FIGURE 1 INTRODUCTION •Acrylic bone cement is widely used for fixation of total joint replacement prostheses. In the United States alone, expenditures on orthopaedic implants exceeded 1.99 billion dollars in 1999 [1]. The lifetime of bone cement is limited by fracture associated with fatigue failure which can ultimately lead to loosening of the implant, and early revision surgery. •Fractographic studies have shown that cracks are associated with defects in the cement, such as voids and large agglomerates of radiopacifier particles [2, 3]. Prevention of such defects is therefore expected to improve the fracture toughness of bone cement. •Commercial cements are "microcomposites" containing 10 weight % barium sulfate radiopacifier particles of 1 m size in a polymethyl methacrylate (PMMA) matrix. In this study, micrometer size barium sulfate particles were replaced with 100nm size barium sulfate particles in a commercial cement. • USAXS was used to quantify the dispersion of both microcomposite and nanocomposite PMMA cements using specific surface area calculations. In vitro fatigue tests were performed to evaluate the performance of each bone cement under simulated in vivo conditions. MATERIALS AND METHODS • CMW bone cements (Johnson & Johnson/Depuy, Warsaw, IN) were mixed using a standard techniques and molded into dumbbell shaped fatigue specimens and 0.5mm thick specimens for USAXS. • 100nm and 1000nm size barium sulfate particles were employed as radiopacifiers (Sachtleben, Duisberg, Germany) The 100nm barium sulfate ultrafine powder contained 2 weight percent sodium citrate anticoagulant. USAXS was performed on microcomposite, FIGURE 3 SUMMARY OF RESULTS: • Scanning electron microscopy of freeze fracture surfaces of CMW1 microcomposite (Fig. 1; left) and the CMW1 nanocomposite (Figure 1; right) showed that in both cases barium sulfate particles were finely dispersed in the cement with no large agglomerates present. • Specific surface, Os, was calculated from the USAXS curves (Figure 2): Os = *K/Q where K = Porod constant, and Q = Invariant Guinier approximation was invoked for calculation of the invariant in the angular region of 0-q min while Porod’s law was used to estimate the invariant for q max -infinity for calculation of Os (see Table below) •Fatigue tests showed that, under the processing and test conditions used in this study, substitution of the micrometer size barium sulfate with the 100nm size particles resulted in a two-fold increase in the number of cycles to failure in CMW1 (Figure 3). REFERENCES [1] Orthopedic Network News, 1999 , 11 (3). [2] Topoleski et al, J Biomed Mater Res, 1990, 24:135 [3] Demian et al, Trans Soc Biomat, 1995, p368 ACKNOWLEDGEMENT We gratefully acknowledge Dr. P. Jemian and Dr. J. Ilavsky for their assistance with USAXS measurements. The UNI-CAT facility at the Advanced Photon Source (APS) is supported by the FIGURE 2 A USAXS STUDY OF NANOCOMPOSITE POLY(METHYL METHACRYLATE) BONE CEMENTS Mary Turell, Heimo Schnablegger*, David Baker**, Lisa Pruitt**, Anuj Bellare Department of Orthopaedic Surgery, Brigham and Women’s Hospital, Harvard Medical School, Boston, MA *University of Hamburg, Hamburg, Germany **Department of Mechanical Engineering, University of California at Berkeley, Berkeley, CA 0 2,000 4,000 6,000 8,000 10,000 12,000 Num ber ofcycles to failure CM W 1 R adiolucent CM W 1 M icrocomposite CM W 1 Nanocom posite Sam ple K [nm -4 cm -1 ] Porod Exponent Os [cm -1 ] Radiolucent 0.02165 3.08 6.16147x10 5 M icrocomposite 0.2166 3.83 2.39404x10 5 Nanocom posite 2.119 3.65 2.19348x10 6

Upload: julia-white

Post on 17-Dec-2015

212 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: FIGURE 1 INTRODUCTION Acrylic bone cement is widely used for fixation of total joint replacement prostheses. In the United States alone, expenditures on

FIGURE 1

INTRODUCTION

•Acrylic bone cement is widely used for fixation of total joint replacement prostheses. In the United States alone, expenditures on orthopaedic implants exceeded 1.99 billion dollars in 1999 [1]. The lifetime of bone cement is limited by fracture associated with fatigue failure which can ultimately lead to loosening of the implant, and early revision surgery.

•Fractographic studies have shown that cracks are associated with defects in the cement, such as voids and large agglomerates of radiopacifier particles [2, 3]. Prevention of such defects is therefore expected to improve the fracture toughness of bone cement.

•Commercial cements are "microcomposites" containing 10 weight % barium sulfate radiopacifier particles of 1 m size in a polymethyl methacrylate (PMMA) matrix. In this study, micrometer size barium sulfate particles were replaced with 100nm size barium sulfate particles in a commercial cement.

• USAXS was used to quantify the dispersion of both microcomposite and nanocomposite PMMA cements using specific surface area calculations.

• In vitro fatigue tests were performed to evaluate the performance of each bone cement under simulated in vivo conditions.

MATERIALS AND METHODS

• CMW bone cements (Johnson & Johnson/Depuy, Warsaw, IN) were mixed using a standard techniques and molded into dumbbell shaped fatigue specimens and 0.5mm thick specimens for USAXS.

• 100nm and 1000nm size barium sulfate particles were employed as radiopacifiers (Sachtleben, Duisberg, Germany) The 100nm barium sulfate ultrafine powder contained 2 weight percent sodium citrate anticoagulant.

• USAXS was performed on microcomposite, nanocomposite and unfilled cement samples at UNICAT using 10 KeV x-rays.

• Fatigue tests were performed on pre-notched samples under load control with an R ratio (minimum load/maximum load) of 0.03 with a stress amplitude of 15MPa. The load was cycled at a frequency of 2 Hz.

• Freeze fracture cross sections were examined using a JEOL 6320FV field emission scanning electron microscope (SEM). FIGURE 3

SUMMARY OF RESULTS:

• Scanning electron microscopy of freeze fracture surfaces of CMW1 microcomposite (Fig. 1; left) and the CMW1 nanocomposite (Figure 1; right) showed that in both cases barium sulfate particles were finely dispersed in the cement with no large agglomerates present.

• Specific surface, Os, was calculated from the USAXS curves (Figure 2):

Os = *K/Q where K = Porod constant, and Q = Invariant

Guinier approximation was invoked for calculation of the invariant in the angular region of 0-qmin while Porod’s law was used to estimate the invariant for qmax-infinity for calculation of Os (see Table below)

•Fatigue tests showed that, under the processing and test conditions used in this study, substitution of the micrometer size barium sulfate with the 100nm size particles resulted in a two-fold increase in the number of cycles to failure in CMW1 (Figure 3).

REFERENCES

[1] Orthopedic Network News, 1999 , 11 (3).[2] Topoleski et al, J Biomed Mater Res, 1990, 24:135[3] Demian et al, Trans Soc Biomat, 1995, p368

ACKNOWLEDGEMENT

We gratefully acknowledge Dr. P. Jemian and Dr. J. Ilavsky for their assistance with USAXS measurements. The UNI-CAT facility at the Advanced Photon Source (APS) is supported by the University of Illinois at Urbana Champaign, Materials Research Laboratory (U.S. Department of Energy, the State of Illinois-IBHE-HECA, and the National Science Foundation), the Oak Ridge National Laboratory (U.S. DOE), the National Institute of Standards and Technology (U.S. Department of Commerce), and UOP LLC. The APS is supported by the U.S. Department of Energy, Office of Science, Office of Basic Energy Sciences, under Contract No. W-31-109-ENG-38.

FIGURE 2

A USAXS STUDY OF NANOCOMPOSITE POLY(METHYL METHACRYLATE) BONE CEMENTS

Mary Turell, Heimo Schnablegger*, David Baker**, Lisa Pruitt**, Anuj Bellare

Department of Orthopaedic Surgery, Brigham and Women’s Hospital, Harvard Medical School, Boston, MA *University of Hamburg, Hamburg, Germany

**Department of Mechanical Engineering, University of California at Berkeley, Berkeley, CA

0

2,000

4,000

6,000

8,000

10,000

12,000

Nu

mb

er

of

cycl

es

to

fai

lure

CMW 1 Radiolucent

CMW 1 Microcomposite

CMW 1 Nanocomposite

Sample K [nm-4cm-1]

Porod Exponent

Os [cm-1]

Radiolucent 0.02165 3.08 6.16147x105 Microcomposite 0.2166 3.83 2.39404x105 Nanocomposite 2.119 3.65 2.19348x106