fabrication and testing of kgmt fsm prototype
DESCRIPTION
Fabrication and testing of KGMT FSM prototype. Ho-Soon Yang, Hak-Yong Kihm, Il-Kwon Moon, Jae-Bong Song, Yun-Woo Lee Korea Research Institute of Standards and Science, Korea Young-Soo Kim Korea Astronomy & Space Science Institute, Korea. Oct. 4. 2010. Contents. Objective of project - PowerPoint PPT PresentationTRANSCRIPT
![Page 1: Fabrication and testing of KGMT FSM prototype](https://reader036.vdocument.in/reader036/viewer/2022081511/56814c62550346895db98703/html5/thumbnails/1.jpg)
K-D-PR-1100-009
Fabrication and testing of KGMT Fabrication and testing of KGMT FSM prototypeFSM prototype
Oct. 4. 2010
Ho-Soon Yang, Hak-Yong Kihm, Il-Kwon Moon, Jae-Bong Song, Yun-Woo Lee
Korea Research Institute of Standards and Science, Korea
Young-Soo KimKorea Astronomy & Space Science Institute, Korea
![Page 2: Fabrication and testing of KGMT FSM prototype](https://reader036.vdocument.in/reader036/viewer/2022081511/56814c62550346895db98703/html5/thumbnails/2.jpg)
Contents
Objective of project Recent experiences in KRISS Manufacturing flow diagram Optical testing Working cell design Future works
2
![Page 3: Fabrication and testing of KGMT FSM prototype](https://reader036.vdocument.in/reader036/viewer/2022081511/56814c62550346895db98703/html5/thumbnails/3.jpg)
Objective
3
Technology development related to manufacturing of FSM for
GMT project in advance
For FSMP
Use a model very close to FSM FSM prototype in KGMT project
Fabrication of FSMP with its own working cell designed by KRISS
Optical testing of FSMP after assembly with a cell (very close to real
FSM cell)
![Page 4: Fabrication and testing of KGMT FSM prototype](https://reader036.vdocument.in/reader036/viewer/2022081511/56814c62550346895db98703/html5/thumbnails/4.jpg)
4
Recent experiences @ KRISS
800 mm lightweighted mirror for space application (1/2)
Gravity
Quilting (30 g/cm2)
And other environmental effects …
![Page 5: Fabrication and testing of KGMT FSM prototype](https://reader036.vdocument.in/reader036/viewer/2022081511/56814c62550346895db98703/html5/thumbnails/5.jpg)
5
Recent experiences @ KRISS
800 mm lightweighted mirror for space application (2/2) - Measurement of wavefront error with two different methods : CGH (computer generated hologram) and conventional null lenses
CGHConventional null lens
/30 rms
![Page 6: Fabrication and testing of KGMT FSM prototype](https://reader036.vdocument.in/reader036/viewer/2022081511/56814c62550346895db98703/html5/thumbnails/6.jpg)
6
1 m off-axis paraboloid for collimator (1/2) - Focal length = 6 m - Distance between the center of OAP and optical axis = 700 mm
Recent experiences @ KRISS
Flat mirror
Interferometer
OAP
![Page 7: Fabrication and testing of KGMT FSM prototype](https://reader036.vdocument.in/reader036/viewer/2022081511/56814c62550346895db98703/html5/thumbnails/7.jpg)
7
1 m off-axis paraboloid for collimator (2/2)
Reference flat Interferometer
OAP mirror
Alignment telescope
Recent experiences @ KRISS
![Page 8: Fabrication and testing of KGMT FSM prototype](https://reader036.vdocument.in/reader036/viewer/2022081511/56814c62550346895db98703/html5/thumbnails/8.jpg)
8
Fabrication facilities in KRISS
Rotating table with aspheric surface under test
Measuring tower with null interferometer
2 m CNC machine
1 m polishing machine
2 m coating chamberAssembly room
0.6 m polishing machine & testing tower
2 m polishing machine
![Page 9: Fabrication and testing of KGMT FSM prototype](https://reader036.vdocument.in/reader036/viewer/2022081511/56814c62550346895db98703/html5/thumbnails/9.jpg)
9
Manufacturing flow diagram
Curve Curve generationgeneration
Curve Curve generationgeneration
Chemical Chemical EtchingEtching
Chemical Chemical EtchingEtching
Assembly Assembly with cellwith cell
Assembly Assembly with cellwith cell FATFATFATFAT
10% larger size 10% larger size Mirror blank Mirror blank procurementprocurement
10% larger size 10% larger size Mirror blank Mirror blank procurementprocurement
LightweightingLightweightingLightweightingLightweighting
PolishingPolishingPolishingPolishing GrindingGrindingGrindingGrinding InspectionInspectionInspectionInspection
InspectionInspectionInspectionInspection
Figuring Figuring (98%)(98%)
Figuring Figuring (98%)(98%)
InspectionInspectionInspectionInspection
Rounding to Rounding to target target
dimension dimension
Rounding to Rounding to target target
dimension dimension
Final Final figuring figuring
Final Final figuring figuring
![Page 10: Fabrication and testing of KGMT FSM prototype](https://reader036.vdocument.in/reader036/viewer/2022081511/56814c62550346895db98703/html5/thumbnails/10.jpg)
10
SpherometerProfilometer
~ mm
Grinding
ProfilometerInterferometer
~ μm
Polishing
Interferometer with CGH
~ nm
Figuring
Scheme of Optical testing
Efficient testing process is a key for successful fabrication of mirror
![Page 11: Fabrication and testing of KGMT FSM prototype](https://reader036.vdocument.in/reader036/viewer/2022081511/56814c62550346895db98703/html5/thumbnails/11.jpg)
11
KRISS profilometer
Measurement accuracy ~ 1 um Compensation of measurement errors of probe system using a high quality reference mirror
Reference mirror
![Page 12: Fabrication and testing of KGMT FSM prototype](https://reader036.vdocument.in/reader036/viewer/2022081511/56814c62550346895db98703/html5/thumbnails/12.jpg)
150 mm
CGH layout :(Line spacing : ~ 20 um)
Adjustment between interferometer and CGH
Test configuration of FSMP off-axis segment (off-axis type CGH )
Interferometric testing using CGH
12
3
4
![Page 13: Fabrication and testing of KGMT FSM prototype](https://reader036.vdocument.in/reader036/viewer/2022081511/56814c62550346895db98703/html5/thumbnails/13.jpg)
Adjustment I between CGH and FSMP segment
Adjustment II between CGH and FSMP segment
CGH alignment
12
3
4
![Page 14: Fabrication and testing of KGMT FSM prototype](https://reader036.vdocument.in/reader036/viewer/2022081511/56814c62550346895db98703/html5/thumbnails/14.jpg)
14
Turn-over tool
KRISS Profilometer
Testing tower setup
![Page 15: Fabrication and testing of KGMT FSM prototype](https://reader036.vdocument.in/reader036/viewer/2022081511/56814c62550346895db98703/html5/thumbnails/15.jpg)
15
Optical testing during fabrication Optical testing after assembly with FSMP cell
Testing configuration
![Page 16: Fabrication and testing of KGMT FSM prototype](https://reader036.vdocument.in/reader036/viewer/2022081511/56814c62550346895db98703/html5/thumbnails/16.jpg)
Concept for working cell design
Minimization of print through Load cell on the back plate of the mirror 24 axial support location Polishing pressure at the pocket of mirror with 0.1 psi
![Page 17: Fabrication and testing of KGMT FSM prototype](https://reader036.vdocument.in/reader036/viewer/2022081511/56814c62550346895db98703/html5/thumbnails/17.jpg)
FE Model of M2
Thin shell elements Material : Zerodur Diameter : 1064 mm Thickness : 140 mm Hole size : 100 mm Mass : 94 kg Face sheet : 20 mm Back plate : 14 mm Rib : 5 mm Web : 10 mm
Top view
bottom view
![Page 18: Fabrication and testing of KGMT FSM prototype](https://reader036.vdocument.in/reader036/viewer/2022081511/56814c62550346895db98703/html5/thumbnails/18.jpg)
Axial support
x
y
support Radius (mm) Force (N)
Ring 1 157.5 21
Ring 2 242.5 33
Ring 3 367.5 42
Ring 4 424.4 58
Support locations
![Page 19: Fabrication and testing of KGMT FSM prototype](https://reader036.vdocument.in/reader036/viewer/2022081511/56814c62550346895db98703/html5/thumbnails/19.jpg)
Print through
Gravity z
Displacement PV 165 nm
Surface error Without correction
PV 166 nmRMS 48 nm
Surface error With correction
PV 50 nmRMS 9 nm
![Page 20: Fabrication and testing of KGMT FSM prototype](https://reader036.vdocument.in/reader036/viewer/2022081511/56814c62550346895db98703/html5/thumbnails/20.jpg)
Print through
Gravity z
10-2
10-1
100
101
102
101
102
103
separation (cm)
rms
wav
efro
nt e
rror
(nm
)
Structure function of M2
M2 structure function
![Page 21: Fabrication and testing of KGMT FSM prototype](https://reader036.vdocument.in/reader036/viewer/2022081511/56814c62550346895db98703/html5/thumbnails/21.jpg)
Surface error
Quilting effect from polishing pressure of 0.1 psi
Surface error Without correction
PV 110 nmRMS 31 nm
Surface error With correction
PV 35 nmRMS 6 nm
Displacement PV 110 nm
![Page 22: Fabrication and testing of KGMT FSM prototype](https://reader036.vdocument.in/reader036/viewer/2022081511/56814c62550346895db98703/html5/thumbnails/22.jpg)
Surface error
Polishing pressure 0.1 psi
10-2
10-1
100
101
102
101
102
103
separation (cm)
rms
wav
efro
nt e
rror
(nm
)
Structure function of M2
M2 structure function
![Page 23: Fabrication and testing of KGMT FSM prototype](https://reader036.vdocument.in/reader036/viewer/2022081511/56814c62550346895db98703/html5/thumbnails/23.jpg)
23
Future works
Manufacturing of two types of CGH Manufacturing of testing tower Optimization of mirror working cell Fabrication of FSMP by the end of next year
We hope to take parts in the fabrication of FSM in future