dry matter density and relative feed value of a...

13
1 Dry Matter Density and Relative Feed Value of a Sensor Developed by Harvest Tec Allen Young Utah State University Introduction Harvest Tec (Hudson, WI) has developed a sensor, which attaches to a baler, and is then designed to measure moisture %, acid detergent fiber % (ADF) and neutral detergent fiber % (NDF) of hay as it is baled. These two values are then be used to determine the relative feed value (RFV) for each bale of hay. An additional piece of equipment can also be attached to the baler that gives the weight of each bale. These pieces of information are entered onto an RFID tag that is attached to the bale. In addition, the location of the field from which the bale came from can be added to the tag. A hand-held or stationary scanner is then used to read the information on each tag. Having this information at the time of baling can save time and money in terms of sampling costs and provides real-time information for the buyer or seller. The purpose of this project was to compare scissor clipped samples from each field (essentially this is the control) with laboratory analysis of each bale, by field and cutting, to help develop and calibrate the sensor to give accurate results in the field. Material and Methods This project was carried out on farms owned and operated by Utah State University Agricultural Experiment Station. Hay samples from 3 fields over 3 cuttings (546 total bales) were used for this project. On the day of cutting, scissor samples were collected at different locations within the field, composited, and then taken to the USU Analytical Laboratory for compositional analysis. The scissor cuts were taken to mimic the height of the cutting bar to give representative samples. For most fields and cuttings, one sample was used to compare with the laboratory samples; however, there were two fields that had 3 and 5 samples taken. These were used to compare statistical differences and extrapolated back to the laboratory samples. Bales from these fields were then sampled and analyzed at the same laboratory to give the hay composition. These values were averaged and compared with the scissor cut sample(s).

Upload: others

Post on 21-Jun-2020

5 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Dry Matter Density and Relative Feed Value of a …harvesttec.com/wp-content/uploads/2016/01/USU-final...1 Dry Matter Density and Relative Feed Value of a Sensor Developed by Harvest

1

Dry Matter Density and Relative Feed Value of a Sensor Developed by Harvest Tec

Allen Young

Utah State University

Introduction

Harvest Tec (Hudson, WI) has developed a sensor, which attaches to a baler, and is then

designed to measure moisture %, acid detergent fiber % (ADF) and neutral detergent fiber %

(NDF) of hay as it is baled. These two values are then be used to determine the relative feed

value (RFV) for each bale of hay. An additional piece of equipment can also be attached to the

baler that gives the weight of each bale. These pieces of information are entered onto an RFID

tag that is attached to the bale. In addition, the location of the field from which the bale came

from can be added to the tag. A hand-held or stationary scanner is then used to read the

information on each tag. Having this information at the time of baling can save time and money

in terms of sampling costs and provides real-time information for the buyer or seller. The

purpose of this project was to compare scissor clipped samples from each field (essentially this is

the control) with laboratory analysis of each bale, by field and cutting, to help develop and

calibrate the sensor to give accurate results in the field.

Material and Methods

This project was carried out on farms owned and operated by Utah State University

Agricultural Experiment Station. Hay samples from 3 fields over 3 cuttings (546 total bales)

were used for this project. On the day of cutting, scissor samples were collected at different

locations within the field, composited, and then taken to the USU Analytical Laboratory for

compositional analysis. The scissor cuts were taken to mimic the height of the cutting bar to

give representative samples. For most fields and cuttings, one sample was used to compare with

the laboratory samples; however, there were two fields that had 3 and 5 samples taken. These

were used to compare statistical differences and extrapolated back to the laboratory samples.

Bales from these fields were then sampled and analyzed at the same laboratory to give the hay

composition. These values were averaged and compared with the scissor cut sample(s).

Page 2: Dry Matter Density and Relative Feed Value of a …harvesttec.com/wp-content/uploads/2016/01/USU-final...1 Dry Matter Density and Relative Feed Value of a Sensor Developed by Harvest

2

Results

Analysis of the scissor cut samples are shown in Table 1. Only averages for Field 1&2 – 1 (n

= 3) and Field 10 – 3 (n = 5) are shown. All others represent one value. Table 2 shows the

averages of the laboratory analysis for samples taken from bales based on field and cutting. The

values are graphed and shown in Figures 1 to 5. The most obvious difference is in moisture %.

This is to be expected and will be discussed later. Figures 6 – 9 are statistical comparisons using

Field 10 – 3 because there were 5 scissor samples. Because of multiple samples, standard

deviations could be compared. There were 49 laboratory samples. A t-test (Two samples

assuming unequal variance) was performed on each of the 4 different composition variables. All

were significantly different except for ADF%. Even though significantly different, from a

practical aspect, the differences were minor. One field was chosen to show the individual

variation between bales for one parameter of interest. Figure 10 shows the RFV for all bales

from one field for one specific cutting. Any field could have been chosen and would have shown

similar variation. For this one field the mean was 184 and ranged from an RFV of 140 to 232.

The important point is that all bales from a field are not uniform and sampling can be a great

source of variation (and possibly contention) if too few samples are taken to give a uniform

sample for analysis.

A subset of the total bales was pair matched to compare the sensor moisture % against the

laboratory analysis moisture %. The results are shown in Table 3. The correlation, as we

expected, was very low (Figure 11). We recognize that the laboratory samples are going to be

more uniform and drier than the bale moisture because of the sample processing used prior to

laboratory analysis. The purpose was to try and develop some type of correction factor that

could be used to come up with a more realistic dry matter value in the field. Baler moisture

percent was corrected based on the mean and standard deviation of the baler moisture content for

the 383 samples. Therefore the baler moistures were adjusted by the following correction

factors: if moisture was <8.21%, then 0.93 was subtracted from the original baler moisture %; if

<12.51, then 0.98 was subtracted; if <16.81, then 1.03 was subtracted; if anything was over

16.81, then 1.08 was subtracted. The results of the correlation between the dry matter of a bale

based on laboratory analysis and the dry matter of a bale based on the corrected moisture from

Page 3: Dry Matter Density and Relative Feed Value of a …harvesttec.com/wp-content/uploads/2016/01/USU-final...1 Dry Matter Density and Relative Feed Value of a Sensor Developed by Harvest

3

the sensor are shown in Figure 12. While not perfect, it shows an improvement in the R2 value

and may be useful in giving more realistic values of actual bale weight on a dry matter basis.

Correction for size of bale based on cubic feet and cubic feet of dry matter probably won’t

improve the value because multiplying by a constant doesn’t change the relative difference

between bales. We could not find an acceptable way of correcting our data set for moisture or

density that would provide an acceptable estimate of RFV. It appears that will have to come

from the laboratory analysis.

Additional figures are listed showing relationships between different compositional variables

that may be of interest in the future. They are: Figure 13 (ADF% compared with NDF% of bale

samples based on laboratory analysis); Figure 14 (RFV compared with NDF%, ADF% and CP%

in bale samples based on laboratory analysis); Figure 15 (RFV compared with NEL in bale

samples based on laboratory analysis); Figure 16 (RFV compared with TDN% of bale samples

based on laboratory analysis); and Figure 17 (NDF% compared with CP%, by cutting, based on

laboratory analysis).

Discussion and Recommendations

1. Using the scissor cut method is a reasonable way to get information for calibrating the

sensor on the baler. Based on the overall averages for all fields and all cuttings, all

composition values were within normal ranges of variation and analytical techniques.

The differences between CP%, ADF% and NDF% were 1.3%, 0.8% and 0.1%,

respectively. The RFV was ~ 2 units different. These are all significantly less than the

standard deviations of the laboratory results for these same variables (1.4%, 2.4% and

3.0%, respectively). The standard deviation for RFV was 19.8 units. This is interesting

because the range in values, using the same sample, for this laboratory is approximately

3.5, 2.1 and 3.5%, respectively (results from an unrelated project). However, there is a

certain amount of built-in variation within the whole system that needs to be kept in mind

such as sampling variation, laboratory variation and within field variation (Fig. 10). The

“big picture” is that this is a reasonable method of calibration. I would recommend that

multiple (at least 3) field samples be collected and averaged for the calibration in order to

Page 4: Dry Matter Density and Relative Feed Value of a …harvesttec.com/wp-content/uploads/2016/01/USU-final...1 Dry Matter Density and Relative Feed Value of a Sensor Developed by Harvest

4

help watch for and deal with field variation. Another aspect to consider would to collect

“scissor” samples by taking several grab samples of the new mown hay immediately after

cutting rather than clipping samples prior to cutting. In my opinion, you would get the

same results.

2. There are large variations in the moisture % of the bale as it comes off the baler. These

do not correlate with laboratory moisture contents. This is to be expected because the

laboratory samples are dried prior to analysis. Because of this variation, the “in field”

dry matter in a bale could be misleading and not be a true reflection of the actual dry

matter. This may reflect “real life” issues of buying and selling hay and probably over or

under estimate the true dry matter in a bale in the field because hay is bought or sold

based on actual weight. Therefore, an effort was made to try and correct the bale dry

matter to closer reflect the actual dry matter based on laboratory moisture. This was

moderately successful, but still only explains ~51% of the variation (this is an

improvement from 25%; Figs. 11 vs 12).

3. Other relationships, based on the laboratory analysis of the samples are shown for the

purpose of providing information that might be useful for incorporation of future

information into the RFID tag.

4. As a final comment, the system seems to work and appears to be a practical way to get a

reasonable approximation of the quality of hay as it comes out of the field. Because of

the variation, I would suggest that there still needs to be some type of process to regularly

calibrate the sensor so that it gives values representative of the hay in question. The

system seems to work well under Utah conditions, using alfalfa hay, but we didn’t test

the system using grass hays or silages. I think that would be the next step in this process.

Page 5: Dry Matter Density and Relative Feed Value of a …harvesttec.com/wp-content/uploads/2016/01/USU-final...1 Dry Matter Density and Relative Feed Value of a Sensor Developed by Harvest

5

Table 1. Scissor clip values by field and cutting for 2013. Field 1&2 – 1 has 3 samples and Field

10 – 3 had 5 samples, but only the averages are shown here.

FIELD Moist. % Bale Weight CP% ADF% NDF% TDN% RFV

Field 1&2 - 1st 5 1466 22.7 28.2 33.0 68.6 188.9

Field 1&2 - 2nd 9 1299 19.9 33.0 39.9 N/A 147.5

Field 1&2 - 3rd 8 1432 25.5 26.1 29.7 70.7 214.7

Field 7 - 1st 10 1577 21.7 26.7 33.0 70.1 191.8

Field 7 - 2nd 12 1424 19.6 32.1 39.3 N/A 151.3

Field 7 - 3rd 17 1404 20.7 30.9 37.6 65.7 160.3

Field 8 - 2nd 5 1245 20.7 28.0 35.5 N/A 176.1

Field 8 - 2nd 5 1245 21.7 24.4 31.0 N/A 209.7

Field 8 - 3rd 10 1417 21.1 30.3 36.3 68.0 167.4

Field 10 - 1st 11 1644 23.3 26.0 31.7 70.9 201.7

Field 10 - 2nd 15 1387 21.4 30.2 36.9 N/A 165.0

Field 10 - 3rd 12 1392 22.3 29.6 36.8 67.1 166.8

Total/Avg 10 1415 21.8 28.7 34.9 68.8 179.5

Table 2. Average values, based on laboratory analysis, of each bale by field and cutting for 2013.

Field

No.

samples

DM

%

Moisture

%

CP

%

ADF

%

NDF

% RFV

Bale

Weight

CJ-1&2 123 92.9 7.1 20.5 28.7 33.2 187.3 1406

CJ-10 8 92.5 7.5 20.7 26.6 31.3 205.0 1513

CJ-7 37 91.2 8.8 21.2 26.8 31.6 201.1 1492

Cutting 1 - Total/Avg 168 92.5 7.5 20.7 28.2 32.8 191.2 1428

CJ-1 78 93.5 6.5 19.7 32.3 38.1 156.5 1310

CJ-10 46 91.8 8.2 21.4 29.1 35.4 174.5 1398

CJ-7 18 92.4 7.6 20.1 32.1 38.1 157.5 1402

CJ 8A 24 93.1 6.9 18.3 30.6 38.1 159.9 1388

CJ 8B 2 92.5 7.5 22.0 24.5 29.1 223.0 1565

Cutting 2 - Total/Avg 168 92.8 7.2 20.0 31.1 37.3 162.8 1358

CJ-1&2 109 92.2 7.8 21.2 28.8 33.8 183.6 1465

CJ-10 49 92.2 7.8 20.7 29.3 34.7 177.5 1412

CJ-7 37 92.4 7.6 20.2 29.6 35.1 175.0 1453

CJ 8A 15 92.6 7.4 18.2 30.6 38.2 159.2 1461 Cutting 3 -

Total/Avg 210 92.3 7.7 20.7 29.2 34.6 178.9 1450

Overall Total/Avg 546 92.5 7.5 20.5 29.5 34.8 177.7 1415

Page 6: Dry Matter Density and Relative Feed Value of a …harvesttec.com/wp-content/uploads/2016/01/USU-final...1 Dry Matter Density and Relative Feed Value of a Sensor Developed by Harvest

6

Table 3. Average and SD of moisture % based on laboratory analysis compared with baler

reading. Values were obtained using bale identification number and match from the two lists.

Field No. samples Lab Moist. % Lab SD Baler Moist. % Baler SD

CJ 8A 23 6.9 0.95 15.8 13.32

CJ-1 78 6.5 0.38 9.5 3.64

CJ-10 127 8.0 0.30 13.4 3.30

CJ-7 47 7.7 0.41 16.2 2.97

CJ-1&2 108 7.8 0.84 13.5 6.88

Total/Avg 383 7.5 0.80 13.1 5.93

Figure 1. Moisture % of samples based on laboratory analysis and scissor cut by field and

cutting.

0.0

2.0

4.0

6.0

8.0

10.0

12.0

14.0

16.0

18.0

Moisture %

Lab Moist% Scissor Moist. %

Page 7: Dry Matter Density and Relative Feed Value of a …harvesttec.com/wp-content/uploads/2016/01/USU-final...1 Dry Matter Density and Relative Feed Value of a Sensor Developed by Harvest

7

Figure 2. Crude protein % of samples based on laboratory analysis and scissor cut by field and

cutting.

Figure 3. Acid detergent fiber % of samples based on laboratory analysis and scissor cut by field

and cutting.

0.00

5.00

10.00

15.00

20.00

25.00

30.00

CP%

Lab CP% Scissor CP %

0.00

5.00

10.00

15.00

20.00

25.00

30.00

35.00

ADF%

Lab ADF% Scissor ADF %

Page 8: Dry Matter Density and Relative Feed Value of a …harvesttec.com/wp-content/uploads/2016/01/USU-final...1 Dry Matter Density and Relative Feed Value of a Sensor Developed by Harvest

8

Figure 4. Neutral detergent fiber % of samples based on laboratory analysis and scissor cut by

field and cutting.

Figure 5. Relative feed value of samples based on laboratory analysis and scissor cut by field and

cutting.

0.005.00

10.0015.0020.0025.0030.0035.0040.0045.00

NDF%

Lab NDF% Scissor NDF %

0.00

50.00

100.00

150.00

200.00

250.00

RFV%

Lab RFV% Scissor RFV %

Page 9: Dry Matter Density and Relative Feed Value of a …harvesttec.com/wp-content/uploads/2016/01/USU-final...1 Dry Matter Density and Relative Feed Value of a Sensor Developed by Harvest

9

Figures 6 – 9. Comparisons between laboratory analysis and scissor cut for Field 10 -3. This

particular field and cutting was selected because it had 5 scissor cut analyses which allows for

standard deviations to be calculated to determine if there are differences between method of

analysis.

20.65 22.30

0.00

5.00

10.00

15.00

20.00

25.00

1

CP%

Lab Analysis Scissor Cut

34.72 36.80

0.00

10.00

20.00

30.00

40.00

50.00

1

NDF%

Lab Analysis Scissor Cut

29.33 29.60

0.00

10.00

20.00

30.00

40.00

1

ADF%

Lab Analysis Scissor Cut

177.50 166.80

0.00

50.00

100.00

150.00

200.00

1

RFV%

Lab Analysis Scissor Cut

Page 10: Dry Matter Density and Relative Feed Value of a …harvesttec.com/wp-content/uploads/2016/01/USU-final...1 Dry Matter Density and Relative Feed Value of a Sensor Developed by Harvest

10

Figure 10. Relative Feed Value (RFV) for each bale of hay from one field at one cutting and the

overall average for that field at that cutting (solid line). Average for all bales was 184 (SD =

14.4) and the values ranged from a maximum of 232 to a minimum of 140 (difference of 92).

Figure 11. Laboratory moisture % compared with baler moisture % from a subset of 383 bale

samples. A perfect correlation would be the line with squares.

120

140

160

180

200

220

240

1 5 9

13

17

21

25

29

33

37

41

45

49

53

57

61

65

69

73

77

81

85

89

93

97

10

1

10

5

RFV

Avg. CJ-1&2, Cutting 3 CJ-1&2, Cutting 3

y = 2.5894x - 7.2637 R² = 0.2507

0

5

10

15

20

25

30

0.0 2.0 4.0 6.0 8.0 10.0 12.0 14.0 16.0

Bal

er

Mo

ist.

%

Lab Moist. %

Lab Moisture% vs Baler Moisture%

Baler Perfect Correlation

Linear (Baler) Linear (Perfect Correlation)

Page 11: Dry Matter Density and Relative Feed Value of a …harvesttec.com/wp-content/uploads/2016/01/USU-final...1 Dry Matter Density and Relative Feed Value of a Sensor Developed by Harvest

11

Figure 12. Correlation between bale dry matter (lb) based on laboratory analysis and bale dry

matter (lb) using sensor moisture that was converted using a correction factor. The trendline

used was a linear relationship.

Figure 13. ADF% compared with NDF% of bale samples based on laboratory analysis.

y = 0.663x + 476.63 R² = 0.5097

800

900

1000

1100

1200

1300

1400

1500

1600

1700

800 900 1000 1100 1200 1300 1400 1500

Bal

e D

M (

lab

)

Corrected Bale DM, lb

Bale DM, lb

y = 1.1993x - 0.4855 R² = 0.8979

20.00

25.00

30.00

35.00

40.00

45.00

50.00

55.00

20.00 25.00 30.00 35.00 40.00 45.00

ND

F%

ADF%

ADF vs NDF

Linear (ADF vs NDF)

Page 12: Dry Matter Density and Relative Feed Value of a …harvesttec.com/wp-content/uploads/2016/01/USU-final...1 Dry Matter Density and Relative Feed Value of a Sensor Developed by Harvest

12

Figure 14. RFV compared with NDF%, ADF% and CP% in bale samples based on laboratory

analysis.

Figure 15. RFV compared with NEL in bale samples based on laboratory analysis.

y = -0.0003x2 + 0.1817x - 0.7226 R² = 0.6529

y = 0.0007x2 - 0.3827x + 82.079 R² = 0.9927

y = 0.0006x2 - 0.3291x + 68.811 R² = 0.9422

0.00

10.00

20.00

30.00

40.00

50.00

60.00

0.00 50.00 100.00 150.00 200.00 250.00

RFV

RFV vs NDF%, ADF% and CP%

CP%

NDF%

ADF%

Poly. (CP%)

Poly. (NDF%)

Poly. (ADF%)

y = -7E-06x2 + 0.0038x + 0.2453 R² = 0.8707

0.40

0.45

0.50

0.55

0.60

0.65

0.70

0.75

0.80

90.00 110.00 130.00 150.00 170.00 190.00 210.00 230.00 250.00

NEL

RFV%

RFV vs NEL

Page 13: Dry Matter Density and Relative Feed Value of a …harvesttec.com/wp-content/uploads/2016/01/USU-final...1 Dry Matter Density and Relative Feed Value of a Sensor Developed by Harvest

13

Figure 16. RFV compared with TDN% of bale samples based on laboratory analysis.

Figure 17. NDF% compared with CP%, by cutting, based on laboratory analysis.

y = 0.2777x2 - 29.561x + 907.59 R² = 0.8783

0.00

50.00

100.00

150.00

200.00

250.00

50.00 55.00 60.00 65.00 70.00 75.00

RFV

TDN%

RELATIVE FEED VALUE (RFV)

y = -1.5181x + 64.181 R² = 0.7522

y = -1.5335x + 67.969 R² = 0.7313

y = -1.4961x + 65.56 R² = 0.8381

22.00

27.00

32.00

37.00

42.00

47.00

52.00

14.00 16.00 18.00 20.00 22.00 24.00 26.00

ND

F%

CP%

NDF% vs CP% by Cutting 1st

2nd

3rd