fer smbt
TRANSCRIPT
6890 FPD
June 1 9, 2003
Agile nt R e stricte d Page 2
6890 FPD - Top Level Assembly
AUX 2 TempHeater
Back DET TempHeater
Vent Tube
Photo-Multiplier Tube (PMT)
6890 FPD
June 1 9, 2003
Agile nt R e stricte d Page 3
6890 FPD Installation
AUX 2 TempHeater Connection
Back DET TempHeater Connection
Vent Tube
Photo-Multiplier Tube (PMT)
Ignitor Connection
High Voltage Connection
Signal Connection
6890 FPD
June 1 9, 2003
Agile nt R e stricte d Page 4
6890 FLAME PHOTOMETRIC DETECTOR
F la m e P h o t o m e t r ic
Product No. : G2333A/G2334A or HP 6890 option 240/241
Single and Dual Signal options.
EPC Only.
Same design as the 5890 FPD . (Most part numbers the same)
Phosphorus and Sulfur filters available.
6890 FPD
June 1 9, 2003
Agile nt R e stricte d Page 5
6890 FLAME PHOTOMETRIC DETECTOR
Installation:
Can be retrofitted on older instruments but may require top cover modification.
PMT mounted left to right Single FPD - Fitted in back position (ease of access to front)Dual FPD - Fitted in back : Uses both detector PCB slots : Cannot use ALS in back position.Uses both DETB and AUX2 heated zones to heat transfer line and heatedblock, but only DET temp entered.Requires AUX actuator solenoid bracket and wiring harness.Requires 2.1x 6890 firmware and 4.02 Chemstation (single) 3.0x 6890 firmware and 5.01 Chemstation (dual)
6890 FPD
June 1 9, 2003
Agile nt R e stricte d Page 6
6890 FLAME PHOTOMETRIC DETECTOR
F lame Photometric
Heater/Sensor connectedto Detector wiring harness
Heater/Sensor connectedto Aux wiring harness
Both heaters are rated 60W, 40VDC
6890 FPD
June 1 9, 2003
Agile nt R e stricte d Page 7
6890 FLAME PHOTOMETRIC DETECTORDual Wavelength option
PMT 2PMT 1
PhosphorusFilter
SulfurFilter
Air/H2
Lines
Jet
6890 FPD
June 1 9, 2003
Agile nt R e stricte d Page 8
6890 FLAME PHOTOMETRIC DETECTORDual Wavelength option
6890 FPD
June 1 9, 2003
Agile nt R e stricte d Page 1 0
6890 FLAME PHOTOMETRIC DETECTOR
Chemiluminescence zone 1st Heat Shield
CH3 - S - CH3 + X2 = Products + hv (light)X2 is the oxidising reagent gas.h is Planck's constant.v is the lights frequency.
6890 FPD
June 1 9, 2003
Agile nt R e stricte d Page 1 1
6890 FLAME PHOTOMETRIC DETECTOR
S 2
HPO
C H C
S F ilter P F ilter
2
0.5
0
0.5
0
0.5
0
0.5
0
300 400 500 600
Relative SpectralResponse
Wavelength (nm)
Element Wavelength (nm)
Colour
Sulphur 393 blue/purplePhosphorus 525 yellow/green
6890 FPD
June 1 9, 2003
Agile nt R e stricte d Page 1 2
6890 FLAME PHOTOMETRIC DETECTORPhoto Multiplier Tube Theory
Electron Multiplication by the DynodesTypical Gain 100,000 - 1,000,000
Photo-emissiveCathode
Anode/Collector
Dynodes
Photon of Light
e -106
Secondary electrons
850 VDC
e -
Photo-electron
DC Voltage
Signal outto electrometer
1 30 90 2500 7500 67500
3 10 270 810 22500 200000
6890 FPD
June 1 9, 2003
Agile nt R e stricte d Page 1 3
6890 FLAME PHOTOMETRIC DETECTOR
Primary - Petroleum & PetrochemicalSulphur in process streams, fuels & oil products
SecondaryEnvironmental
Pesticides - sulphur, phosphorus, tin compounds
Flavors & FragrancesBeverages - sulphur containing compounds
Uses
6890 FPD
June 1 9, 2003
Agile nt R e stricte d Page 1 4
6890 FLAME PHOTOMETRIC DETECTOR
ConditionsColumn: 30m x 250um HP5MS, 2 ml/min constant flowOven program: 60C for 2 minutes then 15C/min to 300C, hold for 2 minFPD: 150 ml/min H2, 100 ml/min air, and 30 ml/min N2 makeup, 250C and data collected at 20 HzInlet: COC run in oven track mode1 ul injected.
20-40 ppb organophosphate pesticides
min11 12 13 14 15 16 17 18
20
30
40
50
60
1-2 ppm organophosphate pesticides
min11 12 13 14 15 16 17 18
150 pA
0
500
1000
1500
2000
6890 FPD
June 1 9, 2003
Agile nt R e stricte d Page 1 5
6890 FLAME PHOTOMETRIC DETECTOR
Varian3400
PE9000XL
Shimadzu GC 17A
Fisons8000
O.I.5380
HP5890
Agilent6890
S MDL pg/sec Dynamic Range S/C Selectivity
<10010^3
10^3/^6
1010^210^4
2010^310^5
510^310^5
110^310^6
2010^310^5
2010^310^5
P MDL pg/sec Dynamic Range S/C Selectivity
<110^5
10^5?
110^310^5
0.910^410^6
0.210^410^6
0.110^410^5
0.910^410^6
0.910^410^6
Upper temp limit 400 450 450 350 420 300 250Unique features Dual
FlamePPC
H2 / airDual PMT Pulsed
FPDDualPMT
EPC
Also PFPD
EPC HP GC one or two
FPDs
DualPMT
Price $11,595 $7245 $7400