Dual-Energy CT Applications in Radiation Therapy THE UNIVERSITY OF WISCONSIN–MADISON - Jessica Miller 1

Upload: others

Post on 19-Apr-2020

5 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

Dual-Energy CT Applications in Radiation Therapy

THE UNIVERSITY OF WISCONSIN–MADISON

- Jessica Miller

1

Page 2: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

• Funding provided by Siemens Medical

2

Disclosures 

Page 3: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

• General principles of dual‐energy CT

• Technical approaches to dual‐energy CT

• Potential applications and challenges of dual‐energy CT in radiation oncology

Learning objectives

3

Page 4: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

What is dual‐energy CT?

80 kVp – Low Energy Image 140 kVp – High Energy Image

4

Page 5: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

• Above k‐edge energy, a linear attenuation can be approximated by the sum of PE and CS (basis pairs)

• + • Where  ∝ and  ∝

• Alternatively, the linear attenuation coefficient for an arbitrary material can be represented as a weighted sum of two independent material’s attenuation coefficients

• + 

Basis pair decomposition

5

Page 6: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

Basis pair decomposition

6

Szczykutowicz, T. in press. “Dual‐Energy and Spectral Imaging.” Comprehensive Biomedical Physics

Page 7: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

Basis pair decomposition

7

Szczykutowicz, T. in press. “Dual‐Energy and Spectral Imaging.” Comprehensive Biomedical Physics

Page 8: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

Dual source DECT technology

8

• Two x‐ray tubes separated by 90 degrees

• Dual‐Energy CT FOV of 33 cm

• Filters can be optimized for spectral separation

Image courtesy of SIEMENS

Page 9: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

• Sequential CT scans: • 140 kV • 80 kV

• Creates a low‐ and high‐energy spectra, sequentially

Image courtesy of SIEMENS

Sequential scans 

140 kV

80 kV

return to starting position

9

Page 10: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

Fast kVp switching

10

• 0.5 ms switching between 80 and 140 kVp to acquire interleaved projection data

• Requires fast generator response and detector response

Image courtesy of GE Healthcare

Page 11: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

Dual‐layer detector

11Cynthia H. McCollough; Shuai Leng; Lifeng Yu; Joel G. Fletcher; Radiology 2015, 276, 637‐653.

• ”Sandwich” scintillation detectors:

• Low energy data collected from the top/proximal layer• High energy data acquired from the bottom/distal layer

Page 12: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

• A removable split‐filter composed of gold (Au) and tin (Sn) which filters a 120 kV x‐ray beam

• Creates a low‐ and high‐energy spectra simultaneously

Images courtesy of SIEMENS

Single‐source DECT TwinBeam system

12

Page 13: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

Potential radiation therapy applications for DECT

Mixed – 120 kVp equivalent  

13

Page 14: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

Potential radiation therapy applications for DECT

True Contrast Image Virtual Non‐contrast

Iodine Map Rho/Z Map

14

Page 15: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

40 keV 55 keV

77 keV 190 keV

Potential radiation therapy applications for DECT

15

Page 16: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

Virtual monoenergetic reconstructions (VMI)

Roele, E.D., Timmer, V.C.M.L., Vaassen, L.A.A. et al. Curr Radiol Rep (2017) 5: 19.

Page 17: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

Metal Artifact Reduction at Higher VMI Energies

Shima Aran, Laleh Daftari Besheli, Musturay Karcaaltincaba, Rajiv Gupta, Efren J. Flores and Hani H. AbujudehAmerican Journal of Roentgenology 2014 202:4, W314‐W324

17

Page 18: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

Improved dose calculation

18

Radiation therapy applications:

Page 19: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

19

Wouter van Elmpt et al. Radiotherapy and Oncology 2016, 3119, 137‐144.

Improved dose calculations

• Brachytherapy

• Protons

• External Beam – Photon Therapy

N. Hudobivnik et al. Med. Phys. 2016, 43, 495.

Page 20: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

Dose calculations – virtual non‐contrast (VNC) images

True Contrast Image Virtual Non‐contrast

20

Page 21: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

21

HU difference map (Mixed – VNC)

Dose calculations – virtual non‐contrast (VNC) images

Images courtesy of Dr. Huang‐Vredevoogd, University of Wisconsin

Page 22: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

22

• Ideal plan – VNC plan• Ideal plan – No override plan

Dose calculations – virtual non‐contrast (VNC) images

Images courtesy of Dr. Huang‐Vredevoogd, University of Wisconsin

Page 23: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

Tumor identification, characterization and delineation

23

Radiation therapy applications:

Page 24: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

Cyst or Carcinoma?

Renal call carcinomaHyperattenuating cyst

Mukta D. Agrawal; Daniella F. Pinho; Naveen M. Kulkarni; Peter F. Hahn; Alexander R. Guimaraes; Dushyant V. Sahani; RadioGraphics 2014, 34, 589‐612. 24

Page 25: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

Calcium or hemorrhage?

Ranliang Hu; Laleh DaftariBesheli; Joseph Young; Markus Wu; Stuart Pomerantz; Michael H. Lev; Rajiv Gupta; Radiology 2016, 280, 177‐183. 25

Page 26: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

Material decomposition – iodine map

Roele, E.D., Timmer, V.C.M.L., Vaassen, L.A.A. et al. Curr Radiol Rep (2017) 5: 19.

Page 27: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

Material decomposition – iodine map

Mukta D. Agrawal; Daniella F. Pinho; Naveen M. Kulkarni; Peter F. Hahn; Alexander R. Guimaraes; Dushyant V. Sahani; RadioGraphics 2014, 34, 589‐612.27

Page 28: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

Tumor delineation

Mukta D. Agrawal; Daniella F. Pinho; Naveen M. Kulkarni; Peter F. Hahn; Alexander R. Guimaraes; Dushyant V. Sahani; RadioGraphics 2014, 34, 589‐612.

28

Page 29: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

Tumor delineation for pancreatic cancer

29

Page 30: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

Tumor delineation for pancreatic cancer

30

Page 31: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

Tumor delineation 

31

Page 32: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

Treatment response assessment 

32

Radiation therapy applications:

Page 33: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

33

Treatment response with DECT

Xu Dai et al. European Journal of Radiology (2013) 82: 327‐334

Apfaltrer et al. Invest Radiol. (2012) 42:(1): 65‐70

Page 34: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

Normal tissue segmentation

34

Radiation therapy applications:

Page 35: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

Normal tissue delineation

35

Postma et al. Dual‐Energy CT: What the Neuroradiologist Should Know. Current Radiology Reports. 2015;3(5):16.

Supratentorialwhite matter/basal ganglia 

Posterior fossa

50 keV to 70 keV

Higher energies

Page 36: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

Functional normal tissue segmentation and toxicity

36

Radiation therapy applications:

Page 37: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

Material decomposition – xenon map & iodine map

37

Zhang, L.J., Zhou, C.S., Schoepf, U.J. et al. EurRadiol (2013) 23: 2666

Ventilation

Perfusion

Xenon inhalation

Iodine injection

Page 38: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

38

Houda Bahig et al., International Journal of Rad. Onc., Biology, Physics. V99,Issue 1, Paes 334‐343 (Oct 2017). 

Dose calculation accounting for functional lung 

Page 39: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

Bone Marrow

39

Fornaro et al. Dual‐ and multi‐energy CT: approach to functional imaging. Insights into Imaging. 2011;2(2):149‐159.

Taiki Magome et al., Int J Radiation OncolBiol Phys, Vol. 96, No. 3, pp. 679‐687, 2016 

Page 40: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

40

Sarah McGuire et al., Radiotherapy and Oncology, Vol. 99, No. 1, pp. 49‐54, 2011 

Bone Marrow

Page 41: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

Conclusions

41

• Improving dose calculation

• Tumor identification and delineation

• Treatment response

• Normal tissue segmentation

• Functional normal tissue toxicities 

Challenges in commissioning and quality assurance of DECT systems for radiation therapy applications

Page 42: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

Thank you

42

Page 43: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

A. Visualization of bone marrow edema via virtual calcium removal

B. Enhancement of CNR between tumor and healthy tissue via iodine uptake 

C. Visualization of tissue metabolic activity via glucose uptake

D. Improvement of dose calculation accuracy with effective atomic number information

Which of the following is NOT a current application of Dual‐energy CT?

43

van Elmpt W, Landry G, Das M, Verhaegen F. Dual energy CT in radiotherapy: Current applications and future outlook. Radiother Oncol. 2016 Apr;119(1):137‐44.

McCollough CH, Leng S, Yu L, Fletcher JG. Dual‐ and multi‐energy CT: principles, technical approaches, and clinical applications. Radiology 2015; 276: 637–53.

Page 44: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

44

Page 45: Dual-Energy CT Applications in Radiation Therapychapter.aapm.org/nccaapm/z_meetings/2017-11-10/04...Dual-Energy CT Applications in Radiation Therapy ... Dual source DECT technology

45

Houda Bahig et al., International Journal of Rad. Onc., Biology, Physics. V99,Issue 1, Paes 334‐343 (Oct 2017).