Echocardiography of Prosthetic Valves
Dr. Tehrani
![Page 2: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/2.jpg)
Different Types of Valves
Homografts (allograft)Cadaveric human aortic and pulmonary valves
Heterograft (Xenograft)Prosthetic Valves
Bioprosthetic valvesPig aortic valve Bovine pericardial (other)
MechanicalUrethane ball in a cageSingle or multiple discs
![Page 3: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/3.jpg)
Homografts (allograft)
![Page 4: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/4.jpg)
Homografts
Homograft ValvesHarvested soon after death w/ the endothelium still viablePreparation for implantation
Storage in ABXCryopreservation (more recently)
No anticoagulationLow incidence of endocarditisFailure due to gradual aortic incompetence
![Page 5: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/5.jpg)
Homografts
PositionMitral
Fitted w/ stent, not proved successful (high failure rate at 5 years)Stentless grafts not an option for MVR
AorticStentless
Subcoronary Root Replacement
![Page 6: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/6.jpg)
Echocardiography of Stentless Aortic Homografts
Doppler flow characteristics similar to native valve.Only 2-D evidence: Increased Echo intensity, and Thickness of aortic annulus.
![Page 7: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/7.jpg)
Stentless Heterografts (Xenograft)
![Page 8: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/8.jpg)
Stentless Heterografts (Xenograft)
Same utility as allografts for AVR:
Subcoronary implantation, andRoot replacement
Advantage over allografts is wider availabilityDurability is at least as good as allografts
![Page 9: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/9.jpg)
Prosthetic Valves
![Page 10: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/10.jpg)
Prosthetic Valves
All ProstheticProsthetic valves valves have a sewing ring anchored to the native tissue with sutures
The occluding portion of the valve:Tissue leaflets Bio-ProstheticProstheticSingle or multiple discs/ Urethane ball in a cage Mechanical Prosthetic
Bio-Prosthetic and Mechanical Prosthetic
![Page 11: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/11.jpg)
Bio-prosthetic Valves
![Page 12: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/12.jpg)
Bioprosthetic Valves
Two types, of occluding mechanism1.Porcine aortic
valve (the valve size of the biggest pig is limiting)
Hancock, andCarpentier-Edwards bioprosthesis
Carpentier-Edwards bioprosthesis
![Page 13: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/13.jpg)
Bioprosthetic valves
2. Bovine Pericadium leaflets are shaped to size.
More choices
Echocardiographiaclly these two valves types are indistinguishable.
Ionescu-Shiley
(1976)
![Page 14: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/14.jpg)
Bioprosthetic valvesMitral Position
2-D ECHOCARDIOGRAPHIC APPEARANCE
![Page 15: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/15.jpg)
Bioprosthetic valvesAortic Position
2-D ECHOCARDIOGRAPHIC APPEARANCE
![Page 16: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/16.jpg)
Bioprosthesis
Used extensively in a variety of sites:AorticMitralTricuspid
Advantage:Low thrombogenicity => No anticoagulation
![Page 17: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/17.jpg)
Bioprosthesis
Disadvantages:Less durable than mechanical prosthesisMitral position worse
Due to greater backpressure gradientDysfunction:
Leaflet thickening, and Ca++ Fracture, tears, or progressive stenosis
In vivo, roughly 10% of normal bioprosthetic valves have some leakage.
![Page 18: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/18.jpg)
Overview of Various Devices
Bio-Prosthetic Valves
Mechanical Prosthetic Valves
![Page 19: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/19.jpg)
Mechanical Vlaves
Ball-and-Cage ValvesTilting disc Prosthesis
Single diskBileaflet
![Page 20: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/20.jpg)
Mechanical Prosthesis
The occluding mechanism dictates both: The echocardiographic appearance of the valve, andThe flow pattern through the valve
To assess performance, the type of valve implanted must be known
![Page 21: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/21.jpg)
Ball-and-Cage Valves
First implanted by starr and Harken in 1960.
![Page 22: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/22.jpg)
Ball-and-Cage Valves
Opening and closure of the ball-valve
![Page 23: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/23.jpg)
Ball-and-Cage Valves
Axisymmetric flow around the valve.Stagnant flow in the shadow of the ball.
![Page 24: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/24.jpg)
Ball-and-Cage Valves
Doppler assessment at the margins of the ball
![Page 25: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/25.jpg)
Ball-and-Cage Valves
M-Mode assessment of Ball-Cage Valve
![Page 26: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/26.jpg)
Ball-and-Cage Valves
DurableMitral position
Satisfactory profile with the largest size (34 or 32 mm diameter devices)Can affect the interventricular septum
Aortic positionSmall prosthesis required, which can be associated with significant gradient
Regurgitation limited to closure backflow.
![Page 27: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/27.jpg)
Tilting Disc Prosthesis
All essentially similar consisting of Circular prosthetic material, andOne or two hinged and mobile disc(s)
Disc attachment to the ring is eccentric
Closure occurs by backpressure on the largest portion of the disk
![Page 28: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/28.jpg)
Single Disc Prosthesis
Single Disc devices:Hall-Medtronics monostrutBjork-Shiley
Opening arch is 55-70 degreesFlow orifice:
Major and minor flow orificesStreamlines of flow passing through the sewing ring and then laterally out and around the prosthetic disc
![Page 29: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/29.jpg)
Single Disc Prosthesis
Bjork-Shiley StandardConvex-concave
Many other variations in the marketAll of these devices have a zone of stagnation behind the disc thrombus formation
![Page 30: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/30.jpg)
Single Disc Prosthesis
Bjork-Shiley in the Mitral position
![Page 31: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/31.jpg)
Single Disc Prosthesis
Leak around: Central strut
Dominant jetBetween the occluding disc and sewing ring.
Two smaller peripheral jets
Normal hemodynamicsReg.Frac. approx. 12%
Tachycardia, and low outputReg.Frac. upto 37%
![Page 32: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/32.jpg)
Single Disc Prosthesis
![Page 33: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/33.jpg)
![Page 34: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/34.jpg)
Single Disc Prosthesis
Dysfunction
Gradual ingrowth of fibrous tissue (panus)
Flow obstructionIntermittent sticking of the valve with associated flash pulmonary edema
![Page 35: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/35.jpg)
Bileaflet Mechanical Prosthesis
St. Jude prosthesisThe most commonly used.
Two equal sized semi-circular leaflets attached by a midline hinge.Discs can tilt in excess of 80 degrees, resulting in larger:
Orifice area
![Page 36: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/36.jpg)
Bileaflet Mechanical Prosthesis
St. Jude prosthesisThe most commonly used.
Two equal sized semi-circular leaflets attached by a midline hinge.Discs can tilt in excess of 80 degrees, resulting in larger:
Orifice area
Regurgitant back flow
![Page 37: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/37.jpg)
Bileaflet Mechanical Prosthesis
Regurgitation occurs at the disc marginsThe regurgitant jets converge toward the center of the valve
![Page 38: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/38.jpg)
Bileaflet Mechanical Prosthesis
St. Jude valve in the mitral position.
![Page 39: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/39.jpg)
Imaging of Prosthetic Valves
![Page 40: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/40.jpg)
Special Problems of 2-D Imaging Artificial Valves
Echocardiographs are calibrated to measure distance based on the speed of sound in tissue.
Prosthetic valves have different acoustic properties than tissue. Hence, distortion of:
SizeLocation, andAppearance, of the prosthesis.
![Page 41: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/41.jpg)
Special problems of artificial valves
Intense reverberation, andShadowingLess gain leads to less:
Reverberation, and Shadowing, as well asBetter visualization of non-biologic components of the valve HOWEVER Decreased definition of cardiac
structures
![Page 42: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/42.jpg)
Special problems of artificial valves
First image at normal settings, then Reduce the gain to interrogate the leaflets of Bio-prosthetic valves.
Utilize multiple views.
![Page 43: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/43.jpg)
Prosthetic Valve Pathology
Prosthetic Valve Stenosis Aortic Mitral
Prosthetic Valve Regurgitation Aortic Mitral
![Page 44: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/44.jpg)
G E N E R A L L Y
Prosthetic Stenosis (and Regurgitation) is:
A question of degree,
Not a question of whether.
![Page 45: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/45.jpg)
Prosthetic Valve Stenosis
Determinants of gradients across normal prosthetic valves include:
Valve type, i.e., ManufacturerValve sizeFlow through the valve
Wide range of “Normals”
![Page 46: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/46.jpg)
Aortic Prosthesis Gradients as a Function of Valve TYPE and SIZE
No.21
No.27
Dependence on:
Valve type, and
Size
![Page 47: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/47.jpg)
Gradient as a Function of Valve Type
Normal Dopplar data in patients with various types of prosthetic valves in the Aortic Position
![Page 48: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/48.jpg)
Gradient as a Function of Valve Size
Valve specifications and doppler echocardiographic data in 67 St. Jude medical valves in the Aortic position
Chafizadeh ER, Circ. 83:213, 1991
![Page 49: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/49.jpg)
Gradient as a Function of Flow
No.21
I. Valve type, i.e., Manufacturer
II. Valve size
III. Flow through the valve
![Page 50: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/50.jpg)
Indicies of Valve Stenosis which are Less Flow Depenent
A. Contour of jet velocityB. Doppler velocity indexC. Effective orifice areaD. Valve resistance
![Page 51: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/51.jpg)
A-Contour of the jet velocity
With prosthetic obstruction there is:
Late peaking of the velocity,More rounded contour,Prolonged ejection.
![Page 52: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/52.jpg)
B-Doppler Velocity Index
DVI= Pk VelLVOT/Pk Veljet
Flow independent 0.2 – 0.27 cutoff for critical stenosis
Caveat: Pressure recoveryTo be discussed …
![Page 53: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/53.jpg)
C-Effective Orifice Area
Continuity Eqn.
Caveat: Pressure recoveryTo be discussed …
jetLVOTLVOT TVITVICSAEOA
![Page 54: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/54.jpg)
D-Valve resistance
At cutoff of 280 dynes.sec.cm5, best at differentiating AS, from NL (Zoghbi et al.)
1.33 x SEPGrad.VR
![Page 55: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/55.jpg)
Special Caveats Re:Overestimation of Gradients
Two scenarios:
I. The velocity upstream from the valve is not negligible in application of the Bernoulli Eqn.
Usually in AV when proximal velocity on the LVOT is > 1.5 m/s
2 2
21 4V V P
![Page 56: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/56.jpg)
…Overestimation of Gradients
II. Central acceleration with the St. Jude valve:
Increase of velocities (and gradients) is created at the level of the valve through the smaller central orifice.Most significant with:
High flow states Small valves
![Page 57: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/57.jpg)
…Overestimation of Gradients
Central acceleration with the St. Jude valve:
CW Doppler records these high velocities.
Catheter-derived gradients show pressure recovery at 30mm downstream from the valve.
![Page 58: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/58.jpg)
Indicies which are LessFlow Dependent, BUT…
A. Contour of jet velocityB. Doppler velocity indexC. Effective orifice areaD. Valve resistance
Clearly, both ofThese ParametersWill be Affected byThe PressureRecovery Phenomenon.
![Page 59: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/59.jpg)
Prosthesis-Patient Mismatch
![Page 60: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/60.jpg)
Mismatch
Rahimtoola 1978: “Mismatch is present when the effective
prosthetic valve area, after insertion into the patient, is less than that of a normal human valve.”
By definition: Some such “mismatch” will almost always be
present.
![Page 61: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/61.jpg)
Mismatch
Literature identifies the above as a cut-off for mismatch
22Index /mcm 0.85 EOA
![Page 62: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/62.jpg)
Mismatch
0.007184 x hgt.wt.BSA 0.7250.425
Next locate the publishedin-vivo EOA of the valve used.
This is the EOA that The patient physiologicallyNeeds.
![Page 63: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/63.jpg)
Mismatch
JACC Review Article, 10/2000
Not the company reported data
![Page 64: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/64.jpg)
Prosthetic Valve Pathology
Prosthetic Valve Stenosis Aortic Mitral
![Page 65: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/65.jpg)
Mitral Prosthesis Stenosis
Parameters used for assessment of function:
A. PHT/Area by PHTB. Effective Orifice Area by
continuityC. Mean gradient
![Page 66: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/66.jpg)
Mitral Prosthesis Stenosis
A-PHT/Area by PHTNot expected to yield accurate valve area
The empiric constant of 220 validated for the geometry of rheumatic MS
Useful in longitudinal follow-up of valve FxShould not be used when diastolic filling period is short (fusion of E and A)
TachycardiaLong first degree block
![Page 67: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/67.jpg)
Mitral Prosthesis Stenosis
B-Effective Orifice Area by continuity Eqn.
One underlying assumption is absence of significant AI or MR
Physiologic prosthetic MR 10-30% (Medtronic-Hall, significant central MR, specific design feature less thrombogenic)
![Page 68: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/68.jpg)
Mitral Prosthesis Stenosis
C-Mean gradient, function of:
SizeType of prostheticFlow
Heart rate (should also bereported when evaluating MVA)
![Page 69: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/69.jpg)
Prosthetic Valve Pathology
Prosthetic Valve Stenosis Aortic Mitral
Prosthetic Valve Regurgitation Aortic Mitral
![Page 70: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/70.jpg)
Prosthetic Valve Regurgitatoin
Two issues:
Physiologic v.s. Pathologic regurgitationTTE v.s. TEE for assessment of regurgitation
![Page 71: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/71.jpg)
Prosthetic Valve Regurgitation
Physiologic Regurgitation
Early onsetEarly onset and brief durationReflects backflow from closing movement of occluding device
Tilting disc and bileaflet valves have additional late backflowlate backflow leakage
Intended to reduce risk of thrombosis
![Page 72: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/72.jpg)
Aortic Prosthesis Regurgitation
Criteria similar to grading native valve AI:
Jet widthPHT < 350Holodiastolic flow reversalRegurgitant fraction>40%
![Page 73: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/73.jpg)
Mitral Prosthesis Regurgitation
TTE of limited value in assess MR due to acoustic shadowing of the LA
Doppler findings suggestive of severe MRE wave > 1.9 m.sPISAShort isovolumetic relaxation timeTVILVOT/TVIPr-MV < 0.4
![Page 74: Echocardiography of Prosthetic Valves Dr. Tehrani](https://reader036.vdocument.in/reader036/viewer/2022062305/56649cbe5503460f94983e7c/html5/thumbnails/74.jpg)