OPERATING INSTRUCTIONS
IMPORTANT
This manual provides operating instructions for the Maximus 5000 series. The instructions contained in this booklet should be thoroughly read and understood before operating the unit and chair. After the installation has been completed, keep this manual in a safe place and refer to it for future maintenance.
Quolis 5000 Dental Unit
Intended Use of the Product
This product is an active therapeutic device intended to administer or exchange energy of electric, air and water for the exclusive use for diagnoses, treatments and relative procedures of dentistry, and its characteristic is not in a potentially hazardous way between such energy and human body, taking account of the nature, the density and site of application of the energy.The product must be operated or handled by the qualified dentists or by dental staffs under the supervision of the dentist.Such dentists or dental staffs should instruct and/or assist the patients to approach to and leave from the product.Patients should not be allowed to operate or handle the product unless he/she is so instructed. The product is supplied together with the handpieces like electric micromotor, air turbine and/or motor, scaler and so on.
Environmental RequirmentsAmbient Temperature Operating 41° to 104° Storage 14° to 122F°Humidity 10 % - 80%Atmospheric Pressure 8.7psi – 15.4psi
Classification of Eqipmenta. Type of shock protection : Class I Equipment
b. Degree of shock protection : Type B Applied part
c. Degree of protection against water ingress : Ordinary Equipment (All Products)
d. Mode of operation : Continuous Operation
e. Flammable Gasses : Not suitable for use in the presence of a flammable anesthetic mixture with air,
oxygen, or nitrous oxide.
f. Earth : Protective earth (ground)
g. Hospital Grade Plug : Grounding reliability can be achieved only when the equipment is connected to an
equivalent receptacle marked hospital only or hospital grade.
Page 1. OVERALL VIEW AND MAJOR COMPONENTS ------------------- 1
2. DIMENSIONS AND SPECIFICATIONS ------------------------------- 2 2-1. DIMENSIONS ------------------------------------------------------- 2 2-2. SPECIFICATIONS -------------------------------------------------- 2
3. OPERATING INSTRUCTIONS ----------------------------------------- 3 3-1. DOCTOR TABLE SECTIONS ------------------------------------ 2 3-2. VACUUM PACK SECTION --------------------------------------- 9 3-3. FOOT CONTROL SECTION -------------------------------------- 10
4. RIGHT / LEFT HANDED DENTISTRY CONVERSION ----------- 11
5. CARE AND MAINTENANCE ------------------------------------------- 12
6. UNIT FLOW DIAGRAM 6-1. DOCTOR TABLE SECTION -------------------------------------- 13 6-2. SWING ARM SECTION ------------------------------------------- 14 6-3.VACUUM PACK SECTION ---------------------------------------- 14 6-4. JUNCTION BOX SECTION --------------------------------------- 15 6-5. OVERALL UNIT FLOW DIAGRAM ---------------------------- 16
7. ELECTRICAL DIAGRAM 7-1. CHAIR SECTION --------------------------------------------------- 17 7-2. UNIT SECTION ----------------------------------------------------- 18
8. LOCATION OF PRODUCT LABELS --------------------------------- 19
TABLE OF CONTENTS
14
15
16
24
17 1918
21
2223
20
1
2
34
78
9
10
11
12
13
56
1. OVERALL VIEW AND MAJOR COMPONENTS
(17) SE Handpiece(18) Assistant Side Syringe(19) Assistant Touch Pad(20) Solid Collector (21) Assistant Swing Arm(22) Assistant Holder Arm(23) Vacuum Pack Control Panel(24) Dental Light
Fig.1-1 Overall View and Major Components
-1-
(1) Control Head (2) Main Control Panel (3) Dr. Table Indicator (4) Handle (5) Dr. Side Handpiece Holder (6) Dr. Side Syringe (7) Handpiece Hose (8) Foot Controller
(9) Balance Arm(10) Dr Swing Arm(11) Water Bottle(12) Junction Box(13) Dental Light Swing Arm(14) Light Pole(15) Assistant Holder(16) Vacuun Handpiece
2. DIMENSIONS AND SPECIFICATIONS 2-1. DIMENSIONS
- Inch /(mm) --Tolerance ±10%-
Power Consumption ------------------------------AC120V 1A
Frequency ------------------------------------------60 Hz
Fuse value ------------------------------------------AC125V 10A
Control Voltage ------------------------------------DC5V
Table Unit Net Weight ----------------------------25 kg
Swing Arm Net Weight ---------------------------25 kg (Without Dental Light)
Vacuum Pack Net Weight ------------------------20 kg
Junction Net Weight ------------------------------ 7 kg
Doctor Table Maximum Load ------------------- 2 kg(Gross weight of the doctor table are 8kg or less, the
spring of the balance arm can be adjusted.)
Operating Pressure -------------------------------- Water 0.2 MPa (30 psi) Max. 0.28 MPa (40 psi)
Air 0.5 MPa (75 psi)
Fig.2-1 Dimensions
-2-
2-2. SPECIFICATIONS
-3-
Fig.3-1-1 Master SwitchTurn off the master switch after daily operation and for long term interval.
3. OPERATING INSTRUCTIONS
3-1. DOCTOR TABLE SECTION
3-1-1. Master Switch (Fig.3-1-1 & Fig.3-1-2) Turn on the master switch located under the doctor table, the power indicator on the main control panel will then illuminate green.
3-1-2. Control Panel Main Control Panel
ONOFF
Fig.3-1-2 Main Control Panel
(1) Power Indicator(2) Function Switch(3) Spray Switch(4) Light Pack Switch(5) Function Indicator (6) Safety Lock Indicator(7) Dental Light Switch
(8) Electric Motor Direction Control Switch (9) Select Switch(10) Decreasing / Increasing Switch(11) Store Switch (12) Timer Switch(13) Chair Manual Control Switches(14) Chair Auto Mode Switch(15) Chair Auto Return Switch
1 2 3 4 5
6 7 8 9 10 11 12
13
15
14
Power Indicator
(1) Power Indicator Turn on the master switch, the power indicator will illuminate green.
(2) Function Switch Provides supplement function (flush out operation, control panel switching sound on/off, fiber optic handpiece illumination timing, changing the maximum speed of micromotor atvariable mode, selection of doctor number, etc.) selected by the function switch. (3) Spray Switch Turning on/off the coolant air/water for instruments. Micromotor can be selected four mode (air and water / air only / water only / none). Air turbine and air motor is available two mode (air and water / none) Electric scaler, only water is supplied.
Function Switch
Spray Switch
(4) Light Pack Switch Pick up the fiber optic handpiece from the holder, momentarily press the light pack switch, the indicator illuminates in green and the handpiece light turns on. To switch off the light pack, simply press the light pack switch again.
(5) Function Indicator Display the status of setting or handpiece. a) Operation after turning on the main switch After main switch is turned on , the following is displayed and jobs are done. 1) All the LED on the control panel lit. Program No.(Ver. No), Doctor No. are displayed. 2) Reading of several data from the memory. Reading of the status of the instrument holders. 3) Unless any promlem is found, the displays go off and only "Belmont" is displayed. 4) After the may SW is turned on, the above-mentioned process takes 5 seconds. Chairs cannot be operated during this period.
b) When instruments are properly put in the holders 1) The controller detect the holder where the instruments are not properly put, then make a beep and display the holder No(s). If you put the instrument properly during the beep, the beep stops and you can use the holder normally. 2) The holder which were detected not to be properly put with instroments cannot be used. In order to put an instrument there, turn off the main switch, put the instrument and then turn on the switch again after 5 seconds.
c) Automatic turning off of the display After several seconds, the displays go off automatically. 1) In 120 seconds after the final normal display (Belmont). 2) In 30 seconds after any operation. After that it display turns on when any switch on the panel is pressed.
(6) Safety Lock Indicator While the handpiece is running, the safety lock indicator illuminates amber and all chair control switches will not operate.
(7) Dental Light Switch Press the dental light switch to turn on/off the dental light.
Light Pack Switch
Light Pack Indicator
Safety Lock Indicator
Dental Light Switch
(8) Electric Motor Rotation Direction Control Switch (Optional) After picking up the electric motor from the holder, the electric motor rotation direction can be changed by momentarily pressing this switch the rotation direction will be indicated by the amber and green LEDs. Indicator in green : Forward Rotation Indicator in amber : Reverse Rotation (Note : Do not change the electric motor direction while the motor is running. When the electric motor with setting Reverse Rotation is returned the holder and picked up again, a buzzer sounds.
(9) Select Switch Selection of operation mode pf micromotor, scaler,. Selection of setting of function.
(10) Decreasing / Increasing Switch Press the decrease/increase switch to change mocromotor speed, scaler power, timer, alame etc.
(11) Store Switch Various setting can be put in memory pressing by store switch. The setting can be used when the main switch is turned on.
(12) Timer Switch Timer can be set maximum 90 mins. 59 secs. in 1 secs. segment. Momentary press the timer switch, and set the time by pressing decrease / increase switch.
(13) Chair Manual Control Switches 1) Seat Lifting -----------------Press the ( ) switch until the seat is lifted up to the desired position. 2) Seat Lowering --------------Press the ( ) switch until the seat is lowered to the desired position. 3) Backrest Reclining ---------Press the ( ) switch until the backrest is reclined to the desired position. 4) Backrest Raising -----------Press the ( ) switch until the backrest is raised up to the desired position.
(14) Chair Auto Mode Switches Preset Operation Momentarily press the preset-1 switch ( ), the chair moves to the preset 1 position and stops automatically. Preset 2 position operated by the preset switch ( ). Preset 3 position operated by the preset switch ( ).
(Note:For auto mode position adjustment refer to chair operation manual )
Electric Motor Direction Control Switch
Green LED(Forward Rotaition)
Amber LED(Reverse Rotation)
Decrease/Increase Switch
Timer Switch
Select Switch
Store Switch
Chair Manual Control Switches
Chair Auto Mode Switches
3-1-3. First priority function of the instruments The priority of the operation is given only to the instrument that has been withdrawed first. Only that instrument will be operable. 1) After all the handpiece are placed in the holders, the first instruments withdrawed will be operable. When more than one instruments are withdrawed and the first one is returned to the holder, “Return to the holder” will be displayed. After the second or third instrument are placed back to the holder, then withdraw agan the instrument that you want to use.
2) The operable instrument number will be displayed (see example)
3) The instrument will not be operable if it is withdrawed with the foot controller being pressed. Make sure to withdraw the handpiece before pressing the foot controller. The orange LED for rotation direction (counterclockwise rotation) will lit.
(15) Chair Auto Return Switch 1) Momentarily press the auto return switch ( ), the chair returns to the initial position (the seat is the lowest position and the backrest is the upright position) and stops automatically.
2) Emergency Stop (Safety Stop) During automatic movements (preset, auto return and last position memory), by momentarily pressing any chair control switch this will cancel the automatic movement immediately.
Auto Return Switch
3-1-4. Unit Supplement Functions A. Spray Switch
1) Turning on/off the coolant air/water for instruments. 1) Pick up the instrument from the holder.
2) Select spray mode
2) Setting and memory 1) Do the above mentioned setting 2) Momentarily press store switch to put it in memory. (Do not press the foot control)
3) The coolant air and water for instrument can be supplied in four modes or two modes. For micromotor, [Four Mode] can be selected. , Air and water / air only / water only / none. For air turbine and air motor, only [Two Mode] is available. Air and water / none. For electric scaler, only water is supplied. 4) Set to four mode or two mode
1)Momentarily press the function switch seven times.
2) Press the decrease switch for two mode 3) Press the increase switch for four mode (Note:Press any of the above keys in 10 seconds after
pressing function switch )
Spray Switch
C. Maximum Rotation Speed of Micromotor (available at variable rotation speed mode) The maximum rotation speed (rotation speed range) can be set. The speed depend on the model of the micromotr. The set speed remains until the man switch is turned off. Press [Store] switch to put it in memory. When two micromotor is used, the setting can be done for each one. 1) Press select switch( ) until “Va” is displayed on top of the display.
2) Press the decrease switch or increase switch ( ) to changing the maximum speed.
for DA-700(No torque adjustment)
for Bien Air MXVariable Mode Maximum Speed
Torque(%)Rotation Mode(Auto Reverse)
The currently adjusted speed
B. Fiber optic handpiece lighting mode (Optional) In case that fiber optic handpiece is installed, the fiber optic turns on when the handpiece is taken out of the holder, and turns off when the handpiece is returned to the holder. This could be changed to fiber optic turns on when the handpiece is taken out of the holder and drive air pedal of foot control is activated.
Setting 1) Press the function switch four times. 2) Press the increase switch ( ) for H (holder) mode. Press the decrease switch ( ) for F (foot control) mode. (Note: Press any of the above keys in 10 seconds after pressing function switch ) 3) Press store switch without rotating the instruments to put it in memory.
Function Switch Decrease/Increase Switch
The Maximum speed of micromotor at variable speed mode The Maximum speed of micromotor at variable speed mode can be selected between “three steps” or “five steps”. At three steps, the maximum speed will be 5000, 10000 and 40000 rpm. At five steps, the maximum speed will be 1000, 5000, 10000, 20000 and 40000rpm. The rotation speed range depends on the model of the micromotor.
1) Momentarily press the function switch six times and press the
decrease switch for three steps.
2) Momentarily press the function switch six times and press the
increase switch for five steps. (Note:Press any of the above keys in 10 seconds after pressing function switch.)
Function Switch Decrease/Increase Switch
3 steps
2000 - 10000
2000 - 20000
2000 - 40000
5 steps
2000 - 5000
2000 - 10000
2000 - 20000
2000 - 30000
2000 - 40000
DA-700
3 steps
100 - 5000
1000 - 10000
1000 - 40000
5 steps
100 - 1000
100 - 5000
1000 - 10000
1000 - 20000
1000 - 40000
MX
-4-
E. Fixed Rotation Speed Mode This function can provide constant speed of electric motor up to three different speed(M1,M2,M3). The set speed remains until the man switch is turned off. Press [Store] switch to put it in memory. NOTE ; Once this function selected, foot controller behaves as on/off switch of electric motor and rotation speed can not be varied by foot controller. 1) Pick up the electric motor from the holder. 2) Momentarily press the select switch( )and repeat it until selected setting figure is displayed in function indicator.
To change the setting speed. 1) Pick up the electric motor from the holder. 2) Momentarily press the electric motor speed setting switch and repeat it until selected setting figure is displayed in function indicator. 3) Set up the required rotation speed by using increasing and/or decreasing switch. 4) Momentarily press store switch.
Rotation Speed Control Volume Knob
D. Rotation Speed Control Volume Knob (Optional) Turning the rotation speed control volume knob to change the rotation speed of micromotor.
(Note: When use the micromotor depressing the foot controller
to obtain enough coolant air. )
F. Torque adjustment of Bien Air MX Torque of rotation (%), Rotation mode (normal, auto forward, auto reverse), time of reverse rotation at auto forward setting and the intensity of the light can be adjusted. The setting can be done both for variable rotation mode and pre-set rotation mode. Torque adjustment and rotation mode change can be done only in low rotation speed (100 to 5000rpm). In high rotation speed, the torque remains 100%.
Setting 1) Withdraw MX motor 2) Press “F” until the display become like the example at right. 3) Press [ Select ] and select the item to adjust. 4) “>” will be displayed at the left side of the item to be adjusted. 5) Press [ - ] [ + ] to adjust. 6) To finish the setting, press [Store] switch. To cancel, press [F].
• Torque Setting The torque can be adjusted in the range of 10 to 100% In high speed rotation mode, the torque stays 100%. The actual torque value may change depending on the power input or the status of the motor.
Decrease/Increase Switch
Select SwitchFunction Switch
Store Switch
G. Timer Timer can be set maximum 90 mins. 59 secs. in 1 secs. segment. Preset time
1) Momentarily press the timer swith , and select the presetting time 0 > 1 > 2 > 3 by pressing the select switch . 2) Press the decrease switch or increase switch to change
the time. Press the select switch to switch the minutes or seconds.
3) Momentarily press the timer switch to start timer. The setting time is indicated on the function indicator. (Note:The presetting time remains until the man switch is turned off. Press Store switch to put it in memory)
Decrease/Increase Switch
Timer Switch Select Switch
Store Switch
• Rotation mode 1)NOR(Normal Mode) The motor remain rotation in onedirection regardless the load to the motor. 2) A. Rev (Auto reverse) The motor stop by a specific (programmable) load and turn to oposit direction. When you stop the motor and the restart, the motor turns in original direction. 3) A.FOW Auto Forward The motor stop by a specific (programmable) load and turn to oposit direction, then after a speficic (programmable) period, it the motor rotates in original direction again.
• Time of opposite rotation in Auto Forward mode The time of rotation in opposite direction in Auto Forward mode. (Note:The actual rotation time in the opposite direction will be longer than the displayed time approximately by 2 seconds.) • Intensity of the light of micromotor The intensity of the light of MX micromotor can be adjusted, within the range of 00 to 15.
Preset Time Adjustment Four preset time can be set. ( 0 > 1 > 2 > 3 )
1) Momentarily press the timer switch.
2) Press the select switch and select the preset number 0.
3) Press the decrease switch or increase switch to change
the time. Press the select switch to switch the minutes or seconds.
4) Press the store switch , then thme is memorized for preset 0. 5) Preset time 1 or 2.3 can be memorized by above 2 to 4.
-5-
H. Selection of Doctor Number Up to four dentists can program their own setting of the followings 1) Pre-set speed of micromotor and on/off of the light. 3 speeds each. (For MX, torque and rotation mode are also pre-programmable) 2) Timer of light of air turbine after releasing the foot pedal 3) Pre-set positions of the chair, 3 positions each. (Note : The chair needs to be able to transmit signals with this PCB and be programmed for four doctors. 4) Initial intensity (high / low) of the dental light. (Note : The dental light needs to be able to transmit signals with this PCB and be programmed for four doctors. 5) The setting of [F] switch.
Setting
1) Press function switch ( ) once.
The current docoto No. will be displayed. 2) Press any [Pre-set] switch 0, 1 2 or 3. of the chair ( ) ( ) ( ) ( ) (Note:Select doctor No. in 10 seconds after pressing function switch)
Function Switch Chair Auto Mode Switch
1) Short time flush out
Momentarily press the function switch two times and
momentarily press the decrease switch. Pick up the handpieces from the holder and set them in the buclet etc. By momentarily pressing the foot controller this starts short time flush out. Water comes out from the handpiece and stops automatically after 40 seconds. During flush out, by momentarily pressing any switch of doctor table or foot controller will cancel flush out immediately. 2) Long time flush out
Momentarily press the function switch two times and
momentarily press the increase switch. Pick up the handpieces from the holder and set them in the buclet etc. By momentarily pressing the foot controller this starts long time flush out for 5 minutes. During flush out, momentarily pressing the any switch of doctor table or foot controller will cancel flush out immediately.
I. Flush out system The QUOLIS is equipped with two types of flush out system. Short time flush out - Cleaning handpiece water lines for 40 secs. Long time flush out - Cleaning handpiece water lines for 5 mins. Function Switch Decrease/Increase
Switch
-6-
J. Control panel switching sound on/off You can turn on/off or change the frequency of the beep when you hear at pressing switch. Volume of tbe beep depends on the frequency.
1) Press function switch three times.
Turn off the beep ---------- press decrease switch Turn on the beep ---------- press increase switch Higher frequency ---------- press up switch Lower frequency ----------- press down switch Default frequency ---------- press 0 switch 2) After changing frequency, press increase switch (Note: Press any of the above keys in 10 seconds after pressing function switch.)
Function Switch Decrease/Increase Switch
Up/Down Switch 0 Switch
K. Alarm sound Type of alarm sound can be selected among four. Changing alarm sound helps dentists to recognize the device making the sound.
Setting
1) Press function switc ( )five times.
2) Select alarm type 0, 1, 2 or 3 by pressing switch for pre-set positions of the chair ( ) ( ) ( ) ( ) (Note:Press any of the above keys in 10 seconds after pressing function switch )
Function Switch Chair Auto Mode Switch
L. Scaler for NSK Varios (Optional) The setting range of ultrasonic scaler can be selected in 3 ranges (Scaling, prosthesis removal, amalgam, plugging / Ultrasonic endodontic treatment / Ultrasonic periodontal treatment).
Select Switch Decrease/Increase Switch
1) Pick up the handpiece of scaler and set the range by
pressing select switch.
2)The power of the scaling can be adjustable by pressing the decrease switch or increase switch.
0 Switch
-8-
Doctor Table Section Control Knobs
3-1-5. Doctor Table Section Control Knobs A. Handpiece Spray Water Control Knobs The handpiece spray water control knobs are located under the doctor table. Each handpiece spray water control knob is 1-4 from the left side HP1,HP2,HP3,... The handpiece spray water volume can be controlled independently.
B. Doctor's Syringe Flow Control Knobs Doctor's syringe flow control knobs are located under the doctor table. The flow control knobs adjust the doctor's syringe air and water flow volume. The yellow capped knob is the air flow control knob, the blue capped knob is the water flow control knob.
Note : Turning the control knob counterclockwise will increase the flow volume and turning clockwise will decrease.
3-1-6. Handpiece pressure Gauge Handpiece drive air pressure gauge is located in the doctor table .While a handpiece is working, the handpiece drive air pressure is indicated on the handpiece pressure gauge.
Control Knob
Increase Decrease(Stop)
Handpiece Pressure Gauge
Balance Arm Air Brake
3-1-7. Balance Arm Air Brake Balance arm air brake button is located on the handle. Grasp the handle and press the air brake button to adjust the table height. Release the air brake button at the desires table position, the balaance arm is locked.
Note : Do not load over 2.0 kg on the table.
Doctor's Syringe Water Control Knobs
HP1
HP2
HP3
Doctor's Syringe Air Control Knobs
Air Brake ButtonAir Brake Button
Handpiece Pressure Gauge
-9-
(1) Assistant Side Control Panel (Fig.3-2-1) Chair auto mode switches and chair manual mode switch are located on the vacuum pack control panel. (2) Vacuum Pack Control Panel (Fig.3-2-1)
A. Service Water Outlet (Fig.3-2-1) The service water outlet provides a quick-connection for water.
B. Service Water Outlet Control Knob (Fig.3-2-1) The water volume from the service water outlet can be adjusted by the service water outlet control knob. Turning the knob counterclockwise will increase the flow volume and turning clockwise will decrease. C. Service Air Outlet (Fig,3-2-1) Service air outlet provides a quick-connection for air.
(3) Vacuum Pack height Adjustment (Fig.3-2-1) Press the lock cancel button to adjust the vacuum pack height. Release the lock cancel button at the desires position, the vacuum pack is locked. (Note: Lift up the vacuum pack a little when press the lock cancel button to adjust height)
3-2. VACUUM PACK SECTION
(4) Assistant’s Syringe Control Knobs(Fig.3-2-2) Assistant’s syringe flow control knobs are located in the cuspidor unit body. The yellow capped knob is to adjust the assistant's syringe air flow volume, and the blue capped knob is to adjust water flow volume.
Note : Turning a knob counterclockwise increase flow volume and turning clockwise will decrease.
Fig.3-2-2 Assistant’s Syringe Control Knob
Chair Auto Mode Switch
Chair Manual Control Switch
Chair Auto Return Switch
Service WaterControl Knob
Service Water Outlet
Service Air Outlet
Lock Cancel Button
Assistant Control Panel
Vacuum Pack Control Panel
Assistant's SyringeFlow Control Knobs
Increase Decrease
Fig.3-2-1 Vacuum Pack Section
-10-
Vacuum Handpiece and Saliva Ejector Handpiece
(5) Assistant Instrument Holder Saliva ejector handpiece and vacuum handpiece have stop valves to close and adjust suction power.
Saliva EjectorHandpiece
Open
Close
Vacuum Handpiece
Open
Close
3-3. FOOT CONTROL SECTION
Foot Control
Pick up a handpiece from the handpiece holder and depress the
foot control, the handpiece starts running and coolant air comes
out from the handpiece.
-11-
Light Pole
Doctor Table
Light Pole
Lock Pin
Doctor Table
4. RIGHT / LEFT HANDED DENTISTRY CONVERSION 1. Bring the doctor table to following position. (Fig.4-1) 2. Pull up the lock pin( ) on the light pole mount arm. Lift up the chair seat and rotate the light pole to the other side and set lock pin on light pole mount arm. (Fig.4-2) 3. Swing the doctor table to other side fully.(Fig.4-3) 4. Swing the vacuum pack to other side.(Fig.4-4)
Fig.4-1 Fig.4-2
Fig.4-3 Fig.4-4
Fig.5-2 Vacuum and Saliva Ejector Handpiece
Fig.5-1 Solid Collector Filter(3) Vacuum Handpiece and Saliva Ejector (Fig.5-2) A. Pull out the Vacuum hand piece from the vacuum hose and clean it. Loosen the vacuum hose nut, the vacuum hose can then be removed from the cuspidor unit.
5. CARE AND MAINTENANCE
Turn OFF the master switch after daily operation and for long term intervals.Turn the main water valve OFF after daily operation and for long term intervals.
CAUTION
CAUTION
All surfaces can be cleaned with DURR FD312 cleaner ( or equivalent).Do not drench the chair and unit.Wipe all surfaces dry after cleaning.
Cleaning Unit
(2) Solid Collector (Fig.5-1) Pull out the solid collector filter and clean it. After cleaning, refit the solid collector firmly.
Solid Collector Filter
B. After daily operation, run two cups of clean water through the handpieces to clean inside. Note : After cleaning the handpieces, apply a white vaseline lightly on the rubber O-Ring to prevent damage to this O-Ring.
-12-
Oil Mist Separator Body
Oil Mist Filter
(4) Oli Mist Separator Handpiece oil mist separator is located under the doctor table. The oil contained in the exhaust air of each handpeace is collected in this container. Once a week or oil comes to the red line, open the oil mist separator and clean the separator body and filter.
Saliva EjectorHandpiece Body
VacuumHandpiece Body
VacuumHose Saliva Ejector
Hose
Vacuum Saliva Ejector
Chip Joint
Lever
Lever
SE Screen
6. UNIT FLOW DIAGRAM6-1. DOCTOR TABLE SECTION
-13-
6-2. SWING ARM SECTION
6-3.VACUUM PACK SECTION
-14-
-15-
6-4. JUNCTION BOX SECTION
AIR
DRAIN
VACUUM
WATER
-16-
6-5. OVERALL UNIT FLOW DIAGRAM
AIR
DRAIN
VACUUM
WATER
-17-
7. ELECTRICAL DIAGRAM7-1.CHAIR SECTION
12
34
56
78
9
84736251
21
21
SV1
SV2
SV4
SV3
1 2 3 4 5 6
12
SV6
98
76
54
32
1
1 2W
hite
Yello
w
XA9P
AN
FU07
A0
11
Trun
k Li
ne B
for S
olen
oid
Valv
e
VL2P
Lock
Red
Brow
n
1051 6 2 7 3 8 4 9
CN9
5566
-10A
(MO
LEX)
43
12
VL2P
Fuse
F 10
ABlac
k
Gre
en/Y
ello
w
Black
Whi
te
Whi
teBl
ack
LN
CN18
IL-5
P-S3
EN2
(JA
E)
54
32
1
FLA
SH
CN17
IL-7
P-S3
EN2
(JA
E)
76
54
32
1
RY-2
005
(1.2
)
(3.4
)
(2.5
)
(3.4
)
3
11V
9V
Trun
k Li
ne fo
r Cap
acito
r
AN
FU12
A0
Brow
n
Yello
w
Ora
nge
Red
Gra
y
Purp
le
Blue
Gre
en
CN1-
2(1-
1)IL
-10P
-S3E
N2
(JA
E)
109
87
65
344
221
CN13
IL-3
P-S3
EN2
(JA
E)
CN14
IL-3
P-S3
EN2
(JA
E)
CN7
IL-3
P-S3
EN2
(JA
E)
CN6
IL-3
P-S3
EN2
(JA
E)
CN5
IL-3
P-S3
EN2
(JA
E)
CN12
IL-3
P-S3
EN2
(JA
E)321
CN15
IL-2
P-S3
EN2
(JA
E)211
CN2-
2(2-
1)IL
-6P-
S3EN
2(J
AE)
CN1-
1(1-
2)IL
-10P
-S3E
N2
(JA
E)
CN16
B03B
-XA
SK-1
CN9
B09B
-XA
SK-1
CN10
IL-1
0P-S
3EN
2(J
AE)
1 2 3 4 5 6
21
1 2 31 2 3
VL8P
56
109
87
65
344
221
1 2 3
Pote
ntio
met
er fo
r Bac
kres
tBr
own
Red
Ora
nge
109
87
65
43
21
Cont
rol P
.C.B
.(AG
FS13
*0)
XX-S
TD20
07/0
0TI
Rela
y P.C
.B.
120V
Yello
w
Blac
k do
t ind
icat
es m
ale
pin
side
[Rem
arks
]Co
nnec
tor S
ymbo
l
XA •
• • X
A C
onne
ctor
VL
• • •
VL
Conn
ecto
r
YL •
• • Y
L Co
nnec
tor
NM
F • •
• N
re M
ini F
it Co
nnec
tor
m •
• •
Min
i Uni
vers
al C
onne
ctor
(AM
P)Cap
Plug
21
Mot
or
123
Up
Dow
n
Rais
e
Recl
inin
g
Pote
ntio
met
er fo
r Sea
tBr
own
Red
Ora
nge
TSM
-CBC
-5LP
-M1-
E
Tran
sfor
mer
TK-1
20V-
2H
Red
Whi
te/R
ed
Brown
SW2
SW3
LIM
NO
L
Lim
itST
ORE
EWS
VDA
S20
E53
M
321
1 2 33211 2 3
Capa
cito
r45
uF
1 5 2 6 3 7 4 8
4
231 6 4 5
2 3 1 2
32
1
54
32
1
CN6
5566
-08A
(MO
LEX)
CN4
B02P
-NV
CN3
B3P-
VH
CN1
B2P-
VH
CN2
B10B
-XA
SK-1
N-A
CN8
B03B
-XA
SK-1
N-A
CN7
B04B
-XA
SK-1
N-A
CN10
5566
-06A
(MO
LEX)
CN5
B04P
-NV
109
87
65
344
221 Brown
544
221
Yello
w
Yello
wA
NFT
58A
0
Trun
k LI
ne fo
r Mot
or /
Capa
cito
rVL
6P
AN
FT57
A0
Tran
sfor
mer
Sec
tion(
120V
)
AN
ET56
A0
Mot
or S
ectio
n (1
15V
)
AN
FT56
A0
AN
DR5
0A0
AN
DR3
6A0
Trun
k Li
ne fo
r Rel
ay
Sole
noid
Val
ve S
ectio
n(1
20V
)
1 2 3 4
Chai
r Dow
n Pr
ohib
ition
Chai
r Ope
ratio
n Pr
ohib
ition
CN4-
1(4-
2)IL
-4P-
S3EN
2(J
AE)
CN8
B04B
-XA
SK-1
12
34
CN3
B05B
-XA
SK-1
SW4
MO
DE
SW1
31
(Pro
port
iona
l Sol
enoi
d Va
lve)
67
89
122
12
34
5
1
Whi
te
Red
10
NM
F2P
N.O
.CO
M
1 3 2
CN11
5556
-04A
Mai
n Sw
itch
L2
N4
24
Up
Dow
n
Rais
e
Recl
inin
g
CN2-
1(2-
2)IL
-6P-
S3EN
2(J
AE)
1 2 3 4 5 6
CN19
B05B
-XA
SK-1
3
1 2 3 4
DN
CS42
A0
CN4-
1(4-
2)IL
-4P-
S3EN
2(J
AE)
N.O
.CO
M
Unu
sed
12
CN9
IL-2
P-S3
EN2
(JA
E)
109
87
65
344
221
1112
13
Arm
rest
Mem
bran
e SW
- L(A
GFS
14A
0)
109
87
65
344
221
1112
13
109
87
65
344
221
1112
13
109
87
65
344
221
1112
1310
98
76
53
4422
111
1213
Dis
trib
utio
n P.C
.B.
(AG
FS12
A0)
(Dis
trib
utio
n P.C
.B. in
the
right
sid
e ar
mre
st)
XA9P
XA2P
Touc
hpad
Sw
itch
(AG
FS16
A0)
To S
eria
l Com
mun
icat
ion
Line
for
Doc
tor T
able
D L
ight
PCB
.
4
Rota
tion
Lock
EWS
VDA
S20
E53
DN
CS42
A0
AN
FT71
A0
Mai
n Po
wer
Sw
itch
Sect
ion
AN
FT64
A0
Trun
k Li
ne fo
r Den
tal L
ight
Pow
er S
ourc
e
AN
FT65
A0
Trun
k Li
ne ff
or C
hair
Pow
er S
ourc
e
AN
FT63
A0
Eart
h
AN
FT55
B0
XA13
P
XA13
PXA
13P
XA13
PXA
13P
AN
FT60
A0
AN
FT61
A0
Trun
k Li
ne B
for M
embr
ane
SWA
NFT
62A
0
Trun
k Li
ne C
for M
embr
ane
SW
AN
FT67
A0
Seria
l Com
mun
icat
ion
Line
AN
FT68
A0
D L
ight
Com
mun
icat
ion
Line
AN
FT69
A0
Touc
hpad
SW
Trun
k Li
ne A
for
Safe
ty S
witc
h fo
r Und
er P
ump
Cove
r
(AG
FS15
A0)
3
122
344
56
78
910
XA10
P
AN
FT76
A0
Ass
ista
nt S
ide
Green
22
44
AN
FT70
A0
Pow
er C
able
Sec
tionSa
fety
Sw
itch
for U
nder
Pum
p Co
ver
Brow
nRe
d
Brow
nRe
d
Chai
r Dow
n Pr
ohib
ition
Chai
r Ope
ratio
n Pr
ohib
ition
Motor
Up
Down
Reclining
Raise
Up/Down Slow Slide
Backrest Slow Slide
Lock Cancel
RedOrangeYellowGreenBluePurpleGray
Black
RedOrangeYellowGreenBluePurpleGray
Pink
WhiteBlack
White/Red
Light Blue
AR
Up
Down
Reclining
Raise
P1
P2
LP
COM
5V
Lock Cancel
Safety
Safety LED.
COM
BrownRed
Seria
l Tra
nsm
issi
on
Seria
l Rec
eptio
n
Seria
l Tra
nsm
issi
on
Seria
l Rec
eptio
n
RedBrown
RedBrown
OrangeYellowGreenBluePurpleGray
Black
Touc
hpad
SW
Trun
k Li
ne B
for
AR
Up
Down
Reclining
Raise
P1
P2
LP
COM
RedBrown
OrangeYellowGreenBluePurpleGrayBlack
RedBrown
OrangeYellowGreenBluePurpleGray
Black
AR
Up
Down
Reclining
Raise
P1
P2
LP
COM
Touc
hpad
SW
Trun
k Li
ne C
for
BrownRedOrangeYellowGreenBluePurpleGrayWhiteBlackPinkLight BlueWhite/Red
BrownRedOrangeYellowGreenBluePurpleGrayWhiteBlackPinkLight BlueWhite/Red
BrownRedOrangeYellowGreenBluePurpleGrayWhiteBlackPinkLight BlueWhite/Red
AR
Up
Down
Reclining
Raise
P1
P2
LP
5V
Lock Cancel
Safety
Safety LED.
COM
AR
Up
Down
Reclining
Raise
P1
P2
LP
5V
Lock Cancel
Safety
Safety LED.
COM
Arm
rest
Mem
bran
e SW
- R
Yello
wYe
llow
Yello
w
Yello
wW
hite
Red
Yello
wYe
llow
Gra
y
Gra
y
Trun
k Li
ne A
for M
embr
ane
SW
BlueBlue
Light Blue
Blue
White
White
Black
Black
White
Blac
kBl
ack
Whi
te
Whi
teYe
llow
Whi
teYe
llow
Whi
teYe
llow
Trun
k Li
ne A
for S
olen
oid
Valv
e
Motor
Up
Down
Reclining
Raise
COM Com
Up/Down Slow Slide
Backrest Slow Slide
COM
COM
Lock Cancel
SV8
SV7
3 2
VL4P4 1
Yello
wW
hite
Yello
wW
hite
Gre
enBl
ue
Yello
wO
rang
e
Up/
Dow
n Sl
ow S
lide
Back
rest
Slo
w S
lide
Back
rest
Slo
w
Up/
Dow
n Sl
ow
1 2
XA2P
Ther
mis
tor
NTC
The
rmis
tor
Red
Brow
nBl
ack
Blac
kTh
erm
isto
r for
Sol
enoi
d Va
lves
Chai
r upp
er fr
ame
Eart
h Li
ne Base
Whi
te
-18-
7-2. UNIT SECTION
12
1 2 3 4
1 2
HP
Hol
der S
W P
CB. F
ourt
h Li
ne
4 3 2 1
21G
ray
Purp
le
14142424
(2.3
)
(1)
CN36
B03B
-XA
SK-1
N-A
321
Tim
erSp
ray
F(Fu
nctio
n)ST
ORE
LED
MA
IN(G
reen
)LO
CK(O
rang
e)Li
ght P
ack(
Gre
en)
Spra
y A
ir(Bl
ue)
ãtì](
û?)
ê?ì](
óŒ)
Operation Switch
-+SE
LECT
Ligh
t Pac
kD
enta
l Lig
htê?
ãtêÿ
ë÷
Chair Operation Switch
LPP2P1AR
Rais
eD
own
UP
Trun
k Li
ne fo
r Dis
play
AN
FT93
A0
NO
COM
Min
i Air
Switc
h(F/
C)
AN
FT97
A0
AN
FT95
A0
AN
FT96
A0
AN
FT83
A0
AN
FT80
A0
AN
FT79
A0
AN
FT87
A0
AN
FT86
A0
AN
FT82
A0
AN
FT85
A0
Tran
sfor
mer
Sec
tion
AN
FT92
A0
Trun
k Li
ne fo
r Den
tal L
ight
Pow
er S
ourc
e (J
BOX)
AN
FT81
A0
21
32
13
21
3
Pow
er C
able
(UL)
JBO
X
AN
FT84
A0
Seria
l Com
mun
icat
ion
Line
AN
FT78
A0
Trun
k Li
ne fo
r HP
Hol
der
AN
FT77
A0
AGFS
17A
0
AGFS
20A
0
Air
SW
Fuse
F 10
A
120V
14V 24V
Tran
sfor
mer
FK-
023
Junc
tion
Box
Sect
ion
AC IN
1 2211 221
21
1 2 3
12
12
34
21
21
34
56
71
2
87654321 4321
8 7 6 5 4 3 2 1 87654321
1 2 3 4 5 6 7 8 1 2 3 4 1 21 221 1 221
1 2 3 4
43211 2
CN31
IL-2
P-S3
EN2
(JA
E)
21
CN3
IL-3
P-S3
EN2
(JA
E)
Seria
l Com
mun
icat
ion
Line
Se
rial T
rans
mis
sion
(1.2
)
Seria
l Rec
eptio
n(3
.4)
CN29
236-
747(
Gre
en)
(WAG
O)
CN26
IL-3
P-S3
EN2
(JA
E)PE
RIO
END
O
CN16
IL-4
P-S3
EN2
(JA
E)
Wat
er
Doc
tor M
embr
ane
SW(A
GFS
18A
0)
12
34
56
78
910
12
344
665
Uni
t P.C
.B.
(AG
FS22
**)
CN21
IL-4
P-S3
EN2
(JA
E)
POW
ER
CN2
IL-2
P-S3
EN2
(JA
E)CN4
IL-2
P-S3
EN2
(JA
E)
CN6
IL-2
P-S3
EN2
(JA
E)
CN7
IL-3
P-S3
EN2
(JA
E)
CN10
IL-2
P-S3
EN2
(JA
E)
CN12
IL-2
P-S3
EN2
(JA
E)
CN17
IL-2
P-S3
EN2
(JA
E)
Air
SW
CN24
IL-4
P-S3
EN2
(JA
E)
CN25
IL-4
P-S3
EN2
(JA
E)
LP1
LP2
LP3
LP4
CN27
236
-401
(Gra
y)W
AGO
12
34
56
78
910
12
344
665
CN14
IL-2
P-S3
EN2
(JA
E)
Air
Red
Blac
kRe
dBl
ack
CN30
IL-2
P-S3
EN2
(JA
E)
Dis
play
(AG
FS19
A0)
12
34
56
78
910
81
210
313
614
7
CN5
IL-1
0P-S
3EN
2(J
AE)
CN18
IL-2
P-S3
CN2
(JA
E)W
hite
Blac
k
HP1
SV
HP2
SV
HP3
SV
HP4
SV
CN8
IL-8
P-S3
EN2
(JA
E)
HP1HP2
HP4
12V
0V0VHP3
CN13
IL-8
P-S3
EN2
(JA
E)
CN1
IL-3
P-S3
EN2
(JA
E)
CN36
236
-401
(Gra
y)W
AGO
CN37
236
-401
(Gra
y)W
AGO
CN38
236
-401
(Gra
y)W
AGO
CN23
IL-4
P-S3
EN2
(JA
E)
CN11
IL-8
P-S3
EN2
(JA
E)
FLA
SH
CN28
IL-7
P-S3
EN2
(JA
E)
CN19
IL-2
P-S3
EN2
(JA
E)
CN9
IL-4
P-S3
EN2
(JA
E)
GN
DCN
22IL
-2P-
S3EN
2(J
AE)
HP4
HP3
HP2
HP1
CN32
IL-2
P-S3
EN2
(JA
E)CN
33IL
-2P-
S3EN
2(J
AE)
CN34
IL-2
P-S3
EN2
(JA
E)CN
35IL
-2P-
S3EN
2(J
AE)
CN20
IL-1
0P-S
3EN
2(J
AE)
CN15
IL-6
P-S3
EN2
(JA
E)
XA-1
3PXA
-6P
AN
FT98
A0
Trun
k Li
ne A
for D
r Mem
bran
e SW
AN
FT99
A0
HP3HP2HP1
HP4
---
+++
-+
8
Sole
noid
Val
ve fo
r Pilo
t Air
Sole
noid
Val
ve fo
r Pilo
tA
ND
T08A
0
-+
-+
25
Sole
noid
Val
ve fo
r Spr
ayA
ND
T07A
0
1 2
12
1 2 3 4
12
12
1 21 21 2H
P H
olde
r SW
PCB
.
L N
Blac
k
VL4P
VL2P
Pow
er S
ourc
e fo
r Den
tal L
ight
AGD
R39A
0
Brow
n
Red
XA2P
XA2P
Gre
enBl
ue
Ora
nge
Yello
w
Brow
nRe
d
Stan
dard
Spe
c.
Syrin
ge T
urbi
ne x
3
Brow
n
AN
FU04
A0
Mai
n A
ir SW
Sec
tion
AGFL
34A
0
AN
FU05
A0
Eart
h (D
r Tab
le)
CN1
PS-L
A-1
4P
45
911
12
CN1
12
1314
Brow
nRe
dO
rang
eG
reen
Yello
w
Blue
Gra
yPu
rple
Whi
te
Dynamic Output 1Dynamic Output 2Dynamic Output 3Dynamic LED 1Dynamic LED 2Dynamic LED 3
Red
BlackWhiteGray
BrownRed
OrangeYellowGreen
Blue
PurpleBlue
GreenYellow
Orange
Brown
Trun
k line
B fo
r 24V
.12V
Pow
er S
ourc
e
Purple
Blue
PurpleBlue
Purple
Blue
Black
Red
Mai
n G
auge
Cla
mp
Red
Blue
Red
Blac
k
Purp
leBl
ue
Trun
k line
A fo
r 24V
.12V
Pow
er S
ourc
e
Fuse
6A
Fuse
2A
Trun
k line
(Blue
) for
Fus
e Ho
lder
Trun
k line
(Red
) for
Fus
e Ho
lder
Gre
en/Y
ello
w
Trun
k line
(Blac
k) fo
r Fus
e Ho
lder
Chair
Pow
er S
upply
Sec
tion
Chair
Pow
er S
upply
Bl
ack
Blac
kBl
ack
Whi
teW
hite
Whi
teG
reen
Blac
kBl
ack
Gra
yW
hite
Whi
te
Whi
te
Gre
en/Y
ello
w
Earth
Purp
le
Purp
leO
rang
e
Ora
nge
Red Blac
kBl
uePu
rple
J-Bo
x G
auge
Cla
mp
Sect
ion
To C
hair
PCB.
(CH
)
Red
Blac
kRe
dBl
ack
Red
Blac
kRe
dBl
ack
HP
Hol
der S
W P
CB.
HP
Hol
der S
W P
CB.
Seria
l Tra
nsm
issi
on
Seria
l Rec
eptio
n
Brow
nBr
own
Brow
nBr
own
Brow
n
BrownRed
OrangeYellowGreen
Blue
WhiteGray
Purple
0V5VVFD-RS
Low
erSp
ray
Wat
er(B
lue)
Trun
k Li
ne B
for D
r Mem
bran
e SW
VFD-R/WVFD-E
VFD-1
VFD-3VFD-2
VFD-0
AN
FU69
A0
Trun
k line
(Blac
k) fo
r Fus
e Ho
lder
HP2
HP1
HP3
Input AInput BInput CInput DInput EInput FInput GInput HDisplay LED.Display LED.
Maximus
ET
L L
IST
ED
CU
SCO
NFOR
MS
TOUL
STD
73CE
RTIF
IED
TOCA
N/CS
A ST
D C2
2.2
NO. 6
8
Maximus
3058278
ETL CLASSIFIED
MEDICAL ANDDENTAL EQUIPMENT
CONFORMS TO UL STD 60601-1
8. LOCATION OF PRODUCT LABELS
(1) Product label & ETL label are located under the doctor table
(2) Product label (SRRIAL No.) is located side of the assistant swing arm.
-19-
(3) Product label (SRRIAL No.) is located side of the assistant swing arm.
NOTE
BELMONT EQUIPMENT, Division of Takara Belmont USA, Inc.101 Belmont Drive Somerset, New Jersey 08873 U.S.A. TEL.:(732) 469-5000 / (800) 223-1192 Fax.:(732)526-6322 / (800) 280-7504
TAKARA CO, CANADA LTD.2076 S. Sheridan Way, Mississauga, Ont., L5J2M4, Can. TEL.:(905) 822-2755 Fax.:(905)822-6203