evaluation of diagnostic platforms for g6pd deficiency -...

1
Average G6PD activity (U/gHb) Trinity G6PD activity (U/gHb) Difference in G6PD activity (U/gHb) + 1.96 SD - 1.96 SD Mean Pearson correlation coefficient [0.7585] Linear fit equation: y = 3.84 +0.89x R 2 = 0.575 0 2 4 6 8 10 12 14 16 18 20 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 R&D G6PD activity (U/gHb) 20 18 16 14 12 10 8 6 4 2 0 4 2 0 -2 6 8 10 conclusions A comparison of two quantitative G6PD tests shows: Both tests equally classify severe G6PD deficient samples. Reference values vary widely depending on the reference test used. The BinaxNOW® test accurately detected all severely deficient patients. The fluorescent spot test only accurately detects all severely deficient specimens if intermediates are considered as G6PD deficient. By cytochemical staining, it was possible to detect heterozygous females in the population. Only three of these would have been excluded from treatment using a moderately deficient cut-off of 60% normal activity. WWW.PATH.ORG introduction Treatment with anti-malarial drugs in the 8-aminoquinolone group (e.g., primaquine, pamaquine and tafenoquine) can cause acute hemolysis in people with glucose-6-phosphate dehydrogenase (G6PD) deficiency.As such, there is a need for a test that is appropriate for G6PD deficiency screening in order to ensure safe administration of these drugs for radical cure of Plasmodium vivax. We evaluated different assays and platforms for determining the G6PD status of a patient. methods Assays used to measure G6PD activity Trinity Biotech spectrophotometer test: A quantitative kinetic assay expressing G6PD activity in units per gram of hemoglobin. R&D Diagnostics Ltd® enzymatic colorimetric method: A quantitative kinetic colorimetric assay expressing G6PD activity in units per gram of hemoglobin. Trinity Biotech Fluorescent spot test: Most common qualitative assay,in which fluorescence of nicotinamide adenine dinucleotide phosphate (NADPH) is directly observed. BinaxNOW® G6PD Test: A qualitative assay that uses a lateral flow test platform. Cytofluorometric assay: Allows observation of mosaic red blood cell populations in females. evaluation of diagnostic platforms for g6pd deficiency Nicole LaRue, Maria Kahn, Michael Kalnoky, Brandon T. Leader, Pooja Bansil, Sarah McGray, Gonzalo J. Domingo table 1: Trinity Biotech Reference Summary of G6PD activity (N = 214) Reference values Total Female Male Adjusted male Number of cases 214 107 107 101 Mean (U/g Hb) 7.17 7.70 6.63 6.99 Standard deviation (U/g Hb) 7.295 7.69 7.06 7.18 Median (U/g Hb) 7.295 7.69 7.06 7.18 Range (U/g Hb) 0.12–14.04 1.03–14.04 0.12–12.26 0.84–12.26 table 2: Number and percent distribution of BinaxNOW® and Trinity Biotech Fluorescent Spot test results (N = 214) Number (N) Percent (%) BinaxNOW® Test Normal (N=201) 182 90.5 Deficient (N=201) 19 9.45 Untreated (N=214) 13 6.07 Trinity Biotech Fluorescent Spot Test (FST) Normal 189 88.32 Deficient 13 6.07 Intermediate 12 5.61 Deficient + intermediate 25 1.68 Normal + intermediate 201 93.93 figure 2: Comparison of G6PD activity classified samples by the Trinity Biotech quantitative assay to two qualitative tests. The shading in the table indicates discrepancies in definition of phenotype between the tests within intermediate ranges. Classification by Trinity Biotech Quantitative Trinity Biotech Fluorescent Spot Test BinaxNOW® Trinity Biotech Quantitative Calculated Activity Sample #147 (female) NORMAL 7.50 Sample #141 (male) NORMAL 8.24 Sample #110 (female) INTERMEDIATE 3.40 Sample #105 (male) DEFICIENT 0.12 Sample #92 (male) DEFICIENT 0.66 Sample #145 (male) DEFICIENT 0.69 figure 1: Pearson’s correlation coefficient and a Bland Altman scatterplot describing the relationship between the Trinity Biotech quantitative and R&D Diagnostics Ltd® quantitative test for G6PD activity. results figure 3: G6PD zygosity measured by mean FITC fluorescence compared to G6PD activity measured by Trinity Biotech. 0 5 10 15 250 300 350 400 450 500 550 600 650 10% 20% 30% 60% Mean FITC Channel Intensity measure by Cytometer G6PD Zygosity measured by mean FITC Fluorescence and Trinity Zygosity determined by Flow Cytometry & Percent Normal determined by Trinity G6PD activity measured by trinity (U/(g/dl)Hb) Normal Heterozygous Deficient 60% normal 30% normal 20% normal 10% normal results continued results continued We thank Amanda Vilbrandt and Scott Brown for their support in the development of this poster. This poster is based on research funded by the Bill & Melinda Gates Foundation. For further information please contact [email protected], PATH, Seattle, WA. table 3: Clinical performance of the BinaxNOW® and Trinity Biotech Fluorescent Spot screening tests for detection of deficient G6PD activity* Cut-off value (% of adjusted normal male value, 7.18) 10% 20% 30% 60% Cut-off value (U/gHb) 0.718 1.436 2.154 4.308 Number of samples with G6PD levels below cut-off (%) 6 (2.80) 16 (7.48) 18 (8.41) 23 (10.75) BinaxNOW G6PD Test Sensitivity % (95%Cl) 100.0 (54.1, 100.0) 100.0 (79.4, 100.0) 100.0 (81.5, 100.0) 82.6 (61.2, 95.0) Specificity % (95%Cl) 93.3 (88.9, 96.4) 98.4 (95.3, 99.7) 99.5 (97.0, 100.0) 100.0 (97.9, 100.0) Trinity Fluorescent Spot Test a Sensitivity % (95%Cl) 100.0 (54.1, 100.0) 100.0 (79.4, 100.0) 100.0 (81.5, 100.0) 91.3 (72.0, 98.9) Specificity % (95%Cl) 90.9 (86.1, 94.4 95.5 (91.5, 97.9) 96.4 (92.8, 98.6) 97.9 (94.7, 99.4) Trinity Fluorescent Spot Test b Sensitivity % (95%Cl) 83.3 (35.9, 94.6) 75.0 (47.6, 92.7) 72.2 (46.5, 90.3) 56.5 (34.5, 76.8) Specificity % (95%Cl) 96.2 (92.6, 98.3) 99.5 (97.2, 100.0) 100.0 (98.1, 100.0) 100.0 (98.1, 100.0) *Gold standard: Trinity Biotech test a Trinity Biotech spot test intermediate test results combined with deficient test results b Trinity Biotech spot test intermediate test results combined with normal test results 1275

Upload: dangkhanh

Post on 22-Mar-2018

228 views

Category:

Documents


3 download

TRANSCRIPT

Average G6PD activity (U/gHb) Trinity G6PD activity (U/gHb)

Di�

eren

ce in

G6P

D a

ctiv

ity

(U/g

Hb)

+ 1.96 SD

- 1.96 SD

Mean

Pearson correlation coe�cient [0.7585]

Linear fit equation: y = 3.84 +0.89xR2 = 0.575

0 2 4 6 8 10 12 14 16 18 20 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15

R&D

G6P

D a

ctiv

ity

(U/g

Hb)

20

18

16

14

12

10

8

6420

4

2

0

-2

6

8

10

conclusions• AcomparisonoftwoquantitativeG6PDtestsshows:

– BothtestsequallyclassifysevereG6PDdeficient samples.

– Referencevaluesvarywidelydependingonthereferencetestused.

• TheBinaxNOW®testaccuratelydetectedallseverelydeficientpatients.

• ThefluorescentspottestonlyaccuratelydetectsallseverelydeficientspecimensifintermediatesareconsideredasG6PDdeficient.

• Bycytochemicalstaining,itwaspossibletodetectheterozygousfemalesinthepopulation.Onlythreeofthesewouldhavebeenexcludedfromtreatmentusingamoderatelydeficientcut-offof60%normalactivity.

w w w . p a t h . o r g

introductionTreatmentwithanti-malarialdrugsinthe8-aminoquinolonegroup(e.g.,primaquine,pamaquineandtafenoquine)cancauseacutehemolysisinpeoplewithglucose-6-phosphatedehydrogenase(G6PD)deficiency.Assuch,thereisaneedforatestthatisappropriateforG6PDdeficiencyscreeninginordertoensuresafeadministrationofthesedrugsforradicalcureofPlasmodium vivax.

WeevaluateddifferentassaysandplatformsfordeterminingtheG6PDstatusofapatient.

methodsAssays used to measure G6PD activity Trinity Biotech spectrophotometer test:AquantitativekineticassayexpressingG6PDactivityinunitspergramofhemoglobin.

R&D Diagnostics Ltd® enzymatic colorimetric method:AquantitativekineticcolorimetricassayexpressingG6PDactivityinunitspergramofhemoglobin.

Trinity Biotech Fluorescent spot test:Mostcommonqualitativeassay,inwhichfluorescenceofnicotinamideadeninedinucleotidephosphate(NADPH)isdirectly observed.

BinaxNOW® G6PD Test:Aqualitativeassaythatusesalateralflowtestplatform.

Cytofluorometric assay:Allowsobservationofmosaicredbloodcellpopulationsinfemales.

evaluation of diagnostic platforms for g6pd deficiency Nicole LaRue, Maria Kahn, Michael Kalnoky, Brandon T. Leader, Pooja Bansil, Sarah McGray, Gonzalo J. Domingo

table 1: trinity Biotech reference Summary of g6pD activity (N = 214)

reference values total Female Male adjusted male

Number of cases 214 107 107 101Mean (U/g Hb) 7.17 7.70 6.63 6.99Standard deviation (U/g Hb) 7.295 7.69 7.06 7.18Median (U/g Hb) 7.295 7.69 7.06 7.18

Range (U/g Hb) 0.12–14.04 1.03–14.04 0.12–12.26 0.84–12.26

table 2: Number and percent distribution of BinaxNow® and trinity Biotech Fluorescent Spot test results (N = 214)

Number (N) percent (%)

BinaxNow® test

Normal (N=201) 182 90.5

Deficient (N=201) 19 9.45

Untreated (N=214) 13 6.07

trinity Biotech Fluorescent Spot test (FSt)

Normal 189 88.32

Deficient 13 6.07

Intermediate 12 5.61

Deficient + intermediate 25 1.68

Normal + intermediate 201 93.93

figure 2: Comparison of g6pD activity classified samples by the trinity Biotech quantitative assay to two qualitative tests. the shading in the table indicates discrepancies in definition of phenotype between the tests within intermediate ranges.

Classification by trinity Biotech Quantitative

trinity Biotech Fluorescent Spot

testBinaxNow®

trinity Biotech Quantitative

Calculated activity

Sample #147 (female) NORMAL 7.50

Sample #141 (male) NORMAL 8.24

Sample #110 (female) INTERMEDIATE

3.40

Sample #105 (male) DEFICIENT 0.12

Sample #92 (male) DEFICIENT

0.66

Sample #145 (male) DEFICIENT

0.69

figure 1: pearson’s correlation coefficient and a Bland altman scatterplot describing the relationship between the trinity Biotech quantitative and r&D Diagnostics Ltd® quantitative test for g6pD activity.

results

figure 3: g6pD zygosity measured by mean FItC fluorescence compared to g6pD activity measured by trinity Biotech.

0 5 10 15250

300

350

400

450

500

550

600

650

10% 20% 30% 60%

Mea

n FI

TC C

hann

el In

tens

ity

mea

sure

by

Cyto

met

er

G6PD Zygosity measured by mean FITC Fluorescence and Trinity Zygosity determined by Flow Cytometry & Percent Normal determined by Trinity

G6PD activity measured by trinity (U/(g/dl)Hb)

NormalHeterozygousDeficient60% normal30% normal20% normal10% normal

results continued results continued

We thank Amanda Vilbrandt and Scott Brown for their support in the development of this poster. This poster is based on research funded by the Bill & Melinda Gates Foundation. For further information please contact [email protected], PATH, Seattle, WA.

L B - 2 0 1 8

table 3: Clinical performance of the BinaxNow® and trinity Biotech Fluorescent Spot screening tests for detection of deficient g6pD activity*

Cut-off value (% of adjusted normal male value, 7.18)

10% 20% 30% 60%

Cut-off value (U/gHb) 0.718 1.436 2.154 4.308

Number of samples with G6PD levels below cut-off (%)

6 (2.80) 16 (7.48) 18 (8.41) 23 (10.75)

BinaxNow g6pD test

Sensitivity % (95%Cl) 100.0 (54.1, 100.0)

100.0 (79.4, 100.0)

100.0 (81.5, 100.0)

82.6 (61.2, 95.0)

Specificity % (95%Cl) 93.3 (88.9, 96.4)

98.4 (95.3, 99.7)

99.5 (97.0, 100.0)

100.0 (97.9, 100.0)

trinity Fluorescent Spot testa

Sensitivity % (95%Cl) 100.0 (54.1, 100.0)

100.0 (79.4, 100.0)

100.0 (81.5, 100.0)

91.3 (72.0, 98.9)

Specificity % (95%Cl) 90.9 (86.1, 94.4

95.5 (91.5, 97.9)

96.4 (92.8, 98.6)

97.9 (94.7, 99.4)

trinity Fluorescent Spot testb

Sensitivity % (95%Cl) 83.3 (35.9, 94.6)

75.0 (47.6, 92.7)

72.2 (46.5, 90.3)

56.5 (34.5, 76.8)

Specificity % (95%Cl) 96.2 (92.6, 98.3)

99.5 (97.2, 100.0)

100.0 (98.1, 100.0)

100.0 (98.1, 100.0)

*Gold standard: Trinity Biotech test a Trinity Biotech spot test intermediate test results combined with deficient test resultsb Trinity Biotech spot test intermediate test results combined with normal test results

1 2 7 5