Measuring system Measuring range[Max]: 1 kN Overload protection (of [Max]): 150 % Precision (of [Max]): 0,005 Readout [d]: 0,500 N Test frequency: 2000 Hz Display Display digit height: 1 cm Functions Data hold function: yes Function to set limits: yes Obtaining average: yes Peak function: yes Statistical function: yes Environmental conditions Maximum operating temperature: 40 O C Minimum ambient temperature: 10 O C Power supply Charging time: 4 h Input voltage: 100 V - 240 V n/a: CH EURO n/a: Mains adapter external Power supply Operating time (Backlight off): 12 h Service DAkkS Calibration certificate 963-262 (Pressure): DAkkS Calibration certificate 963-162 (Tensile): DAkkS Calibration certificate 963-362 (Tensile/Pressure): ISO calibration certificate: 961-362 ISO Calibration certificate (Pressure): 961-262 ISO Calibration certificate (Tensile): 961-162 ISO Calibration certificate (Ten- 961-362 sile/Pressure): Category Brand: Category: Forcegauges Product Group: Electronic force-measuring device Packaging & shipping Delivery: 24 h Dimensions packaging(WxDxH): 325 x 185 x 165 mm Gross weight: 2,734 kg Net weight: 1,556 kg Shipping method: Parcel service UN number: 3496 Construction Cablelength: 3 m Dimensions of display device (WxDxH): 230 x 66 x 36 mm Overall dimensions mounted (WxDxH): 230 x 66 x 36 mm X-SENSORS GmbH l Steinbrechstr. 9, 71106 Magstadt, Deutschland l Tel: +49 7159/49697-0 l [email protected] l www.sensorshop.com FH-M X-SENSORS GmbH Magstadt, Deutschland l Tel: +49 7159/49697-0 [email protected] l www.sensorshop.com Kompetenz in Messtechnik und Kundenbetreuung

Upload: others

Post on 01-Sep-2019

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: FH-M fileSTANDARD OPTION Model Measuring Tension] range [Max KN Readout [d N Option ISO Calibration Certificate Compression Tension/Compression FH 11K. 0,5 961-162 262 362

Measuring system

Measuring range [Max]: 1 kN

Overload protection (of [Max]): 150 %

Precision (of [Max]): 0,005

Readout [d]: 0,500 N

Test frequency: 2000 Hz

Display

Display digit height: 1 cm

Functions

Data hold function: yes

Function to set limits: yes

Obtaining average: yes

Peak function: yes

Statistical function: yes

Environmental conditions

Maximum operating temperature: 40 OC

Minimum ambient temperature: 10 OC

Power supply

Charging time: 4 h

Input voltage: 100 V - 240 V

n/a: CH EURO

n/a: Mains adapter external

Power supply

Operating time (Backlight off): 12 h

Service

DAkkS Calibration certificate 963-262

(Pressure):

DAkkS Calibration certificate 963-162

(Tensile):

DAkkS Calibration certificate 963-362

(Tensile/Pressure):

ISO calibration certificate: 961-362

ISO Calibration certificate (Pressure): 961-262

ISO Calibration certificate (Tensile): 961-162

ISO Calibration certificate (Ten- 961-362

sile/Pressure):

Category

Brand:

Category: Force gauges

Product Group: Electronic force-measuring

device

Packaging & shipping

Delivery: 24 h

Dimensions packaging (WxDxH): 325 x 185 x 165 mm

Gross weight: 2,734 kg

Net weight: 1,556 kg

Shipping method: Parcel service

UN number: 3496

Construction

Cable length: 3 m

Dimensions of display device

(WxDxH):

230 x 66 x 36 mm

Overall dimensions mounted

(WxDxH):

230 x 66 x 36 mm

X-SENSORS GmbH l Steinbrechstr. 9, 71106 – Magstadt, Deutschland l Tel: +49 7159/49697-0 l [email protected] l www.sensorshop.com

FH-M X-SENSORS GmbH

Magstadt, Deutschland l Tel: +49 7159/49697-0 [email protected] l www.sensorshop.com Kompetenz in Messtechnik und Kundenbetreuung

Page 2: FH-M fileSTANDARD OPTION Model Measuring Tension] range [Max KN Readout [d N Option ISO Calibration Certificate Compression Tension/Compression FH 11K. 0,5 961-162 262 362

STANDARD OPTION

Model

Measuring range

[Max] KN

Readout

[d]

N

Option ISO Calibration Certificate

Tension Compression Tension/Compression

FH 1K. 1 0,5 961-162 961-262 961-362

FH 2K. 2 1 961-162 961-262 961-362

FH 5K. 5 1 961-163 961-263 961-363

FH 10K. 10 5 961-163 – –

FH 20K. 20 10 961-164 – –

FH 50K. 50 10 961-165 – –

FH 100K. 100 50 961-166 – –

Digital force gauge FH-M

Force measuring devices with RS-232 data interface and

with external measuring cells

Features

•Turnable display with backlight

•Digital force gauge with remote sensor

•Cable length: approx. 3 m

•Data interface RS-232, included

• Delivered in a hard carrying case

•Selectable measuring units: N, lb, kg, KN, t •Peak-Hold function to capture peaks (measure-

ment result will be “frozen” for a short time) or

Track function mode for a continuous measure-

ment indication (period of time approx. 10 s)

•Function to set limits, programming of

Max./Min., in pull and push direction, with

output of acoustic and optical signal. Ideal

mode for efficient and accurate testing of

standard parts

•Auto-Power-Off

•Internal memory for up to 10 measurements

•Mini Statistics Kit: calculates the average

result from up to ten stored single results,

min., max., n

Technical data

•High resolution: up to 10,000 points

(total measuring range)

•Measuring frequency: 2000 Hz

•Precision: 0,5 % of [Max]

•Overload protection: 150 % of [Max]

•Dimensions housing WxDxH 66x36x230 mm •Rechargeable battery pack integrated,

standard, operating time up to 12 h without

backlight, charging time approx. 4 h

• Tension loops and compression plates are

included in delivery

FH 1K.– FH 20K.:

•Dimensions sensor WxDxH 51x76,2x19 mm

•Thread: M12

FH 5K.– FH 20K.: •Dimensions sensor WxDxH

76,2x50,8x28,2 mm

•Thread: M12

FH 50K.:

•Dimensions sensor WxDxH 76,3x108x25,5 mm

•Thread: M18

FH 100K.:

•Dimensions sensor WxDxH 125,2x178x51,3 mm

•Thread: M30

Accessories

•Relais module, serves to amplify the output

signal of the dynamometer to control direct

actions, AFH-02

•Force-time data transfer software for

graphical representation on the PC and data

transfer to Microsoft Excel,

AFH FAST •Force-displacement data transfer software

with graphic display of the measurement

process, AFH FD

•Thermal printer, YKB-01N

X-SENSORS GmbH l Steinbrechstr. 9, 71106 – Magstadt, Deutschland l Tel: +49 7159/49697-0 l [email protected] l www.sensorshop.com

FH-M X-SENSORS GmbH

Magstadt, Deutschland l Tel: +49 7159/49697-0 [email protected] l www.sensorshop.com Kompetenz in Messtechnik und Kundenbetreuung

Page 3: FH-M fileSTANDARD OPTION Model Measuring Tension] range [Max KN Readout [d N Option ISO Calibration Certificate Compression Tension/Compression FH 11K. 0,5 961-162 262 362

Adjusting program (CAL):

For quick setting of the balance’s accuracy.

External adjusting weight required.

Calibration block:

standard for adjusting or correcting

the measuring device.

Peak hold function:

capturing a peak value within a

measuring process.

Scan mode:

continuous capture and display

of measurements.

Push and Pull:

the measuring device can capture tension

and compression forces.

Length measurement:

captures the geometric dimensions of

a test object or the movement during

a test process.

Focus function:

increases the measuring accuracy of a

device within a defined measuring range.

Internal memory:

to save measurements

in the device memory.

Data interface RS-232:

bidirectional, for connection

of printer and PC.

Data interface USB:

To connect the balance to a printer,

PC or other peripheral devices.

Data interface Infrared:

To transfer data from the balance to a

printer, PC or other peripheral devices.

Control outputs

(optocoupler, digital I/O):

to connect relays, signal lamps, valves, etc.

Analogue interface:

to connect a suitable peripheral device for

analogue processing of the measurements.

Statistics:

using the saved values, the device

calculates statistical data, such as

average value, standard deviation etc.

PC Software:

to transfer the measurements from

the device to a PC.

Printer:

a printer can be connected to the

device to print out the measurements.

GLP/ISO record keeping:

of measurements with date, time and

serial number. Only with printers.

Measuring units:

Weighing units can be switched to

e.g. non-metric at the touch of a key.

Please refer to website for more details.

Measuring with tolerance range:

Upper and lower limiting can

be programmed individually,

e.g. for sorting and dosing.

ZERO:

Resets the display to “0”.

Battery operation:

Ready for battery operation.

The battery type is specified

for each device.

Rechargeable battery pack:

rechargeable set.

Mains adapter:

230V/50Hz in standard version for EU.

On request GB, AUS or USA version

available.

Power supply:

Integrated, 230V/50Hz in EU.

More standards e.g. GB, AUS

or USA on request.

Motorised drive:

The mechanical movement is carried out

by a motorised drive.

Fast-Move:

the total length of travel can be covered

by a single lever movement.

ISO Calibration:

The time required for ISO calibration

is shown in days in the pictogram.

Package shipment:

The time required for internal

shipping preparations is shown

in days in the pictogram.

Pallet shipment:

The time required for internal

shipping preparations is shown

in days in the pictogram.

Warranty:

The warranty period is shown

in the pictogram.

Pictograms:

X-SENSORS GmbH l Steinbrechstr. 9, 71106 – Magstadt, Deutschland l Tel: +49 7159/49697-0 l [email protected] l www.sensorshop.com

FH-M X-SENSORS GmbH

Magstadt, Deutschland l Tel: +49 7159/49697-0 [email protected] l www.sensorshop.com Kompetenz in Messtechnik und Kundenbetreuung

Page 4: FH-M fileSTANDARD OPTION Model Measuring Tension] range [Max KN Readout [d N Option ISO Calibration Certificate Compression Tension/Compression FH 11K. 0,5 961-162 262 362

User's manual

AFH-02

AFH-02 relay module

Machine control by means of control outputs

Features

•In order to control the machine operations with the use of a dynamometer it is necessary to process the dynamometer control signals in terms of the machine control system.

•As the output signals of FH dynamometer are too low for direct control of the

operations, a relay module is used for their processing which allows for

operation with constant/variable voltage from 20 to 35 V.

•The dynamometers of FH series of company make it possible after setting

the values in the device to control various outputs. It is possible to control 3

outputs. 1. the measurement result is lower than the lower limit value (LoLt option in FH dynamometer menu), 2. the measurement result is higher than the upper limit (HILt option in FH dynamometer menu), 3. the measurement result exceeds Stop value (StoP option in FH dynamometer menu).

• The system of the relay module connections allows setting 4 control options: - value below the lower limit,

- value is OK (between the lower and upper limit), - value above the upper limit, - exceeding of the Stop value.

•The relay module is intended to be fixed on 35/15 mm rails which conform

to DIN 50022 standard as well as for universal assembly in the control cabinet.

•The relay module is equipped in 4 relays, each with one potential-free

change-over.

•The dynamometer is connected to the relay module with the use of SUB-D plug-in socket .

• The scope of delivery : a relay module, a connection cable ca. 1.5m long.

Assignment of terminals

Break contact Common potential- free contact

Make contact

Lower limit value

Ö1 C1 S1

OK Ö2 C2 S2

Upper limit value

Ö3 C3 S3

Stop Ö4 C4 S4

Technical data

Input voltage Relay connection voltage Dimensions

from 20 to 35 V, constant/variable voltage 250 V/10 A 90 × 77 × 41 mm

Compatibility with the equipment software of FS dynamometer from S 5.28 version with full range of functions, below with limited operation (no useful 'Stop' signal).

X-SENSORS GmbH l Steinbrechstr. 9, 71106 – Magstadt, Deutschland l Tel: +49 7159/49697-0 l [email protected] l www.sensorshop.com

FH-M X-SENSORS GmbH

Magstadt, Deutschland l Tel: +49 7159/49697-0 [email protected] l www.sensorshop.com Kompetenz in Messtechnik und Kundenbetreuung

Page 5: FH-M fileSTANDARD OPTION Model Measuring Tension] range [Max KN Readout [d N Option ISO Calibration Certificate Compression Tension/Compression FH 11K. 0,5 961-162 262 362

Thank you for buying a force gauge. We hope you are

pleased with your high quality force gauge with its big

functional range. If you have any queries, wishes or helpful

suggestions, do not hesitate to call our service number.

„Sensor Outside“ means the cell is outside the body.

1. Included in Delivery

- FH, incl. rechargeable battery

- Carrying Case

- Charger

- 5 pcs. M3x 8 screws

Important annotation:

By pressing the RESET key (on the right side of housing,

see illustration on the left), individual settings and

memorised values can be re-set or erased, in example for

a new start of the instrument after an operating error.

Data in mm

Length of cable between display and sensor: 2500 mm

Measurement Terms:

Track

Peak

= continuous measurement

= Capture of the maximum value

Dimensions of Sensors (from FH200k on inquiry)

2. Working Conditions

10°C to 30°C / 15% up to 80% humidity

RS 232 port

Set

Pilot lamp

LCD

Force sensor

Force sensor, M6- thread

4-pin Sensor bushing

Instruction Manual

FH

Sensor

Outside

FH_SO-BA-e-1217 1

X-SENSORS GmbH l Steinbrechstr. 9, 71106 – Magstadt, Deutschland l Tel: +49 7159/49697-0 l [email protected] l www.sensorshop.com

FH-M X-SENSORS GmbH

Magstadt, Deutschland l Tel: +49 7159/49697-0 [email protected] l www.sensorshop.com Kompetenz in Messtechnik und Kundenbetreuung

Page 6: FH-M fileSTANDARD OPTION Model Measuring Tension] range [Max KN Readout [d N Option ISO Calibration Certificate Compression Tension/Compression FH 11K. 0,5 961-162 262 362

Instruction Manual

FH 3. Electrical Power Supply

Either by rechargeable battery or current power supply

Current power supply:

are charged

- Connection by power adapter

- Rechargeable batteries

simultaneously

Rechargeable battery pack for mobile applications:

- Type: Ni 8.4V / 600 mAh

- Charging time: approx 1 hour

4. Technical Data

- Accuracy: ± 0,5 % of Capacity

- Data Sampling Rate: 2.000 Hz

- Weight: 640 g

5. Operation

Display

(1) Measuring Result

(2) Measuring Units

(3) Activation of PRINT Function

(4) Indication of power charging status

PEAK or AUTO-PEAK Mode

(5) Average value of stored peak values

(6) Force direction

(7) Occupancy of storing spaces

(8) AVERAGE- or Saving Mode

Operating keys

ON / OFF:

ON / OFF key

(For ON, press 1 sec.)

UNIT:

- Press short:

- Press for 2 sec.:

Select unit:

N, kg or lb

Display return

ZERO:

Three functions:

- Zeros the measuring result (Tara function)

- Cleans the peak value (in Peak mode)

- Saves a setting (in SET mode

SET:

- 1 x Press: Upper Limit [HidT / Hi Lt]. To

change press: ▲ or ▼ (see section

High/Low Limit function)

- 1 x Press: Lower Limit [LodT]. To change

press ▲ or ▼ (see section High/Low Limit

function )

- 1 x Press: Minimum limit to save peak values

in the instrument’s memory. [LE.SET / Lo

Pe]. Please see section „Min Limit Peak

Save“. To change press: ▲ or

▼ (Only in „Peak-Mode“ active) (see

section min. limit peak save)

- 1 x Press: Sets the Auto-STOP force value

that stops the movement of a test stand.

Adjustment of the force value. To change

press: ▲ or ▼ (only active in „Battery-

Mode“)

- 1 x Press: Sets the AUTO-OFF force value that

stops the movement of a test stand. To

change press: ▲ or ▼ (Only in „Battery-

Mode“ active)

- 1 x Press: Peak-Freeze-Time [PE.2E / A.PE /

HoldT]: Time period in which a peak value

is being shown in the display in sec. To

change press: ▲ or ▼

Hi Lt

Lo dT

Lo Pe

P.OFF

StoP

FH_SO-BA-e-1217 2

X-SENSORS GmbH l Steinbrechstr. 9, 71106 – Magstadt, Deutschland l Tel: +49 7159/49697-0 l [email protected] l www.sensorshop.com

FH-M X-SENSORS GmbH

Magstadt, Deutschland l Tel: +49 7159/49697-0 [email protected] l www.sensorshop.com Kompetenz in Messtechnik und Kundenbetreuung

Page 7: FH-M fileSTANDARD OPTION Model Measuring Tension] range [Max KN Readout [d N Option ISO Calibration Certificate Compression Tension/Compression FH 11K. 0,5 961-162 262 362

Instruction Manual

FH - 1 x Press: Data Transfer [rS232] to PC (PC) or

to the printer (Print) or (in version U 5.1) to

the test stand (stand)

- PC: Pressing SET once: Saves the Settings

- PRINT: Pressing SET twice: sending data to the printer

- STAND: Pressing SET three times: sending a signal to

the Test Stand to stop its movement (in version U 5.1)

Back Light:

Three functions

- Track mode (continuous measurement)

- Peak mode (capture of maximum values)

- Auto-Peak mode, same as Peak-mode, only

without the „Min limit peak save“ function

MEMORY:

Saves the peak values to calculate the average

value (please see section “Memory mode”

DELETE Function

Deletes stored peak values (only in „Memory

mode“ active)

PRINT:

Sends the stored peak values to a PC or Printer

(please see section „Data Output”)

High / Low limit function

LEDs to display OK / NOT OK Tests

OK

Lower than lower limit

OK

Higher than higher limit

This function allows efficient testing of OK / NOT OK

measurements of similar or identical testing objects.

A lower and an upper limit value can be defined. The

instrument compares the individual measuring results with

these limit values and shows the OK or NOT OK result by

green or red light diode and by sound.

To set these limit values, please see the SET Menu in

section „Operating keys“.

Measurement (Track Modus)

Display (1) shows the continuous force in a defined

direction (6)

To zero the display, press:

Peak-Hold Function (Peak Mode)

Please press:

PEAK:

Auto-Peak-Hold-Function(Auto-Peak Mode)

Please press:

Min. limit peak save

This function allows to eliminate unwanted „Pre-Peak

values“ that are lower than the main peak value (Fp). The

“Min limit peak save” value (Fo) takes care, that these

“Pre-Peak values” are not saved.

The „Min limit peak save“ function is only possible in Peak-

Mode.

To set this Min limit value, please see the SET Menu in

section „Operating keys“.

Memory mode and average value (from up to 10 peak

values)

Saving peak values in the instrument

Activating the „AUTO PEAK Function“ by PEAK key

Deactivating the „Average Function“ by MEMORY key

Now, all peak values are stored automatically in the instrument

HoldT

rS232

FH_SO-BA-e-1217 3

X-SENSORS GmbH l Steinbrechstr. 9, 71106 – Magstadt, Deutschland l Tel: +49 7159/49697-0 l [email protected] l www.sensorshop.com

FH-M X-SENSORS GmbH

Magstadt, Deutschland l Tel: +49 7159/49697-0 [email protected] l www.sensorshop.com Kompetenz in Messtechnik und Kundenbetreuung

Page 8: FH-M fileSTANDARD OPTION Model Measuring Tension] range [Max KN Readout [d N Option ISO Calibration Certificate Compression Tension/Compression FH 11K. 0,5 961-162 262 362

Instruction Manual

FH To browse through the stored values, please use the

▲ or ▼ keys. (The values will be shown in the upper display segment)

By pressing the MEMORY key, the average value of the stored peak values can be shown (in the upper

display segment)

To delete every stored value, press the▼-key in the AVERAGE-Mode

6. Configuration of RS 232

SUB-D 9pm

FH_SO-BA-e-1217 4

Pin Signal Illustration

2 TxD Output signal

3 RxD Input signal

5 GND Ground

6 +1.6 to + 2 V Over upper limit

7 +1.6 to + 2 V Lower than lower limit

8 +1.6 to + 2 V OK

6.1 Output Protocol

RS-232 Parameter

Baud rate:

Data-Bit:

Parity:

Stop-Bit:

9600

8

none

1

The measured value is requested by the PC by the ASCII

Sign “9”.

The measured value that comes from the instrument has

this format:

e.g. 0011.70 means -11,70 Newton, if

Newton is the selected unit

||-----|

| | > the other 6 places describe the

measured value as ASCII-Signs

| > the first place describes the

direction of the force (0 = minus = Pressure;

1 = plus = Tension)

or: 1021.15 means +21,15 N (Tension)

7. Warning

Intended use

The instrument you have acquired serves to determine the

measuring value of the material to be measured. It is

intended to be used as a “non-automatic“ instrument, i.e.

the material to be measured is manually and carefully

attached at the instrument. The measuring value can be

read off after a stable measuring value has been obtained.

Inappropriate use

Do not use the instrument for medical measurements. In

the case that small quantities are removed or added to the

material to be measured, incorrect measuring results can

be displayed due to the “stability compensation“ inside of

the instrument. (Example: Slow draining off of liquid from a

container suspended from the instrument). Do not attach a

continuous load. This can damage the measuring unit as

well as the parts, relevant to safety.

Prevent jolts, torsion and oscillation (e.g. by appending

slopingly) of all kinds. Be sure to prevent overloading the

instrument in excess of the stated maximum load (max.),

minus any tare weight that may possibly exist. This could

damage the instrument (risk of breakage).

Important:

•Always make sure that there are no people or materials

below the load that could be injured or damaged!

•The instrument is not suitable for measuring people. Do

not use as baby scales!

•The instrument does not comply with the medical product

law (MPG).

Never operate the instrument in hazardous locations. The

series design is not explosion-proof. Structural alterations

may not be practised to the instrument. This can lead to

incorrect measuring results, faults concerning safety

regulations as well as to destruction of the instrument. The

instrument may only be used in compliance with the

described guidelines. Varying areas of application/

planned use must be approved by in writing.

Guarantee

The guarantee is not valid following

• non-observation of our guidelines in the operating

instructions

• alteration to or opening of the device

• mechanical damage and damage caused by media,

liquids

• natural wear and tear

• inappropriate erection or electric installation

• overloading of the measurement equipment

Monitoring the test substances

The metrology features of the instrument and any possible

available adjusting weight must be checked at regular

X-SENSORS GmbH l Steinbrechstr. 9, 71106 – Magstadt, Deutschland l Tel: +49 7159/49697-0 l [email protected] l www.sensorshop.com

FH-M X-SENSORS GmbH

Magstadt, Deutschland l Tel: +49 7159/49697-0 [email protected] l www.sensorshop.com Kompetenz in Messtechnik und Kundenbetreuung

Page 9: FH-M fileSTANDARD OPTION Model Measuring Tension] range [Max KN Readout [d N Option ISO Calibration Certificate Compression Tension/Compression FH 11K. 0,5 961-162 262 362

Instruction Manual

FH intervals within the scope of quality assurance. For this

purpose, the answerable user must define a suitable

interval as well as the nature and scope of this check.

Information is available on

the home page (www. -sohn.com) with regard to the

monitoring of instrument test substances and the test

weights required for this. Test weights and instruments can

be adjusted quickly and at a reasonable price in ’s

accredited DKD calibration laboratory (return to national

normal).

Fundamental safety information

Do not use the hanging instrument to transport loads.

Prevent jolts, torsion and oscillation (e.g. by appending

slopingly) of all kinds.

Never use the hanging instrument over the maximum

permitted weight (!!Danger of breaking!!).

Always make sure that there are no living beings or

materials below the load that could be injured or damaged.

The hanging electronic instruments from the instrument

are only suitable for hand-held use or use in a test stand.

They are not suitable for hanging from a mechanical hook,

e.g. a crane hook.

Observe the information in the operating instructions

Please read the operating instructions carefully before

setting up and commissioning, even if you already have

gained experience with instruments.

Staff training

The device may only be operated and looked after by

trained members of staff.

Major display deviations (incorrect measuring results) are

possible if electromagnetic fields occur as well as due to

static charging and instable power supply. It is then

necessary to change the location.

Unpacking

Carefully remove the instrument from its packaging,

remove the plastic wrapping and position the instrument in

its intended working location.

8. CE Declaration of conformity

FH_SO-BA-e-1217 5

X-SENSORS GmbH l Steinbrechstr. 9, 71106 – Magstadt, Deutschland l Tel: +49 7159/49697-0 l [email protected] l www.sensorshop.com

FH-M X-SENSORS GmbH

Magstadt, Deutschland l Tel: +49 7159/49697-0 [email protected] l www.sensorshop.com Kompetenz in Messtechnik und Kundenbetreuung

Page 10: FH-M fileSTANDARD OPTION Model Measuring Tension] range [Max KN Readout [d N Option ISO Calibration Certificate Compression Tension/Compression FH 11K. 0,5 961-162 262 362

Instruction Manual

FH

FH Adjustment Procedure

FH_SO-BA-e-1217 6

1 Start the

instrument

Press ON/OFF The green

light is on

2 Enter the

adjustment

procedure

Directly after pressing

ON/OFF, press PEAK

and PRINT together for

more times and very

short intervals until the

red light is on.

The lower

red light is

on

3 Instrument

Type

Press SET

Directly after pressing

PEAK + PRINT

XXX N is

shown

a (Back to the

normal

mode ???)

(If you are in the

operation mode again,

the instrument has to be

shut off. Start again with

1. Eventually pressing

the buttons faster)

4 Select the

instrument

type

Press ▼▲ The

correspondi

ng value to

the instr.

appears in

the display

5 Save

instrument

type

Press SET

6 Enter the

adjustment

sequence

Press MEMORY The upper

red light is

on

7 Choose the

adjustment

weight

Press UNIT and

with▼▲ specify the

calibration weight in

Newton (X kg * 9,81)

The weight

is written in

the display

8 Save the

adjustment

weight

Press SET and UNIT

simultaneously

9 Attach the

weight to the

instrument

(by hanging)

Press ZERO The

instrument

changes to

operation

mode.

Adjustment

procedure is

finished

X-SENSORS GmbH l Steinbrechstr. 9, 71106 – Magstadt, Deutschland l Tel: +49 7159/49697-0 l [email protected] l www.sensorshop.com

FH-M X-SENSORS GmbH

Magstadt, Deutschland l Tel: +49 7159/49697-0 [email protected] l www.sensorshop.com Kompetenz in Messtechnik und Kundenbetreuung

Page 11: FH-M fileSTANDARD OPTION Model Measuring Tension] range [Max KN Readout [d N Option ISO Calibration Certificate Compression Tension/Compression FH 11K. 0,5 961-162 262 362

FH Schnittstellenbeschreibung

17.05.2016 FH-SCHNITTSTELLEN BESCHREIBUNG DP

Senden vom PC Gerät

ASCⅡ Dezimal Hex

2 50 32 Anzeige auf „Null“ stellen

3 51 33 Einheit umstellen auf "KN"

4 52 34 Einheit umstellen auf "TF"

5 53 35 Einheit umstellen auf "KLBF"

6 54 36 Schaltet Gerät in den Track Modus

7 55 37 Schaltet Gerät in den Peak Modus

9 57 39 Gerät sendet aktuellen wert an den PC

Sendeformat

1. bei der Einheit kN sieht das Daten Byte folgender maßen aus:

bit 1 bit 2 bit 3 bit 4 bit 5 bit 6 bit 7

Kraft-

richtung

data 1

data 2

Dezimal

punkt

data 3

data 4

data 5

Bit 1 Kraftrichtung

1: Kraftrichtung “Zug”

0: Kraftrichtung “Druck”

Wenn das Gerät 0,005 kN in “Zugrichtung misst sieht

das Datenbyte folgender maßen aus:

bit 1 bit 2 bit 3 bit 4 bit 5 bit 6 bit 7

1 0 0 . 0 0 5

Wenn das Gerät 0,005 kN in “Druckrichtung misst sieht das Datenbyte folgender maßen

aus:

bit 1 bit 2 bit 3 bit 4 bit 5 bit 6 bit 7

0 0 0 . 0 0 5

- Baudrate: 9600 - Daten-Bit: 8 - Parität: keine - Stop-Bit: 1

Beispiel:

Der Messwert wird durch das ASCII-Zeichen „9“ angefordert.

Der zurückgelieferte Messwert sieht folgendermaßen aus:

bedeutet -11,70 Newton, wenn Newton als Einheit eingestellt ist z.B. 0011.70

||-----|

| | >

| >

die restlichen 6 Stellen beschreiben den Messwert als ASCII-Zeichenkette

erstes Zeichen beschreibt das Vorzeichen (0 = minus = Druck; 1 = plus = Zug)

oder: 1021.15 bedeutet +21,15 N (Zugkraft)

X-SENSORS GmbH l Steinbrechstr. 9, 71106 – Magstadt, Deutschland l Tel: +49 7159/49697-0 l [email protected] l www.sensorshop.com

FH-M X-SENSORS GmbH

Magstadt, Deutschland l Tel: +49 7159/49697-0 [email protected] l www.sensorshop.com Kompetenz in Messtechnik und Kundenbetreuung

Page 12: FH-M fileSTANDARD OPTION Model Measuring Tension] range [Max KN Readout [d N Option ISO Calibration Certificate Compression Tension/Compression FH 11K. 0,5 961-162 262 362

FH Data Connection protocol

17.05.2016 FH-SCHNITTSTELLEN BESCHREIBUNG DP

Senden vom PC Device

ASCⅡ decimal hex

3 50 32 Set the unit to “zero”

3 51 33 make the unit as "KN"

4 52 34 make the unit as "TF"

5 53 35 make the unit as "KLBF"

6 54 36 make the gauge on the tracking mode

7 55 37 make the gauge on the peak mode

9 57 39 send the value data to pc

format of sending

1. when the unit is " KN", the format sending as:

bit 1 bit 2 bit 3 bit 4 bit 5 bit 6 bit 7

bit of

force

direction

data 1

data 2

point

data 3

data 4

data 5

bit of force direction

1: force direction is pull 0:

force direction is press

if the pull force is 0.005KN, and the unit is "KN" the

format of sending is:

bit 1 bit 2 bit 3 bit 4 bit 5 bit 6 bit 7

1 0 0 . 0 0 5

if the pressure is 0.005KN, and the unit is "KN"

the format of sending is:

bit 1 bit 2 bit 3 bit 4 bit 5 bit 6 bit 7

0 0 0 . 0 0 5

- Baudrate: 9600 - Data-Bit: 8 - Parity: none - Stop-Bit: 1

Example: The measured value requested through sending ASCII-sign „9“ from the PC to the force gauge

The requested data looks like:

z.B. 0011.70 means -11.70 N if the force gauge is set to unit Newton

||-----|

| | > the other 6 digits are the measured value in ASCII

| > the first sign in the byte is the pre sign (0 = minus = comprehension;

1 = plus = Tension)

or: 1021.15 means +21,15 N

X-SENSORS GmbH l Steinbrechstr. 9, 71106 – Magstadt, Deutschland l Tel: +49 7159/49697-0 l [email protected] l www.sensorshop.com

FH-M X-SENSORS GmbH

Magstadt, Deutschland l Tel: +49 7159/49697-0 [email protected] l www.sensorshop.com Kompetenz in Messtechnik und Kundenbetreuung

X-SENSORS GmbH l Steinbrechstr. 9, 71106 – Magstadt, Deutschland l Tel: +49 7159/49697-0 l [email protected] l www.sensorshop.com

FH-M X-SENSORS GmbH

Magstadt, Deutschland l Tel: +49 7159/49697-0 [email protected] l www.sensorshop.com Kompetenz in Messtechnik und Kundenbetreuung

Page 13: FH-M fileSTANDARD OPTION Model Measuring Tension] range [Max KN Readout [d N Option ISO Calibration Certificate Compression Tension/Compression FH 11K. 0,5 961-162 262 362

Kommandos zum steuern der Prüfstände

17.05.2016 FH-SCHNITTSTELLEN BESCHREIBUNG DP

Commands to manage the test stands

TVM-N Startet Bewegung “AUF” ASCII 7c

TVM-N Startet Bewegung “AB” ASCII 7d

TVM-N STOP ASCII 7e

TVM-N Start movement UP ASCII 7c

TVM-N Start movement DOWN ASCII 7d

TVM-N STOP ASCII 7e

X-SENSORS GmbH l Steinbrechstr. 9, 71106 – Magstadt, Deutschland l Tel: +49 7159/49697-0 l [email protected] l www.sensorshop.com

FH-M X-SENSORS GmbH

Magstadt, Deutschland l Tel: +49 7159/49697-0 [email protected] l www.sensorshop.com Kompetenz in Messtechnik und Kundenbetreuung

Page 14: FH-M fileSTANDARD OPTION Model Measuring Tension] range [Max KN Readout [d N Option ISO Calibration Certificate Compression Tension/Compression FH 11K. 0,5 961-162 262 362

FH External Sensor Attachement

Dimensions of the standard attachements that come with the external sensors of FH force

gauges.

Model FH 1k – FH 20K FH 50K FH 100k

Screw attachement to the sensor (outside thread)

M 12

M 18

M 30

- 2 x Heavy-duty eye-attachement for pull-tests

Inner element with moveable bearing

Model FH 1k – FH 20K FH 50K FH 100k

Inside bearing (X)

Wall thickness (Y)

Outer diameter (Z)

Length of thread (A)

Overall length (B)

12 mm

10 mm

32 mm

32 mm

70 mm

18 mm

15 mm

48 mm

40 mm

94 mm

30 mm

20 mm

70 mm

68 mm

146 mm

(Z)

(Y)

(X)

(A) (B)

AS, 23.05.2008 FH Sensor Outside Standard Attachm UK.doc Page 1 of 2

X-SENSORS GmbH l Steinbrechstr. 9, 71106 – Magstadt, Deutschland l Tel: +49 7159/49697-0 l [email protected] l www.sensorshop.com

FH-M X-SENSORS GmbH

Magstadt, Deutschland l Tel: +49 7159/49697-0 [email protected] l www.sensorshop.com Kompetenz in Messtechnik und Kundenbetreuung

Page 15: FH-M fileSTANDARD OPTION Model Measuring Tension] range [Max KN Readout [d N Option ISO Calibration Certificate Compression Tension/Compression FH 11K. 0,5 961-162 262 362

FH External Sensor Attachement

- 2 x Pressure plates for push-tests

(D)

(A) (C) (B)

1. 2. 1. 2.

1. Screw-body with moveable plate

The pressure-plate is placed freely on the screw body. The plate can twist to a certain degree against

the screw-body.

AS, 23.05.2008 FH Sensor Outside Standard Attachm UK.doc Page 2 of 2

2. Fixed screw-plate, screwable

The screw-plate and the screw body is one fix piece.

Model FH 1k – FH 20K FH 50K FH 100k

Length of thread (D) for 1. and 2.

Total length incl. plate for 1. (A)

Total length incl. plate for 2. (A)

Diameter of plate for 1. (B)

Diameter of plate for 2. (B)

Thickness of plate for 1. (C)

Thickness of plate for 2. (C)

12 mm

34 mm

29 mm

31 mm

31 mm

18 mm

11 mm

15 mm

47 mm

35 mm

58 mm

48 mm

15 mm

15 mm

18 mm

51 mm

40 mm

58 mm

48 mm

27 mm

27 mm

X-SENSORS GmbH l Steinbrechstr. 9, 71106 – Magstadt, Deutschland l Tel: +49 7159/49697-0 l [email protected] l www.sensorshop.com

FH-M X-SENSORS GmbH

Magstadt, Deutschland l Tel: +49 7159/49697-0 [email protected] l www.sensorshop.com Kompetenz in Messtechnik und Kundenbetreuung

Page 16: FH-M fileSTANDARD OPTION Model Measuring Tension] range [Max KN Readout [d N Option ISO Calibration Certificate Compression Tension/Compression FH 11K. 0,5 961-162 262 362

Dimensionen FH externe Kraftmesszellen

Höchstlast HxLxW T Kabellänge

1 kN 76,2x50,8x19,0 mm M12 x 1,75

Ca. 3m

2 kN 76,2x50,8x19,0 mm M12 x 1,75

5 kN 76,2x50,8x28,2 mm M12 x 1,75

10 kN 76,2x50,8x28,2 mm M12 x 1,75

20 kN 76,2x50,8x28,2 mm M12 x 1,75

50 kN 108x76,2x25,4 mm M18 x 1,5

100 kN 177,8x125,0x50,8 mm M30 x 2,0

FH-ZB-d-1410 1

X-SENSORS GmbH l Steinbrechstr. 9, 71106 – Magstadt, Deutschland l Tel: +49 7159/49697-0 l [email protected] l www.sensorshop.com

FH-M X-SENSORS GmbH

Magstadt, Deutschland l Tel: +49 7159/49697-0 [email protected] l www.sensorshop.com Kompetenz in Messtechnik und Kundenbetreuung

Page 17: FH-M fileSTANDARD OPTION Model Measuring Tension] range [Max KN Readout [d N Option ISO Calibration Certificate Compression Tension/Compression FH 11K. 0,5 961-162 262 362

Dimension FH external load cells

Maximum Load HxLxW T Cable length

1 kN 76,2x50,8x19,0 mm M12 x 1,75

Ca. 3m

2 kN 76,2x50,8x19,0 mm M12 x 1,75

5 kN 76,2x50,8x28,2 mm M12 x 1,75

10 kN 76,2x50,8x28,2 mm M12 x 1,75

20 kN 76,2x50,8x28,2 mm M12 x 1,75

50 kN 108x76,2x25,4 mm M18 x 1,5

100 kN 177,8x125,0x50,8 mm M30 x 2,0

2 FH-ZB-d-1410

X-SENSORS GmbH l Steinbrechstr. 9, 71106 – Magstadt, Deutschland l Tel: +49 7159/49697-0 l [email protected] l www.sensorshop.com

FH-M X-SENSORS GmbH

Magstadt, Deutschland l Tel: +49 7159/49697-0 [email protected] l www.sensorshop.com Kompetenz in Messtechnik und Kundenbetreuung

Page 18: FH-M fileSTANDARD OPTION Model Measuring Tension] range [Max KN Readout [d N Option ISO Calibration Certificate Compression Tension/Compression FH 11K. 0,5 961-162 262 362

FH-ZB-d-1410 1

Dimensionen FH externe Kraftmesszellen FH 10 EXT bis FH 500 EXT

Abmessungen

L x B x H:

76 x 65 x 22 mm

G: Gewinde M6

Kabellänge: 2 m

Externer Sensor mit Zubehör für Drucktests.

100 mm

L

B G

G

H

X-SENSORS GmbH l Steinbrechstr. 9, 71106 – Magstadt, Deutschland l Tel: +49 7159/49697-0 l [email protected] l www.sensorshop.com

FH-M X-SENSORS GmbH

Magstadt, Deutschland l Tel: +49 7159/49697-0 [email protected] l www.sensorshop.com Kompetenz in Messtechnik und Kundenbetreuung

Page 19: FH-M fileSTANDARD OPTION Model Measuring Tension] range [Max KN Readout [d N Option ISO Calibration Certificate Compression Tension/Compression FH 11K. 0,5 961-162 262 362

2 FH-ZB-d-1410

Dimension FH external load cells FH 10 EXT to FH 500 EXT

Dimensions of the sensor

L x B x H:

76 x 65 x 22 mm

G: Thread M6

Cable length: 2 m

External sensor with accessories for push tests.

100 mm

L

B G

G

H