Efficient workflow and short learning curve for cataract surgery with FEMTO LDV Z8 1 Background The purpose of this study is to assess the efficiency and effectiveness of the FEMTO LDV Z8 laser during cataract surgery in an outpatient cataract surgery center in Switz erland. The speed and precision with which a laser can achieve a high quality cataract removal is essential for positive clinical results and the overall satisfaction of the patient. Methods This prospective, observational study was performed at the Swiss Eye Research Foundation, Eye Clinic ORASIS, Reinach, Switzerland. Fourteen patients were subjected to unilateral FLACS which consisted of capsulotomy and lens fragmentation performed with the Ziemer LDV Z8 femtosecond laser system. Ease of phacoemulsification (4point scale), completeness of capsulotomy (10point scale), time for preparation of femtosecond laser (min), effective phacoemulsification time (sec), total duration of surgery (min), and safety of the procedure were evaluated. All surgeries were performed under topical anesthesia. Pupil dilation during surgery was achieved by combined use of phenylephrine and tropicamide eye drops. The integrated optical coherence tomography system imaged the ocular structures, and treatment parameters were determined in a customized manner using the laser plat form settings wizard. Laser treatment started with lens fragmentation (an eightpiece pie cut pattern) followed by an anterior capsulotomy (5.0 mm diameter). The mean patient age was 73.5 ± 10.6 years (70.2 ± 8.3 years for males and 74.9 ± 11.3 years for females). Unilateral FLACS was performed in eight right eyes and six left eyes. Eight patients had grade 2 cataract and six patients had grade 3 cataract according to the Lens Opacities Classification System III. Results Laser preparation time before each operation was 3.6 ± 0.7 min. Effective phacoemulsification time was 2.5 ± 3.1 sec. The total duration of the FLACS operation was 16.3 ± 4.5 min. The total duration of the procedure for the dif ferent phases of the learning curve was 21.9 ± 1.8 min for the first surgery session (three patients), 16.0 ± 2.7 min for the second session (six patients) and 12.5 ± 1.1 min for the third session (five patients). Ease of fragmentation was 3.9/4 and completeness of capsulotomy was 9.9/10, as evaluated by the surgeon. No major complications, such as anterior capsule tears, posterior capsule tears or dropped nuclei occurred during the study. Moreover, none of the patients developed subconjunctival hemorrhage postoperatively. Conclusions In our clinic, using the LDV Z8, the learning curve to achieve high quality FLACS with no complications was less than 10 patients with an average total surgery time of around 12 min. The most important features of FLACS include the creation of precise, perfectly centered anterior capsuloto mies, effective lens fragmentation and minimal complica tions, all key pieces to achieving a good surgical outcome and patient satisfaction. Due to the high frequency femto second LDV Z8 laser system, we were able to visualize the anterior corneal surface and posterior lens capsule regardless of cataract grade. In addition, the fluidfilled interface does not lead to formation of marked corneal folds which allow an uninterrupted laser treatment. These features enabled precise and complete removal of the capsule button. References: 1 Pajic B, Vastardis I, Gatzioufas Z, and PajicEggspuehler B. First experience with the new highfrequency femtosecond laser system (LDV Z8) for cataract surgery. Clinical Ophthalmology. 2014; 8: 2485–2489 Clinical study FEMTO LDV Z8 Cataract 5 0 10 15 20 25 n=3 n=6 n=5 Total duration of surgery (min) Operative sessions 1 2 3 * * * Fig 1. Learning curve of surgeries with the Z8. ANOVA reveals significance between each group (p<0.001) Ziemer Ophthalmic Systems AG, CH-2562 Port, Switzerland | www.femtoldv.com | [email protected] The FEMTO LDV Z8 is CE marked and FDA cleared. For some countries, availability may be restricted due to regulatory requirements; please contact Ziemer for details.

Upload: others

Post on 17-Oct-2020

7 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: FEMTO LDV Z8 Cataract - Ziemer USA · 2016. 10. 10. · folds which allow an uninterrupted laser treatment. These features enabled precise and complete removal of the capsule button

Efficient workflow and short learning curve for cataract surgery with FEMTO LDV Z81

BackgroundThe purpose of this study is to assess the efficiency and effectiveness of the FEMTO LDV Z8 laser during cataract surgery in an outpatient cataract surgery center in Switz­erland. The speed and precision with which a laser can achieve a high quality cataract removal is essential for positive clinical results and the overall satisfaction of the patient.

MethodsThis prospective, observational study was performed at the Swiss Eye Research Foundation, Eye Clinic ORASIS, Reinach, Switzerland. Fourteen patients were subjected to unilateral FLACS which consisted of capsulotomy and lens fragmentation performed with the Ziemer LDV Z8 femtosecond laser system. Ease of phacoemulsification (4­point scale), completeness of capsulotomy (10­point scale), time for preparation of femtosecond laser (min), effective phacoemulsification time (sec), total duration of surgery (min), and safety of the procedure were evaluated. All surgeries were performed under topical anesthesia. Pupil dilation during surgery was achieved by combined use of phenylephrine and tropicamide eye drops. The integrated optical coherence tomography system imaged the ocular structures, and treatment parameters were determined in a customized manner using the laser plat­form settings wizard. Laser treatment started with lens fragmentation (an eight­piece pie cut pattern) followed by an anterior capsulotomy (5.0 mm diameter). The mean patient age was 73.5 ± 10.6 years (70.2 ± 8.3 years for males and 74.9 ± 11.3 years for females). Unilateral FLACS was performed in eight right eyes and six left eyes. Eight patients had grade 2 cataract and six patients had grade 3 cataract according to the Lens Opacities Classification System III.

ResultsLaser preparation time before each operation was 3.6 ± 0.7 min. Effective phacoemulsification time was 2.5 ± 3.1 sec. The total duration of the FLACS operation was 16.3 ± 4.5 min. The total duration of the procedure for the dif­ferent phases of the learning curve was 21.9 ± 1.8 min for the first surgery session (three patients), 16.0 ± 2.7 min for the second session (six patients) and 12.5 ± 1.1 min for the third session (five patients). Ease of fragmentation

was 3.9/4 and completeness of capsulotomy was 9.9/10, as evaluated by the surgeon. No major complications, such as anterior capsule tears, posterior capsule tears or dropped nuclei occurred during the study. Moreover, none of the patients developed subconjunctival hemorrhage post­operatively.

ConclusionsIn our clinic, using the LDV Z8, the learning curve to achieve high quality FLACS with no complications was less than 10 patients with an average total surgery time of around 12 min. The most important features of FLACS include the creation of precise, perfectly centered anterior capsuloto­mies, effective lens fragmentation and minimal complica­tions, all key pieces to achieving a good surgical outcome and patient satisfaction. Due to the high frequency femto­second LDV Z8 laser system, we were able to visualize the anterior corneal surface and posterior lens capsule regardless of cataract grade. In addition, the fluid­filled interface does not lead to formation of marked corneal folds which allow an uninterrupted laser treatment. These features enabled precise and complete removal of the capsule button.

References:

1 Pajic B, Vastardis I, Gatzioufas Z, and Pajic­Eggspuehler B. First experience with the new high­frequency femtosecond laser system (LDV Z8) for cataract surgery. Clinical Ophthalmology. 2014; 8: 2485–2489

Clinical study

FEMTO LDV Z8Cataract

5

0

10

15

20

25n=3

n=6

n=5

To

tal d

urat

ion

of

surg

ery

(min

)

Operative sessions

1 2 3

*

*

*

Fig 1. Learning curve of surgeries with the Z8. ANOVA reveals significance between each group (p<0.001)

Ziemer Ophthalmic Systems AG, CH-2562 Port, Switzerland | www.femtoldv.com | [email protected] The

FEM

TO L

DV

Z8

is C

E m

arke

d an

d FD

A c

lear

ed. F

or s

ome

coun

trie

s, a

vaila

bilit

y m

ay b

e re

stric

ted

due

to r

egul

ator

y re

quire

men

ts; p

leas

e co

ntac

t Zi

emer

for

det

ails

.