FDR MS-3500 Service Manual026-201-05(1)E Spec-11
2. Dimensions and Weight of the Machine
External Dimensions and Weight
Width Depth Height Weight
Exposure stand Approx. 624 mm
(24.57 in.)
Approx. 1270 mm (50 in.)
(Base plate approx. 787 mm (30.98 in.))
Approx. 1974 mm 2285 mm
(77.72 in. to 89.96)
Approx. 370 kg
(815.7 lb.)
Control cabinet Approx. 205 mm (8.07 in.)
Approx. 503 mm (19.8 in.)
Approx. 530 mm (20.87 in.)
Approx. 20 kg
(44.09 lb)
Generator (X-ray high voltage unit)
Approx. 445 mm
(17.52 in.)
Approx. 315 mm (12.4 in.)
Approx. 825 mm (32.48 in.)
Approx. 70 kg
(154.3 lb)
Positioner control cabinet Approx. 160 mm (6.3 in.)
Approx. 350 mm (13.78 in.)
Approx. 473 mm(18.62 in.)
Approx. 17 kg
(37.48 lb)
Hardware specification, dimensions and weight are subject to change without notice.
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-12
2.1 External View
Controller unitPositioner control cabinet
(Accessory sold separately)
Generator (X-ray high voltage unit)
Install the controller unit and generator within 5 m (196.9 in.) of the exposure stand. Install the positioner control cabinet within 3 m (118.1 in.) of the exposure stand.
Exposure stand
DRMS1001_0_E.ai
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-13
2.2 Mammographic Stand
DRMS1001_01_1_E.ai
Units: mm (in.)
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-14
2.3 Control Unit
DRMS1001_221_01_E ai
Units: mm (in.)
2.4 Control Pad
DRMS1001_222_E.ai
Units: mm (in.)
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-15
2.5 Foot Switch
DRMS1001_223_E.ai
Units: mm (in.)
2.6 Generator (X-ray High Voltage Device)
825(
32.4
8)
DRMS1001_224_E.ai
Units: mm (in.)
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-16
2.7 AWS: PC and Keyboard
DRMS1001_225_E ai
Units: mm (in.)
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-17
2.8 HUB Unit
DRMS1001_226_E.ai
Units: mm (in.)
The external view and dimensions are just examples, as this is general-purpose electric equipment/component.
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-18
2.9 24x30 Compression Plate (High)
183(
7.20
)
341(13.43)
DRMS1001_227_H_E.ai
Units: mm (in.)Permissible range: ±10%
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-19
2.10 Operation Table
Characteristics - Can store the AWS PC.
DRMS1001_228_E.ai
Units: mm (in.)
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-20
Protective Shield (Optional)
DRMS1001_229_E.ai
Units: mm (in.)
Characteristics Comes with protective glass (equivalent to 0.5 mm Pb), and can be used in the study room. Can store the AWS PC.
600(23.6)
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-21
Protective Plate (Optional), Storage Rack (Accessory Sold Separately)
Characteristics Comes with protective glass (equivalent to 0.5 mm Pb), and can be used in the study room. Can store the AWS PC. Can store each compression plate or accessories during exposure.
600(23.6)
DRMS1001_230_E.ai
Units: mm (in.)
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-22
Protective Glass Cover for Additional Height (0.5 mm Pb Equivalent) (Optional) (Outside Japan)
609(23.98)
310(
12.2
)
23(0.91)
1900
(74.
8) 2140
(84.
25)
DRMS1001_231_E.ai
Units: mm (in.)
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-23
2.11 FDR MS-3500 Options
(1)-1 24x30 compression plate (Low) (1) Compression plate
(1)-2 24x30 compression plate (Shift)
DRMS1001_232_E.ai
Units: mm (in.)
341(13.4)
183(7.20)
341(13.4)
183(7.20)
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-24
(1)-3 24x30 compression plate (SS)
(1)-4 24x30 compression plate (Flex)
314(13.4)
183(7.20)
341(13.4) 313(12.3)
DRMS1001_233_E.ai
Units: mm (in.)
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-25
DRMS1001_234_E.ai
Units: mm (in.)
313(12.3)341(13.4)
215(8.46)
(1)-5 24x30H compression plate (FS)
(1)-6 18x24 compression plate (Low)
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-26
(1)-7 18x24 compression plate (High)
(1)-8 18x24 compression plate (Small)
276(10.9)312(12.3)
183(7.20)
276(10.9)
183(7.20)
311(12.2)
DRMS1001_236_E ai
Units: mm (in.)
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-27
(1)-9 18x24 compression plate (Flex)
(1)-10 18x24L compression plate (FS)
215(8.46)
310(12.2)276(10.9)
276(10.9)
DRMS1001_236_E ai
Units: mm (in.)
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-28
123(4.84)
183(7.20)
313 (12.3) 264 (10.4)
183 (7.20)
(1)-11 Rectangular spot compression plate
(1)-12 Compression plate for axilla radiography
DRMS1001_237_E.ai
Units: mm (in.)
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-29
(1)-13 Compression plate for magnifications
(1)-14 Rectangle spot magnification compression plate
251(9.88)
126(4.96)
311(12.2)
123(4.84)
311(12.2)
DRMS1001_238_E.ai
Units: mm (in.)
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-30
(1)-15 2D biopsy compression plate
183(7.20)
DRMS1001_239_E.ai
Units: mm (in.)
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-31
(2) Magnification table
Magnification table (when stored)
DRMS1001_240_E.ai
Units: mm (in.)
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-32
(3) Remote exposure hand switch
(4) Remote exposure foot switch
(5) Shielding plate
Units: mm (in.) DRMS1001_241_E.ai
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-33
(6) Under-armrest
Installed
DRMS1001_242_E.ai
Units: mm (in.)
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-34
(7) Compression plate storage holder
* Can be attached to the wall surface in either horizontal or vertical orientation.
57(2
.25)
DRMS1001_243_E.ai
Units: mm (in.)
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-35
2.12 Accessories Sold Separately(1) Biopsy Positioner
(1)-1 Positioner
(1)-2 Positioner Control Cabinet
DRMS1001_246_E.ai
Units: mm (in.)
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-36
(1) -3 Control panel
(1)-4 Biopsy compression plate
DRMS1001_247_E.ai
Units: mm (in.)
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-37
(1)-5 Needle holder C
(1)-6 Needle holder M
DRMS1001_248_E.ai
Units: mm (in.)
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-38
(1)-7 Needle holder Y
(1)-8 Needle holder M8
Units: mm (in.) DRMS1001_249_E ai
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-39
(1)-9 Needle holder E
(1)-10 Biopsy spacer
Units: mm (in.) DRMS1001_250_E ai
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-40
(1)-11 Needle guide cover
Units: mm (in.) DRMS1001_251_E ai
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-41
(2) Lateral approach kit (2)-1 Lateral adapter
(2)-2 Lateral biopsy compression plate
Units: mm (in.) DRMS1001_252_E ai
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-42
(3) Decoration label
594 23.39
1670(65.75)
Decoration label attachment surface
DRMS1001_254_E ai
Units: mm (in.)
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-43
3. Installation Space and Servicing Space
■ Mammographic Stand/Control Unit
● Front/rear direction
R700(27.6)R200(7.87)
DRMS1002_1_01_E.ai
Units: mm (in.)
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-44
● Height direction
26
5( 1
0.43
)
C arm variable range
DRMS1003_1_01_E.aiUnits: mm (in.)
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-45
4. Range of Movement■ Mammographic Stand
REFERENCE
Although the machine is set in the factory so that the C arm stops at 2,350 mm when it is moved up to its uppermost position, the uppermost position of the C arm can be adjusted within a range of 2,050 to 2,350 mm by repositioning the C arm elevation upper-limit switch.The corresponding adjustment range of the uppermost position of the FPD upper surface is 1,200 to 1,500 mm.
DRMS1060_1 ai
Unit:mm
1935
~228
5(fa
ctor
y de
faul
t)
+120°
Arm uppermost adjustable range
350
460~
810
-120°
224
C arm adjustable range
+90° -90°
690
1150
~150
0
Lowermost position of FPD upper surface
Uppermost position of FPD upper surface
974
1503
+180° -180°
Lowermost position at which c arm can swivel 180°
0°
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-46SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-47
5. Anchoring Locations
5.1 Mammographic Stand
DRMS1003_21_01_E.aiUnits: mm
Through-hole
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-48
5.2 Generator (X-ray High Voltage Unit)300(11.81) recommended
(Min. 200(7.87))
DRMS1003_22_E.ai
Units: mm (in.)
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-49
5.3 Control Unit
DRMS1003_21__01_E.ai
Units: mm (in.)
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-50
5.4 Operation Desk and Storage Rack
DRMS1003_23_E.ai
Units: mm (in.)
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-51
6. System Configuration Example CAUTION
Do not use a multi-socket/extension cord for the quality and safety of the product.
NOTES - The FDR-3500DRLH can be installed in an exposure chamber (patient
environment).- Appropriately furnish the exposure chamber and the operation chamber to
visually and acoustically check the patient conditions.
6.1 Installing only the Mammographic Stand in the Exposure Chamber
DRMS1005_1_E.ai
Do not place obstacles on the flow line of a patient.
Do not place peripheral devices in the safety guaranteed area.
Control Pad Cable: approx. 9m
Power Supply, I/F cable: approx. 4m
2503 or more
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-52
6.2 Installing Everything Except the Operation Desk in the Exposure Chamber
Length of the cable for the control pad : Approx. 10m
Power and I/F cable length : Approx. 5m
Power and I/F cable length : Approx. 5m
Power and I/F cable length : Approx. 11m
3090 or more
No peripheral devices are allowed to be installed in the safety guaranteed area.
No obstacle is allowed to be installed on the flow line of a patient.
DRMS1005_2_en.ai
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-53
6.3 Installing the Operation Desk in the Exposure Chamber
WARNINGWhen the operation desk is to be installed in the exposure chamber, put the optional protective board for radiation protection. Place the operation desk at a distance of 1,500 mm or more from the X-ray focus.
DRMS1008_1_E ai
Unit:mmFocus has to be on the same vertical line of center of the protective board
Do not place peripheral devices in the safety guaranteed area
Do not place obstacles on the flow line of a patient.
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-54SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-55
7. Machine SpecificationsSpecifications of the FDR MS-3500 including processing capabilities are described below.They are subject to change without prior notice.
7.1 Exposure Unit Specifications
DRMS1008_2_E ai
Power RequirementsInput voltage Input current (standby) Input current (normal) Input current (peak) Frequency
AC200V (Single phase) 2.4A
16.5A 90A 50 Hz/60 HzAC208V (Single phase) 2.3AAC220V (Single phase) 2.18AAC230V (Single phase) 2.09AAC240V (Single phase) 2.0A
Requirements for Maximum Allowable Impedance between Substation Equipment and Circuit Breaker
Input voltage Maximum allowable impedance Input voltage Maximum allowable
impedanceAC200V (Single phase) 0.143Ω AC230V (Single phase) 0.165ΩAC208V (Single phase) 0.150Ω AC240V (Single phase) 0.172ΩAC220V (Single phase) 0.158Ω
Requirements for Circuit Breaker Connected to Equipment
Number of electrodes 2Rated insulation voltage Ui (V) 600 (V) or moreRated current (A) 50 (A) (AC240V)
* Install a breaker upstream of the device that does not trip when
or 130A flows for 10 seconds, and that trips when 1500A flows continuously for 0.1 seconds.
Insulation Distance between L and NCreepage 3mm (0.11 in.) or longerClearance 1.6mm (0.06 in.) or longer
Operating Characteristic Curves
Operating characteristic curves
Maximum
Maximum (60A~63A)
Maximum (40A~50A)
Current (Percentage of rated current)
Minimum
Operation tim
e
Temperature Characteristic Curve
Percentage of operation
time variation (%
)
Reference am
bient tem
perature
Ambient temperature (°C)
Reference ambient temperature ( 0°C)Reference ambient temperature ( 5°C)
* Recommended product : NF63-CV (50A) (Mitsubishi)
Total shutdown time
Delayed trip Instantaneous trip
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-56
DRMS1008_3_E.ai
Environmental Requirements(1) Operating conditions
Temperature : 20°C to 30°CHumidity : 30% RH to 75% RH (No dew condensation)Air pressure : 750 hPa to 1060 hPa
(2) Non-operating conditionsTemperature : 15°C to 35°CHumidity : 10% RH to 80% RH (No dew condensation)Air pressure : 750 hPa to 1060 hPa
* This equipment cannot be used at an altitude of over 3000 m (9842 ft).
Attenuation Equivalent• Collimator : At least 0.4 mmAl (to the conditions of 49kV radiation)• Exposure table : 0.1 mmAl (to the conditions of 30kV radiation)
: 0.21 mmAl (to the conditions of 30kV radiation)• Biopsy protection cover : 0.18 mmAl (to the conditions of 30kV radiation)
Operating ModeContinuous operation with intermittent loading
Continuous Operation60 shots/hour (28kV, W/Rh, 71mA, 4 consecutive hours)40 shots/hour (28kV, W/Rh, 71mA, 8 consecutive hours)However, maintain the interval shown in the “Minimum exposure interval for energy input to X-ray tube” graph at least.
SID (Focal Spot to Image Receptor Distance)65 cm (25.59 in.)
Characteristics of X-ray ExposureTube voltage range 22kV to 49kV (at 1kV increments)Tube voltage accuracy ±1kVmAs setting range Large focal spot: 2mAs to 600mAs
Small focal spot: 2mAs to 300mAsTomosynthesis: 16mAs to 90mAs(Depending on load conditions such as tube voltage)
mAs setting accuracy ±10%+0.2mAsTube current Large focal spot: 20mA to 200mA
Small focal spot: 20mA to 61mANominal maximum output power 7kW (35kV x 200mA) * Load time: 40ms
Nominal maximum output power (at 30kV, 40ms) 5kW (168mA)
Maximum output power for making exposures at 1-minute intervals
2.8kW (28kV x 100mA) Load time 1s
Total allowable mAs per hour (at 35kV) total 7560mAs
Anode material 10%Re/Tungsten face on molly alloy
X-ray tube focal spot sizeLarge focal spot: 0.3Small focal spot: 0.1* Per IEC 60336, determined with the slit camera method.
0.63mm Be
W/50μmRh, W/700μmAl
SPEC
FDR MS-3500 Service Manual026-201-05(1)E Spec-57
DRMS1008_5_E.ai
Cooling curve of the X-ray system (tube)----- Tube [°C]----- Ambient temperature
Time [min]
Temperature [°C
)]
Minimum exposure interval for energy input to X-ray system (tube)
■ 2D imaging● Tomosynthesis imaging
Min. exposure interval [sec]
Input for exposure [kV x mA x sec]
Condition: 25°C, large focal spot
X-ray Conversion MethodDirect conversion
X-ray Conversion Materiala-Se
236.4 mm (9.31 in.) (vertical) x 296.4 mm (11.7 in.) (horizontal)
Pixel Resolution50 x 50 μm/pix
Number of Output Pixels4728 (vertical) x 5928 (horizontal) = 28,027,584 pixels
Output gradation(1) FDR-3000DRL
16bit(2) FDR-3000AWS
SPEC