![Page 1: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/1.jpg)
HOSPI TALS
Experience with Rubidium-82 PET Myocardial Perfusion Scintigraphy in
Clinical Routine and Research
Dr John C. Dickson, Medical Physicist
Institute of Nuclear Medicine, University College London Hospitals and University College London
United Kingdom
![Page 2: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/2.jpg)
HOSPI TALS
Myocardial Perfusion Imaging
![Page 3: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/3.jpg)
HOSPI TALS
Myocardial Perfusion Imaging
MRI
SPECT
• Thallium
• Tc-99m Sestamibi, Tetrafosmin
PET
• O-15 Water, N-13 Ammonia
• Rb-82 Chloride
![Page 4: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/4.jpg)
HOSPI TALS
Why perform Rubidium MPI?
• The patient– Stress and Rest Imaging in less than 30 minutes
– Reduced Radiation Burden
• The technologist
• The scientist– Potential for Quantitative Flow Measurements
• The clinician– Quick and complete picture of gross and detailed flow
– Potential to combine with simultaneous CT investigations
– Rapid Access
• The accountant– Cost effective for high throughput centres
![Page 5: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/5.jpg)
HOSPI TALS
![Page 6: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/6.jpg)
HOSPI TALS
Rubidium Chloride
• Monovalent potassium analog
• Partially extracted by myocardial cells via Na/K adenosine triphosphatase pump
• Extraction is inversely and non-linearly proportional to perfusion
![Page 7: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/7.jpg)
HOSPI TALS
Production of Rubidium-82
• Rubidium-82 produced from Strontium-82 generator– Rb-82 T1/2 75 Seconds
– Sr-82 T1/2 25.55 Days
• Rubidium given as an i.v. infusion– 60 mCi maximum (2D Mode)
– 35 seconds maximum
– Approximately 30 mls
– 3D Mode give 40mCi
• Generator lifetime 4 weeks
![Page 8: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/8.jpg)
HOSPI TALS
Production of Rubidium-82
![Page 9: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/9.jpg)
HOSPI TALS
Production of Rubidium-82
![Page 10: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/10.jpg)
HOSPI TALS
Production of Rubidium-82
• QC of product
– Column Wash
– Breakthrough Test (Sr82/Sr85)
– Infusion System Calibration
![Page 11: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/11.jpg)
HOSPI TALS
Production of Rubidium-82
Interference with PET QA
0
2
4
6
8
10
12
14
0 50 100 150 200 250 300 350 400
Time after start of infusion (sec)
Detected
Cou
nts (
kcp
s)
![Page 12: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/12.jpg)
HOSPI TALS
Rubidium Imaging
![Page 13: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/13.jpg)
HOSPI TALS
Basic Imaging Protocol
Stress CTAC
Stress PET
Adenosine
Stress
Rest PET
Rest CTAC
Scout
(Topogram)
![Page 14: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/14.jpg)
HOSPI TALS
Rubidium PETCT Protocol - CT
• Scout
– AP view
– 120 kVp, 10 mA
• CT
– 140 kVp, 50 mA, 0.8s Pitch 1.75
– Assessment of calcification and AC
• Often need two CT’s for attenuation correction
– Can capture heart end-systoleStress CTAC
Stress PET
Adenosine
Stress
Rest PET
Rest CTAC
Scout
(Topogram)
![Page 15: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/15.jpg)
HOSPI TALS
The need for Attenuation Correction
![Page 16: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/16.jpg)
HOSPI TALS
The mis-registration problem
![Page 17: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/17.jpg)
HOSPI TALS
Misregistration of Emission and CTAC
Software related reduction techniques
![Page 18: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/18.jpg)
HOSPI TALS
Misregistration of Emission and CTAC
Protocol related reduction techniques• Cine CT (Pan et al, Med Phys 2006)
– Series of CTs performed over 6 seconds
– Low mA - CT AC use only (5mGy)
![Page 19: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/19.jpg)
HOSPI TALS
Rubidium PETCT Protocol - PET
• Rb82 infusion - 60 mCi (or no more than 35 seconds)
• Data acquired 2D in list mode (BGO)
• Acquisition– Delay before imaging (2:30 stress, 2:40 rest)
– 5 minute (8 bin) gated data for both stress and rest
– Additionally dynamic data for flow estimation
• Reconstruction OSEM
• Display GE Myometrix– QGS difficulties with base
Stress CTAC
Stress PET
Adenosine
Stress
Rest PET
Rest CTAC
Scout
(Topogram)
![Page 20: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/20.jpg)
HOSPI TALS
Rubidium PETCT Protocol - PET
0
1
2
3
4
5
6
0 10 20 30 40 50 60
Time (sec)
mC
i / s
ec
Week 1
Week 4
50.95 mCi
60.04 mCi
• Rb82 infusion - 60 mCi (or no more than35 seconds)
• At end of generator, delivery of 60mCiresults in too much decay of deliveredtracer
• Delay before imaging (2:30 stress,2:40 rest)
• To start earlier adds more bloodpoolcounts than myocardial counts
Decay Corrected
![Page 21: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/21.jpg)
HOSPI TALS
Processing Rb-82 data
• Base plane difficult to define
• Standard QGS, ECToolbox, 4D-MSPECT make assumptions on data reconstruction
• Better solutions
– QPET
– GE CardIQPhysio (Adw)
– GE Myometrix (Xel)
![Page 22: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/22.jpg)
HOSPI TALS
ECG Gating
• ECG Gated Bins
– SPECT (16 bins)
– PET (8 bins)
• Better temporal resolution→ More accurate LVEFs ECG Bins Segmentation
Failure (%)
6 0
8 5
10 35
12 35
16 45
6 bin
LVEF: 58%16 bin
LVEF: 70%
• Rubidium data – noisy
• To calculate LVEF→ Segment Myocardium
![Page 23: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/23.jpg)
HOSPI TALS
Rubidium PETCT ProtocolAdenosine Stress Test
• 6 minute stress test
• Second iv access for Adenosine
• Rubidium infusion starts two minutes into Adenosine
• We don’t use Dobutamine with Rubidium imagingStress CTAC
Stress PET
Adenosine
Stress
Rest PET
Rest CTAC
Scout
(Topogram)
![Page 24: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/24.jpg)
HOSPI TALS
The result
Stress CTAC
Stress PET
Adenosine
Stress
Rest PET
Rest CTAC
Scout
(Topogram)
![Page 25: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/25.jpg)
HOSPI TALS
Advanced Imaging Protocols
Stress CTAC
Stress PET
Adenosine
Stress
Rest PET
Rest Ca Score
(CTAC)
Scout
(Topogram)
Ca Score
+
MPI
Rest CT CA
Rest CTAC
Rest PET
Stress PET
Adenosine
Stress
Stress CTAC
Scout
(Topogram)
MPI
+
CT CA
Rest Ca Score
(CTAC)
Rest PET
Stress PET
Adenosine
Stress
Stress CTAC
Scout
(Topogram)
Ca Score
+
MPI
+
CT CA
![Page 26: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/26.jpg)
HOSPI TALS
Rubidium MPI + CT CA
Rest CT CA
Rest CTAC
Rest PET
Stress PET
Adenosine
Stress
Stress CTAC
Scout
(Topogram)
![Page 27: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/27.jpg)
HOSPI TALS
Radiation Protection Issues• Patient Dose
– PET – 5.5 mSv (2 x 60 mCi)
• Tl-201 – 14 mSv (80 MBq)
• Tc-99m Tetrafosmin – 12 mSv (2 x 800 MBq)
• Tc-99m Sestamibi – 16 mSv (2 x 800 MBq)
– CT CA - 6-12 mSv
• Newer technology allowing lower (sub 5mSv) doses
• Staff Dose
– Less than that seen with Tc-99m MPI
– Patient managed remotely from control room
![Page 28: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/28.jpg)
HOSPI TALS
Quantitative myocardial perfusion imaging (Q MPI)
![Page 29: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/29.jpg)
HOSPI TALS
What do we mean by Q MPI
![Page 30: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/30.jpg)
HOSPI TALS
Why perform Quantitative MPI?
• Triple vessel disease
– Standard imaging compares relative flow
– Balanced disease may appear normal
• Microcirculatary disease
– Can’t visualise microcirculature
– Use Myocardial Blood Flow (MBF) or Coronary Flow Reserve (CFR) as an indicator
MBFRestMBFStressCFR
![Page 31: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/31.jpg)
HOSPI TALS
Why perform Quantitative MPI?
• Detection of early disease
• Disease progression
– Treatment, or lifestyle changes e.g. Smoking, diet
• Evaluation of endothelial function
– Regulatory mechanism for changes in flow
![Page 32: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/32.jpg)
HOSPI TALS
![Page 33: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/33.jpg)
HOSPI TALS
![Page 34: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/34.jpg)
HOSPI TALS
Models
![Page 35: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/35.jpg)
HOSPI TALS
Rubidium Chloride
• Monovalent potassium analog
• Partially extracted by myocardial cells via Na/K adenosine triphosphatase pump
• Extraction is inversely and non-linearly proportional to perfusion
![Page 36: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/36.jpg)
HOSPI TALS
Three Compartment Model
PScap – Permeability surface products of capillaries PScell – Permeability surface product of cell wallsVi – Fractional volumes of compartmentsF – specific volume blood flow
Physiological Model (almost)[heterogeneous flow rates, variable capillary level, axial diffusion]
![Page 37: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/37.jpg)
HOSPI TALS
Two Compartment Model
Plasma and capillary compartments are lumpedAssumption: They both equilibrate rapidly
Interstitial
Space
Intercel.
Space
![Page 38: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/38.jpg)
HOSPI TALS
One Compartment Model
Uptake and washout parameters only
![Page 39: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/39.jpg)
HOSPI TALS
Which Compartmental Model?
• Rubidium PET data is noisy
• Three and two compartment models difficult to implement– Wavelet based noise reduction can help
– Needs a priori values of some model parameters
• One compartment model– Good reproducibility at rest and stress
– Better fit than two compartment model
• Two and one compartment model– K1 dependent on both blood flow and extraction
– Requires non-linear extraction correction
![Page 40: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/40.jpg)
HOSPI TALS
Non-linear extraction model
Lortie M 2007:
EJNMMI
![Page 41: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/41.jpg)
HOSPI TALS
Alternatives
Factor Analysis (GFADS)(El-Fakhiri et al 2005)
• GFADS used to estimate LV and RV TACs
– Estimate k1 and k2
• Accounts spillover from LV and RV blood
• More robust to noise
Retention Model(Herrero et al 1990)
• Division of stable uptake in myocardium by integrated blood input gives k1
• Derive flow from k1
• Assumes tracer doesn’t leave myocardium during acquisition
![Page 42: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/42.jpg)
HOSPI TALS
Technical Issues
![Page 43: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/43.jpg)
HOSPI TALS
High Count Rates
• Randoms, Deadtime & Decay Correction
– Very high count rates (>3 Mcps) at start of dynamic acquisition
– Short half-life with relatively long frames can cause decay correction problems
![Page 44: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/44.jpg)
Spillover and partial volume effects
• Rb-82– Positron energy 3.4 MeV– Average Range 5.9 mm
• Myocardial Wall– 8-12 mm thickness
![Page 45: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/45.jpg)
HOSPI TALS
Software
![Page 46: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/46.jpg)
Imagen Q (CVIT, Kansas: Bateman et al)
![Page 47: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/47.jpg)
Carimas (Turku PET Centre, Knuuti et al)
![Page 48: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/48.jpg)
![Page 49: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/49.jpg)
HOSPI TALS
Case Studies
![Page 50: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/50.jpg)
HOSPI TALS
![Page 51: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/51.jpg)
HOSPI TALS
![Page 52: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/52.jpg)
HOSPI TALS
![Page 53: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/53.jpg)
HOSPI TALS
![Page 54: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/54.jpg)
HOSPI TALS
![Page 55: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/55.jpg)
HOSPI TALS
Indications
• 2 previous PCIs to Cx and LAD.
• New onset chest pain ?cause.
• Not exertional but previous episodes have been atypical.
![Page 56: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/56.jpg)
HOSPI TALS
PET findings
• There is no inducible ischaemia during adenosine vasodilatation.
• There has been good revascularization.
• The proximal LAD disease is not flow limiting.
![Page 57: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/57.jpg)
HOSPI TALS
CT Findings
![Page 58: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/58.jpg)
HOSPI TALS
CT findings
• LEFT MAIN STEM: Normal origin and course. The LMS is plaque free.
• LEFT ANTERIOR DESCENDING: There is diffuse heterogenous plaque of the proximal LAD, causing mild to moderate (up to 50%) stenosis. The mid-LAD stent is patent. The distal LAD is plaque free, albeit small calibre. The 1st & 2nd DIAGONALs are plaque free.
• LEFT CIRCUMFLEX: The proximal LCx is plaque free. The stent in the AV segment is patent, and takes an OBTUSE MARGINAL course. The distal OM is plaque free. The distal LCx is plaque free, albeit small calibre.
• RIGHT CORONARY: Normal origin and course. There is a small eccentric calcified plaque in mid-RCA, not causing any stenosis. The RCA is dominant, supplying the PDA which is plaque free
![Page 59: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/59.jpg)
HOSPI TALS
Flow Findings
![Page 60: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/60.jpg)
HOSPI TALS
Summary
• MPI with Rubidium PET/CT:
– is fast and effective (Stress and Rest < 30 min)
– allows simple add-ons for CT CA, or Ca Score
– Can provide true quantitative data to assess balanced disease, early disease, and/or compromised micro-circulature and endothelial function
![Page 61: Experience with Rubidium-82 PET Myocardial Perfusion ... · PDF fileMyocardial Perfusion Scintigraphy in Clinical Routine and Research ... •Display GE Myometrix ... compromised micro-circulature](https://reader031.vdocument.in/reader031/viewer/2022030410/5a9822b37f8b9aba4a8d1344/html5/thumbnails/61.jpg)