Page 1: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Electron Energy-Loss Spectrometry

Walid Hetaba28.10.2016

Page 2: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Energy loss spectrum

Williams, Carter: Transmission Electron Microscopy, Springer 2009

Page 3: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Energy loss spectrum

Williams, Carter: Transmission Electron Microscopy, Springer 2009

Page 4: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Ti L2,3-edge O K-edge

NiO, WIEN2k TiO2, exp.

Hetaba et al., PRB 85 (2012), 205108

Energy loss near edge spectrum

ELNES

Page 5: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Schrödinger equation:

Formalism

Bra-Ket formalism:

Page 6: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Double differential scattering cross-section

Page 7: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Double differential scattering cross-section

Page 8: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Double differential scattering cross-section

Page 9: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Double differential scattering cross-section

Page 10: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Double differential scattering cross-section

Dynamic form factor (DFF)

Page 11: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Demtröder, Experimentalphysik 3, Springer 2010

Interferometric EELS (the mixed DFF)

Page 12: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Nelhiebel, PhD thesis, 1999Demtröder, Experimentalphysik 3, Springer 2010

Kohl & Rose, 1985

Interferometric EELS (the mixed DFF)

Page 13: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Nelhiebel, PhD thesis, 1999Demtröder, Experimentalphysik 3, Springer 2010

Kohl & Rose, 1985

Interferometric EELS (the mixed DFF)

Page 14: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Nelhiebel, PhD thesis, 1999Demtröder, Experimentalphysik 3, Springer 2010

Kohl & Rose, 1985

Interferometric EELS (the mixed DFF)

Page 15: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Page 16: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

The MDFF for crystals

Page 17: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

The MDFF for crystals

Page 18: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Experimental realisation

Page 19: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34 Ti 1.00 ± 0.000 55.66

Experimental spectra

Page 20: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.47 ± 0.067 32.01 Ti 1.00 ± 0.000 67.99

Experimental spectra

Page 21: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Experimental spectra

Differences in fine structure: different electronic structure

Page 22: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Page 23: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Payne et al., Rev. Mod. Phys. 64, 1045 (1992)

Calculate starting density from atomic densities:

Calculate potential from density

Solve Kohn-Sham equation:

Calculate density from gained wave functions:

New density = old density?

Input: Unit cell with atomic positions

no

yes

WIEN2k simulation

Page 24: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Payne et al., Rev. Mod. Phys. 64, 1045 (1992)

Calculate starting density from atomic densities:

Calculate potential from density

Solve Kohn-Sham equation:

Calculate density from gained wave functions:

New density = old density?

Input: Unit cell with atomic positions

no

yes

WIEN2k simulation

Page 25: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

TELNES – calculation of spectra

Page 26: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

θ

k1k0

q

Effects of anisotropy

Page 27: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

θ

k1k0

q

Effects of anisotropy

Page 28: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

θ

k1k0

q

Hébert-Souche et al., UM 83 (2000), 9-16

Effects of anisotropy

Page 29: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Orientation: c-axis parallel to electron-beam

Anisotropy – experiment 1

Page 30: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Orientation: c-axis perpendicular to electron-beam

Anisotropy – experiment 2

Page 31: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

• EMCD, 2006 • Polarisation vector momentum transfer vector

Energy-loss magnetic chiral dichroism

Page 32: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

kq‘

EE‘

q

• EMCD, 2006 • Polarisation vector momentum transfer vector

Energy-loss magnetic chiral dichroism

Page 33: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

• EMCD, 2006 • Polarisation vector momentum transfer vector

Energy-loss magnetic chiral dichroism

Page 34: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

• EMCD, 2006 • Polarisation vector momentum transfer vector

Energy-loss magnetic chiral dichroism

Page 35: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

EMCD – example 1

Page 36: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

EMCD – example 1

Page 37: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

EMCD – example 2

Page 38: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Channelling-Formalismus/Sims

ELCE, 1982

Energy losses by channelled electrons

Page 39: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Channelling-Formalismus/Sims

Krivanek et al., Ultram 9 (1982), 249

ELCE, 1982

Energy losses by channelled electrons

Page 40: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Incoming wave, scattering, outgoing wave Input: crystal, exp. paramters, scattering model Output: Bloch waves, intensity maps, thickness dependence

Löffler & Schattschneider, Ultram 110 (2010), 831 Hetaba, Diploma-Thesis, 2011

ELCE simulations

Page 41: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Incoming wave, scattering, outgoing wave Input: crystal, exp. paramters, scattering model Output: Bloch waves, intensity maps, thickness dependence

Löffler & Schattschneider, Ultram 110 (2010), 831 Hetaba, Diploma-Thesis, 2011

ELCE simulations

Page 42: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Incoming wave, scattering, outgoing wave Input: crystal, exp. paramters, scattering model Output: Bloch waves, intensity maps, thickness dependence

Löffler & Schattschneider, Ultram 110 (2010), 831 Hetaba, Diploma-Thesis, 2011

ELCE simulations

Page 43: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Incoming wave, scattering, outgoing wave Input: crystal, exp. paramters, scattering model Output: Bloch waves, intensity maps, thickness dependence

Löffler & Schattschneider, Ultram 110 (2010), 831 Hetaba, Diploma-Thesis, 2011

ELCE simulations

Page 44: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Incoming wave, scattering, outgoing wave Input: crystal, exp. paramters, scattering model Output: Bloch waves, intensity maps, thickness dependence

Löffler & Schattschneider, Ultram 110 (2010), 831 Hetaba, Diploma-Thesis, 2011

ELCE simulations

Page 45: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Ti

O

OAxis color code: x, y, z

Hetaba, Micron 63 (2014), 15

ELCE experiments & simulations

Page 46: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Ti

O

OAxis color code: x, y, z

Hetaba, Micron 63 (2014), 15

ELCE experiments & simulations

Page 47: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Ti

O

OAxis color code: x, y, z

Hetaba, Micron 63 (2014), 15

ELCE experiments & simulations

Page 48: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Hetaba, Micron 63 (2014), 15

ELCE experiments & simulations

Page 49: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

Hetaba, Micron 63 (2014), 15

ELCE experiments & simulations

Page 50: Electron Energy-Loss  · PDF fileElectron Energy-Loss Spectrometry ... Relative quantification: Elem. Atomic ratio (/Ti) Percent content N 0.80 ± 0.113 44.34

Max-Planck-Institut für Chemische Energiekonversion

• EELS comparable to XAS • Available at lower costs • Almost all TEMs have EELS-capability • Usually less energy-resolution but higher spatial resolution

• EELS is a versatile analytical methods • Information about:

• Elemental composition • Thicknes • Electronic structure • Optical properties • Magnetic properties

Summary