exchange principles - stanford universityesi.stanford.edu/respiration/cclabmanual.pdf6 logged on at...
TRANSCRIPT
1
Lab # 1
Exchange principles con- and counter-current systems
Michael Axelsson University of Gothenburg
Department of Zoophysiology
2
INTRODUCTION
Life depends on the exchange of mass and energy so a constant and
permanent transfer of molecules such as gases (oxygen and carbon dioxide), ions
(protons, cations like sodium, potassium and calcium or anions like chloride and
bicarbonate) and energy (heat) is taking place between organisms and their
environment.
Mass transfer occurs by molecular diffusion while heat transfer occurs by
conduction, convection and radiation.
The flow of either mass or heat depends on the properties of the media
where exchange is taking place, the difference in temperature or concentration
and the distance. Because of this, mass and thermal exchange between an animal
and its environment will always occur. Such ubiquity constitutes an important
selection pressure towards the appearance of specialized exchange structures
designed to avoid or improve the exchange of heat or molecules with the
environment.
The following are examples of exchange structures designed to avoid loss of heat
or molecules:
• Vascular design of blood flow in the appendages of aquatic animals or animals
living in cold climates to avoid heat loss
• Vasa recta in kidneys to supply nutrients to the inner medulla without removing
the osmotic gradient
• Vascular heat traps that allow the maintenance of specific organ systems at
temperatures above the temperature of the rest of the body of the animal
• Airway structure in desert animals to avoid excessive evaporative water loss
The following are examples of exchange structures designed to facilitate heat or
mass transfer:
• Salt glands in birds and reptiles that aid in the excretion of excess salts (mostly
NaCl)
• Rete mirabile in fish aided to filling the swim bladder with oxygen at high
pressures
• Fish gills that uptake oxygen from water and download carbon dioxide
3
• Brain cooling mechanisms associated with respiratory passages in some birds
and mammals
• Extensively vascularized and expanded respiratory epithelia in tetrapods to
allow the exchange of respiratory gases
With the exception of the last example, all others are based on a common
scheme that has appeared recurrently in the evolution of many animal groups: the
countercurrent exchanger.
The purpose of this laboratory exercise is to study the basic function of the
countercurrent exchanger and to compare it with its alternative arrangement: the
concurrent exchanger. The system used for this lab is an obvious simplification of
the anatomical arrangements encountered in nature and it is based on two copper
tubes soldered together along its entire length (1 m). When the water inside the
tubes flows in opposite directions, we talk about countercurrent exchange. When
water flows in the same direction we talk about concurrent exchange.
DESCRIPTION OF THE SYSTEM The actual system is located in the Department of Zoology at the University
of Göteborg but it can be controlled remotely from any location in the world. A
camera placed over the system allows direct visualization of the actual lab setting.
Below is a schematic drawing of the setup with numbers that will be referred
to in the text. Thermostated baths (1) are used to simulate two different
4
temperatures; in this case one bath is set to 5°C and the other to 37°C. Two
peristaltic pumps (2) are used to pump water from the two baths through the heat-
exchanger model (3, see below for more details). The temperature in the
exchanger is measured at 4 locations along each of the two tubes in the heat
exchanger model (at 4 cm, 36 cm, 65 cm and 96 cm respectively) plus at the very
entrance at three locations. The temperature is measured by an interface box (4,
PICO TC08 thermocouple interface) and the information is entered into a
computer (6). The computer controls the two pumps via a control box (5) and also
sends an alarm signal to an external alarm box (7) every time a person logs on
remotely.
The computer records the temperature along the exchanger and controls the
two peristaltic pumps (see below for more information about the program).
OPERATION OF THE SYSTEM If you are going to run the lab from a remote location (any location outside
the lab were the equipment is setup) you need a computer with access to the
Internet and a minimum connection speed of 126 kbits/second (ISDN or higher).
Log on to the following address
http://vivaldi.zool.gu.se/Labview_WWW/Index.htm.
A window looking like
the picture to the right will
show. On this page there
are links to different things
that you will need during
the lab. “Heat exchanger” is the link to the program
that runs the lab. If you
5
click on this link another window will open. If this is the first time you log on it will
take some time before the program-window is shown (up to several minutes). The
first time you log on the program automatically downloads two files from National
Instruments, first the run-time module that let you run the program through the web
browser, the second is a plug-in that let your web browser show the program.
During this process a window will be shown asking if you would like to install the
run-time module, just click the “Yes” button. When the run-time module and plug-in
are installed the program will be shown (as in the figure below) and you can see
the temperature in the heat exchanger.
In the window that opened (and also in the initial window), there is a link to a
web-camera that will allow you to see the lab. You can control the camera via you
web browser. The are several preset positions that you can use to see different
areas of the lab, and you can also
use the PAN, TILT and ZOOM
functions to navigate the lab
yourself.
There is also a link to a chat
that will help you to communicate
with other students that will be
6
logged on at the same time (if you are doing the lab alone or if all people in the
lab-group is present in front of the same computer you will not need this). When
you activate the chat you will be asked for a login name and then the window with
the chat will open.
OPERATION OF THE PROGRAM To run the lab you need control of the setup, one person at the time can be in
control and to gain control you “right click” on the program-window and select
“Request control of VI” (VI stands for Virtual Instrument) in the window that pops
up. You will then be granted control unless somebody else has control of the
program at the moment. If so, you will be placed in a queue. When you have
control you can use the mouse to click on the various control buttons (see
description of the program below).
In the top part of the program, information about the pumps are shown, in the
case below both pumps are on and pumping and the direction of warm and cold
water through the exchanger will be indicated by arrows (blue for the cold and red
for the warm water). In the middle the temperature data are shown. The thick
lines/dots indicate the temperature at each measuring point along the heat
exchanger while the thin dotted lines indicate the temperature of the water
entering the heat exchanger.
There are four
different tabs that
you can select,
“Time diagram”,
“Temperature diagram” “Pump settings” and
“Logg”, this will be
explained later. In
the lower part there
is one button “Save data”, it is used to
7
start or stop saving data to the hard disk. At the right information about the time to
the next measurement is shown; this is set to 5 seconds between each measuring
point.
In the next tab
“Temperature diagram” the
temperature data
is shown inside 4
circles that
represent the 4
measuring points
along each tube
and the flow
direction is
indicated by the
large arrows or in
the case when one or both pumps are off a white area and the text “Pump X is off”
The temperature of the water going in to the heat exchanger is also shown outside
of the
corresponding
arrow.
When you
click on the button
“Pump settings” a
window will show
with the controls
for the pumps as
shown in the figure
to the right. You
can turn on and off
the pumps
individually,
change the speed of each pump and change the flow set-up between counter
current and con current. The speed of the pumps can be set to 1, 2 or 3
8
corresponding to low medium and high flow. Once you finished setting the pumps
you close the window by clicking “Close window”. Below is s schematic drawing
showing how the
water (cold in blue and
warm in red) is
pumped through the
exchanger during
counter-current and
con-current setup The
fourth tab “Log”
shows a list of the
systems that are logged on to the computer. Every time someone logs in an alarm
in the lab is set of and a green small windows saying “New user logged on”
shows for 5 seconds
in the lower white
field next to the
“Save data” button“.
In the “Log” tab
there is also a button
called “Get attention” that when
pressed sounds an
audio/visual alarm in
the lab to get
attention from the
person that is locally
responsible for the
lab. This alarm is to inform the person responsible for the lab that somebody has
logged in so that we can make sure that everything is working in the lab and we
can also join the chat to help you with any problem.
You can then follow the development of the temperature either by looking at
the “Time diagram” or “Temperature diagram”.
9
Saving data
When the temperature readings have stabilized i.e. there is not further changes
you should save a short period of data that will be used for the calculations
afterwards. To start saving data to the hard disk click on the “Save data” button
now every 5 seconds one set of temperature data will be saved. After 1-2 minutes
you can stop saving data by clicking on the same button that now reads “Stop
saving”.
You have to download the data to you local hard disk (or floppy) for later
processing and to do this you “right click” on the diskette symbol to the left of the
program and select “Save Target As” in the pop-up window, this will open up a
window and you can now select were on your local machine you want to save the
file and what you would like to call it. By default this file is called DATAFILE.TXT
and you should rename it (give it a name that indicate what type of experiment you
were running and speed of pumps and so on)
The data files that you have saved are plain ASCII files and can be open in Excel
or any other program that can open ASCII (text) files. To open the files in Excel go
to “Files” > “Open” and then to the sub directory were you have saved the file and
then in the tab “File of types” select “All files *.*”, this is important otherwise
Excel only looks for its own files with the extensions “.exl”.
10
DESCRIPTION OF THE LAB EXERCISE When reading about countercurrent exchangers in textbooks the first concept
that usually appears is that countercurrent exchangers are more efficient than
concurrent exchangers.
Think about this and come up with a way to quantify the exchanger
efficiency. Efficiency is usually expressed as a percentage. Try to calculate the
efficiency using the temperature data you obtain after running the heat exchanger
in the countercurrent and the concurrent modes.
When running the heat exchanger, wait until the temperature values stabilize.
You can keep track of this using the Time diagram. Once the temperatures
stabilize, write the numbers down in your notebook, plot them in an Excel graph
and compare both exchange modes. Or use the Save data function and download
the data file to your computer
Now that you have done this exercise, try to answer the following questions:
1. Did you achieve a 100% efficient exchange in any mode?
2. Can countercurrent exchange have an efficiency of 100%? How?
3. Can concurrent exchange have a 100% efficiency? How?
Flow through the exchanger tubes can be changed as specified in the
previous section. Do you think that flow can affect efficiency? Try to make a
prediction before running the exchanger at lower or higher flows and verify if you
prediction is right.
Describe how efficiency varies with alterations of flow. Plot your data in
graphs to demonstrate your results and discuss the possibility that efficiency in
counter current exchangers is similar to the efficiency in concurrent exchangers.
Let’s consider a real scenario, the fish gill
11
The two panels below shows one single gill lamella inserted between to Y-axis
with oxygen partial pressure. The task is to draw the lines for the partial pressure
of oxygen in the water and in the blood in the two different situations (counter-
current to the left and con-current to the right). The water and venous blood
oxygen tension is given in the top of the figure.
12
References
Illustration for this manual is from the following sources, some have been redrawn
while some have just been scanned.
• Animal Physiology, Mechanisms and adaptions, R. Eckert & D. Randall,
W.H. Freeman and Company, San Fransisco, ISBN 0 7167 1423 X
• Biology of fishes.2 nd edition. Q Bone, NB Marshall and JHS Blaxter. ISBN
0 412 741140 7
• Comparative Animal physiology, P. C. Withers, Suanders College
Publishing, Forth Worth, Philadelphia, ISBN 0-03-012847-1