en 60601-1 “the third edition” 60601-1 “the third edition ... for 60601-1 expected mid-2012)...
TRANSCRIPT
Christine Ruther NSAI Medical Device Seminar, Galway
www.nsai.ie 2011‐12‐01 1
EN 60601-1 “The Third Edition”
Christine Ruther
© Copyright 2011
Family TTree:
2
Christine Ruther NSAI Medical Device Seminar, Galway
www.nsai.ie 2011‐12‐01 2
Transition Transition Information
3
3rd Edition Transition(aka: why I should care)
Product Status
US – FDA(AAMI ES 60601-1)
Health Canada EU - MDD OthersStatus (AAMI ES 60601 1) Canada
NewDevice
mandatory (1JUL13)
mandatory (1JUN12)
mandatory (1JUN12*) ...
Existing Device
(new sale)
okay as is okay as is mandatory (1JUN12*) ...
4
In the Field
(repairs)okay as is okay as is okay as is ...
* * -- see next slidesee next slide
Christine Ruther NSAI Medical Device Seminar, Galway
www.nsai.ie 2011‐12‐01 3
What’s the *
• In Europe, transition depends on particular standardsparticular standards
• Applicable particular standard?• Yes - date of transition of the
particular standard applies• 60601-2-29 radiotherapy simulators -
1NOV111NOV11• 60601-2-37 ultrasound - 1OCT10
• No - date in table applies5
Heads-Up
• All 60601 standards are being updatedAli i ith 3 d diti• Aligning with 3rd edition
• Normal technical revisions (amendment pending for 60601-1 expected mid-2012)
• Some standards are renumbered• EN ISO 80601-2-X
ll h• Still have most as EN 60601-2-X
• Standards are receiving transition dates
6
Christine Ruther NSAI Medical Device Seminar, Galway
www.nsai.ie 2011‐12‐01 4
Do I have to?
• Standards are voluntary but …MDD A I 2 “ l ti d t d b • MDD Annex I.2 “solutions adopted by the manufacturer for the design and construction of the devices must conform to safety principles, taking account of the generally acknowledged state of the art”state of the art
• MDD Article 5.1 gives presumption of conformity for harmonized standards
7
OverviewOverview
8
Christine Ruther NSAI Medical Device Seminar, Galway
www.nsai.ie 2011‐12‐01 5
Origin of Changes
• Experience with 2nd edition• Info in particular standards• County specific deviations
• US deviations strongly considered
• Other standards• EN 60950 heavily considered
9
So, what’s new?
• Collaterals are included when published• Some requirements have been amended• Some requirements have been amended• New requirements have been added• Focus today, a few points regarding:
• Labeling• Risk Management• Mechanical SafetyMechanical Safety• Electrical Safety• Collaterals
10
Christine Ruther NSAI Medical Device Seminar, Galway
www.nsai.ie 2011‐12‐01 6
LabelingLabeling
11
Marking
• On equipment:Name or trademark of MANUFACTURER• Name or trademark of MANUFACTURER
• Model or type reference• If single use, ‘do not reuse’ or symbol
• Applies to:• Equipment
Its detachable components and• Its detachable components, and• Accessories
12
Christine Ruther NSAI Medical Device Seminar, Galway
www.nsai.ie 2011‐12‐01 7
Indicator Lights
Color 2nd Edition 3rd Edition
Red: warning of danger and/or a need for urgent action
warning – immediate response by the operator is
required
Yellow: caution or attention required caution – prompt response by the operator is required
Green: ready for action ready for use
Oth meaning other than that of meaning other than that of Other: gred or yellow
gred, yellow or green
Symbols and Safety Signs
• New meaning for :• was “Attention: Consult accompanying • was Attention: Consult accompanying
documents“• is now “Caution”
• New “consult instructions for use” symbols:• when recommended
• when mandatory• Emergency stop symbol:
14
Christine Ruther NSAI Medical Device Seminar, Galway
www.nsai.ie 2011‐12‐01 8
Risk Risk Management
15
Risk Management File
• “Risk” appears 666 times• Compliance to ISO 14971 is required• Specific items must be addressed to
define test or test parameters
16
Christine Ruther NSAI Medical Device Seminar, Galway
www.nsai.ie 2011‐12‐01 9
Examples
• Clause 11.6.3 – Spill TestWas: 200 mL tap water ≤5cm 15 sec• Was: 200 mL, tap water, ≤5cm, ~15 sec
• Is: define fluid type, volume & duration for spill test through risk analysis
• Clause 11.1.1 – Pt. Applied Part Temp• Was: 41C• Is: 41C* if exposure <10% of body on
healthy adult
17
* can be 42C with justification & disclosure in IFU
Other Key Items
• Essential performance• Expected service life• Equivalent safety• Parts that contact the patient• Single fault condition• High-integrity components
18
Christine Ruther NSAI Medical Device Seminar, Galway
www.nsai.ie 2011‐12‐01 10
Review
• General ISO 14971 compliance• Details of RMF for 601-1 items• Expect review by
• Regulatory Authority• Notified Body• Test Lab
19
Mechanical Mechanical Safety
20
Christine Ruther NSAI Medical Device Seminar, Galway
www.nsai.ie 2011‐12‐01 11
Pinching/Trapping
• Gaps considered NOT to present a trapping zone hazardtrapping zone hazard
• Distances for adult & child (ISO 13851-1)
a
a
a
a
21
Body Head Leg Foot
a
50 max.
Toe
a
Arm
a
Hand
a
Finger
Safe Distances
• Sufficient distance between people and trapping zone hazard
22
Christine Ruther NSAI Medical Device Seminar, Galway
www.nsai.ie 2011‐12‐01 12
Mechanical Tests
23
Mechanical Tests
• Push: 250 N force over 30 mm dia. area for 5 seconds
• Impact: UL “ball impact test”
• Drop: 1 m (hand-held) / 2, 3, 5 cm dependent on mass
• Rough Handling: ascending (hit & stop), descending step (40 mm) and threshold descending step (40 mm) and threshold -“up & over” barrier
• Mould Stress: “bake” plastic enclosure24
Christine Ruther NSAI Medical Device Seminar, Galway
www.nsai.ie 2011‐12‐01 13
Continuous Activation
• Control system with momentary t t it hcontact switch
• Operator reaction and part momentum such that operator can release switch before injury
• Control must be single fault tolerant to stop motion (e.g. emergency stop stop motion (e.g. emergency stop switch), and prevent unwanted movement (e.g. enable switch).
25
Use of Emergency Stops
• Must:• Consider proximity & response capability of Consider proximity & response capability of
operator.• Be accessible• Can not be part of normal use• Can not introduce added hazards• Must completely open related electrical circuit• Require a single action• Be red & marked with E-Stop Symbol
26
Christine Ruther NSAI Medical Device Seminar, Galway
www.nsai.ie 2011‐12‐01 14
Mechanical Safety Factors
TENSILE SAFETY FACTOR
A B
Not impaired by wear Ductile 2.5 4
Not impaired by wear Brittle 4 6
I i d b D til 5 8Impaired by wear Ductile 5 8
Impaired by wear Brittle 8 12
A – Quantifiable load and material properties with high confidence
B – Other than A
TENSILE SAFETY FACTOR
A B
Impaired by wear w/MECHANICAL SAFETY
DEVICEDuctile 2.5 4
Impaired by wear w/MECHANICAL SAFETY
DEVICEBrittle 4 6
MECHANICAL SAFETYMECHANICAL SAFETY DEVICE 2.5 4
A – Quantifiable load and material properties with high confidence
B – Other than A
Christine Ruther NSAI Medical Device Seminar, Galway
www.nsai.ie 2011‐12‐01 15
Patient/Operator Support Systems
Surfaces that can be used as foot rests• A mass equivalent to twice the intended
person load is be applied to a 0.1 m2 area
Areas a PATIENT or OPERATOR can sit• A mass equivalent to the SAFE WORKING LOAD
representing the PATIENT or OPERATOR is representing the PATIENT or OPERATOR is dropped onto seating surfaces from 150mm
29
Oxygen Enriched Atmospheres
• A source of ignition exists if:The temperature exceeds the ignition point of the p g pmaterial (in oxygen).Temperatures could affect the solder such that loosening could occur.Components that could cause sparks (in failure) would crack or change their outer shape resulting in temperatures exceeding 300º C.Temperatures exceed 300º C in NC.Sparks can be generated in NC or SFC.
• If exists, then mitigate by containing or preventing fire
30
Christine Ruther NSAI Medical Device Seminar, Galway
www.nsai.ie 2011‐12‐01 16
Other Items to Consider
• Vibration (8 hour limit of 2.5m/s2)• Noise (continuous or impulse)• Thermal (time and material type)• Component heating (within rating)• Exposure to cold (include in RMF)
31
Electrical SafetyElectrical Safety
32
Christine Ruther NSAI Medical Device Seminar, Galway
www.nsai.ie 2011‐12‐01 17
MUST Specify• Classification – shock protection
I – relies on protective earth
II – no reliance on protective earth, double insulated
• Applied Part Type – shock & leakage current
B – basic, may be ground referenced (limited use)
BF – basic floating, higher protection than B
CF – cardiac floating, acceptable for direct cardiac contact
• Defibrillation Proof Rating (yes or no)
• Method of Sterilization33
MUST Specify (con’t)
• Ingress Protection (EN 60529) – IPXYX – protection from particles (uncommon to use)Y – protection from fluids (common but not req’d)Examples:IPX0 (normal),IPX1 (drip-proof),IPX7 (watertight)
NOTE: Adequate X rating may allow smaller spacings!
• Use in an oxygen rich environment(concentrations over 25 % at one atmosphere or partial pressures up to 27.5 kPa for higher atmospheric pressures)
• Anesthetic Use (generally no, if yes then Annex G) AP – flammable anesthetic mixture with airAPG – flammable anesthetic mixture with oxygen or nitrous oxide
34
Christine Ruther NSAI Medical Device Seminar, Galway
www.nsai.ie 2011‐12‐01 18
MUST Specify (new)
• High Altitude Use (yes or no)
• Over 2,000 m
• See requirements in clause 8.9.1.5
• Pollution degree (clause 8.9.1.8)
• Overvoltage category (clause g g y (8.9.1.11)
(mains transient peak voltages)35
Pollution Environments
(also all apply to micro-environments)
• Degree 1: sealed to exclude dust and moisture
• Degree 2: generally only non-conductive pollution
Degree 3: d ti ll ti• Degree 3: some conductive pollution
• Degree 4: continuously conductive pollution
36
Christine Ruther NSAI Medical Device Seminar, Galway
www.nsai.ie 2011‐12‐01 19
Overvoltage Category
(refers to max. mains voltage transient equipment is design to withstand)q p g )
37
Sample Product• Device defined as
• Class I (relies on protective earth)• Type BF (applied part)• Type BF (applied part)• Not defib proof• Not AP nor APG• IPX0• Non-sterile• High altitude (8,000 ft = 2.44 km)• Pollution degree 2Pollution degree 2• Overvoltage category II
• Example would be a stimulator suitable for use in aircraft
38
Christine Ruther NSAI Medical Device Seminar, Galway
www.nsai.ie 2011‐12‐01 20
Block Diagram
39
Patient Applied Parts
40Parts Need to Treat asPatient Applied Parts
Christine Ruther NSAI Medical Device Seminar, Galway
www.nsai.ie 2011‐12‐01 21
Dielectric Withstand = Means of Protection• Rather than A-a1 …, B-a …
• MOOP (means of operator protection)• MOPP (means of patient protection)
• 2 MOOP/MOPP = double (or reinforced) insulation
• 1 MOOP/MOPP = basic insulation1 MOOP/MOPP basic insulation• Supplementary insulation in 2nd ed. still
in 3rd ed. (but meaning is simpler)41
Creepage / Clearance
(same meaning but new values)
42
Christine Ruther NSAI Medical Device Seminar, Galway
www.nsai.ie 2011‐12‐01 22
ComparisonID description
Vtest cr (mm) cl (mm)
2nd Ed. 3rd Ed. 2nd Ed. 3rd Ed. 2nd Ed. 3rd Ed.
A Line to neutral 1,500 1,500 3 3 1.6 1.9
B Mains to accessible parts that are 4 000 3 000 8 4 8 5 4 6B pnot protectively earthed 4,000 3,000 8 4.8 5 4.6
C Mains to accessible parts that are protectively earthed 1,500 1,500 4 2.4 2.5 2.3
D Mains input to low voltage 4,000 4,000 8 7.9 5 5
E Low voltage to accessible parts that are not protectively earthed 500 n/a 4 2.4 2 2.3
F Low voltage to accessible parts that are protectively earthed 500 n/a 2 1.2 1 1.2that are protectively earthed
G Low voltage to treatment voltage 3,000 2,000 4.6 5 2.4 3.2
HTreatment voltage to all accessible parts (whether or not protectively earthed)
1,500 1,500 4 4 2.5 2.5
500 1,000 4.6 4.4 2.4 2.4
43
CollateralsCollaterals
44
Christine Ruther NSAI Medical Device Seminar, Galway
www.nsai.ie 2011‐12‐01 23
Incorporated
• EN 60601-1-1• Medical electrical systems• Now clause 16
• EN 60601-1-4• Programmable medical electrical systems• Now clause 14• Now clause 14
45
Still Standing Alone
• EN 60601-1-2* EMCd• EN 60601-1-3 Radiation Protection
• EN 60601-1-6 Usability• EN 60601-1-8* Alarms• EN 60601-1-9 Env Conscious Design
EN 60601 1 10 Phys Closed Loop• EN 60601-1-10 Phys Closed Loop• EN 60601-1-11 Home Use
46* revisions pending
Christine Ruther NSAI Medical Device Seminar, Galway
www.nsai.ie 2011‐12‐01 24
Coming Soon
• EN 60601-1-12 Emergency Medical Environment
47
BonusBonus
48
Christine Ruther NSAI Medical Device Seminar, Galway
www.nsai.ie 2011‐12‐01 25
RoHS – MDD & IVDD
• 2011/65/EU –– Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment Directive
• Transition:• MDD Devices: 22 July 2014• IVDD Devices: 22 July 2016• IVDD Devices: 22 July 2016• Parts for repair may continue past this date
• AIMDD devices remain exempt49
RoHS
• Concerns are for:• Lead
M• Mercury• Cadmium• Hexavalent chromium• Polybrominated biphenyls (PBB)• Polybrominated diphenyl ethers (PBDE)
Exemptions
50
• Exemptions• 32 current exemptions• Reviewed every 5 years
Christine Ruther NSAI Medical Device Seminar, Galway
www.nsai.ie 2011‐12‐01 26
Significant Points
• Requires CE marking (and declaration)
b l• Labeling (mostly covered by MDD)
• Record retention = 10 years• Technical documentation, register of non-
conforming product and product recalls• List of any economic operator who has supplied
you with electrical or electronic equipment andyou with electrical or electronic equipment, and• List of any economic operator to whom you have
supplied any electrical or electronic equipment.
51
Showing Showing Compliance to the 3rd Edition
52
Christine Ruther NSAI Medical Device Seminar, Galway
www.nsai.ie 2011‐12‐01 27
Basic Options
• Complete re-evaluation• New test report• Ensures complete coverage
• Gap analysis• Addendum to existing report• Must be careful to:• Must be careful to:
• Catch everything• Address changes
53
Guidance
• NB-MED guidance is pending• Draft guidance includes:
• Presumption of conformity needs to be reconsideration when standard changes
• Yes, re-testing (re-evaluation) is required• Power supply must have 3rd edition eval. Power supply must have 3 edition eval.
(re-testing or gap analysis + testing)
• Standard was published in 2006 …
54