External Defibrillators: Recalls, Inspections, and the Quality System Regulation Melissa Torres Office of Compliance December 15, 2010

Upload: cory-green

Post on 23-Dec-2015

217 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: External Defibrillators: Recalls, Inspections, and the Quality System Regulation Melissa Torres Office of Compliance December 15, 2010

External Defibrillators:Recalls, Inspections, and the Quality

System Regulation

Melissa TorresOffice of Compliance

December 15, 2010

Page 2: External Defibrillators: Recalls, Inspections, and the Quality System Regulation Melissa Torres Office of Compliance December 15, 2010

2

Overview

• Recalls – Definitions– Analysis of external defibrillator recalls

• Common Quality System (QS) regulation deficiencies identified in recalls and inspections.– Purchasing Controls and Acceptance Activities– Design Controls– Corrective and Preventive Action (CAPA)

Page 3: External Defibrillators: Recalls, Inspections, and the Quality System Regulation Melissa Torres Office of Compliance December 15, 2010

3

What is a Recall?• Voluntary action taken by a firm to address a

problem with a device that violates FDA law, such as:– returning devices to recalling firm– performing a field repair or modification– modifying device labeling – notifying patients of a problem– advising physicians to monitor patients for health issues

• Effective method to remove or correct consumer products from the market place.

Page 4: External Defibrillators: Recalls, Inspections, and the Quality System Regulation Melissa Torres Office of Compliance December 15, 2010

4

How to Determine if you have a Medical Device Recall

• Removal– Physical confiscation from where the device is

used or sold, to some other location for repair, modification, adjustment, relabeling, destruction, or inspection.

– A removal is not part of regularly scheduled maintenance!

• Correction– On site activities which can include repair,

modification, adjustment, relabeling, destruction, inspection, or patient monitoring.

Page 5: External Defibrillators: Recalls, Inspections, and the Quality System Regulation Melissa Torres Office of Compliance December 15, 2010

5

Recall Analysis

• 70 voluntary recalls conducted by manufacturers of external defibrillators from January 2005 – November 2010.– 17 were classified as Class I – 50 were classified as Class II– 1 was classified as Class III– 2 were safety alerts

• FDA has seen an increase in the amount of recalls.

*Recall analysis only includes product code MKJ – Automated External Defibrillator.

Page 6: External Defibrillators: Recalls, Inspections, and the Quality System Regulation Melissa Torres Office of Compliance December 15, 2010

6

External Defibrillator Recall TrendsNumber of Recalls per Year

1/2005 – 11/2010

11

1717

97

9

0

24

68

1012

1416

18

2005 2006 2007 2008 2009 2010*

Year

Nu

mb

er o

f R

ecal

ls

* 2010 data only includes recalls classified between Jan.- Nov. 2010

Page 7: External Defibrillators: Recalls, Inspections, and the Quality System Regulation Melissa Torres Office of Compliance December 15, 2010

7

Quality System (QS) Regulation 21 CFR 820

• Provides framework of basic requirements for medical device manufacturers to follow to ensure that devices are adequately designed and manufactured.

• Manufacturers must develop a Quality System (QS) commensurate with the:– Risk presented by the device;– Extent of the activities to be carried out;– Complexity of the device and manufacturing

processes; and – Size and complexity of manufacturing facility.

Page 8: External Defibrillators: Recalls, Inspections, and the Quality System Regulation Melissa Torres Office of Compliance December 15, 2010

8

What are Purchasing Controls?21 CFR 820.50

• Requirements for manufacturers to ensure that purchased or received product and services from suppliers, contractors, and consultants conform to specified requirements.– Evaluation and selection of suppliers– Control over suppliers– Records of acceptable suppliers– Purchasing data

Page 9: External Defibrillators: Recalls, Inspections, and the Quality System Regulation Melissa Torres Office of Compliance December 15, 2010

9

What are Acceptance Activities?21 CFR 820.80

• Requirements for manufacturers to ensure that received, in-process, and finished devices are tested and accepted based on established requirements and specifications. – Receiving– In-process– Final

Page 10: External Defibrillators: Recalls, Inspections, and the Quality System Regulation Melissa Torres Office of Compliance December 15, 2010

10

Purchasing Controls and the Link to Acceptance Activities

Manufacturers must establish an

appropriate mix of assessment and

receiving acceptance to ensure products

and services are acceptable for their

intended uses...

... The extent of incoming acceptance activities can be based, in part, on the degree to which the supplier has demonstrated a capability to provide quality products or services.Preamble to the 1996 QS Regulation, Comments #99 and #106

Page 11: External Defibrillators: Recalls, Inspections, and the Quality System Regulation Melissa Torres Office of Compliance December 15, 2010

11

What is Unique About Medical Devices?

• Wide range in type of supplied products and services.– Raw materials, components, software,

laboratories, sterilizers, etc.

• Wide range in complexity in supplied products.– From components to finished devices.

• Wide range in risk associated with supplied products and services.– Same supplied product or service may have

different risks based on use.– Same supplier may have different risks for

different supplied product or service.

Page 12: External Defibrillators: Recalls, Inspections, and the Quality System Regulation Melissa Torres Office of Compliance December 15, 2010

12

Why is FDA Concerned about Purchasing Controls?

• There is an increased importance of purchasing controls for both quality attributes as well as product specifications.– Increase in outsourcing of critical components

and manufacturing of entire devices.– Increasing recalls and product problems

associated with purchased components and devices.

Page 13: External Defibrillators: Recalls, Inspections, and the Quality System Regulation Melissa Torres Office of Compliance December 15, 2010

13

Purchasing Controls: Most Commonly Identified Deficiencies

• Inappropriate evaluation of suppliers. • Changes to the components without

notification to the manufacturer.• Problems with the manufacturing of a

component/subassembly. • Inadequate incoming acceptance activities

to detect nonconformities.• Inadequate or lack of process validation

activities at the supplier.

Page 14: External Defibrillators: Recalls, Inspections, and the Quality System Regulation Melissa Torres Office of Compliance December 15, 2010

14

What are Design Controls?21 CFR 820.30

• Requirements for manufacturers to control the design and development process through procedures and processes to assure that devices meet user needs, intended uses, and specified requirements.• Design and

Development Planning • Design Input• Design Output • Design Review

• Design Verification• Design Validation • Design Transfer • Design Changes • Design History File

Page 15: External Defibrillators: Recalls, Inspections, and the Quality System Regulation Melissa Torres Office of Compliance December 15, 2010

15

Design Controls

• It is up to a firm to decide what design tools and practices they will employ.

• The QS regulation focuses on the general procedures and processes used to design the devices.

Page 16: External Defibrillators: Recalls, Inspections, and the Quality System Regulation Melissa Torres Office of Compliance December 15, 2010

16

Design Controls: Most Commonly Identified Deficiencies

• Problems with the initial design of the device.

• Inadequate verification/validation of design changes (including changes in the actual design, process, testing, labeling, etc.) prior to implementation.

• Inadequate software validation.

• Inadequate risk analysis/management.

Page 17: External Defibrillators: Recalls, Inspections, and the Quality System Regulation Melissa Torres Office of Compliance December 15, 2010

17

What is Corrective and Preventive Action? 21 CFR 820.100

• Requirements for manufacturers to – Collect and analyze information.

– Identify and investigate existing

and potential product and quality

problems.

– Take appropriate, effective, and comprehensive corrective and/or

preventive actions.

Page 18: External Defibrillators: Recalls, Inspections, and the Quality System Regulation Melissa Torres Office of Compliance December 15, 2010

18

Corrective and Preventive Action: Most Commonly Identified Deficiencies

• “Fix on Fail” - Fixing individual devices through servicing activities as they fail rather than correcting a group of known defective devices.

• Inadequate analysis of data sources including evaluation of servicing activities within the CAPA system to identify potential trends that may be a recall situation.

• Inadequate investigations to determine the cause of nonconformities.

Page 19: External Defibrillators: Recalls, Inspections, and the Quality System Regulation Melissa Torres Office of Compliance December 15, 2010

19

Servicing Activities

• What is considered by a manufacturer to be routine servicing and preventive maintenance must be defined up front in the design control process.

• If the servicing and maintenance being performed are not defined and documented as expected actions, then it is considered to be a correction to the device.

• Service records should be analyzed in the CAPA system. [21 CFR 820.100(a)(1)]

Page 20: External Defibrillators: Recalls, Inspections, and the Quality System Regulation Melissa Torres Office of Compliance December 15, 2010

20

References

• 21 CFR Part 820– http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfcfr/CFRsear

ch.cfm?CFRPart=820

• Preamble to the QS Regulation Final Rule– http://www.fda.gov/downloads/MedicalDevices/

DeviceRegulationandGuidance/PostmarketRequirements/QualitySystemsRegulations/MedicalDeviceQualitySystemsManual/UCM122806.pdf

• CDRH Learn– http://www.fda.gov/Training/CDRHLearn/default.htm

• Device Advice– http://www.fda.gov/MedicalDevices/

DeviceRegulationandGuidance/default.htm

Page 21: External Defibrillators: Recalls, Inspections, and the Quality System Regulation Melissa Torres Office of Compliance December 15, 2010

21

References, cont.

• GHTF Quality Management System – Medical Devices – Guidance on the Control of Products and Services Obtained from Suppliers – http://www.ghtf.org/documents/sg3/sg3final-N17.pdf

• GHTF Implementation of Risk Management Principles and Activities Within a Quality Management System– http://www.ghtf.org/documents/sg3/sg3n15r82005.pdf

• GHTF Quality management system - Medical Devices - Guidance on corrective action and preventive action and related QMS processes - Draft– http://www.ghtf.org/sg3/sg3-proposed.html

Page 22: External Defibrillators: Recalls, Inspections, and the Quality System Regulation Melissa Torres Office of Compliance December 15, 2010

22

Thank You!