Dual-axis, Duo LateralPosition Sensitive Detectors
Robin Dienhoffer
State University of New York at Oswego
Advisor: Dr. Sherry Yennello
Cyclotron Institute, Texas A&M University
![Page 2: Dual-axis, Duo Lateral Position Sensitive Detectors Robin Dienhoffer State University of New York at Oswego Advisor: Dr. Sherry Yennello Cyclotron Institute,](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d3a5503460f94a14ffb/html5/thumbnails/2.jpg)
Outline
Particle Detectors Silicon Detectors Position Sensitive Detectors now New Position Sensitive Detector
![Page 3: Dual-axis, Duo Lateral Position Sensitive Detectors Robin Dienhoffer State University of New York at Oswego Advisor: Dr. Sherry Yennello Cyclotron Institute,](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d3a5503460f94a14ffb/html5/thumbnails/3.jpg)
Particle Detectors Materials and forms used
Solid, liquid & gasVarious elements
• Germanium, Cesium iodide, silicon, plastic… Various Attributes
Stopping power• ΔE vs. E plots
Energy and position resolutionCostEase of handling
Combining Detectors
![Page 4: Dual-axis, Duo Lateral Position Sensitive Detectors Robin Dienhoffer State University of New York at Oswego Advisor: Dr. Sherry Yennello Cyclotron Institute,](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d3a5503460f94a14ffb/html5/thumbnails/4.jpg)
Silicon Detectors
Not overly sensitive to humidity and operates well at room temperatures
Delicate, but not impossible to handle Still expensive, but becoming more
widely used Band Gap
![Page 5: Dual-axis, Duo Lateral Position Sensitive Detectors Robin Dienhoffer State University of New York at Oswego Advisor: Dr. Sherry Yennello Cyclotron Institute,](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d3a5503460f94a14ffb/html5/thumbnails/5.jpg)
Band Gap
Energy difference between valance band and conduction band Silicon is a good material to detect the
particles that we work with To get to the conduction band,
approximately 2.8 eV are needed To get a readable charge, electrons
must be excited enough to “jump” to the conduction band
This charge is then collected by the charge collection strips to be converted to signals
Valance Band
Conduction Band
Band Gap
![Page 6: Dual-axis, Duo Lateral Position Sensitive Detectors Robin Dienhoffer State University of New York at Oswego Advisor: Dr. Sherry Yennello Cyclotron Institute,](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d3a5503460f94a14ffb/html5/thumbnails/6.jpg)
What are PSDs?
PSD stands for Position Sensitive Detector
Collects deposited charge in such a way that the position of the particle can be determined
This will allow us a more thorough understanding of the underlying reaction mechanism that produced the charged particles
![Page 7: Dual-axis, Duo Lateral Position Sensitive Detectors Robin Dienhoffer State University of New York at Oswego Advisor: Dr. Sherry Yennello Cyclotron Institute,](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d3a5503460f94a14ffb/html5/thumbnails/7.jpg)
FAUST
The Forward Array Using Silicon Technology
Currently uses discrete detectorsPosition information
restricted to which detector was hit
Hoping to upgrade to PSDs, giving better position resolution
![Page 8: Dual-axis, Duo Lateral Position Sensitive Detectors Robin Dienhoffer State University of New York at Oswego Advisor: Dr. Sherry Yennello Cyclotron Institute,](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d3a5503460f94a14ffb/html5/thumbnails/8.jpg)
Some PSDs in Use
Pixilated DetectorsStrip detectors that use separated pads of
Silicon to detect where the particles hitThe number of channels of electronics limit
the feasibility of this option Resistive Detectors
Dual-axisTetra-lateralDADL
![Page 9: Dual-axis, Duo Lateral Position Sensitive Detectors Robin Dienhoffer State University of New York at Oswego Advisor: Dr. Sherry Yennello Cyclotron Institute,](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d3a5503460f94a14ffb/html5/thumbnails/9.jpg)
Dual-axis
Signal pulled from the edge centers
This results in circular pulling that must be adjusted through correcting algorithms
![Page 10: Dual-axis, Duo Lateral Position Sensitive Detectors Robin Dienhoffer State University of New York at Oswego Advisor: Dr. Sherry Yennello Cyclotron Institute,](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d3a5503460f94a14ffb/html5/thumbnails/10.jpg)
Dual-axis Tetra Lateral
Dual-axis Tetra lateral detector collects charge at all four of the corners on one face of the detector
This causes a pin-cushion effect which must, again, be accounted for using correction algorithm
![Page 11: Dual-axis, Duo Lateral Position Sensitive Detectors Robin Dienhoffer State University of New York at Oswego Advisor: Dr. Sherry Yennello Cyclotron Institute,](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d3a5503460f94a14ffb/html5/thumbnails/11.jpg)
Four Corner PSD – Dr. Tribble’s Group
Resistive sheet of Si on front and a non-resistive sheet on the back
resistive strips on front edges correct pincushion effect
Five signals- four on front and one on back (total energy)
![Page 12: Dual-axis, Duo Lateral Position Sensitive Detectors Robin Dienhoffer State University of New York at Oswego Advisor: Dr. Sherry Yennello Cyclotron Institute,](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d3a5503460f94a14ffb/html5/thumbnails/12.jpg)
Dual-axis, Duo Lateral
Dual-axis, duo lateral PSDs collect charge in a way which give position from both the front (horizontally) and the back (vertically). This method will lead to much less distortion than in previous models.
![Page 13: Dual-axis, Duo Lateral Position Sensitive Detectors Robin Dienhoffer State University of New York at Oswego Advisor: Dr. Sherry Yennello Cyclotron Institute,](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d3a5503460f94a14ffb/html5/thumbnails/13.jpg)
F1
F2
B1
B2
Guide wires for
charge
Front 1 Back 2
Front 2Back 1
Charge collecting strips
![Page 14: Dual-axis, Duo Lateral Position Sensitive Detectors Robin Dienhoffer State University of New York at Oswego Advisor: Dr. Sherry Yennello Cyclotron Institute,](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d3a5503460f94a14ffb/html5/thumbnails/14.jpg)
Detecting Location
Charge Collection
Guide Wires
L1 L2
L2L1
L2L1
Edge 1 Edge 2
When L1 = L2, both edges receive the equal charge, or Q1 = Q2When L1 > L2, Edge 1 receives less charge, or Q1 < Q2When L1 < L2, Edge 1 receives more charge, or Q1 > Q2
![Page 15: Dual-axis, Duo Lateral Position Sensitive Detectors Robin Dienhoffer State University of New York at Oswego Advisor: Dr. Sherry Yennello Cyclotron Institute,](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d3a5503460f94a14ffb/html5/thumbnails/15.jpg)
/\/\/\ /\/\/\ /\/\/\VoV1 V2
Vo = V1 + V2
Another Perspective
![Page 16: Dual-axis, Duo Lateral Position Sensitive Detectors Robin Dienhoffer State University of New York at Oswego Advisor: Dr. Sherry Yennello Cyclotron Institute,](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d3a5503460f94a14ffb/html5/thumbnails/16.jpg)
Settings
Found that biasing the front to 40.0 volts and the back guard ring to 5.0 volts is the most effective to get a clear energy spectra and position graph
Shaping time of 1 μs or 3 μs gives both good energy and position resolution
![Page 17: Dual-axis, Duo Lateral Position Sensitive Detectors Robin Dienhoffer State University of New York at Oswego Advisor: Dr. Sherry Yennello Cyclotron Institute,](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d3a5503460f94a14ffb/html5/thumbnails/17.jpg)
228Th Energy Spectra
5.432
5.686
6.051
6.288
6.778
8.784
![Page 18: Dual-axis, Duo Lateral Position Sensitive Detectors Robin Dienhoffer State University of New York at Oswego Advisor: Dr. Sherry Yennello Cyclotron Institute,](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d3a5503460f94a14ffb/html5/thumbnails/18.jpg)
228Th Position Graph
![Page 19: Dual-axis, Duo Lateral Position Sensitive Detectors Robin Dienhoffer State University of New York at Oswego Advisor: Dr. Sherry Yennello Cyclotron Institute,](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d3a5503460f94a14ffb/html5/thumbnails/19.jpg)
Beam Time: Ag + natAuΔE-E Graph
Z = 1
Z = 2
Z = 3 Proton
Deuteron
Triton
![Page 20: Dual-axis, Duo Lateral Position Sensitive Detectors Robin Dienhoffer State University of New York at Oswego Advisor: Dr. Sherry Yennello Cyclotron Institute,](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d3a5503460f94a14ffb/html5/thumbnails/20.jpg)
Beam Time: Ag + natAuPosition Graph
![Page 21: Dual-axis, Duo Lateral Position Sensitive Detectors Robin Dienhoffer State University of New York at Oswego Advisor: Dr. Sherry Yennello Cyclotron Institute,](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d3a5503460f94a14ffb/html5/thumbnails/21.jpg)
Hole Ratios
Mask Measured
3 mm 250 Chan.
2.5 mm 200 Chan.
1.5 mm 125 Chan.
Red = 1.20, 1.25 Green
Red = 0.50, 0.50Blue
Green = 1.67, 1.60 Blue
![Page 22: Dual-axis, Duo Lateral Position Sensitive Detectors Robin Dienhoffer State University of New York at Oswego Advisor: Dr. Sherry Yennello Cyclotron Institute,](https://reader036.vdocument.in/reader036/viewer/2022062714/56649d3a5503460f94a14ffb/html5/thumbnails/22.jpg)
Future Use
This PSD design is currently being optimized for the future FAUST upgrade
Micron Semiconductor is offering this detector to the scientific community at large