exceed abt acetabular system - surgitech exceed abt...surgical technique surgical protocols by mr e...
TRANSCRIPT
Surgical Technique
Surgical Protocols by Mr E J Smith BSc MB BCh FRCS
Exceed ABT Acetabular System
Over 1 million times per year, Biomet helps one surgeon
provide personalized care to one patient.
The science and art of medical care is to provide the right
solution for each individual patient. This requires clinical
mastery, a human connection between the surgeon and the
patient, and the right tools for each situation.
At Biomet, we strive to view our work through the eyes of
one surgeon and one patient. We treat every solution we
provide as if it’s meant for a family member.
Our approach to innovation creates real solutions that assist
each surgeon in the delivery of durable personalized care
to each patient, whether that solution requires a minimally
invasive surgical technique, advanced biomaterials or a
patient-matched implant.
When one surgeon connects with one patient to provide
personalized care, the promise of medicine is fulfilled.
One Surgeon. One Patient.
Table of ContentsExceed ABT Acetabular System
Preoperative Planning ............................................................................................................................................5Manual Preoperative PlanningDigital Preoperative Planning
Surgical Exposure ..................................................................................................................................................6 Preparation of the Acetabulum
Gauging the Acetabulum .......................................................................................................................................7
Preliminary Trial Reduction (without implants) .......................................................................................................7Trial Reduction with Trial BIOLOX® delta or E1 Tapered BearingTrial Reduction with Trial E1 RingLoc-X Bearing
Insertion of Acetabular Shell ................................................................................................................................10RingLoc-X Impaction PlatesRingLoc-X Ring RetainersFinal Seating of the Shell
Supplementary Screw Fixation ............................................................................................................................14Exceed ABT Multi-hole/Revision ShellScrew Hole Preparation
Secondary Trial Reduction with the Trial Liners (with implants) ..........................................................................15
Insertion of Acetabular Bearing ............................................................................................................................16Insertion of E1 RingLoc-X BearingInsertion of BIOLOX® delta or E1 Tapered Bearing
Modular Head Selection and Impaction ..............................................................................................................17
BIOLOX® delta Option Femoral Heads ................................................................................................................18Head Removal and Inspection of the Stem TaperAssembly Instructions
Removal of the Acetabular Bearing and Shell .....................................................................................................20Removal of BIOLOX® delta or E1 Tapered BearingRemoval of E1 RingLoc-X BearingRemoval of Exceed ABT Acetabular Shell
Ordering Information ............................................................................................................................................22
The Exceed ABT Acetabular System comprises fully instrumented acetabular components designed specifically for
cementless application. Each acetabular component comprises a titanium alloy shell and acetabular bearing. Bearings
for the Exceed ABT acetabular system are available in Vitamin E stabilised highly crosslinked polyethylene (E1), BIOLOX®
delta ceramic and in articulation diameters of 22.22, 28, 32, 36 and 40 mm. The initial press-fit fixation and later
secondary fixation of the Exceed ABT shell, into the prepared acetabulum, is enhanced by Biomet’s proprietary Porous
Plasma Spray (PPS) coating, a clinically proven closed pore structure. The additional nano-crystalline hydroxyapatite
coating, BoneMaster, retains the surface topography of PPS and further improves both bone/implant contact and
density.
For details of the complete product range including instrumentation, please refer to pages 22-28 of this document.
Exceed ABT Acetabular System
5
Figure 1
Preoperative PlanningSelection of the correct acetabular component is attained through careful preoperative planning. This can be achieved manually by means of x-ray templates, or digitally by means of a Picture Archiving and Communication System. (PACS)
Manual Preoperative PlanningThe Exceed ABT Acetabular system provides a comprehensive selection of acetabular x-ray templates in 100% and 115% magnification.
These templates are positioned over the anterior/posterior x-ray to best decide the correct implant size, centre of rotation and whether Standard, Hi-Wall, 10 degree inserts or angled shells are required to restore the patient’s natural anatomy.
Digital Preoperative PlanningThe Exceed ABT Acetabular System digital templates are available through various digital template providers. When using digital templating for a primary total hip replacement, it is necessary to use a magnification marker with a known dimension. This is required in order for the system to calculate the correct magnification.
As soon as the correct magnification has been determined, the system can be used to best decide the correct implant size, centre of rotation and whether Standard, Hi-Wall, 10 degree inserts or angled shells are required to restore the patient’s natural anatomy.
Figure 2
Exceed ABT Acetabular System
6
Preparation of the AcetabulumIt is important to remove all articular cartilage and any soft tissue protruding into the acetabulum. In order to seat the acetabular component, some loss of subchondral bone may be inevitable. However, it is important to note, that subchondral bone, particularly peripheral bone, should be preserved.
The acetabular preparation (Figure 4) is undertaken using the grater reamers supplied. The axis of the grater reamer should be positioned at 40 to 45 degrees from the vertical axis and in 20 to 25 degrees of anteversion (Figure 5). Begin reaming with the smallest available grater reamer, increasing the diameter of each sequential grater reamer until the required diameter is achieved. As the shell is hemispherical, it is recommended to under ream the bone by 1-2 mm depending on bone quality.
E.g. 54 mm diameter Exceed ABT acetabular shell requires ream diameter of 52/53 mm.
NOTE: If acetabular osteotomes are used, final shaping must still be achieved using the hemispherical grater reamer to ensure a congruent fit between the shell and the floor of the acetabulum.
Figure 3
Figure 4
Figure 5
Surgical ExposureThe Exceed ABT system can be implanted using any of the standard approaches for total hip replacement. The aim of the approach selected is to provide adequate visualization of both the acetabulum and proximal femur (Figure 3).
7
Gauging the AcetabulumThoroughly clean and dry the acetabulum ensuring that any overhanging soft tissue has been excised. Attach the monoblock insertion handle to the trial shell/gauge. Care must be taken to thread the handle perpendicular to the face of the gauge to avoid risks of cross-threading or tip breakage upon impaction. Insert the trial shell/gauge into the reamed acetabulum (Figure 6). The diameter should correspond to the diameter of the implant to be used*. These trial shells/gauges are used to, a) check the stability of the final implant diameter prior to insertion, and b) to ensure the acetabulum is hemispherical. Should the trial shell/gauge be unstable, or if there are gaps between the trial shell/gauge and the acetabulum, it may be necessary to increase the diameter of the final grater reamer. However, in some instances it may not be possible to increase the reamed diameter. If this is indeed the case, then supplementary screw fixation will be necessary. Please refer to page 14 for method of supplementary screw fixation.
* i.e 1-2 mm greater than the last grater reamer used
Preliminary Trial Reduction (without implants)The preliminary trial reduction of the hip joint can only be carried out once the acetabulum and femur have been fully prepared in accordance with the relevant surgical technique. With the trial shell/gauge firmly located within the acetabulum, a preliminary trial reduction can be completed in order to assess the stability of the joint and to revalidate the component type and size selected at the preoperative planning stage (Figure 7 & 8). For information on head to shell ratios please see Table 1 below.
Table 1. Exceed ABT Acetabular Shell System
Bearing Material
Shell Diameter (mm)
40 42 44 46 48 50 52 54 56 58 60 62 64 66 68 70
E1 RingLoc-X Liners
22.22
32
2836
- 40
BIOLOX® delta & E1 Tapered
Liners- 28 32
36
- 40
Figure 6 Figure 7 Figure 8
Exceed ABT Acetabular System
8
Trial Reduction with Trial BIOLOX® delta or E1 Tapered BearingThe Standard Exceed ABT acetabular shell should be used where adequate femoral head coverage and joint stability is achieved with the component at 40–45 degrees of inclination (Figure 9). The 15 degree Exceed ABT acetabular shell should be used to improve femoral head coverage and stability should the standard shell not suffice. Each 15 degree component is identified with the word ‘cranial’ located superiorly on the lip of the 15 degree feature and highlights the segment of the shell that should be positioned approximately 45 degrees superior/posterior within the acetabulum. It is also important to seat the inferior lip of the component within the rim of the acetabulum (Figure 10). Failure to follow these instructions may result in impingement between the neck of the femoral component and the 15 degree feature or the inferior rim of the component.
With the trial shell in place, select the correct diameter trial tapered insert (28, 32, 36 or 40 mm) to match the diameter of the shell. By selecting the appropriately sized trial tapered bearing, the head diameter is automatically selected (See Table 1 on page 7).
The trial tapered bearings without captive screw can be placed into the trial shell/gauge by hand or using screw forceps provided. The trial tapered bearings with captive screw are placed by hand into the gauge and then locked with a 3.5 mm hex-driver.
A trial reduction can then take place as long as the femur contains the relevant components and trial head.
To remove the trial without captive screw, the screw forceps are engaged into the apical holes and gripped. The trial can then be removed. To remove the trials with captive screws use the 3.5 mm hex-driver to release.
NOTE: The trial tapered liners with captive screws should only be used in conjunction with the trial shell gauges and not with the definitive Exceed ABT shell.
Figure 9 Figure 10
9
Trial Reduction with Trial E1 RingLoc-X BearingStandard E1 RingLoc-X bearings should be used where adequate femoral head coverage and joint stability is achieved. The Hi-Wall components provide additional head coverage and security against dislocation. They are positioned with the Hi-Wall segment of the liner aligned approximately 45 degrees superior/posterior within the acetabulum. The 10 degree component is used to change the inclination or anteversion of the bearing surface.
The E1 +3 mm RingLoc-X trial liner is used to lateralise the femoral head by 3 mm. It can be selected as an alternative to a modular femoral head with a longer neck.
Once the adequate trial components are selected and inserted, the trial reduction can be carried out.
Figure 11
NOTE: RingLoc-X trial reduction can only be completed with the definitive implant. Please refer to page 15.
Care must be taken to ensure that the correct joint biomechanics are restored. Soft tissue tension, risk of impingement, range of motion and leg length should be assessed. The trial reduction procedure should be repeated until satisfactory joint stability and biomechanics are achieved (Figure 11).
Exceed ABT Acetabular System
10
Insertion of Acetabular ShellUpon correct sizing, attach the appropriate Exceed ABT shell directly to the inserter handle. Check that the strike plate is connected to the spring loaded inner shaft by attempting to pull it free. If the strike plate is not connected, firmly push and twist it downwards. When fully seated (Figure 12), the threaded tip of the inner shaft will protrude from the outer shaft of the handle. Turn the strike plate clockwise to firmly tighten the spring loaded threaded tip in the shell apical hole. Hold the shell with the other hand whilst engaging the thread to prevent the shell from rotating. Once assembled to the shell attach the alignment guide to the handle to aid accurate placement (Figure 13, 14 & 15).
IMPORTANT: It is advised to check that the threaded tip is always fully engaged in the cup apical hole when impacting. This helps minimise any risks of threaded tip damage.
For correct placement and orientation of the shell, the dome hole cluster should be positioned in the superior/posterior quadrant. The shell should be impacted at 40 to 45 degrees of inclination from the horizontal axis of the pelvis (Figure 14) and with 20-25 degree anteversion (Figure 15).
When implanting the RingLoc-X shell, it is recommended to use the Exceed ABT impaction plates or ring retainers.
Figure 12 Figure 13 Figure 14 Figure 15
45°
45°
11
RingLoc-X Impaction PlatesThe impaction plates help prevent potential risks of locking ring disengagement from the shell upon impaction.
Select the correct sized impaction plate and position it onto the face of the definitive implant. The plates cut-outs are positioned on the lugs of the cup (Figure 16).
Check that the compression sleeve on the impaction handle is in its upper position. Place the handle through the impaction plates and attach the handle securely to the implant. Turn the compression sleeve clockwise until it tightly clamps the impaction plate against the rim of the shell. Turn the handle strike plate to securely tighten the assembly (Figure 17).
IMPORTANT: Ensure impaction plate is fully seated on shell face to avoid damage to locking ring/locking ring groove (Figure 18).
Figure 16
Impaction plate is seated firmly against rim of cup
Space between impaction plate and rim of cup should not be present
Handle strike plate
Impaction Plate
Compression Sleeve
Figure 17 Figure 18
Incorrect seating of impaction plates.Correct seating of impaction plates.
Exceed ABT Acetabular System
12
RingLoc-X Ring RetainersThe ring retainers provide an alternative solution to prevent potential risks of locking ring disengagement from the shell upon impaction.
Select the correct sized ring retainer and position it into the shell (Figure 19).
Place the handle through the ring retainer and attach the handle securely to the implant (Figure 20).
Following impaction, removal of the ring retainer is possible using the removal forceps provided (Figure 21).
Figure 19 Figure 20 Figure 21
IMPORTANT: When implanting the RingLoc-X shell, it is important to check that the locking ring is fully engaged with the locking groove before impaction begins. If the impaction plates or ring retainers are not used for shell impaction and should the locking ring dissociate from the cup during impaction (but remain in the sterile field), fully seat the cup into the acetabulum. The locking ring can then be repositioned by hand into the cup locking groove. In case the locking ring has fallen outside the sterile field, or is deemed to be damaged by its disengagement from the locking groove, it must be replaced with a new spare locking ring.
13
Final Seating of the ShellIMPORTANT: The position of the shell is crucial in the use of hard on hard bearings (BIOLOX® delta liners) to reduce the risk of impingement.
Check via the apical hole to determine whether the shell is in full contact with the floor of the acetabulum. If not, the impactor handle must be re-attached to the shell and further impaction is required until the shell is fully seated. Failure to fully seat the shell into the acetabulum
may compromise the quality of fixation.
When utilising the curved shell impaction handle, to facilitate tightening or loosening of the thumb wheel, it may be necessary to utilise the 3.5 mm hex-driver contained within the Exceed ABT general instrument tray.
Once the Exceed ABT shell is fully seated, the apical hole is closed using the apical screw provided. The apical screw is secured into position by means of the 3.5 mm hex-driver (Figure 22).
The Exceed ABT acetabular shell is supplied as a solid shell with the option of removing titanium-blanking plugs intra-operatively, should supplementary screw fixation be necessary.*
*All Exceed ABT standard and 15 degree 3 hole shells are provided with titanium blanking plugs inserted in the Exceed ABT shell.
Figure 22
Exceed ABT Acetabular System
14
Supplementary Screw FixationFor primary cases where good bone stock is present and the shell is firmly seated within the acetabulum, the use of fixation screws is generally unnecessary. However, in cases where motion can be detected between the shell and the acetabulum, or where the bone quality is not optimal, supplementary screw fixation is advised. When dome screws are utilized they should be placed posterosuperiorly into the thick part of the ilium (Figure 23). Screw placement must be chosen carefully to avoid injury to neurovascular structures. Care should also be exercised when supplementary screw fixation is required, to avoid damaging or scratching the internal surfaces of the Exceed ABT acetabular components. Should screw fixation be necessary, the blanking screws are removed by means of a 3.5 mm hex-driver (Figure 24).
NOTE: Screws should never be placed in the anteromedial area of the acetabulum.
Exceed ABT Multi-hole/Revision ShellIn cases where extreme motion can be detected between the shell and the acetabulum, where the bone quality is not optimal or a revision shell is required, multi-screw fixation is advised. Screw placement must be chosen carefully to avoid injury to neurovascular structures. Care should also be exercised when supplementary screw fixation is required, to avoid damaging or scratching the internal surfaces of the Exceed ABT acetabular components.
NOTE: Titanium blanking plug screws are not provided as standard in the Exceed ABT multi-hole shell but all holes should be filled before inserting the bearing liner.
Additional titanium blanking plugs are available to fill spare screw holes.
Figure 23 Figure 24
15
Screw Hole PreparationPrepare the screw holes utilizing the quick connect drill bits, the drill guide and the flexible drill bit shaft (Figure 25). When drilling into the posterior/superior quadrant, place a finger posteriorly into the sciatic notch to ensure the screw cannot penetrate too deeply. Measure the length of the drill holes with the flexible depth gauge (Figure 26). Insert the low profile dome screws using the forceps and universal joint screwdriver (Figure 27).
NOTE: To avoid impingement of the acetabular insert, check that all screw heads are seated below the inner surface of the shell. Only use Biomet 6.5 mm low profile dome screws.
Secondary Trial Reduction with the Trial LinersFollowing the insertion and fixation of the acetabular shell, select the correct diameter trial acetabular bearing (22.22, 28, 32, 36 or 40 mm) that matches the diameter of the shell. By selecting the appropriately sized trial bearing, the head diameter is automatically selected (See Table 1 on page 7). This trial insert is then placed into the acetabular component by hand. A trial reduction can then take place. The trial reduction procedure should be repeated until satisfactory joint stability and biomechanics are achieved (Figure 28).
The removal forceps can be utilised with the trial tapered liners (without captive screws) to take them out more easily. Care must be taken not to scratch the shell taper when doing so.
NOTE: The trial tapered liners with captive screws should not be used in conjunction with the definitive Exceed ABT acetabular implant.
Figure 25
Figure 28
Figure 27
Figure 26
Exceed ABT Acetabular System
16
Insertion of Acetabular BearingThe definitive acetabular bearing should only be inserted into the acetabular shell once the shell is deemed to be stable and placed in the correct inclination and orientation. Failure to follow this instruction may cause premature implant failure.
Insertion of E1 RingLoc-X BearingThe definitive E1 RingLoc-X bearing may now be introduced. The bearing should be positioned initially by hand into the shell ensuring that it is correctly aligned both radially and longitudinally. The bearing is then finally impacted in position using the appropriately sized ball impactor and impactor handle (Figure 29). A visual indicator window is located within the shell to highlight whether or not the bearing is successfully restrained within the shell.
Insertion of BIOLOX® delta or E1 Tapered BearingThe definitive BIOLOX® delta or E1 Tapered bearing may now be positioned with the aid of the insertion device and handle, being careful not to misalign the insert (Figure 30). The surgeon should place their fingers around the face of the shell to ensure that the bearing is properly aligned. The edge of the bearing must be flush with the edge of the acetabular shell prior to applying any impaction force to the bearing (Figure 31).
Figure 29 Figure 30 Figure 31
17
Several moderate impactions of the bearing are necessary, utilising the appropriately sized ball impactor to ensure stable seating of the bearing (Figure 32). This is especially important when inserting ceramic bearings. Failure to follow these instructions may result in damaged ceramic bearings.
IMPORTANT: When inserting the trial or definitive bearing, the interior of the acetabular shell should be carefully cleaned and dried. The taper region of the acetabular insert should also be dry before insertion into the acetabular shell. Care should be taken not to scratch the taper surface of the BIOLOX® delta or E1 tapered bearing or the inner taper.
NOTE: The below table (Table 2) shows the different articulations available within the Exceed ABT system. BIOLOX® delta liners can only be used in conjunction with the Exceed ABT TaperFit shell. E1 can be used with the Exceed ABT TaperFit and RingLoc-X shell.
Modular Head Selection & ImpactionWith the definitive acetabular bearing in place and upon completion of femoral reconstruction and trial reduction, the corresponding modular head can now be selected (see Table 1, page 7). Position the modular head onto a dry and clean surface of the trunnion. Fully seat the modular head by means of firm axial impaction utilizing the femoral head pusher and mallet (Figure 33).
IMPORTANT: Ensure all taper surfaces are clean and dry before seating the modular head on the stem taper. It is important that the stem and cup tapers are new, as a used taper can reduce the fatigue strength of ceramic components.
Once the correct modular femoral head has been attached to the femoral component, the hip joint can be reduced (Figure 34).
Table 2: Liners
BIOLOX® delta E1
HeadsBIOLOX® delta Yes Yes
CoCrMo No Yes
Figure 32 Figure 33 Figure 34
Exceed ABT Acetabular System
18
BIOLOX® delta Option Femoral HeadsThe BIOLOX® delta Option modular femoral head system is available for use within the Exceed ABT acetabular system. The BIOLOX® delta Option heads can be used in conjunction with the BIOLOX® delta liners and E1 bearings. Please see Table 2, page 17.
The BIOLOX® delta Option heads can be used as a primary implant or as a revision option where a ceramic head is required to replace a ceramic head on a slightly damaged stem taper.
Definition regarding stem taper’s condition: Intact stem taper: condition of the stem taper after removal of the intact head.
Slightly damaged taper: condition of the stem taper after revision of the femoral head.
Head Removal and Inspection of the Stem TaperIn case of revision surgery, extract the remaining femoral head (and adapter, if applicable) with a suitable extraction instrument to avoid unnecessary damage to the stem taper.
Inspection of the stem taper and decision:
Pristine taper (see Figure 35A).
Tolerable condition with scratches of less than 0.25 mm (see Figure 35B).
Intolerable condition (see Figure 35C through to 35E).
Do not use the BIOLOX® delta Option modular femoral head system with tapers where damage is the same or exceeds that illustrated in Figure 35C to 35E.
Figure 35A
Pristine Taper
Figure 35B
Scratches less
than 0.25 mm
Figure 35C
Intolerable
Figure 35D
Intolerable
Figure 35E
Intolerable
19
Figure 36 Figure 37 Figure 38
Assembly InstructionsThe modular head components must be assembled together before positioning them onto the stem. Verify that the appropriate head size and matching tapers are being utilized before assembly.
The modular head components are assembled by placing the head onto the sleeve as shown in Figures 36 and 37. Ensure the tapers are clean and dry and they are aligned axially before applying pressure. The tapers are engaged once resistance is felt.
Place the BIOLOX® delta Option head assembly on the stem taper with a twisting motion, while applying manual pressure until it locks. Impact the modular head components onto the stem with several brisk mallet strikes using a plastic head impactor only (Figure 38). Metal impactors or any other metallic objects may scratch or crack the modular head bearing surface and, therefore, should not be used.
If the modular ceramic head becomes scratched or cracked, the head and sleeve must be replaced.
NOTE: In cases in which the BIOLOX® delta Option system cannot be put on the stem taper using more pressure, due to scratches or warping of more than 0.25 mm, the BIOLOX® delta Option system should not be used.
NOTE: In vitro testing has shown that the burst strength of BIOLOX® delta Option heads meet current standards when used on defects up to 0.25 mm. In vitro testing on trunnions with greater defects demonstrated a reduced burst strength for the ceramic head which would indicate an increased risk of failure. If the femoral stem is well fixed but the trunnion is damaged and the defects may be greater than 0.25 mm then the surgeon should consider the risks associated with use of the BIOLOX® delta Option heads against the risks associated with removal of the femoral component.
Exceed ABT Acetabular System
20
Removal of Acetabular Bearing & ShellShould it be necessary to remove a bearing from an Exceed ABT shell, you must first endeavour to identify the type of acetabular component inserted as the two shells have differing methods of bearing extraction.
Removal of BIOLOX® delta or E1 Tapered BearingShould it be necessary to remove a BIOLOX® delta or E1 tapered bearing, the insert remover can be used in conjunction with the insertion device. First attach the insertion device to the articulation surface of the bearing and impact the rim of the shell with the insert remover, being careful not to impact the bearing itself. The vibration created from the impact will loosen the bearing from the shell and allow it to be removed (Figure 39).
Removal of E1 RingLoc-X Bearing Should it be necessary to remove an E1 RingLoc-X bearing from an Exceed ABT shell, it will be necessary to locate the open end of the locking ring located within the shell. This open end of the locking ring is located inside the window situated on the mouth of the acetabular shell. To open the ring, utilize the circlip pliers to widen the opening and with the liner extraction tool, remove the liner from the shell (Figure 40).
Figure 39Figure 40
21
Removal of Exceed ABT Acetabular ShellOnce the bearing has been removed as described on page 20, unscrew the apical hole blanking screw with a 3.5 mm hex-driver. Remove any supplementary screw fixation to ensure any bone/implant fixation has been severed (Figure 41). Low profile screws are removed with a 3.5 mm hex-driver, whilst the bone/implant interface is severed with curved osteotomes.
Thread the tip of the monoblock cup insertion handle into the cup apical hole. Remove the cup from the acetabulum. Take extreme care when extracting the cup as too much force may damage the acetabular floor, peripheral rim and anterior and posterior columns of the acetabulum.
NOTE: The spring loaded impaction handle cannot be used for implant removal as the strike plate will detach.
Figure 41
22
Exceed ABT Acetabular System
ImplantsExceed ABT TaperFit Acetabular Components
Exceed ABT TaperFit Acetabular Liners
Size (mm)
15 Degree TaperFit Shells
PC HA/PC BM/PC
Ø46 233746 233746HA 233746BM
Ø48 233748 233748HA 233748BM
Ø50 233950 233950HA 233950BM
Ø52 233952 233952HA 233952BM
Ø54 234454 234454HA 234454BM
Ø56 234456 234456HA 234456BM
Ø58 234858 234858HA 234858BM
Ø60 234860 234860HA 234860BM
Ø62 235262 235262HA 235262BM
Ø64 235264 235264HA 235264BM
Ø66 235266 235266HA 235266BM
Ø68 235268 235268HA 235268BM
Ø70 235270 235270HA 235270BM
Size (mm)
BIOLOX® delta Liners
Ø28 Ø32 Ø36 Ø40
Ø46650-0790
Ø48
Ø50650-0791
Ø52
Ø54650-0795
Ø56
Ø58650-0796 650-0798
Ø60
Ø62
650-0797 650-0799
Ø64
Ø66
Ø68
Ø70
The colours correspond to those on the shell, liner labels and related trial liners.
Size (mm)
E1 Tapered Liners
Ø28 Ø32 Ø36
Std Std Std
Ø46E1-002837
Ø48
Ø50E1-003239
Ø52
Ø54E1-003644
Ø56
Ø58E1-003648
Ø60
Ø62
E1-003652
Ø64
Ø66
Ø68
Ø70
Size (mm)
Standard TaperFit Shells
PC HA/PC BM/PC
Ø46 123746 123746HA 123746BM
Ø48 123748 123748HA 123748BM
Ø50 123950 123950HA 123950BM
Ø52 123952 123952HA 123952BM
Ø54 124454 124454HA 124454BM
Ø56 124456 124456HA 124456BM
Ø58 124858 124858HA 124858BM
Ø60 124860 124860HA 124860BM
Ø62 125262 125262HA 125262BM
Ø64 125264 125264HA 125264BM
Ø66 125266 125266HA 125266BM
Ø68 125268 125268HA 125268BM
Ø70 125270 125270HA 125270BM
23
Implants
Exceed ABT E1 RingLoc-X Acetabular Liners and E1+3mm RingLoc-X Liners
Exceed ABT RingLoc-X Acetabular Shells
Size (mm)
3- Hole Shells Multi-Hole Shells
PC HA/PC BM/PC PC HA/PC
Ø40 131340 131340HA 131340BM - -
Ø42 131342 131342HA 131342BM - -
Ø44 131344 131344HA 131344BM - -
Ø46 131346 131346HA 131346BM - -
Ø48 131348 131348HA 131348BM - -
Ø50 131350 131350HA 131350BM 130450 130450HA
Ø52 131352 131352HA 131352BM 130452 130452HA
Ø54 131354 131354HA 131354BM 130454 130454HA
Ø56 131356 131356HA 131356BM 130456 130456HA
Ø58 131358 131358HA 131358BM 130458 130458HA
Ø60 131360 131360HA 131360BM 130460 130460HA
Ø62 131362 131362HA 131362BM 130462 130462HA
Ø64 131364 131364HA 131364BM 130464 130464HA
Ø66 131366 131366HA 131366BM 130466 130466HA
Ø68 131368 131368HA 131368BM 130468 130468HA
Ø70 131370 131370HA 131370BM 130470 130470HA
Size(mm)
Ø22.22 mm bearing Ø28 mm bearing
Std Hi-Wall 10o Std Hi-Wall 10o
Ø40 EP-042240 EP-052240 EP-062240
Ø42 EP-042242 EP-052242 EP-062242
Ø44 EP-042244 EP-052244 EP-062244
Ø46 EP-042846 EP-052846 EP-062846
Ø48 EP-042848 EP-052848 EP-062484
The colours correspond to those of the trial liners. The items boxed in pink are the E1 + 3mm RingLoc-X liners
Size (mm)Ø32 mm bearing Ø36 mm bearing Ø40 mm bearing
Std Hi-Wall 10o Std Hi-Wall 10o Std Hi-Wall 10o
Ø46 EP-073246 EP-083246
Ø48 EP-073248 EP-083248
Ø50 EP-043250 EP-053250 EP-063250 EP-073650 EP-083650
Ø52 EP-043252 EP-053252 EP-063252 EP-073652 EP-083652
Ø54 EP-043254 EP-053254 EP-063254 EP-043654 EP-053654 EP-063654 EP-074054 EP-084054
Ø56 EP-043256 EP-053256 EP-063256 EP-043656 EP-053656 EP-063656 EP-074056 EP-084056
Ø58 EP-043258 EP-053258 EP-063258 EP-043658 EP-053658 EP-063658 EP-044058 EP-054058 EP-064058
Ø60 EP-043260 EP-053260 EP-063260 EP-043660 EP-053660 EP-063660 EP-044060 EP-054060 EP-064060
Ø62 EP-043262 EP-053262 EP-063262 EP-043662 EP-053662 EP-063662 EP-044062 EP-054062 EP-064062
Ø64 EP-043264 EP-053264 EP-063264 EP-043664 EP-053664 EP-063664 EP-044064 EP-054064 EP-064064
Ø66 EP-043266 EP-053266 EP-063266 EP-043666 EP-053666 EP-063666 EP-044066 EP-054066 EP-064066
Ø68EP-043268 EP-053268 EP-063268 EP-043668 EP-053668 EP-063668 EP-044068 EP-054068 EP-064068
Ø70
24
Exceed ABT Acetabular System
Modular Heads - Biomet 12/14 Taper
Modular CoCr Heads - Biomet Type 1 Taper
Head Offset
CoCrMo Modular Heads
22.22 mm
28 mm 32 mm 36 mm 40 mm
-6mm - 650-0863 650-0870 650-0839 650-1030
-5mm 164441 - - - -
-3mm 163653 650-0864 650-0871 650-0840 650-1031
0mm 164440 650-0865 650-0872 650-0841 650-1032
+3mm - 650-0866 650-0873 650-0842 650-1033
+6mm - 650-0867 650-0874 650-0843 650-1034
+9mm - - - - 650-1035
+12mm - - - - 650-1036
HeadOffset
CoCrMo Modular HeadsBIOLOX® delta Ceramic
Modular Heads
BIOLOX® delta Option Ceramic Heads
28 mm 32 mm 36 mm 40 mm
650-1055 650-1056 650-1057 650-1058
22.22 mm 28 mm 32 mm 36 mm 40 mm 28 mm 32 mm 36 mm BIOLOX® delta Option Taper Sleeves (12/14)
-4 mm - - 650-0882 650-0887 650-1020 - 650-0833 650-0836 -
-3.5 mm - 650-0877 - - - 650-0830 - - -
-3 mm - - - - - - - - 650-1060
-2 mm 164133 - - - - - - - -
0 mm 164132 650-0878 650-0883 650-0888 650-1021 650-0831 650-0834 650-0837 650-1061
+2 mm 164131 - - - - - - - -
+3.5 mm - 650-0879 - - - 650-0832 - - -
+4 mm - - 650-0884 650-0889 650-1022 - 650-0835 650-0838 650-1062
+7 mm - - - - - - - - 650-1063
+8 mm - - - 650-0890 650-1023 - - 650-0667 -
+12 mm - - - - 650-1024 - - - -
Head Offset
BIOLOX® delta Ceramic Modular Heads
BIOLOX® delta Option Ceramic Heads
28 mm 32 mm 36 mm 40 mm
650-1055 650-1056 650-1057 650-1058
28 mm 32 mm 36 mm BIOLOX® delta Option Taper Sleeves (Type 1)
-6mm - - - 650-1064
-3mm 650-1159 650-1163 650-0660 650-1065
0mm 650-1158 650-1162 650-0661 650-1066
+3mm 650-1157 650-1161 650-0662 650-1067
+5mm - - - -
+6mm - 650-1160* 650-0663 650-1068
Modular BIOLOX® delta Ceramic Heads - Biomet Type 1 Taper
*The 32+6mm modular head is used ONLY with Biomet’s Ti6Al4V Type 1 taper stems
Implants
25
Implants
Product Part Number Size Description
650-1000650-1001650-1002650-1003650-1004650-1005650-1006650-1007650-1008650-1009650-1010650-1011650-1012650-1013650-1014650-1015
Ø40 mmØ42 mmØ44 mmØ46 mmØ48 mmØ50 mmØ52 mmØ54 mmØ56 mmØ58 mmØ60 mmØ62 mmØ64 mmØ66 mmØ68 mmØ70 mm
Exceed ABT Spare Locking Ring
103530103531103532103533103534103535103536103537103538103539
Ø6.5 mm x 15 mmØ6.5 mm x 20 mmØ6.5 mm x 25 mmØ6.5 mm x 30 mmØ6.5 mm x 35 mmØ6.5 mm x 40 mmØ6.5 mm x 45 mmØ6.5 mm x 50 mmØ6.5 mm x 60 mmØ6.5 mm x 70 mm
LP Cancellous Bone Screw
123741 Exceed ABT Apical Plug - Spare
123743 Exceed ABT Threaded Plug - Spare
26
Exceed ABT Acetabular System
Instrumentation
ProductPart
NumberSize Description
31-601120 —Exceed ABT General Instrument Tray
Assembly
31-601122 —Exceed ABT Supplementary
Screw Fixation Tray Assembly
31-601124 —Exceed ABT Trial
Standard Shell and Liner Tray Assembly
31-601126 —Exceed ABT Trial
15° Shell and Liner Tray Assembly
31-60150131-60150231-601503
—
Exceed ABT RingLoc-X Standard Trial Instr Tray & Instr
Exceed ABT RingLoc-X Hi-Wall Trial Instr Tray & Instr
Exceed ABT RingLoc-X 10° Trial Instr Tray & Instr
31-601518 —Exceed ABT RingLoc-X
+3 mm Liners Instr Tray Complete
E1-100000 —Exceed ABT E1 Tapered Trial Liners
Complete Set
Exceed ABT Acetabular System Instrumentation
27
Instrumentation
ProductPart
NumberSize Description
31-600055 —Grater Reamer Tray Assembly
(1 mm Increments)
31-60111231-10059231-60111431-100593
Ø3.2 mm x 40 mmØ3.2 mm x 50 mmØ3.2 mm x 60 mmØ3.2 mm x 70 mm
Stainless Steel Twist Drill
31-601128 22-36 mmExceed ABT Trial Modular
Head Tray Assembly Type 1
31-601130 22-36 mmExceed ABT Trial Modular
Head Tray Assembly (12/14 Taper)
31-601374 40/44 mmExceed ABT Trial Modular
Head Type 1 Tray Complete
31-601376 40/44 mmExceed ABT Trial modular
Head (12/14) Tray Complete
28
Exceed ABT Acetabular System
Instrumentation
Product Part Number Size Description
31-601621 31-60162231-60162331-60162431-60162531-60162631-60162731-60162831-60162931-60163031-60163131-60163231-60163331-60163431-601635
Ø40 mmØ42 mmØ44 mmØ46 mmØ48 mmØ50 mmØ52 mmØ54 mmØ56 mmØ58 mmØ60 mmØ62 mmØ64 mmØ66 mm
Ø68/70 mm
RingLoc-X/MH Impaction Plate
Exceed ABT RingLoc-X Impaction Plate
Exceed ABT RingLoc-X Ring Retainers
Product Part NumberSize
RingLoc-X/RingLocDescription
110007401
110007412
110007415
110007406
110007408
110007409
110007402
110007403
110007404
110007411
110007405
110007407
110007413
110007414
110007410
Ø40 mm/20
Ø42 mm
Ø44 mm/21
Ø46 mm/22
Ø48 mm
Ø50 mm/23
Ø52 mm
Ø54 mm/24
Ø56 mm
Ø58 mm/25
Ø60 mm/26
Ø62 mm
Ø64 mm/27
Ø66 mm
Ø68/70 mm/28
RingLoc-X Ring Retainer
29
The Complete Exceed ABT Family
BIOLOX® delta Ceramic Liner E1 STD Liner E1 HW Liner E1 10˚ Liner
CoCr Modular HeadBIOLOX® delta Ceramic Head
E1 Flanged Cemented Cup E1 Non Flanged Cemented Cup E1 Muller Cemented Cup
Exceed ABT Standard TaperFit Shell
Exceed ABT 15° TaperFit Shell
Exceed ABT RingLoc-X Shell
Exceed ABT RingLoc-X Multi-Hole Shell
E1 Tapered STD Liner
BIOLOX® delta Option
30
Notes
Notes
Legal ManufacturerBiomet UK, Ltd.Waterton Industrial EstateBridgend, South WalesCF31 3XA UK
www.biometeurope.com
0086
©2014 Biomet Orthopedics • Form No. BMET0326.3-ENG • REV1214
All content herein is protected by copyright, trademarks and other intellectual property rights owned by or licensed to Biomet Inc. or its affiliates unless otherwise indicated, and must not be redistributed, duplicated or disclosed, in whole or in part, without the express written consent of Biomet.
This material is intended for health care professionals and the Biomet sales force. Distribution to any other recipient is prohibited.
For complete product information, including indications, contraindications, warnings, precautions, and potential adverse effects, see the package insert and www.Biomet.com” where appropriate.
BIOLOX® delta and BIOLOX® delta Option are trademarks of CeramTec AG
For product information see package insert and Biomet’s website. Check for local product clearances and reference product specific instructions for use.
Note: Biomet UK may not be the legal manufacturer of all instruments listed in this surgical technique. Refer to the instrument labeling for Legal Manufacturers name and address.