SPECT CT: HOW MUCH RADIATION DOSE
Thanks to:
![Page 2: SPECT/CT: HOW Much Radiation Dose CT Constitute](https://reader034.vdocument.in/reader034/viewer/2022051123/58ee1ec81a28ab09488b46a1/html5/thumbnails/2.jpg)
SPECT CT HYBRID
![Page 3: SPECT/CT: HOW Much Radiation Dose CT Constitute](https://reader034.vdocument.in/reader034/viewer/2022051123/58ee1ec81a28ab09488b46a1/html5/thumbnails/3.jpg)
Material and Methodology
Reference 1: The Mnaual of SYMBIA SPECT/CT:SIEMENS, GERMANY.
We analyzed 151 patients from a dual-headed SIEMENSโ SYMBIA
T-16 unit with an integrated 16-slice CT scanner.
This CT portion of this system has,
variable tube current (20โ 345 mA),
slice thickness of 0.6โ10 mm,
rotational speed of 0.6โ1.5 secs.
tube voltage up to 130 kVp.
![Page 4: SPECT/CT: HOW Much Radiation Dose CT Constitute](https://reader034.vdocument.in/reader034/viewer/2022051123/58ee1ec81a28ab09488b46a1/html5/thumbnails/4.jpg)
Effective Dose
Reference 2: E. J. Hall and A. J. Giaccia, Radiolobiology for the Radiologist, Lippincott Williams &
Wilkins, Philadelphia, Pa, USA, 6th edition, 2005.
The quantity Effective Dose can have practical value for comparing
the relative doses related to the stochastic effects and comparing
different technologies [2]
![Page 5: SPECT/CT: HOW Much Radiation Dose CT Constitute](https://reader034.vdocument.in/reader034/viewer/2022051123/58ee1ec81a28ab09488b46a1/html5/thumbnails/5.jpg)
Radiation Dose Conversation Mechanism
Reference 3: The Essential Physics of Medical Imaging, Third Edition Third, North American Edition Edition by Jerrold T. Bushberg (Author),
J. Anthony Seibert , Edwin M. Leidholdt Jr. , John M. Boone
Game of 10-10 sec [3]
![Page 6: SPECT/CT: HOW Much Radiation Dose CT Constitute](https://reader034.vdocument.in/reader034/viewer/2022051123/58ee1ec81a28ab09488b46a1/html5/thumbnails/6.jpg)
Risk Factor
Reference 2: E. J. Hall and A. J. Giaccia, Radiolobiology for the Radiologist, Lippincott Williams &
Wilkins, Philadelphia, Pa, USA, 6th edition, 2005.
According to ICRP, the risk of developing a lethal cancer is
approximately 4% per Sievert. The lifetime risk for cancer has been
shown to be greater in the younger patients.
![Page 7: SPECT/CT: HOW Much Radiation Dose CT Constitute](https://reader034.vdocument.in/reader034/viewer/2022051123/58ee1ec81a28ab09488b46a1/html5/thumbnails/7.jpg)
Dose Conversion
Reference 4: (AAPM, 2008), Report, Shope et al, 1981
CT dose index Volume (CTDIv) and Dose Length Product (DLP) are
two basic quantities used to estimate the average absorbed dose in
the organs and tissues based for the helical (volumetric) CT
scanners.[4]
![Page 8: SPECT/CT: HOW Much Radiation Dose CT Constitute](https://reader034.vdocument.in/reader034/viewer/2022051123/58ee1ec81a28ab09488b46a1/html5/thumbnails/8.jpg)
Dose Length Product & CT Dose Indexvolume
![Page 9: SPECT/CT: HOW Much Radiation Dose CT Constitute](https://reader034.vdocument.in/reader034/viewer/2022051123/58ee1ec81a28ab09488b46a1/html5/thumbnails/9.jpg)
CT Dose Calculations
Reference 5: International Commission on Radiological Protection (ICRP) Publication 53 and 80.
๐ท๐๐ ๐ ๐ฟ๐๐๐๐กโ ๐๐๐๐๐ข๐๐ก = ๐ถ๐๐ท๐ผ๐ฃ โ ๐๐๐ก๐๐๐๐ก ๐ฟ๐๐๐๐กโ
๐ท๐ฟ๐ = ๐ถ๐๐ท๐ผ๐ฃ โ ๐ฟ
DLP is measured in mGy*cm
![Page 10: SPECT/CT: HOW Much Radiation Dose CT Constitute](https://reader034.vdocument.in/reader034/viewer/2022051123/58ee1ec81a28ab09488b46a1/html5/thumbnails/10.jpg)
CT Dose Calculations
Reference 5: International Commission on Radiological Protection (ICRP) Publication 53 and 80.
๐ธ๐๐๐๐๐ก๐๐ฃ๐ ๐ท๐๐ ๐ = ๐ท๐๐ ๐ ๐ฟ๐๐๐๐กโ ๐๐๐๐๐ข๐๐ก . ๐๐๐๐ฃ๐๐๐ ๐๐๐ ๐น๐๐๐ก๐๐
Effective dose is measured in mSv
๐ธ๐๐๐ = ๐ท๐ฟ๐ . ๐
![Page 11: SPECT/CT: HOW Much Radiation Dose CT Constitute](https://reader034.vdocument.in/reader034/viewer/2022051123/58ee1ec81a28ab09488b46a1/html5/thumbnails/11.jpg)
CT Dose Conversion Factors
Reference 5: International Commission on Radiological Protection (ICRP) Publication 53 and 80.
Reference 6: Doses to Computed Tomography Examinations in the UK โ 2003 Review (Table 3)
Region of Body Effective Dose per DLP for adult
(mSv per mGy.cm)
Head & Neck 0.0031
Head 0.0021
Neck 0.0059
Chest 0.014
Abdomen & Pelvis 0.015
Trunk 0.015
![Page 12: SPECT/CT: HOW Much Radiation Dose CT Constitute](https://reader034.vdocument.in/reader034/viewer/2022051123/58ee1ec81a28ab09488b46a1/html5/thumbnails/12.jpg)
Nuclear Dose Conversion
Reference 7 : โRadiation dose to patients from radiopharmaceuticals ICRP Publication 53,โ Annals of the ICRP, vol. 18, no. 1โ4, 1987.
J. Valentin, โRadiation dose to patients from radiopharmaceuticals (Addendum 2 to ICRP publication 53) ICRP publication
80 approved by the commission in September 1997,โ Annals of the ICRP, vol. 28, no. 3, pp. 1โ126, 1998.
Effective dose per unit administered activityโ conversion factors
listed in the International Commission on Radiological Protection
(ICRP) Publication 53 and 80.
![Page 13: SPECT/CT: HOW Much Radiation Dose CT Constitute](https://reader034.vdocument.in/reader034/viewer/2022051123/58ee1ec81a28ab09488b46a1/html5/thumbnails/13.jpg)
Results
Reference 8: F. A. Mettler, W. Huda, T. T. Yoshizumi, and M. Mahesh, โEffective doses in radiology and diagnostic nuclear medicine: a
catalog,โ Radiology, vol. 248, no. 1, pp. 254โ263, 2008.
Primary Scan99mTc-MDP Bone Mean: 868 MBqAverage DLP: 46.98
% increase in patient dose due to CT Component
CT Dose
from
SPECT
8Diagnostic
CT Dose
NM Dose NM + CT Dose % increase
w.r.t. RP Dose
Chest 0.65 7
4.95
5.6 13.13
Abd & Pelvis 0.87 8-14 5.75 16.17
Trunk 10.45 8-12 15.4 211.11
Neck 0.29 3 5.24 5.85
All doses are in milli-Sievert (mSv)
![Page 14: SPECT/CT: HOW Much Radiation Dose CT Constitute](https://reader034.vdocument.in/reader034/viewer/2022051123/58ee1ec81a28ab09488b46a1/html5/thumbnails/14.jpg)
Results
Primary Scan99mTc-MIBI ParathyroidMean: 842.5 MBqAverage DLP: 135
% increase in patient dose due to CT Component
CT Dose
from
SPECT
8Diagnostic CT
Dose
NM Dose NM + CT
Dose
% increase
w.r.t. RP Dose
Neck 0.8 3 7.58 8.37 10.55
All doses are in milli-Sievert (mSv)
![Page 15: SPECT/CT: HOW Much Radiation Dose CT Constitute](https://reader034.vdocument.in/reader034/viewer/2022051123/58ee1ec81a28ab09488b46a1/html5/thumbnails/15.jpg)
Results
Primary Scan99mTc-ThyroidMean: 165.5 MBqAverage DLP: 63.5
% increase in patient dose due to CT Component
CT Dose
from
SPECT
8Diagnostic
CT Dose
NM Dose SPECT + CT
Dose
% increase w.r.t.
NM Dose
CT Region: Neck 0.01 3 2.15 2.16 0.46
All doses are in milli-Sievert (mSv)
![Page 16: SPECT/CT: HOW Much Radiation Dose CT Constitute](https://reader034.vdocument.in/reader034/viewer/2022051123/58ee1ec81a28ab09488b46a1/html5/thumbnails/16.jpg)
Results
Primary Scan131I-Post-therapyMean: 2790 MBqAverage DLP: 63.5
% increase in patient dose due to CT Component
CT Dose
from
SPECT
8Diagnostic CT
Dose
NM Dose NM + CT
Dose
% increase
w.r.t. RP Dose
Neck 0.43 3 2790 2790.43 0.015
All doses are in milli-Sievert (mSv)
![Page 17: SPECT/CT: HOW Much Radiation Dose CT Constitute](https://reader034.vdocument.in/reader034/viewer/2022051123/58ee1ec81a28ab09488b46a1/html5/thumbnails/17.jpg)
Secret to low dose
As images are not meant to be diagnostic in the SPECT mounted
CT consequently Effective Dose would be expected to be lower
than typical diagnostic values โ This is the secret (lower mA).
![Page 18: SPECT/CT: HOW Much Radiation Dose CT Constitute](https://reader034.vdocument.in/reader034/viewer/2022051123/58ee1ec81a28ab09488b46a1/html5/thumbnails/18.jpg)
Cancer Risk from SPECT
In our calculations the increased risk of cancer by additional
exposure of low does CT ranges from 0.01% to 0.04 %.
![Page 19: SPECT/CT: HOW Much Radiation Dose CT Constitute](https://reader034.vdocument.in/reader034/viewer/2022051123/58ee1ec81a28ab09488b46a1/html5/thumbnails/19.jpg)
Cancer Risk from SPECT
Although imparted CT dose is quite lower however this lower dose
can be balanced by decreasing the injected radiotracer dose at the
cost of increased image time.
![Page 20: SPECT/CT: HOW Much Radiation Dose CT Constitute](https://reader034.vdocument.in/reader034/viewer/2022051123/58ee1ec81a28ab09488b46a1/html5/thumbnails/20.jpg)
Thanks