equipment standards for cardiology

18
Equipment Standards for Interventional Cardiology Author: A. Dowling, A. Gallagher, C. Walsh, T. Kenny, J. Malone Lead partner : Houghton Institute, St. James’s Hospital, Dublin, Ireland Introduction: Interventional radiology (IR) has seen rapid growth in the field of cardiology over the past decade and often represents an alternative to more hazardous surgery [1-6]. Interventional cardiac procedures are currently the most common group of interventional procedures performed in Europe [1]. Such procedures are complex and may involve prolonged irradiations, which may subject patients and operators to higher levels of risk than those, which normally prevail [1-5]. This has prompted many international bodies to issue special advice and initiate research in this area [3]. Currently interventional cardiology contributes over 10% to annual collective dose in the UK in spite of contributing to a total annual examination frequency of 0.68% [7]. The EU Medical Exposures Directive 97/43/Euratom and consequent national legislation identifies interventional radiology as an area of special concern [8,9]. Advances in imaging technology have facilitated the development of increasingly complex radiological equipment for interventional cardiology [2,4]. The development of the technology is industry and end user driven and is developing at a rate that is ahead of supporting research, equipment standards and the regulatory framework [1]. Consequently, there is a need for definitive equipment requirements and standardisation in the design, manufacture, acceptance and maintenance of this 1

Upload: simon23

Post on 03-Jun-2015

488 views

Category:

Documents


3 download

TRANSCRIPT

Page 1: Equipment standards for cardiology

Equipment Standards for Interventional Cardiology

Author: A. Dowling, A. Gallagher, C. Walsh, T. Kenny, J. Malone

Lead partner: Houghton Institute, St. James’s Hospital, Dublin, Ireland

Introduction:

Interventional radiology (IR) has seen rapid growth in the field of cardiology over the past decade and often represents an alternative to more hazardous surgery [1-6]. Interventional cardiac procedures are currently the most common group of interventional procedures performed in Europe [1]. Such procedures are complex and may involve prolonged irradiations, which may subject patients and operators to higher levels of risk than those, which normally prevail [1-5]. This has prompted many international bodies to issue special advice and initiate research in this area [3]. Currently interventional cardiology contributes over 10% to annual collective dose in the UK in spite of contributing to a total annual examination frequency of 0.68% [7]. The EU Medical Exposures Directive 97/43/Euratom and consequent national legislation identifies interventional radiology as an area of special concern [8,9].

Advances in imaging technology have facilitated the development of increasingly complex radiological equipment for interventional cardiology [2,4]. The development of the technology is industry and end user driven and is developing at a rate that is ahead of supporting research, equipment standards and the regulatory framework [1]. Consequently, there is a need for definitive equipment requirements and standardisation in the design, manufacture, acceptance and maintenance of this equipment. Part of the work of the EU DIMOND III research programme is to develop material for standards in this area [1].

Methods:

It is a requirement of the Medical Exposures Directive, that acceptance testing be carried out on radiological equipment before its first use for clinical purposes and performance testing performed on a regular basis thereafter [8]. A Commissioning / Quality Assurance survey of 12 interventional cardiac systems was carried out by the Department of Medical Physics and Bioengineering, St. James’s Hospital, Dublin. The survey aims to assess the requirements for equipment standards to address the imbalance between the advancing technology and existing standards. The systems are listed in Table 1 and consisted of 15 Image Intensifier-TV chains in total (3 Bi-plane systems). Published results for the GE Innova 2000 flat panel system are included for comparison purposes [10].

Six of the imaging chains were new, 6 were approx. 5 years old and the remaining 3 were greater than 10 years old. All systems had a nominal detector diameter of 23cm with the exception of the GE Advantx (36cm) and GE Innova (20cm x 20cm). Testing was performed in

1

Page 2: Equipment standards for cardiology

line with current international and national guidelines and standards [11-29]. Testing included assessing the performance of the X-ray tube and generator, measuring dose levels in fluoroscopy and digital acquisition modes and a subjective assessment of image quality using the Leeds test objects [11-29]. Dose levels were measured under Automatic Exposure Control (AEC) in fluoroscopy and digital acquisition modes by measuring the detector and patient entrance dose rates / dose per exposure, using published methodology [11-29]. Electrical, mechanical and general radiation safety were also assessed.

No. Manufacturer System1 Philips Integris Allura (Bi-plane)1 Philips Integris2 Philips DCI4 Siemens Coroscop1 Siemens Axiom Artis2 Siemens Bicor (Bi-plane)1 GE Advantx1 GE Innova 2000

Table 1. Systems included in Commissioning / QA Survey

Results:

Dose Levels:The majority of the systems tested had a wide range of user selectable dose options available including a range of pulsed fluoroscopy modes, digital acquisition frame rates, AEC curves and spectral filtration settings. The options available in the fluoroscopy mode on one typical cardiac system are presented in figure 1. This system has two fluoroscopy modes, ‘low/normal’ and ‘high’, three field sizes and the fluoroscopy pulse rate is selectable from 3.12 to 25 pulses per second. The Image Intensifier/detector entrance dose rate is thus user selectable from 0.06 to 2.19 Gy/sec on a single spectral filtration setting. The pulsed fluoroscopy mode most frequently used on the equipment tested was 12.5 or 15 pulses per second (pps).

Figure 1: Fluoroscopy Detector Entrance Dose Rates on a typical cardiac system

2

Im age Intensifie r Entrance Dose Rates

0

0.5

1

1.5

2

2.5

1 2 3 4 5 6 7 8

Pulses/sec

uG

y/s

ec Full Field

Mag 1

Mag 2Low

High

3.12 6.23 12.5 25 3.12 6.25 12.5 25

0.06 – 2.19 Gy/sec

Im age Intensifie r Entrance Dose Rates

0

0.5

1

1.5

2

2.5

1 2 3 4 5 6 7 8

Pulses/sec

uG

y/s

ec Full Field

Mag 1

Mag 2

Full Field

Mag 1

Mag 2Low

High

3.12 6.23 12.5 25 3.12 6.25 12.5 25

0.06 – 2.19 Gy/sec

3.12 6.25 12.5 25 3.12 6.25 12.5 25

0.06 – 2.19

Page 3: Equipment standards for cardiology

The measured detector entrance dose rates for the systems tested, on the 12.5/15pps setting, are presented in figure 2. The systems are grouped according to manufacturer. From figure 2 it is evident that the detector entrance dose rates ranged from 0.2 – 0.7 Gy/sec across the systems. It is also evident that the detector entrance dose rates are manufacturer dependent. The results from one equipment manufacturer were consistently found to be approximately double, or more, that of another manufacturer for both the old and new systems. In figure 2, systems 4 and 5 have no pulsed fluoroscopy option and system 11 was configured incorrectly at commissioning in that the 12.5pps option was incorrectly operating at 25pps.

Figure 2: Fluoroscopy Detector Entrance Dose Rates – 12.5/15pps

Detector entrance dose rates in the continuous fluoroscopy mode (25/30pps) (normal and high fluoroscopy modes) are presented in figure 3. The detector entrance dose rates range from 0.3 – 1.3 Gy/sec. It is also evident in the continuous fluoroscopy mode that the detector entrance dose rates for one manufacturer are consistently notably higher than those for another manufacturer and that the dose rates on one manufacturers ‘normal’ setting are equivalent to, and in some cases higher than, the dose rates on another manufacturers ‘high’ setting.

Figure 3: Fluoroscopy Detector Entrance Dose Rates – 25/30pps

The majority of systems operated with a detector entrance dose rate of 0.4 – 0.6 Gy/sec, in the continuous fluoroscopy mode (25/30 pps) as presented in figure 4.

3

0

0.2

0.4

0.6

0.8

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

Gy/s

ec

* * * *

X

System

Detector Entrance Dose Rates - Grouped By Manufacturer

* : Continuous Fluoro only

0

0.2

0.4

0.6

0.8

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

Gy/s

ec

* * * *

X

System

Detector Entrance Dose Rates - Grouped By Manufacturer

0

0.2

0.4

0.6

0.8

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

Gy/s

ec

* * * *

X

System

Detector Entrance Dose Rates - Grouped By Manufacturer

* : Continuous Fluoro only

00.2

0.4

0.6

0.8

1

1.2

1.4

1 3 5 7 9 11

13

15

ContHigh

Gy/s

ec

System

Detector Entrance Dose Rates Grouped by Manufacturer

00.2

0.4

0.6

0.8

1

1.2

1.4

1 3 5 7 9 11

13

15

ContHighContHigh

Gy/s

ec

System

Detector Entrance Dose Rates Grouped by Manufacturer

0

2

4

6

8

10

12

1 2 3 4 5 6 7

No of Systems

** * * *

Page 4: Equipment standards for cardiology

Figure 4: Fluoroscopy Detector Entrance Dose Rates – 25/30pps

Patient entrance dose rates, measured using the methodology published by Martin et al [17], are presented in figures 5, 6 and 7. Current guidelines are presented beneath the figures.

Figure 5: Fluoroscopy Patient Entrance Dose Rates – 12.5 / 15 pps

Figure 6: Fluoroscopy Patient Entrance Dose Rates – 25 / 30 pps

4

0

5

10

15

20

25

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

Martin: Low < 15 mGy/min; Normal 15-30 mGy/min; High 31 – 45 mGy/min IPEM: 50 mGy/min; IAEA: 25 mGy/min, WHO: 17 mGy/min

* * * *

X

mG

y/m

in

System

Patient Entrance Dose Rates Grouped by Manufacturer

0

5

10

15

20

25

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

Martin: Low < 15 mGy/min; Normal 15-30 mGy/min; High 31 – 45 mGy/min IPEM: 50 mGy/min; IAEA: 25 mGy/min, WHO: 17 mGy/min

* * * *

X

mG

y/m

in

System

Patient Entrance Dose Rates Grouped by Manufacturer

0

1020

3040

5060

70

1 3 5 7 9 11

13

15

Con t

High

Martin: Low < 15 mGy/min; Normal 15-30 mGy/min; High 31 – 45 mGy/min IPEM: 50 mGy/min; High > 1.5 x; IAEA: 25 mGy/min; WHO: 17 mGy/min; RSNA High: 20-43

Systems

Patient Entrance Dose Rates - Grouped by Manufacturer

mG

y/m

in

0

1020

3040

5060

70

1 3 5 7 9 11

13

15

Con t

High

Martin: Low < 15 mGy/min; Normal 15-30 mGy/min; High 31 – 45 mGy/min IPEM: 50 mGy/min; High > 1.5 x; IAEA: 25 mGy/min; WHO: 17 mGy/min; RSNA High: 20-43

Systems

Patient Entrance Dose Rates - Grouped by Manufacturer

mG

y/m

in

0.0-0.2 0.2-0.4 0.4-0.6 0.6-0.8 0.8-1.0 1.0-1.2 1.2-1.4 Gy/sec

Page 5: Equipment standards for cardiology

The patient entrance dose rates in the pulsed fluoroscopy mode (12.5/15pps) ranged from 2.5 – 21.5 mGy/min and in the continuous mode (25/30pps) ranged from 5 – 42 mGy/min. On the high fluoroscopy setting a maximum patient entrance dose rate in excess of 60 mGy/min was measured for an old system. As with the detector entrance dose rates, patient entrance dose rates were found to be manufacturer dependent. Measured patient entrance dose rates for one manufacturer were found to be consistently double, or more, those measured for another manufacturer, for both old and new systems. As with the detector entrance dose rates, patient entrance dose rates on one manufacturers ‘normal’ setting were equivalent to or higher than another manufacturers ‘high’ setting.

For the majority of systems tested, the patient entrance dose rate was 10 – 20 mGy/min. A breakdown is presented in figure 7.

Figure 7: Fluoroscopy Patient Entrance Dose Rates – 25/30 pps.

Patient entrance dose rates were also measured using the IEC protocol [24]. Using the IEC protocol, the measured entrance dose rate is typically higher that that measured using the Martin BJR protocol [17] as the phantom is positioned closer to the X-ray tube. A comparison of results from one cardiac system, using the two protocols [17,24] is presented in figure 8.

Figure 8: Fluoroscopy Patient Entrance Dose Rates – comparison of IEC [24]and Martin, BJR [17] protocols.

Detector entrance dose per exposure measurements in the digital acquisition mode are presented in figure 9.

5

0

0 .05

0 .1

0 .15

0 .2

0 .25

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

Gy/

exp

osu

re

System

Detector Entrance Dose / Exposure Grouped by Manufacturer

0

0 .05

0 .1

0 .15

0 .2

0 .25

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

Gy/

exp

osu

re

System

Detector Entrance Dose / Exposure Grouped by Manufacturer

0

2

4

6

8

1 2 3 4 5

0

20

40

60

80

100

120

140

160

1 2 3 4 5 6

IEC

Martin, BJR

0 – 10 10 – 20 mGy/min 20 – 30 30 – 40 40 – 50

No of Systems

Low FF High FF Low M1 High M1 Low M2 High M2

Page 6: Equipment standards for cardiology

Figure 9: Digital Acquisition Mode – Detector Entrance Dose / Exposure

The detector entrance dose per exposure ranged from 0.06 – 0.2 Gy/exposure on the systems tested. As in fluoroscopy mode, the detector entrance dose per exposure in the digital acquisition mode was found to be consistently higher for one manufacturer over another, for old and new systems. For the majority of systems, the detector entrance dose per exposure was in the range 0.1 – 0.15 Gy/exposure, figure 10.

Figure 10: Digital Acquisition Mode – Detector Entrance Dose / exposure.

Patient entrance dose per exposure measurements in the digital acquisition mode are presented in figure 11. This ranged from 0.03 – 0.12 mGy/exposure, with the majority of systems operating in the range 0.1 – 0.2 mGy/exposure, figure 12.

6

0

1

2

3

4

5

6

7

1 2 3 40.05 - 0.1 0.1 – 0.15 Gy/exp 0.15 – 0.2 0.2 – 0.25

No. of Systems

Page 7: Equipment standards for cardiology

Figure 11: Digital Acquisition Mode, Patient Entrance Dose per exposure.

Figure 12: Digital Acquisition Mode, Patient Entrance Dose per exposure.

Image Quality:

Image Quality was assessed using the Leeds test objects and published protocols [12-16]. Threshold Contrast results are presented in figure 13. As with figures 2-12, these results are grouped by manufacturer. All of the systems had a measured threshold contrast of less than 4%, but 66% of new systems had a threshold contrast of greater than 2.7% which is the maximum threshold contrast expected for new systems of this type [12-16]. In general, an improvement in image quality was not apparent for the systems operating at higher dose levels.

7

00 .020 .040 .060 .08

0 .10 .120 .14

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

mG

y/ex

po

sure

System

Patient Entrance Dose / Exposure Grouped by Manufacturer

00 .020 .040 .060 .08

0 .10 .120 .14

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

mG

y/ex

po

sure

System

Patient Entrance Dose / Exposure Grouped by Manufacturer

0

0.5

1

1.5

2

2.5

3

3.5

4

4.5

1 2 3 40.02 – 0.04 0.04 – 0.08 mGy/exp 0.08 – 0.1 0.1 – 0.12

No. of Systems

Page 8: Equipment standards for cardiology

Figure 13: Fluoroscopy Mode – Threshold Contrast

Limiting spatial resolution results are presented in figure 14. Fifteen out of the 16 systems has a measured spatial resolution of greater than 1.25 lp/mm which is expected for systems if this type [12-16]. One old system had a resolution lower than 1.25 lp/mm.

Figure 14: Fluoroscopy Mode – Limiting Spatial Resolution

8

0

1

2

3

4

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

6 (40%) < 2.7%

% T

hre

sh

old

Co

ntr

as

tFluoroscopy - Low Contrast / Noise N3

Systems

0

1

2

3

4

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

6 (40%) < 2.7%

% T

hre

sh

old

Co

ntr

as

tFluoroscopy - Low Contrast / Noise N3

Systems

0

1

2

3

1 2 3 4 5 6 7 8 9 1 0 1 1 1 2 1 3 1 4 1 5 1 6

Lp /

mm

System 15 > 1.25 lp/mm

Fluoroscopy - Spatial Resolution

0

1

2

3

1 2 3 4 5 6 7 8 9 1 0 1 1 1 2 1 3 1 4 1 5 1 6

Lp /

mm

System 15 > 1.25 lp/mm

Fluoroscopy - Spatial Resolution

Page 9: Equipment standards for cardiology

Threshold Contrast Detail Detectability curves in the digital acquisition mode are plotted in figure 15. These curves are plotted against the published results of a system in good adjustment [29].

Figure 15: Digital Acquisition Mode – Threshold Contrast Detail Detectability

9

1 1 0

1

1 0

1 0 0 T h r e s h o l d D e t e c t i o n I n d e x

Th

resh

old

Det

ecti

on I

ndex

, H

T(A)

S q u a r e R o o t D e t a i l A r e a , A1/ 2

Page 10: Equipment standards for cardiology

Discussion:

From the results presented, it is evident that many interventional cardiac systems offer a wide range of selectable dose rates to the operator. On many systems, the selected mode is not prominently displayed to the operator although the difference in dose delivered from one mode to another may be quite substantial. More recently, an audible alarm has been fitted to some systems to alert the operator when the high fluoroscopy mode is in operation.

It is also evident from the results for new and old systems that dose levels are manufacturer dependent. Dose rates were found to be consistently higher for one manufacturer than for another manufacturer (100% more or greater), in both the fluoroscopy and digital acquisition modes. Justification for this in terms of image quality was not evident for the standard range of quality assurance tests [11-29].

Dose levels classified as ‘high’ by one manufacturer were found to be equivalent to those classified as ‘normal’ by another manufacturer.

Various faults were found on the systems during acceptance testing. On one system, the ‘normal’ fluoroscopy mode of operation was configured to 12.5pps but configured to 25pps on the equivalent system in the room next door. The default configurations were not consistent across the systems tested with 20% of the systems automatically defaulting to the high mode of operation. Four of the systems tested had no pulsed fluoroscopy option available. Of these, three systems were old systems but the other system was a newly purchased system and although the option had been purchased, there had been a failure to install it.

On the vast majority of systems tested, there was no indication of the spectral filter is use. In general, the spectral filters were found to operate under automatic control, without the operator’s knowledge, and on one new system, the filter became jammed out of the beam during commissioning testing.

Several issues relating to DAP meters were identified and in general it was found that the DAP meters were not calibrated to the radiology systems in question and account was not taken of the attenuation of X-ray table nor the spectral filtration.

On one old system, the irradiated field size was found to be substantially greater than the imaged field size, ranging from 60% to 320% greater depending on the field size selected. In general, it was found that there was a requirement for a more meaningful display to the operator of radiation risk or injury to the patient.

In terms of image quality, all systems were found to have a measured threshold contrast of < 4% although 66% of the systems commissioned had a measured threshold contrast of > 2.7% which was higher than would have been expected for new systems of this type. The spatial resolution, visible on the diagnostic monitor, was as expected on all systems with the exception of one old system. No improvement in image quality was detected, using published methodology [12-16], for the systems operating at higher dose levels.

The imaged field size was found to be less than 85% of the detector nominal diameter in more than 90% of systems tested. This may be the result of a magnification effect due to the positioning of crash barriers at the detectors. 66% of systems commissioned demonstrated electrical safety faults.

10

Page 11: Equipment standards for cardiology

Conclusions:

The results of this study are consistent with findings from previous commissioning surveys [30]. Problems were identified with all radiological systems commissioned, with approximately 70% of systems demonstrating significant problems. This emphasises the importance of commissioning radiological equipment and interventional cardiac equipment in particular given the associated levels of risk [1-5]. It also demonstrates the importance of including electrical safety testing in a commissioning programme.

The findings of this study, particularly in the area of dose and image quality, demonstrate the need for definitive equipment requirements and standardisation in the design, manufacture, acceptance and maintenance of equipment for interventional cardiology.

The existing interventional standard IEC60601-2-43, 2000, deals with requirements for the safety of X-ray equipment for interventional procedures [24]. Dedicated interventional equipment for cardiology poses its own unique requirements over and above those of IR equipment in general and warrants consideration to ensure that the particular requirements are well considered and properly reflected in the equipment manufactured.

Proposals for updates to international equipment standards have been prepared in order to address the current imbalance between the advancing technology in the area of interventional cardiology and existing standards. Contact has been made with the International Electrotechnical Commission (IEC), the body with responsibility for standards in this area and a formal proposal has been submitted. The proposed additions are most notably in the areas of Dose and Image Quality, for both Image Intensifier and flat panel technology, which are not addressed by the existing standards. It is essential that research keeps pace with developments in digital imaging technology if Quality Assurance efforts are to be properly guided and sufficiently comprehensive, and if equipment related standards are to remain relevant in this rapidly changing environment.

This study was partly funded by the European Commissions 5th Framework Programme, Nuclear fission and Radiation Protection Contract, DIMOND III (FIGM-CT-2000-00061).

References :

[1] EU DIMOND III project proposal, Measures for Optimising Radiological Information Content and Dose in Digital Imaging and Interventional Radiology, 1999.[2] Radiological Society of North America, Categorical Course in Diagnostic Radiology Physics: Cardiac Catheterisation Imaging, 1998.[3] ICRP 85, Avoidance of Radiation Injuries from Medical Interventional Procedures, Vol 30, 2, 2000[4] Balter S., Interventional fluoroscopy, physics, technology and safety, Wiley-Liss, 2001.[5] Neofotistou V, Vano E, Padovani R, Kotre J, Dowling A et al, Preliminary reference levels in interventional cardiology, European Radiology, Springer-Verlag 2003, 10.1007/s00330-003-1831-x[6] Timmis A, Nathan A, Essentials of Cardiology. Blackwell Scientific Publications, 1993. [7] NRPB, Radiation Exposure to the UK population from Medical and Dental X-ray Examinations, NRPB-W4, D. Hart and B. Wall, 2002. [8] Council Directive 97/43/Euratom on health protection of individuals against the dangers of ionising radiation in relation to medical exposure, 1997.

11

Page 12: Equipment standards for cardiology

[9] Statutory Instruments SI 478 of 2002, European Communities (Medical Ionising Radiation Protection) Regulations, 2002, Stationery Office, Dublin. [10] Medical Devices Agency, Cardiovascular Imaging Systems, update report, GE Medical Systems Innova 2000, Evaluation Report MDA 01144, 2001.[11] Cardiac Catheterisation Equipment Performance AAPM Report No. 70, 2001.[12] British Institute of Radiology (BIR), Assurance of Quality in the Diagnostic Imaging Department (2nd Edition) (2001).[13] Institute of Physics and Engineering in Medicine (IPEM), Recommended Standards for the Routine Performance Testing of Diagnostic X-ray Imaging Systems, Report No. 77 (1997).[14] Institute of Physics and Engineering in Medicine (IPEM), Measurement of the Performance Characteristics of Diagnostic X-ray Systems used in Medicine, Report No. 32, Second Edition. Part 1: X-ray Tubes and Generators. (1996).[15] IPEM, Measurement of the Performance Characteristics of Diagnostic X-ray Systems used in Medicine, Part II, X-ray Image Intensifier Television Systems, Report no. 32, 2nd Edition 1996. [16] MDA Evaluation Report, The Testing of X-ray Image Intensifier Television Systems: 1998, MDA/98/68.[17] Martin CJ, Sutton, Workman A, Shaw A, Temperton D, Protocol for measurement of patient entrance surface dose rates for fluoroscopic equipment, The British Journal of Radiology, 71, (1998), 1283 – 1287.[18] Institute of Physical Sciences in Medicine (IPSM), Data for Estimating X-ray Tube Total Filtration, Report No. 64.[19] European Commission (EC), Radiation Protection 91, Criteria for Acceptability of Radiological (including radiotherapy) and Nuclear Medicine Installations (Luxembourg: Office for Official Publications of the European Communities) (1997).[20] National Radiological Protection Board (NRPB), National Protocol for Patient Dose Measurements in Diagnostic Radiology, Institute of Physical Sciences in Medicine, (1992).[21] AAPM, Quality Control in Diagnostic Radiology, Report of task group no 12, Diagnostic X-ray Imaging Committee, AAPM report no 74, 2002.[22] NCRP, Quality Assurance for Diagnostic Imaging, report no 99, 1997. [23] IEC 61223-3-1 Evaluation and routine testing in medical imaging departments: Acceptance tests – imaging performance of X-ray equipment for radiographic and radioscopic systems. [24] IEC 60601-2-43 International Standard, Medical Electrical Equipment, Part 2-43 Particular requirements for the safety of X-ray equipment for interventional procedures, 2000.[25] International Electrotechnical Commission (IEC), Medical Electrical Equipment, Part 1: General requirements for safety, 60601-1 (1988).[26] Medical Devices Agency, KCARE, Comparative Specification of Single Plane Cardiac Systems, Evaluation Report MDA 01034, 2001.[27] International Standard IEC-60580, Medical Electrical Equipment - Dose Area Product Meters, (IEC 60580:2000(E)).[28] Institute of Physics and Engineering in Medicine (IPEM), Medical and Dental Guidance Notes, A Good Practise Guide on all aspects of Ionising Radiation Protection in the Clinical Environment, (2002).[29] Medical Devices Agency, Cardiovascular Imaging Systems, A comparative report, Evaluation Report MDA 01143, 2001.[30] Dowling A, Kenny T, Malone J, A Critical Overview of Acceptance Testing using various Measured Indices, Radiation Protection Dosimetry, p53 – 59, Vol 94, 1-2, 2001.

12

00.2

0.4

0.6

0.8

1

1.2

1.4

1 3 5 7 9 11 13 15

Gy/

sec

System

Detector Entrance Dose Rates Grouped by Manufacturer0

5

10

15

20

25

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

Martin: Low < 15 mGy/min; Normal 15-30 mGy/min; High 31 – 45 mGy/min IPEM: 50 mGy/min; IAEA: 25 mGy/min, WHO: 17 mGy/min

* *

X

mG

y/m

in

System

Patient Entrance Dose Rates Grouped by Manufacturer

0

1020

3040

5060

70

1 3 5 7 9 11 13 15

Cont

High

Martin: Low < 15 mGy/min; Normal 15-30 mGy/min; High 31 – 45 mGy/min IPEM: 50 mGy/min; High > 1.5 x; IAEA: 25 mGy/min; WHO: 17 mGy/min; RSNA High: 20-43

Systems

Patient Entrance Dose Rates - Grouped by Manufacturer

mG

y/m

in

0

0.05

0.1

0.15

0.2

0.25

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

Gy/

exp

osu

re

System

Detector Entrance Dose / Exposure Grouped by Manufacturer

0

1

2

3

4

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

6 (40%) < 2.7%

% T

hre

sho

ld C

on

tras

t

Fluoroscopy - Low Contrast / Noise N3

Systems

0

1

2

3

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

Lp

/ m

m

System 15 > 1.25 lp/mm

Fluoroscopy - Spatial Resolution

00.2

0.4

0.6

0.8

1

1.2

1.4

1 3 5 7 9 11 13 15

Gy/

sec

System

Detector Entrance Dose Rates Grouped by Manufacturer

00.2

0.4

0.6

0.8

1

1.2

1.4

1 3 5 7 9 11 13 15

Gy/

sec

System

Detector Entrance Dose Rates Grouped by Manufacturer

00.20.40.60.811.21.41357911 1315

ContHigh G

y/se

c

SystemDetector Entrance Dose Rates Grouped by ManufacturerFigure 3.

00.2

0.4

0.6

0.8

1

1.2

1.4

1 3 5 7 9 11 13 15

Gy/

sec

System

Detector Entrance Dose Rates Grouped by Manufacturer

00.2

0.4

0.6

0.8

1

1.2

1.4

1 3 5 7 9 11 13 15

Gy/

sec

System

Detector Entrance Dose Rates Grouped by Manufacturer

00.2

0.4

0.6

0.8

1

1.2

1.4

1 3 5 7 9 11 13 15

Gy/

sec

System

Detector Entrance Dose Rates Grouped by Manufacturer

00.2

0.4

0.6

0.8

1

1.2

1.4

1 3 5 7 9 11 13 15

Gy/

sec

System

Detector Entrance Dose Rates Grouped by Manufacturer

00.2

0.4

0.6

0.8

1

1.2

1.4

1 3 5 7 9 11 13 15

Gy/

sec

System

Detector Entrance Dose Rates Grouped by Manufacturer0

5

10

15

20

25

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

Martin: Low < 15 mGy/min; Normal 15-30 mGy/min; High 31 – 45 mGy/min IPEM: 50 mGy/min; IAEA: 25 mGy/min, WHO: 17 mGy/min

* *

X

mG

y/m

in

System

Patient Entrance Dose Rates Grouped by Manufacturer

00.2

0.4

0.6

0.8

1

1.2

1.4

1 3 5 7 9 11 13 15

Gy/

sec

System

Detector Entrance Dose Rates Grouped by Manufacturer0

1020

3040

5060

70

1 3 5 7 9 11 13 15

Cont

High

Martin: Low < 15 mGy/min; Normal 15-30 mGy/min; High 31 – 45 mGy/min IPEM: 50 mGy/min; High > 1.5 x; IAEA: 25 mGy/min; WHO: 17 mGy/min; RSNA High: 20-43

Systems

Patient Entrance Dose Rates - Grouped by Manufacturer

mG

y/m

in

0

1

2

3

4

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

6 (40%) < 2.7%

% T

hre

sho

ld C

on

tras

t

Fluoroscopy - Low Contrast / Noise N3

Systems