dpg digital pressure gauge instructions - winters...

2
DPG Digital Pressure Gauge Instructions Tel: 1-800-WINTERS / www.winters.com WINTERS INSTRUMENTS Menu Button Press - Changes pressure units (psi, mH 2 O, mmHg, mbar, kPa ranges 100 psi, psi, bar, kg/cm 2 , MPa, atm ranges 160 psi) Hold for 4 - 5 Seconds and Release - Reset to zero (This process will take a few seconds) Power Button Press - Turns gauge ON for about 1 minute and displays battery reading. Backlight* will stay ON for 10 sec - Turns gauge OFF (hold for 2 sec) Press While ON - Turns backlight* back ON for 10 sec * backlight automatically comes ON when pressure changes ±10% of scale (when gauge is ON) LCD Display Battery Reading Pressure Reading Pressure Scale (units) 1. 2. 3. 4. 5. To Change Batteries 1. Remove Rubber Boot 2. Remove Top Phillips Screw to Open Cover 3. Exchange Batteries (Two AAA Batteries) 4. Close Cover and Re-install Screw 5. Re-install Rubber Boot

Upload: hakien

Post on 09-Mar-2018

214 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: DPG Digital Pressure Gauge Instructions - Winters …winters.com/wp-content/uploads/DPG-Instructions.pdf · DPG Digital Pressure Gauge Instructions Tel: 1-800-WINTERS / WINTERS INSTRUMENTS

DPG Digital Pressure Gauge Instructions

Tel: 1-800-WINTERS / www.winters.com WINTERS INSTRUMENTS

Menu Button• Press -Changespressureunits (psi,mH2O,mmHg,mbar,kParanges≤100psi, psi,bar,kg/cm2,MPa,atmranges≥160psi)

Hold for 4 - 5 Seconds and Release• - Resettozero (Thisprocesswilltakeafewseconds)

Power Button• Press -TurnsgaugeONforabout1minuteanddisplays batteryreading.Backlight*willstayONfor10sec -TurnsgaugeOFF(holdfor2sec)

• Press While ON -Turnsbacklight*backONfor10sec

*backlightautomaticallycomesONwhenpressure changes±10%ofscale(whengaugeisON)

LCD Display

Battery Reading

Pressure Reading

Pressure Scale (units)

1. 2.

3. 4.

5.

To Change Batteries

1. Remove Rubber Boot

2. Remove Top Phillips Screw to Open Cover

3. Exchange Batteries (Two AAA Batteries)

4. Close Cover and Re-install Screw

5. Re-install Rubber Boot

Page 2: DPG Digital Pressure Gauge Instructions - Winters …winters.com/wp-content/uploads/DPG-Instructions.pdf · DPG Digital Pressure Gauge Instructions Tel: 1-800-WINTERS / WINTERS INSTRUMENTS

DPGDigital Pressure Gauge Instructions

WINTERS INSTRUMENTS Tel: 1-800-WINTERS / www.winters.comMKT5164

12/14

Setup Options

HoldMenu Buttonforabout6seconds

A) Displaywillshow“tP-nP”

B) PressMenu Buttonanddisplaywillshow“Pot=X”.Thisselectsthenumberofdigitsafterthedecimalpoint.Press Power Buttontoselect“2or3”(forranges≤100psi)or“3or4”(forranges≥160psi).

≤100psi

Pot=2: -Nodigitsafterdecimalpoint(mmHg,mbar) -Onedigitafterdecimalpoint(kPa) -Twodigitsafterdecimalpoint(psi,mH2O)

Pot=3: -Onedigitafterdecimalpoint(mmHg,mbar) -Twodigitsafterdecimalpoint(kPa) -Threedigitsafterdecimalpoint(psi,mH2O)

≥160psi

Pot=3: -Nodigitsafterdecimalpoint(psi) -Onedigitafterdecimalpoint(bar,kg/cm2,atm) -Twodigitsafterdecimalpoint(MPa)

Pot=4: -Onedigitafterdecimalpoint(psi) -Twodigitsafterdecimalpoint(bar,kg/cm2,atm) -Threedigitsafterdecimalpoint(MPa)

C) PressMenu ButtonandDisplaywillshow“OFF=0”.ThisselectstheAutoorManualShutOffmodes.PressPower Buttontoselect“0”or“1”. OFF=0:AutoShutOffmode;DisplayautomaticallyshutsOFFafterabout1minute OFF=1:ManualShutOffmode;DisplaywillremainconstantlyONuntilShutOffmanually(thelifeofthebatteriesisabout 15days).

D) PressMenu ButtonandDisplaywillshow“HL=0”.ThisselectstheDisplaymode.PressPower Buttontoselect“0”or“1”.

HL=0: Displayshowsactualandcontinuousprocesspressure HL=1: Displayshowsonlypeakprocesspressure*

E) PressMenuButtontoreturntoprocesspressuredisplay

*GaugeneedstobecycledON/OFFtobereset