![Page 1: SURGICAL PROCEDURE GUIDE FOR AR SYSTEM USING THE UNIVERSAL KIT](https://reader036.vdocument.in/reader036/viewer/2022062501/5681625c550346895dd2ad88/html5/thumbnails/1.jpg)
SURGICAL PROCEDURE GUIDEFOR AR SYSTEM USING THE UNIVERSAL KIT
![Page 2: SURGICAL PROCEDURE GUIDE FOR AR SYSTEM USING THE UNIVERSAL KIT](https://reader036.vdocument.in/reader036/viewer/2022062501/5681625c550346895dd2ad88/html5/thumbnails/2.jpg)
SURGICAL KIT – UNIVERSAL
4
9 8
7 6
3
510
11
11 – Guide Drill2 – Pilot Drills 3 – Twist Drills4 – Cortical Drills5 – Drill Extension6 – Fixture Driver (Manual)7 – Fixture Driver (Machine)8 – Torque Wrench9 – 1.2mm Hex Driver10 – Parallel Pin11 – 1.2mm Hex Screw
2
![Page 3: SURGICAL PROCEDURE GUIDE FOR AR SYSTEM USING THE UNIVERSAL KIT](https://reader036.vdocument.in/reader036/viewer/2022062501/5681625c550346895dd2ad88/html5/thumbnails/3.jpg)
GUIDE DRILL
BIOTEM Guide Drill• No slipping• Precise Positioning
Round Drill• Risk of slipping may lead
to: • Less precise
positioning• Damaging bone
tissue
Recommended rotational speed1200 rpm
Round Drill vs. Point Drill
Item Code
PD2029
![Page 4: SURGICAL PROCEDURE GUIDE FOR AR SYSTEM USING THE UNIVERSAL KIT](https://reader036.vdocument.in/reader036/viewer/2022062501/5681625c550346895dd2ad88/html5/thumbnails/4.jpg)
PILOT DRILLS
Recommended rotational speed 700-800 rpm
Marker signals depth reached
Groove disperses cooling liquid
Diameter (Ø)
2.0mm 2.9mm
Item Code BTD2015 BTD2915
![Page 5: SURGICAL PROCEDURE GUIDE FOR AR SYSTEM USING THE UNIVERSAL KIT](https://reader036.vdocument.in/reader036/viewer/2022062501/5681625c550346895dd2ad88/html5/thumbnails/5.jpg)
TWIST DRILLS
Recommended rotational speed 700-800 rpm
Marker signals depth reached
Groove disperses cooling liquid
Diameter (Ø)
3.3mm 3.6mm 4.0mm 4.6mm 5.6mm
Item Code BTD3315 BTD3615 BTD4015 BTD4615 BTD5615
![Page 6: SURGICAL PROCEDURE GUIDE FOR AR SYSTEM USING THE UNIVERSAL KIT](https://reader036.vdocument.in/reader036/viewer/2022062501/5681625c550346895dd2ad88/html5/thumbnails/6.jpg)
CORTICAL DRILLS
Recommended rotational speed 700-800 rpm
Diameter (Ø)
3.4mm 3.7mm 4.0mm 5.0mm 6.0mm
Item Code CDA3411S CDA3711S CDA4011S CDA5011S CDA6011S
Groove Line sets the limit of drilling
![Page 7: SURGICAL PROCEDURE GUIDE FOR AR SYSTEM USING THE UNIVERSAL KIT](https://reader036.vdocument.in/reader036/viewer/2022062501/5681625c550346895dd2ad88/html5/thumbnails/7.jpg)
DRILL EXTENSION
Robust connective capsule design with easy assembly / disassembly
Common connective groove designs create a risk for bending
Item Code
ETB4526
![Page 8: SURGICAL PROCEDURE GUIDE FOR AR SYSTEM USING THE UNIVERSAL KIT](https://reader036.vdocument.in/reader036/viewer/2022062501/5681625c550346895dd2ad88/html5/thumbnails/8.jpg)
PARALLEL PIN
Item Code
DGA1620
Insert this side into Ø 2.0mm osteotomy (after using the straight drill)
Insert this side into the implant screw hole
![Page 9: SURGICAL PROCEDURE GUIDE FOR AR SYSTEM USING THE UNIVERSAL KIT](https://reader036.vdocument.in/reader036/viewer/2022062501/5681625c550346895dd2ad88/html5/thumbnails/9.jpg)
HAND DRIVERS
Hex Driver (for torque wrench)
Hex (Ø) 1.2mm 1.2mm
Length 8mm (Short)
15mm (Long)
Item Code
HDT1208 HDT1215
Hex Driver (Manual)
Hex (Ø) 1.2mm 1.2mm
Length 8mm (Short)
15mm (Long)
Item Code
HDB1208 HDB1215
![Page 10: SURGICAL PROCEDURE GUIDE FOR AR SYSTEM USING THE UNIVERSAL KIT](https://reader036.vdocument.in/reader036/viewer/2022062501/5681625c550346895dd2ad88/html5/thumbnails/10.jpg)
FIXTURE DRIVERS
Platform Mini Regular / Wide
Mini Regular / Wide
Color Code
Yellow Green Yellow Green
Short FDAM2108C FDAM2508C FDAH2108 FDAH2508
Long FDAM2112C FDAM2512C FDAH2115 FDAH2515
Contra-Angle Torque Wrench
![Page 11: SURGICAL PROCEDURE GUIDE FOR AR SYSTEM USING THE UNIVERSAL KIT](https://reader036.vdocument.in/reader036/viewer/2022062501/5681625c550346895dd2ad88/html5/thumbnails/11.jpg)
Joining the driver with the implant – Simple and Safe
1. Insert driver into the appropriate connective space while applying pressure to the top of the implant holder.
2. Apply pressure at the bottom of the holder in order to release the implant-driver assembly.
![Page 12: SURGICAL PROCEDURE GUIDE FOR AR SYSTEM USING THE UNIVERSAL KIT](https://reader036.vdocument.in/reader036/viewer/2022062501/5681625c550346895dd2ad88/html5/thumbnails/12.jpg)
DRILLING SEQUENCEAR FIXTURE DRILLING SEQUENCE
Guide Drill BTD2015 BTD2915 BTD3315 BTD3615 BTD4015 BTD4615 BTD5615
- Ø 2.0mm Ø 2.9mm Ø 3.3mm Ø 3.6mm Ø 4.0mm Ø 4.6mm Ø 5.6mm
CDA3413S
Mini Ø 3.4mm
CDA3713S
Regular Ø 3.7mm
CDA4013S
Regular Ø 4.0mm
CDA4513S
Regular Ø 4.5mm
CDA5013S
Wide Ø 5.0mm
CDA6013S
Wide Ø 6.0mm
1
1
2
2
3
4
5
6
3 4 5 6
![Page 13: SURGICAL PROCEDURE GUIDE FOR AR SYSTEM USING THE UNIVERSAL KIT](https://reader036.vdocument.in/reader036/viewer/2022062501/5681625c550346895dd2ad88/html5/thumbnails/13.jpg)
SURGICAL PROCEDURE GUIDEFOR AR SYSTEM USING THE EXTENDED KIT
![Page 14: SURGICAL PROCEDURE GUIDE FOR AR SYSTEM USING THE UNIVERSAL KIT](https://reader036.vdocument.in/reader036/viewer/2022062501/5681625c550346895dd2ad88/html5/thumbnails/14.jpg)
SURGICAL KIT - EXTENDED
21
4 65
7
108
9
1113
1 – Guide Drill2 – Spare Hole3 – Straight Drills4 – Tapered Drills5 – Marking Drills6 – Cortical Drills7 – Parallel Pins8 – Hand Drivers9 – Torque Wrench10 – Drill Extension11 – Spare Holes12 – Implant Drivers (2 Stage)13 – Implant Drivers (1 Stage)
12
3
Item Code
KIT0700
![Page 15: SURGICAL PROCEDURE GUIDE FOR AR SYSTEM USING THE UNIVERSAL KIT](https://reader036.vdocument.in/reader036/viewer/2022062501/5681625c550346895dd2ad88/html5/thumbnails/15.jpg)
GUIDE DRILL
BIOTEM Guide Drill• No slipping• Precise Positioning
Round Drill• Risk of slipping may lead
to: • Less precise
positioning• Damaging bone
tissue
Recommended rotational speed1200 rpm
Round Drill vs. Point Drill
Item Code
PD2029
![Page 16: SURGICAL PROCEDURE GUIDE FOR AR SYSTEM USING THE UNIVERSAL KIT](https://reader036.vdocument.in/reader036/viewer/2022062501/5681625c550346895dd2ad88/html5/thumbnails/16.jpg)
STRAIGHT DRILLS
Recommended rotational speed 1200 rpmHarvest bone at low rotational speed (30-50 rpm)
Length 8.5mm 10mm 11.5mm 13mm 15mm
Item Code BSD2008 BSD2010 BSD2011 BSD2013 BSD2015
Ø 2.0mm
Irrigation groove for easy penetration of cooling media
Stopper for increased drilling precision (prevents over-drilling)
Straight drill design
![Page 17: SURGICAL PROCEDURE GUIDE FOR AR SYSTEM USING THE UNIVERSAL KIT](https://reader036.vdocument.in/reader036/viewer/2022062501/5681625c550346895dd2ad88/html5/thumbnails/17.jpg)
TAPERED DRILLS (1/2)
Recommended rotational speed 1200 rpmHarvest bone at low rotational speed (30-50 rpm)
Irrigation groove for easy penetration of cooling media
Stopper for increased drilling precision (prevents over-drilling)
Taper angle of 15°
![Page 18: SURGICAL PROCEDURE GUIDE FOR AR SYSTEM USING THE UNIVERSAL KIT](https://reader036.vdocument.in/reader036/viewer/2022062501/5681625c550346895dd2ad88/html5/thumbnails/18.jpg)
TAPERED DRILLS (2/2)
Length 8.5mm 10mm 11.5mm 13mm 15mm
Item Code BTD3008 BTD3010 BTD3011 BTD3013 BTD3015
Ø 3.0mm
Length 8.5mm 10mm 11.5mm 13mm 15mm
Item Code BTD3308 BTD3310 BTD3311 BTD3313 BTD3315
Length 8.5mm 10mm 11.5mm 13mm 15mm
Item Code BTD3608 BTD3610 BTD3611 BTD3613 BTD3615
Length 8.5mm 10mm 11.5mm 13mm 15mm
Item Code BTD3808 BTD3810 BTD3811 BTD3813 BTD3815
Length 8.5mm 10mm 11.5mm 13mm 15mm
Item Code BTD4208 BTD4210 BTD4211 BTD4213 BTD4215
Ø 3.6mm
Ø 3.3mm
Ø 3.8mm
Ø 4.2mm
![Page 19: SURGICAL PROCEDURE GUIDE FOR AR SYSTEM USING THE UNIVERSAL KIT](https://reader036.vdocument.in/reader036/viewer/2022062501/5681625c550346895dd2ad88/html5/thumbnails/19.jpg)
MARKING DRILLS
Recommended rotational speed 700-800 rpmHarvest bone at low rotational speed (30-50 rpm)
Diameter (Ø)
2.7mm 4.6mm 5.0mm 5.6mm
Item Code MDR2713 MDR4613 MDR5013 MDR5613
Irrigation groove for easy penetration of cooling media
Stopper for increased drilling precision (prevents over-drilling)
Lateral microgrooves enhance cooling media dispersion
13 11.5
10 8.5 7
![Page 20: SURGICAL PROCEDURE GUIDE FOR AR SYSTEM USING THE UNIVERSAL KIT](https://reader036.vdocument.in/reader036/viewer/2022062501/5681625c550346895dd2ad88/html5/thumbnails/20.jpg)
CORTICAL DRILLS
Recommended rotational speed 700-800 rpm
Diameter (Ø)
3.4mm 3.7mm 4.0mm 4.5mm 5.0mm 6.0mm
Item Code
CDA3411 CDA3711 CDA4011 CDA4511 CDA5011 CDA6011
Marker signals Limit of groove line
Groove Line sets the limit of drilling
![Page 21: SURGICAL PROCEDURE GUIDE FOR AR SYSTEM USING THE UNIVERSAL KIT](https://reader036.vdocument.in/reader036/viewer/2022062501/5681625c550346895dd2ad88/html5/thumbnails/21.jpg)
DRILL EXTENSION
Robust connective capsule design with easy assembly / disassembly
Common connective groove designs create a risk for bending
Item Code
ETB4526
![Page 22: SURGICAL PROCEDURE GUIDE FOR AR SYSTEM USING THE UNIVERSAL KIT](https://reader036.vdocument.in/reader036/viewer/2022062501/5681625c550346895dd2ad88/html5/thumbnails/22.jpg)
PARALLEL PIN
Item Code
DGA1620
Insert this side into Ø 2.0mm osteotomy (after using the straight drill)
Insert this side into the implant screw hole
![Page 23: SURGICAL PROCEDURE GUIDE FOR AR SYSTEM USING THE UNIVERSAL KIT](https://reader036.vdocument.in/reader036/viewer/2022062501/5681625c550346895dd2ad88/html5/thumbnails/23.jpg)
HAND DRIVERS
Hex Driver (for torque wrench)
Hex (Ø) 1.2mm 1.2mm
Length 8mm (Short)
15mm (Long)
Item Code
HDT1208 HDT1215
Hex Driver (Manual)
Hex (Ø) 1.2mm 1.2mm
Length 8mm (Short)
15mm (Long)
Item Code
HDB1208 HDB1215
![Page 24: SURGICAL PROCEDURE GUIDE FOR AR SYSTEM USING THE UNIVERSAL KIT](https://reader036.vdocument.in/reader036/viewer/2022062501/5681625c550346895dd2ad88/html5/thumbnails/24.jpg)
FIXTURE DRIVERS
Platform Mini Regular / Wide
Mini Regular / Wide
Color Code
Yellow Green Yellow Green
Short FDAM2108C FDAM2508C FDAH2108 FDAH2508
Long FDAM2112C FDAM2512C FDAH2115 FDAH2515
Contra-Angle Torque Wrench
![Page 25: SURGICAL PROCEDURE GUIDE FOR AR SYSTEM USING THE UNIVERSAL KIT](https://reader036.vdocument.in/reader036/viewer/2022062501/5681625c550346895dd2ad88/html5/thumbnails/25.jpg)
Joining the Driver with the Implant Fixture
1. Apply pressure on the top part of the fixture mount then insert driver into the appropriate connective space
2. Apply pressure on the bottom part of the fixture mount in order to release the implant-driver assembly.
Ergonomic mount design allows for a simple assembly with the driver as well as a safe manipulation.
![Page 26: SURGICAL PROCEDURE GUIDE FOR AR SYSTEM USING THE UNIVERSAL KIT](https://reader036.vdocument.in/reader036/viewer/2022062501/5681625c550346895dd2ad88/html5/thumbnails/26.jpg)
DRILLING SEQUENCE
AR FIXTURE DRILLING SEQUENCE (1/2)
Point Drill
Straight Drill
Tapered Drill
Cortical Drill
Implant Tapered Drill
Cortical Drill
Implant Tapered Drill
Cortical Drill
Implant
- Ø 2.0mm Ø 3.0mm Ø 3.4mm Ø 3.4mm Ø 3.3mm Ø 3.7mm Ø 3.7mm Ø 3.6mm Ø 4.0mm Ø 4.0mm
![Page 27: SURGICAL PROCEDURE GUIDE FOR AR SYSTEM USING THE UNIVERSAL KIT](https://reader036.vdocument.in/reader036/viewer/2022062501/5681625c550346895dd2ad88/html5/thumbnails/27.jpg)
DRILLING SEQUENCE
AR FIXTURE DRILLING SEQUENCE (2/2)
Tapered Drill
Cortical Drill
Implant Marking Drill
Cortical Drill
Implant Marking Drill
Cortical Drill
Implant
Ø 4.2mm Ø 4.5mm Ø 4.5mm Ø 4.6mm Ø 5.0mm Ø 5.0mm Ø 5.6mm Ø 6.0mm Ø 6.0mm