibiotic
Suture removal
Mobilization of adjacent joint
Weight bearing
Post operative observations:
Wound healing.
Edema of part.
Changes of sudeck osteodystrophy.
Joint and complications.
41
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
46/51
Shoulder Stiffness
Painful arc syndrome
Active abductionElbow Stiffness
Myositis Ossificans
Range of flexion
Range of extension
Wrist Stiffness
Deformity
Dorsiflexion
Palmer flexion
Sudeck dystrophyHip Stiffness
Active flexion
Active abduction
Active adduction
Active internal rotation
Active external rotation
Knee Swelling
Stiffness
Flexion deformity
Active extension
Effusion
Crepitus
Patellar tap
Ankle Stiffness
Sudeck dystrophy
Deformity
Active plantar flexionActive dorsiflexion
Deformity- Equinus
42
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
47/51
Pre operative x -ray
Intra-operative MIPO Technique
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
48/51
post operative X-RAY
Pre-operative x-ray fracture proximal end ulna
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
49/51
Intra - operative
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
50/51
Post operative x-ray
-
7/31/2019 Locking Compression Plate in Fractures of Osteoporotic Bones 2012 -Article
51/51
Post operative ROM