Effects of Erythrocyte Aggregation on

Blood Rheology in Regard to Future

Sepsis Diagnosis Application

Darren Burley

Courtney Campagna

Aislinn Harte

Samantha Kelly

Max Spiegelhoff

Advisors:

Prof Raymond Page

Prof Ahmet Can Sabuncu

May 7th, 2020

Page 2: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

2

Contents

Authorship .................................................................................................................................................... 1

Acknowledgements ...................................................................................................................................... 2

Abstract ........................................................................................................................................................ 3

Chapter 1: Introduction ................................................................................................................................ 4

Chapter 2: Background ................................................................................................................................. 5

2.1 Sepsis Overview .................................................................................................................................. 5

2.2 Current Diagnostic Methods ............................................................................................................... 7

2.2.1 Blood Culture ............................................................................................................................... 7

2.2.2 Protein Analysis ............................................................................................................................ 8

2.2.3 Prothrombin and Partial Thromboplastin Time/Platelet Counts ................................................. 8

2.3 Potential Biological Tests for Sepsis Diagnosis ................................................................................... 9

2.3.1 Biomarkers ................................................................................................................................... 9

2.4 Potential Mechanical Tests for Sepsis Diagnosis .............................................................................. 16

2.4.1 Red Blood Cell Deformability ..................................................................................................... 16

2.4.2 RBC Aggregation ........................................................................................................................ 20

2.4.3 White Blood Cells ....................................................................................................................... 23

Chapter 3: Project Strategy ........................................................................................................................ 26

3.1 Client Statement ............................................................................................................................... 26

3.2 Design Requirements ........................................................................................................................ 27

3.2.1 Standards ................................................................................................................................... 27

3.3 Revised Project Statement ................................................................................................................ 29

Chapter 4: Design Process .......................................................................................................................... 30

4.1 Needs Analysis .................................................................................................................................. 30

4.2 Conceptual Designs and Prototype Testing ...................................................................................... 31

4.2.1 Erythrocyte Sedimentation Rate ................................................................................................ 32

4.2.2 Capillary Fill ................................................................................................................................ 34

4.2.3 Vibration Syllectometry ............................................................................................................. 36

4.3 Alternative Designs ........................................................................................................................... 37

Page 3: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

3

4.4 Final Design Selection ....................................................................................................................... 38

4.4.1 Blood Preparation ...................................................................................................................... 38

4.4.2 Micro-ESR Methodology & Materials ........................................................................................ 39

4.4.3 Capillary Fill Methodology & Materials...................................................................................... 40

4.4.4 Laser Syllectometry Methodology & Materials ......................................................................... 42

Chapter 5: Design Verification ................................................................................................................... 44

5.1 Micro-ESR Results ............................................................................................................................. 44

5.2 Capillary Fill Results........................................................................................................................... 46

5.3 Laser Syllectometry Results .............................................................................................................. 47

Chapter 6: Final Design and Validation ..................................................................................................... 55

6.1 Final Design CAD Prototype .............................................................................................................. 55

6.2 Standards .......................................................................................................................................... 56

6.3 Economics ......................................................................................................................................... 57

6.4 Environmental Impact ....................................................................................................................... 58

6.5 Societal Influence .............................................................................................................................. 58

6.6 Political Ramifications ....................................................................................................................... 59

6.7 Ethical Concerns ................................................................................................................................ 59

6.8 Health and Safety .............................................................................................................................. 59

6.9 Manufacturability ............................................................................................................................. 60

6.10 Sustainability ................................................................................................................................... 61

Chapter 7: Discussion/Future .................................................................................................................... 62

7.1 Micro-ESR .......................................................................................................................................... 62

7.1.1 Implications ................................................................................................................................ 62

7.1.2 Limitations .................................................................................................................................. 62

7.1.3 Future ......................................................................................................................................... 63

7.2 Capillary Fill ....................................................................................................................................... 64

7.2.1 Implications ................................................................................................................................ 64

7.2.2 Limitations .................................................................................................................................. 64

7.2.3 Future ......................................................................................................................................... 65

7.3 Syllectometry .................................................................................................................................... 66

Page 4: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

4

7.3.1 Implications ................................................................................................................................ 66

7.3.2 Limitations .................................................................................................................................. 67

7.3.3 Future ......................................................................................................................................... 68

7.4 Device Discussion .............................................................................................................................. 69

Chapter 8: Conclusions and Recommendations ........................................................................................ 70

8.1 Conclusions ....................................................................................................................................... 70

8.2 Recommendations ............................................................................................................................ 70

References .................................................................................................................................................. 72

Appendix A – Interviews ............................................................................................................................ 80

Appendix B – Pros and Cons of Different Testing Methods ...................................................................... 85

Appendix C – Pairwise Analysis ................................................................................................................. 86

Appendix D – Lab Gantt Chart ................................................................................................................... 87

Appendix E – Micro-ESR Data and Analysis ............................................................................................... 88

Appendix F – Capillary Fill Data and Analysis ............................................................................................ 94

Appendix G – Laser Syllectometry Data and Analysis ............................................................................... 99

Page 5: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

1

Authorship

The five project team members, Courtney Campagna, Darren Burley, Aislinn Harte,

Samantha Kelly, and Max Spiegelhoff, contributed equally to the content found in this report.

Extensive collaboration was demonstrated within each section while writing, revising,

compiling, correcting, and finalizing the report.

Page 6: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

2

Acknowledgements

The project team would like to thank the following group of individuals and

organizations for their assistance in completing this project:

Lisa Wall, Laboratory Manager, for her training and support of all lab-based activities

Dr. Michael Puskarich, Dr. Kate Madden, and the doctors at Brigham/Faulkner

hospital for helping the team to better understand sepsis from a certified medical

perspective as well as inform us of current sepsis diagnostic methods used within the

field today

Funding from the Biomedical Engineering Department at WPI

WPI Tinkerbox I & E Program for the additional monetary support

Most importantly, the team would like to thank our MQP advisors Professor Raymond Page,

PhD and Professor Ahmet Can Sabuncu, PhD. Their guidance and expertise over the past

academic year have not only helped enhance the quality of our work but also provided invaluable

professional engineering experience and education that will serve us for years to come.

Page 7: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

3

Abstract

Sepsis is a condition resulting from the overreaction of the body to an infectious agent

and results in millions of deaths worldwide each year. Sepsis has a very short infection-to-

mortality time and can be hard to detect and treat before related afflictions become permanent. A

device that could quickly and effectively diagnose sepsis could be incredibly beneficial to

ensuring proper treatment is given to septic patients. Furthermore, providing alternative methods

of testing than the currently used standard would allow for additional in-depth analysis and the

possibility of increased diagnostic success. This paper works to therefore describe the design of

a prototype, the included tests, and the reason for use in regard to future septic diagnosis.

Page 8: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

4

Chapter 1: Introduction

Every year, at least 1.7 million adults develop sepsis within the United States of America

and almost 270,000 die as a result [1]. Across the globe, almost 30 million people are diagnosed

with sepsis which leads to approximately 6 million deaths per year, averaging a mortality rate

between 15-30%. In severe cases of sepsis, the mortality rate can jump up to anywhere between

40-60%, mainly amongst infants and those in intensive care units [2]. Despite such jarring death

rates, there are currently no methods to effectively treat patients who have developed severe

cases of sepsis; if the disease isn’t treated within the first few days of diagnosis, sepsis can lead

to kidney failure, immune system disruption, and permanent organ damage [1]. Currently, many

methods of testing sepsis are based around clinical observations rather than a direct diagnosis.

According to Dr. Kate Madden, a doctor in critical care medicine at Boston Children’s Hospital:

We have many patients whom we may suspect [have] sepsis, but a large proportion do

not end up having a clear diagnosis. For research and quality improvement purposes, we

use the judgement of the clinical team, usually the attending physician, as well as specific

treatments that go along with sepsis. [Appendix A]

This inability to accurately test for sepsis can make it incredibly difficult to ensure the

patient is getting the proper treatment they need; the quick-acting nature of the disease makes

early diagnosis critical for survival. Therefore, the goal of our project is to create a device that

allows for an accurate and specific diagnosis of sepsis that can be used by the medical

community. To do so, we plan on designing a small device capable of using a sample of the

patient’s blood and performing a series of quick turnaround and low-cost tests that can provide

better clarity for a sepsis-positive prognosis to a physician. Our developed device will be

inexpensive and require little-to-no user training to successfully perform a test. It is our hope that

production and distribution of the device will help medical centers worldwide to better support

patients who could potentially be battling this life-threatening disease.

Page 9: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

5

Chapter 2: Background

2.1 Sepsis Overview

Sepsis is a serious medical condition commonly characterized as an overwhelming

immune response to an infection and is a common complication of medical surgery or poor

treatment of a wound. In many cases, the first stage of sepsis is presented by Systemic

Inflammatory Response Syndrome, or SIRS, in which the body begins to overcompensate for an

infectious agent. According to the America College of Physicians/Society of Critical Care, SIRS

often presents itself through “extreme increases or decreases in body temperature, expedited

heartrate, and increased respiratory rate usually greater than 20 breaths/minute” [3]. Early

treatment of SIRS is a necessity to minimize further medical complications that can occur.

However, while almost all septic patients have SIRS, the presence of SIRS does not

indicate that a person will automatically have sepsis. Sepsis is primarily characterized by the

presence of an infectious agent entering the blood stream, commonly referred to as “blood

poisoning” [4]. In the instances of surgery, or in the event of an open wound occurring on the

body, improper cleaning and management of the wound will increase the probability of an

infectious agent entering the body. Once the agent begins to disperse throughout the blood

stream - and it becomes harder for the immune system to combat the pathogen - is when SIRS

begins to transition to officially being declared as “sepsis”. This usually occurs around 3-5 days

after the initial wound or infection occurs. A timeline of sepsis stages can be seen with Figure 1.

Page 10: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

6

Figure 1: SIRS-sepsis patient timeline

Once immunosuppression begins and the body starts being unable to fight back the

increased concentration of pathogens in the blood, the severity of sepsis increases dramatically.

During this stage, colloquially referred to as “severe sepsis”, the natural physiology of the body

becomes incredibly hampered. From the period of approximately five to fifteen days after the

initial point of infection, patients diagnosed with sepsis can see decreased levels of urination,

more prevalent respiratory issues like pneumonia or shortness of breath, chills or fevers, and

patches of discolored skin [5]. Internally, sepsis can be categorized through decreased blood

platelet count, increased formation of blood clots, abnormal heart pumping, and drops in blood

pressure. Proper treatment of sepsis using antibiotics and vasoactive medications during this

phase is critical; if left unchecked for too long, the mortality rate of sepsis jumps up to almost

50% as many patients become afflicted with “septic shock” [6]. At this stage of the disease,

organ failure is incredibly common. Lactate levels begin to rise within the body and the

combination of decreased blood pressure and blood clotting leads to gangrene within body tissue.

Page 11: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

7

The disease becomes almost untreatable at this point. While “severe sepsis” commonly produces

a negative blood culture (due to the delicacy of the test and blood sensitivity during

immunization), “septic shock” commonly results in a renewal of positive blood cultures [7]. This

is assumed to be because the body develops an inability to overcome infection at this point as

total immune system shutdown occurs, allowing microorganisms to once again prevalently exist

in the blood [8]. Around approximately 28 days from the starting point of infection - or 20 days

from the diagnosis of severe sepsis - the body begins total shutdown and patient death occurs.

According to the UK Sepsis Trust, if a patient is able to recover from sepsis, it can take “up to 18

months before the survivor starts to feel like their normal self” with “around 50 percent of

survivors dealing with post-sepsis syndrome (PPS)” that can include insomnia, fatigue, damaged

organs and a severely hampered mental state [9]. Patients who recover from sepsis also have an

increased chance of becoming affected again due to their previously weakened immune system.

2.2 Current Diagnostic Methods

2.2.1 Blood Culture

Currently, there are no truly effective methods for diagnosing sepsis; the primary method

for sepsis evaluation lies within blood cultures. Being that sepsis primarily spawns from bacteria

entering and dispersing itself throughout the blood stream, samples of blood taken can

commonly be tested for bacteremia that might be present within the blood. In many instances

surrounding blood cultures, tests can take anywhere from 2-5 days to develop and don’t always

guarantee positive outcomes [10]. Furthermore, testing for blood cultures does not directly

signify the specific bacteria or fungi that exists within the body. In the case of sepsis, where the

dispersion of bacteria throughout the blood occurs rapidly from the point of infection, patients

may already be in the stage of severe sepsis before positive blood culture results occur [11]. As

mentioned, blood cultures only work for testing bacteremia found in the blood; in the event that a

fungi or virus was the cause of sepsis, these tests would be completely ineffective. Therefore,

while certainly fundamental for the most part, it is the rather long testing periods coupled with

the delicate specificity that hinder blood cultures from being the superior diagnostic method.

Page 12: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

8

2.2.2 Protein Analysis

Since sepsis starts with a massive inflammation of the body, it is common for increased

concentration of inflammation proteins like C-reactive protein (or CRP) to be present. In many

instances, the increased concentration of C-reactive proteins allows for medical doctors to catch

sepsis during the earlier stages of testing. The problem lies in the fact that protein increases are

not always consistent. In a case study performed at Westminster Hospital in London, 43 out of

49 patients exhibited increased concentration of C-reactive protein but only by a maximum of

about 25%, a relatively small change that could be attributed to other factors like age and gender

[12].

According to Dr. Michael Puskarich, an associate professor at the University of

Minnesota medical school:

Clinically SIRS, "sepsis-3", [and] procalcitonin occasionally have all been tried and they

are not consistently successful. These tools are primarily ways for clinicians to make sure

they are "getting the points" for treatment of sepsis patients and meeting CMS [Centers

for Medicare and Medicaid Services] core measure success but suffers from terrible

usability. For true patient care, good clinicians rely on gestalt and experience combined

with overall lab and diagnostic test results to diagnose sepsis more than any single tool,

which in and of itself is also problematic due to the highly variable nature and clinical

presentation of the disease. Sepsis is also confusing because it exists on a spectrum - it's

not a "yes / no" despite many people wanting it to be. You don't go from routine flu to

"sepsis," it exists on a spectrum with other infectious diseases that the body does or does

not control on its own. [Appendix A]

2.2.3 Prothrombin and Partial Thromboplastin Time/Platelet Counts

Sepsis can have a major impact on the ability of blood to clot within the body.

“Prothrombin” and “Partial Thromboplastin Time” refer to how long it takes for clotting within

the body to form, usually measured within seconds [13]. More specifically, “Prothrombin time”

measures the integrity of the extrinsic system as well as factors common to both systems while

“Partial Thromboplastin Time” measures the integrity of the intrinsic system and the common

components. On average, clotting time generally takes about 20 to 35 seconds for a healthy

person but can be significantly lower within patients who have sepsis, indicating excess blood

clotting [10, 14]. In contrast, another common symptom of sepsis is a decrease in the number of

Page 13: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

9

platelets that enter the body. Platelets are smaller cells that are instrumental in the formation of

clotting blood; if platelet counts are low, red blood cells are less likely to clot at the presence of

an open wound [10]. This vicious contradictory cycle therefore causes increased blood clotting

internally within the body amongst red blood cells rather than at the presence of an open wound

or injury. This is further elaborated on in section “2.4.2 RBC Aggregation”. While such testing

via a d-dimer test (which indicates clot levels in the blood) can be done as early as the starting

stages of SIRS, there is currently not enough evidence to determine if testing for clots is an

effective method for properly diagnosing sepsis. These challenges are once again further

elaborated on within section 2.4.2.

2.3 Potential Biological Tests for Sepsis Diagnosis

2.3.1 Biomarkers

In regard to all possible diagnostic methods for sepsis, biomarkers have found prevalence

for their diversity and adaptability. Biomarkers, as defined by the National Cancer Institute, are

“biological molecule[s] found in blood, other bodily fluids, or tissues that [are] a sign of a

normal or abnormal process, or of a condition or disease” [15]. Of all the possible biomarkers,

the three that emerged to have the most promise in the diagnosis of sepsis (based off extensive

literature review) include procalcitonin, Interleukin- 6 and presepsin.

2.3.1.1 Procalcitonin

Procalcitonin is currently the most studied of the three biomarkers and has been

recognized for its potential in diagnosing sepsis within patients. Procalcitonin (PCT) is a

propetide of calcitonin consisting of 116 amino acids and produced by the parafollicular cells of

the thyroid and parenchymal cells found in the lungs, liver, kidney, adipocytes and muscle

during normal production [16, 17]. Normally, PCT does not make it into the bloodstream

because it is converted to calcitonin leaving the blood concentration at about 0.05 ng/mL [18].

The levels of PCT released in the bloodstream are induced by bacterial endotoxins, cytokines,

lipopolysaccharides and other inflammatory signals. PCT production is reduced when interferon

- signaling proteins produced and released by host cells in the presence of viruses and other

pathogens - are present meaning that, during viral infections, the concentration levels do not

Page 14: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

10

increase significantly or at all [19]. Therefore, PCT can only be used for sepsis brought upon by

bacterial infections and to differentiate between bacterial and viral sepsis [20].

The current methods to measure PCT concentrations in blood use immunoassays. PCT is

the only one of the three biomarkers discussed with an FDA approved assay for sepsis risk

assessment in the ICU. The VIDAS BRAHMS PCT test is an automated test that measures the

procalcitonin concentrations from a patient’s blood sample. It performs enzyme-linked

fluorescent immunoassay (ELFA) on 200 μl samples of whole human blood with a limit of

detection of 0.03ng/mL and a range of 0.05-200ng/mL [21]. ELFA is a newer version of the

enzyme-linked immunosorbent assay (ELISA). The main difference is that it allows for a much

shorter testing time. All steps can be completed, and results are produced in about twenty

minutes, though it is noted that this test should be used in conjunction with other laboratory tests

for more accurate results [20]. Currently, this device is being sold as a kit including all the

reagents needed and the machine needed to carry out the tests [21].

Many research studies have found that PCT allows for the early detection of sepsis. PCT

level changes are detectable between two to four hours after inflammation begins and will reach

their peak concentration at about fourteen hours, remaining elevated for about twenty-four hours

after the inflammatory stimulus [17]. This allows for quick testing of the patient which is a

necessity due to the rapid attack of sepsis on the body. According to a study published by the

American Thoracic Journal, there were significant differences between the varying severities

(SIRS, sepsis, severe sepsis and septic shock) and the concentration of PCT in the plasma of

those patients. The study was conducted with 78 patients -18 of whom were diagnosed with

SIRS, 14 with sepsis, 21 with “severe sepsis”, and 25 with “septic shock” [22]. The study found

that the PCT concentrations for SIR, sepsis, “severe sepsis”, and “septic shock” “(ng/mL) were

0.6 (0 to 5.3) for SIRS; 3.5 (0.4 to 6.7) for sepsis, 6.2 (2.2 to 85) for severe sepsis; and 21.3 (1.2

to 654) for septic shock (p < 0.001)” [22]. By being able to determine the severity of the patient’s

condition, a recommendation on treatment will be more accurate. As seen in Figure 2, based off

of the PCT concentration in the blood sample, antibiotics use is either encouraged or

discouraged. Antibiotics, if prescribed when unnecessary, may cause additional harm, delaying

proper treatment and simultaneously increasing resistance of antibiotics among present bacteria

[23]. The same study found that procalcitonin had a specificity of 78%, a sensitivity (the

Page 15: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

11

proportion of people who test positive relative to those who have the disease) of 97%, and a

positive and negative accuracy predictive value (which determines how accurate positive and

negative tests outcomes are) of 94% and 88% respectively [22]. This reliability to differentiate

between severities allows doctors to decide on the type and aggressiveness of their treatment,

hopefully saving those patients.

Figure 2: Shows the different severities of illness related to sepsis and the associated PCT blood concentration

values. Also illustrates what values signify the use of antibiotic treatment [18].

Page 16: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

12

Though many studies show the reliability of PCT, concentrations can be increased in the

absence of sepsis. According to a study done by the Journal of Applied Laboratory Medicine,

PCT concentrations can increase in the blood:

In many clinical settings in the absence of infection, including after major surgery,

transplant, trauma, and severe burns, prolonged or severe cardiogenic shock, severe organ

perfusion anomalies, autoimmune disorders, malignancies and metastasis, noninfectious

systemic inflammation, chronic kidney disease (CKD), and physiologically in the neonate

[24].

Fungal and malarial infections (along with certain medications) can also increase PCT

concentrations by simulating the release of cytokines [25]. A meta-analysis study in 2007 found

that the mean values of sensitivity and specificity were around 71% meaning that the diagnostic

value of sepsis was low and that it cannot reliably differentiate sepsis from other conditions [22].

Not only can different diseases increase PCT concentrations but so can the age of patients;

neonates’ baseline PCT levels are higher than that of adults during the first 48 hours of life [25].

These differences in PCT concentrations must be studied further to determine how much of a

difference they make in patients’ PCT blood concentrations and how this affects their sepsis

diagnosis.

2.3.1.2 Interleukin-6

Interleukin-6 (IL-6) has garnered attention from scientists and researchers as one of the

next best sepsis diagnostic tools. IL-6 is one of the most commonly used cytokines in diagnosing

bacterial infections and is produced by T cells, macrophages, fibroblasts, endothelial cells and

more in response to infection/tissue trauma [26]. IL-6 helps to defend the host by stimulating

acute phase responses, hematopoiesis and immune reactions and “is strictly controlled by

transcriptional and posttranscriptional mechanisms, dysregulated continual synthesis of IL-6

plays a pathological effect on chronic inflammation and autoimmunity” [27]. IL-6 plays an

important part in homeostasis of the body; these molecules regulate serum, iron, and zinc

concentrations through control of their transporters, promote platelet creation in the bone

marrow, induce the production of phase proteins, reduce production of fibronectin, albumin and

Page 17: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

13

transferrin (other proteins with important regulatory functions), promote differentiation of T-

cells, and other effects that occur in many chronic diseases [27].

IL-6 has been found to be a reliable biomarker for diagnosing, determining severity, and

predicting the outcome of patients with sepsis. IL-6 concentration increases within two hours

from point of infection and can continue increasing for up to twenty-four hours after [28]. This

means that sepsis can be diagnosed early and allows for a longer testing window after the initial

infection. According to a meta-data analysis of 21 studies, IL-6 was found to have a sensitivity of

68% and a specificity of 73% [30]. This provides moderately accurate results for the diagnosis of

sepsis. Though not as helpful in differentiating severities of sepsis as PCT, IL-6 is an effective

alternative in diagnosing different stages of sepsis [28]. This can be seen below in Figure 3:

Figure 3: IL-6 concentrations as severities of sepsis increase [28].

2.3.1.3 Presepsin

Presepin, also known as CD14, is another biomarker that shows great promise in

diagnosing sepsis. Presepsin is a glycoprotein that is a receptor of the lipopolysaccharide-

lipopoly-saccharide (LSP-LBP) binding protein [31]. This receptor can activate signal

transduction pathways which trigger systemic inflammatory responses. Presepsin has two forms -

mCD14 and sCD14 - which are membrane and soluble forms respectively. Membrane bound

Page 18: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

14

presepsin is found on the surfaces of monocytes and macrophages and sometimes on neutrophils.

Likewise, sCD14 is created from mCD14 falling-off or secretion by these cells [32]. When used

in diagnosing sepsis, sCD14 is the preferred form of measurement [31].

Immunoassays are primarily used to measure the concentration of presepsin in the blood.

Most studies use chemiluminescent immunoassays, which combine chemiluminescence and

immunochemical reactions to produce variable light emission, to measure the presepsin

concentration in whole blood samples [33]. When performed in a test setting, the blood sample is

collected and mixed with a reagent before being diluted. Afterwards, a luminescent substrate is

added and automatically processed with an immunoassay analyzer to measure how vibrant the

given light is, showing overall presepsin concentration [34]. Currently, there is an immunoassay

analyzer called PATHFAST created by the Mitsubishi Chemical Medicine Corporation of Japan

that can analyze samples in just seventeen minutes and costs anywhere from $800 to $1,200 [35].

Presepsin is a fast-appearing and reliable biomarker for diagnosing severity of sepsis and

estimating 28-day mortality for patients. Presepsin concentrations rise in the first two hours after

infection, peak at 4 hours, and decrease between 4-8 hours after infection [32]. This quick

timeframe allows for early diagnosis of sepsis but can also be deemed too quick to effectively

work. A study conducted by Ma et. al, with 41 patients found significance between the presepsin

concentrations between different severities of infection [34]. The study identified ranges for

normal, SIRS, local infection sepsis and “severe sepsis” at 294.2±121.4 pg/mL, 721±611pg/mL,

333.5±130.130.6 pg/mL, 817.9±572.7 pg/mL and 1992.9± 1509.2 pg/mL respectively [34]. As

seen in Figure 4, all of the values are significantly different from one another, allowing analysis

of severity. The same study found that presepsin has a sensitivity of 91.9% and another study

found that presepsin has a sensitivity of 85.7% in predicting sepsis [31, 34].

Page 19: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

15

Figure 4: Comparison of presepsin concentrations in different severities of sepsis and infection [34].

Although presepsin does have lots of promise, there are some limitations with its use as a

biomarker for sepsis. Though many studies found a high sensitivity and specificity, a meta-data

analysis conducted by Zhang et al., of eight studies with 1757 patients found that it may not be as

useful as once thought [36]. This meta-analysis found that the pooled sensitivity was 0.77 and

the sensitivity was 0.73 meaning that, from these eight studies, there was only a moderate

diagnostic capacity for diagnosing sepsis [36]. As LPS, the binding protein, is a component of

gram-positive bacteria, presepsin is much more sensitive to gram-positive bacterial infections

than gram-negative bacterial infections with a sensitivity of 95.5% versus 77.8% [32]. Presepsin,

as it is produced in response to bacterial infection, will therefore not be a successful predictor of

viral infection which can also lead to sepsis. Lastly, in patients with renal failure compared to

those without, there was no significance between the presepsin concentration values [32]. This

may create false results if the patient has any disease which caused kidney failure and needs to

be considered when deciding upon which test is the best.

Page 20: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

16

2.4 Potential Mechanical Tests for Sepsis Diagnosis

2.4.1 Red Blood Cell Deformability

Analyzing red blood cell deformability could be helpful as a means of indirectly

detecting sepsis. A technique known as laser diffraction, a relatively recent development, offers a

way to detect the shape and size of cells.

Laser diffraction is based upon the basic principles of the way light interacts with

surfaces. There are four ways in which light interacts when it strikes the surface of an object: the

light could be diffracted, refracted, reflected, or absorbed (as seen in Figure 5). Reflected light is

light that hits the object and gets “reflected” back directly along the path that the light traveled to

the object on, making analysis for this purpose nonexistent. Diffraction makes relatively acute

angles in reference to the horizontal plane. In contrast, higher intensities caused by refraction

makes wider angles along the positive x axis [37]. As “intensity of light” is easier to measure

from a fundamental standpoint, it is the recommended to use diffraction when analyzing smaller

particles.

Figure 5: Schematic of refracted, reflected, and absorbed light [37].

Fundamentally, information can be obtained about a particle by observing the intensity of

scattered light and the angle of the diffracted light.

Page 21: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

17

Figure 6: Scattering of light from particles of different sizes [38].

Scattered light is essentially a bunch of randomly directed light rays whereas diffraction

is used to delineate how the light rays are spread out by passing across an edge at a certain angle.

An image of the shape of a cell can be extrapolated from the pattern that the diffracted light rays

form [39]. The pattern that is generated by the scattered light is analogous to how a shadow can

be formed on a wall when you shine a light on an object. The shadow could be seen as the

pattern being made by the scattered light except for in laser diffraction, the actual light (laser) is

the pattern. Figure 7 is a simple diagram of a diffraction pattern.

Figure 7: Schematic of laser passing through a sample to form a light pattern [39].

Page 22: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

18

Here, the blue arrow is a suspension of particles, the red arrow is a laser beam, and the

red circles are simplified diffraction patterns from the diffracted scattering of the laser. As the

pattern is from a sample of particles as opposed to a single particle, it is therefore possible to

calculate each particle’ size based on the projected light pattern. From there, it is then possible to

distinguish the amount of particles that exist of a specific size based on the particle’s respective

light intensity. Similarly, once raw data has been collected from the scattered laser light through

means of sensors, the data can then be analyzed to derive a particle size. The data is analyzed

using software such as LA-960 software from HORIBA. The general theory for how particle size

can be derived is depicted in the schematic below in Figure 8. In Figure 8 a circular aperture is

cut into a thin 2D medium, as shown by the grey square. In the equations, “theta” is the angle

relative to the positive x axis that a given laser beam ray makes from the aperture to the point

that the ray hits on the projecting screen behind the aperture, “d” is the diameter of the circular

aperture, “lambda” is the wavelength of the laser light, “D” is the distance between the aperture

and the projecting screen (the black square), “y” is the distance between the center of the

projecting screen and each point of contact on the projecting screen by the laser from the

diffracted rays, and “m” is the refractive index. “d” which is the diameter of the circular aperture,

would be representative of the particle diameter and therefore the diameter of the particle could

be found from calculating “d” using the equations below.

Here, the blue arrow is a suspension of particles, the red arrow a laser beam, and the red

circles are simplified diffraction patterns from the diffracted scattering of the laser. As the

resulting pattern is from a sample of particles as opposed to a single particle, it is therefore

possible to calculate each particle’s size based on the projected light pattern. From this

calculation, one can then calculate the quantity of particles of a set size based on the intensity of

light within the pattern. In summary, the pattern itself indicates the size of the particles and the

intensity of light in the pattern indicates how many particles there are.

Once raw data has been collected from the scattered laser light through means of sensors,

the data can then be analyzed to derive a particle size. Commonly, the data is analyzed using

software such as LA-960 software from HORIBA. In Figure 8, a circular aperture is cut into a

thin 2D medium as shown by the grey square. In the presented equations, “theta” is the angle

relative to the horizontal plane that a given laser beam ray makes from the aperture to the point

Page 23: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

19

that the ray hits on the projecting screen behind the aperture. “d” is the diameter of the circular

aperture, “lambda” is the wavelength of the laser light, “D” is the distance between the aperture

and the projecting screen (the black square), “y” is the distance between the center of the

projecting screen and each point of contact on the projecting screen by the laser from the

diffracted rays, and “m” is the refractive index. As “d” would also be representative of the

particle diameter, the diameter of the calculated particle could be found from calculating “d”

using the equations below:

Figure 8: Fraunhofer circular aperture diffraction theory [40].

The basic concept of how a “particle size analyzer” or a laser diffraction device works is

that it first measures scattered light intensity and angle and then secondly it transforms the

scattered data into a particle size distribution [41]. Figure 9 shows an example plot of raw data

from different particle sizes. In Figure 9 the vertical axis, “I”, is intensity and the horizontal axis,

“alpha”, is the distance from the center of the diffraction pattern. The “y” shown in Figure 8

above is a good representation of “alpha”.

The basic components of how a “particle size analyzer” or a laser diffraction device

works is that it first measures scattered light intensity and angle and then secondly it transforms

the scattered data into a particle size distribution [41]. Figure 9 shows an example plot of raw

data from different particle sizes. In figure 9, the vertical axis, “I”, is intensity and the horizontal

axis, “alpha”, is the distance from the center of the diffraction pattern. The “y” shown in figure 8

above is seen as a good representation of “alpha”.

Page 24: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

20

Figure 9: Intensity plot: Overlapping diffraction patterns of a sample containing particles of different sizes (left),

and a sum of diffraction patterns, i.e. intensities measured by the detector (right) [39].

Laser diffraction is useful in detecting deformability or non-deformability of red blood

cells. Normal red blood cells in healthy humans will readily deform when under shear stress. In

sepsis, red blood cell deformability has been observed to decrease significantly [42]. Red blood

cell deformability has also been known to be the cause of surprisingly low viscosity at high shear

rates [43]. For this reason, an effort can be made to study both the deformability of red blood

cells and its correlation to viscosity.

2.4.2 RBC Aggregation

RBC aggregation is promoted when inflammation, infection, or trauma occur in the body.

In low shear conditions, RBCs tend to aggregate into stacks resembling a roll of coins. These

stacks, called rouleaux, can further aggregate parallel to one another to create larger aggregates.

In normal blood, shear rates of 7-10 s-1 are sufficient to disperse these aggregates [44].

Aggregate formation is also resisted by RBC surface charge since cells are electrostatically

repelled. RBC aggregation is largely a result of macromolecules in plasma, especially fibrinogen

[43]. The University of Southern California Medical School conducted a study of experimental

sepsis in rats that examined the differences in RBC aggregation between healthy rats, rats who

had undergone a cecum ligation/puncture to induce sepsis, and rats who experienced sham

operations to cause inflammation without infection [42]. Blood samples were collected from

each group, with sham-operated and septic samples collected 18 hours after the operation.

Fibrinogen levels and RBC aggregation were found to be significantly higher in the sham

operated and septic rats, compared to control. When blood plasma was replaced by a 3% dextran

70 solution, RBC aggregation in the septic rats was again found to be significantly higher than

Page 25: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

21

the control, as seen in Figure 11. Furthermore, aggregation was found to be significantly higher

in the septic group than the sham-operated group. Without the effects of fibrinogen from the

plasma, this difference must be the result of cellular properties such as RBC surface charge [42].

This experiment displays potential towards differentiating between patients with sepsis compared

to those with non-septic inflammation.

Figure 11: RBC aggregation indexes (M1) measured using a Myrenne aggregometer [42]

A similar study of experimentally induced sepsis in pigs at the University of Debrecen by

Nemeth et. al. found more controversial results [45]. Five pigs were administered Escherichia

coli intravenously in increasing amounts over three hours while four pigs were given a similar

volume of an isotonic saline solution. Blood samples were collected from all pigs at the

beginning of the experiment and every second hour afterwards for eight hours. The intravenous

administration of E. coli initially caused bacteremia. This then developed into fulminant (sudden

or severe) sepsis. Of the five pigs in the septic group, two died in 3-4 hours and the remaining

three died in 6-7 hours. Instead of the expected increase in RBC aggregation in the sepsis group,

RBC aggregation decreased over time. Nemeth et. al. speculated this result may have been due to

the experiment investigating the early hours of fulminant sepsis rather than a study of a slower-

developing sepsis over a longer period of time, or from the effects of the specific E. coli strain

used, which contained hemolysin. This hemolysin affects the RBC membrane and causes

swelling, which could affect its aggregability [45]. The results may also have been affected by

the deaths of pigs in the septic group. If some factors contributed both towards pigs surviving

Page 26: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

22

longer and towards decreased RBC aggregation, the early deaths of some pigs would skew

results towards traits of the pigs that survived longer.

RBC aggregation can be measured in a variety of ways. The erythrocyte sedimentation

rate (ESR) is a widely used test in which a sample of blood is placed in a vertical glass tube and

the rate at which RBCs, also known as erythrocytes, settle to the bottom of the tube is observed.

The greater the aggregation, the faster the RBCs will sediment. This test takes about an hour to

perform [43]. Another test uses low shear viscometry to measure the viscosity of blood, which

increases with RBC aggregation, but is also affected by hematocrit. RBCs can also be observed

directly using a microscope and can be used in conjunction with an image analysis program to

track aggregation over time [44]. Finally, a test called “syllectometry” analyzes the light

scattering of RBC suspensions; as light is shone through a thin layer of blood, the resultant rays

impact a photodetector on the other side that records the light intensity. If blood is disaggregated,

each RBC scatters light and the intensity that reaches the photodetector will be low. As RBCs

aggregate, the light will be able to shine more directly through to the photodetector. Therefore,

the intensity of light measured by the photodetector can be used to determine RBC aggregation

[46].

These two experiments embody the challenge in developing a conclusive diagnostic test

for sepsis. As sepsis is very broad, it may develop suddenly or slowly. Furthermore, whether it is

bacterial or viral, as well as specifics of the infection and the patient, can all have effects on

blood. While results may seem promising in one experiment, a similar experiment may have

entirely different results. In a hospital setting, patients will be of different ages and backgrounds,

with different comorbidities to consider. For example, patients with diabetes mellitus are likely

to have increased RBC aggregation in the absence of infections or inflammation [42]. There is

also the question of how well experiments on animal models translate to humans. Further studies

focused on humans are therefore required.

Page 27: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

23

2.4.3 White Blood Cells

White blood cells have been recognized for their use as indicators in the diagnosis of

sepsis. The primary function of white blood cells is to “fight bacteria, viruses, and other

organisms your body identifies as a danger” [10]. A high white blood cell count is generally an

indication of infection as a greater amount of white blood cells are needed to keep the body

healthy.

Neutrophil to lymphocyte ratio (NLR) is a specific method based on white blood cells

that may be used for early sepsis diagnosis. Neutrophils, a type of granulocyte, are the first white

blood cells to respond and travel to sites of acute inflammation in the body [47]. Due to this fast

response time, neutrophils act as the first line of defense, “directly phagocytosing and killing

microorganisms” while lymphocytes “determine the specificity of the immune response to

infectious microorganisms and other foreign substances” [48, 49]. The primary function of

lymphocytes is to bind to receptors to detected antigens and aid in their removal from the

patient’s body. Lymphocytes undergo cloning to produce multiple identical cells and to fight

against the detected infection. Sepsis generally causes an increase in neutrophil count; this is due

in part to neutrophil dysfunction, “a hallmark of sepsis, contributing to weak immune responses

to the causative infections, as well as additional off-target organ damage” [50]. This dysfunction

affects neutrophil response to the chemical stimulus of infection and negatively affects

antimicrobial activity. Since preliminary neutrophils cannot properly fulfill their functions, more

neutrophils are therefore needed to quell the spread of pathogens, weakening the stopping power

of the body in fighting off newer infections. The lifespan of these neutrophils thereby increases,

increasing the number of immature neutrophils circulating in the body [51]. On the contrary,

sepsis generally causes a decrease in lymphocyte count. As sepsis progresses and neutrophils

cannot properly control responses against infection, lymphocyte apoptosis occurs. This induces

“a state of 'immune paralysis' that renders the host vulnerable to invading pathogens” [52]. The

death of these lymphocyte cells causes a lower lymphocyte count.

NLR is calculated as the ratio of neutrophil count to lymphocyte count in a blood sample.

This blood sample is collected from a simple blood differential test, which is relatively easy to

perform and inexpensive compared to other methods of diagnosis. Normal NLR values for an

adult in good health are between 0.78 and 3.53 [53]. A higher NLR value corresponds to a

Page 28: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

24

greater amount of physiological stress and systemic inflammation within the patient's body.

However, there is currently no standardized level of measurement demonstrating the significance

of a value when it is higher than that of the average healthy patient. Greater physiological stress

and systemic inflammation may be a result of sepsis, coronary interventions, coronary artery

bypass grafting, inflammatory conditions, etc. Therefore, utilizing only NLR as the primary

method of diagnosis for sepsis is not ideal as NLR cannot currently diagnose a specific

condition. However, NLR in combination with another method of diagnosis may be more

promising.

Designing a microfluidic device based on spontaneous neutrophil motility may also be a

promising strategy in the diagnosis of sepsis. Motility can be described as the ability of an

organism to move by itself, utilizing its own metabolic energy [54]. Patients with sepsis

generally have high spontaneous motility when compared to healthy patients. Along with

motility, the effectiveness of neutrophils in attacking pathogens is important to observe as sepsis

negatively affects the ability of neutrophils to respond properly to chemotactic signals. In one

specific study, Dr. Muldur of Harvard Medical School and Massachusetts General Hospital and

his team created a microfluidic device to measure spontaneous neutrophil migration from a

single drop of diluted blood [55]. The device contained eight migration mazes with each maze

consisting of several migration channels and a red blood cell (RBC) filter. An image of this

microfluidic device is shown below in Figure 12 [55].

Figure 12: The microfluidic device with a detailed view of the maze [55]

Page 29: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

25

Spontaneous neutrophil motility is then detected through microscopy techniques and

tracked through ImageJ/Fiji analysis software. The “Sepsis Score” can then be calculated using

the equation below:

Sepsis Score=N*(O+P+R+AD)/103 Eq. 1

In this equation, “… N, is the total number of migrated neutrophils, O, the total number

of oscillation phenotype, defined as ‘the total number of cells that switch direction twice in a

channel and migrate for more than 15μm in each segment’, P, the total number of arrest

phenotype, defined as ‘a cell with zero velocity’, R, the total number of retrotaxis phenotype,

defined as ‘any cell that leaves the maze back into the central chamber’, and AD, average

distance of all neutrophils traveled in the two mazes, divided by four” [55]. A patient is

considered septic if their “Sepsis Score” is greater than 30. This microfluidic device proved to be

very specific to sepsis with 98% specificity and 97% sensitivity [55]. This device also proves to

be simple to use as an operator would only need to be trained for preparing and loading the

patient’s blood sample, as data analysis and the “Sepsis Score” results are automated [50].

Page 30: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

26

Chapter 3: Project Strategy

3.1 Client Statement

While the scope of the project is vast, the initial client statement from Professor Sabuncu,

a project advisor, is as follows:

Red Blood Cell (RBC) aggregation is central to the study and diagnosis

of sepsis. At low shear rates and during hemostasis, RBCs pile together to form a

“rouleau”. At high shear rates, these aggregates are dispersed, whereas, in the

microcirculation, shear forces are not strong enough to dissociate the aggregates.

In infectious diseases, where the RBC aggregation is elevated, the local viscosity

increases in the microvasculature and this increase contributes to low perfusion and

tissue ischemia. Therefore, knowledge of the amount of RBC aggregation at low

shear rates could allow early diagnosis of septic shock and organ failure. Currently,

the gold standard clinical methods to identify sepsis rely on microbiological

techniques that require 2-3 days to complete. However, a point of care technology

that can diagnose sepsis from the RBC physical properties does not exist. In

addition, the current techniques do not allow for comprehensive characterization of

RBC aggregation. For a complete analysis of RBC aggregation, RBC aggregate

size distribution, aggregate resistance to disaggregation (shear stress), aggregate

morphology, and aggregation kinetics need to be quantified. Also, a comprehensive

technique should be able to distinguish between plasmatic and cellular factors that

lead to aggregation.

Shortly after the initial client statement was provided, a decision was made between the

team and the project advisors, Professor Sabuncu and Professor Page, to examine the issue of

diagnosing sepsis more broadly. The team is to determine the most important criteria for a sepsis

diagnostic test, devise a test/series of tests that best meets those criteria, and design a device to

conduct said test(s).

Page 31: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

27

3.2 Design Requirements

3.2.1 Standards

Throughout the design process, the team must ensure that all requirements from relevant

industry standards are met.

ISO 13485:2016 specifies requirements for a quality management system where an

organization needs to demonstrate its ability to provide medical devices and related services that

consistently meet customer and applicable regulatory requirements. Such organizations can be

involved in one or more stages of the lifecycle, including design and development, production,

storage and distribution, installation, or servicing of a medical device and design and

development or provision of associated activities (e.g. technical support). ISO 13485:2016 can

also be used by suppliers or external parties that provide product, including quality management

system-related services to such organizations [56].

Another important standard that is relevant to our design is ISO 11737-2:2009 which

addresses standard requirements for the sterilization of medical devices. ISO 11737-2:2009

“specifies the general criteria for tests of sterility on medical devices that have been exposed to a

treatment with the sterilizing agent reduced relative to that anticipated to be used in routine

sterilization processing” [57]. Sterilization is used to kill bacteria and other biological agents that

may have been in contact with the device in order to protect the user. ISO 11737-2:2009 further

defines a medical device as an “instrument, apparatus, implement, machine, appliance, implant,

in vitro reagent or calibrator, software, material or other related article, intended by the

manufacturer to be used, … for human beings for one or more of the specific purpose(s) of: -

diagnosis, prevention, monitoring, treatment or alleviation of disease …” [57]. The team’s

design will therefore fall under the definition of medical device provided in this and the prior

standard. As the team will be trying to diagnosis sepsis using an instrument, the design will be

considered a medical device. As the device will be interacting with human patients and blood,

the device must be sterilized as to keep all patients safe.

ISO 10993-1 is based on biocompatibility and biological evaluations of medical devices.

The major goal of this standard is to protect humans from biological risks associated with

medical devices. This standard provides requirements on how to develop a risk management

Page 32: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

28

process and assess all potential biological hazards in order to keep all users safe [58]. ISO

10993-1 will be incorporated into the design as it will be interacting with human patients and

blood and the team will need to perform biocompatibility evaluations in order to guarantee safety

to all users.

ISO/TC 76 is another relevant standard that is centered around blood processing

equipment. ISO/TC 76 is a “standardization of containers (such as infusion bottles and bags,

injection vials, ampoules, glass cylinders, cartridges, prefillable syringes, etc.) application

systems (such as giving sets, non-electrically driven portable infusion devices, blood collection

systems, etc.) and accessories for infusion, transfusion, injection and blood processing in blood

banks, terms, definitions, requirements and test methods for these devices, specifications and test

methods for quality and performance of their materials and components (such as elastomeric

closures, caps and ports, pipettes, etc.)” [59]. This standard is relevant to the team’s design and

will be incorporated into a system that will collect and process a human patient’s blood.

ISO 10993-11 is based upon biological evaluation of medical devices in regard to

systemic toxicity and adverse systemic reactions. This standard provides requirements for

developing a process to evaluate systemic toxicity of medical devices to determine whether the

medical device is safe to use on humans [60]. ISO 10993-11 will be incorporated into the design

as it will be utilized with human patients and blood and the team will need to perform

biocompatibility and toxicity evaluations in order to guarantee the highest safety to all users.

In addition to these industry standards, the team will also adhere to all relevant standards

related to ethics. One specific ethical standard is the HIPAA Privacy Rule. This rule “…

establishes national standards to protect individuals’ medical records and other personal health

information …” [61]. This rule preserves the privacy of personal health information obtained

from patients. Information that is protected under this law includes “information your doctors,

nurses, and other health care providers put in your medical record, conversations your doctor has

about your care or treatment with nurses and others, information about you in your health

insurer’s computer system, billing information about you at your clinic, and most other health

information about you held by those who must follow these laws” [62]. The team will

incorporate the HIPAA Privacy Rule by keeping all patient information safe and private unless

given patient authorization.

Page 33: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

29

3.3 Revised Project Statement

From our research on the methods and limitations of sepsis testing, a revised project

statement was developed. The goal of this project is “to development a point-of-care device that

prioritizes accuracy, a rapid testing time, and a low cost.” The device will be used when sepsis is

suspected in patients in order to provide strong evidence towards or against a diagnosis of sepsis.

By obtaining results in hours or minutes rather than days, sepsis can be treated earlier in order to

increase chances of survival.

Page 34: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

30

Chapter 4: Design Process

4.1 Needs Analysis

Table 1: Pairwise Analysis for Device Designs

When it comes to designing the final device, it is important to consider what both the

client and the team find valuable in terms of design elements and overall purpose. If the team

focuses on elements that emphasize different components from the client, the device would be

considered a failure regardless of its effectiveness in tackling the problem. As a result, we

decided to create a pairwise analysis of the different components that the client desired to

determine which aspects of the device should be more focused on than others. A pairwise

analysis matrix works by comparing and ranking a design element (low cost, easy to

manufacture, etc.) and comparing it against the following columns of the same criteria. If the

former characteristic is more important than its column counterpart, it is given a value of “1”. If

equal, it is assigned a value of “1/2” and if less important, a value of “0”. All values are then

totaled following the end of the process and those with the highest total score are therefore the

most important traits to focus on when designing our device.

In this instance, comparing the qualities found in Table 1 to those desired by the client, it

was determined that “effectiveness in diagnosing” sepsis within the patients was the most

important task for our device to accomplish, followed closely by a “low cost” solution and the

ability to provide a “quick testing time”. The ability to effectively diagnose the disease captures

the highest position since the device’s primary function is to help characterize those with sepsis

as accurately as possible. The concept of a low cost yet quick testing design also plays into the

notion that sepsis has a very small window in which the disease can be properly monitored

Page 35: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

31

before negative effects begin hampering the body. If a procedure takes too long to perform or

costs too much that it might propose doubt in the mind of a doctor to use, it would also serve as

an ineffective diagnostic method.

Keeping in mind what is important in the device design, it is now possible for our team to

test and adapt methods that focus on achieving these three aspects. Possible methods used in

conjunction with the device are discussed in detail in the next section “4.2 Conceptual Designs

and Prototype Testing”.

4.2 Conceptual Designs and Prototype Testing

Following the pairwise analysis, a concept map was created to allow deeper analysis into

potential methods for sepsis diagnosis. Based on the concept map seen in Figure 13, the team

then narrowed our focus to RBC aggregation due to its variable testing methods, speed of testing

time, and for being the most specific in relation to a septic diagnosis than all other methods. In

choosing RBC aggregation, we are further ensuring that we meet the criteria set by our pairwise

analysis. Methods to measure RBC aggregation in relation to sepsis are discussed in further

detail in the following sections.

Figure 13: Concept map

Page 36: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

32

4.2.1 Erythrocyte Sedimentation Rate

4.2.1.1 ESR

One diagnostic blood test that can be used to compare septic and non-septic blood is the

erythrocyte sedimentation rate (ESR), first mentioned in section “2.4.2 RBC Aggregation”. ESR

measures the rate at which erythrocytes in an anti-coagulated whole blood sample settle in a test

tube within the time frame of one hour, expressed in units of millimeters [63]. Throughout the

test, the test tube must remain perfectly vertical as the resultant values are determined by the

height of plasma at the top of the tube after the set one-hour period [64]. Primarily, ESR can be

used as an indicator of inflammation; patients with inflammation have an increase in the amount

of proteins in their blood, which causes red blood cells to clump together [64]. Due to this

clumping, red blood cell aggregates within the patient’s blood will settle at the bottom of the

tube at a faster rate than those that remain independent. Therefore, a patient with inflammation

will have a greater ESR level. Figure 14 shows how red blood cells aggregate when introduced

to inflammation. Sepsis involves widespread inflammation in the body so it can be assumed that

septic patients will have high ESR levels.

Figure 14: Normal erythrocyte sedimentation compared to inflamed [64]

The Westergren method is considered the gold standard for ESR. Materials required for

the Westergren method include anti-coagulated blood, sodium citrate solution, a tube support

Page 37: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

33

rack, and “standardized colorless, circular glass or plastic tubes, with an inner diameter of at least

2.55 mm and sufficient length to include a 200 mm sedimentation scale” [63]. A representation

of three different situations is shown in Figure 15 below. Sample A shows the whole blood

sample mixed with sodium citrate at the start of the ESR test. Sample B shows normal results for

ESR at the end of the 60-minute test time. Sample C shows results for a higher ESR level at the

end of the 60-minute test time.

Figure 15: Three possible ESR situations [63]

When using the Westergren method, normal ESR values are ≤15 mm/hr for men and ≤20

mm/hr for women [63]. ESR may increase with age and the highest ESR values generally come

from people between 65-74 years of age [63]. Although ESR is a non-specific marker, it will be

beneficial to study as there may be a distinguishing feature between septic and non-septic blood

that can be measured using this method.

Page 38: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

34

4.2.1.2 Micro-ESR

Micro-ESR is another diagnostic blood test that can be used to compare septic and non-

septic blood and, similar to its full-sized companion, Micro-ESR requires capillary blood

samples, micro hematocrit capillary tubes, a lancet, laboratory slides, sodium citrate, antiseptic

solution, and a rack to hold all of the tubes [65].

The biggest comparison between ESR and micro-ESR tests are their testing timeframes.

During a specific procedure performed by Reza Hashemi, a scientist of the Department of

Internal Medicine at Shohadaye Tajrish Hospital in Iran, a patient’s fingertip was carefully

punctured with a lancet after properly cleaning with antiseptic solution [65]. Four drops of blood

were collected and added to a laboratory slide with a single drop of sodium citrate. The blood

and sodium citrate would then be gently mixed on the slide. Next, “a 7.5-centimeter capillary

tube was placed on the slide immediately with a 30 to 45-degree angle” [65]. The blood sample

would then rise through the capillary tube. When the blood sample would reach the 7-centimeter

mark, the tube would be placed vertically and placed into a tube rack to ensure all tubes would

be vertical [65]. Based on the statistical analysis of Hashemi’s results, a correlation was made

between micro-ESR and traditional ESR. Hashemi’s results proved that micro-ESR results can

be successfully interrupted at 20 minutes [65]. Therefore, micro-ESR is faster than traditional

ESR, which requires 60 minutes to complete, while still using less materials to perform.

However, like ESR, micro-ESR is still a non-specific marker of inflammation. Therefore, this

teams feels it will be beneficial to study as there may be a distinguishing feature between septic

and non-septic blood that can be measured using this method.

4.2.2 Capillary Fill

Capillary action is “the tendency of a polar liquid to rise against gravity into a small-

diameter tube” [66]. There are two opposing forces that drive capillary action – cohesion and

adhesion. Cohesion results from intermolecular forces, and there is a correlation between these

intermolecular forces and a liquid’s viscosity. Adhesion results from attractive forces between

the liquid and another surface. In capillary action, adhesion seeks to maximize the amount of

capillary surface the liquid touches, while cohesion seeks to minimize the liquid’s surface area

[66].

Page 39: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

35

For a capillary tube placed into a liquid at an angle, the liquid will rise along the tube to a

specific height. As it does so, it experiences a force that is described by the following equation,

illustrated in Figure 16 [67]:

𝐹 = 2𝜋𝑟𝛾 cos 𝜃 − 𝜌𝑔𝜋𝑟 cos 𝛽 𝑠(𝑡), Eq. 2

where

F is the force the liquid exerts on the capillary tube

γ is the surface tension of the liquid

θ is the contact angle between the liquid and the capillary

ρ is the density of the liquid

r is the radius of the capillary tube

β is the angle of the capillary tube with respect to the vertical

s(t) is the distance traveled by the liquid along the capillary as a function of time t

Figure 16: Liquid rising in a capillary tube [66]

The first term in this equation describes the force upward due to surface tension, and the

second term describes the force downward as a result of gravity. The upward force features two

properties of the liquid – surface tension γ and contact angle θ. Surface tension increases with

larger intermolecular forces. The contact angle is determined by the balance between cohesive

and adhesive forces, with a smaller contact angle contributing to a larger upward force [68].

Page 40: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

36

As red blood cells aggregate, blood viscosity increases [43]. It is suspected that, as red blood

cells aggregate, cohesive forces in the blood will also increase [68]. Since there is a correlation

between viscosity and the intermolecular forces that affect capillary action, it is possible that

different levels of aggregation will cause differences in the speed at which blood will fill a

capillary tube [66]. Measurements of this speed can be made and compared to the relative

aggregation of these blood samples to investigate a potential relationship between aggregation

and capillary fill speeds.

4.2.3 Vibration Syllectometry

Most commercial aggregometers today disaggregate RBCs using a shearing system

which can be very expensive. In order to simplify aggregometer design, reduce costs, and allow

for simpler cleaning between tests, a team at Kyungpook National University developed an

instrument that disaggregates blood through vibrations, then measures aggregation through

syllectometry, as seen in Figure 17 [69]. The disaggregation mechanism consists of a function

generator, amplifier and speaker. A jig is attached to the speaker diagram such that the jig will

vibrate when the speaker is turned on. A glass test slide to hold a small amount (around 10 μl) of

blood is then fixed to the jig [69]. This slide has a cavity for the blood and can be easily disposed

of after use [69, 70].

Figure 17: Vibration syllectometry device using backscattered light [69]

Page 41: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

37

The syllectometry is conducted using a laser diode (650 nm, 1.5-5mW) that emits a laser

beam through the blood sample. A photo diode then measures the intensity of light transmitted

through or backscattered by the blood [69, 70]. Measurements from the photo diode are recorded

to a computer and the resultant intensity is then used to describe the degree of aggregation in the

RBCs. For measurements of transmitted light, a higher intensity indicates a higher degree of

aggregation [70]. In backscattered light measurements, a lower intensity indicates a higher

degree of aggregation [69].

Following the procedure laid out by Shin et. al, a blood sample is vibrated for 40 seconds

at a frequency and amplitude of 150 Hz and 0.5 mm, respectively [69]. The exact timing,

frequency, and amplitude can be varied based on the results observed in order to refine the

resultant values. The duration should be long enough that the RBCs completely disaggregate

while the vibration frequency and amplitude should be mild enough to avoid damaging the cells.

Once the blood is fully disaggregated, measurements of the variation in light intensity over time

are collected and the application of a curve-fitting program is used to quantify aggregation [69].

For collection of research data, especially when comparing septic and healthy blood, it is helpful

to use samples with a similar hematocrit since this affects light intensity.

4.3 Alternative Designs

As RBC aggregation is our main focus, any additional methods need to also focus on

RBC aggregation as the main conditional testing criteria. Referring back to our concept map

from section “4.2 Concept Map”, our main alternative methods would be incorporating either a

standard ESR test or the use of image tracking. Standard ESR tests, while requiring more blood

and a longer testing time, are still effective at looking at erythrocyte sedimentation rate. In regard

to image tracking, red blood cells can be analyzed over a set period of time to determine how

long it takes for rouleaux to form. Specific RBCs can be placed with a virtual marker and tracked

as aggregation occurs through a series of photos, making use of continual time periods to detail

rouleaux formation.

Page 42: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

38

4.4 Final Design Selection

4.4.1 Blood Preparation

The preparation of the blood dextran solutions for the micro-ESR, capillary fill, and

vibration syllectometry were all done as one big batch. Blood was obtained from AllCells, LLC, a

reputable blood and marrow cell supplier from Quincy, Massachusetts. The blood was fresh

whole blood shipped in a small vial, mixed with an anti-coagulant (Citrate Dextrose Solution –

A), and placed in a climate-controlled shipping container before being stored in a fridge. To

make our dextran-blood solutions, we used dextran with a 50k molecular weight (MW), whole

human blood, and phosphate-buffered saline (PBS). To create the concentrations, five different

masses of Dextran 50k MW, obtained from Sigma-Aldrich, were measured using an automated

scale. The actual masses of dextran, in comparison to the desired thresholds, can be seen in Table

2. These masses were measured on an electronic scale with an accuracy of +0.1mg:

Table 2: Shows the nominal and actual masses of dextran used in blood sample preparation

Mass of Dextran (grams) Actual Mass of Dextran (grams)

0.025 0.025

0.05 0.051

0.10 0.099

0.15 0.152

0.20 0.201

These measured dextran amounts were each placed in their own 15mL conical tube and

labelled with the dextran mass added. Using a graduated cylinder, 10mL of PBS were measured

and added to each conical tube. The PBS-dextran solutions were then stirred until all the dextran

dissolved in the PBS. To sterilize the PBS-dextran solutions, each solution was passed through a

sterile vacuum pump filter inside of a laminar hood. These solutions were then placed in a new

sterile 15mL conical tube and labelled by the amount of dextran they contained. Next, inside the

same laminar hood, 4mL of 40% hematocrit blood obtained from All Cells was measured with a

serological pipette and placed into a separate15mL conical tube. This was repeated for six

conical tubes total. Those conical tubes filled with 4mL of blood were then placed inside a

centrifuge and spun at 200xg for 10 minutes to allow the full separation of plasma from the

Page 43: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

39

blood cells. After centrifuging, the plasma was then aspirated from each of the conical tubes until

there was about 1.6mL of hematocrit left inside the conical tubes. To maintain a 40% hematocrit

and produce 4mL of each test, 2.4mL of the dextran-PBS solutions were added to five of the

conical tubes containing blood and appropriately labelled. For the sixth tube, 2.4mL of pure PBS

was added for a control of 0 grams of dextran. These dextran-blood solutions were then mixed.

All solutions were stored in a refrigerator at four degrees Celsius for the duration of the

experiment.

4.4.2 Micro-ESR Methodology & Materials

This set-up included a clay capillary tube sealer stand, a blank backdrop made from paper

towels, and a sterile cover placed far above the tubes. To begin the Micro-ESR test, a capillary

tube filled with blood was placed vertically within a clay-sealant stand, as seen in Figure 18. The

capillary tubes were previously sterilized by leaving them in 70% isopropanol for 10 minutes and

then allowing them to dry for an hour, until no more isopropanol was present. The cover was

used to prevent the downward air of the laminar hood from evaporating or drying out the plasma

layer.

Figure 18: Capillary tubes in a clay stand for micro-ESR

Throughout the experiment, pictures were taken every thirty minutes for four hours,

starting at time zero, using the phone’s camera. Each timer was started immediately after the

capillary tube was placed into the clay sealing stand and ended after the test was completed.

Once all pictures had been collected, the blood inside the capillary tube was aspirated and the

capillary tube was disposed of inside the bio-sharps container while the images were prepped to

Page 44: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

40

be analyzed using the ImageJ software. Using the backdrop in conjunction with ImageJ, the

known length of the stand was used to determine the height of the plasma layer in mm. The

plasma was measured from the bloodline to the horizontal height between the two sides of the

meniscus. This was done to prevent confusion between shadows and the line of the meniscus.

This experiment was then repeated three times for each different dextran-blood mixture

concentration. The trials were then averaged to find the average aggregation rate of the blood-

dextran solution.

4.4.3 Capillary Fill Methodology & Materials

Figure 19: A capillary fill test in progress

In order to complete the capillary fill tests, a stand for the experiment was constructed

using two conical tube stands of known dimensions, a piece of cardboard, and a large latex

glove, shown in Figure 19. The cardboard was cut to be the same size as the top of the stand. It

was then placed on the top of the stand and covered with the large latex glove, which acted as a

hydrophobic layer during the testing period. This layer was further controlled by sterilization

using 70% isopropanol before it was placed in the laminar hood. A second sterile stand of the

same size was placed in the laminar hood directly in front of the stand with the glove. A phone,

which was used to record the experiment, was then set-up in a way that allowed for both stands

to be fully in view outside the laminar hood. The rest of the experiment was then conducted

inside the hood. The blood solutions that were to be tested were then removed from the fridge

and given time to rise to room temperature, which was about 30 minutes. The conical tube with

the blood-dextran sample was then lightly shaken to breakup any rouleaux that had formed while

resting.

Page 45: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

41

Using a similar method as before, a capillary was sterilized with 70% isopropanol. Using

a P2000 pipette, 500 microliters of the blood sample were then removed and placed on the latex

glove covered stand. The tube was filled by manually laying it between the two stands, which

created a slope of negative five degrees from the plane (cardboard) the blood sample laid on. The

capillary tube was then pushed into the blood droplet and allowed to fill from one end to another.

Prior to completion, the tube was then plugged with the experimenter's gloved thumb and

removed from between the stands. The filled capillary tube was then placed vertically into the

clay stand for the micro-ESR experiment. The purpose of joining these experiments was done to

conserve blood.

The video of the experiment, taken on the phone, was then uploaded to a computer. The

video was cut into four frames where the blood had flowed to different locations. These frame

images were then analyzed using the prior ImageJ software to determine how far the blood had

flowed from the opening of the tube that was inserted into the blood droplet. The length was

determined by finding the conversion from pixels to mm from the known length of the stand.

The flow rate was then calculated by using the frames/second of the video to convert each frame

to seconds before timed relative to each other. For example, if the video had a frame rate of two

seconds, and frame one was the tenth frame of the video and frame two was the twentieth frame

of the video, frame one would be converted to zero seconds and frame two would be at time ten

seconds. This time, and the length travelled, was then used to determine flow rate of the blood

sample. The flow rates of all three trials were averaged to determine the average flow rate for

each dextran-blood solution. This was repeated for each additional dextran-blood mixture.

It is important to note that the capillary tubes were placed at a negative angle, unlike the

positive angles described in section “4.2.2 Capillary Fill Test”. This angle was chosen because

initial trials found that the dextran-blood solutions did not rise far enough into the capillary

tubes. Suitable measurements could not be made at a positive angle. This is likely a result of

replacing blood plasma with PBS, which has different properties. For example, PBS has a lower

viscosity, which suggests weaker intermolecular forces [71]. From equation (2), increasing the

surface tension but maintaining the same contact angle allows a liquid to rise further in a tube

[67]. These different properties of PBS meant that angles typically used to draw whole blood into

a capillary tube were not viable for this test.

Page 46: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

42

4.4.4 Laser Syllectometry Methodology & Materials

Unlike the micro-ESR and capillary fill tests, the test for laser syllectometry required

more personalized equipment than what was present within our laboratory. To begin, a 3D stand

was modelled using SolidWorks, sitting 6 inches tall, 4.02 inches wide, and 2.42 inches deep.

Three notches were made at 1.67, 2.18, and 3.17 inches respectively and served as holders for an

attachment piece meant to house the slide within the device itself. Cut directly into the device’s

top was a hole for the placement of a DZS Elec 650nm laser diode (part number 5VLD-R650-

5MW-12). This diode was aligned with a Texas Instruments OPT101P photodiode with an on-

chip trans-impedance amplifier that sat within the middle of a breadboard holstered at the bottom

of the device. The breadboard, photodiode, and laser diode were all connect to an Arduino Mega

2560 system at the back of the device that regulated voltage control, controlled the laser diode,

and recorded output from the photodiode. The Arduino was attached directly to a laptop for

power and data transfer. The 3D drawing of our design can be seen below in Figure 20:

Figure 20: The stand designed for syllectometry

Using the prepared blood samples mentioned in section “4.4.1 Blood Preparation”, a

small drop of about 100 µL was placed into the center of an AmScope BS-C12 deep-welled

slide. The slide was then fixed with a glass cover and placed within the slide holder before being

inserted into the device. Prior to activation, two BestTong vibrational coin motors (part number

Page 47: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

43

A00000117) were placed on the left and right of the slide and ran at a frequency of 150hz for

approximately 10 seconds. Once the blood had disaggregated, the laser was activated and the

signal output from the photodiode was recorded for approximately 5-7 minutes. Following test

completion, the slide was disposed of in the proper biohazard container and another was

prepared. Five tests were performed for each dextran solution concentration using the same

methods.

Page 48: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

44

Chapter 5: Design Verification

The team tested the three separate methods of measuring RBC aggregation – micro-ESR,

capillary fill, and syllectometry – to validate how effective they would be in achieving the

project goal. The first test, micro-ESR, measured the plasma height of blood in a capillary tube

as RBCs settled over a period of four hours. The capillary fill test measured the flow rate of

blood in a capillary tube held at an angle. Finally, the syllectometry test measured the intensity of

a laser shone through a few drops of blood as its RBCs aggregated.

5.1 Micro-ESR Results

Figure 21: Buffer height of all amounts of added Dextran 50K MW

Page 49: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

45

Figure 22: Comparison of buffer heights of all Dextran 50K MW amounts

As seen in Figure 21 and 22, the slope of the linear trend lines increases with an increase

in dextran. The only exception is 0.1 g/dL of dextran since it has a smaller slope than that of 0.05

g/dL of dextran. Otherwise, there is a positive correlation between the amount of dextran and

plasma height. The R-squared value of all the linear trend lines for 0, 0.05, 0.1, 0.15 and 0.2

grams are above 0.94 meaning that there is very low variation between trials. For 0.025 grams,

the R-squared value is 0.8252 which is still relatively high and gives confidence in the found

trend.

Page 50: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

46

5.2 Capillary Fill Results

Table 3: Average flow rate of different grams of Dextran

Grams of Dextran Average Flow Velocity

(mm/s)

Standard Deviation (mm/s)

0.025 1.03 +0.209

0.05 2.73 +0.497

0.1 1.02 +0.150

0.15 2.40 +0.486

0.2 1.21 +0.595

Table 3 shows all of the calculated flow rates of dextran amounts 0.025 g/dL to 0.2 g/dL.

For all solutions, three trials were conducted except for 0.1 and 0.2 grams, where only two trials

were usable due to angle differences. The data shows that the average flow rate for each

trial varies from trial to trial. In this testing period, 0 g/dL of dextran was unable to be tested and

is further elaborated on in section “7.2.2 Limitations”.

Page 51: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

47

5.3 Laser Syllectometry Results

Figure 23: Laser Syllectometry Initial Setup

Our testing of aggregation via laser syllectometry yielded interesting results. The main

notion of testing is that, for each blood sample at a different dextran concentration, there should

be an increase in the rate of change and larger difference between the maximum and minimum

values of the sampled values. These values can be seen in Table 4 below:

Page 52: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

48

Table 4: Average Testing Values

Each row of Table 4 represents the average value as a result of all tests performed at the

noted dextran concentration (g/dL of PBS) with the “Control” row referring to directly extracted

whole blood with plasma still intact. After the tests were conducted, the resulting values were

imported into an Excel spreadsheet and individually calculated for the same values as shown

above. As the input into the system was originally a 1024-ADC byte count, each byte was

converted to a voltage resolution before being processed. From the newly acquired voltage, the

minimum and maximum values were taken immediately following the start of the test until the

final recorded value, ignoring the starting values from before the sample was properly situated

within the system (these pre-test values are more clearly show in the data presented in Appendix

H). After the minimum and maximum for each of the five tests per data set were recorded, the

difference was taken between the two values. From there, the rate of change for each test was

taken using Excel’s SLOPE() function with the y-values corresponding to the voltage and the x-

values corresponding to time.

The range of initial and ending voltage values were different depending on how precisely

the laser was interacting with the blood; samples that were not aligned exactly the same as others

Page 53: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

49

might see increased or decreased voltage values, hence why it was essential to look at the change

in voltages over the time period. When looking at the dextran solutions ranging from 0.025g/dL

to 0.05g/dL, there is a slight increase between the average min-max voltage difference (a change

of ~0.001 volts or 1 mV) and a similar change between the rates of change (~0.000025V/s or

0.025mV/s). This trend continues with each increase in dextran concentration for both the min-

max difference and rate of change. It is important to note that, while these trends increase with

each concentration, the number of tests within each dextran sample isn’t the same. This

limitation is further discussed in section “7.3.1 Limitations”.

The individual values for each test, as well as graphs displaying their rates of change, can

be seen in the Figures 24 to 29 and Table 5 to 10 below:

Table 5: Control Tests Information

Figure 24: Control Tests Rate of Change

Page 54: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

50

Table 6: 0.025g/dL Tests

Figure 25: 0.025g/dL Tests Rate of Change

Page 55: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

51

Table 7: 0.05g/10mL Tests

Figure 26: 0.05g/dL Tests Rate of Change

Page 56: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

52

Table 8: 0.1g/dL Tests

Figure 27: 0.1g/dL Tests Rate of Change

Page 57: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

53

Table 9: 0.15g/dL Tests

Figure 28: 0.15g/dL Tests Rate of Change

Page 58: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

54

Table 10: 0.2g/dL Tests

Figure 29: 0.2g/dL Tests Rate of Change

Page 59: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

55

Chapter 6: Final Design and Validation

6.1 Final Design CAD Prototype

The development of the final device design was in large part a combination of what was

initially used for testing and awareness of the problems that were faced when performing said

tests for the first time. With reference to the syllectometry test, issues with the voltage reading at

which to start recording data, the ability to watch the test in action while minimizing the amount

of excess light on the photodiode, and wiring were all present. Keeping these aspects in mind, as

well as other smaller additional features, we developed our final design shown in Figure 30

below:

Figure 30: Final Design Front (L) and Back (R)

As a comparison to the initial design, it is clear that both the width, height, and depth of

the device has been increased. That said, the device was still kept relatively small to minimize

production cost and increase portability, having a new width of 7.87 inches, height of 11 inches,

and depth of 4.72 inches. One of the most prominent new features is the attached 5-inch LCD

touchscreen display at the top of the device. Connecting directly to the Arduino, the touchscreen

allows control of the device and its parameters without the need of a computer such as modifying

Page 60: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

56

testing parameters or displaying live data during a testing period. To further this notion of a

“stand-alone device”, an additional opening has been created in the back of the device to allow

the connection of a USB drive for data storage.

On both the front and back of the device, there exists a one-way acrylic pane. This allows

the user to ensure that, during syllectometry testing, the laser is properly aligned with the blood

sample without increasing ambient light to the photodiode. This notion is further supported by a

voltage-recording threshold where data won’t be recorded until a proper starting voltage is

achieved. Additional holes have also been cut in both the top and back paneling for better wire

management.

On the side of the device sits an opening attachment to a brushless DC stepper motor.

Stepper motors allow for a high degree of modular control when it comes to angular analysis and

can be easily adjusted when attached to our main Arduino control system. This motor, when

connected to a small capillary holder, allows easy modulation in controlling the starting angle

during a capillary fill test. Furthermore, this same motor can be used in our micro-ESR testing.

By plugging the capillary tube with clay, the motor can rotate the tube to ensure it is perfectly

vertical during the recording period. Beside the stepper motor also exists a small spot for holding

an internal battery in the event the device cannot be connected to a continuous power source.

6.2 Standards

Our design incorporated many engineering, industry, and manufacturing standards. Our

team incorporated SolidWorks drafting standards when creating the stand for vibration

syllectometry testing and our final design using computer-aided design. Our team also

incorporated IEEE standards for computer and electronic equipment, such as the 1641-2010

IEEE standard for signal and test definition. Our team incorporated this standard when

performing vibration syllectometry experiments, which involved electrical signals.

However, as the team was not able to fully manufacture our final design due to time

constraints and issues regarding the pandemic, most of the other applicable standards would only

be incorporated if the design were to be manufactured and marketed in the industry. For

example, the HIPAA Privacy Rule would only be fully incorporated when patients are actually

using our device. ISO 10993-11 also would not be fully incorporated until the final design is

Page 61: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

57

manufactured and patients are using the device since it requires biocompatibility and toxicity

evaluation on the device. These evaluations cannot be conducted until the device is fabricated.

ISO 13485:2016 focuses on quality management systems of medical devices, with the goal of

continuously meeting customer and regulatory requirements, and would not be fully incorporated

until device manufacturing. ISO 11737-2:2009 also would not be fully incorporated until the

final design is manufactured and is put into use in a hospital setting since this standard involves

the sterilization of medical devices. ISO/TC 76 also would not be fully incorporated until the

final design is being used with patient blood samples since this standard is centered around blood

processing equipment.

6.3 Economics

As sepsis is a serious condition requiring a hospital stay, it can be very costly. According

to the Agency for Healthcare Research and Quality, sepsis is “the most expensive condition

treated in U.S. hospitals, costing nearly $24 billion in 2013” [1]. The lasting effects and chances

of reoccurrence of sepsis can lead to expensive re-hospitalization of patients two to three times

as frequently as other conditions (such as pneumonia or heart failure). The sooner a patient

receives treatment, the more likely they will be to recover without lasting symptoms or requiring

readmission. Since current tests can take time, patients are treated when sepsis is suspected,

while awaiting test results.

There are also issues with false positive tests, with one study finding that “a single false-

positive blood culture event results in an additional 2.4 days stay in the hospital for the patient”

[72]. This is estimated to cost the U.S. healthcare system $7.5 billion per year. The short time

required for our tests to achieve results, and the ability for them and other tests to validate one

another, can help reduce delays in treatment or extended hospital stays, providing significant

savings for patients and hospitals.

Page 62: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

58

6.4 Environmental Impact

The final design will have environmental impacts due to the materials our team has

chosen to utilize. The final design includes an LCD touchscreen display. LCD screens contain

toxic substances that could have negative effects on ecosystems if they are not disposed of

properly. Additionally, the manufacturing process of LCD screens “requires sulfur hexafluoride,

a chemical substance that is believed to be responsible for 29 percent of all global warming”

[73]. The final design also contains acrylic panels. Acrylic plastic material is not easy to recycle.

In some cases, “large pieces can be reformed into other useful objects if they have not suffered

too much stress, crazing, or cracking, but this accounts for only a very small portion of the

acrylic plastic waste” [74]. Acrylic plastic is also not easily biodegradable. Considering these

factors, the final design’s use of acrylic may have a negative impact on the environment.

However, the final design will also contain PLA, which is recyclable and biodegradable. PLA

would not have any negative impacts on the environment.

6.5 Societal Influence

As a service, being able to diagnose sepsis with any form of confidence can certainly be

beneficial to helping save the lives of millions of people and would, therefore, be an incredibly

helpful addition to the world of healthcare. While not directly revolutionizing the global market,

our device certainly can help in furthering the investigation in sepsis diagnosis, laying the

groundwork for devices of similar nature. Furthermore, the limited amount of blood needed for

testing samples ensures that the risk of increased infection remains limited in an already

immuno-compromised patient.

The effect on the individual should also not be ignored. Awaiting a diagnosis can be

stressful for patients and loved ones. Faster results lessen the time spent fearing the unknown,

and ensure proper treatment is provided to increase chances of recovery. Currently, when sepsis

is suspected, treatment is provided before testing is completed. This is due to the fast-acting

nature of sepsis, where every hour of delayed treatment may increase the odds of a poor outcome

by 3-7% [75]. Receiving faster results will prevent unnecessary treatment for non-septic patients,

allowing hospital resources to be used for other patients, positively influencing others’ outcomes

when resources are limited.

Page 63: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

59

6.6 Political Ramifications

While sepsis can affect anyone, diagnosis and treatment largely depends on one’s

socioeconomic status. A recent study from January 2020 estimated the global deaths from sepsis

to be 11 million – twice the previously-believed amount. This revised estimate accounts for new

data from areas with a lower socio-demographic index (SDI), while previous estimates were

based on hospital data from high-income countries. The burden of sepsis is higher in areas with a

lower SDI, where many cases occur outside of hospitals and are unrecorded [76]. In areas of

poverty or with limited resources, our device can help in reducing the amount of sepsis-based

fatilities and can decrease the cost and turnaround time of septic tests.

6.7 Ethical Concerns

The major goal of our product is to help give patients a good and satisfying life by

accurately diagnosing sepsis in a safe and efficient way. Diagnosing sepsis early on could save

many patient lives. Although there may be concerns regarding the ethics of our product since it is

a medical device, our team followed ethical guidelines and moral principles to make sure our

product was centered around meeting the needs of patients. Our underlying motivation was to do

what was morally right and good for patients. Another important goal was to prevent harm. Our

team made sure that our product met technical standards and safety standards so that our product

was completely safe for patient use. Our product was also designed to respect the autonomy of

patients. Patients will have the right to deny using our product. Patient data will never be

released and will always remain protected under the HIPAA Privacy Rule.

6.8 Health and Safety

One of the most important tasks of our device is to help better diagnose sepsis within a

potential patient. Through the usage of the tests incorporated into our device, we can say that

such an achievement is possible in a variety of ways. As a septic patient might already be in a

weakened state, avoiding as much additional stress as possible on a person is essential. As our

system requires a relatively small blood sample size to perform tests and has a fast turnaround

time, we can ensure that no additional harm would come to a patient during the testing period.

Page 64: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

60

6.9 Manufacturability

If the device we are designing were to be marketed, we recommend that a good way to

manufacture it would be through injection molding and/or compression molding. Injection

molding is a fast manufacturing process and when mass producing a product in this manner, the

cost per item produced is relatively low. Initially the cost of the machines and the creation of

molds for startup companies is high, but once the right investments have been made, products

can be produced at a quick rate. Compression molding is another manufacturing option which

could potentially be used. One benefit of compression molding is that tooling costs are cheap,

making it good for small production runs. Also, compression molding can produce very durable

products. However, compression molding is slow and cannot be used for intricate geometry. The

device does not need to be highly durable as it will only be required to withstand small force

loads. Therefore, the most advisable method for manufacturing would be injection molding. In

order to produce the device inexpensively but also meet the design requirements, we would

advise that the device be made of plastic. Injection molding is primarily used with plastics and

thus our device could easily be made from plastic. For the initial prototyping of our device, we

would use an additive manufacturing method such as 3D printing. 3D printing can produce the

prototype model that we need, and the cost of a 3D printer is small compared to the machinery

required for mass producing. The material we would use would be PLA (Polylactic acid-

biodegradable filament) or ABS (Acrylonitrile Butadiene Styrene). These are basic and common

3D printing materials that are low cost and would be sufficient for producing a working model of

our device. We would probably opt to use PLA for prototyping as PLA tends to be more precise

because it prints at lower temperatures making it less likely to warp. The electronic hardware

involved with our device such as the vibration syllectometry circuitry and laser diodes, the 5-

inch LCD touchscreen display, and the brushless DC stepper motor, are things that have already

been produced and thus we do not need to worry about manufacturing the electronic hardware

ourselves.

Page 65: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

61

6.10 Sustainability

A great benefit of making our device primarily out of PLA is that PLA, which is a plastic,

is easily recyclable, so our product could be made at least in part from recycled material. Another

benefit of making our device primarily out of PLA is that it is biodegradable, so it is safe for the

environment. The energy costs associated with our device are very low. For one, the tests that

require electrical power can be performed relatively quickly. Secondly, the power usage required

during that time is negligible. Therefore, it is unlikely that any piece of electrical machinery will

burn out within a short period of time.

Page 66: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

62

Chapter 7: Discussion/Future

7.1 Micro-ESR

7.1.1 Implications

The micro-ESR data shows a strong positive correlation between an increase in the

amount of dextran and an increase in plasma height. Plasma height is representative of the

amount of space where fluid is present, but blood cells are not. Therefore, the higher the plasma

height the higher the aggregation. As a result of these tests, it can be concluded that an increase

in dextran increases aggregation of the red blood cells and that the micro-ESR test is capable of

recognizing this increased aggregation. For this reason, it is believed that micro-ESR is a

beneficial method in determining aggregation rate for the potential diagnosis of sepsis.

7.1.2 Limitations

Errors and limitations in the micro-ESR testing could be results from bubbles within the

dextran-blood sample, the placement of the capillary tubes within the clay stand, tools used to

measure the plasma height, the amount of time and labor it took for each test, and lastly, the

limited number of tests conducted. As the same tubes and samples used in the capillary fill test

were used for the micro-ESR tests, the movement during filling caused some of the samples to

include air bubbles. These air bubbles cause sections of the entire blood sample to aggregate

separately. If the dextran were unevenly split between sections, this could have caused more

aggregation in one section than the other.

After the capillary tubes were filled, they were pushed into a stand containing clay that

would seal the bottom of the capillary tube. During placement into the clay, the capillary tubes

were not always completely vertical. This angle could have caused the blood to aggregate at an

angle that would not be reflected during measurements. The entire test itself took four hours with

an individual taking images of the capillary tubes every thirty minutes. This always required

someone to be present in the lab and to keep track of multiple timers at once. The juggling of

trials and long hours is not ideal for someone who wants to run many trials at

once. Furthermore, as we used a phone camera for all our image collection, the resolution was

Page 67: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

63

not extremely high. This made it difficult to differentiate the top of the plasma and the start of

the aggregated blood from reflections and shadows in the capillary tube. Similar to the capillary

fill experiments, all measurements were made using ImageJ. As image length measurements

were made manually, human error in choosing locations to start and stop

measurements introduces a new area for error to be introduced. Lastly, although most results

supported the trend that an increase in dextran amount increased aggregation, the 0.10 g/dL

dextran test did not follow this trend. This anomaly could be due to any of the errors mentioned

before or could be due to not properly mixing the dextran-blood sample before use. Lastly, as

there were only three trials for all six tested dextran amounts (except for 0g/dL of dextran which

only had two) there is a limited sample size to statistically determine the accuracy of this test.

7.1.3 Future

For the future, further testing and refinement of the tests should be completed to ensure

the results of this study are reproducible. To decrease the inconsistencies between tests, the

capillary tubes should be filled in a way that ensures little to no air bubbles get caught in the

samples. This will allow for even aggregation throughout tube and more accurate measurements

of aggregation. After using the clay stand to plug the end of the capillary tubes, the capillary tube

should be placed on a stand that will ensure that the tube is held completely vertical for the

duration of the experiment. This will make the blood aggregation flat within the tube which will

allow for accurate measurements. Additionally, the camera used for tracking the height of

plasma should have a resolution high enough to allow for minute detail changes within the tube

to be tracked which will also allow for more accurate measurements. To reduce the amount of

required labor time, either the dimension of the capillary tubes could be changed, or the

collection periods could be changed. To change the collection period to every hour instead of

every 30 minutes would allow fewer required hours in the lab. Otherwise, instead of frequent

imaging, a time-lapse video recording of the capillary tube could be taken over the four hours. A

software could then be used to autonomously track the aggregation of red blood cells within the

tube. This would allow for more data points per trial and significantly reduce labor time. The

inner diameter of the capillary tube could also be changed to require a smaller amount of blood

and or to reduce the amount of blood in contact with the wall which reduces forces against

aggregation. This would also decrease the amount of labor time required. Lastly, more tests

Page 68: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

64

should be conducted. Currently, there are only three trials for each dextran-blood solution. By

increasing the number of trials, this will increase the statistical significance of results and ensure

the reliability of the findings.

7.2 Capillary Fill

7.2.1 Implications

The capillary fill tests had little to no correlation between the amount of dextran and the

flow rate time. The results were inconclusive as there were increases and decreases between the

flow rates as the amount of dextran increased. As an increase in the concentration of dextran in

the blood increases aggregation of the red blood cells, it can be concluded that there is no

relationship between aggregation and flow rate within our current setup. Therefore, it is

recommended that further analysis into the angles involved in the capillary fill test be conducted

using the final design prototype to determine the viability of this method in terms of RBC

aggregation.

7.2.2 Limitations

As previously explained, the capillary fill tests cannot be applied to the final design as the

results could not provide any correlation between aggregation and flow rate. This inability to

find a correlation could be a result of inconsistencies due to the materials used, human error, and

the limited number of tests conducted. For our tests, we intended to mechanically set the angle

each tube was held at. The angle for testing was intended to be positive. But in preliminary tests,

it was found that to keep the angle consistent between all tests, the angle had to be negative to

allow for capillary flow. To create this negative testing angle, the capillary tube was held by

hand between two stands. This did allow for an angle that was generally consistent, but variation

was still present that could have been minimized with a mechanical device. This also introduced

increased movement of the capillary tube as it was held by a human. This movement could have

disrupted flow, changing the results of the study. The capillary tubes could have also had

variation in their hydrophilic properties. Although each tube was sterilized, manufacturing of

each tube could have been inconsistent in the coatings. Thus, different capillary tubes could have

had different hydrophilic properties which would have helped or hindered capillary flow of the

blood in the tube.

Page 69: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

65

Placing the blood on a latex glove for a hydrophobic surface was also not ideal. The

latex glove was originally hydrophobic, though it still has some pores. But after repeated use, the

hydrophobicity of the gloves decreased, causing the contact angle of the blood droplet to

decrease between tests. During some tests, bubbles were present in the blood that disrupted flow.

For some other tests, the capillary tube had to be placed into the blood droplet multiple times

before flow could be generated. In relation to the specific dextran tests, the 0 g/dL tests were

unable to be conducted and, due to large differences in angles, only two tests, out of three, could

be analyzed for 0.2g/dL and 0.1g/dL concentrations.

Lastly, there were only three trials for each tested dextran solution which is not a large

enough study to provide statistically significant results. The measuring of each test was done by

measuring the travelled distance of the blood in capillary tubes in four frames from the recorded

video. The video of the trial’s resolution was too low to see minute changes thereby making the

measuring of the blood’s location difficult. As the measuring was done using ImageJ, each line

was drawn by a person. As people are not precise with any consistency, the exact measuring of

the blood location was a variable to consider. These inconsistencies, and limited trials in testing

the dextran-blood solutions, were mainly due to the shortened testing times resulting from the

shutting of the lab because of the COVID-19 pandemic. Without this interruption, many changes

to our testing protocol would have been made to improve reliability of the study.

7.2.3 Future

For future testing of the relationship between capillary fill and aggregation, many

changes would be made to the materials and automation of testing would have been used. To

address the inconsistent angle and location of the capillary tube, a mechanical stand could be

built to introduce the capillary tube to the blood droplet at a set angle and then hold the capillary

tube still for the entirety of the trial, similar to what is present in the final device prototype. This

would eliminate the effect of movement of the capillary tube on the flow rate of the blood and

ensure no air bubbles would be introduced to the samples. The tube itself could also be plasma

cleaned to create consistent surfaces between capillary tubes. This would have ensured that the

blood stuck to the tube to relatively the same degree between each test providing the same

potential for each blood sample to flow. To increase hydrophobicity and consistency of the

surface that the blood is placed on, a glass slide coated with Acryl-Glide or a similar agent that

Page 70: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

66

creates a hydrophobic surface should also be used. Using this glass slide would ensure a flat

surface for every test. The coating would also induce a consistent contact angle for the blood

samples; by having a consistent contact angle, the pressure difference within the capillary tube

would be the same for every test.

The measuring of the capillary fill should also be automated. This could be done by using

a higher resolution camera, allowing for clearer pictures to be analyzed, and in conjunction with

a software that could track the location of the blood in the tube throughout the video. This would

also significantly reduce the amount of time to analyze each trial within each dextran-blood

solution test. Therefore, more tests could be conducted and analyzed in a shorter amount of

time. This would also reduce variation between test operators and allow the test to be conducted

with less operator influence. In the future, the testing should encompass all the dextran-blood

solutions. By including the 0 g/dL dextran-blood solution, there would be more comparable data.

This would allow for any anomalies or trends to be spotted. Lastly, every dextran-blood solution

should include more than three trials. A larger sample size would create statistically significant

results. With these more accurate and precise tools, the inconsistencies between tests could be

significantly decreased. With this new set of data, the usefulness of capillary fill in determining

increased aggregation would be re-evaluated.

7.3 Syllectometry

7.3.1 Implications

The main focus of the laser syllectometry test was to develop a way of measuring blood

aggregation via an experimental and unconventional method of RBC disaggregation and use of

light scattering. The notion present was that, if blood aggregation was occurring at a faster rate,

you would see the intensity of transmitted light increase at a faster rate and therefore a higher

voltage climb would be recorded. Being that the resolution for capturing data was relatively low

and, initially, very sensitive to outside light sources, the average min-max difference as well as

the rate of change of volts over time was key for our analysis.

Page 71: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

67

Referring back to Table 4 (which summarized the syllectometry results and can be found

on pg. 48), it can clearly be seen that increasing the concentration of the dextran solution leads to

an increase in the min-max voltage difference and in the rate of change. As dextran has been

found to increase RBC aggregation, when the concentration of dextran increases, an expected

increase in aggregation should also be present. If aggregation is occurring at a faster rate, an

increase in rouleaux should also be expected, causing less light to be scattered through the

sample and more directly hitting the photodiode. Therefore, based on the data present in Table

11, we can say that laser syllectometry is a valid method for testing blood aggregation.

It is important to note that, while informative, the “control” row isn’t accurately involved

with the data present in the 0.025g/dL tests to the 0.2g/dL. Ideally, this data was meant to serve

as a basis for confirmed septic blood; this will be further discussed in “7.3.2 Limitations” and

“7.3.3 Future”.

7.3.2 Limitations

One of the largest limitations on the syllectometry tests, as with many of the other tests,

was the inability to be present in the lab during the entirety of D-term due to the COVID-19

global pandemic. Given that lab time was hindered during the final term, limited testing data was

recorded; this can be seen in Appendix H for all the test data, noting how some tested dextran

concentrations have one- or two-sample tests. Another hindrance, in relation to the pandemic,

was the inability to re-test data using the improved device design. Some of the flaws recognized

within the design were corrected – as mentioned in section “6.1 Final Design CAD Prototype” –

and would therefore allow for more precision with the produced results. An improved blackout

system, threshold voltage control, and resolution adjustments would have made for significantly

more stable outcomes with each dextran solution. Unfortunately, this could not be achieved and

would certainly be a recommended point of continuation if possible.

Finally, in relation to the syllectometry test itself, while the data present does show an

increase in blood aggregation as reflected by a change in voltage over time, it doesn’t speak to

any certainty on whether a patient absolutely has sepsis. The blood sample used has the potential

to display a variety of other conditions that a patient might be dealing with besides sepsis. This

Page 72: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

68

test is meant to work in conjunction with the other tested methods to help improve the accuracy

of a sepsis diagnosis and is therefore not recommended as a standalone test for diagnosis.

7.3.3 Future

When conducting additional tests using laser syllectometry, there are a few instances that

the team would like to test. For one, the inability for the team to acquire valid samples of septic

blood was a major hindrance and is what resorted to the use of dextran solutions. Being able to

use clinically certified septic blood in conjunction with that received from AllCells would allow

the team to speak with more certainty regarding the success of the laser syllectometry test.

Furthermore, if testing were to continue, it would be ideal to use the updated device design for

both increased precision and control of testing parameters, leading to fewer variables present in

the data.

Page 73: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

69

7.4 Device Discussion

As mentioned in chapter 4, a few of the tests carried out did not require extensive

fabrication of a device or system. However, the final device design provides a platform to help

carry out the tests in a more controlled fashion. For instance, the final design allows us to hold

capillary tubes at desirable angles with increased precision than what someone would be able to

perform manually. Furthermore, the increased parameters regarding the syllectometry test allow

more control over the data being recorded. There are, however, some limitations to the amount of

control the device has on certain things. One such limitation is that the device is subject to

whatever external conditions are present within the testing environment. These include, but are

not limited to, temperature, moisture, ambient light, and vibrations. Therefore, it is recommended

that tests be conducted using the device in a controlled laboratory setting to minimize extraneous

effects during the testing periods. Another limitation to the device is that it doesn’t have any

features that guarantee keeping a person sterile from the blood they are working on. When

conducting tests with our device, users would need to take the necessary precautions to make

sure they are working safely with a blood sample. These precautions include wearing gloves and

cleaning surfaces that may need to be cleaned. The single-use disposable nature of the glass

slides and capillary tubes do help to reduce the cleaning required between tests, allowing for a

faster testing turnabout than a device that needs a deep cleaning with each use.

Page 74: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

70

Chapter 8: Conclusions and Recommendations

8.1 Conclusions

Sepsis is a life-threatening condition that kills millions each year. Many physicians

consider diagnosis the most difficult aspect of sepsis, as due to its wide range of symptoms with

high variability, there exists no conclusive test for the disease. Existing tests face the issues of

low sensitivity and specificity, slow turnaround times, and uncertain results. Physicians must

therefore rely on experience and a range of tests and symptoms when attempting to diagnose a

patient. Backed by extensive literature review, the team investigated the possibility of measuring

erythrocyte aggregation as a sepsis diagnosis tool.

Of the three aggregation measurement methods tested, micro-ESR and laser

syllectometry were capable of differentiating levels of aggregation among blood samples

containing various concentrations of dextran in PBS with a degree of confidence. The final

design focused on building around laser syllectometry (as it was the most complex in terms of

tools and equipment) while also featuring an attachment that would assist in further testing the

methods of micro-ESR and capillary fill with greater precision. Therefore, using at least two of

the three proposed testing methods, the final device design creates a system that allows for the

possible diagnosis of sepsis via red blood cell aggregation.

8.2 Recommendations

The team was successful in identifying two methods capable of measuring blood

aggregation, but these results could be further improved through recommended adjustments to

the experimental procedures and direct validation with septic blood.

For the capillary fill tests, further testing had been planned using plasma-treated capillary

tubes and drops of blood placed on glass slides coated in Acryl-Glide, a hydrophobic substance.

These were expected to increase blood adhesion to the capillary tubes and drive capillary action

more strongly, possibly allowing measurements to be made at a positive angle. This would allow

for existing theory to quantitatively describe blood flow, and better differentiate the effects of

Page 75: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

71

capillary action from gravity. This may lead to a correlation between RBC aggregation and flow

rate whereas none was observed in the current tests.

The final design contained improvements to all three tests, allowing more accurate and

precise control of capillary tube angles for the micro-ESR and capillary fill tests and

improvements on laser alignment for the syllectometry test. The team was unable to fabricate

and validate this design to assess its functionality or use it to conduct further trials. This,

combined with the limited number of tests that could be performed, leads to our recommendation

that further testing be conducted to validate the ability of each test to measure RBC aggregation

using the newly created final design.

Finally, the team recommends validation of these tests with septic blood. The team could

not obtain septic blood for testing within a reasonable time period and instead induced

aggregation using dextran. Ideally, tests would be conducted with septic and non-septic blood

samples from a variety of people, including those with non-septic inflammatory conditions, to

determine the specificity and sensitivity of our tests.

Page 76: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

72

References

[1] National Institute of General Medical Sciences. (2019, September 27). Sepsis. Retrieved

October 11, 2019, from

https://www.nigms.nih.gov/education/pages/factsheet_sepsis.aspx

[2] Davis, C. (2019, July 12). Sepsis (Septicemia) Diagnosis, Causes, Treatment &

Symptoms. Retrieved from https://www.medicinenet.com/sepsis/article.htm

[3] Chakraborty, R. & Burns, B. (2020, March 28). Systemic Inflammatory Response

Syndrome. StatPearls - PubMed-NCBI. Retrieved from

https://www.ncbi.nlm.nih.gov/books/NBK547669/

[4] Cirino, E. (2017, July 18). Blood Poisoning: Symptoms, Signs, Causes, and Treatment.

Retrieved from https://www.healthline.com/health/blood-poisoning

[5] O'Connell, K., & Higuera, V. (2018, August 31). Sepsis: Symptoms, Causes, Treatment,

Risks & More. Retrieved from https://www.healthline.com/health/sepsis

[6] Leon, A. (2013). Clinical course of sepsis, severe sepsis, and septic shock in a cohort of

infected patients from ten Colombian hospitals (doi: 10.1186/1471-2334-13-345).

Retrieved from https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3727953/

[7] De Prost, N., Razazi, K., & Brun-Buisson, C. (2013). Unrevealing culture-negative

severe sepsis. Critical Care, 17(5), 1001. doi:10.1186/cc13022/

[8] Bone, R.C. et al. (1992). Definitions for sepsis and organ failure and guidelines for the

use of innovative therapies in sepsis. The ACCP/SCCM Consensus Conference

Committee... - PubMed - NCBI (DOI: 10.1378/chest.101.6.1644). Retrieved from

https://www.ncbi.nlm.nih.gov/pubmed/1303622

[9] UK Sepsis Trust. (2018, June 17). Post Sepsis Syndrome. Retrieved from

https://sepsistrust.org/get-support/support-for-survivors/post-sepsis-syndrome/

[10] "Testing For Sepsis," 2019. [Online]. Available: https://www.sepsis.org/sepsis-

basics/testing-for-sepsis/

Page 77: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

73

[11] Cantey, J. B., & Baird, S. D. (2017). Ending the Culture of Culture-Negative Sepsis in

the Neonatal ICU. Pediatrics, 140(4), e20170044. doi:10.1542/peds.2017-0044

[12] Matson, A., Soni, N., & Sheldon, J. (1991). C-reactive Protein as a Diagnostic Test of

Sepsis in the Critically Ill. Anaesthesia and Intensive Care, 19(2), 182-186.

doi:10.1177/0310057x9101900204 [11]

[13] McGill Physiology Virtual Lab. (n.d.). Blood Laboratory: Hemostasis: PT and PTT tests.

Retrieved from https://www.medicine.mcgill.ca/physio/vlab/bloodlab/PT_PTT.htm

[14] “Partial Thromboplastin Time (PTT) Test: Purpose and Procedure.” [Online]. Available:

https://www.healthline.com/health/partial-thromboplastin-time-ptt. [Accessed: 06-May-

2020].

[15] National Institutes of Heath, U.S. Department of Health and Human Services, National

Cancer Institute, USA.gov, Biomarker, National Cancer Institute.

[16] M. D. Gonsalves and Y. Sakr, “Early Identification of Sepsis,” Current Infectious

Disease Reports, vol. 12, no. 5, pp. 329–335, Sep. 2010.

[17] M. Barati et al., "Comparison of WBC, ESR, CRP and PCT serum levels in septic and

non-septic burn cases," in Burns, vol. 34, Tehran, Iran University of Medical Science,

2008, pp. 770-774.

[18] "Procalcitonin: a promising diagnostic marker for sepsis and antibiotic therapy," Journal

of Intensive Care, pp. 5-51, 2017.

[19] Roche Diagnostics International Ltd., Biomarkers and Sepsis, Switzerland: COBAS,

2012.

[20] Biomerieux, "FDA.gov," 10 November 2016. [Online]. Available:

https://www.fda.gov/media/100863/download. [Accessed 28 September 2019].

[21] "VIDAS B.R.A.H.M.S PCT," BioMerieux Diagnostics, 2019. [Online]. Available:

https://www.biomerieux-diagnostics.com/vidasr-brahms-pct. [Accessed 28 September

2019].

Page 78: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

74

[22] S. Harbarth et al., "Diagnostic Value of Procalcitonin, Interleukin-6, and Interleukin-8 in

Critically Ill Patients Admitted with Suspected Sepsis," American Thoracic Society

Journal, vol. 164, no. 3, pp. 396-402, 2001.

[23] K. MM, R. Kusters, I. MJ and S. LM, "A PCT algorithm for discontinuation of antibiotic

therapy is a cost-effective way to reduce antibiotic exposure in adult intensive care

patients with sepsis," Journal of Medical Economics, pp. 944-53, 2015.

[24] A. Farooq and J. M. Colon-Franco, "Procalcitonin and its Limitations: Why a

Biomarker's Best Isn't Good Enough," The Journal of Applied Laboratory Medicine,

January 2019.

[25] I. Samsudin and S. D. Vasikaran, "Clinical Utility and Measurement of Procalcitonin,"

The Clinical Biochemist Reviews, pp. 59-68, April 2017.

[26] R. Venkataraman and J. A. Kellum, "Sepsis: Update in Management," Advances in

Chronic Kidney Disease, vol. 20, no. 1, pp. 6-13, 22 December 2012.

[27] ACRO Biosystems, "Il-Molecule Information," ACRO Biosystems, 2017. [Online].

Available: https://www.acrobiosystems.com/L-326-IL-6.html. [Accessed 5 October

2019].

[28] T. Tanaka, M. Narazaki and T. Kishimoto, "Il-6 in Inflammation, Immunity and

Disease," Cold Spring Harbor Perspectives in Biology, vol. 6, no. 10, October 2014.

[29] Roche Diagnostics International Ltd., Biomarkers and Sepsis, Switzerland: COBAS,

2012.

[30] L. Ma et al., "Role of interleukin-6 to differentiate sepsis from non-infectious systemic

inflammatory response syndrome," Cytokine, vol. 88, pp. 126-135, December 2016.

[31] L. Bo, C. Yun-Xia, Y. Qin, Z. Yun-Zhou and L. Chun-Sheng, "Diagnostic value and

prognostic evaluation of Presepsin for sepsis in an emergency department," Critical Care,

vol. 17, no. 5, 2013.

[32] Q. Zou, W. Wen and X.-c. Zhang, "Presepsin as a novel sepsis biomarker," World

Journal of Emergency Medicine, vol. 5, no. 1, pp. 16-19, 2014.

Page 79: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

75

[33] "Creative Diagnostics," Creative Diagnostics Inc, 2009. [Online]. Available:

https://www.creative-diagnostics.com/Chemiluminescence-Immunoassay-guide.htm.

[Accessed 29 September 2019].

[34] T. Shozushima, G. Takahashi, N. Matsumoto, M. Kojika, E. Shigeatsu and Y. Okamura,

"Usefulness of presepsin (sCD14-ST) measurements as a marker for the diagnosis and

severity of sepsis that satisfied diagnostic criteria of systemic inflammatory response

syndrome," Journal of Infection and Chemotherapy, vol. 17, no. 6, pp. 764-769, 2011.

[35] LSI Medience Corporation, "The Sepsis Marker: PATHFAST Presepsin," Mitsubishi

Chemical Europe, [Online]. Available: https://www.pathfast.eu/presepsin. [Accessed 29

September 2019].

[36] Z. Zheng, L. Jiang, L. Ye, Y. Gao, L. Tang and Z. Zhang, "The accuracy of presepsin for

the diagnosis of sepsis from SIRS: a systemic review and meta-analysis," Annals of

Intensive Care, pp. 5-48, 8 December 2015.

[37] 2020. [Online]. Available: https://www.researchgate.net/figure/Incident-light-on-a-

particle-Scattering-occurs-by-refraction-diffraction-and-reflection_fig6_266890765.

[Accessed: 14- May- 2020]

[38] "Membrane Characterisation Equipment | Particle Size Measurements > 100 nm |

European Membrane Institute (EMI) - University of Twente", Universiteit Twente, 2020.

[Online]. Available: https://www.utwente.nl/en/tnw/emi/characterisation/particle-size-

large/. [Accessed: 08- May- 2020]

[39] “Laser diffraction for particle sizing :: Anton Paar Wiki,” Anton Paar. [Online].

Available: https://wiki.anton-paar.com/en/laser-diffraction-for-particle-sizing. [Accessed:

07-May-2020].

[40] "Circular Aperture Diffraction", Hyperphysics.phy-astr.gsu.edu, 2020. [Online].

Available: http://hyperphysics.phy-astr.gsu.edu/hbase/phyopt/cirapp2.html. [Accessed:

07- May- 2020]

[41] Horiba.com. (2019). Fundamental Principles of Laser Diffraction - HORIBA. [online]

Available at: https://www.horiba.com/us/en/scientific/products/particle-

Page 80: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

76

characterization/technology/laser-diffraction/fundamentals-of-diffraction/ [Accessed 6

Oct. 2019].

[42] O. Baskurt, A. Temiz Artmann, and H. Meiselman, “Red blood cell aggregation in

experimental sepsis,” J. Lab. Clin. Med., vol. 130, pp. 183–90, Aug. 1997.

[43] O. K. Baskurt and H. J. Meiselman, “Blood Rheology and Hemodynamics,” Semin

Thromb Hemost, vol. 29, no. 05, pp. 435–450, Nov. 2003.

[44] M. Piagnerelli, K. Z. Boudjeltia, M. Vanhaeverbeek, and J.-L. Vincent, “Red blood cell

rheology in sepsis,” Intensive Care Med., vol. 29, no. 7, pp. 1052–1061, Jul. 2003.

[45] N. Németh et al., “Early hemorheological changes in a porcine model of intravenously

given E. coli induced fulminant sepsis,” Clin. Hemorheol. Microcirc., vol. 61, no. 3, pp.

479–496, Jan. 2016.

[46] Z. Isiksacan, O. Erel, and C. Elbuken, “A portable microfluidic system for rapid

measurement of the erythrocyte sedimentation rate,” Lab. Chip, vol. 16, no. 24, pp.

4682–4690, 2016.

[47] M. Eberl and M. Davey, "Neutrophils," 2019. [Online]. Available:

https://www.immunology.org/public-information/bitesized-

immunology/cells/neutrophils.

[48] J. Cavenagh, "White Blood Cells," Surgery, pp. 61-64, 2007.

[49] "Lymphocyte: Blood Cell," 2019. [Online]. Available:

https://www.britannica.com/science/lymphocyte.

[50] F. Ellet et al., "Diagnosis of sepsis from a drop of blood by measurement of spontaneous

neutrophil motility in a microfluidic assay," Nature Biomedical Engineering, 2018.

[51] X.-F. Shen, K. Cao, J.-p. Jiang, W.-X. Guan and J.-F. Du, "Neutrophil dysregulation

during sepsis: an overview and update," Journal of Cellular and Molecular Medicine , pp.

1687-1697, 2017.

[52] J. D. Lang and G. Matute-Bello, "Lymphocytes, apoptosis and sepsis: making the jump

from mice to humans," BioMed Central, 2009.

Page 81: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

77

[53] P. Forget, C. Khalifa, J.-P. Defour, D. Latinne, M.-C. Van Pel and M. De Kock, "What is

the normal value of the neutrophil-to-lymphocyte ratio?," BMC Research Notes, 2017.

[54] "Motility," Biology Dictionary , [Online]. Available:

https://biologydictionary.net/motility/. [Accessed 5 October 2019].

[55] S. Muldur, A. L. Marand, F. Ellett and D. Irimia, "Measuring spontaneous neutrophil

motility signatures from a drop of blood using microfluidics," Methods Cell Biology ,

vol. 147, pp. 93-107 , 2018.

[56] “ISO 13485:2016,” ISO, 21-Jan-2020. [Online]. Available:

https://www.iso.org/standard/59752.html. [Accessed: 01-May-2020].

[57] International Organization for Standardization, "ISO 11737-2:2009 - Sterilization of

medical devices — Microbiological methods — Part 2: Tests of sterility performed in the

definition, validation and maintenance of a sterilization process," 2009. [Online].

Available: https://www.iso.org/obp/ui/#iso:std:iso:11737:-2:ed-2:v1:en.

[58] International Organization For Standardization, "ISO 10993-1:2018 - Biological

evaluation of medical devices — Part 1: Evaluation and testing within a risk management

process," 2018. [Online]. Available: https://www.iso.org/obp/ui/#iso:std:iso:10993:-1:ed-

5:v2:en.

[59] International Organization For Standardization, "ISO/TC 76 - Transfusion, infusion and

injection, and blood processing equipment for medical and pharmaceutical use," [Online].

Available: https://www.iso.org/committee/50044.html.

[60] International Organization For Standardization, "ISO 10993-11:2017 - Biological

evaluation of medical devices — Part 11: Tests for systemic toxicity," 2017. [Online].

Available: https://www.iso.org/obp/ui/#iso:std:iso:10993:-11:ed-3:v1:en.

[61] U.S. Department of Health & Human Services (HHS), "The HIPAA Privacy Rule,"

[Online]. Available: https://www.hhs.gov/hipaa/for-professionals/privacy/index.html.

[62] U.S. Department of Health & Human Services (HHS), "Your Rights Under HIPAA,"

[Online]. Available: https://www.hhs.gov/hipaa/for-individuals/guidance-materials-for-

consumers/index.html.

Page 82: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

78

[63] “A classic, gold standard: The Westergren method for ESR measurement,” RR

Mechatronics, Aug. 2015.

[64] “Erythrocyte Sedimentation Rate (ESR),” Lab Tests Online. [Online]. Available:

https://labtestsonline.org/tests/erythrocyte-sedimentation-rate-esr. [Accessed: 18-Nov-

2019].

[65] R. Hashemi, A. Majidi, H. Motamed, A. Amini, F. Najari, and A. Tabatabaey,

“Erythrocyte Sedimentation Rate Measurement Using as a Rapid Alternative to the

Westergren Method,” Emerg, vol. 3, no. 2, pp. 50–53, 2015.

[66] Libretexts, “Some Properties of Liquids,” Chemistry LibreTexts, 13-Apr-2020. [Online].

Available:

https://chem.libretexts.org/Bookshelves/General_Chemistry/Map:_Chemistry_-

_The_Central_Science_(Brown_et_al.)/11:_Liquids_and_Intermolecular_Forces/11.3:_S

ome_Properties_of_Liquids. [Accessed: 06-May-2020].

[67] J. L. G. Santander and G. Castellano, “Determination of the kinematic viscosity by the

liquid rise in a capillary tube,” Revista Brasileira de Ensino de Física, vol. 35, no. 3,

2013.

[68] D. J. Patel and R. N. Vaishnav, “Some elementary hemodynamic concepts,” in Basic

hemodynamics and its role in disease processes. Baltimore: University Park Press, 1980,

ch. 2, pp. 83–90.

[69] S. Shin, J. H. Jang, M. Park, Y. Ku, and J. S. Suh, “A noble RBC aggregometer with

vibration-induced disaggregation mechanism,” 2005.

[70] S. Shin, J. H. Jang, M. S. Park, Y. H. Ku, and J. S. Suh, “Light-transmission

aggregometer using a vibration-induced disaggregation mechanism,” Review of Scientific

Instruments, vol. 76, no. 1, p. 016107, Dec. 2004.

[71] E. Yeom, Y. J. Kang, and S.-J. Lee, “Changes in velocity profile according to blood

viscosity in a microchannel,” Biomicrofluidics, vol. 8, no. 3, 2014.

[72] BioSpace, “Reducing False Positive Results for Sepsis with Steripath Improves

Economic and Clinical Outcomes, Study Reports,” BioSpace, 17-Apr-2019. [Online].

Page 83: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

79

Available: https://www.biospace.com/article/releases/reducing-false-positive-results-for-

sepsis-with-steripath-improves-economic-and-clinical-outcomes-study-reports/.

[Accessed: 06-May-2020].

[73] Makandi, P. (2020). Negative Effects of LCD Monitors. Retrieved from

https://itstillworks.com/negative-effects-lcd-monitors-7478611.html

[74] How Products Are Made (2020). Acrylic Plastic. Retrieved from

http://www.madehow.com/Volume-2/Acrylic-Plastic.html

[75] C. W. Seymour et al., “Delays From First Medical Contact to Antibiotic Administration

for Sepsis,” Critical Care Medicine, vol. 45, no. 5, pp. 759–765, 2017.

[76] K. E. Rudd et. al., “Global, regional, and national sepsis incidence and mortality, 1990–

2017: analysis for the Global Burden of Disease Study,” The Lancet, vol. 395, no. 10219,

Jan. 2020.

Page 84: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

80

Appendix A – Interviews

Dr. Kate Madden, MD, MMSc Associate in Critical Care Medicine Boston

Children’s Hospital and Instructor in Anesthesia, Harvard Medical School What is your experience with sepsis or SIRS-related illnesses in?

As an attending physician in a pediatric intensive care unit (PICU), I care for many

patients with suspected or proven sepsis. I have also been involved in some multicenter

initiatives to increase consistency of response to suspected sepsis in terms of the actions of

doctors and nurses. Through these collaborative, we have instituted tools to help doctors and

nurses identify sepsis, communicate about the plan for care, and make informed decisions about

the management of suspected sepsis.

What has your work revolved around?

Mostly covered above. PICU and hospital-wide education about sepsis and need for

timely care, development of tools to help nurses and doctors identify sepsis and communicate

about the plan.

Can you please explain what are how sepsis is currently diagnosed in the hospital? What steps

are involved? And who makes the final call?

Sepsis is a set of clinical signs, and not truly a diagnosis per se. We have many patients in

whom we may suspect sepsis, but a large proportion do not end up having a clear diagnosis.

Only a small proportion have a bacteria or virus grow from a lab test that confirms the diagnosis.

For research and quality improvement purposes, we use the judgement of the clinical team,

usually the attending physician, as well as the specific treatments that go along with sepsis.

Who performs these steps? What qualifications are required?

Not sure if above answers this? What steps specifically?

How successful are these current protocols in diagnosing sepsis?

It depends how you define success. We currently want to screen a large number

of patients for a few that might really have sepsis, as we know it is a very morbid condition and

we don’t want to miss any patients. As a result, we screen many patients who do not really have

sepsis, just fever and maybe one other SIRS criteria, but in the end will not develop severe sepsis

or septic shock.

What are the challenges with these current hospital protocols?

As with any large hospital initiatives – getting engagement from all different disciplines

involved, changing culture, the burden of documentation on nursing especially, communication

challenges.

What are the limitations when working with patients when diagnosing sepsis? How do infants

differ from adults (if any)?

Not sure exactly what this refers to? Infants are certainly much more difficult to assess in

terms of mental status, focal symptoms, etc. We tend to screen very broadly for all infants who

might have sepsis, because it can be quite easy to miss in young infants.

Page 85: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

81

What is the largest challenge when diagnosing sepsis in a hospital setting?

Many challenges, but probably the biggest is distinguishing fever and tachycardia (very

common) from true sepsis that might progress to severe sepsis or septic shock.

What would you like to see in future sepsis diagnostic tools?

More objective tools to diagnose sepsis at the bedside – measures of oxygen delivery to

the tissues, organ function, and rapid microbiologic diagnostics. In pediatrics, we have a lot of

challenges assessing mental status and thus brain perfusion, especially in young patients, so

some bedside objective measure would be very helpful. Folks are working on bedside tools to

evaluate for response from fluid resuscitation, including ultrasound tools, which could be helpful

as we see the side effects of excessive fluid resuscitation in these patients as well. Knowing

specific microbiological data would help tailor therapy and anticipate outcomes.

How would patient care be changed with faster diagnostic methods?

More accurate and timely management.

Dr. Jouha Min, PHD, Postdoctoral Fellow in Center for Systems Biology at

Mass General Hospital What is your experience with sepsis or SIRS-related illnesses?

 Sepsis is an often-fatal condition that arises when the body launches an overwhelming immune

response to an infection that causes more damage to the body than the infection itself. A critical

unmet need in combating sepsis is the lack of accurate early biomarkers that can alert clinicians

to this potential life-threatening situation and allow them to take preventative action. 

Why did you choose to research sepsis?

When I joined the Center for Systems Biology at MGH, I learned about the IL-3 work

that just got published by another PI at the Center, Dr. Filip Swirski, and I as an engineer thought

that it would be a straightforward project to design and develop a point-of-care device to

measure IL-3 levels in the patient’s blood to “predict” sepsis.

How did you choose what biomarker to use in your diagnostic method?

Did you consider IL-6,8 or 10? Or Procalcitonin?

 The IBS (integrated biosensor for sepsis) prototype measures levels of a protein called cytokine

interleukin-3 (IL-3) in the blood. We (our collaborator at the center) have IL-3 as an independent

predictor of septic shock and death being produced by innate response activator (IRA) B cells

following TLR activation. IL-3 operates upstream of key cytokines including TNFα, IL-1β and

IL-6; high IL-3 level and can trigger a detrimental cytokine storm. Measuring IL-3 thus can give

an early window to monitor immune responses through minimally invasive blood testing. 

Why did you choose Interleukin-3 as your diagnostic method?

Refer to my answer to Question #3. 

Where did you get your ideas for your Interleukin-3 device?

How it works:  Simply put, the platform uses magnetic beads to directly and quickly

extract target protein IL-3 from blood samples. The beads are adorned with antibodies and

Page 86: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

82

electron mediators, and through a series of reactions, generate an electrical current which

provides an analytical readout of IL-3 levels. 

Why magnetic beads approach: We advanced a portable biosensor consisting of a

disposable kit for blood processing and an electrical detection system. The kit is used to capture

IL-3 on magnetic beads and label it for electrochemical reaction; the detector then measures

electrical currents for IL-3 quantification. This strategy has practical advantages: i) target protein

(IL-3) can be enriched directly from blood; ii) the assay achieves high detection sensitivity

through magnetic enrichment and enzymatic signal amplification; iii) based on the electrical

measurements, the sensor can be easily miniaturized and easy-to-use.

What skills sets are required to use your device and who was the intended user?

It was originally intended to be used by doctors and/or nurse practitioners in the

emergency room or intensive care unit. But with further improvement (i.e., full automation), any

technician (or even a caregiver) would be able to use it. 

Would you consider combining different methods (such as Procalcitonin and IL-3)?

Yes, our next generation device will be able to measure multiple markers (e.g., IL-3,

TNFα, IL-1β, PCT, IL-6) for even more robust characterization of host response, further

improving diagnostic accuracy.

What do you feel are aspects of sepsis that get overlooked?

Medical doctors, scientists, and engineers are working their best to overcome the unmet

needs in combating sepsis. But its such complex pathophysiology of systemic nature hampers us

from inventing new ways to diagnose and/or treat. Long story short, we need a better scientific

understanding of sepsis (i.e., mechanisms/pathways, new biomarkers). 

How did you decide what needs were to be fulfilled by your project?

Based on discussions with some doctors specializing infectious diseases as well as

literature research, it was so evident what the unmet need is for sepsis.

What is the goal for sepsis diagnostics in the future?

The immediate need for sepsis diagnostics is to develop a reliable clinical tool that could

be readily integrated into clinical workflows, enabling timely diagnosis and proactive treatment

of sepsis. 

Dr. Michael Puskarich, MD, Associate Professor, University of Minnesota What is your experience with sepsis or SIRS-related illnesses? 

Clinically I am an emergency physician, so I see these patients regularly and have

to differentiate patients with sepsis from those with non-infectious etiologies on every one of my

shifts.  This has become an increasingly important problem not only for patient care, but also

with the advent of the CMS Sepsis Core measure. I have been intimately involved in

developing internal EMR-based screening and recognition tools for sepsis which are terribly un-

specific. I have also published extensively on the relationship between SIRS, sepsis, and various

consensus definitions of the disease. I have also been involved in many sepsis trials that use

these criteria for inclusion in the study.

Page 87: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

83

What has your work revolved around? 

 I am most interested in the role of metabolism in sepsis as a therapeutic target, related to my

work describing the prognostic value of lactate. With this, as well as clinical trial work in the

field in general, I have observed (along with many in the field) that the clinical consensus

definitions suffer from poor specificity and variable sensitivity for the disease.  The clinical

presentation of 

sepsis differs from animal models leading to failure to translate animal results to improvements

in human health. Part of my work is involved in finding ways to disambiguate early sepsis

presentations from non-infectious etiologies, so as to only select patients most likely to benefit

from a therapy prior to enrollment in a clinical trial. 

What are the current methods utilized to diagnose sepsis? Is it successful? 

Clinically SIRS, "sepsis-3", procalcitonin occasionally have all been tried and they are

not consistently successful. These tools are primarily ways for clinicians to make sure they are

"getting the points" for treatment of sepsis patients and meeting CMS core measure success,

but suffers from terrible usability. For true patient care, good clinicians rely on gestalt and

experience combined with overall lab and diagnostic test results to diagnose sepsis more than

any single tool, which in and of itself is also problematic due to the highly variable nature and

clinical presentation of the disease. Sepsis is also confusing because it exists on a spectrum - it's

not a "yes / no" despite many people wanting it to be.  You don't go from routine flu to "sepsis,"

it exists on a spectrum with other infectious diseases that the body does or does not control on its

own. 

Who carries out these sepsis diagnostic tests? What skills sets are required? 

Generally, the used tools are based on vital signs and laboratory tests, including white

blood cell count, measure of organ failure (creatinine, bilirubin, platelets), mental status, lactate

as an indicator of disease severity, and others. An astute clinician is needed for excellent

diagnostic acumen. Sometimes labs and vitals outperform the clinician, sometimes vice versa. 

Set standardized criteria almost always suffer from a severe over sensitivity and poor specificity.

The consequence of this if applied too broadly is that by diagnosing patients without sepsis with

sepsis ("just to be safe"), it diverts attention and resources of the doctor towards these patients 

who do not necessarily have sepsis and diverts it from other sick patients being cared for

simultaneously by the doctor. In the emergency department, we frequently managed 10-20

patients simultaneously, so poor specificity tests have a real cost to OTHER patients in the

department that is not considered in sepsis care mandates. Meanwhile, poor sensitivity tests lead

to potentially missing patients that can have consequences that include death, as we know sepsis

is a time sensitive disease process. 

What is the goal for sepsis diagnostics in the future? 

Tests with an excellent balance of both sensitivity and specificity to know who time sensitive

treatment must be delivered to ASAP. Alternatively, diagnostic tests that reliably determine

the causative organism and / or pathophysiologic process that predominates in the patient at any

given time could lead to tailored / personalized therapies. 

Page 88: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

84

 What is the largest issue when diagnosing sepsis in a hospital setting? What are the biggest

challenges with the current hospital methods? 

See above. 

What symptoms prompt initial sepsis testing? 

 Fever (or hypothermia) with hypotension is the most obvious.  Fever with other signs of

hemodynamic instability (tachycardia, high respiratory rate, high or low white blood cell count)

can increase clinical suspicion, particularly if it persists after giving IV fluids. In elderly patients

and immunosuppressed patients there is not always a fever, and malaise or fatigue can be the

first symptoms. Some patients have respiratory distress without fever.  Most patients with a

fever, however, do not have sepsis.  In general, sepsis is considered in patients having: a) a fever,

b) hypothermia, c) hemodynamic instability, and d) vague symptoms in an elderly

or immunocompromised patient.  Unfortunately, this means 1/4 to 1/3 of all patients in the

emergency department (nearly 100 patients per day in my ED) received at least a superficial

consideration of sepsis, while 95% of these likely will not have it. 

What would you like to see in a sepsis diagnostic tool? What functions would be useful? Are

there any variables that need to be considered when creating such as tool such as time or

resources? 

1. specificity 

2. sensitivity 

3. rapid (has to have results in 15-30 minutes at longest). A point of care bedside device would

be ideal but not necessary 

4. likely blood based 

5. simplicity - need to be able to be run by a routine laboratory technician or nurse with minimal

training. 

6. ease of interpretation - generally a yes / no, or yes / maybe / no answer will be required to

ensure clinical uptake by clinicians. This is a sticking point, as I had mentioned previously,

sepsis is not a yes / no disease. However, a "go/no go" answer will be required for broad uptake

by clinicians downstream. 

Page 89: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

85

Appendix B – Pros and Cons of Different Testing

Methods

Page 90: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

86

Appendix C – Pairwise Analysis

Page 91: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

87

Appendix D – Lab Gantt Chart

Page 92: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

88

Appendix E – Micro-ESR Data and Analysis

0 g/dL

Trial 1

Trial 2

Average Results

Page 93: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

89

0.025 g/dL

Trial 1

Trial 2

Trial 3

Average Results

0.05 g/dL

Page 94: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

90

Trial 1

Trial 2

Trial 3

Average Results

0.10 g/dL

Trial 1

Page 95: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

91

Trial 2

Trial 3

Average Results

0.15 g/dL

Trial 1

Page 96: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

92

Trial 2

Trial 3

Average Results

0.20 g/dL

Trial 1

Page 97: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

93

Trial 2

Trial 3

Average Results

Page 98: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

94

Appendix F – Capillary Fill Data and Analysis

0.025 g/dL

Trial 1

Trial 2

Trial 3

Average Results

Page 99: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

95

0.05 g/dL

Trial 1

Trial 2

Trial 3

Average Results

Page 100: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

96

0.10 g/dL

Trial 1

Trial 2

Average Results

Page 101: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

97

0.15 g/dL

Trial 1

Trial 2

Trial 3

Average Results

Page 102: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

98

0.20 g/dL

Trial 1

Trial 2

Average Results

Page 103: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

99

Appendix G – Laser Syllectometry Data and Analysis

Page 104: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

100

Control Values

Page 105: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

101

Page 106: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

102

Page 107: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

103

0.025 g/dL Values

Page 108: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

104

Page 109: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

105

Page 110: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

106

0.05 g/dL Values

Page 111: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

107

Page 112: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

108

Page 113: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

109

0.10 g/dL Values

Page 114: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

110

0.15 g/dL Values

Page 115: Effects of Erythrocyte Aggregation on Blood Rheology in ...€¦ · Effects of Erythrocyte Aggregation on Blood Rheology in Regard to Future Sepsis Diagnosis Application Darren Burley

111

0.20 g/dL Values