Innovative solutions in resuscitation therapyMultifunction Electrode Pads
452296291631.indd 1 6/17/13 10:44 AM
Multifunction electrode pads2
Enhance resuscitation therapy and workflowWith pads for emergency use, monitoring, and special procedures, there are HeartStart multifunction electrode pads for every need – sized and designed for a wide variety of environments, caregivers, treatments, and patients alike.
452296291631.indd 2 6/17/13 10:44 AM
Multifunction electrode pads 3
Rugged and dependableConstructed of a thin flexible conductor layered between a protective polymer backing and hydro-gel, Philips pads withstand the rigors of CPR while enhancing pre-hospital to hospital workflow.
Enhance resuscitation therapy and workflowWhen connected to a manual defibrillator with advanced capabilities, HeartStart multifunction pads can provide ECG monitoring, external pacing, synchronized cardioversion, and defibrillation. The pads can facilitate ventilation detection when used with the Q-CPR measurement and feedback option on the Philips HeartStart MRx monitor/defibrillator. They are easy to use, compliant with internationally-recognized standards, and conform to the body’s contours for proper current distribution. When combined, these features may enhance resuscitation therapy.
All pads are optimized for Philips defibrillators to decrease the need to change pads if you have to switch defibrillators. Most pads are compatible with defibrillators from other manufacturers using Philips HeartStart pads adapters. These adapters provide simple inter-connectivity and allow for easy standardization of pads, whether defibrillators are biphasic or monophasic, manual or automated. The end result of using Philips pads and adapters can be a more seamless workflow.
452296291631.indd 3 6/17/13 10:44 AM
Multifunction electrode pads4
Adult/Child pads• For patients ≥ 22 pounds (10kg) using manual defibrillators or 55 pounds (25kg) using AEDs
or AED mode• Conductor surface area 75cm2 and gel surface area 102cm2
• Multifunction – defibrillation, pacing, monitoring, and cardioversion• Hydro-gel thickness and adhesion suited for emergency situations• 24 month shelf life
HeartStart pads for emergency and general use
Adult/Child with barrel connector (M3501A)• General purpose• Designed for manual defibrillators• 24” (60cm) lead wire. Requires hands-free cable when
connected to a manual defibrillator.• Packed 10/case
Adult/Child Defibrillator Pads (989803158211/989803158221)SMART Pads II (989803139261) SMART Pads III (989803149981/989803149991)• General purpose • Designed for AED use• Suitable for both BLS- and ACLS-trained responders• 48" (120cm) lead wire. Requires hands-free cable when
connected to a manual defibrillator.• Wide storage temperature range• Packed 1 or 5/case
Adult/Child Plus (M3713A)Adult/Child Preconnect (989803166021)• General purpose • Designed primarily for manual defibrillator use, and compatible
with AEDs• Package and instructions optimized for ACLS-trained responders• 48” (120cm) lead wire. Requires hands-free cable when connected
to a manual defibrillator.• Packed 10/case
452296291631.indd 4 6/17/13 10:44 AM
Multifunction electrode pads 5
Infant pads• For patients < 22 pounds (10kg)• Specially formulated gel hydrates skin, yet provides adhesion suitable for emergency situations• Pads sized for infant patients – conductor surface area 32cm2 and gel surface area 44cm2
• Multifunction – defibrillation, pacing, monitoring, and cardioversion• Recommended anterior-posterior placement depicted on pad’s outer surface• Long shelf life
Infant Plus (M3717A)• General purpose• Designed only for use with manual
defibrillators (in manual mode); can use with HeartStart XL+ in AED mode if Infant/Child category is selected
• Not to be used with AEDs • Package and instructions optimized for
ACLS-trained responders• 24" (60cm) lead wire. Requires hands-
free cable.• Packed 5/case
Infant with barrel connector (M3504A)• General purpose• 24" (60cm) lead wire. Requires hands-
free cable.• No adapter required• Not to be used with AEDs
FR2 Infant/Child (M3870A)• Designed only for use with FR2 and
FR2+ AEDs • Unique bear-shaped connector allows
users to distinguish the pads• 48" (120cm) lead wire• Pad circuitry attenuates delivered shock to
a therapy appropriate for patients under 8 years of age or 55 pounds (25kg)
• Suitable for both BLS- and ACLS-trained responders
• Packed 1/case
452296291631.indd 5 6/17/13 10:44 AM
Multifunction electrode pads6
Specialty pads• Optimized uses – radiological procedures in the catheterization lab, electrophysiology lab, critical
care unit, emergency department, telemetry, and step-down cardiac unit• Multifunction – defi brillation, pacing, monitoring, and cardioversion• Package and instructions optimized for ACLS-trained caregivers
Adult/Child Radiotransparent (M3718A)Reduced Skin Irritation• Ideal for radiological procedures requiring
increased visibility• Completely transparent pad under typical
X-ray levels• Thick gel layer provides enhanced
patient contact• Hydro-gel with high water content allows
for easy pad removal, excellent skin hydration, and reduced skin irritation
• Hydrophilic properties allow electrode to perform under diaphoretic conditions
• Adhesive properties prevent pad migration• Lessens skin irritation by neutralizing
changes in pH conditions produced during pacing
• For use on patients ≥ 22 pounds (10kg) when used with a manual defi brillator
• Conductor surface area and gel surface area are 97cm2
• 48" (120cm) lead wire requires hands-free cable when connected to a manual defi brillator
• Packed 10/case
Thermogram comparison*
HeartStart Radiotransparent PadsProvide a more uniform distribution of current under the electrode during use
Other Manufacturer’s PadsExhibit concentrated energy or “hot spots” that could cause skin irritation
HeartStart pads for special procedures and environments
* Thermograms were taken 3.6 seconds after
defi brillation using a bi phasic defi brillator
452296291631.indd 6 6/17/13 10:44 AM
Multifunction electrode pads 7
Infant Radiotransparent (M3719A)Reduced Skin Irritation• For use on smaller patients, weighing < 22
pounds (10kg)• Conductor surface area and gel surface
area are 50cm2
• 24" (60cm) lead wire• Not to be used with AEDs; can use with
HeartStart XL+ in AED mode if Infant/Child category selected
• Designed primarily for manual defi brillators• Packed 5/case
HeartStart Adult/Child Radiolucent Pads• Cast slight shadow under
typical adult X-ray levels
HeartStart Adult/Child Radiotransparent Pads• Completely transparent under
typical adult X-ray levels
Adult/Child Radiolucent (M3716A)• Suitable for patients who will likely require
chest X-rays• Pad casts slight shadow under typical adult
X-ray levels• Moist, long-lasting hydro-gel with 50%
more water content than HeartStart Adult/Child Plus pads, yet provides adhesion suitable for emergency situations
• Package and instructions optimized for ACLS-trained responders
• 48” (120cm) lead wire requires hands-free cable when connected to a manual defi brillator
• Indicated for patients ≥ 22 pounds (10kg) when used with a manual defi brillator
• Packed 10/case
Radiolucent vs. Radiotransparent comparison
452296291631.indd 7 6/17/13 10:44 AM
Multifunction electrode pads8
AdaptersTo facilitate connection of Philips HeartStart pads to other defi brillators in your facility, we offer HeartStart adapters. A built-intether secures the adapter to the defi brillator’s hands-free cable, providingits presence when needed.
Technical specifi cations
Clear, concise labeling on the package and pads guides experienced and infrequent caregivers alike through pad selection, placement, and connection.
989803166021 M3713A M3716A M3717A M3718A
Description/name Adult/Child Preconnect Pads
Adult/Child Plus Pads
Adult/Child Radiolucent Pads
Infant Plus Pads Adult/Child Radio-transparent
PadsRadio-transparent
Pads per case 10 10 10 5 10
Indications External pacing, defi brillation, monitoring, and cardioversionPatientInfant <10kgChild ≥10kg – <25 kgAdult ≥25kg
ChildAdult
ChildAdult
ChildAdult
InfantChildAdult
Conductor surface area (cm2) 75 75 75 32 97Approximate lead wire length (cm)
120* 120 120 60 120
No. of hours on skin 24 24 24 24 24Number of defi brillation shocks 50 50 50 50 50Duration of pacing (hours) at 100mA, 80 bpm
8 8 8 8 3
Temperature range for storage 15° – 35°C 15° – 35°C 15° – 35°C 15° – 35°C 15° – 35°CShelf life (in months) from date of manufacture (may be less upon receipt)
24 24 24 24 24
Philips adapter compatibility 05-1000005-1010005-10200
05-1000005-1010005-10200
05-1000005-1010005-10200
05-1000005-1010005-10200
05-1000005-1010005-10200
* (approx. 30 outside package)
Adapters are available for connecting HeartStart pads, (except those specifi ed NA above) to the devices listed to the right.
Model 05-10200Adapter fi ts Philips hands-free cables M3507A or M1750A/B for connection toHewlett-Packard CodeMaster 100, XE, XL, and XL+; Agilent Heartstream/PhilipsHeartStart XL and XLT; Philips HeartStart MRx and XL+; and Laerdal HeartStart 4000 defi brillator/monitors
452296291631.indd 8 6/17/13 10:44 AM
Multifunction electrode pads 9
Model 05-10000Removable adapter for Medtronic Physio-ControlLifePak 9, 10C, 11, 12, 20 and 500 defi brillators
M3719A M3501A M3504A 989803158211 (1 set)989803158221 (5 sets)
989803139261 989803149981 (1 set)989803149991 (5 sets)
M3870A
Radio-transparent Infant
Radio-transparent Pads
Adult/Child Pads with
barrel connector
Infant Pads with barrel connector
Defi brillator Pads SMART Pads II SMART Pads III FR2 Infant/Child Pads
5 10 5 1 or 5 1 1 or 5 1External pacing, defi brillation, monitoring, and cardioversion Defi brillation
InfantChildAdult
InfantChildAdult
InfantChildAdult
InfantChildAdult
InfantChild
50 75 32 85 80 80 8560 60 60 120 120 120 120
24 24 24 24 24 24 2450 50 50 50 50 50 501.5 8 8 8 8 8 N/A
15° – 35°C 15° – 35°C 15° – 35°C 0° – 50°C 0° – 50°C 0° – 50°C 0° – 32°C24 24 24 30 30 30 30
05-1000005-1010005-10200
N/A N/A 05-1000005-1010005-10200
05-1000005-1010005-10200
N/A N/A
Model 05-10100Removable adapter for Zoll 1200, 1400, 1600 and M-Series defi brillators
Pace, monitor, cardiovert, and defi brillate with a single pad family, as HeartStart adapters work with many popular defi brillators. When transferring a patient from one defi brillator to another, simply
plug the pad connector into the HeartStart adapter, then plug the adapter into your defi brillator’s hands-free cable. (Note: The standard plug-style connector fi ts Philips HeartStart hands-free cable, model M3508A; no adapter necessary.)
defi brillators
452296291631.indd 9 6/17/13 10:44 AM
Multifunction electrode pads10
Defi brillator compatibility
989803166021 M3713A M3716A M3717A M3718A
Description/name Adult/Child Preconnect Pads
Adult/Child Plus Pads
Adult/Child Radiolucent Pads
Infant Plus Pads Adult/Child Radio-transparent
PadsRadio-transparent
Advanced Life Support Defi brillators (ALS)XL+861290
M3508A Therapy Cable, plug connector
● ● ● ● ●
M3507A Therapy Cable, barrel connector
MRx/XL/XLTM3535A, M3536A M4735A, M3500B
M3508A Therapy Cable, plug connector
●2 ●2 ●2 ●1 ●2
M3507A Therapy Cable, barrel connector
Automated External Defi brillators (AEDs)FR35
861388, 861389●3,4 ●3,4 ●3,4
FRx861304, 861305
●3,4 ●3,4 ●3,4
FR2/FR2+M3840A, M3841A, M3860A, M3861A, 861458, 861459
● ● ● ●
1 Not in AED mode, as the energy automatically delivered is 150 joules and therefore too high for an infant (<10kg)
2 In AED mode for adults (≥25kg) only, as the energy automatically delivered is 150 joules and therefore too high for a
child (<25kg) or infant (<10kg)
3 Pads can be used on adults and children ≥ 25kg or 8 years old.
4 Pads will operate with device for therapy, but cannot be left attached while device is in standby mode or while turned
off due to self-test activity of the device
5 Not available in all geographic areas. Please check availability with your Sales and Marketing Center contact.
452296291631.indd 10 6/17/13 10:44 AM
Multifunction electrode pads 11
M3719A M3501A M3504A 989803158211 (1 set)989803158221 (5 sets)
989803139261 989803149981 (1 set)989803149991 (5 sets)
M3870A
Radio-transparent Infant
Radio-transparent Pads
Adult/ChildPads with
barrel connector
Infant Pads with barrel connector
Defi brillator Pads SMART Pads II SMART Pads III FR2 Infant/Child Pads
● ● ● ●
● ●
●1 ●2 ●2 ●2
●2 ●1
●3 ● ●●3 ● ●3
● ● ● ●
452296291631.indd 11 6/17/13 10:44 AM
© 2013 Koninklijke Philips Electronics N.V.All rights are reserved.
Philips Healthcare reserves the right to make changes in specifications and/or to discontinue any product at any time without notice or obligation and will not be liable for any consequences resulting from the use of this publication.
Printed in The Netherlands4522 962 91631 * JUN 2013
Please visit www.philips.com/HeartStart/Supplies
To order HeartStart pads and other resuscitation supplies and accessories: http://www.philips.com/resuscitationsupplies
Philips Healthcare is part of Royal Philips Electronics
How to reach uswww.philips.com/[email protected]
Q-CPR is a trademark of Laerdal Medical Corporation
452296291631.indd 12 6/17/13 10:44 AM