![Page 1: Nitrogen Dioxide (NO PDA: A Global 2 Sterilization and ... · PDF file2) Sterilization and Decontamination Systems . November 4, ... • Surface sterilization process ... • Common](https://reader030.vdocument.in/reader030/viewer/2022021504/5aaaf6307f8b9a693f8b47da/html5/thumbnails/1.jpg)
PDA: A Global Association
Nitrogen Dioxide (NO2) Sterilization and Decontamination Systems
November 4, 2014
PDA Europe Parenterals 2014
![Page 2: Nitrogen Dioxide (NO PDA: A Global 2 Sterilization and ... · PDF file2) Sterilization and Decontamination Systems . November 4, ... • Surface sterilization process ... • Common](https://reader030.vdocument.in/reader030/viewer/2022021504/5aaaf6307f8b9a693f8b47da/html5/thumbnails/2.jpg)
Objectives
• Review NO2 processes – Sterilization – Decontamination
• Give examples of applications – Potential benefits – Limitations
2
![Page 3: Nitrogen Dioxide (NO PDA: A Global 2 Sterilization and ... · PDF file2) Sterilization and Decontamination Systems . November 4, ... • Surface sterilization process ... • Common](https://reader030.vdocument.in/reader030/viewer/2022021504/5aaaf6307f8b9a693f8b47da/html5/thumbnails/3.jpg)
Nitrogen Dioxide Gas Sterilant
• Nitrogen Dioxide (NO2) developed for sterilization and decontamination
• Implemented by multiple companies (3 systems on order, 8 cycle validation projects for drug delivery systems)
• Demonstrated efficacy • NO2 has broad application to prefilled syringes
– Decontamination of components – Surface sterilization of packaged syringes and delivery
systems (with and without vacuum – Depyrogenation of components and filling lines
3
![Page 4: Nitrogen Dioxide (NO PDA: A Global 2 Sterilization and ... · PDF file2) Sterilization and Decontamination Systems . November 4, ... • Surface sterilization process ... • Common](https://reader030.vdocument.in/reader030/viewer/2022021504/5aaaf6307f8b9a693f8b47da/html5/thumbnails/4.jpg)
NO2 Strengths and Limitations
• Surface sterilization process – Material permeation is low compared to EO (leads
to faster aeration) – Low permeability limits device configurations
(structures must be open to gas) • Excellent results with long lumens, complicated
geometries, etc. • Most materials are compatible with the NO2
process – Exceptions: Paper, Delrin, Nylon, Polyurethane
4
![Page 5: Nitrogen Dioxide (NO PDA: A Global 2 Sterilization and ... · PDF file2) Sterilization and Decontamination Systems . November 4, ... • Surface sterilization process ... • Common](https://reader030.vdocument.in/reader030/viewer/2022021504/5aaaf6307f8b9a693f8b47da/html5/thumbnails/5.jpg)
NO2 Strengths and Limitations
• Biocompatibility – No increase in cytotoxicity • Common packaging materials are compatible
including Tyvek, Mylar, trays and tubs, etc. • Consumables available including: BI’s, CI’s,
NO2 detectors, humidity detector, scrubber material, sterilant, etc.
5
![Page 6: Nitrogen Dioxide (NO PDA: A Global 2 Sterilization and ... · PDF file2) Sterilization and Decontamination Systems . November 4, ... • Surface sterilization process ... • Common](https://reader030.vdocument.in/reader030/viewer/2022021504/5aaaf6307f8b9a693f8b47da/html5/thumbnails/6.jpg)
Safety of Sterilant Gases Safety-related Properties
NO2 Hydrogen Peroxide
Ethylene Oxide
Color Reddish-Brown
Colorless Colorless
Odor Threshold 0.1 ppm Odorless 200 – 400 ppm
OSHA PEL (Exposure Limit)
5 ppm 1 ppm 1 ppm
NFPA: Health 3 3 3
NFPA: Flammability
0 0 4
NFPA: Instability 0 1 3
6
NO2 can be smelled at concentration levels below harmful levels
![Page 7: Nitrogen Dioxide (NO PDA: A Global 2 Sterilization and ... · PDF file2) Sterilization and Decontamination Systems . November 4, ... • Surface sterilization process ... • Common](https://reader030.vdocument.in/reader030/viewer/2022021504/5aaaf6307f8b9a693f8b47da/html5/thumbnails/7.jpg)
NO2 Sterilization Overview
Feature Parameter Range Nitrogen dioxide gas sterilant Boiling point 21°C
Low sterilant concentration 10 mg/L – 15 mg/L (less than 1%)
Room temperature process Consistent results from 10°C to 65°C
Humidity reduces cycle time 70% - 80% RH provides rapid lethality, 0% RH requires longer exposure time
Fast cycle exposure times Typically less than 20 minutes
Low sterilant residuals Residuals often not measureable, no increase in cytotoxicity observed
7
![Page 8: Nitrogen Dioxide (NO PDA: A Global 2 Sterilization and ... · PDF file2) Sterilization and Decontamination Systems . November 4, ... • Surface sterilization process ... • Common](https://reader030.vdocument.in/reader030/viewer/2022021504/5aaaf6307f8b9a693f8b47da/html5/thumbnails/8.jpg)
Typical Vacuum Cycle
The number of pulses is a cycle variable and depends on the specific load
1st Half-Cycle 2nd Half-Cycle
8
![Page 9: Nitrogen Dioxide (NO PDA: A Global 2 Sterilization and ... · PDF file2) Sterilization and Decontamination Systems . November 4, ... • Surface sterilization process ... • Common](https://reader030.vdocument.in/reader030/viewer/2022021504/5aaaf6307f8b9a693f8b47da/html5/thumbnails/9.jpg)
Typical Without Vacuum
• The gas fill and aeration depend on load conditions. • This cycle is for surface sterilization of syringes. Isolator
decontamination requires less dwell time 9
![Page 10: Nitrogen Dioxide (NO PDA: A Global 2 Sterilization and ... · PDF file2) Sterilization and Decontamination Systems . November 4, ... • Surface sterilization process ... • Common](https://reader030.vdocument.in/reader030/viewer/2022021504/5aaaf6307f8b9a693f8b47da/html5/thumbnails/10.jpg)
Mechanism of Sterilization • The NO2 sterilant has a much lower oxidation potential than other
sterilants (not principle mechanism of inactivation) • The NO2 process degrades DNA. The degradation is not reversible • NO2 exposure is not carcinogenic or mutagenic for humans
Biological Process NO2 Results
DNA Integrity Fragmented
ATP Production Inhibited
Germination Process Disrupted
Reproduction None
Oxidation Potential of Selected Gases
Oxidant Oxidation potential, (V)
Flourine 3.0 Hydroxyl radical 2.8 Ozone 2.1 Hydrogen peroxide 1.8 Noxilizer’s Sterilant, NO2
-0.8
10
![Page 11: Nitrogen Dioxide (NO PDA: A Global 2 Sterilization and ... · PDF file2) Sterilization and Decontamination Systems . November 4, ... • Surface sterilization process ... • Common](https://reader030.vdocument.in/reader030/viewer/2022021504/5aaaf6307f8b9a693f8b47da/html5/thumbnails/11.jpg)
y = -0,5626x + 6,3453 R² = 0,9842
0
1
2
3
4
5
6
7
0 2 4 6 8 10
Log
Surv
ivor
s (C
FU)
Time (min.)
Microbial Inactivation: D-Value • Linear inactivation of microbes with NO2
– Typical methods of D-value, SAL, etc. apply – Geobacillus stearothermophilus is the indicator organism used
D-Value = 1.77 minutes
11
![Page 12: Nitrogen Dioxide (NO PDA: A Global 2 Sterilization and ... · PDF file2) Sterilization and Decontamination Systems . November 4, ... • Surface sterilization process ... • Common](https://reader030.vdocument.in/reader030/viewer/2022021504/5aaaf6307f8b9a693f8b47da/html5/thumbnails/12.jpg)
Material Compatibility
Materials tested for multiple cycles with no observed degradation
* Depends on grade.
Compatible Materials Stainless Steel Polyethylene Polyetherimide Thermoplastic
Elastomers Aluminum Polypropylene Polycarbonate
Gold (Plating) PET / PETG Cyclic Olefins Polylactides Bioresorbables Glass / Ceramic Polystyrene PVC*
Fluoropolymers Polysulfones Silicone* This list is not exhaustive. Viton (Gaskets) PEEK / PAEK Hypalon
12
![Page 13: Nitrogen Dioxide (NO PDA: A Global 2 Sterilization and ... · PDF file2) Sterilization and Decontamination Systems . November 4, ... • Surface sterilization process ... • Common](https://reader030.vdocument.in/reader030/viewer/2022021504/5aaaf6307f8b9a693f8b47da/html5/thumbnails/13.jpg)
Materials Compatibility
*Requires lower humidity cycles. • NOTE: Each device needs to be evaluated on an individual basis, as geometry
and load configuration can affect sterilization.
Noxilizer is not compatible with a few medical device materials.
Incompatible Material Design Alternatives Polyurethane Thermoplastic Elastomers (TPE)
Nylon Polyester, Polyolefin
Delrin (polyacetal) PEEK, PSU, PEI
Cellulose-based (some paper) Polyester or styrene label stock
Copper-containing alloys* Stainless steel
13
![Page 14: Nitrogen Dioxide (NO PDA: A Global 2 Sterilization and ... · PDF file2) Sterilization and Decontamination Systems . November 4, ... • Surface sterilization process ... • Common](https://reader030.vdocument.in/reader030/viewer/2022021504/5aaaf6307f8b9a693f8b47da/html5/thumbnails/14.jpg)
Applications • Sterilization of vials prior to filling • Syringe tub decontamination system • Surface sterilization of prefilled syringes • Needle lumen sterilization (drug delivery
system) • Sterilizer unit for medical devices
14
![Page 15: Nitrogen Dioxide (NO PDA: A Global 2 Sterilization and ... · PDF file2) Sterilization and Decontamination Systems . November 4, ... • Surface sterilization process ... • Common](https://reader030.vdocument.in/reader030/viewer/2022021504/5aaaf6307f8b9a693f8b47da/html5/thumbnails/15.jpg)
Sterilization of Syringes and Vials
• Prior to filling, syringes in tubs and vials in trays can be rapidly sterilized or packaging decontaminated.
• Important considerations are: – Minimal sterilant residuals on surfaces that
contact the product – Processing speed to maintain production rate
• Test for sterilant residuals on surface of vials and syringes
15
![Page 16: Nitrogen Dioxide (NO PDA: A Global 2 Sterilization and ... · PDF file2) Sterilization and Decontamination Systems . November 4, ... • Surface sterilization process ... • Common](https://reader030.vdocument.in/reader030/viewer/2022021504/5aaaf6307f8b9a693f8b47da/html5/thumbnails/16.jpg)
Cycle Parameters
• Double the cycle used to sterilize vials: worst-case scenario
Results • Ion chromatography on water
extract * – Below LOD: [NO3
-] < 2.0 µg/mL
• pH measurement on 0.01 M NaCl extract – Converted to NO2 on surface:
0.49 nmol NO2 / cm2
• Residuals are below NO3-
amount permitted in WFI
Parameter Set Point
NO2 Concentration 10 mg/L
Relative Humidity 60 % RH
Dwell Time (per pulse) 10 min.
Number of Pulses 4
PassPort Residuals Test
16
![Page 17: Nitrogen Dioxide (NO PDA: A Global 2 Sterilization and ... · PDF file2) Sterilization and Decontamination Systems . November 4, ... • Surface sterilization process ... • Common](https://reader030.vdocument.in/reader030/viewer/2022021504/5aaaf6307f8b9a693f8b47da/html5/thumbnails/17.jpg)
Cycle Parameters Results
• Non-cytotoxic response – COC vials – Butyl rubber stoppers
Test COC Vial Result
Butyl Rubber Stopper Result
MEM Elution
Cytotox. 0 0
Direct Contact Cytotox.
0 0
• Cyclic olefin copolymer (COC) vials, 3-mL
• West butyl rubber stoppers
Parameter Set Point
NO2 Concentration 10 mg/L
Relative Humidity 60 % RH
Dwell Time (per pulse) 10 min.
Number of Pulses 4
PassPort Residuals Test
17
![Page 18: Nitrogen Dioxide (NO PDA: A Global 2 Sterilization and ... · PDF file2) Sterilization and Decontamination Systems . November 4, ... • Surface sterilization process ... • Common](https://reader030.vdocument.in/reader030/viewer/2022021504/5aaaf6307f8b9a693f8b47da/html5/thumbnails/18.jpg)
PassPort 60 System • Nominal 60 liter chamber • Accommodates 4 tubs per cycle.
15 minute cycle = 16 tubs / hr = 1600 syringes per hour
18
![Page 19: Nitrogen Dioxide (NO PDA: A Global 2 Sterilization and ... · PDF file2) Sterilization and Decontamination Systems . November 4, ... • Surface sterilization process ... • Common](https://reader030.vdocument.in/reader030/viewer/2022021504/5aaaf6307f8b9a693f8b47da/html5/thumbnails/19.jpg)
Noxilizer PassPort System • This system is designed to be used when introducing
work items into an isolator • Versions can have various sizes, from 60 L volume
(large enough for two syringe tubs) to 300 L (for up to 16 syringe tubs)
• Versions can be designed with or without vacuum capability
• Requirements: – Less than 15 minute process – Six log reduction – Aerate to < 5 ppm – No residuals
19
![Page 20: Nitrogen Dioxide (NO PDA: A Global 2 Sterilization and ... · PDF file2) Sterilization and Decontamination Systems . November 4, ... • Surface sterilization process ... • Common](https://reader030.vdocument.in/reader030/viewer/2022021504/5aaaf6307f8b9a693f8b47da/html5/thumbnails/20.jpg)
PassPort System Cycle Time
• This cycle is recorded from a 50 L chamber used for testing
• A scrubber is used in the exhaust path (between the chamber and the pump) to remove NO2
0
200
400
600
800
0 1 2 3 4 5 6 7 8 9 10 11 12 13
Pres
sure
(Tor
r)
Time (Minutes)
20
![Page 21: Nitrogen Dioxide (NO PDA: A Global 2 Sterilization and ... · PDF file2) Sterilization and Decontamination Systems . November 4, ... • Surface sterilization process ... • Common](https://reader030.vdocument.in/reader030/viewer/2022021504/5aaaf6307f8b9a693f8b47da/html5/thumbnails/21.jpg)
PassPort Lethality Test • Two types of BI’s used: ‘normal’ BI’s (in pouch)
and coupon BI’s • Coupon BI’s are cut from the tub rim • To form the biological indicator (BI), the spore
suspension, with > 106 spores, was placed under the Tyvek flap that remained on the tub rim and allowed to dry
Tyvek
Tub Rim
Spores Coupon BI
21
![Page 22: Nitrogen Dioxide (NO PDA: A Global 2 Sterilization and ... · PDF file2) Sterilization and Decontamination Systems . November 4, ... • Surface sterilization process ... • Common](https://reader030.vdocument.in/reader030/viewer/2022021504/5aaaf6307f8b9a693f8b47da/html5/thumbnails/22.jpg)
• Single exposure pulse used for this lethality ranging study • A single exposure at 5 minutes with 5 mg/L NO2 concentration will
decontaminate tubs
Exposure Time
BI Results (# Pos./# Tested)
Tub BI Results (# Pos./# Tested)
2 Min 0/5 5/5
3 Min 0/5 1/5
4 Min 0/5 3/5
5 Min 0/50 0/5
6 Min 0/40 0/5
Enumeration Results Fraction Negative Results (5 mg/L)
PassPort Lethality Test
22
![Page 23: Nitrogen Dioxide (NO PDA: A Global 2 Sterilization and ... · PDF file2) Sterilization and Decontamination Systems . November 4, ... • Surface sterilization process ... • Common](https://reader030.vdocument.in/reader030/viewer/2022021504/5aaaf6307f8b9a693f8b47da/html5/thumbnails/23.jpg)
Pre-filled Syringes
• After filling, syringes are packaged and sterilized* for ophthalmic applications, used during surgery, etc.
*Only the surface of the container are sterilized. The compound in the container is not sterilized.
• Important considerations are: – Temperature of the sterilization process – Contamination of the API due to sterilant
ingress (permeability of the closure system) • Test ingress with WFI filled syringes
23
![Page 24: Nitrogen Dioxide (NO PDA: A Global 2 Sterilization and ... · PDF file2) Sterilization and Decontamination Systems . November 4, ... • Surface sterilization process ... • Common](https://reader030.vdocument.in/reader030/viewer/2022021504/5aaaf6307f8b9a693f8b47da/html5/thumbnails/24.jpg)
Pre-filled Syringes
Cycle Parameters
• Approximately double the cycle used to sterilize syringes – Worst-case scenario
Results • Ion chromatography on water
extract * – Below LOD: [NO3
-] < 2.0 µg/mL
• pH measurement on 0.01 M NaCl extract – Control = 6.89, Exposed = 6.65 – Converted to ppm NO3
- : 0.002 ppm NO3
-
• Residuals are low in concen.: – WFI limit: 0.2 ppm NO3
-
Parameter Set Point
NO2 Concentration 15 mg/L
Relative Humidity 60 % RH
Dwell Time 75 min.
* Test was performed by an external laboratory. 24
![Page 25: Nitrogen Dioxide (NO PDA: A Global 2 Sterilization and ... · PDF file2) Sterilization and Decontamination Systems . November 4, ... • Surface sterilization process ... • Common](https://reader030.vdocument.in/reader030/viewer/2022021504/5aaaf6307f8b9a693f8b47da/html5/thumbnails/25.jpg)
Syringe and Vial Summary
• NO2 is a new method of sterilization/decontamination
• This method provides a viable method for sterilizing syringes, vials, etc.
• Results show: – Sterility assurance level of 10-6 possible – Material compatibility – Low level of sterilant residuals (WFI)
• Rapid cycle time for efficient manufacturing – In-line sterilization or batch processing
25
![Page 26: Nitrogen Dioxide (NO PDA: A Global 2 Sterilization and ... · PDF file2) Sterilization and Decontamination Systems . November 4, ... • Surface sterilization process ... • Common](https://reader030.vdocument.in/reader030/viewer/2022021504/5aaaf6307f8b9a693f8b47da/html5/thumbnails/26.jpg)
Isolator Decontamination and Aeration
Data recorded in a heavily loaded isolator made be Walker Barrier Systems 26
![Page 27: Nitrogen Dioxide (NO PDA: A Global 2 Sterilization and ... · PDF file2) Sterilization and Decontamination Systems . November 4, ... • Surface sterilization process ... • Common](https://reader030.vdocument.in/reader030/viewer/2022021504/5aaaf6307f8b9a693f8b47da/html5/thumbnails/27.jpg)
NO2 Aeration of 800 L Isolator • Aeration begins at 8.5 minutes • Concentration in the isolator is below 1 ppm in 5
minutes of aeration and below 100 ppb in 13 minutes (Draeger tubes and EC Cells)
0,01
0,1
1
10
100
1000
10000
0,0 5,0 10,0 15,0 20,0 25,0
Conc
entr
atio
n (p
pm)
Time (Minutes)
FTIR Data
EC Cell
Draeger Tube
Cycle # NO2 Concentration Dwell (Minutes)
Positive BI's SSTL
20140808A 14 mg/L 7.5 0/12
27
![Page 28: Nitrogen Dioxide (NO PDA: A Global 2 Sterilization and ... · PDF file2) Sterilization and Decontamination Systems . November 4, ... • Surface sterilization process ... • Common](https://reader030.vdocument.in/reader030/viewer/2022021504/5aaaf6307f8b9a693f8b47da/html5/thumbnails/28.jpg)
Syringe Needle Sterilization • Needle lumen is most challenging location • Demonstrate inoculation and recovery in syringe needles • Average inoculation = 1.09 x 106 per syringe • Needles had between 104 and 105 spores in the lumen • 12 inoculated syringes in each cycle, in tub, three cycles • Cycle parameters were: 6 min. dwell,10 mg/L, 75% RH
• All 36 syringes and syringe
needles are sterilized • SAL requires a second
half-cycle)
28
![Page 29: Nitrogen Dioxide (NO PDA: A Global 2 Sterilization and ... · PDF file2) Sterilization and Decontamination Systems . November 4, ... • Surface sterilization process ... • Common](https://reader030.vdocument.in/reader030/viewer/2022021504/5aaaf6307f8b9a693f8b47da/html5/thumbnails/29.jpg)
Endotoxin Dose Response
• 1 pulse, 60 minutes, 75% RH (cycle with no vacuum) • 4 pulses, 10 minutes, 75% RH (fast cycle)
• Cycle validated on an aseptic filling line 29
![Page 30: Nitrogen Dioxide (NO PDA: A Global 2 Sterilization and ... · PDF file2) Sterilization and Decontamination Systems . November 4, ... • Surface sterilization process ... • Common](https://reader030.vdocument.in/reader030/viewer/2022021504/5aaaf6307f8b9a693f8b47da/html5/thumbnails/30.jpg)
Vacuum Sterilization Units • RTS 360: Simple, effective and economical in-house
sterilization – Operates at Room Temperature – Maintains Material Properties – Shorter Cycle Time – No Pre-Conditioning – No Lengthy Aeration Required – No Residuals – Free Standing/Non-Hazardous Bi-Product – Safer than Ethylene Oxide – Fully Scalable Multi-pallet system available Nox. RTS-360 Typical EO System
Chamber Capacity 360 Liters 2200 Liters
Standard Cycle Time 80 Minutes 12-18 Hours *
Manufacturing to Release Time
On-Site, Immediate Use
Off-Site, 7-25 Day Turnaround
Patented Noxilizer Sterilization System
30
![Page 31: Nitrogen Dioxide (NO PDA: A Global 2 Sterilization and ... · PDF file2) Sterilization and Decontamination Systems . November 4, ... • Surface sterilization process ... • Common](https://reader030.vdocument.in/reader030/viewer/2022021504/5aaaf6307f8b9a693f8b47da/html5/thumbnails/31.jpg)
NO2 Processes
• NO2 process offers a new solutions
• Contact details David Opie
European Contact:
Stephen Morley [email protected]
31