hyperion colorimeter
specification
2
Contents 1 Hyperion .............................................................................................. 3
2 Highlights ............................................................................................. 3
3 General specifications ......................................................................... 4
4 Typical spectral sensitivity ................................................................... 4
5 10mm measurement specifications .................................................... 5
6 5mm improved measurement specifications ....................................... 6
7 10mm improved measurement specifications ..................................... 7
8 10mm WA measurement specifications .............................................. 8
9 20mm measurement specifications .................................................... 9
10 20mm improved measurement specifications ............................... 10
11 10mm measurement data ............................................................. 11
12 10mm improved measurement data ............................................. 13
13 20mm measurement data ............................................................. 15
14 20mm improved & 10mm WA measurement data ........................ 17
15 Xtreme low luminance performance data ..................................... 19
15.1 Hyperion 10mm ......................................................................... 19
15.2 Hyperion 10mm improved ......................................................... 19
16 Hyperion 10mm dimensions .......................................................... 20
17 Hyperion 10mm & 5mm improved dimensions ............................. 21
18 Hyperion 10mm WA dimensions ................................................... 22
19 Hyperion 20mm dimensions .......................................................... 23
20 Hyperion 20mm improved dimensions .......................................... 24
21 Hyperion M8 fibre dimensions ....................................................... 25
3
1 Hyperion The Hyperion colorimeter offers a unique combination of high speed and
accurate colour measurement capabilities packed in a robust package. In
comparison to its predecessors, the Hyperion has a significant improvement
on filter characteristics and an incredible speed upgrade. It is actually four
times faster than previous models making accurate colour measurements
possible at 0.3cd/m2 in only 50ms.
The Hyperion colorimeter is available with a 10mm spot size. A fiber version
is also available with several optics, custom optics can be applied on
request. Added to the filter characteristics the high sensitivity, ultra-low
noise electronics and a huge dynamic range make it the ideal device for
display measurements even at low levels.
2 Highlights Highly accurate colour measurement according to human eye
(CIE1931)
Fast colour measurement even at low luminance level
Flicker luminance (Y) function: 2000 samples/second
Auto-range function
Powerful MCU enables internal JEITA flicker calculation
Mechanical shutter
USBMTC standard compliant
Windows, Linux and MAC OSX compatible
Directly supported in Labview, LabWindows, Visual Studio via
VISA library
4
3 General specifications Interfaces
USB 2.0 USBMTC compliant, SCPI command set, high speed device
RS 232 For PC and embedded purposes, using same command set as USB
Trigger in & out 5V compliant
Power ratings
Min voltage
Typical voltage
Max voltage
Max current
USB power 4.75V 5.00V 5.25V 300mA
9V power 8.75V 9.00V 9.25V 300mA
Measurement system
Photo detector 3 silicon photo diode using XYZ interference filter
Spectral response Approximates CIE 1931 colour matching functions
Measurement parameters
XYZ, Yxy, Yuv, correlated colour temperature (CCT), dominant wavelength DWL, Flicker, Response time
Size (HxWxD) 53x70x147 mm (without lens system)
Weight 670 gram
Mounting 12 M3 threat holes spread over four sides
4 Typical spectral sensitivity
5
5 10mm measurement specifications
Measurement system
Optical system Acceptance angle 5º (+/- 2.6)
Measurement spot size 14 mm at 50 mm distance 19 mm at 100 mm distance
Flicker measurement speed (sample mode)
Luminance 2000 samples / second, XYZ 2000 samples / second Correct detected frequency of at least 1 Hz
Colour measurement speed Colour 22 ms or higher, depending on luminance level (including communication) 150 cd/m² with DC level light at 16 ms. PWM requires longer integration (multiple frames)
Sample mode signal frequency response
Parameter F3dba
Gain 1 DC – 500 Hz
Gain 2 DC – 500 Hz
Gain 3 DC – 500 Hz
Colorimeter specification
Parameter Range Accuracy Repeatability
Resolution 16 bit for X, Y and Z >80 dB without averaging
Luminance (Y) 0.005 cd/m² - 20,000 cd/m² integration time between 0.5ms – 1s
+/-2% of measured value, measured at white image of LCD display, Luminance of app. 150 cd/m², x=0.300 y= 0.325
Y +/- 0.5% at 0.1 cd/m² ¹ Y +/- 0.2% at 1 cd/m² ¹ Y +/- 0.15% at 5 cd/m² ¹ Y +/- 0.1% at 150 cd/m² ¹
Chromaticity (x,y) +/- 0.001 after calibration, measured at white image of LCD display, Luminance of app. 150 cd/m², x=0.300 y= 0.325
x,y +/- 0.001 for Y at 0.1 cd/m² ¹ x,y +/- 0.0002 for Y at 1 cd/m² ¹ x,y +/- 0.0002 for Y at 5 cd/m² ¹ x,y +/- 0.0002 for Y at 150 cd/m² ¹
Measurement speed 4-10 samples / s for Y at 0.1 cd/m² ¹ 10-20 samples / s for Y at 1 cd/m² ¹ 40 samples / s for Y at 5 cd/m² ¹ 40 samples / s for Y at 150 cd/m² ¹
Flicker (contrast method) 5 cd/m² or higher +/- 0.3% flicker frequency 30Hz AC/DC 10% sine wave. Sine wave at 10 cd/m²
+/- 0.2%
Flicker (JEITA method) 5 cd/m² or higher +/- 0.3dB flicker frequency 30Hz AC/DC 10% sine wave. Sine wave at 10 cd/m²
+/- 0.2dB
Operating temperature 10-35°C ²
Shutter lifetime >1000000
Shutter speed 250-300 ms depending on temperature and lifetime
1 All measurements are performed 20 times on a LED display with sufficient signal noise ratio, value is based on 2 sigma. Sample speed depends on the measured sample.
If the sample uses PWM, it will take longer. Use of lower rated values is strongly recommended to ensure repeatability.
2 Operating temperature reaches from 0 to 40 degrees. Dark level compensation is optimized for operating in temperatures between 10 and 35 degrees.
Other temperature ranges can be calibrated using the mechanical shutter if necessary.
6
6 5mm improved measurement specifications
Measurement system
Optical system Acceptance angle 5º (+/- 2.5)
Measurement spot size 9.6 mm at 50 mm distance 14 mm at 100mm distance 6.8 mm at 20 mm distance 5.2 mm at 1 mm distance
Flicker measurement speed (sample mode)
Luminance 2000 samples / second, XYZ 2000 samples / second Correct detected frequency of at least 1 Hz
Colour measurement speed Colour 22 ms or higher, depending on luminance level (including communication) 150 cd/m² with DC level light at 16 ms. PWM requires longer integration (multiple frames)
Sample mode signal frequency response
Parameter F3dba
Gain 1 DC – 500 Hz
Gain 2 DC – 500 Hz
Gain 3 DC – 500 Hz
Colorimeter specification
Parameter Range Accuracy Repeatability
Resolution 16 bit for X, Y and Z >80 dB without averaging
Luminance (Y) 0.012 cd/m² - 48,000 cd/m² integration time between 0.5ms – 1s
+/-2% of measured value, measured at white image of LCD display, Luminance of app. 150 cd/m², x=0.300 y= 0.325
Y +/- 0.3% at 0.4 cd/m² ¹ Y +/- 0.15% at 4 cd/m² ¹ Y +/- 0.1% at 20 cd/m² ¹ Y +/- 0.1% at 150 cd/m² ¹
Chromaticity (x,y) +/- 0.001 after calibration, measured at white image of LCD display, Luminance of app. 150 cd/m², x=0.300 y= 0.325
x,y +/- 0.00075 for Y at 0.4 cd/m² ¹ x,y +/- 0.0003 for Y at 4 cd/m² ¹ x,y +/- 0.0003 for Y at 20 cd/m² ¹ x,y +/- 0.0002 for Y at 150 cd/m² ¹
Measurement speed 4-10 samples / s for Y at 0.4 cd/m² ¹ 10-20 samples / s for Y at 4 cd/m² ¹ 40 samples / s for Y at 20 cd/m² ¹ 40 samples / s for Y at 150 cd/m² ¹
Flicker (contrast method) 15 cd/m² or higher +/- 0.3% flicker frequency 30Hz AC/DC 10% sine wave. Sine wave at 10 cd/m²
+/- 0.2%
Flicker (JEITA method) 15 cd/m² or higher +/- 0.3dB flicker frequency 30Hz AC/DC 10% sine wave. Sine wave at 10 cd/m²
+/- 0.2dB
Operating temperature 10-35°C ²
Shutter lifetime >1000000
Shutter speed 250-300 ms depending on temperature and lifetime
1 All measurements are performed 20 times on a LED display with sufficient signal noise ratio, value is based on 2 sigma. Sample speed depends on the measured sample.
If the sample uses PWM, it will take longer. Use of lower rated values is strongly recommended to ensure repeatability.
2 Operating temperature reaches from 0 to 40 degrees. Dark level compensation is optimized for operating in temperatures between 10 and 35 degrees.
Other temperature ranges can be calibrated using the mechanical shutter if necessary.
7
7 10mm improved measurement specifications
Measurement system
Optical system Acceptance angle 5º (+/- 2.5)
Measurement spot size 14.4 mm at 50 mm distance 18.7 mm at 100 mm distance 10.2 mm at 1 mm distance
Flicker measurement speed (sample mode)
Luminance 2000 samples / second, XYZ 2000 samples / second Correct detected frequency of at least 1 Hz
Colour measurement speed Colour 22 ms or higher, depending on luminance level (including communication) 150 cd/m² with DC level light at 16 ms. PWM requires longer integration (multiple frames)
Sample mode signal frequency response
Parameter F3dba
Gain 1 DC – 500 Hz
Gain 2 DC – 500 Hz
Gain 3 DC – 500 Hz
Colorimeter specification
Parameter Range Accuracy Repeatability
Resolution 16 bit for X, Y and Z >80 dB without averaging
Luminance (Y) 0.003 cd/m² - 12,000 cd/m² integration time between 0.5ms – 1s
+/-2% of measured value, measured at white image of LCD display, Luminance of app. 150 cd/m², x=0.300 y= 0.325
Y +/- 0.3% at 0.1 cd/m² ¹ Y +/- 0.15% at 1 cd/m² ¹ Y +/- 0.1% at 5 cd/m² ¹ Y +/- 0.1% at 150 cd/m² ¹
Chromaticity (x,y) +/- 0.001 after calibration, measured at white image of LCD display, Luminance of app. 150 cd/m², x=0.300 y= 0.325
x,y +/- 0.00075 for Y at 0.1 cd/m² ¹ x,y +/- 0.0003 for Y at 1 cd/m² ¹ x,y +/- 0.0003 for Y at 5 cd/m² ¹ x,y +/- 0.0002 for Y at 150 cd/m² ¹
Measurement speed 4-10 samples / s for Y at 0.1 cd/m² ¹ 10-20 samples / s for Y at 1 cd/m² ¹ 40 samples / s for Y at 5 cd/m² ¹ 40 samples / s for Y at 150 cd/m² ¹
Flicker (contrast method) 5 cd/m² or higher +/- 0.3% flicker frequency 30Hz AC/DC 10% sine wave. Sine wave at 10 cd/m²
+/- 0.2%
Flicker (JEITA method) 5 cd/m² or higher +/- 0.3dB flicker frequency 30Hz AC/DC 10% sine wave. Sine wave at 10 cd/m²
+/- 0.2dB
Operating temperature 10-35°C ²
Shutter lifetime >1000000
Shutter speed 250-300 ms depending on temperature and lifetime
1 All measurements are performed 20 times on a LED display with sufficient signal noise ratio, value is based on 2 sigma. Sample speed depends on the measured sample.
If the sample uses PWM, it will take longer. Use of lower rated values is strongly recommended to ensure repeatability.
2 Operating temperature reaches from 0 to 40 degrees. Dark level compensation is optimized for operating in temperatures between 10 and 35 degrees.
Other temperature ranges can be calibrated using the mechanical shutter if necessary.
8
8 10mm WA measurement specifications
Measurement system
Optical system Acceptance angle 10º (+/- 5)
Measurement spot size 17.4 mm at 50 mm distance 26.1 mm at 100 mm distance
Flicker measurement speed (sample mode)
Luminance 2000 samples / second, XYZ 2000 samples / second Correct detected frequency of at least 1 Hz
Colour measurement speed Colour 22 ms or higher, depending on luminance level (including communication) 150 cd/m² with DC level light at 16 ms. PWM requires longer integration (multiple frames)
Sample mode signal frequency response
Parameter F3dba
Gain 1 DC – 500 Hz
Gain 2 DC – 500 Hz
Gain 3 DC – 500 Hz
Colorimeter specification
Parameter Range Accuracy Repeatability
Resolution 16 bit for X, Y and Z >80 dB without averaging
Luminance (Y) 0.0004 cd/m² - 4,000 cd/m² integration time between 0.5ms – 1s
+/-2% of measured value, measured at white image of LCD display, Luminance of app. 150 cd/m², x=0.300 y= 0.325
Y +/- 0.1% at 0.1 cd/m² ¹ Y +/- 0.1% at 1 cd/m² ¹ Y +/- 0.1% at 5 cd/m² ¹ Y +/- 0.1% at 150 cd/m² ¹
Chromaticity (x,y) +/- 0.001 after calibration, measured at white image of LCD display, Luminance of app. 150 cd/m², x=0.300 y= 0.325
x,y +/- 0.0005 for Y at 0.1 cd/m² ¹ x,y +/- 0.00025 for Y at 1 cd/m² ¹ x,y +/- 0.0002 for Y at 5 cd/m² ¹ x,y +/- 0.0002 for Y at 150 cd/m² ¹
Measurement speed 4-10 samples / s for Y at 0.1 cd/m² ¹ 10-20 samples / s for Y at 1 cd/m² ¹ 40 samples / s for Y at 5 cd/m² ¹ 40 samples / s for Y at 150 cd/m² ¹
Flicker (contrast method) 4 cd/m² or higher +/- 0.3% flicker frequency 30Hz AC/DC 10% sine wave. Sine wave at 10 cd/m²
+/- 0.2%
Flicker (JEITA method) 4 cd/m² or higher +/- 0.3dB flicker frequency 30Hz AC/DC 10% sine wave. Sine wave at 10 cd/m²
+/- 0.2dB
Operating temperature 10-35°C ²
Shutter lifetime >1000000
Shutter speed 250-300 ms depending on temperature and lifetime
1 All measurements are performed 20 times on a LED display with sufficient signal noise ratio, value is based on 2 sigma. Sample speed depends on the measured sample.
If the sample uses PWM, it will take longer. Use of lower rated values is strongly recommended to ensure repeatability.
2 Operating temperature reaches from 0 to 40 degrees. Dark level compensation is optimized for operating in temperatures between 10 and 35 degrees.
Other temperature ranges can be calibrated using the mechanical shutter if necessary.
9
9 20mm measurement specifications
Measurement system
Optical system Acceptance angle 3.5º (+/- 1.75)
Measurement spot size 25 mm at 50 mm distance 28 mm at 100 mm distance
Flicker measurement speed (sample mode)
Luminance 2000 samples / second, XYZ 2000 samples / second Correct detected frequency of at least 1 Hz
Colour measurement speed Colour 22 ms or higher, depending on luminance level (including communication) 150 cd/m² with DC level light at 16 ms. PWM requires longer integration (multiple frames)
Sample mode signal frequency response
Parameter F3dba
Gain 1 DC – 500 Hz
Gain 2 DC – 500 Hz
Gain 3 DC – 500 Hz
Colorimeter specification
Parameter Range Accuracy Repeatability
Resolution 16 bit for X, Y and Z >80 dB without averaging
Luminance (Y) 0.0005 cd/m² - 5,000 cd/m² integration time between 0.5ms – 1s
+/-2% of measured value, measured at white image of LCD display, Luminance of app. 150 cd/m², x=0.300 y= 0.325
Y +/- 0.1% at 0.1 cd/m² ¹ Y +/- 0.1% at 1 cd/m² ¹ Y +/- 0.1% at 5 cd/m² ¹ Y +/- 0.1% at 150 cd/m² ¹
Chromaticity (x,y) +/- 0.001 after calibration, measured at white image of LCD display, Luminance of app. 150 cd/m², x=0.300 y= 0.325
x,y +/- 0.0005 for Y at 0.1 cd/m² ¹ x,y +/- 0.00025 for Y at 1 cd/m² ¹ x,y +/- 0.0002 for Y at 5 cd/m² ¹ x,y +/- 0.0002 for Y at 150 cd/m² ¹
Measurement speed 4-10 samples / s for Y at 0.1 cd/m² ¹ 10-20 samples / s for Y at 1 cd/m² ¹ 40 samples / s for Y at 5 cd/m² ¹ 40 samples / s for Y at 150 cd/m² ¹
Flicker (contrast method) 5 cd/m² or higher +/- 0.3% flicker frequency 30Hz AC/DC 10% sine wave. Sine wave at 10 cd/m²
+/- 0.2%
Flicker (JEITA method) 5 cd/m² or higher +/- 0.3dB flicker frequency 30Hz AC/DC 10% sine wave. Sine wave at 10 cd/m²
+/- 0.2dB
Operating temperature 10-35°C ²
Shutter lifetime >1000000
Shutter speed 250-300 ms depending on temperature and lifetime
1 All measurements are performed 20 times on a LED display with sufficient signal noise ratio, value is based on 2 sigma. Sample speed depends on the measured sample.
If the sample uses PWM, it will take longer. Use of lower rated values is strongly recommended to ensure repeatability.
2 Operating temperature reaches from 0 to 40 degrees. Dark level compensation is optimized for operating in temperatures between 10 and 35 degrees.
Other temperature ranges can be calibrated using the mechanical shutter if necessary.
10
10 20mm improved measurement specifications
Measurement system
Optical system Acceptance angle 5º (+/- 2.5)
Measurement spot size 22.1 mm at 50 mm distance 26.7 mm at 100 mm distance
Flicker measurement speed (sample mode)
Luminance 2000 samples / second, XYZ 2000 samples / second Correct detected frequency of at least 1 Hz
Colour measurement speed Colour 22 ms or higher, depending on luminance level (including communication) 150 cd/m² with DC level light at 16 ms. PWM requires longer integration (multiple frames)
Sample mode signal frequency response
Parameter F3dba
Gain 1 DC – 500 Hz
Gain 2 DC – 500 Hz
Gain 3 DC – 500 Hz
Colorimeter specification
Parameter Range Accuracy Repeatability
Resolution 16 bit for X, Y and Z >80 dB without averaging
Luminance (Y) 0.0004 cd/m² - 4,000 cd/m² integration time between 0.5ms – 1s
+/-2% of measured value, measured at white image of LCD display, Luminance of app. 150 cd/m², x=0.300 y= 0.325
Y +/- 0.1% at 0.1 cd/m² ¹ Y +/- 0.1% at 1 cd/m² ¹ Y +/- 0.1% at 5 cd/m² ¹ Y +/- 0.1% at 150 cd/m² ¹
Chromaticity (x,y) +/- 0.001 after calibration, measured at white image of LCD display, Luminance of app. 150 cd/m², x=0.300 y= 0.325
x,y +/- 0.0004 for Y at 0.1 cd/m² ¹ x,y +/- 0.0002 for Y at 1 cd/m² ¹ x,y +/- 0.00016 for Y at 5 cd/m² ¹ x,y +/- 0.00016 for Y at 150 cd/m² ¹
Measurement speed 4-10 samples / s for Y at 0.1 cd/m² ¹ 10-20 samples / s for Y at 1 cd/m² ¹ 40 samples / s for Y at 5 cd/m² ¹ 40 samples / s for Y at 150 cd/m² ¹
Flicker (contrast method) 4 cd/m² or higher +/- 0.3% flicker frequency 30Hz AC/DC 10% sine wave. Sine wave at 10 cd/m²
+/- 0.2%
Flicker (JEITA method) 4 cd/m² or higher +/- 0.3dB flicker frequency 30Hz AC/DC 10% sine wave. Sine wave at 10 cd/m²
+/- 0.2dB
Operating temperature 10-35°C ²
Shutter lifetime >1000000
Shutter speed 250-300 ms depending on temperature and lifetime
1 All measurements are performed 20 times on a LED display with sufficient signal noise ratio, value is based on 2 sigma. Sample speed depends on the measured sample.
If the sample uses PWM, it will take longer. Use of lower rated values is strongly recommended to ensure repeatability.
2 Operating temperature reaches from 0 to 40 degrees. Dark level compensation is optimized for operating in temperatures between 10 and 35 degrees.
Other temperature ranges can be calibrated using the mechanical shutter if necessary.
11
11 10mm measurement data
Luminance Setting 1: max integration time 100ms Setting 3: max integration time 240ms Setting 2: max integration time 1000ms
Measurement range from (cd/m²) to (cd/m²)
level (%)
level (cd/m²) from (cd/m²)
to (cd/m²)
level (%)
level (cd/m²)
from (cd/m²)
to (cd/m²)
level (%)
level (cd/m²)
0.005 20000 0.005 20000 0.005 20000
Accuracy (for white)
0.01 0.1 0.3 1 5 15 50 100
- - - - - - - 20000
+/-5% +/-5% +/-4% +/-3% +/-3% +/-2% +/-2% +/-2%
- - - - - - - -
0.01 0.1 0.3 1 5 15 50 100
- - - - - - - 20000
+/-5% +/-5% +/-4% +/-3% +/-3% +/-2% +/-2% +/-2%
- - - - - - - -
0.01 0.1 0.3 1 5 15 50 100
- - - - - - - 20000
+/-5% +/-5% +/-4% +/-3% +/-3% +/-2% +/-2% +/-2%
- - - - - - - -
Repeatability (2 sigma)
0.01 0.1 0.3 1 5 15 50 100
- - - - - - - 20000
+/-4% +/-1% +/-1.2% +/-0.6% +/-0.3% +/-0.3% +/-0.2% +/-0.3%
+/-0.0005 +/-0.001 +/-0.0036 +/-0.006 +/-0.015 +/-0.045 +/-0.10 +/-0.30
0.01 0.1 0.3 1 5 15 50 100
- - - - - - - 20000
+/-3% +/-0.6% +/-0.5% +/-0.5% +/-0.3% +/-0.3% +/-0.12% +/-0.08%
+/-0.0003 +/-0.0006 +/-0.0015 +/-0.005 +/-0.015 +/-0.045 +/-0.06 +/-0.08
0.01 0.1 0.3 1 5 15 50 100
- - - - - - - 20000
+/-1% +/-0.5% +/-0.2% +/-0.2% +/-0.15% +/-0.13% +/-0.07% +/-0.07%
+/-0.0001 +/-0.0005 +/-0.0006 +/-0.002 +/-0.0075 +/-0.0195 +/-0.035 +/-0.07
12
Chromaticity Setting 1: max integration time 100ms Setting 3: max integration time 240ms Setting 2: max integration time 1000ms
Measurement range from (cd/m²)
to (cd/m²)
level from (cd/m²)
to (cd/m²)
level from (cd/m²)
to (cd/m²)
level
0.1 20000 0.1 20000 0.1 20000
Accuracy (for white)
0.1 0.3 1 5 15 50 100
- - - - - - 20000
+/-0.005 +/-0.0025 +/-0.002 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015
0.1 0.3 1 5 15 50 100
- - - - - - 20000
+/-0.005 +/-0.0025 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015
0.1 0.3 1 5 15 50 100
- - - - - - 20000
+/-0.003 +/-0.002 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015
Repeatability (2 sigma)
0.1 0.3 1 5 15 50 100
- - - - - - 20000
0.006 0.0026 0.00056 0.00066 0.00035 0.00017 0.00014
0.1 0.3 1 5 15 50 100
- - - - - - 20000
0.0037 0.0014 0.00061 0.00067 0.00028 0.00012 0.00012
0.1 0.3 1 5 15 50 100
- - - - - - 20000
0.001 0.00022 0.00033 0.00066 0.00033 0.00022 0.0001
Speed Setting 1: max integration time 100ms Setting 3: max integration time 240ms Setting 2: max integration time 1000ms
from (cd/m²)
to (cd/m²)
times/sec from (cd/m²)
to (cd/m²)
times/sec from (cd/m²)
to (cd/m²)
times/sec
xyLv 0.1 0.3 1 5 15 50 100
- - - - - - 20000
8 8 10 40 40 40 40
0.1 0.3 1 5 15 50 100
- - - - - - 20000
3.7 3.7 10 40 40 40 40
0.1 0.3 1 5 15 50 100
- - - - - - 20000
0.6 – 0.9 0.6 – 0.9 10 40 40 40 40
13
12 10mm improved measurement data
Luminance Setting 1: max integration time 100ms Setting 3: max integration time 240ms Setting 2: max integration time 1000ms
Measurement range from (cd/m²)
to (cd/m²)
level (%)
level (cd/m²) from (cd/m²)
to (cd/m²)
level (%)
level (cd/m²)
from (cd/m²)
to (cd/m²)
level (%)
level (cd/m²)
0.003 12000 0.003 12000 0.003 12000
Accuracy (for white)
0.01 0.1 0.3 1 5 15 50 100
- - - - - - - 12000
+/-5% +/-5% +/-4% +/-3% +/-3% +/-2% +/-2% +/-2%
- - - - - - - -
0.01 0.1 0.3 1 5 15 50 100
- - - - - - - 12000
+/-5% +/-5% +/-4% +/-3% +/-3% +/-2% +/-2% +/-2%
- - - - - - - -
0.01 0.1 0.3 1 5 15 50 100
- - - - - - - 12000
+/-5% +/-5% +/-4% +/-3% +/-3% +/-2% +/-2% +/-2%
- - - - - - - -
Repeatability (2 sigma)
0.01 0.1 0.3 1 5 15 50 100
- - - - - - - 12000
+/-3% +/-0.9% +/-0.9% +/-0.6% +/-0.3% +/-0.3% +/-0.2% +/-0.3%
+/-0.0003 +/-0.0009 +/-0.0027 +/-0.006 +/-0.015 +/-0.045 +/-0.10 +/-0.30
0.01 0.1 0.3 1 5 15 50 100
- - - - - - - 12000
+/-2% +/-0.5% +/-0.5% +/-0.4% +/-0.3% +/-0.3% +/-0.12% +/-0.08%
+/-0.0002 +/-0.0005 +/-0.0015 +/-0.004 +/-0.015 +/-0.045 +/-0.06 +/-0.08
0.01 0.1 0.3 1 5 15 50 100
- - - - - - - 12000
+/-0.6% +/-0.4% +/-0.2% +/-0.15% +/-0.13% +/-0.11% +/-0.06% +/-0.06%
+/-0.0001 +/-0.0004 +/-0.0006 +/-0.0015 +/-0.0065 +/-0.062 +/-0.030 +/-0.06
14
Chromaticity Setting 1: max integration time 100ms Setting 3: max integration time 240ms Setting 2: max integration time 1000ms
Measurement range from (cd/m²)
to (cd/m²)
level from (cd/m²)
to (cd/m²)
level from (cd/m²)
to (cd/m²)
level
0.003 12000 0.003 12000 0.003 12000
Accuracy (for white)
0.1 0.3 1 5 15 50 100
- - - - - - 12000
+/-0.005 +/-0.0025 +/-0.002 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015
0.1 0.3 1 5 15 50 100
- - - - - - 12000
+/-0.005 +/-0.0025 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015
0.1 0.3 1 5 15 50 100
- - - - - - 12000
+/-0.003 +/-0.002 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015
Repeatability (2 sigma)
0.1 0.3 1 5 15 50 100
- - - - - - 12000
0.004 0.0020 0.00041 0.00056 0.00032 0.00017 0.00014
0.1 0.3 1 5 15 50 100
- - - - - - 12000
0.0025 0.0009 0.00043 0.00040 0.00028 0.00012 0.00012
0.1 0.3 1 5 15 50 100
- - - - - - 12000
0.00075 0.00022 0.00033 0.00066 0.00033 0.00022 0.0001
Speed Setting 1: max integration time 100ms Setting 3: max integration time 240ms Setting 2: max integration time 1000ms
from (cd/m²)
to (cd/m²)
times/sec from (cd/m²)
to (cd/m²)
times/sec from (cd/m²)
to (cd/m²)
times/sec
xyLv 0.1 0.3 1 5 15 50 100
- - - - - - 12000
8 8 10 40 40 40 40
0.1 0.3 1 5 15 50 100
- - - - - - 12000
3.7 3.7 10 40 40 40 40
0.1 0.3 1 5 15 50 100
- - - - - - 12000
1.25 1.5 15 40 40 40 40
15
13 20mm measurement data Luminance Setting 1: max integration time 100ms Setting 3: max integration time 240ms Setting 2: max integration time 1000ms
Measurement range from (cd/m²) to (cd/m²)
level (%)
level (cd/m²) from (cd/m²)
to (cd/m²)
level (%)
level (cd/m²) from (cd/m²)
to (cd/m²)
level (%)
level (cd/m²)
0.0005 5000 0.0005 5000 0.0005 5000
Accuracy (for white)
0.01 0.1 0.3 1 5 15 50 100
- - - - - - - 5000
+/-5% +/-5% +/-4% +/-3% +/-3% +/-2% +/-2% +/-2%
- - - - - - - -
0.01 0.1 0.3 1 5 15 50 100
- - - - - - - 5000
+/-5% +/-5% +/-4% +/-3% +/-3% +/-2% +/-2% +/-2%
- - - - - - - -
0.01 0.1 0.3 1 5 15 50 100
- - - - - - - 5000
+/-5% +/-5% +/-4% +/-3% +/-3% +/-2% +/-2% +/-2%
- - - - - - - -
Repeatability (2 sigma)
0.01 0.1 0.3 1 5 15 50 100
- - - - - - - 5000
+/-2% +/-0.45% +/-0.45% +/-0.3% +/-0.1% +/-0.1% +/-0.1% +/-0.1%
0.01 0.1 0.3 1 5 15 50 100
- - - - - - - 5000
+/-1% +/-0.2% +/-0.2% +/-0.15% +/-0.1% +/-0.1% +/-0.1% +/-0.1%
0.01 0.1 0.3 1 5 15 50 100
- - - - - - - 5000
+/-0.3% +/-0.2% +/-0.1% +/-0.1% +/-0.1% +/-0.1% +/-0.1% +/-0.1%
16
Chromaticity Setting 1: max integration time 100ms Setting 3: max integration time 240ms Setting 2: max integration time 1000ms
Measurement range from (cd/m²)
to (cd/m²)
level from (cd/m²)
to (cd/m²)
level from (cd/m²)
to (cd/m²)
level
0.0005 5000 0.0005 5000 0.0005 5000
Accuracy (for white)
0.1 0.3 1 5 15 50 100
- - - - - - 5000
+/-0.005 +/-0.0025 +/-0.002 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015
0.1 0.3 1 5 15 50 100
- - - - - - 20000
+/-0.005 +/-0.0025 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015
0.1 0.3 1 5 15 50 100
- - - - - - 5000
+/-0.003 +/-0.002 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015
Repeatability (2 sigma)
0.1 0.3 1 5 15 50 100
- - - - - - 5000
0.001 0.0002 0.0002 0.0002 0.0002 0.0002 0.0001
0.1 0.3 1 5 15 50 100
- - - - - - 5000
0.0003 0.0002 0.0002 0.0002 0.0002 0.0002 0.0001
0.1 0.3 1 5 15 50 100
- - - - - - 5000
0.0003 0.0002 0.0002 0.0002 0.0002 0.0002 0.0001
Speed Setting 1: max integration time 100ms Setting 3: max integration time 240ms Setting 2: max integration time 1000ms
from (cd/m²)
to (cd/m²)
times/sec from (cd/m²)
to (cd/m²)
times/sec from (cd/m²)
to (cd/m²)
times/sec
xyLv 0.1 0.3 1 5 15 50 100
- - - - - - 5000
8 8 10 40 40 40 40
0.1 0.3 1 5 15 50 100
- - - - - - 5000
3.7 3.7 10 40 40 40 40
0.1 0.3 1 5 15 50 100
- - - - - - 5000
3.7 10 40 40 40 40 40
17
14 20mm improved & 10mm WA measurement
data
Luminance Setting 1: max integration time 100ms Setting 3: max integration time 240ms Setting 2: max integration time 1000ms
Measurement range from (cd/m²) to (cd/m²)
level (%)
level (cd/m²) from (cd/m²)
to (cd/m²)
level (%)
level (cd/m²) from (cd/m²)
to (cd/m²)
level (%)
level (cd/m²)
0.0004 4000 0.0004 4000 0.0004 4000
Accuracy (for white)
0.01 0.1 0.3 1 5 15 50 100
- - - - - - - 5000
+/-5% +/-5% +/-4% +/-3% +/-3% +/-2% +/-2% +/-2%
- - - - - - - -
0.01 0.1 0.3 1 5 15 50 100
- - - - - - - 5000
+/-5% +/-5% +/-4% +/-3% +/-3% +/-2% +/-2% +/-2%
- - - - - - - -
0.01 0.1 0.3 1 5 15 50 100
- - - - - - - 5000
+/-5% +/-5% +/-4% +/-3% +/-3% +/-2% +/-2% +/-2%
- - - - - - - -
Repeatability (2 sigma)
0.01 0.1 0.3 1 5 15 50 100
- - - - - - - 5000
+/-1.5% +/-0.45% +/-0.45% +/-0.3% +/-0.1% +/-0.1% +/-0.1% +/-0.1%
0.01 0.1 0.3 1 5 15 50 100
- - - - - - - 5000
+/-0.6% +/-0.1% +/-0.1% +/-0.1% +/-0.1% +/-0.1% +/-0.1% +/-0.1%
0.01 0.1 0.3 1 5 15 50 100
- - - - - - - 5000
+/-0.2% +/-0.1% +/-0.1% +/-0.1% +/-0.1% +/-0.1% +/-0.1% +/-0.1%
18
Chromaticity Setting 1: max integration time 100ms Setting 3: max integration time 240ms Setting 2: max integration time 1000ms
Measurement range from (cd/m²)
to (cd/m²)
level from (cd/m²)
to (cd/m²)
level from (cd/m²)
to (cd/m²)
level
0.0004 4000 0.0004 4000 0.0004 4000
Accuracy (for white)
0.1 0.3 1 5 15 50 100
- - - - - - 5000
+/-0.005 +/-0.0025 +/-0.002 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015
0.1 0.3 1 5 15 50 100
- - - - - - 20000
+/-0.005 +/-0.0025 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015
0.1 0.3 1 5 15 50 100
- - - - - - 5000
+/-0.003 +/-0.002 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015
Repeatability (2 sigma)
0.1 0.3 1 5 15 50 100
- - - - - - 5000
0.001 0.0002 0.0002 0.0002 0.0002 0.0002 0.0001
0.1 0.3 1 5 15 50 100
- - - - - - 5000
0.0003 0.0002 0.0002 0.0002 0.0002 0.0002 0.0001
0.1 0.3 1 5 15 50 100
- - - - - - 5000
0.0003 0.0002 0.0002 0.0002 0.0002 0.0002 0.0001
Speed Setting 1: max integration time 100ms Setting 3: max integration time 240ms Setting 2: max integration time 1000ms
from (cd/m²)
to (cd/m²)
times/sec from (cd/m²)
to (cd/m²)
times/sec from (cd/m²)
to (cd/m²)
times/sec
xyLv 0.1 0.3 1 5 15 50 100
- - - - - - 5000
8 8 10 40 40 40 40
0.1 0.3 1 5 15 50 100
- - - - - - 5000
3.7 10 40 40 40 40 40
0.1 0.3 1 5 15 50 100
- - - - - - 5000
3.7 10 40 40 40 40 40
19
15 Xtreme low luminance performance data In below sub-chapters the repeatability data for extreme low luminance
can be found. This data is non-binding, so no claims can be made based
on this data, this data is purely for indication. When measuring assumed is
that dark is kept under control
15.1 Hyperion 10mm
15.2 Hyperion 10mm improved
Chromaticity Setting 3: max integration time 1000ms
Measurement range at (cd/m²) level
Accuracy (for white)
0.005 0.0015
- -
+/-0.02 +/-0.07
Repeatability (2 sigma)
0.005 0.0015
- -
0.012 0.050
Luminance Setting 3: max integration time 1000ms
Measurement level at (cd/m²) level (%)
level (cd/m²)
Accuracy (for white)
0.005 0.0015
- -
+/-6% +/-10%
- -
Repeatability (2 sigma)
0.005 0.0015
- -
+/-2% +/-5%
- -
Luminance Setting 3: max integration time 1000ms
Measurement level at (cd/m²) level (%)
level (cd/m²)
Accuracy (for white)
0.005 0.0015
- -
+/-8% +/-15%
- -
Repeatability (2 sigma)
0.005 0.0015
- -
+/-4% +/-11%
- -
Chromaticity Setting 3: max integration time 1000ms
Measurement range at (cd/m²) level
Accuracy (for white)
0.005 0.0015
- -
+/-0.01 +/-0.07
Repeatability (2 sigma)
0.005 0.0015
- -
0.005 0.015
20
16 Hyperion 10mm dimensions
21
17 Hyperion 10mm & 5mm improved dimensions
22
18 Hyperion 10mm WA dimensions
23
19 Hyperion 20mm dimensions
24
20 Hyperion 20mm improved dimensions
25
21 Hyperion M8 fibre dimensions
26
Admesy B.V. Sleestraat 3 6014 CA Ittervoort The Netherlands T +31 (0)475 600 232 F +31 (0)475 600 316 www.admesy.com [email protected]
The material in this document is subject to change. No rights can be derived from the content of this document. All rights reserved. No part of this document may be reproduced, stored in a database or retrieval system, or published in any form or way, electronically, mechanically, by print, photo print, microfilm or any other means without prior written permission from the publisher.
Version 1.2.1 02/2019