Download - UNDERSTANDING NEONATAL WAVEFORM GRAPHICS
![Page 1: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/1.jpg)
Kuehne 2011
UNDERSTANDING NEONATAL WAVEFORM GRAPHICS
Brandon Kuehne, MBA, RRT-NPS, RPFT Director- Neonatal Respiratory Services
![Page 2: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/2.jpg)
Kuehne 2011
Disclosures
Purpose: To enhance bedside staff’s knowledge of ventilation and oxygenation support to the neonate. To support and encourage the use of the proximal flow probe via the explanation of theory of operation. To encourage the practical application of waveform analysis associated with neonatal pressure ventilation.
Objectives: Identify and discuss the various clinical types of graphic waveforms provided at bedside to the neonatal
caregiver. Describe the common types neonatal ventilatory complications that can be diagnosed and corrected with the
proper application of graphic waveform identification and analysis.
The Planning Committee and Faculty of this activity have no disclosed conflicts of interest related to this content.
Completion Criteria: In order to receive Continuing Nursing Education (CNE) credit, you must attend 80% of the program.
No commercial support was received for this program Nationwide Children’s Hospital’s accreditation as a provider refers to recognition of educational activities
only and does not imply ANCC Commission on Accreditation, Ohio Board of Nursing, ONA or Children’s Hospital’s approval or endorsement of any product.
![Page 3: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/3.jpg)
Kuehne 2011
What is Respiration?
Ventilation The removal of CO2
Oxygenation The uptake of 02
![Page 4: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/4.jpg)
Kuehne 2011
How do I achieve these items, Mechanically? Ventilation component is comprised of two parts
Tidal Volumes RateTogether they make:
Vt X RR = minute volume
In Neonate approx. 200ml/kg/min• Oxygenation component consists of a
combination of: Fractional inspired Oxygen (Fi02) Alveolar pressure (Mean Airway Pressure)
![Page 5: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/5.jpg)
Kuehne 2011
Oxygenation
FiO2 21% - 100%
Alveolar Pressure (MAP) Distending the alveoli allows oxygen exchange to
take place over greater period of time. The biggest components that affect the MAP are
PEEP CPAP
PIP and Rate and Insp. Time affect MAP to a much lesser extent
![Page 6: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/6.jpg)
Kuehne 2011
Ventilation
Tidal Volume- Effective Vt =
Exhaled Vt – ((PIP – PEEP) x Tubing compliance) wt.(kg)
Most ventilators can display this number in real time if the proper patient weight is inputted
(targets are usually 4-6ml/kg for newborns)
![Page 7: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/7.jpg)
Kuehne 2011
Ventilation
Respiratory Rate Frequency that alveoli expand and contract
Good starting numbers Newborn – 3 months 30 - 40bpm 3 months to 2 years 20bpm 2 years – adult 12bpm
EXCEPTION … disease process involving air trapping
![Page 8: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/8.jpg)
Proximal Airway Flow Monitoring
![Page 9: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/9.jpg)
Kuehne 2011
What is measured and where? Pressure- is measured back
at machine Time – is measured back at
machine Flow is measured at patient
(only with flow probe in place) Volumes (Vti and Vte) are
derived from flows read at patient wye via flow sensor
(deriving volumes from flow signals is a process called integration)
![Page 10: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/10.jpg)
Kuehne 2011
Proximal Airway Monitoring
Hotwire flow sensor is required in order to run ventilator properly in Neonatal Mode
Provides accurate two-way flow monitoring at patient’s airway
Critical for volume measurements on VLBW neonates
![Page 11: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/11.jpg)
Kuehne 2011
+ -
Heated Anemometers (flow sensor) How Do They Work?
Heated element (gold wire) Measures the current necessary to
maintain the temperature constant (cooling effect of gas flow)
Lack of moving parts Fast and sensitive response
(Electrical signal –response slightly under speed of light)
Virtually no resistance Very Accurate
![Page 12: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/12.jpg)
Kuehne 2011
Flow Sensor Issues
Humidity Water will create significant fluctuations of
accuracy. Secretions- surfactant
Reading above or below baseline in the presence of zero flow
Very delicate-breaks easily Wears-out due to processing and age
![Page 13: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/13.jpg)
Kuehne 2011
Calibration of Heated Wire Flow Sensor
Occasionally RTs will need to disconnect flow sensor from the patient and perform zero flow calibration maneuver in order to reset flow reading to baseline.
![Page 14: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/14.jpg)
Kuehne 2011
Scalar Identification and Analysis
Pressure
Flow
Volume
0+
_
Insp. Exp.
PIP
Exp.
PEEP
Insp.
![Page 15: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/15.jpg)
Kuehne 2011
Waveform Identification and Analysis
Pressure
Flow
0+
_
Insp. Exp.
PIP
Exp.
PEEP
Insp.Red indicates machine initiated breath
Flow Baseline
![Page 16: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/16.jpg)
Kuehne 2011
SIMV with Spontaneous Pressure Supported Breaths
Yellow indicates patient triggered breath
![Page 17: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/17.jpg)
Kuehne 2011
Paw
Paw
Synchronized Intermittent Ventilation (SIMV)
SIMV: Mandatory (patient or machine init Spontaneous breaths
Are there any spontaneous breaths here?
![Page 18: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/18.jpg)
Kuehne 2011
Paw CPAP
Paw PS
Pressure Support Ventilation
![Page 19: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/19.jpg)
Kuehne 2011
Increased Expiratory Resistance
Prolonged expiratory flow indicates an obstruction to exhalation and may be caused by obstruction of a large airway, bronchospasm, or secretions
![Page 20: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/20.jpg)
Kuehne 2011
Increased Expiratory Resistance
Normal Resistance
Increased Resistance
Possible Intrathoracic Obstruction- i.e. bronchospasm or secretions
![Page 21: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/21.jpg)
Kuehne 2011
Variable Airway Obstruction-Secretions or water in tubing (extra-thoracic)
Jagged flow signal on inspiration
![Page 22: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/22.jpg)
Kuehne 2011
Airway Obstruction- Flow-Volume Loop before and after removing
water from tubing
BE
BEFORE SXN AFTER SXN
![Page 23: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/23.jpg)
Kuehne 2011
Insufficient Expiratory Time
Expiratory flow is unable to return to baseline prior to the initiation of the next mechanical breath
Incomplete exhalation causes gas trapping, dynamic hyper-expansion and the development of intrinsic PEEP (aka “Auto-PEEP” or “Breath-stacking”)
Can be fixed by decreasing I-time
![Page 24: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/24.jpg)
Kuehne 2011
Gas Trapping with Inappropriate Inspiratory Time-Insufficient Exp.Time
Inspiratory Time 0.5 s
Inspiratory Time 0.3 sInspiration beginning before flow returns to baseline
Plenty of time to exhale at this I-Time
![Page 25: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/25.jpg)
Kuehne 2011
Air leaks
![Page 26: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/26.jpg)
Kuehne 2011
Air leak- Related to ET tubes or circuit
Volume
Pressure
Flow
Volume
Time
Volume never returns to baseline
![Page 27: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/27.jpg)
Kuehne 2011
Air leak- Related to ET tubes or Circuit
Volume Flow
Volume
Time
![Page 28: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/28.jpg)
Kuehne 2011
“Sawtooth" Pattern
![Page 29: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/29.jpg)
Kuehne 2011
Baseline Flow
![Page 30: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/30.jpg)
Kuehne 2011
![Page 31: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/31.jpg)
Kuehne 2011
Patient Lockout
![Page 32: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/32.jpg)
Kuehne 2011
Trigger Sensitivity-Inappropriate Flow Trigger
e
Sensitivity level
Time
Flow
TimeLook at all this unsupported patient effort
![Page 33: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/33.jpg)
Kuehne 2011
Trigger Sensitivity- Appropriate Flow Trigger
![Page 34: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/34.jpg)
Kuehne 2011
Auto Cycling
![Page 35: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/35.jpg)
Kuehne 2011
Autocycling- Secondary to Leak
![Page 36: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/36.jpg)
Kuehne 2011
Resolved Autocycle- Flow Trigger Increased
Trigger ↑’d to 0.7 L/M -auto cycle ended
A short expiratory hold maneuver revealed no patient effort
![Page 37: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/37.jpg)
Kuehne 2011
Flow Starvation
![Page 38: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/38.jpg)
Kuehne 2011
Insufficient PSV
Figure “8”
Just a little more pressure (indirectly increasing flow) and flow starvation goes away
![Page 39: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/39.jpg)
Kuehne 2011
Pressure Volume Loops to Assess RDS
![Page 40: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/40.jpg)
Kuehne 2011
Graphical Analysis of RDS
Pre Surfactant
![Page 41: UNDERSTANDING NEONATAL WAVEFORM GRAPHICS](https://reader030.vdocument.in/reader030/viewer/2022020622/61ebb06da437095ca0449c2c/html5/thumbnails/41.jpg)
Kuehne 2011
Pressure- Volume Graphical Analysis of RDS
Six Hours Post Surfactant
Nice football shape @ 45° angle