Rebekah Livesley, 2008
Analysis of Vitamin C in Fruit Juices
InstrumentMetrohm Polarograph
![Page 2: Rebekah Livesley, 2008 Analysis of Vitamin C in Fruit Juices Instrument Metrohm Polarograph](https://reader035.vdocument.in/reader035/viewer/2022062516/56649d8d5503460f94a7575a/html5/thumbnails/2.jpg)
Vitamin C (Ascorbic Acid)
formula weight 176.13g
Water Soluble
melting point 190-192° C,
unstable in solutions containing oxygen as it is readily oxidized
![Page 3: Rebekah Livesley, 2008 Analysis of Vitamin C in Fruit Juices Instrument Metrohm Polarograph](https://reader035.vdocument.in/reader035/viewer/2022062516/56649d8d5503460f94a7575a/html5/thumbnails/3.jpg)
Why is Vitamin C important? It is involved in over 300 biological processes
in the body for example Antioxidant Collagen manufacture & neurotransmitters Immune function
It is a Nutrient not synthesized by the body Complete deficiency results in scurvy Deficiency symptoms
Bleeding gums Joint pain Bruising & wounds that don’t heal Fatigue & muscular weaknesses
![Page 4: Rebekah Livesley, 2008 Analysis of Vitamin C in Fruit Juices Instrument Metrohm Polarograph](https://reader035.vdocument.in/reader035/viewer/2022062516/56649d8d5503460f94a7575a/html5/thumbnails/4.jpg)
Recommended Daily intake is 30-60mg
The RDI is what you need to consume to prevent a deficiency
100ml (which is less than ½ a cup) of the Popper Juice orange Juice is all you would need to consume
excess vitamin C is usually eliminated through the urine
An excess of Vitamin C can cause diarrhea as water is pulled into the colon from the circulatory system
![Page 5: Rebekah Livesley, 2008 Analysis of Vitamin C in Fruit Juices Instrument Metrohm Polarograph](https://reader035.vdocument.in/reader035/viewer/2022062516/56649d8d5503460f94a7575a/html5/thumbnails/5.jpg)
Fruit JuicesExpected concentrations of Vitamin C in Fruit Juices
0
100
200
300
400
500
600
700
McCance andWiddoson’s TheComposition of
Foods (7)
Berlitz, Grosh,Scheiberle Food
Chemistry (8)
StaplesDrinks (9)
Exp
ecte
d c
on
cen
trat
ion
Apple Juice
Orange Juice
![Page 6: Rebekah Livesley, 2008 Analysis of Vitamin C in Fruit Juices Instrument Metrohm Polarograph](https://reader035.vdocument.in/reader035/viewer/2022062516/56649d8d5503460f94a7575a/html5/thumbnails/6.jpg)
Samples
Popper Juice Apple and Blackcurrant
Popper Juice Orange Juice
Popper Juice Apple Juice
Popper Juice Apple And Mango Juice
WESTCLIFF Apple Juice
And V8 Vegetable Juice
![Page 7: Rebekah Livesley, 2008 Analysis of Vitamin C in Fruit Juices Instrument Metrohm Polarograph](https://reader035.vdocument.in/reader035/viewer/2022062516/56649d8d5503460f94a7575a/html5/thumbnails/7.jpg)
Metrohm Polarograph
Photographed by Shay
![Page 8: Rebekah Livesley, 2008 Analysis of Vitamin C in Fruit Juices Instrument Metrohm Polarograph](https://reader035.vdocument.in/reader035/viewer/2022062516/56649d8d5503460f94a7575a/html5/thumbnails/8.jpg)
ProceduresDifferential Pulse Polarography
Standard Addition Method Dilute sample accurately to approximate concentration of
40mg/L
Place in sample cell with water and buffer solution
Run scan -0.4 to +0.1 V
3 standard additions
Normal Calibration Method Prepare 0, 5, 10, 25mg/L standard solutions in buffer
solution Dilute sample accurately to an approximate concentration
of 10mg/L in buffer Runs scans as above
![Page 9: Rebekah Livesley, 2008 Analysis of Vitamin C in Fruit Juices Instrument Metrohm Polarograph](https://reader035.vdocument.in/reader035/viewer/2022062516/56649d8d5503460f94a7575a/html5/thumbnails/9.jpg)
Differential Pulse Polarography
Standard Addition Results
Sample WESTCLIFF Apple Juice
200mg/L standard Solution of Ascorbic acid
Average concentration
263 mg/L 235 mg/L
Recovery Check 174.1 % recovery
Standard Deviation 68% 51%
Relative Error 17%
![Page 10: Rebekah Livesley, 2008 Analysis of Vitamin C in Fruit Juices Instrument Metrohm Polarograph](https://reader035.vdocument.in/reader035/viewer/2022062516/56649d8d5503460f94a7575a/html5/thumbnails/10.jpg)
Differential Pulse Polarography
Normal Calibration Results
Sample Average concentration
(mg/L)
Recovery Check
(% recovery)
Standard Deviation
WESTCLIFF Apple Juice 273 117 6%
JUST JUICE Apple Juice 174 91 -
POPPER JUICE – Orange Juice 811 108 -
POPPER JUICE – Apple and Blackcurrant Juice
633 - -
POPPER JUICE – Apple and Mango Juice
731 - -
POPPER JUICE – Apple Juice 662 - -
V8 – Vegetable Juice 332 155 -
![Page 11: Rebekah Livesley, 2008 Analysis of Vitamin C in Fruit Juices Instrument Metrohm Polarograph](https://reader035.vdocument.in/reader035/viewer/2022062516/56649d8d5503460f94a7575a/html5/thumbnails/11.jpg)
Com parisson of Results to the concentration of vitam in c specified in the Nutritional Inform ation
0
100
200
300
400
500
600
700
800
900
Co
nc
en
tra
tio
n m
g/L
average concentration mg/L nutritional information concentration mg/L
![Page 12: Rebekah Livesley, 2008 Analysis of Vitamin C in Fruit Juices Instrument Metrohm Polarograph](https://reader035.vdocument.in/reader035/viewer/2022062516/56649d8d5503460f94a7575a/html5/thumbnails/12.jpg)
Concentration Range of Vitamin C
To Find the Concentration Range for the Metrohm Polarograph
(values from 0.1 - 100mg/L)
y = 29.61x - 66.298
R2 = 0.9865
-1000
0
1000
2000
3000
4000
0 20 40 60 80 100 120
Concentration (mg/L)
Pea
k H
eig
ht
(nA
)
![Page 13: Rebekah Livesley, 2008 Analysis of Vitamin C in Fruit Juices Instrument Metrohm Polarograph](https://reader035.vdocument.in/reader035/viewer/2022062516/56649d8d5503460f94a7575a/html5/thumbnails/13.jpg)
To Show Rate of Oxidation of 20 times diluted westcliff apple juice samples
185
190
195
200
205
210
0 10 20 30 40 50 60 70
Time (minutes)
Con
cent
ratio
n (m
g/L)
![Page 14: Rebekah Livesley, 2008 Analysis of Vitamin C in Fruit Juices Instrument Metrohm Polarograph](https://reader035.vdocument.in/reader035/viewer/2022062516/56649d8d5503460f94a7575a/html5/thumbnails/14.jpg)
Chart of the concentration of WESTCLIFF apple juice in an open popper as a funtion of time
266268270272274276278280
0 20 40 60 80 100 120
time (minutes)
Conc
entra
tion
(mg/
L)
![Page 15: Rebekah Livesley, 2008 Analysis of Vitamin C in Fruit Juices Instrument Metrohm Polarograph](https://reader035.vdocument.in/reader035/viewer/2022062516/56649d8d5503460f94a7575a/html5/thumbnails/15.jpg)
Choice of method
normal calibration simpler for sample delivery (no pipetting required) more consistent results better RC values
![Page 16: Rebekah Livesley, 2008 Analysis of Vitamin C in Fruit Juices Instrument Metrohm Polarograph](https://reader035.vdocument.in/reader035/viewer/2022062516/56649d8d5503460f94a7575a/html5/thumbnails/16.jpg)
What went wrong
Standard Addition Method
Lack of precision in the results obtained Caused by
Micropipettes Human error 3 aliquots to be transferred to sample cell per run
![Page 17: Rebekah Livesley, 2008 Analysis of Vitamin C in Fruit Juices Instrument Metrohm Polarograph](https://reader035.vdocument.in/reader035/viewer/2022062516/56649d8d5503460f94a7575a/html5/thumbnails/17.jpg)
What went wrong
Normal Calibration Method An aliquot of the V8 vegetable juice could not be taken by
pipette as the juice was so viscous Had to weigh a mass and do a density calculation before
diluting the juice JUST JUICE – Apple Juice concentration was outside the
concentration range of the standards used in the calibration graph
Easily fixed – just recalibrate
![Page 18: Rebekah Livesley, 2008 Analysis of Vitamin C in Fruit Juices Instrument Metrohm Polarograph](https://reader035.vdocument.in/reader035/viewer/2022062516/56649d8d5503460f94a7575a/html5/thumbnails/18.jpg)
Polarography
Disadvantages
Not time efficient
Must use a buffer solution
![Page 19: Rebekah Livesley, 2008 Analysis of Vitamin C in Fruit Juices Instrument Metrohm Polarograph](https://reader035.vdocument.in/reader035/viewer/2022062516/56649d8d5503460f94a7575a/html5/thumbnails/19.jpg)
Conclusion The normal Calibration polarographic method for the
analysis of vitamin C is better than the standard addition method
All sample analyzed had considerably more vitamin C than specified on the packet except Just Juice apple juice which had about a 1/5